Theory and Utility of FT-NIR Spectroscopy
An Introduction to Nicolet Antaris FT-NIR Analyzers
![Page 2: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/2.jpg)
2
The study of the interaction of electromagnetic waves and matter
(Includes UV, visible, NIR, IR, Raman etc.)
MOLECULAR SPECTROSCOPYMOLECULAR SPECTROSCOPY
![Page 3: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/3.jpg)
3
Transitions Guide to the EM Spectrum
![Page 4: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/4.jpg)
4
Radiowave NMRMicrowave Rotation of molecules, ESRIR Fundamental molecular
vibrationsNIR Overtones and combinations of
Mid-IRUV / Visible Electronic transitions, energy of
electrons raised to an excited state
X-Ray Core electronic transitions in atom
SPECTRAL ABSORPTIONSSPECTRAL ABSORPTIONS
![Page 5: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/5.jpg)
5
Vibrational Spectroscopic Regions
• Mid-IR Spectral Region
–4,000 - 400 cm -1 (wavenumber)
–2,500 - 25,000 nm (wavelength of light)
• NIR Spectral Region
–12,820 - 4,000 cm-1 (wavenumber)
–780 - 2,500 nm (wavelength of light)
To convert between nm and wavenumbers, simply take the inverse of the number in nm and multiply by 107. For example, to convert 1000 nm into
wavenumbers, take 1000, invert it and multiply by 107. This gives 10,000 cm-1.
![Page 6: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/6.jpg)
6
What Information Comes From the NIR ?
• Mid-IR spectroscopy makes use of fundamental vibrations• Near-IR spectroscopy uses Overtones and Combination
Bands• Overtones and Combination Bands are types of vibrations
related to the fundamental vibrations seen in the Mid-IR
3000 cm-16000 cm-19000 cm-112000 cm-1
3rd Overtone2nd Overtone
1st Overtone
Fundamental Vibration
![Page 7: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/7.jpg)
7
Nature of Overtones
• Musical instruments like guitars and pianos make use of fundamentals and overtones just like in spectroscopy
• Overtones occur at about 2 and 3 times the frequency of the fundamental vibration. Absorption intensity decreases with increasing overtones
• Band overlap increases with increasing overtones
![Page 8: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/8.jpg)
8
Fundamental and Overtones – A Quantum ViewFundamental and Overtones – A Quantum View
-50015003500550075009500
11500135001550017500
-0.75 -0.25 0.25 0.75Displacement from Equilibrium Bond Length (Angstroms)
Ener
gy (c
m-1
)
2nd Overtone v = 0 → 3
1st Overtone v = 0 → 2
Fundamental v = 0 → 1
ν=3
ν=2
ν=1
ν=0
![Page 9: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/9.jpg)
9
Occur Close to Integer Multiples Of Fundamental Bands.
For example: C-H Overtones Will Occur Near:
First Overtone2960cm-1 (C-H Stretch) * 2 = 5920 cm-1
3378 nm (C-H Stretch) * 1/2 = 1689 nm
Second Overtone2960cm-1 (C-H Stretch) * 3 = 8880 cm-1
3378 nm (C-H Stretch)* 1/3 = 1126 nm
OvertonesOvertones
![Page 10: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/10.jpg)
10
Combination Bands
• There are vibrations in the Near-IR other than overtones of fundamentals. These are called Combination Bands.
• COMBINATION Bands are the sum of several fundamentals from different vibrations. One photon of light excites two vibrations.
• Combination Bands are typically found at lower energies than overtones.
![Page 11: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/11.jpg)
11
COMBINATION Bands Appear Near The Sum Of Two Or Three FUNDAMENTAL Bands
For Example: A C-H Combination Will Occur Near…
2960cm-1 (C-H Stretch) + 1460cm-1 (C-H Bend) = 4420cm-1
= 2262nm
Combination BandsCombination Bands
![Page 12: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/12.jpg)
12
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
1.1
1.2
Abso
rban
ce
4000 5000 6000 7000 8000 9000 10000 11000 Wavenumbers (cm-1)
Example NIR Spectrum
3rd Overtone2nd Overtone
1st Overtone
Combination
![Page 13: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/13.jpg)
13
NIR Characteristics
• R-H groups have the strongest overtones due to hydrogen’s degree of anharmonicity.
• O-H, N-H, C-H, S-H bonds etc. strong NIR absorbers
• H2 and O2 has no change in dipole moment, hence no NIR absorption
• R-H stretch or R-H stretch/bend form most NIR bands
![Page 14: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/14.jpg)
14
Typical Absorptions
Frequency (cm-1)14,300 11,111 9090 7692 6666 5882 5263 4761 4347 4000
700 900 1100 1300 1500 1700 1900 2100 2300 2500
Wavelength (nm)
![Page 15: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/15.jpg)
15
Typical FT-NIR Measurements
• Qualitative and quantitative analysis– Identification– Qualification and quantitation– Determination of change– Process monitoring and control
• Can measure all sample typesSolids Softgels FilmsLiquids Grains SuspensionsTablets Pastes SyrupsPowders
![Page 16: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/16.jpg)
16
Uses for NIR Spectroscopy
• NIR spectroscopy finds widespread use as an excellent QA/QC analysis tool in industrial applications:
– Pharmaceutical
– Chemical
– Petrochemical
– Food and Beverage
– Polymer
– Agriculture
• NIR is a proven and popular technique because it is rapid, accurate, non-destructive and can go through glass and plastic to analyze the material inside
![Page 17: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/17.jpg)
17
The FT-NIR Analysis Advantage
• Large pathlength – Analyze more sample
• Sampling in situ - through glass, packaging materials
– No sample preparation required
• Accurate and precise• Fast - Results in seconds• No Hazardous reagents• Non-destructive• Remote sampling capabilities
![Page 18: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/18.jpg)
18
•ASTM NIR Method E 1944•USP Chapter <1119>•European Pharmacopeia (Ph.Eur.) 2.2.40 1997 (European Directorate for the Quality of Medicines)
Regulatory Acceptance for Near IR
![Page 19: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/19.jpg)
19
Types of Near IR Instrumentation
• Filter – Wavelength filters used to select specific wavelengths for analysis
• Dispersive – Gratings or prisms in combination with slits
• Fourier Transform (Nicolet Antaris) – Interferometer and internal HeNe laser
![Page 20: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/20.jpg)
20
• A complete analyzer for the pharmaceutical, food & beverage, chemical, and polymer industries
Nicolet AntarisTM FT-NIR Analyzer
![Page 21: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/21.jpg)
21
FT-NIR Spectrometer
Detector
IR Source
Fixed mirror
x 0 -x
Beamsplitter
Laser diodeHe-Ne laser
Sample
Moving mirror
![Page 22: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/22.jpg)
22
• Based on the same interferometer technology used for mid-infrared and FT-Raman
• Utilizes all the advantages of FT based instruments– Internal calibration is derived from He-Ne laser
(precision = 0.01 cm-1)– High resolution (4 cm-1)– All wavelengths are measured simultaneously– Higher energy throughput (larger apertures)– Excellent stability and reproducibility– Dynamic alignment
FT Technology for Near Infrared
![Page 23: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/23.jpg)
23
Limitation of Old Dispersive Technology
• Mechanically complicated• Grating filter and chopper
• Poor resolution• 16 cm-1 or worse• Loss of spectral information
• Wavelength inaccuracy• Due to prism or grating
• Stray light effects• Difficult to transfer methods
due to wavelength inaccuracy
![Page 24: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/24.jpg)
24
Traditional NIR Problems Poor Wavelength Accuracy and Precision
• FT-NIR advantage: data repeatability and method reliability due to superior wavelength accuracy and precision
– Internal frequency reference with HeNe laser
7299.02500
Water Vapor Peak Bench S/N AFA00
7299.02100Water Vapor Peak Bench S/N AFA0
1300
1350
1400
1450
1500
1550
1600
1650Si
ngle
Bea
m
7300 Wavenumbers (cm-1)
![Page 25: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/25.jpg)
25
Traditional NIR Problems Low Resolution
• Information can be hidden at lower resolution– Need resolution to distinguish two closely related materials
• FT-NIR advantage – simpler models with fewer calibration spectra possible with greater specificity
32 cm-1
2 cm-1
![Page 26: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/26.jpg)
26
Traditional NIR Problems Poor System Stability and Artifacts
• Stability – consistency in measurement results
• Artifacts – peaks or other anomalies in the spectrum not due to the sample
• Problems eliminated with Nicolet Antarisdesign
![Page 27: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/27.jpg)
Sampling with an FT-NIR Analyzer
![Page 28: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/28.jpg)
28
Measurement Techniques
• Transmission– Liquid analysis
• Quantitative• Qualitative
– Films– Tablets– Semi-solids
• Diffuse reflectance– Powders– Solids – Tablets– Fibers – Pastes– Suspensions with high
solids contents– Opaque semi-solids
![Page 29: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/29.jpg)
29
Nicolet Antaris FT-NIR Analyzer - Overview
Fiber Optic
Integrating Sphere
Transmission
Method Development Sampling System (MDS) shown above
Dedicated configurations also available for specific analyses
![Page 30: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/30.jpg)
30
• MDS System – All techniques on one
system• Other Dedicated Systems
– RTS System– Liquid Analyzer– Solid Sampling System– Tablet Analyzer– Fiber Optic System– Fiber Optic Multiplexer
Tablet Analyzer
Fiber Optic
Integrating Sphere
Transmission
Nicolet Antaris FT-NIR Analyzers
![Page 31: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/31.jpg)
31
Nicolet Antaris Optical Bench
• Rugged industrial platform
• Dynamic Alignment
• User-replaceable and pre-aligned consumables
• Modular design flexibility with stability of fixed beampaths
• Integrated validation (Built to meet or exceed USP monograph)
![Page 32: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/32.jpg)
32
Near IR Quartz Halogen Source
• External replacement with only 3 screws
• No wires
• Source change in seconds
• Pinned-in-place aluminum housing for lock-and-key fit
• High throughput
![Page 33: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/33.jpg)
33
Internal Referencing For All Collection Types
• Protected internal reference material simplifies background collection
– Spectralon
– Diffuse Gold Flag
• Eliminates errors due damaged reference materials
• Collect backgrounds with sample already loaded
• Easy to use and automated
![Page 34: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/34.jpg)
34
Reflection Sampling – Integrating Sphere
• Ideal collection environment for diffusely reflected light
• High efficiency collection
• Internal gold background material
• Sample simply needs to be placed on the instrument for data collection
![Page 35: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/35.jpg)
35
Liquid Sampling - Transmission Compartment
• Collect data on liquid, film or paper
• Accommodates vials, tubes, cuvettes and card holders
• Temperature control
• Automatic internal background position in transmission cell
• Reproducible positioning
![Page 36: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/36.jpg)
36
Fiber Optic Sampling – SabIR and 3rd Party Probes
• Transflectance or reflection sampling
• Fiber optic port uses standard SMA connectors
• Remote sampling capability for At-line, near-line, on-line
• Ideal for incoming raw material identification and in-process analysis
![Page 37: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/37.jpg)
37
Solid Transmission Sampling – Tablet Analyzer
• Thick, opaque tablets give quantitative results for
– Content uniformity– API or excipient concentrations– Physical properties like hardness
• Integrating sphere reflection data gives tablet coating information
• Collect reflection and transmission data without removing the tablet
• Dedicated softgel detector
Transmission
Diffuse Reflection
![Page 38: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/38.jpg)
38
Antaris II
• Method Development System - All sampling types for NIR– Transmission, tablets, softgels, diffuse reflection, fiber optics– Automatic sampling of solid dosage forms - 40 tablets at once– Simultaneous reflection and transmission measurement of tablets
• Configurable base system with selection of sampling modules
![Page 39: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/39.jpg)
39
Remote Sampling – Fiber Optics – Antaris MX
Antaris MX
• Simultaneous multi-channel measurement
• Simultaneous internal backgrounds• No mechanical multiplexing• Dual trigger probe capability• Programmable analog/digital I/O
options with RESULT Interface• Spectral corrections adjust for fiber
optic differences
![Page 40: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/40.jpg)
40
Analyzer performance verification
• Traditionally done with probe pulled off process
• Internal reference channel can qualify operation while probe is inline
• Traceable reference standards for spectroscopic performance checks according to USP <1119>
![Page 41: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/41.jpg)
41
Fiber optic probes
• High quality probes with proprietary window mounts in custom configurations
• Specialty probes– Fermentation – dual mode
reflection and transmission– Retractable dryer probe with self
cleaning tip and wash chamber
![Page 42: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/42.jpg)
42
High Throughput Sampling – Autosampler RS and Multipro Autosampler
• Multiple samples with no operator interaction
• Eliminates sample-to-sample operator error
• Useful for analyzing tablets, softgels, solids or powders in vials
• Transfer already existing methods for seamless scale-up
![Page 43: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/43.jpg)
43
Heterogeneous Sampling – Sample Spinners
• Obtain spectra of heterogeneous samples like
Polymer pellets Grains
Coarse powders Silage
Granular solids Corn
Crystalline materials
• Eliminate the need to collect multiple spectra on the same sample
![Page 44: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/44.jpg)
44
Antaris EX: A complete process analyzer
• Hazardous environments • Integrated computing• Integrated process
communications• Same sampling technology
and advantages as MX– Simultaneous measurements
with internal referencing and qualification
• Not a box that you have to put inside of another box!
• Complete new lineup of probes
![Page 45: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/45.jpg)
45
Hazardous environments
• Three EX models:– EX-Z1 for Class I Division 1/Zone 1
Environments– EX-Z2 for Class I Division 2/Zone 2
Environments– EX-IP65 for
• Enclosures are:– Hose washable, Stainless steel,
corrosion proof– Individually certified
• Vortex cooling provides operating temperature range of -20 to 40 C
• Purge systems (different for each environment) linked to power of system
![Page 46: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/46.jpg)
46
On Blender Analysis
![Page 47: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/47.jpg)
47
Blending in Development – Multiple sizes
• Instrumentation must work with a wide range of blender sizes• Issues of scale with equipment size and weight
![Page 48: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/48.jpg)
48
Monitoring the Blend with NIR
• By scanning the sample (pharmaceutical powder blend) through a window in the modified lid, the progression of mixing and concentration of components can be measured
• Data is triggered when bin is upside down (powder in contact with sampling window)
• Powered by rechargeable battery pack• Diffuse reflection sampling• Wireless communications to remote PC with RESULT
software for control/DAQ• Standard deviation measurement can be used to determine
when mix is fully achieved
![Page 49: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/49.jpg)
49
The Spectroscopy Technology
Optical BenchOptical Bench
Optical Module Sub-AssemblyOptical Module Sub-Assembly
Spectrometer AssemblySpectrometer Assembly•Source – Semiconductor based NIR tunable laser
Not a light bulb/globar – does not go “out” or fadeRated for a 25 year lifetime (Telcordia Spec.)
•Wavelength selection - High resolution (2 nm / 8cm-1)•Fabry-Perot tunable filter
Tuning range of hundreds of nmScans full range in tens of millisecondsInsensitive to vibration
•Internal Wavelength and Amplitude ReferenceEnsures optimum wavelength calibration over time
![Page 50: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/50.jpg)
50
Instrumentation and Sampling Interface• Detector: Single element InGaAs photodiode
– Optimized detector element sized to collect highest solid angle formaximum efficiency
• Baseline s/n of 0.0002• Wavelength ranges available
– 1350-1550 nm, 1550-1800 nm and1800 – 1970 nm
• Fast Scanning (~80 ms / scan)• Sample sensing area of 40 mm
(coincides with ~600 mg sample size)• Smaller spot size available (~10 mm) for
“intra-dosage” uniformity investigation• Wireless communication (~15 meters)• Battery operation (2 – 3 hours)
Fiber Optic InputFrom Spectrometer
Collimating LensAssembly
SapphireWindowAssembly
Detector LensAssembly
Extended Range HighPerformance SingleElement InGaAs Detector
Diffuse Gold Reference
Validation Sample Position
Fiber Optic InputFrom Spectrometer
Collimating LensAssembly
SapphireWindowAssembly
Detector LensAssembly
Extended Range HighPerformance SingleElement InGaAs Detector
Fiber Optic InputFrom Spectrometer
Collimating LensAssembly
SapphireWindowAssembly
Detector LensAssembly
Extended Range HighPerformance SingleElement InGaAs Detector
Diffuse Gold Reference
Validation Sample Position
Diffusely reflected energy
![Page 51: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/51.jpg)
51
Summary
• On Blender Real-time analysis provides a rapid and calibration free method of getting blend information and performing experiments
• With the correct type of optical interface the instrumentation can be quickly moved between blender sizes and types for scale-up development
• Instrumentation is process ‘friendly’– Small– Lightweight– Easy to set up– Long battery life– Robust
![Page 52: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/52.jpg)
RESULT Software Suite for the Nicolet Antaris Analyzer
![Page 53: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/53.jpg)
53
RESULT Software Suite for Near IR
• RESULT is the software package made for collecting Near IR data with the Nicolet Antaris Analyzer
• Provides– Real time data acquisition– Sample prediction using developed models– Chemometrics tools for method development– Full suite of compliance tools
![Page 54: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/54.jpg)
54
RESULT Provides Tools for all Users
• RESULT Operation allows users with no science training to run routine analyses
– Minimal training time and one-click use
• RESULT Integration provides the tools to develop workflows with integrated SOPs for workroom deployment
– Design simple workflows like Raw Material ID
– Create sophisticated experiments with logical gates, delays and repeats
![Page 55: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/55.jpg)
55
RESULT Provides Tools for all Users
• TQ Analyst is a powerful chemometrics package– Easy to use
– Fully integrated with spectral collection software
– No secondary programs needed
• Administrator Mode– Creates new users
– Maintains system integrity
– Sets workflow access privileges for all users
![Page 56: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/56.jpg)
56
ValPro Instrument Qualification Platform
• Design qualification (DQ) documentation provided
• Detailed Installation Qualification (IQ) procedure
• USP-based Operational Qualification (OQ) tests provided– Traceable & serialized internal validation
wheel
• Tools for Performance Qualification (PQ)
![Page 57: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/57.jpg)
57
21 CFR Part 11 Compliance with RESULT
• Full Suite of Part 11 Tools– Full operation auditing– SOP enforcement and
automation– Digital signatures– Security, access control, and
privilege and workflow assignment
![Page 58: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/58.jpg)
58
Software Validation
• TQ Analyst Algorithms– Fully Documented– Externally reproduced for verification in Excel– Demonstrates internal validation
![Page 59: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/59.jpg)
59
Process Analytical Technology (PAT)
• PAT is a system for designing, analyzing, and controlling manufacturing through timely measurements (i.e., during processing) of critical quality and performance attributes of raw and in-process materials and processes with the goal of ensuring final product quality
• Seamless migration of FT-NIR technology from the laboratory to the manufacturing floor
![Page 60: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/60.jpg)
60
Time for a Paradigm Shift - PAT
• In-process testing that currently exists in pharmaceutical manufacturing consists of
– Physical measurements
– Results from samples sent to the lab while processes were halted or continued “blindly”
• Current testing is not real-time, nor “whole batch”
• Non-optimal conditions can still occur (over-blending for example)
• Testing and waiting for results is costly
![Page 61: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/61.jpg)
61
Typical FT-NIR PAT Applications
• Blending and mixing• Reaction monitoring• Fermentation• Drying• Post tableting tests
![Page 62: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/62.jpg)
62
• Traditionally, optimal mixing time was determined during process development and scale-up
• Avoids problems of incomplete or over-blending (de-blending)
• Improves efficiency of production processes and equipment usage
• Assures product uniformity• Goals
– Establish adequacy of mixing– Eliminate thieving/process
interference– Eliminate error and recipe
approach– Accurate endpoint determination
NIR in the Manufacturing Process
![Page 63: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/63.jpg)
63
Integration with manufacturing systems
• RESULT OPC (OLE for Process Control) Server facilitates powerful communication with manufacturing systems
• Send and receive commands automatically
• Trigger runs, endpoints, valves, etc.
![Page 64: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/64.jpg)
64
Communication is Essential to Process Control
• Communication system must be able to bridge– Disparate technologies from various manufacturers
– Compatibility with both high and low level interfaces
– Opposing computer platforms
– Geographic separation from target
– Interface with business software to expand the process control loop
• Data communication architecture must be inclusive, not exclusive
• Existing and modern facilities
![Page 65: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/65.jpg)
65
OPC in Process: Dryer Monitor
Analyzer Data Processing
OPC Server Output
SCADA or DCS OPC Input
Dryer Data Collection
![Page 66: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/66.jpg)
66
Typical FT-NIR Pharmaceutical Applications
• Raw Material ID• Moisture Content• Content Uniformity• Tablet API• Coating Thickness• Process Analytical Technology (PAT)
![Page 67: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/67.jpg)
67
Pharmaceutical Application: Raw material ID
• Apply NIR to inspect incoming raw materials quickly at the loading dock
• Equip the receiving area with an NIR to be used by technicians who are not trained as scientists
• Confirm the identity of each container of material that is received, without lab wet chemistry techniques that are time-consuming and expensive
![Page 68: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/68.jpg)
68
Library Raw Material Samples
• One lot each of the following substances was provided for the construction of the raw material library
d-Glucosed-Fructose
Sucrosed-Mannitold-Sorbitol
α-d-Lactose MonohydrateAcetylsalicylic Acid
AcetaminophenL-Ascorbic Acid
Citric Acid
![Page 69: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/69.jpg)
69
Validation Samples
d-Glucose - Second lot from a different manufacturer
d-Fructose - Second lot from a different manufacturer
Acetaminophen - Second lot from a different manufacturer
Salicylic Acid - Degradation product of Acetylsalicylic Acid
2-acetamidophenol - Constitutional isomer of acetaminophen
α-d-Lactose anhydrous - Dehydrated form of library material
Black items are positive challenge samples (Should Pass) Red items are negative challenge samples (Should Fail)
![Page 70: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/70.jpg)
70
External Challenge Validation Results
Compound Score Result
Glucose (Lot #2) 99.9 Pass
Fructose (Lot #2) 98.0 Pass
Acetaminophen (Lot #2) 100.0 Pass
Salicylic Acid 55.2 Fail
2-Acetamidophenol 19.3 Fail
Lactose Anhydrous 68.9 Fail
![Page 71: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/71.jpg)
71
FT-NIR Success stories
• Customer had a plant with diverse array of incoming raw materials that needed to be classified accurately.
• Implemented NIR testing for over 200 materials• Based upon technician time savings alone, paid back the
investment in about 7 months• Needed to identify a small number of materials at very
high volume composing the bulk of their test load• Implemented 100 % inspection for ID and qualification• Dropped their testing time on each incoming shipment of
raw material by 68 %
![Page 72: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/72.jpg)
72
Lyophilized Material Analysis
• Lyophilized materials are an excellent application for analysis by NIR
• Packaging cannot be opened without corrupting the product inside
• Can apply to food, pharmaceutical, fine chemical or biotech industries
![Page 73: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/73.jpg)
73
Current Testing for Lyophilized Products
• Current testing methods like Karl Fisher titration, LOD or GC
– cumbersome, time consuming and costly
– destructive
• NIR does not need solvents or carrier gases
• NIR calibrations for lyophilized materials have been shown to be as accurate as the reference method
• Other components can also be measured using the same NIR spectrum (Buffer, API, Cryoprotectant)
![Page 74: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/74.jpg)
74
Case Study For Lyophilized Sample Analysis
• Primary component of interest in any lyophilized materials is moisture (0.5% - 2.0% typically)
• Sample is Thrombin, a topical coagulant used in the medical and dental fields
• Thrombin loses its efficacy if exposed to high heat or wet conditions. Current application will look at moisture and potency in Thrombin
![Page 75: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/75.jpg)
75
Moisture First Overtone Region
![Page 76: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/76.jpg)
76
Close-up of First Moisture Overtone Region
![Page 77: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/77.jpg)
77
Calibration Curve for Moisture
![Page 78: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/78.jpg)
78
Calibration for Potency
![Page 79: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/79.jpg)
Quantification of Active Pharmaceutical Ingredient (API)
in Tablets
![Page 80: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/80.jpg)
80
Show Concentration Differences in Tablets
• Distinguish different clinical tablet formulations by amount of active ingredient
• Finished product is a tablet and manufacturing protocols need independent verification of amount of active ingredient
• Understand the differences between transmission and reflection analysis of tablets
![Page 81: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/81.jpg)
81
Reflectance Spectra for Clinical Tablets
0.250.35
0.45
0.55
0.65
0.75
Log(
1/R
)
5000 6000 7000 8000 9000
-2.2-1.8-1.4-1.0-0.6-0.20.20.61.01.4
Log(
1/R
)
5400 5500 5600 5700 5800 5900 6000
Wavenumbers (cm-1)
Expanded Second
Derivative
![Page 82: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/82.jpg)
82
Tablet Reflection Results
0
10
20
30
0 10 20 30
Label Claim (Rel. Units)
NIR
Cal
cula
ted
(Rel
. Uni
ts)
Calibration
Validation
R = 0.99742RMSEC = 0.606
![Page 83: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/83.jpg)
83
Tablet Transmission Results
0
10
20
30
0 10 20 30
Label Claim (Rel. Units)
NIR
Cal
cula
ted
(Rel
. Uni
ts)
Calibration
Validation
R = 0.99965RMSEC = 0.0561
![Page 84: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/84.jpg)
84
API Quantification Can Also Be Done Using Qualitative Methodology - Discriminant Analysis
PC1
PC3
0124121624
• Principal component scores plot reveals quantitative method may not be necessary
• The required information is obtained with excellent results using qualitative analysis (Discriminant)
![Page 85: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/85.jpg)
85
NIR in Pharmaceuticals - Summary
• NIR has proven useful in the following applications:– Raw Material ID
– Raw Materials - Particle Size
– Analysis of Parenteral Drugs - Lyophilized Materials
– Quantification of API in Tablets
– Blend Homogeneity
– Lactose Polymorph Screening
– SoftGel Analysis
– Enzyme Activity
![Page 86: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/86.jpg)
86
Near IR in the Chemical Industry
• Near IR analysis has proven worth across the chemical industry
–Fine chemicals
–Polymers
–Petrochemicals
–Paints, coatings and adhesives
–Soaps and cleaners
–Synthetic materials
• At-line, near-line and on-line
![Page 87: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/87.jpg)
Polymerization Monitoring:Acid Value and Hydroxide
![Page 88: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/88.jpg)
88
HoVal Analysis in NIR - Purpose
• HoVal is short for Hydroxyl Value, an important metric in polymerization reactions
• Measures both starting materials (Acid and OH) simultaneously from the same spectra
• Condensation polymers (Polyesters) show decrease in acid and alcohol concentration with time
• Use NIR to monitor this decrease will show when polymerization is finished
![Page 89: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/89.jpg)
89
Experimental
• Samples were placed into 6 mm disposable vials• Analysis using Nicolet Antaris Transmission module
![Page 90: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/90.jpg)
90
Spectral Data for HoVal Analysis
0.4
0.8
1.2
1.6
2.0
2.4
2.8
5000 6000 7000 8000 9000 10000
Wavenumbers (cm-1)
![Page 91: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/91.jpg)
91
Hydroxyl Value Correlation
Acid Number Correlation
NIR Replaces Two Wet Chemical Methods
![Page 92: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/92.jpg)
Plasticizer Content in PVC
![Page 93: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/93.jpg)
93
Plasticizer Content in PVC
• Plasticizers are common additives to polymers.• The quantity of plasticizer determines the properties of
the polymer.• The plasticizer contents of polymers differ depending
on the application.• Primary methods often involve tedious and time-
consuming extractions.
![Page 94: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/94.jpg)
94
Samples
• Samples of polyvinyl chloride (PVC) with varying amounts of dioctylphthalate (DOP) as plasticizer
• Three types of samples:
– Opaque PVC plates with 5 - 50% DOP
– Transparent PVC plates with 5 - 50% DOP
– PVC Films with 9 - 40% DOP
![Page 95: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/95.jpg)
95
FT-NIR Measurements
• As is measurements (no sample preparation or modification).
• Data collected from 4,000 cm-1 to 10,000 cm-1
• 4 cm-1 resolution.• 90 scans• One (1) minute collection time.• All measurements done on
Integrating Sphere
![Page 96: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/96.jpg)
96
Non-Transparent Plates
![Page 97: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/97.jpg)
97
Calibration Parameters
MSC4125-39464296-4263Norris 5,52nd
Derivative3Films
MSC4772-45205935-5322Norris 5,52nd
Derivative4Transparent
MSC8982-3920Norris 9,02nd
Derivative4Opaque
PathlengthRegion (cm-1)SmoothingData
FormatFactors
![Page 98: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/98.jpg)
98
Non-Transparent Plates
![Page 99: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/99.jpg)
99
Calibration Results
0.4070.9990.1740.999Films
1.450.9960.5910.999Transparent Plates
1.490.9950.3750.999OpaqueRMSECVCorr. Coeff.RMSECCorr. Coeff.
Cross ValidationCalibration
![Page 100: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/100.jpg)
100
Conclusions on NIR of Plasticizer Content
• Additives such as plasticizers can be measured in various forms of plastic materials
• These materials can be finished or unfinished– Measurements can be in-process or QC
measurements
• The analyses can be done easily, rapidly and non-destructively
• Current methodology for extraction of matrix elements in polymers is extremely time consuming
![Page 101: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/101.jpg)
101
Near IR and Food
• Near IR analysis of food and food products has found extensive use in all segments of the industry
– Processed foods
– Beverages
– Grain
– Fruit
– Meat
• Common components like protein, moisture, ash, starch and fat are measured simultaneously without destroying the sample
![Page 102: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/102.jpg)
102
FT-NIR As A Process Analytic in Food
• Develop methods for any type of food product
• High resolution advantage of FT technology keeps number of standards low
• Many different sampling accessories for food products– Sample spinner
– SabIR Fiber Optic Probe
– Cheese and Viscous Liquid Samplers
– Softgel detector (Gum)
• OPC Output for analyzer incorporation into Process Control Systems
![Page 103: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/103.jpg)
103
NIR Analysis in the Food Industry
• Moisture and Protein in Barley
• Viscous Liquid Analysis
• Oil in Rice Powder
• Amylose Levels in Corn
• Multi-component Analysis of Wine
• Fat and Dry Matter in Cheese
• Discrimination of Jellybeans
• Choline Chloride Levels in Silage
![Page 104: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/104.jpg)
104
• FT-NIR Nicolet Antaris Spectrometer Platform
• RESULT Analyzer Software Platform
• TQ Analyst Chemometric Software Platform
• ValPro Validation/Qualification Platform
• Support “Pyramid of Support” products and servicesSales
ApplicationsService
Customer
Nicolet Antaris Product Line – Complete Solution
![Page 105: Theory and Utility of FT-NIR Spectroscopy An Introduction to Nicolet](https://reader031.vdocuments.mx/reader031/viewer/2022021804/620d64799d59d8769f434a56/html5/thumbnails/105.jpg)
105
The FT-NIR Advantage
The Bottom Line: Save Time and Money.
Use FT-NIR to its maximum capability.