Annexure-A Technical Specifications of Obstetrics and Gynecology Equipments Contents Equip sr. no Equipment name Quantity 1 Colposcope 5 2 Operating Laparoscopy Set 6 3 LCD screen (Monitor) with trolley 10 4 Digital CO2 insufflator 7 5 Laparoscopy trolley and documentation System 7 6 Hysteromat Set 8 7 Vaginal Hysterctomy Set 24 8 Digital High Definition Video Recording System and documentation system 8 9 Three Chip camera head & focusing coupler 6 10 Shadow less mobile side lamp for O.T 28 11 Ultrasound with Doppler and in-built printing and recording system 5 12 Delivery Set 62 13 Obstetric labour table 46 14 Spermfuge Machine 5 15 IVF Instrument 3 16 LLETZ 3

Upload: others

Post on 29-Jul-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Technical Specifications of Obstetrics and Gynecology ......short Cleaning Adaptor and Colptomizer.USFDA / Eiropean CE approved.QTY-01 20 Macro needle holder with tungsten carbide

Annexure-A

Technical Specifications of Obstetrics and Gynecology Equipments

Contents

Equip sr. no

Equipment name Quantity

1 Colposcope

5

2 Operating Laparoscopy Set

6

3 LCD screen (Monitor) with trolley

10

4 Digital CO2 insufflator

7

5 Laparoscopy trolley and documentation System 7

6 Hysteromat Set

8

7 Vaginal Hysterctomy Set

24

8 Digital High Definition Video Recording System and documentation system

8

9 Three Chip camera head & focusing coupler

6

10 Shadow less mobile side lamp for O.T

28

11 Ultrasound with Doppler and in-built printing and recording system

5

12 Delivery Set 62

13 Obstetric labour table

46

14 Spermfuge Machine

5

15 IVF Instrument 3

16 LLETZ

3

Page 2: Technical Specifications of Obstetrics and Gynecology ......short Cleaning Adaptor and Colptomizer.USFDA / Eiropean CE approved.QTY-01 20 Macro needle holder with tungsten carbide

Annexure-A

1. Colposcope

Equipment Name Technical Specifications

Colposcope The video colposcope must have magnification from min.

1x to max. 40x.

Magnification Indicator should be on the video

colposcope

Resolution must be > 825 lines (Std.) > 1,000 lines

(Gamma on)

No of pixel should be 12,00,000

The image sensor must be 1/6 “ microlens color CCD with

high speed Digital Signal Processing

High MCD super bright white shadow less LED light for

true colour reproduction

Colour temperature should be 70000 K, light source life

should be minimum 5000 hrs

Auto focus range should be up to 30-40 cm

Facility for fast focusing zooming, light source, image

freeze using thump on the hand held unit itself

Should be with E-flip & Mirror image function to ensure

extended attention and perfect diagnosis

Acetic test timer and magnification indicator should be

displayed on screen

There must be Electronic Green filter in the hand held unit

without decrease in illumination

Control panel should have feather touch and water proof

buttons. Facility for fast auto/manual focusing. Internal

image freeze function facility. There should be two built

in Video output – BNC & SVHS on the unit. Company

should provide Colposcopy image Management Software

with computer with following facilities. Upgradable

software. Masking facilities. Image capturing while

recording/playing. Final reports with one, two, three &

four images with facility to adjust height & width of

images. Reference: images with finding for comparison.

Facility to save & send the report through e-mail in pdf

format. Facility to get referral linked images. Online

support facility (through internet) for software

Page 3: Technical Specifications of Obstetrics and Gynecology ......short Cleaning Adaptor and Colptomizer.USFDA / Eiropean CE approved.QTY-01 20 Macro needle holder with tungsten carbide

Annexure-A

Colposcopy software should run on both window XP &

Vista operating systems

Colposcopy assisted dynamic cases

Facility to take colposcopy images with the colposcopy

report on hard copy Facility to store still images, cine loop

or procedure on CD

Company have to provide trolley for personal computer

Colposcope should be CE, ERTL & ISO certified

Colposcope should be operated through remote control

also

Colposcope should be equipped with Gamma Processor

to enchance vascular structure

Colposcope should have Remote Control facility

2. Operating Laparoscopy Set

Sr. No. Description of Technical Specification

1 Straight Forward Telescope 0° enlarged view, diameter 10 mm, length 31 cm, autoclavable,

fiber optic light transmission incorporated.connection for fiber optic light cable offset by

90°, USFDA / European CE approved.Telescope should have definind length and capability

ofautoclavable for Sterilization QTY-01

2 Forward-Oblique Telescope 30° enlarged view, diameter 5 mm, length 29 cm, autoclavable,

fiber optic light transmission incorporated, connection for fiber optic light cable offset by

90°, USFDA / Eiropean CE approved..Telescope should have definind length and capability

of autoclavable for Sterilization QTY-01

3 Specification rod lens system Scratch resistant Adaptability to high intensity Xenon light

300 watts

4 Trocar -5/6mm diameter; should have multifunctional valve to prevent damage of sharp

instruments and tip lens while passing through the cannula valve. It should have stopcock

for CO2 insufflation. Trocar should have pyramidal tip with pin holes near the tip for safety

outlet of CO2 gas. The working length of the cannula should be 105 mm.USFDA / Eiropean

CE approved. should have multifunctional valve to prevent damage of sharp instruments

and tip lens while passing through the cannula valve. QTY-04

Page 4: Technical Specifications of Obstetrics and Gynecology ......short Cleaning Adaptor and Colptomizer.USFDA / Eiropean CE approved.QTY-01 20 Macro needle holder with tungsten carbide

Annexure-A

5 Trocar -10/11 mm diameter; should have multifunctional valve to prevent damage of sharp

instruments and tip lens while passing through the cannula valve. It should have stopcock

for CO2 insufflation. Trocar should have pyramidal tip with pin holes near the tip for

safety outlet of CO2 gas. The working length of the cannula should be 105 mm. USFDA /

Eiropean CE approved.USFDA / Eiropean CE approved. should have multifunctional valve

to prevent damage of sharp instruments and tip lens while passing through the cannula

valve. QTY-03

6

Reducer 11/5 mm-USFDA / Eiropean CE approved. QTY-02

7 Coagulating & Dissectuing Electrode, L- shaped, with connector pin for unipolar

coagulation.USFDA / Eiropean CE approved. QTY-02

8 Three part laparoscopic autoclavable Grasping Forcep with ratchet, Tiger Jaws, 2X4 teeth,

360 degree rotational sheeth, with connector pin for unipolar coagulation, size 5 mm, length

36 cm, long, double action jaws, with ergonomic plastic handle with larger contact area at

the finger ring to avoid pressure sores, can be dismantled with the press of a button, USFDA

/ Eiropean CE approved. QTY-01

9 Three part laparoscopic autoclavable KELLY Dissecting and Grasping Forcep, 360 degree

rotational sheeth, with connector pin for unipolar coagulation, size 5 mm, length 36 cm,

long, double action jaws, with ergonomic plastic handle with larger contact area at the finger

ring to avoid pressure sores, can be dismantled with the press of a button, USFDA /

Eiropean CE approved. QTY-01

10 Three part laparoscopic Bipolar Maryland Dissecting & Grasping Forceps with overload

tissue protection, Wide jaws, with connector pin for bipolar coagulation, 360 degree

rotational sheeth, size 5 mm, length 36 cm, long, single action jaws, ergonomic Plastic

handle with larger contact area, can be dismantled with the press of a button. USFDA /

Eiropean CE approved. QTY-01

11 high frequency cord for bipolar instruments;USFDA / Eiropean CE approved. QTY-02

12 Fiber optic light cable, size 4.8 mm, length 300 cm.USFDA / Eiropean CE approved. QTY-

01

Page 5: Technical Specifications of Obstetrics and Gynecology ......short Cleaning Adaptor and Colptomizer.USFDA / Eiropean CE approved.QTY-01 20 Macro needle holder with tungsten carbide

Annexure-A

13

Three Chip Camera System Specifications:Digital Three-Chip Medical-Video-Camera –

color system PAL, with integrated Image Processing Module. Special Features:-• Should

be fully compatible to the color systems PAL & NTSC.All the necceasry setting required to

be done during surgery in the camera should be under sterile area. • Should have the

facility of Digital Image Processing by means of an integrated Image Processing Module.

Multiple settings allow the user to select the preferred level of image enhancement.

-Digital contrast enhancement -Digital anti-moiré/anti-grid filter for use with

fiberscopes.•Should have Parfocal Optical Zoom lens system.• Should have the additional

facility of Electronic 2x zoom, selectable in 4 steps.Integrated Still Image function.• Digital

Video output DV. For direct hook up with Lap top • Should have the Auto Exposure

System for consistent illumination level.• High horizontal image resolution of more than

750 lines, therefore even the finest variations in tissue structures are perceivable on a high-

resolution monitor. • Automatic white balance with memory function for two settings.

• Multilingual menu system.• Character generator • Composite, S-VHS and RGB

compatibility• World power supply • 2 programmable function keys on the camera head for

control of camera functions or video printer/recorder functions & other peripheral units. •

Adaptable to an operating microscope by means of a special quick adapter. • Camera head

fully soakable for disinfection. USFDA / Eiropean CE approved. QTY-01

14 Suction and Irrigation Tube, anti-reflex surface with two-way stopcock, for single hand

control, size 5 mm and 10mm, length 36 cm. USFDA / Eiropean CE approved. QTY-01

EACH

15 Veress Pneumoperitoneum needle with spring loaded blunt inner cannula, LUER-Lock,

autoclavable, diameter 2.1mm , working length 10- 18 cms. USFDA / Eiropean CE

approved.Veress Pneumoperitoneum needle with spring loaded blunt inner cannula, LUER-

Lock, autoclavable, diameter 2.1mm , working length 13 & 15 cms. QTY-01 EACH

16 METZENBAUM Scissors, 360 degree rotational sheeth, with connector pin for unipolar

coagulation, size 5 mm, length 36 cm, long, double action jaws, with ergonomic plastic

handle with larger contact area at the finger ring to avoid pressure sores, can be dismantled

with the press of a button. and Straight Scissor also.USFDA / Eiropean CE approved. QTY-

02 EACH

17 Monopolar high frequency cord-USFDA / Eiropean CE approved.QTY-02

18

PCOD Needle-USFDA / Eiropean CE approved.QTY-02

19 Uterine Manipulator, consisting of: Handle with fixation screw Manipulator Rod 2 Sealing

Cylinder , Silicone Seal, package of 3 (3 sizes) Sheath ,Working Insert, conical, with thread,

medium Working Insert, atraumatic, diameter 7 mm, length 50 mm Working Insert, with

connector for chromopertubation, atraumatic, Ø 4 mm, length 40 mm , Anatomical Blade,

short Cleaning Adaptor and Colptomizer.USFDA / Eiropean CE approved.QTY-01

20 Macro needle holder with tungsten carbide insert, dismantling into three parts namely, outer

tube, handle and inserts, ergonomic pistol handle, with disengageable ratchet, jaw curved to

left and Stright, size: 5mm, length : 33cm for use with suture material size: 0/0 to

7/0.USFDA / Eiropean CE approved.QTY-01 EACH

21 Post closure USFDA / European CE approved.QTY-01

Page 6: Technical Specifications of Obstetrics and Gynecology ......short Cleaning Adaptor and Colptomizer.USFDA / Eiropean CE approved.QTY-01 20 Macro needle holder with tungsten carbide

Annexure-A

22 Knot Pusher USFDA / European CE approved. QTY-01

23 BABCOCK Grasping Forceps, rotating, with connector pin for unipolar coagulation, size 10

mm, length 36 cm, double action jaws. USFDA / Eiropean CE approved. QTY-01

24 Xenon light source of 300 Watts : Lamp type:- Xenon 300 Watt • Should be able to

produce color temperature of 6000 K, • Light Outlets – 1, • Should have continuous manual

adjustement of light output or automatically by camera head. • Should be supplied with one

Light cable Diameter 4.8mm, Length 300cm. Should be offered with one spare xenon lamp

300 watts. USFDA / Eiropean CE approved.QTY-01 .

25 CO2 Electronic Insufflator: Electronic CO2 insufflators Should have an adjustable flow

rate Gas flow/min 0-20 ltrs, Pressure (mmHg) 0-30 (mmHg), Intra-abdominal pressure

guage 0-50 (mmHg), Power supply 100-240 VAC (50/60 Hz), Dimension 305 x 164 x 233

mm (w x h x d), Weight 6 kg, Certified to IEC 601 – 1, CE label according to MDD. -High

degree of patient safety. -Clear, adjacent display for set value and actual value allow easy

monitoring of insufflations process, - Touch keys for precise pre-selection of set values. -•

Can be control with camera head, it makes operating for the surgeon a highly ergonomic and

pleasant working experience.Should have High Pressure Hose, American connection / pin-

Index connection (CO2), length 102cm.Safety – constant monitoring of Intra-abdominal

pressure; any overpressure is reduced immediately. USFDA / Eiropean CE approved. QTY-

01 .

26 MEDICAL GRADE 19 Inch Table Mounted LCD VIDEO MONITOR, Special features

: Multi-norm monitor for PAL & NTSC with automatic Switchover. Liquid Crystal

Display.Composite, S-Video, RGB,SDI/HD-SDI,DVI-D,VGA, compatible. Compact &

Light weight design.Drip Water protected, dustproof housing.On-Screen menu for Monitor

setting (in multiple language possibility).Brightness 450 cd/m2.Max Viewing Angle 178

degree Vertical.Pixel Distance: 0.294mm.Contrast: 600: 1.Input Signal Level: 0,7 Vp –

P.Rated Power: 40 Watt.Operating conditions: 0 – 38 degree Centigrade.Storage: At 20 – 60

degree Centigrade.Relative Humidity: 5 – 90% non-condensing.Dimensions: W x H x D:

465 x 400 x 98.Power Supply: 100 – 240 VAC, 50/60 Hz.Certification: EN 60601,

protection class IPX1.Compatible with 4:3,5:4,16:9 and 16:10 signal format display Picture

in picture Certified to EN60601-1,UL60601-1,MDD93/42/EECProtection class IPX1.

USFDA / Eiropean CE approved. QTY-01 .

27 Equipment Cart:- Equipment Cart wide, high , rides on 4 antistatic dual wheels equipped

with locking brakes, mains switch on cover, central beams with integrated electrical sub

distributors with 12 sockets, grounding plugs, Dimensions in mm (w x h x d):Equipment

cart: 830 x 1474 x 730,Shelf: 630 x 25 x 510 Caster diameter: 150mm ,consisting of:- Beam

Package, equipment cart high 3 X Shelf, wide , Drawer Unit with Lock, wide, - 2X

Equipment Rail , long, - Camera holder. Monitor Holder, Monitor Holder , height adjustable

, swiveling and tilting , swivel range approx. 360 degrees.Load capacity max 18 kg, with

monitor fixation VESA 75/100.USFDA / Eiropean CE approved. QTY-01

Page 7: Technical Specifications of Obstetrics and Gynecology ......short Cleaning Adaptor and Colptomizer.USFDA / Eiropean CE approved.QTY-01 20 Macro needle holder with tungsten carbide

Annexure-A

3. LCD screen (Monitor) with trolley

Equipment Name Technical Specification

LCD screen (Monitor)with trolley

MEDICAL GRADE 19 Inch Table Mounted LCD VIDEO MONITOR, Special features : Multi-norm monitor for PAL & NTSC with automatic Switchover. Liquid Crystal Display.Composite, S-Video, RGB,SDI/HD-SDI,DVI-D,VGA, compatible. Compact & Light weight design.Drip Water protected, dustproof housing.On-Screen menu for Monitor setting (in multiple language possibility).Brightness 450 cd/m2.Max Viewing Angle 178 degree Vertical.Pixel Distance: 0.294mm.Contrast: 600: 1.Input Signal Level: 0,7 Vp – P.Rated Power: 40 Watt.Operating conditions: 0 – 38 degree Centigrade.Storage: At 20 – 60 degree Centigrade.Relative Humidity: 5 – 90% non-condensing.Dimensions: W x H x D: 465 x 400 x 98.Power Supply: 100 – 240 VAC, 50/60 Hz.Certification: EN 60601, protection class IPX1.Compatible with 4:3,5:4,16:9 and 16:10 signal format display Picture in picture Certified to EN60601-1,UL60601-1,MDD93/42/EECProtection class IPX1. USFDA / Eiropean CE approved. QTY-01

4. Digital CO2 insufflator

Equipment Name Technical Specification

Digital CO2 insufflator CO2 Electronic Insufflator: Electronic CO2 insufflators

Should have an adjustable flow rate Gas flow/min 0-20

ltrs, Pressure (mmHg) 0-30 (mmHg), Intra-abdominal

pressure guage 0-50 (mmHg), Power supply 100-240

VAC (50/60 Hz), Dimension 305 x 164 x 233 mm (w x h

x d), Weight 6 kg,

Certified to IEC 601 – 1, CE label according to MDD. -

High degree of patient safety. -Clear, adjacent display for

set value and actual value allow easy monitoring of

insufflations process, - Touch keys for precise pre-

selection of set values. -

• Can be control with camera head, it makes operating for

the surgeon a highly ergonomic and pleasant working

experience.Should have High Pressure Hose, American

connection / pin-Index connection (CO2), length

102cm.Safety – constant monitoring of Intra-abdominal

pressure; any overpressure is reduced immediately.

USFDA / European CE approved. QTY-01

Page 8: Technical Specifications of Obstetrics and Gynecology ......short Cleaning Adaptor and Colptomizer.USFDA / Eiropean CE approved.QTY-01 20 Macro needle holder with tungsten carbide

Annexure-A

5. Laparoscopy trolley and documentation System

Equipment Name Technical Specification

Laparoscopy trolley and documentation System

Equipment Cart, wide, high , rides on 4 antistatic dual wheels equipped with locking brakes, mains switch on cover, central beams with integrated electrical sub distributors with 12 sockets, grounding plugs, Dimensions in mm (w x h x d):Equipment cart: 830 x 1474 x 730,Shelf: 630 x 25 x 510 Caster diameter: 150mm ,consisting of:- Beam Package, equipment cart high 3 X Shelf, wide , Drawer Unit with Lock, wide, - 2X Equipment Rail , long, - Camera holder. Monitor Holder, Monitor Holder , height adjustable , swiveling and tilting , swivel range approx. 360 degrees.Load capacity max 18 kg, with monitor fixation VESA 75/100. USFDA / European CE approved.

6. Hysteromat Set

Equipment Name

Technical Specification

Hysteromat Set

Forward-Oblique Telescope 30°, enlarged view, diameter 2.9-3.0 mm, length

30 cm, autoclavable, fiber optic light transmission incorporated.QTY-01,

Hysteroscope sheath for diagonostic continuous irrigation size 4.5-4.6mm outer

sheath and 3.8-3.9mm inner sheath for use with 2.9-3.0mm hysteroscope

telescope.QTY-01 Hysteroscope sheath for Operative continuous irrigation

size 5.0-5.1mm outer sheath with operating channel 5Fr, leur lock adaptor and

4.3-4.4mm inner sheath for use with 2.9-3.0mm hysteroscope telescope.

With stopcock & leur lock adaptor and inflow and outflow channel.QTY-01.

Scissors semi rigid, blunt tips, 5 Fr., length 33-36cm, single action jaws-- Qty 02,

Scissors semi rigid, pointed jaws, 5 Fr., length 33-36cm, single action jaws,

semi-rigid ---Qty 02. Biopsy- and Grasping Forceps semi rigid, 5 Fr. ,

length 33-36cm, double action jaws---Qty 02, Unipolar Needle Electrode ,

5 Fr., length 34 cm,-Qty 02, Bipolar Dissection Electrode, 5Fr, 36cm,

Bipolar---Qty 02 , High Frequency Cord with 4 mm plug HF-unit, length of

cable should be minimum 300 cm ---Qty 01, Bipolar High Frequency Cord,

should fit in most of the standard electrosurgical unit having banana plug one

side, length should be not less than 250cm.

Page 9: Technical Specifications of Obstetrics and Gynecology ......short Cleaning Adaptor and Colptomizer.USFDA / Eiropean CE approved.QTY-01 20 Macro needle holder with tungsten carbide

Annexure-A

BIPOLAR WORKING ELEMENT SET:- BIPOLAR Working Element to be used with

22Fr Resectoscope sheath: Motion by means of a spring. The thumb support is movable.

Return of the loop is controlled by the thumb and in rest position the electrode should rest

inside the operating sheath, to be used with 2.9mm hysteroscopy telescope. Should work

in saline, BIPOLAR Cutting loop 21 Fr should work in saline, Cutting Electrode 21Fr,

bipolar, pointed, should work in saline, Continuous Flow Resectoscope Sheath 22 Fr.,

including connection tubes for in- and outflow, 2 LUER-lock adaptors, 22Fr, oblique beak,

fixed inner tube, with ceramic insulation, for use with working element , Continuous

Flow Resectoscope Sheath 22 Fr., for Bi-Polar, including connection tubes for in- and

outflow, 2 LUER-lock adaptors, 22Fr, oblique beak, rotating inner tube, with ceramic

insulation, for use with working element. should work in saline, Obturator, for use with

the Resectoscope sheath. QTY-01

CONTROLLED IRRIGATION AND SUCTION UNIT: FOR OPERATING

HYSTEROSCOPE AND TCRE PROCEDURE TO MAINTAIN INTRAUTERINE

CAVITY PRESSURE AS DESIRABLE BY SURGEON. PRESSURE OF IRRIGATION

TO BE MAINTAINED BETWEEN:(PRESSURE 0-200mmHg).FLOW TO BE

MAITAINED BETWEEN: (Flow 0-500ml/L.)SUCTION PRESSURE TO BE

MAINTAINED BETWEEN:Suction controlled pressure between 0-(-50kPa)Should have

foot switch for controlling various functions.consisting of: MAIN UNIT WITH DIGITAL

DESPLAY, PUMP HEAD, INDICATOR AND PRESET AND ACTUAL VALUE FOR

PRESSURE, FLOW,SUCTION WITH SENSOR ATTACHMENT FOR REUSABLE

DOME.AUTOMATIC DOME DETECTION FACILITY.Should be supplied along with

following accessories: HYSTEROSCOPE reusable Tubing Set, 2 Nos.Reusable Pressure

dome 2 Nos.Suction bottle 5.0L 2 Nos.Suction bottle cap with stand , Suction tubing set

reusable 2 Nos.,foot switch, 1 Nos,. USFDA / Eiropean CE approved. QTY-01

7. Vaginal Hysterctomy Set

Equipment Name Technical Specification

Vaginal Hyserctomy Set Galicups (M) – 2

Kidney Trays (M+ Big)- 2

Sponge holders- 3

Needle Holder 2 (1 S+1 Big)- 2

6” curved artery forceps 1 set (6 No)-6

6” curved artery forceps 1 set (6 No)-6

6” Allys forceps 1 set (6 No)-6

Babcocks 2 No.s- 2

Kockers (cd) forceps 2 Nos, straight 2-4 long cd artery (6

No)- 6

Long Allys forceps 1 set (6 No)- 6

Page 10: Technical Specifications of Obstetrics and Gynecology ......short Cleaning Adaptor and Colptomizer.USFDA / Eiropean CE approved.QTY-01 20 Macro needle holder with tungsten carbide

Annexure-A

Deavers retractor Small – 2

Kellys vaginal retractor -1

Suction Cannula – 1

Myoma Screw-1

Cross action towel clips- 8

Mohnians towel clips – 6

BP handels- 2 (22-1)

Nontoothed dissecting forceps- 2

Ntoothed dissecting forceps- 2

Mayo’s curved scissors- 2

Mayo’s straightscissors- 1

Mosquito curved artery forceps – 6 nos

Mosquito Straight artery forceps – 6 nos

Metal Urethral Catheter- 1

Bladder sound 1

Sim’s Speculum- 1

Miyas Hook

Illuminated Sims Speculum wth fibre optic attachement

wth fiber optic cable

Light Source (Halogen) 250 Watt

8. Digital High Definition Video Recording System and documentation system

Equipments Name

Technical Specification

Digital High Definition Video Recording System and documentation system

Documentation system for digital recording of still images and videos:

Pre-configured complete system for intra-operative recording of still

images and videos and for entering of patient data.Characteristics of the

system::Structured and well-arranged control menus and user

guidance.Optimized for control via the front plate of the system and via a

keyboard.Possibility of control from sterile field via camera head buttons

for recording still images and videos.Recording from clinical video

sources:Still images and videos up to FULL HD resolution Digital inputs:

Page 11: Technical Specifications of Obstetrics and Gynecology ......short Cleaning Adaptor and Colptomizer.USFDA / Eiropean CE approved.QTY-01 20 Macro needle holder with tungsten carbide

Annexure-A

DVI and HD-SDI.Analog inputs: Y/C (S-Video), RGB and Composite.

Storage of data on external USB storage medias (external hard disk or

stick), internal hard disk or a network.Quickprint function for fast print of

images.Recorded data can be edited.Possibility of cart installation with a

screen.

Technical data:Input voltage: 100-240 VAC.Input frequency: 50/60 Hz.

Power input: 1.6A (250 VAC).Power output: 25 Watt.Potential

equilization: according to DIN 42801.Potential equilization: according to

DIN 42801.Potential equilization: according to DIN 42801.Hard disk:

320GB.USB connections: USB2.0, High-Speed (1 front connection, 2

connections on the rear).Printer support: Color printer Sony UP-DR80MD

for medical applications.Networking: RJ45 10/100/1000 Ethernet.

External storage: USB-drives (sticks oder hard disks), networks.

Image formats: BMP, JPG, TIFF.Video formats: MPEG-4 (H.264).

Supported resolutions: 1920x1080, 1280x1024, 1280x720, 1024x768,

800x600, 640x480.Video signal inputs: DVI-D, HD-SDI, Y/C (S-Video),

RGB, Composite.Color systems: PAL, NTSC.Resolution images and

videos: up to 1920x1080, aspect ratio 16:9. Dimensions: 305mm x 75,

5mm x 233mm (w x h x d). Weight: 1.8 kg.Type approval: IEC 60601-1-1

, EN60601-1, EN60601-2.Classification: CE, MDD.

Type of protection: IP20. Protection class: 1/B.Medical product: yes.

Consisting of: Control unit. Compact keyboard with touchpad. DVI-D

cable.USB stick.Power cable. USFDA / Eiropean CE approved.QTY-01

Certificates: USFDA/European CE

9. Three Chip camera head & focusing coupler

Equipment Name Technical Specification

Three Chip camera Head & focusing coupler

Three Chip Camera System Specifications:Digital Three-

Chip Medical-Video-Camera – color system PAL, with

integrated Image Processing Module.

Special Features:-• Should be fully compatible to the color

systems PAL & NTSC.All the necceasry setting required

to be done during surgery in the camera should be under

sterile area.

•Should have the facility of Digital Image Processing by

means of an integrated Image Processing Module.

Multiple settings allow the user to select the preferred

level of image enhancement. -Digital contrast

enhancement -Digital anti-moiré/anti-grid filter for use

with fiberscopes.

Page 12: Technical Specifications of Obstetrics and Gynecology ......short Cleaning Adaptor and Colptomizer.USFDA / Eiropean CE approved.QTY-01 20 Macro needle holder with tungsten carbide

Annexure-A

•Should have Parfocal Optical Zoom lens system.• Should

have the additional facility of Electronic 2x zoom,

selectable in 4 steps.Integrated Still Image function

.• Digital Video output DV. For direct hook up with Lap

top

• Should have the Auto Exposure System for consistent

illumination level.

• High horizontal image resolution of more than 750 lines,

therefore even the finest variations in tissue structures are

perceivable on a high-resolution monitor.

• Automatic white balance with memory function for two

settings.

• Multilingual menu system.

• Character generator

• Composite, S-VHS and RGB compatibility

• World power supply

• 2 programmable function keys on the camera head for

control of camera functions or video printer/recorder

functions & other peripheral units.

• Adaptable to an operating microscope by means of a

special quick adapter.

• Camera head fully soakable for disinfection. USFDA /

European CE approved.QTY-01

10. Shadow less mobile side lamp for O.T

Equipment Name Technical Specification

Shadow less mobile side lamp for OT Mobile LED OT light completely spring balance with

40,000 HRS. lamp life. Inbuilt Power supply with battery

backup for emergency use. It should provide cool white

light. Light intensity should be 80,000 Lux.

Product should be European CE/ US FDA approved.

11. Ultrasound with Doppler and in-built printing and recording system

Equipment Name Technical Specification

Ultrasound with Doppler and in-built printing and recording system

Capable of processing and displaying 3D rendering of 2D

images

Page 13: Technical Specifications of Obstetrics and Gynecology ......short Cleaning Adaptor and Colptomizer.USFDA / Eiropean CE approved.QTY-01 20 Macro needle holder with tungsten carbide

Annexure-A

Broadband multifrequency Tranducers-convex

(abdominal) and TVS

Include a full complement of measurement and

calculation package

Digital Ultrasonic System

T issue Harmonic Imaging

Software package along with format for Ob & Gy

Real time dynamic ROF, RDA, DFP, DRA.

*permanent inbuilt + USB images storage

>256 frame line loop.S

Ergonomic keyboard

-back lit keyboard + direct printing, 6 STC slides for easy

adjustment

-multifunction knob + toggle for quick adjustment

Scanning method – Electronic convex / Linear /

Microconvex

Display modes – B, B/B,UB, M, B/M

Gray scale

Probe frequency 2.5 MHz – 9.0 MHz

Monitor - Zoom , Multi step

Image processing & reversing

Power supply Ac 230v + 10%, 50 Hz

Standard - main unit + convex printer + 2 probe connector

+ USB ports + VGA Port (extra transvaginal probe) +

inkjet printer

Trolley, Biopsy kit

Multidisciplinary ultrasound color Doppler with pulse

wave -spare part with ultra sound modem

-Beam former board

-Transmitter board

-Receiver board

-Digital scan converter board

-Power supply unit

-Pobes X 2

Certificates: USFDA/European CE Approved

Page 14: Technical Specifications of Obstetrics and Gynecology ......short Cleaning Adaptor and Colptomizer.USFDA / Eiropean CE approved.QTY-01 20 Macro needle holder with tungsten carbide

Annexure-A

12. Delivery Set

Equipment Name Technical Specifications

Delivery set Episiotomy Scissor - 1

Long Artery Forceps- 2

Sponge Holding – 2

Tooth Forceps Needle Holder – 1

Small tray for making delivery kits(SS)

Certificate : USFDA/European CE approved

13. Obstetric labour table

Equipment Name Technical Specifications

Obstetric Labour table Overall approx extended size 1880 mm length*9000 mm

width*800mm height

Three section-top backrest adjustable on rachet

-middle section with’ U’ cut

-Leg end section can slide completely

under main frame

Gas sparing assisted trendelenberg position

SS head side railing,SS handgrips ,rubber padded SS

lithotomy rods and SS bowl

Certification : USFDA/European

14. Spermfuge Machine

Equipment Name Technical Specifications

Spermfuge Machine The ultrasonic dissector should have following features

Shears must be capable of dissecting, cutting and

coagulating vessels upto 7 mm.

Cutting & coagulating with ultrasonic energy

Transducer for Lap and open surgery

Tissue gasping with using Parallel Jaw closure

Should be compatible with 5mm instruments

The system should have stand by mode

Page 15: Technical Specifications of Obstetrics and Gynecology ......short Cleaning Adaptor and Colptomizer.USFDA / Eiropean CE approved.QTY-01 20 Macro needle holder with tungsten carbide

Annexure-A

Power should be adjustable from 1-5

The ultrasonic dissector should have following instrument

to handle following open and laparoscopic surgery

instruments such as :

Shears of 5mm thickness and shaft length 180 mm- 390

mm

Shears of 5mm thickness and shaft length 320mm-

380mm

Shears of 5mm thickness and shaft length 90mm-170mm

Ball probe

Hook probe of 5 mm thickness and shaft length 300mm -

350mm

Cable for ultrasonic transducer -1no.

ultrasonic transducer -1no.

Foot pedal for ultrasonic generator-1no.

Cerification :USFDA/European CE Apporoved

15. IVF Instrument

Equipment Name Technical Specifications

IVF Instruments 1- Air Jacked CO2 Incubator with IR Sensor Special : Capacity 150 Ltrs Four Inner Glass Doors to saveloss of CO2 gas System to achieve Maximum relative Humidity & Dry Inner Walls Stainless Steel interior Standard Inner 4 Door Fan free Interior – Prevents building up of Germs Perforated shelves for uniform heat distribution Air Jacketed LCD Display Hot Air auto sterilization (180 C) 2. Integrated Laminar flow hood with heated table the integrated Laminar system is Often built in vertical laminar flow cabinet. The utensils are warmed by heating A part of the table by means of special heat Exchanger controlled

Page 16: Technical Specifications of Obstetrics and Gynecology ......short Cleaning Adaptor and Colptomizer.USFDA / Eiropean CE approved.QTY-01 20 Macro needle holder with tungsten carbide

Annexure-A

3. Stereo Zoom Microscope Zooming Body With Diopter adjuster, rubber eye shield, reticle lead Diascopic stand C-Ds Diascopic stand S with reflecting lead Power Cord BE 6 V- 20 W halogen lamp with reflector CDs- 33 mm NCB 11 filter 4. Rocket of London Craft Ovum Aspiration pump with Foot operated Switch with Vacuum gauge & regulator Specification Low Pressure: 0-550mm Hg= 0-666 Mbar Volume of overflow vessel: 2Χ100 Connecting for aspiration tubing 2 to 4mm Power Supply : 230 V Power Frequency : 50/60Hz Power Point : 50 VA Nominal Current: 220 mA Power Line isolation : 0.315 5. Indian Pressurizing Module Pressuring Module is used to create the positive inside culture lab Tech specification Pressure: 250 CFM Construction: Liminated board Filtration: 0.3 μ Efficiency : 99.97% Construction : S.S body 6. Digital Cell Transporter with steel body heating Block for heating of test tubes in removable, autoclave anodized aluminum blocks for maintenance of required temp. Digital display & control of temp. with + 0.2C accuracy from ambient to 11 c, 3 anodized blocks of 75Χ 50Χ50 mm to accommodate 12Χ13 ml and 12Χ6 ml tubes . 7. Petri –Plate Prewarmer with Steel Body for preheating of slides, Petri dishes, pipettes etc. of desired temp. Anodized aluminum heating plate (Approx dimension 300 Χ100 Χ40mm) Temp range: from ambient to 1000C with accuracy of + 0.20C 8. Microscope Stage Warmer with Steel Body for maintaining the temp of critical biological specimens at desired level during microscopic observation. An amodized plate of 10 mm thickness with 50 mm central grove with glass & 175 Χ 155 mm dimension which should fit on most miscroscope stages. Temp range ambient to 550C with accuracy of + 02.0C 9. Coda Aero Air purification unit Max Air flow: 530 CFM Effective coverage: up to 350 sq feet Application : more than 200 sq feet

Page 17: Technical Specifications of Obstetrics and Gynecology ......short Cleaning Adaptor and Colptomizer.USFDA / Eiropean CE approved.QTY-01 20 Macro needle holder with tungsten carbide

Annexure-A

10. CL2200 Freeze control from CRYOLOGIC PTY LTD. Controlled Rate Preprogrammed Freezer for freezing Sperms, Oocytes, Embros complete with Preprogrammed Controller LN2 bath (Capacity 1.5 L) Carry case Standard Cryochamber to freeze 23 Straws of 0.5 cc & 46 Straws of 0.25 cc Specification Temprature Controller Controlled Temperature range: 200C & 430C Temperature warning ( ED or Sound): 1.50 C deviation Temperature display : Digital. LCD, 0.10C resolution Timing : Digital, Quartz Crysteal Internal Programms: Maximum number of fixed programs: 8 nos Others: continuous LN2 consumption/less than 1 Liter per hour Power consumption/less than 60 Watt Sperm/Embryo freezing

a) TA- 26CRYOCAN, IBP make, 25.7 ltrCapicity with rubber ring & pad b) B) BA- 11 CCRYOCAN, IBP make 10.7 ltr capacity with canisters (For

large capacity) c) BA- 3 11 CCRYOCAN, IBP make 10.7 ltr capacity with canisters (For

Carrying frozen specimen 12. Doctors friendly Software Reprosoft SOFTWARE is designed to cover important areas in a fertility clinic It can be used in a LAN environment with multiple users. Degrees of access to individual user 13. Inverted Research Microsoft: with Hoffman Modulation Contrast Inverted Microsoft basic unit (100-240 V) 12 V- 100W Lamp house- 2 BL with pre-centered 100 W helogen Main Body, Side port Coaxial Coarse/ fine focus W /tension adjustment, Light distribution T-DH 100W Dia illumination Pillar supplied with detachable condenser carrier, diffuser, 45 mm filter slots. LamphouserERmote Cable Helogen Lamp- 12 -100 W LL Power Cord BE 25 Inclination ube D 25 Inclination angle, Turret selector incorporated B/2.5*C/0 HEPA filter Cerification :USFDA/European CE Apporoved

16. LLETZ

Equipment Name Technical Specification

LLETZ Non conductive instruments can be used with any LLETZ/LLEP

Page 18: Technical Specifications of Obstetrics and Gynecology ......short Cleaning Adaptor and Colptomizer.USFDA / Eiropean CE approved.QTY-01 20 Macro needle holder with tungsten carbide

Annexure-A

system Highly reliable,autoclavable Built in smoke tubes on all ‘duckbil’ specula. Electrosurgery unit features 5 treatment modes –purecut ,blend,coagulation,fulgrationand bipolar powerful 120 watts purecut and 80watt coagulation modes reduce electrode draw and assure quicker coagulation Square LLETZ electrodes 1cm*0.4cm sterile,disposable,box of 5;shaft :13cm(5”) Loop LLETZ electrodes 0.5*0.5cm , sterile,disposable,box of 5;shaft :13cm(5”) Loop LLETZ electrodes 1*0.5 cm, sterile,disposable,box of 5;shaft :13cm(5”) Loop LLETZ electrodes 1*1 cm, sterile,disposable,box of 5;shaft :13cm(5”) Loop LLETZ electrodes 1.5*0.5 cm, sterile,disposable,box of 5;shaft :13cm(5”) Loop LLETZ electrodes1.5*1 cm, sterile,disposable,box of 5;shaft :13cm(5”) Loop LLETZ electrodes1.5 *1.5cm, sterile,disposable,box of 5;shaft :13cm(5”) Loop LLETZ electrodes2*1cm, sterile,disposable,box of 5;shaft :13cm(5”) Ball LLETZ electrodes,3mm dia, sterile,disposable,box of 5;shaft :13cm(5”) Ball LLETZ electrodes,5mm dia, sterile,disposable,box of 5;shaft :13cm(5”) Cerification :USFDA/European CE Apporoved