tc convencional vs tc helicoidal
TRANSCRIPT
![Page 1: TC CONVENCIONAL VS TC HELICOIDAL](https://reader036.vdocuments.mx/reader036/viewer/2022081417/55d59665bb61ebc8138b45c2/html5/thumbnails/1.jpg)
TC CONVENCIONAL VRS TC HELICOIDAL
MARLON MIZAEL GARCIA HERNANDEZRESIDENTE IDT
![Page 2: TC CONVENCIONAL VS TC HELICOIDAL](https://reader036.vdocuments.mx/reader036/viewer/2022081417/55d59665bb61ebc8138b45c2/html5/thumbnails/2.jpg)
TOMOGRAFIA
Tomos=corte
Grafos= escritura, imagen, gráfico
![Page 3: TC CONVENCIONAL VS TC HELICOIDAL](https://reader036.vdocuments.mx/reader036/viewer/2022081417/55d59665bb61ebc8138b45c2/html5/thumbnails/3.jpg)
CONCEPTOS
PIXEL Y VOXEL
![Page 4: TC CONVENCIONAL VS TC HELICOIDAL](https://reader036.vdocuments.mx/reader036/viewer/2022081417/55d59665bb61ebc8138b45c2/html5/thumbnails/4.jpg)
CONCEPTOS
GROSOR DE CORTE: DETERMINA EL VOLUMEN DEL VOXEL, LO QUE ES LO MISMO LA ANCHURA DEL CORTE.
![Page 5: TC CONVENCIONAL VS TC HELICOIDAL](https://reader036.vdocuments.mx/reader036/viewer/2022081417/55d59665bb61ebc8138b45c2/html5/thumbnails/5.jpg)
CONCEPTOS
RESOLUCION ESPACIAL: capacidad para detectar y representar de forma gráfica, estructuras anatómicas o lesiones pequeñas que se encuentran muy cercanas entre si.
la mínima distancia que debe haber entre dos puntos de un objeto, para poder identificarlos como imágenes independientes
![Page 6: TC CONVENCIONAL VS TC HELICOIDAL](https://reader036.vdocuments.mx/reader036/viewer/2022081417/55d59665bb61ebc8138b45c2/html5/thumbnails/6.jpg)
ESPESOR DE CORTE O COLIMACION
Los cortes finos aumentan la resolución espacial y disminuyen el efecto de volumen parcial. Sin embargo aumentan el ruido y la cantidad de radiación.
![Page 7: TC CONVENCIONAL VS TC HELICOIDAL](https://reader036.vdocuments.mx/reader036/viewer/2022081417/55d59665bb61ebc8138b45c2/html5/thumbnails/7.jpg)
CONCEPTOS
Campo de visión (F O V = Field of view)
Determina el diametro del corte y depende de la zona de estudio.
Cuanto más amplio sea el FOV más pequeña se verá la imagen en la pantalla que al ampliarla perderá resolución.
![Page 8: TC CONVENCIONAL VS TC HELICOIDAL](https://reader036.vdocuments.mx/reader036/viewer/2022081417/55d59665bb61ebc8138b45c2/html5/thumbnails/8.jpg)
PITCH
pitch = Movimiento de la mesa en mm x giro (segundo) / Grosor de corte
![Page 9: TC CONVENCIONAL VS TC HELICOIDAL](https://reader036.vdocuments.mx/reader036/viewer/2022081417/55d59665bb61ebc8138b45c2/html5/thumbnails/9.jpg)
PITCH
El pitch determina la separación de las espirales, de tal manera que a 10mm de desplazamiento de la mesa por segundo, si cada giro dura un segundo, y el grosor de corte fuese de 10mm correspondería un pitch 1
pitch = (10mm x 1s)/5 mm = 2 ;es decir el índice de pitch sería de 2
![Page 10: TC CONVENCIONAL VS TC HELICOIDAL](https://reader036.vdocuments.mx/reader036/viewer/2022081417/55d59665bb61ebc8138b45c2/html5/thumbnails/10.jpg)
PITCH
Cuanto mayor es el valor del pitch, más estiradas estarían las espirales, mayor sería su cobertura, menor la radiación del paciente, pero menor sería la calidad de las imágenes obtenidas
![Page 11: TC CONVENCIONAL VS TC HELICOIDAL](https://reader036.vdocuments.mx/reader036/viewer/2022081417/55d59665bb61ebc8138b45c2/html5/thumbnails/11.jpg)
UH
Expresa de forma numérica, por cada centímetro y para cada tejido que atraviesa, la atenuación en la intensidad que experimenta el haz de rayos, desde que sale del tubo, hasta que llega atenuado a los detectores.
![Page 12: TC CONVENCIONAL VS TC HELICOIDAL](https://reader036.vdocuments.mx/reader036/viewer/2022081417/55d59665bb61ebc8138b45c2/html5/thumbnails/12.jpg)
ELEMENTO Y UH
Metal Hueso Cortical Hueso Esponjoso Contrastes Yodados Hematomas Micro calcificaciones Partes Blandas Agua Lípidos Aire
+ 1100 +4000 + 1000 + 1800 + 400 + 700 + 100 + 250 + 75 + 50 + 60 + 85 + 30 + 50 0 - 10 - 120 - 1000
![Page 13: TC CONVENCIONAL VS TC HELICOIDAL](https://reader036.vdocuments.mx/reader036/viewer/2022081417/55d59665bb61ebc8138b45c2/html5/thumbnails/13.jpg)
![Page 14: TC CONVENCIONAL VS TC HELICOIDAL](https://reader036.vdocuments.mx/reader036/viewer/2022081417/55d59665bb61ebc8138b45c2/html5/thumbnails/14.jpg)
![Page 15: TC CONVENCIONAL VS TC HELICOIDAL](https://reader036.vdocuments.mx/reader036/viewer/2022081417/55d59665bb61ebc8138b45c2/html5/thumbnails/15.jpg)
TC CONVENCIONAL VRS TC HELICOIDAL
EN TC CONVENCIONAL SE ADQUIERE SECUENCIALMENTE UNA SERIE DE IMÁGENES SEPARADAS POR ESPACIOS IGUALES.
![Page 16: TC CONVENCIONAL VS TC HELICOIDAL](https://reader036.vdocuments.mx/reader036/viewer/2022081417/55d59665bb61ebc8138b45c2/html5/thumbnails/16.jpg)
TC CONVECIONAL CARACTERISTICAS
El grosor del corte y el intervalo de solapamiento se seleccionan al comienzo del estudio.
Los datos de cada nivel de imagen se almacenan por separado.
la pequeña pausa entre las secciones permite al paciente respirar sin causar artefactos respiratorios importantes
![Page 17: TC CONVENCIONAL VS TC HELICOIDAL](https://reader036.vdocuments.mx/reader036/viewer/2022081417/55d59665bb61ebc8138b45c2/html5/thumbnails/17.jpg)
puede durar varios minutos, según la región corporal y el tamaño del paciente.
En TC convencional el tubo de rayos x, recibe la energía para una rotación.
![Page 18: TC CONVENCIONAL VS TC HELICOIDAL](https://reader036.vdocuments.mx/reader036/viewer/2022081417/55d59665bb61ebc8138b45c2/html5/thumbnails/18.jpg)
CARACERISTICAS DE LA TC HELICOIDAL CON RESPECTO A LA TC COVENCIONAL.
Disminución de artefactos por movimiento, porque elimina los errores de registro, debido a la respiración.
![Page 19: TC CONVENCIONAL VS TC HELICOIDAL](https://reader036.vdocuments.mx/reader036/viewer/2022081417/55d59665bb61ebc8138b45c2/html5/thumbnails/19.jpg)
Mejor detección de lesiones, no hay pérdida de información, por no existir intervalos de corte.
![Page 20: TC CONVENCIONAL VS TC HELICOIDAL](https://reader036.vdocuments.mx/reader036/viewer/2022081417/55d59665bb61ebc8138b45c2/html5/thumbnails/20.jpg)
![Page 21: TC CONVENCIONAL VS TC HELICOIDAL](https://reader036.vdocuments.mx/reader036/viewer/2022081417/55d59665bb61ebc8138b45c2/html5/thumbnails/21.jpg)
![Page 22: TC CONVENCIONAL VS TC HELICOIDAL](https://reader036.vdocuments.mx/reader036/viewer/2022081417/55d59665bb61ebc8138b45c2/html5/thumbnails/22.jpg)
Mejor resultado diagnóstico por contraste intravenoso, por su bajo tiempo de barrido, podemos seleccionar el pico de actividad del contraste.
Menor cantidad de contraste.
Realización de Angio Tc.
![Page 23: TC CONVENCIONAL VS TC HELICOIDAL](https://reader036.vdocuments.mx/reader036/viewer/2022081417/55d59665bb61ebc8138b45c2/html5/thumbnails/23.jpg)
Inconvenientes de Tc helicoidal
Ligera reducción de la resolución axial, por los algoritmos de reconstrucción.
Ligero aumento del ruido
Aumento del tiempo de procesado de datos, al haber mayor información y más datos.
![Page 24: TC CONVENCIONAL VS TC HELICOIDAL](https://reader036.vdocuments.mx/reader036/viewer/2022081417/55d59665bb61ebc8138b45c2/html5/thumbnails/24.jpg)
BIBLIOGRAFIA
Hofer, Matias. Manual de Tomografía computada. 4ta edición. 6 – 10.
Bushong Stewart. Manual de Radiología para técnicos. 6ta edición. 395 – 407.