T H E E C H O D O E S N ’ T L I E

D E I R D R E M U R P H Y A L F R E D I C U

Page 2: The Echo Doesn't Lie by Murphy

E C H O D O E S N ’ T L I E B U T I T C A N B E N D T H E T R U T H A L I T T L E

Page 3: The Echo Doesn't Lie by Murphy

E C H O I S B A S E D O N A N U M B E R O F A S S U M P T I O N S

Page 4: The Echo Doesn't Lie by Murphy

E C H O I S B A S E D O N A N U M B E R O F A S S U M P T I O N S

Sound travels at same speed though all tissues (It doesn’t)

Echoes are generated from centre of the beam

!

Page 5: The Echo Doesn't Lie by Murphy

Z O E T R O P E T Y T H E TA S M A N I A N T I G E R

Page 6: The Echo Doesn't Lie by Murphy

Important in utilising the technology that we understand the limitations so that we can use it to its full potential

Page 7: The Echo Doesn't Lie by Murphy

C O M M O N H A E M O D Y N A M I C I N F O R M A T I O N

and the pitfalls in their measurement..

Page 8: The Echo Doesn't Lie by Murphy

C A R D I A C O U T P U T

fundamental measure in critically ill patient

Page 9: The Echo Doesn't Lie by Murphy

LVOT method

Assumes the LVOT is a cylinder

We can measure the VTI of blood flow in the LVOT by placing a pulse wave doppler gate in LVOT

This gives us the LVOT VTI

Page 10: The Echo Doesn't Lie by Murphy

M A T H S …

Volume =CSA X height (distance)

!

Stroke volume= π r2 X VTI

!

Cardiac output= SV x heart rate

Page 11: The Echo Doesn't Lie by Murphy

any measurement error will be

squared !

CSA= π r2

Page 12: The Echo Doesn't Lie by Murphy

Major pitfalls

!

Flow acceleration at valve- measure 1 cm back

Ensure line up with cursor- inaccurate if >10°

Page 13: The Echo Doesn't Lie by Murphy

U S E V I E W W I T H T H E B E S T D O P P L E R L I N E U P - D I F F E R E N T F O R D I F F E R E N T PA T I E N T S

5 chamber view

3 chamber view

Page 14: The Echo Doesn't Lie by Murphy

H O W D O Y O U K N O W Y O U ’ V E G O T A G O O D D O P P L E R T R A C E

PW Doppler spectral outline

Trace not “filled in”- in moving front of blood flow

Not jagged feathery ends

Page 15: The Echo Doesn't Lie by Murphy

Measurement of VTI or stroke distance

Average 3 in sinus rhythm Average 5 if arrhythmia

Page 16: The Echo Doesn't Lie by Murphy
Page 17: The Echo Doesn't Lie by Murphy

T H E LV O T I S E L L I P S O I D

! !

! !

Page 18: The Echo Doesn't Lie by Murphy

A G R E E M E N T W I T H T H E R M O D I L U T I O N

Reasonable

Operator needs to be aware of the sources of error

!

Page 19: The Echo Doesn't Lie by Murphy

Patient commenced on adrenaline after study

LVOT measured at 1.0 cm, VTI N

LVOT in adult 1.8-2.6

Page 20: The Echo Doesn't Lie by Murphy

R V S P ( PA P R E S S U R E E S T I M A T I O N )

Based on Bernoulli equation

Pressure gradient =4V2

RVSP= PG + CVP

PA pressure = RVSP

Page 21: The Echo Doesn't Lie by Murphy

W H A T A R E T H E R E Q U I R E M E N T S ?

Need to evaluate in a number of views to get the best line up with the colour jet

Page 22: The Echo Doesn't Lie by Murphy

R S I D E D C A R D I A C O U T P U T

Useful to quantify shunts,

MCS RVOT inflow

velocites lower and vary with

respiration

Page 23: The Echo Doesn't Lie by Murphy

PA Diastolic (PR jet) 8 mm (+ RAP) mmHg

PA pressure estimation from Pulmonary Acceleration time

Page 24: The Echo Doesn't Lie by Murphy

W H A T A B O U T V O L U M E S TA T E A N D E C H O ?

Page 25: The Echo Doesn't Lie by Murphy

E A R LY S T U D I E S L O O K E D A T LV E D A

Problems as doesn’t take into

account compliance

afterload states

Page 26: The Echo Doesn't Lie by Murphy

S TA T I C E S T I M A T E S O F R A T R I A L P R E S S U R E A N D E C H O

IVC dimension (spontaneous breathing) and collapsibility

!

IVC = <2.1 and varies > 50% Estimated RAP =3

In between =8

IVC = > 2.1 and doesn’t vary Estimated RAP =15

Page 27: The Echo Doesn't Lie by Murphy

L H E A R T P R E S S U R E SEcho assessment of left atrial pressure

!

Mitral valve E/e’

E/A > 2 PAOP >18

!

E/e’ > 15 PAOP > 18

Page 28: The Echo Doesn't Lie by Murphy

S O M U C H F O R S TA T I C PA R A M E T E R S . .

Page 29: The Echo Doesn't Lie by Murphy

F U N C T I O N A L H A E M O D Y N A M I C S

Describing the effects of cardiorespiratory interactions in positive pressure ventilation

Page 30: The Echo Doesn't Lie by Murphy

IVC distensibility index

Change in IVC with positive pressure

!

> 18% significant !

Sensitivity 90% Specificity 100%

– Cut off of 18%

–Max IVC D-min IVC D/ Mean IVC D Max IVC diameter-min IVC diameter/ mean IVC diameterFeissel et al ICM 2004

Page 31: The Echo Doesn't Lie by Murphy

SVC collapsibility

V useful as intrathoracic

TOE

>36% significantMax-Min/Max

value

Viellard-Baron et al ICM 2004

Page 32: The Echo Doesn't Lie by Murphy

P U L S E P R E S S U R E VA R I A B I L I T Y / S T R O K E V O L U M E VA R I A B I L I T Y

Can assess with echo

Need to be v entilated

Sinus rhythm

Page 33: The Echo Doesn't Lie by Murphy

PA S S I V E L E G R A I S E

Page 34: The Echo Doesn't Lie by Murphy

Volume responsiveness and echo using passive leg raise

Change in VTI (SV) of 12% predicts fluid responsiveness

Lamia et al ICM 2007. Monnet at al CCM 2006

VTI =19 VTI =27

45%

Page 35: The Echo Doesn't Lie by Murphy

Mandeville. Can Transthoracic Echo be Used to Predict Fluid

Responsiveness in Critically Ill? Crit Care Research and Practice 2012

Page 36: The Echo Doesn't Lie by Murphy

3 HEART BEATS

INSPIRATION

POST INSPIRATORY DROP IN LV OUTPUT ONLY IF VOLUME RESPONSIVE

Page 37: The Echo Doesn't Lie by Murphy

B E N E F I T O F U S I N G E C H O

assess for false positives

Page 38: The Echo Doesn't Lie by Murphy

W O U L D Y O U G I V E F L U I D T O E I T H E R O F T H E S E PA T I E N T S ?

Page 39: The Echo Doesn't Lie by Murphy

A V O L U M E R E S P O N S I V E N E S S S T U D Y W I L L T E L L Y O U B O T H

Page 40: The Echo Doesn't Lie by Murphy

Increase intrathoracic pressure

Increase RV after load

Decreased RV stroke volume

Decrease LV stroke volume

Page 41: The Echo Doesn't Lie by Murphy

V E N T R I C U L A R S Y S T O L I C F U N C T I O N

Page 42: The Echo Doesn't Lie by Murphy

E C H O A S S E S S M E N T O F LV F U N C T I O N

Page 43: The Echo Doesn't Lie by Murphy

F R A C T I O N A L S H O R T E N I N G

!

many assumptions

inaccurate if wall motion abnormality

any errors in measurement will be cubed for EF measurement

Page 44: The Echo Doesn't Lie by Murphy

B I P L A N E S I M P S O N ’ S

1. Trace ED area A4C 1. Trace ES area A4C

Page 45: The Echo Doesn't Lie by Murphy

B I P L A N E S I M P S O N ’ S

3. Trace ED area A2C 4. Trace ES area A2C

Page 46: The Echo Doesn't Lie by Murphy

E Y E B A L L M E T H O D

Page 47: The Echo Doesn't Lie by Murphy

3 D E C H O

Impressive pictures and more accurate quantification

Page 48: The Echo Doesn't Lie by Murphy

R I G H T V E N T R I C L E

Page 49: The Echo Doesn't Lie by Murphy

R V S Y S T O L I C F U N C T I O N

TAPSE >1.6cm

(Tricuspid Annular Plane Systolic Excursion)

Page 50: The Echo Doesn't Lie by Murphy

R V S Y S T O L I C F U N C T I O N

S’ > 10 cm/S N

Page 51: The Echo Doesn't Lie by Murphy

S O W H Y U S E E C H O A S A H A E M O D Y N A M I C T O O L ? ?

Tells you what the problem is currently (not just the haemodynamic effects of the problem)

!

What’s causing it

!

If what you are doing about it helps

Page 52: The Echo Doesn't Lie by Murphy

C A S E S T U D Y

72 yo man post CAGs X 2 and AVR

“Good” LV intra-operative

Hypotensive

MAP 65 PAC: CO 3.6/ C.I 1.8

PA pressures 56/30

CVP 18

Management?

Inotropes and vasopressor: Milrinone 10 mcg/min, adrenaline 7 mcg/min, Noradenaline 17 mcg/min

Page 53: The Echo Doesn't Lie by Murphy

D I F F E R E N T I A L

? Tamponade

? Graft ischaemia

Page 54: The Echo Doesn't Lie by Murphy
Page 55: The Echo Doesn't Lie by Murphy
Page 56: The Echo Doesn't Lie by Murphy
Page 57: The Echo Doesn't Lie by Murphy
Page 58: The Echo Doesn't Lie by Murphy
Page 59: The Echo Doesn't Lie by Murphy

Dx LV outflow tract obstruction (with

SAM)

Rx: Avoid hypovolaemia

Avoid inotropy Maintain afterload

Page 60: The Echo Doesn't Lie by Murphy
Page 61: The Echo Doesn't Lie by Murphy
Page 62: The Echo Doesn't Lie by Murphy

D Y N A M I C LV O T O

Seen after cardiac surgery classically AVR

Seen in non-cardiac surgery patients also esp elderly females with hx HTN and DM

Haemodynamic situation worsened by inotropes and can contribute to downward spiral

Page 63: The Echo Doesn't Lie by Murphy

S U M M A R Y

Echo plays key role in assessment of haemodynamics

Helps identify false positives in terms of volume responsiveness

Adds a subtlety to the haemodynamic assessment

Is user dependant and like any tool is more powerful when used optimally