Page 1: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

Temperature Monitor with SMS Alerts

Page 2: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers Issue 1 Page 2 of 24

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

Quick Start Guide

1. Check that there is adequate mobile signal strength from the appropriate provider at the location where you intend to use the unit.

2. Download and install the “EasyLog SMS” App from Google Play store or the App Store.

3. CarefullyremovethebatteryplateandfityourSIMcardintothe microSIMholder.

4. Insert4xAAalkalinebatteriesintotheunitandreplacethebatteryplate.

5. Connectyourprobeintheexternaltemperatureprobeport.

6. Open the App on your smartphone or tablet and and add a new device. UsethemobilenumberthatwassuppliedwithyourSIMcard.

7. IfusingAndroid,selectthedevicejustaddedandpressthe“Setup”button.

8. Configuretheparameterstosuityourintendedapplication,thenpress the “Start” button (Android) or “Send SMS Message” button (iOS) and followtheon-screeninstructions,whichwillpreparetheunittoreceivethe configurationSMSandsendthemessage.

9. Oncetheconfigurationhasbeencompleted,youwillreceivea “Setup complete” message and the unit will begin monitoring.

@@@ [insert line diagram]

1

2

3

4

5

6

7

89

1 Status button2 Stop button3 Status LEDs4 Netlight5 External temperature probe port

6 Sounder7 Battery Bay8 Reset switch9 SIM card holder

Page 3: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers Issue 1 Page 3 of 24

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

EL-SMS App User Guide - Android Page No.Features 4Compatibility/Limitations 4App Use 4

EL-SMS App User Guide - iOSFeatures 7Compatibility/Limitations 7App Use 7

EL-SMS-2G-TP - User GuideIntroduction 9Basic Operation 9

SIMCard 9Thermistor Probe 10TemperatureMonitoringInterval 10Multiple SMS Destinations 10Alarm Delay Operation 10AlarmDelayExample 10Operating Modes 11Standby Mode 11Setup Mode 11ConfiguringtheUnit 11Run Mode 13Stop Monitoring 13SMS Retry Process 13Security 14

Table of Contents

Hardware Operation 15Power Up LED Sequence 15LED Flash Sequences 15Netlight LED Operation 15Sounder 15Battery 16Aerial and Signal Strength 16Real Time Clock 17Time Zone 17Message Groups 17NotificationTypes 18“Setup Complete” Message 18“NoSMSRxd”Message 18“Temperature Alarm” Message 19“Summary” Message 19“Low Battery” Message 19

Probe Errors 20SIMCardMissing/ErrorDetection 21Disclaimers 21

SMS Limit 21SMS Delivery 21Warnings 21

Troubleshooting 22Other Information 23Basic Specifications 23Glossary of Terms 24

Page 4: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers Issue 1 Page 4 of 24

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

EL-SMS App User Guide - Android

Features ● Simple management of your EL-SMS devices.

Compatibility/Limitations ● CompatiblewithAndroidversions2.3.3+(API10). ● Android device must have mobile network service and the ability to send/ receivetextmessages.

App Use (App images are subject to change) 1. DownloadandinstalltheAndroidAppfromGooglePlay. 2. Open the App (see Fig. 1). 3. Add the EL-SMS device to your list by clicking the “Add logger button” at the lower right of the screen. 4. InputtheEL-SMSdevicemobilenumber.Optionallyaddadevicename andnotesregardingthedevice’slocation,useetc.thenpress“Save” (see Fig. 2).

Thedevicenumbercanconsistofthefollowingcharacters:“+0123456789”andmustbe7digitsormore.

The device name can consist of the following characters shown between the double quote marks below:

“abcdefghijklmnopqrstuvwxyzABCDEFGHIJKLMNOPQRSTUVWXYZ0123456789”

Thenamecanbeuptoamaximumof16characters.Ifleftemptythedefaultname“EasyLog SMS” will be used.

Fig. 1 Fig. 2

Page 5: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers Issue 1 Page 5 of 24

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

App Use (Continued)

The EL-SMS device is now in your device list. (See Fig. 3).

5. Toconfigurethedevice,tapthedevicenameinthelisttobringupthe optionsmenu(SeeFig.4).

a. SETUP–AllowsyoutoentertheFullSetupmenu,seesection6below. b. EDIT/VIEW – Edit or view the basic details of the device. c. CHANGE ICON – Allows you to change the icon for the device to a photo or reset it back to the default. d. DELETE DEVICE – Removes the device from the App.

6. FullSetup

Tap the setting you wish to change (See Fig. 5).

a. Probe selection - Allows you to select the correct probe type for your device. Select Type 2 for the standard probe (-TP model) or Type A for the high accuracy probe (-TP+ and -PROBE-G models). b. Temperature Scale – Choose between Celsius and Fahrenheit temperature measurements (See Fig. 6). c. Temperature Alarms – Set the high and low alarm thresholds you require (See Fig. 7).

Fig. 3

Fig. 4

Fig. 5

Fig. 6

EL-SMS App User Guide - Android

Page 6: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers Issue 1 Page 6 of 24

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

App Use (Continued)

d. AlarmDelay–Configurehowmanyreadingsmustoccurinanalarm conditionbeforeyouarenotifiedandhowmanyreadingsafterthe temperature normalises before a new alarm condition will be sent to you. (See Fig. 8). e. SummaryInterval–TheintervalatwhichtheEL-SMSdevicewillsend you information regarding its status. (See Fig. 9). f. InputuptothreecontactnumbersthatthisEL-SMSdeviceshouldnotify, and select which messages they will receive (See Fig. 10). g. Pressthestartbuttontosendtheconfigurationmessagetothe EL-SMS device.

7. Viewingnotifications–Fromthehomepageswipeleftortap “Notifications”toviewthenotificationlist(SeeFigs.11and12).

This list contains the messages you have received from EL-SMS devices in your list.Tapanotificationtoviewthemessagedetails.Messagescanbedeletedbypressingthetrashcanbuttononthenotification,orbylong-pressingonthenotificationinthenotificationlist.

Fig. 7

Fig. 10

EL-SMS App User Guide - Android

Fig. 8 Fig. 9

Fig. 11 Fig. 12

Page 7: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers Issue 1 Page 7 of 24

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

EL-SMS App User Guide - iPhone and iPad

Features ● Simple management of your EL-SMS devices.

Compatibility/Limitations ● CompatiblewithiOSversions9.2andabove. ● Apple device must have mobile network service and the ability to send/ receivetextmessages.

App Use (App images are subject to change) 1. DownloadandinstalltheAppfromtheAppStore/iTunes.

2. OpentheApp(seeFig.1).FullSetup–Configurethedevicebytapping the options in the list. 3. Select and enter the EL-SMS device mobile number. Optionally add a device name then press “Save” (see Fig. 2). 4. ProbeSelection-Selecttheappropriateprobetype.SelectType2for the standard probe (-TP model) or Type A for the high accuracy probe (-TP+ and -PROBE-G models). 5. TemperatureScale–ChoosebetweenCelsiusandFahrenheittemperature measurements (See Fig. 3).

Fig. 1

Fig. 2

Fig. 3

Page 8: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers Issue 1 Page 8 of 24

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

6.TemperatureAlarms-Setthehighandlowalarmthresholds(SeeFig.4).

7.AlarmDelay–Configurehowmanyreadingsmustoccurinanalarmcondition beforeyouarenotifiedandhowmanyreadingsafterthetemperaturenormalises beforeanewalarmconditionwillbesenttoyou(SeeFig.5).

8.SummaryInterval–TheintervalatwhichtheEL-SMSdevicewillsendyou informationregardingitsstatus.(SeeFig.6).

9.ContactInformation–InputuptothreecontactnumbersthatthisEL-SMSdeviceshouldnotifyandselectwhichmessagestheywillreceive(SeeFig.7).

10.Pressthe“SendSMSMessage”buttontosendtheconfigurationmessageto theEL-SMSdevice(SeeFig.1).

EL-SMS App User Guide - iPhone and iPad

Fig. 4

Fig. 5

Fig. 6

Fig. 7

Page 9: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers Issue 1 Page 9 of 24

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

IntroductionThe EL-SMS-2G range is a family of battery powered temperature alerts that will sendSMS(ShortMessageService)notificationmessagestodesignatedmobilephonenumberswheneverthetemperaturemonitoredexceedsaprogrammedupper or lower temperature limit.

Thetemperaturelimitsalongwithotherconfigurableparametersaresenttothe unitviaaconfigurationSMSwhichiscreatedusingthe“EL-SMS”App.

OnceconfiguredtheEL-SMSunitscansendperiodicsummarymessages toconfirmthattheyarerunningnormally.Summarymessagescontaintheminimum,maximumandaveragetemperaturesincetheprevioussummarymessage,alongwithahistoryofthethreemostrecenttemperaturelimitbreachesthat the unit has detected.

These products use the GSM (Global System for Mobile Communications) network todelivertheSMSnotificationmessagesandmustbefittedwithamicroSIMcardto allow them to connect to a GSM network and register. These units use a built-in aerial to allow communication with the GSM network.

Notificationmessagescanbesenttouptothreeprogrammedmobilephonenumbers.Eachnumbercanbeindividuallyconfiguredtoreceivespecificalerts,i.e.Alarm,LowBatteryorSummarymessages.

Basic OperationSIM Card

User Guide

TheunitsrequireamicroSIMcardtooperate.ItisimportantthatnoPINcodeisenabledasthiswillpreventoperation.Ifnecessary,anormalmobilephonecanbeusedtodisabletheSIMPINcodelock.BeawarethattheproductwilleraseanyexistingSMSmessagesthatarealreadystoredontheSIMcard.

TheSIMcardmustbefittedtooperatenormally.

All2GmicroSIMcardsshouldworkaslongasthePINcodefunctionisdisabled.Visithttp://maps.mobileworldlive.comtocheckforGSMcoverageinyourgeographic location.

Page 10: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers Issue 1 Page 10 of 24

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

Thermistor ProbeThese units are supplied with a thermistor probe to sense temperature. The EL-SMS-2G-TP is supplied by default with a Type 2 (Standard) thermistor probe. TheEL-SMS-2G-TP+issuppliedbydefaultwithaTypeA(HighAccuracy)probeand the EL-SMS-2G-PROBE-G with a Type A probe inserted into a glycol bottle.

Thethermistorprobesconnecttotheunitviathe3.5mmjacksocket.

Thethermistorprobetypecanbeselectedwhentheunitisconfiguredusing the App.

Temperature Monitoring IntervalThese units will take a temperature reading every minute. For this reason it may take up to one minute before the unit responds to a change in temperature.

Multiple SMS DestinationsTheunitscansendnotificationstoamaximumofthreemobilephonenumbers.

Itshouldbenotedthatoperatinganyunitwithmorethanonedestinationmobilephonenumberwillsignificantlyreducetheexpectedbatterylifeandmayincuradditional costs as more messages are sent.

When multiple destination numbers are used these units will attempt to send notificationmessagestoeachphonenumber.Iftheunitisabletosuccessfullytransfer the SMS to the mobile network for at least one of the destination numbers then it will assume that the delivery of the message was successful. Be aware however that actual delivery of the message is reliant on the mobile network and mobile operator’s services operating normally.

User Guide

Alarm Delay OperationThese units allow a separate pre and post alarm delay to be applied to any set temperaturealarms.Thisisusefultopreventsendingexcessivetextmessagesifthemeasuredtemperatureisfluctuatingaroundeitherthehighorlowalarmlevels.

Whenaknock-indelayisused,theseunitswillwaituntilthatnumberofconsecutivetemperature readings have all been in an alarm condition before sending a temperaturealertmessage.Ifaknock-outdelayisused,theywillstayinanalarmstateuntilthespecifiednumberofconsecutivereadingsarealloutsideofanalarmcondition.Whentheknock-outdelayhascompleted,theywillonceagainrespondto new alarm conditions when they occur.

Ifthealarmdelayfeatureisnotrequiredthenthedelayvaluesshouldbesettoone.

Alarm Delay ExampleConsiderthefollowingsetupconfiguration:

Lowtemperaturealarmsetat12°Cusingan“Indelay”of15andan“Outdelay” of30.

Withthesesettingsthetemperaturemustdropbelow12°Candstaybelowfor 15minutesbeforeanalarmconditionwillberegistered.Duringthis15minuteperiodifanytemperaturereadingisabove12°Cthenthedelayisreset.Thismeansthatthetemperaturemustthenremainbelow12°Cforafurther15minutesbeforethe alarm is registered.

Oncethealarmhasbeenregistered(after15consecutiveminutesbelow12°C)alowtemperatureSMSnotificationwillbesent.

Page 11: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers Issue 1 Page 11 of 24

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

Inorderforthealarmconditiontobeclearedthetemperaturemustriseandstayabove12°Cforatleast30minutes.

Operating ModesTheEL-SMS-2Gdeviceshavethreeoperatingmodes,theseare:

Standby mode Setup mode Run mode

Standby ModeStandbymodeisthemodethattheunitwillbeinwhenitisidle,i.e.whentheunitis not being setup or running. This state consumes minimal power. A one second pressofthe“Status”buttonwillconfirmiftheunitisinthisstate.TheunitwillshoweitherfourgreenorfiveredLEDflashes.FourgreenLEDflashesmeantheunitisreadytobeginitsfirstsetupattempt.FiveredLEDflashesmeanthatthelastattempt to set up the unit failed but it is ready for another attempt.

The unit will enter standby mode under the following conditions:

● Whenitispoweredonandtheunitisnotconfigured.

● Whenitispoweredonandtheunitwasconfiguredbuttherealtimeclock hasresetduetoaninterruptiontothepower(e.g.ifthebatteriesranflat during operation).

● When an attempt to setup the unit has been made and does not complete successfully.

User Guide

● When the unit is in run mode and the stop monitoring button sequence is pressed.

Setup ModeThisisthemodethatallowsaconfigurationSMStobesenttotheunitviatheApp.

Theunitwillnotreceiveaconfigurationmessageuntilyouhavefollowedthestepsbelowandtheunitisreadytoreceiveit.Ifyoutrysendingaconfigurationmessagebeforethis,theunitwillnotreceiveit.Theconfigurationprocesscanbestartedfromeither standby or run mode.

Configuring the Unit ● Use the App to select the parameters you require.

Note that the Apps do not save settings for each device. The Android App will remember the settings used for the last device configured.

● IntheApp,pressthe“Start”button(Android)or“SendSMSMessage”(iOS) at the bottom of the “Setup” screen.

● Pushthe“Status”buttonontheEL-SMSunitforfivesecondsandthen releasethebutton.TheredLEDshouldbeginflashingonceeveryfive seconds. Wait until you hear a triple beep from the sounder. When this happenstheLEDwillchangetoadoublegreenLEDflasheveryfive seconds.

● TheEL-SMSunitisnowreadytoreceivetheconfigurationSMS.Goahead

Page 12: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers Issue 1 Page 12 of 24

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

and press the “Send” button in the App.

● Once you hear a double beep from the sounder you know the unit has successfullyreceivedtheconfigurationmessage.Iftheconfiguration SMSwasnotreceivedwithintwominutesyouwillhearfiveshortbeeps from the sounder to signal that the setup did not complete successfully.

● Iftheconfigurationcompletedsuccessfullyyouwillreceivea “Setup complete” SMS on the mobile device that was used to send theconfigurationmessagetotheunit.

● TheunitwillthenenterRunmodewhereitwillbeginmonitoringtheexternal temperatureforalarmbreaches.Thiscanbeconfirmedbypressingthe status button for one second and checking that the LED’s operate as per the LED Flash Sequences.

Please note that the unit will not respond to button pushes once the setup sequence has started.

The length of time required for the unit to register on the mobile network can vary andisdependentontheunithavingsufficientsignal.Iftheunitisunabletoregisterwithintwominutesitwillexitthesetupattempt.ThiswillbeindicatedbyfiveshortbeepsfromthesounderandfiveredLEDflasheswhenthestatusbuttonispushed.

SometimestheunitmaynotbelisteningforSMSmessageswhentheconfigurationSMS is sent. This could be due to delays inherent in the GSM network such as networkoutages,oversubscription,routinemaintenancework,messagecentretechnicalissuesetc.Ifasetupattemptdoesfailyoucansimplyrepeatthesetupprocess.Ifrepeatedattemptstosetuptheunitarefailingthenyoushouldcheckthatavailablesignalstrengthissufficientandtherearenoknownnetworkissuesthat

User Guide

could be causing the problem.

IfasetupattemptfailedbecausetheunitdidnotreceiveavalidconfigurationSMS:

1. Iftheunitwasnotpreviouslyconfiguredtheunitwillenterstandbymode again.WhenthestatusbuttonispusheditwilldisplayfiveredLEDflashes showingthatthelastattempttoconfiguretheunitfailed.

2. Iftheunitwaspreviouslyconfigureditwillsenda“NoSMSRxd”notification tothemobilephonenumbersthatwerepreviouslyconfiguredtoreceive “Summary”groupnotifications.TheunitwillthengobackintoRunmode usingitspreviouslyconfiguredsettings.

IftheunitisconfiguredandoperatinginRunmodeitshouldbenotedthatmakinganewattempttoconfiguretheunitwillstopitmonitoringtemperature.Ifthesetupdoesnotcompletesuccessfullythenitwillreverttorunmodeagain,however,thefollowing parameters will have changed:

● Summarymessageswillbedelayedbyuptofiveminutes. ● Temperature delay counts will be cleared. ● Low and high temperature alarm history will be cleared. ● Minimum,maximumandaveragetemperatureswillbereset. ● Anypendingbatteryloworsummarynotificationswillbecleared.

Page 13: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers Issue 1 Page 13 of 24

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

Run ModeFollowingasuccessfulconfigurationoftheunititwillenterintoRunmode.Inthismodetheunitcheckstoseewhetheranynotificationmessagesareduetobesent(i.e.alarmbreach,batteryloworsummarynotifications)andthenprocessesthenotificationsinorderoftheirpriority:

● Temperature alarms have the highest priority ● Low battery alerts have medium priority ● Summary messages have the lowest priority

InRunmodethetemperatureismeasuredonceeveryminuteusingtheexternalprobe and the value is checked to see whether either the upper or lower temperaturealarmlimitshavebeenbreached.Ifanalarmlimitisbreached(seealarmdelayoperation)theunitwillsendanalertnotificationmessagetothecontacts registered to receive the “alarm” group messages.

The battery condition is checked each time a temperature measurement is made. Ifalowbatteryconditionisdetectedabatterylownotificationmessagewillbesent.

The buttons are disabled whenever the unit is actively sending an SMS.

Stop MonitoringWhen the unit is in Run mode it can be put back into a Standby mode by pressing the “Stop monitoring” button combination. This should be done whenever you want to stop the unit from monitoring the temperature and sending alert messages.

Tostoptheunit,pressthestatusbuttonforfivesecondswhilstholdingdownthe“Stop” button for at least one second. The unit will show the green LED for three seconds after which it will reset. The power up LED sequence will then be displayed

User Guide

andtheunitwillbeinthestandbymodeinan“unconfigured”state.

To start monitoring again you will need to repeat the setup process.

SMS Retry Process

Start SMS alert

Alert success

Continue monitoring

Allow up to 2 minutes to register on mobile network

Success/Failure?

No

Yes

No

Yes

Make 2 attempts to send alert to each stored number

At least 1 SMSdelivered to network?

Was this a retry attempt?

5 Minute holdoff period then retry

attempt.

No more retries. Alert failure. Alert will

be dropped.

Page 14: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers Issue 1 Page 14 of 24

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

User Guide

SMS Retry Process (Continued) The unit uses the retry sequence shown in the above diagram when attempting to send either a temperature alert or battery low message:

Duringtheretryprocess,anyotherpendingnotificationmessageswillbepostponed.

Theunitwillassumethatanotificationmessagehasbeensuccessfullydeliveredaslong as the message was successfully transferred to the mobile network for at least one of the programmed numbers.

Ifafterfollowingtheaboveretrysequence,anotificationmessagestillcan’tbesenttothemobilenetwork,thenthealertwillbecancelledandtheunitwillcontinuenormaloperationlookingfornewalertconditionstooccur.Ifthecancelledmessagewasanalarmnotificationthealarmconditionmaystillbereportedinthefollowing:

1)Thealarmhistoryinthenextalarmmessage. 2)Mostrecentalarminthenextsummarymessage.

Theunitwillonlymakeoneattempttosendasummarymessage.Iftheattemptis unsuccessful then the summary message will be cancelled. Regardless of the outcometheminimum,maximumandaveragetemperaturedatafortheperiodwillbe reset.

1EasyLog Office; ,6 ,5*#,55*5000000000000000000 00000000000000000 0000000000000000007*77224+2+225+297222272292,*222292 792222222*25*757+5#2 7522622725229225222

Example screen shot of encrypted text

SecurityTheonlywaytochangetheconfigurationofarunningEL-SMSunitistophysicallygototheunittoinitiateanewconfigurationsequence.Ifsecurityintheenduserapplication is important then this should be considered when installing the unit.

Setting up the unit requires that someone is physically with the unit to begin the setupsequencebypressingthe“Status”buttonforfiveseconds.Theconfigurationmessage itself could be sent from any location so long as the person with the unit canpressthebuttonincoordinationwiththepersonsendingtheconfigurationSMS.

TheunitwillacceptaconfigurationmessagefromanymobiledevicerunningtheApp.Onceasetupattempthasfinished,theunitwillnolongerrespondtoincomingmessagesunlessanewsetupisattempted.Theconfigurationmessagesenttotheunit from the app is sent partially encrypted although the unit name is readable.

Page 15: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers Issue 1 Page 15 of 24

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

Hardware Operation

Power up LED sequenceWhentheunitispoweredon,thefollowingLEDflashsequenceisdisplayed:

RedLED,GreenLED,NoLED’s

LED Flash Sequences

Start up flash sequence RedLED,GreenLED-0.5seachStatus display (standby mode)Short push of status button.

4greenflashes=Setupnotyetattempted.5redflashes=Setupattemptedbutfailed.1.5sredLED=Lowbattery

Setup mode sequence5 second push of status button

1redflashevery5seconds=NotreadytoreceiveSMS.2greenflashesevery5seconds=Readyto receive SMS.

Status display (Run mode) Push of status button

2redflashes=Inalarmcondition(Highestpriority) 3redflashes=Lowbattery(Mediumpriority) 1greenflash=Running,noalarmconditions(Lowpriority)

Stop monitoring GreenLEDonfor3sthenofffor3sfollowedbyreset. Unit“Configured”statusiscleared.

SIM Card Error 4redflasheseverysecondforapproximately 10minutes.

User Guide

Netlight LED OperationThe orange Netlight status LED will show you the status of the internal 2G GSM modulewhenitisinuse.TheflashsequencesdisplayedbytheNetlightareasfollows:

Off GSMmoduleispoweredoff.1 second flash GSM module is not registered on a mobile network.2 second flash GSM module is registered on a mobile network.

SounderTheunitisfittedwithasounderwhichisusedtoindicatevariousoperatingconditions as follows:

Ready for configuration message Triple beepConfiguration message received Double beepNo configuration message received OR message was not compatible with the unit

Five short beeps

Page 16: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers Issue 1 Page 16 of 24

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

BatteryFor reliable operation of the unit it is recommended that only high quality alkaline batteries are used. This will ensure the best possible battery life.

WhentheunitisinStandbymode,thebatterystatuswillbecheckedonceeverythreeminutes.Ifthebatteryislow,theredLEDwilllightfor1.5swhenthe “Status” button is pushed. The unit will not send a low battery alert when it is in Standby mode.

WhentheunitisinRunmode,itwillsendalowbatteryalertwhenitdetectsalowbattery condition.

Itshouldbenotedthatthebatterycapacitywillbereducedwhentheunitisoperating at lower temperatures. Also be aware that setting multiple destination phonenumberstoreceivenotificationmessageswilldrainthebatterymuchfasterand as such it is recommended that only one of the programmed phone numbers isconfiguredtoreceivethelowbatteryalert.

Iftheunitisnotgoingtobeusedforanextendedperiodoftimethebatteries should be removed from the product.

Theexpectedbatterylifeatroomtemperature,sendingadailysummarymessageanduptoonealarmnotificationperweek,isoneyear.

Thisestimateassumesthatonlyonedestinationnumberhasbeenconfigured.Usingadditionaldestinationnumberswillreducetheexpectedbatterylife.

User Guide

Aerial and Signal StrengthThe unit uses a built-in aerial to connect to the mobile network. The reliability of the unit for sending messages is dependent on the signal strength available at the locationofoperation.Thephysicallocationoftheunitwillalsoaffecttheavailablesignal strength and should be considered when situating the unit.

Themostrecentsignalstrengthisdisplayedineachnotificationmessagethatissent by the unit. For reliable operation the signal strength should be above three bars. Operating below this signal strength may prevent the unit from being able to send messages. Note that the unit may not always connect to the same service provideriftheSIMcardpermitsroamingaccess.

The following table shows how the unit will report and display 2G signal strengths.

Signal strength

(bars)

Signal strength (dBm)

Signal strength (ASU, 0 - 31)

Usability

1 -115to-107 0to3 WeakSignal/Intermittentoperation2 -106to-99 4to7 PoorSignal/Intermittentoperation3 -98to-87 8to13 Good / Reliable operation4 -86to-77 14to18 Good / Reliable operation5 -76to-53 19to31 Excellent/Reliableoperation

FailuretoregisterontoamobilenetworkwouldusuallyindicateeitherinsufficientsignalstrengthorthatthemobilenetworkhasfailedtovalidatetheSIMcard(e.g.barredorblacklistedSIMnumbers).

Page 17: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers Issue 1 Page 17 of 24

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

Real Time ClockThe unit contains a real time clock which is used to add a timestamp to the outgoing messages.Theclockissetautomaticallywhentheunitreceivesaconfigurationmessage from the App.

Itshouldbenotedthattherewillbeadelayfromtheconfigurationmessagebeingsent from the App to being received by the unit. This delay is largely dependent on the mobile network and service provider and will therefore be unpredictable. This means that the clock in the unit may lag behind the actual time of the device usedtosendtheconfigurationmessagebytheamountthattheconfigurationmessage was delayed.

Also note that the real time clock will be reset if the batteries are removed. TheclockissetbyreconfiguringtheEL-SMSunitusingtheApp.

Time ZoneOutgoingnotificationmessagessentbytheunitcontainthetimezoneaspartofthetimestampinformation.Thetimezoneissettobethesameasthelocaltimezonesettingonthemobiledevicethatisusedtoconfiguretheunit.

Message GroupsThe message groups allow each programmed mobile phone number to receive differentcombinationsoftheavailablenotificationmessages.Eachmobilephonenumber can be subscribed to any of the following message groups:

User Guide

● Alarms ● Low Battery ● Summary

Page 18: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers Issue 1 Page 18 of 24

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

User Guide

EasyLog SMSSetup complete05/06/2017, 09:17 (UTC+00)Cur: +23.8 CSig: 4 (-80 dBm)Bat: OKv1.2Type: 002SN: 876543212345678

EasyLog SMSNo SMS Rxd05/06/2017, 09:17 (UTC+00)Cur: +23.8 CSig: 4 (-80 dBm)Bat: OKv1.2Type: 002SN: 876543212345678

Notification Types“Setup complete” Message

The“Setupcomplete”messageisreceivedwhentheunithasbeenconfiguredandconfirmsthat setup of the unit was successful. This message contains the following information:

● Date/Timefieldforconfirmationthattherealtimeclockhasbeenset. ● Current temperature. ● Signalstrengthinbars(whereoneispoorsignalandfiveisexcellentsignal). ● Programmedfirmwareversion. ● Unit type number which is set at the factory and is used internally by the unit. ● Serial number of the unit.

“No SMS Rxd” Message

This“NoSMSRxd”messageisreceivedifasetupattemptfailsbecausenoconfigurationmessagewasreceived.ItisonlysentiftheunitwaspreviouslyinRunmode.Themessagewill be sent to phone numbers that are subscribed to the “Summary” message group.

Thisnotificationmessagecontainstheunique15digitserialnumberfortheunitandalsoshows the measured signal strength.

Page 19: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers Issue 1 Page 19 of 24

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

User Guide

EasyLog SMSA1: L -17.0 F05/06/2017, 12:15 (UTC+00)A2: PrbErr OA3: H +72.8 FSig: 4Bat: OK

EasyLog SMS06/06/2017, 10:00 (UTC+00)Cur: +23.8 CMin: -17.0 CMax: +24.1 CAvg: +12.4 CA1: L -17.0 C05/06/2017, 12:15Sig: 4Bat: OK

“Temperature Alarm” Message

The temperature alarm message shows a history of the last three temperature alarm conditions. The temperature reading shown with each alarm in the history is the temperature reading taken at the time the alarm was registered by the unit. Thetime,dateandtimezoneinformationisshownonlyforthemostrecentalarm.Hightemperaturealarmsareindicatedbythe‘H’character,lowalarmsareindicatedby‘L’.Probeerrors are shown with the message PrbErr followed by O or C for missing or broken probes.

“Summary” Message

Thesummarymessageissentattheintervalspecifiedwhentheunitwasconfigured.

The following information is shown in the summary message:

● Theminimum,maximumandaveragetemperaturereadingssincethelast summary interval. ● The most recent alarm condition and the date and time at which it was triggered. ● The most recent signal strength and battery status.

Notethattheminimum,maximumandaveragetemperaturereadingsareresetateachsummary message interval even if the summary message itself was unable to be sent due to lackofsignal,networkoutageetc.

Summary messages may be delayed if the unit is sending higher priority messages such as the temperature alarm or low battery messages.

“Low Battery” Message

The low battery message will be sent when the unit detects a low battery condition.

EasyLog SMSBat: Low27/09/2018, 11:43 (UTC+00)Sig: 4

Page 20: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers Issue 1 Page 20 of 24

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

User Guide

Example summary message with short circuit probe error

Example setup complete message with short circuit error

EasyLog SMS07/06/2017, 10:00 (UTC+00)Cur: PrbErr CMin: ---.-Max: ---.-Avg: PrbErrA1: PrbErr C07/06/2017, 03:30Sig: 4Bat: OK

EasyLog SMSSetup complete05/06/2017, 09:17 (UTC+00)Cur: PrbErr ClosedSig: 4 (-80 dBm)Bat: OKv1.2Type: 002SN: 876543212345678

Probe ErrorsIfthethermistorprobegoeseitheropenorshortcircuitduringnormaloperationthiswill be reported in the alarm and summary messages as follows:

PrbErr O is displayed for a disconnected probe error (open circuit). PrbErr C is displayed for a short circuit probe error (probably caused by a faulty probe).

SetupComplete/NoSMSRxdmessagereporttheprobeerrorsasfollows:

PrbErr Open is displayed for a disconnected probe error (open circuit). PrbErr Closed is displayed for a short circuit probe error.

Whenaprobeerroroccurs,theunitinternallyregistersthefollowingtemperatureforeach condition:

OpenCircuitProbeError -51degC ShortCircuitProbeError -50degC

Forthisreason,toreceiveprobeerroralarms,itisnecessarytohavethelowalarmenabled.

Whenaprobeerrorconditionisdetected,themin/maxtemperaturereadinginthenextsummarymessagewithbeshownas“---.-“,indicatingthataprobeerrorwasencounteredduringthelastsummaryinterval.Also,theaveragetemperaturereading will be shown as “PrbErr” since the average temperature reading for the summary interval will be invalid.

Page 21: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers Issue 1 Page 21 of 24

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

User Guide

SIM Card Missing / SIM Card Fault ErrorIfduringnormaloperationtheunitdetectseitheramissingorfaultySIMcard,theunitwillflashtheredledfourtimesasecondfor10minutestoindicatetheerror.The unit will then automatically reset.

Ifthiserrorisobserved,checkthataSIMcardisfittedandtheSIMcardcontacts are clean.

DisclaimersSMS LimitThis product does not limit the number of messages that can be sent in a given timeperiod.Thisshouldbeconsideredwhensettinguptheunit.Forexample,ifthemeasuredtemperatureislikelytofluctuatearoundtheprogrammedalertlevelthenyou should consider using the alarm delay feature to ensure that you do not receive notificationmessageseachtimeanalarmthresholdiscrossed.Similarlywhenyouhavefinishedusingtheunit,remembertousethe“Stopmonitoring”buttonsequencetoensurethattheunitwillnotsendanymorenotificationmessages. YoucanconfirmthattheunithasstoppedandisinStandbymodebypushingthestatus button and checking the LEDs.

Failuretoadequatelyconfiguretheproductmayleadtoexcessiveoperatingcosts.

SMS DeliverySMStextmessagingisa“storeandforward”messagingsystemwhichusesmessage service centres to temporarily store messages until they can be delivered. Whenusingthisproductbeawarethatthetextmessagingnetworkservicedoes

not guarantee delivery of the messages that are sent to it. Sent messages may not reach their intended destination and this should be considered when planning howthisproductwillbeusedinyourapplication.Youshouldalsobeawarethatthe delay between the unit sending a message and it being received can vary significantly.

Typicalfactorsthatcanaffectmessagedeliveryortiminginclude:

● Available credit.● Signal strength and quality.● Cell tower available capacity.● Sendingmessagesbetweendifferentnetworkoperators.● Agreements in place between network operators where roaming is used.● Available capacity at the sending and receiving message centres.● Network outage or planned maintenance.● Receivingdeviceswitchedoff,flatbatteryorno-signal(willeventually be discarded by the network if they cannot be delivered. The length of time varies depending on the service provider used).

WarningsAseparationdistanceof20cm(7.87”)ormoremustbemaintainedbetweentheantennaofthisproductandallpersonstocomplywithFCCRFexposurerequirements.

Page 22: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers Issue 1 Page 22 of 24

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

User Guide

Troubleshooting

Problem Possible Causes SolutionMy unit is not responding to button pushes. A setup attempt is underway. Wait until the setup attempt is complete and try again.

Unit has malfunctioned. Remove batteries. Wait one minute. Re-insert batteries to restart the unit.

Flat battery. Replacebatteriesandre-configuretheunit.A GSM operation is underway. Waituntilthenetlightstopsflashingandtryagain.

I’mnotgettingsummarynotificationsanymore. Flat battery. Replacebatteriesandre-configuretheunit.The unit has no signal. Check signal strength using a mobile phone and if necessary

re-position the unit.Possible mobile network outage / maintenance work. Check the signal strength shown in the last summary

messagewassufficientforreliableoperation.SMS message centre issues. Normal operation will resume when the network service

disruption has ended.Incorrectdestinationnumberenteredwhenunitwasconfigured.

Check destination number in the App and ensure it is enabled toreceivesummarymessage.Re-configuretheunitif necessary.

SIMcardregistrationproblem. EnsurethattheSIMcardaccountisactiveandhascredit.Itriedtosetupmyunitbutthesetuphasfailed. Insufficientmobilesignal. Check signal strength using a mobile phone and if necessary

re-position the unit.SIMcardnotfitted. Removebatteries,insertaSIMcardandthenre-insertthe

batteries.Network congested delaying reception of the message. Wait and try again.

Ican’tgetmyunitintosetupmode. “Status” button has not been pressed for long enough. Ensurethe“Status”buttonhasbeenpressedfor5secondsbefore being released.

Page 23: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers Issue 1 Page 23 of 24

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

User Guide

Other InformationMobileVirtualNetworkOperators(MVNO’s)resellaccesstothemobileinfrastructureofthemajornetworkoperatorsandgenerallyoffertheirowncustomerservice,billingsupportsystems,marketingandsalespersonnel.MVNO’scantypicallyofferbetterdealsthanyoumightgetthroughthemajornetworksdirectly.

Basic SpecificationsThermistor probe types Type2(Standard)orA(HighAccuracy)Thermistor probe measurement range

-40to+125°C(-40to+257°F)

Operating Temperature Range -20to+40°C(-4to+104°F)Resolution 0.1°C(0.2°F)Accuracy ±0.1°C(±0.1°F)*Temperature Monitoring Interval Once per minuteDimensions 105x105x29mm(4.1x4.1x1.1”)Weight (excl batteries) 128g(4.5oz)GSM Module QuadBand2GModule

Class4(2W@850/900MHz)Class1(1W@1800/1900MHz)

Internal Antenna Penta-band Embedded Antenna50OhmImpedanceLinear Polarisation3dBi GainVSWR<3:1

Battery Life 1Year(at25°C,dailysummarymessagetoonephone number)

IP Rating TheproducthasanIPratingofIP53provid-ing the thermistor probe is connected. This gives:- Protection against solid bodies larger than 1mm.-Protectionagainstwaterdropletsdeflectedupto15°fromvertical.

Ifyouwishtousetheunitoutdoors,thether-mistor probe must be connected and the unit should be positioned in a sheltered location.

* At 25°C. See probe accuracy curve on data sheet. Important, quoted accuracy applies only when the unit is operating within the specified operating temperature range. Thermistor probe errors are not included and should also be taken into consideration.

Supplied ‘Type 2’ Standard Thermistor ProbePart Number: EL-PROBE2-1.0M-TP

Supplied‘TypeA’HighAccuracyThermistorProbePart Number: EL-PROBEA-3.0M-TP+

Page 24: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers Issue 1 Page 24 of 24

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

Glossary of TermsGlobal System for Mobile Communications [GSM]StandarddevelopedbytheEuropeanTelecommunicationsStandardsInstitutetodescribeprotocols for second-generation (2G) digital cellular networks used by mobile devices.

Mobile Virtual Network Operator [MVNO] Thesearevirtualnetworkoperatorswhoresellaccesstothemajornetworkoperators’infrastructureandgenerallyoffertheirowncustomerservice,billingsupportsystems,marketing and sales personnel.

Short Message Service [SMS] Textmessagingserviceallowingshorttextmessagestobesentbetweendevicesconnected via a GSM network.

Subscriber Identity Module [SIM] Integratedcircuitthatsecurelystorestheinternationalmobilesubscriberidentity(IMSI)and the related key used to identify and authenticate subscribers on mobile telephony devices.Insimpleterms,achipwhichidentifieswhoispayingthebillforanynetworkservices that are used.

Page 25: Temperature Monitor with SMS Alerts · microSIM holder. 4. Insert 4 x AA alkaline batteries into the unit and replace the battery plate. 5. Connect your probe in the external temperature

www.lascarelectronics.com/data-loggers

EL-SMS-2G-TP / TP+ / PROBE-G USER MANUAL

Temperature Monitor with SMS Alerts