syringe pump manufacturer | chemyx - fusion series · a syringe pump is a small,...

57
FUSION SERIES USER MANUAL

Upload: others

Post on 02-Aug-2020

7 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

FUSION SERIESUSER MANUAL

Page 2: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

2

EU Declaration of Conformity (DoC) 6

Manual Description 7

General Information 7

Safety Information 7

Warranty and Repair Information 9

Limited Warranty 9

Repair Facilities and Parts 9

Dead Pixel Policy 8

Serial Numbers 8

Calibrations 8

Product Overview 9

The Fusion Series Syringe Pumps 9

Principle of Operation 9

Features 11

Pump 11

Controls 14

Power and Data Connectors 15

Audible Alerts 15

Stall Detection 15

Specifications 16

Pump Setup 17

PoweringOn/OffthePump 17

Loading a Syringe 18

Fusion 100 18

Fusion 200 19

Fusion 400 20

Page 3: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

3

Adjusting the Safety Collar 20

Fusion 100 and 200 21

Fusion 400 21

Pump Operation 22

Selecting Items 22

Navigating Between Screens 23

Entering Values 23

Entering Numbers 23

Entering Letters 23

Mode Selection Screen 24

Basic Mode 24

Syringe Selection 25

Target Transfer Volume 26

Infusion/Withdrawal (Fusion 200 only) 26

Flow Rate 27

Start-Time Delay 27

Options 28

Flow-Rate Units 28

Saving Run Parameters 29

Priming/Bolus Rate 29

Enable Volume Lock 30

Multi-Step Mode 31

Setup 31

Syringe Selection 31

Total Steps 31

Loop All 32

Page 4: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

4

Sub-Loops 32

Flow-Rate Units 33

Step Parameters 33

Target Transfer Volume 33

Infusion/Withdrawal (Fusion 200 only) 34

Flow Rate 34

Static Flow Rate 35

Variable Flow Rate 35

Start-Time Delay 35

Step Loop 36

Running the Pump 37

Basic Mode System Status 38

Multi-Step Mode System Status 39

System Load Run 42

System Settings 43

Touchscreen Toggle 43

Baud Rate 43

Power Settings 44

Motor Power` 44

LCDPower-OffTime 44

Pump Control by Computer 46

Cable Requirements 46

Connecting 47

Settings 47

Pump Commands 47

Chemyx Pump Controller Programs 54

Page 5: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

5

Pump Maintenance 54

Lubricating the Pump 54

Replacing the Fuse 56

Pump Accessories 57

Footswitch 57

Connecting the Footswitch 57

Using the Footswitch 57

10-Channel Syringe Rack (Fusion 200 only) 58

Attaching the Rack 59

Loading Syringes 61

Appendices 62

Appendix A: Syringe Volume/Diameter Reference Table 62

Appendix B: Available Parts 56

Fusion Series Syringe Pumps 56

Options and Accessories 56

Stainless-Steel Syringes 56

Appendix C: Contact Information 57

INTELLECTUAL PROPERTYAllIntellectualProperties,asdefinedbelow,ownedbyorwhichisotherwisethepropertyofChemyxInc.oritssuppliersrelatingtotheChemyxsyringepumps,includingbutnotlimitedto,accessories,

partsorsoftwarerelatingthereto(ChemyxSyringePumps),areproprietarytofederalandstatelaws,andinternationaltreatyprovisions.IntellectualPropertyincludesbutisnotlimitedto,inventions

(patentableorunpatentable),patents,tradesecrets,copyrights,software,firmware,computerprograms,andrelateddocumentationandotherworksofauthorship.Moreover,youagreethatyouwill

not, and will not attempt to, modify, prepare derivative works of, reverse engineer, disassemble the Chemyx syringe pumps, decompile or otherwise attempt to create source code from the related

software/firmware.NotitletoorownershipinIntellectualPropertyistransferredtoyou.AllapplicablerightsoftheIntellectualPropertyshallremainwithChemyxanditssuppliers.

Page 6: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

6

Declaration of Conformity

ThePDFversionofthelatestDoCcanbedownloadedatwww.chemyx.com/support

Page 7: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

7

General Description

PleasereadthefollowingsafetyprecautionstoensurepersonalsafetyandoperationallongevityoftheChemyxsyringepump.Chemyx is not responsible for the equipment if used in amanner not specified by themanufacturer; warranty coverageprovidedbytheequipmentmaybedroppedasaresult.

CHEMYX PRODUCTS ARE NOT APPROVED FOR CLINICAL USE ON HUMANS.

Chemyx isnotresponsible for theuseofpowersuppliesoutsidethestatedelectricalspecificationsor failuretoswitchthepowerconverterfrom220Vto110Vwhileinthe220Venvironmentorviceversa.

Theproductshouldbeproperlygrounded.

Use Proper Power Supply

Ground Product

Safety Information

WarrantycoveragewillbelostifthepumpisopenedwithoutauthorizationfromChemyx.Donottouchanyelectricalconnectorsonorintheproduct.

Even though the pump can operate at extremely fast speeds, the user must determine the properflowrateforanygivenapplication.Forinstances,pumpingat90mL/minusinga20-gaugeneedlewillcausestallsand/orpotentialburstingofthesyringe.Chemyxisnotresponsibleforanydamagethatmightresultfromsituationssimilartotheexampleabove.

Do Not Open The Pump

Do Not Operate With Suspected Failures

Page 8: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

8

Donotplacefingersbetweenthepusherblockandtheendblockwhilethepumpisrunning.

Readalllabelsontheproducttoensureproperusage.

The user is responsible for wetting ground glass syringes and setting and tightening the safety collar/barappropriately.

Pinch Hazard

Observe all Warning labels on Product

Chemyx is not Responsible for Syringe Damage

Page 9: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

9

You must contact Chemyx (call +1 281-277-5499 or visit www.chemyx.com/contact-us)beforereturningaproduct.ChemyxwillissueaReturnAuthorization(RA)numbertoyou.

Return products to:

ChemyxInc.

10905CashRoad|Stafford,TX77477USA.

Warranty and Repair InformationWarranty InformationChemyxprovidesatwo-year limitedwarranty(1) first-year for parts and labor (2) second-year for parts only fromtheshipmentdateforitspumpsagainstdefectsinmaterialsandworkmanship.Chemyxwillrepairanyproductthatprovesdefectiveduringitsstatedwarrantyperiod.

The foregoing warranty will not apply to damage resulting from:

Improperorinadequatemaintenanceoroperation.

Unauthorizedmodificationormisuseoftheproduct.

Operationoutsidetheelectricalspecificationsfortheproduct.

Operationoutsidethetemperaturespecificationsfortheproduct.

User-inducedinternalandexternalcontaminationsoftheinstrument.

Failuretousepropersurgeprotection.

Improper product return, packaging, and shipping

Removing serial number from syringe pump

Page 10: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

10

Repair Facilities and PartsChemyxcanrepairanysyringepumpwithout major damage.YoumustcontactChemyxcall+1281-277-5499oremailusatinfo@chemyx.combeforeshippingaproductforrepair.ChemyxwillissueaReturnAuthorization(RA)numbertoyou.

Return products to:

Chemyx Returns

10905CashRoad|Stafford,TX77477USA

Dead Pixel PolicyDuringtheLCDMonitormanufacturingprocess,itisnotuncommonforoneormorepixelstobecomefixedinanunchangingstate.Thevisibleresultisafixedpixelthatappearsasanextremelytiny,darkorbrightdot.Inalmosteverycase,thesefixedpixelsarehardtoseeanddonotdetractfromdisplayqualityorusability.Adisplaywiththreetosevenbrightordarkdotsisconsiderednormalandwithinindustrystandards.Ifyourscreendisplaysmorethansevendeadpixelsduringthewarrantyperiod, your systemwill qualify forwarranty replacement. Please see the LimitedWarranty section for details on how toactivateawarrantyclaim.

Serial NumbersTheserialnumberislocatedontheback,topleftcornerorcenterofthepumpunderasmallbarcode.Removaloftheserialnumberlabelwillvoidthewarranty.

CalibrationsChemyxpumpsarepre-calibrateduponarrivaltoyoursite.Allcalibratedparametersarewithinstatedaccuracyandprecisionspecifications of the pump. Although the pump is highly accurate, syringes of differentmaterials can attribute significantvolume errors. Typically, these errors can range from 1% for glass syringes to 5% for plastic syringes. Chemyx is notresponsibleforerrorsgeneratedfromsyringes.

Page 11: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

11

Product Overview

ChemyxFusionSeriessyringepumpsaredesignedtohandleeverytypeofhigh-precisiondosingapplication.Theunits’flexibilitywith syringe types and sizes and an advanced software interface make this line of high-precision syringe pumps integrable intoanyoperationorlaboratoryworkflow.Thoroughlydurablebothinsideandout,theChemyxFusionsyringepumpsaredesignedtoprovideyearsofconstantandreliableservice.

Asyringepumpisasmall,positive-displacementpumpusedtograduallytransferprecisevolumesoffluid.AllChemyxFusion-seriessyringepumpsaredrivenviaasteppermotor.Thissteppermotorpreciselyturnsaleadscrewthatisthreadedthroughapusherblock,whichcausesthepusherblocktomove.Whenthepumpisininfusionmode,thepusherblockpushesagainsttheplungerofasecuredsyringe,causingthefluidtobeejectedatanaccurateandpreciserate.

Ifthepumpiscapableofwithdrawal,theplungerofthesyringeisheldbybracketsonthepusherblock.Whenthesteppermotorturnsintheoppositedirection,thepusherblockmovessuchthatthesyringeplungerispulled,thusdrawingfluidintothesyringe.

FortheFusionSeriessyringepumps,thepusherblockmovestotherightforinfusionandtotheleftforwithdrawal.

Principle of Operation

Page 12: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

12+1281-277-5499|www.chemyx.com

AllFusionSeriespumpshaveasingleleadscrewthroughwhichthepusherblockisthreaded.Thepusherblockmoveswhentheleadscrewturns.Thetwoguiderodskeepthepusherblockhorizontalandperfectlyperpendiculartotheleadscrew.Theblock-release button disengages the pusher block from the lead screw, which allows the pusher block to be easily moved to a newpositionaslongasthebuttonisheld.Releasingthebuttonwilllockthepusherblockbackinplace.

Syringesareplacedonthesyringe-holderblockinthev-shapedgrooves,orchannels,andheldsecurelyusingsyringeclamps.TheFusion100and200bothhavetwosyringechannels,whiletheFusion400hasfoursmallersyringechannels.

The safety collar of the Fusion 100 and 200 helps protect the integrity of syringes by keeping the pusher block from pushing onacompletelydepressedsyringe.

Controlled Elements

Block Release Button Syringe Holder BlockSafety Collar

Lead Screw Guide Rods Pusher Block Syringe Clamp Syringe Channels

Fusion 100 pump components

Page 13: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

13+1281-277-5499|www.chemyx.com

TheFusionSeriesusesabright,4.25”,resistive,LCDtouchscreenwithanintuitivegraphicaluserinterface(GUI)forsettingupdetailedpumprunmethods.ButtonsandentryfieldswithintheGUIareallselectableusingthetouchscreen.Asaresistive,LCDtouchscreen,theinterfacecanbeoperatedwithanyobject,notjustthetouchofafingertip.Selectingelementsbytouchonlyrequiresasmallamountofpressure;however,pleasedonotusesharpobjectsastheymaydamagethescreen.

The Fusion series also includes physical navigation controls, allowing users to manually maneuver between selectable GUI elementswithouttouchingthescreen.InputtingnumbersandletterswithintheGUIisdoneusingthephysicalnumerickeypad.Thephysicalpumpcontrolkeysareusedtostart,pause,andstoppumpactivity.

The power LED and pump-control LEDs allow users to easily discern whether the pump is on and whether it is currently running, paused,orstopped.

Navigation Keypad

Front Panel of Pump

Controls

Power LED Pump Controls LCD Touchscreen Numeric Keypad

Page 14: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

14+1281-277-5499|www.chemyx.com

In addition to having all of the same features and components of the Fusion 100, the Fusion 200 is capable of both infusion andwithdrawal.Becauseof thisdifference, theFusion200uses adjustable retaining brackets on the pusher block and the syringe-holder block that keep the syringe in place during bothinfusionandwithdrawal.

In addition to having all of the same features and components of the Fusion 100, the Fusion 200 is capable of both infusion andwithdrawal. Because of this difference, the Fusion 200uses adjustable retaining brackets on the pusher block and the syringe-holder block that keep the syringe in place during bothinfusionandwithdrawal.

Bracket Release Knob

Bracket Release Knob

Flange Retaining Bracket

Flange Retaining Bracket

Plunger Retaining Bracket

Plunger Retaining Bracket

Bracket Release Knob

Bracket Release Knob

Fusion 200 Specific Components

Fusion 400 Specific Components

Page 15: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

15

Specifications

Specification Fusion 100 Fusion 200 Fusion 400

Movement Modes Infusion Infusion/

Withdrawal Infusion

Syringe Sizes W(Min/Max) 0.5μL/60mL 0.5μL/60mL 0.5mL/10mL

Flow Rate Range (Smallest Syringe)

0.0001to1.5679μL/min

0.0001to1.5679μL/min

2 to 10.536mL/min

Flow Rate Range (Largest Syringe)

0.0002 to 157 mL/min

0.0002 to 85 mL/min

4.83pL/minto2.1mL/min

Drive Mecha-nism

1.8°steppermotorwith1/32 microstepping

1.8°steppermotorwith1/32 microstepping

1.8°steppermotorwith1/32 microstepping

Step Resolution 0.098μm/μstep 0.046μm/μstep 0.016μm/μstep

Nominal Linear Force 16 kg (35 lbs) 23 kg (50 lbs) 23 kg (50 lbs)

Accuracy ±<0.04%error(99.96%) ±<0.04%error(99.96%) ±<0.04%error(99.96%)

Reproducibility ±<0.05%deviation(99.95%) ±<0.05%deviation(99.95%) ±<0.05%deviation(99.95%)

Operation Temp 4°Cto40°C(40°Fto104°F)

4°Cto40°C(40°Fto104°F)

4°Cto40°C(40°Fto104°F)

Storage Temp -10°Cto70°C(14°Fto158°F)

-10°Cto70°C(14°Fto158°F)

-10°Cto70°C(14°Fto158°F)

Page 16: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

16

Specifications (Continued)

Specification Fusion 100 Fusion 200 Fusion 400

Relative Humidity 20%-80% 20%-80% 20%-80%

Data Ports RS-232 (DB9 and USB-B), TTL (USB-B)

RS-232 (DB9 and USB-B), TTL (USB-B)

RS-232 (DB9 and USB-B), TTL (USB-B)

Height 11cm(4.4in.) 12cm(4.7in.) 13cm(5.1in.)

Width (Length) 24cm(9.5in.) 24cm(9.5in.) 24cm(9.5in.)

Depth 17cm(6.8in.) 17cm(6.8in.) 17cm(6.8in.)

Weight 3 kg (7 lb) 3 kg (7 lb) 3 kg (7 lb)

Screen 4.25”diagonal3.7”X2.1”

4.25”diagonal3.7”X2.1”

4.25”diagonal3.7”X2.1”

Page 17: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

17

Pump SetupToprovidepowertothepump,anIEC60320C13cableshouldbepluggedintothepowersocketonthebackofthepump.Thecableshouldbepluggedintoanappropriatepowersource(110Vor220V)withpropergrounding.Adjustthevoltage-converterswitchonthebackofthepumpsuchthatthevaluedisplayedontheswitchcorrespondswiththevoltageofthepowersupply.

Thepowerswitchforthepumpislocatedontheback,abovethepowersocket.Uponflippingtheswitchtothe“on”position,theblueLEDpowerlightonthefrontofthepumpwillimmediatelyturnon,alongwiththeredLEDlightontheStopkey.Inafewseconds,abeepwillsound,andtheLCDscreenwilldisplayaninitialstartupscreen.

The startup screen displays the version of the firmwarecurrentlyinstalledonthepumpatthetopofthescreen.Thestartupscreenwilldisplayforapproximatelyfivesecondsasindicatedbythecountdownontherightofthestartupscreen.

After the startup screen, the LCD screen will display the last modethepumpwasin.

Powering On/Off the Pump

Page 18: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

18

The Fusion series can accommodate up to two syringes in the Fusion 100 and 200 models, while the Fusion 400 can house up tofoursyringes.TheFusion200canbeexpandedto10channelsusingtheChemyx10-ChannelSyringeRackaccessory.Thevolumeofthesyringescanvaryfromaminimumof0.5μLuptoamaximumof60mL(10mLmaximumfortheFusion400).Syringesofanymaterial(e.g.,glass,plastic,stainlesssteel,etc.)canbeused;however,someusecasesmayrequiresyringesthataremoreaccurateanddurable.

1. If necessary, move and tighten the infusion safety collar (seeAdjustingtheSafetyCollar).

2. While pressing in the lock-release button, slide the pusher blockallthewaytotheleft.

3. Pull up on the spring-loaded syringe clamp and place the syringe in one of the channels of the syringe-holder block.Ensurethattheflangeofthesyringebarrelisflushagainsttheedgeofthesyringe-holderblock.Slowlylowerthesyringeclamptosecurethesyringeinplace.

Loading a Syringe

Fusion 100

2 3

4

1

4. Press the lock-release button and slide the pusher block totherightuntilitisflushagainstthesyringeplunger.

Page 19: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

19

1. If necessary, move and tighten the infusion safety collar (seeAdjustingtheSafetyCollar).

2. While pressing in the block release button, slide the pusherblockallthewaytotheleft.

3. Loosenthescrewknobsfortheflangeretainingbracketandadjustthebracketsuchthattheflangeofthesyringebarrelwillfitinbetween.Donottightenthebracket,yet.

4. Pull up on the spring-loaded syringe clamp and place the syringeinoneofthechannelsoftheholderblock.Ensurethattheflangeofthesyringebarrel isflushagainsttheedgeoftheholderblock.Slowlylowerthesyringeclamptosecurethesyringeinplace.

Fusion 200

2 31 4

5

7

6

8

5. Loosen the bracket clamp knob for the plunger retaining bracket and adjust the gap such that the plunger cap of thesyringemayfit.

6. Press the block release button and slide the pusher block to the rightuntil it isflushagainst thesyringeplunger.The plunger cap for the syringe should fit in the gapbetweenthebracketandthepusherblock.

7. Tighten the bracket clamp knob for the plunger retaining brackettosecurethesyringeplungerinplace.

8. Movetheflangeretainingbracketflushagainsttheflangeof the syringe barrel and tighten the bracket clamp knobs tosecurethesyringebarrelflangeinplace.

Page 20: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

20

1. If necessary, move and tighten the infusion safety collar (seeAdjustingtheSafetyCollar).

2. While pressing in the lock-release button, slide the pusher blockallthewaytotheleft.

3. Pull up on the spring-loaded syringe clamp and place the syringe in one of the channels of the syringe-holder block.Ensurethattheflangeofthesyringebarrelisflushagainsttheedgeofthesyringe-holderblock.Slowlylowerthesyringeclamptosecurethesyringeinplace.

4. Press the lock-release button and slide the pusher block totherightuntilitisflushagainstthesyringeplunger.

Fusion 400

2 1 3

5 6 7 8

5. Loosen the bracket clamp knob for the plunger retaining bracket and adjust the gap such that the plunger cap of thesyringemayfit.

6. Press the lock-release button and slide the pusher block to the rightuntil it isflushagainst thesyringeplunger.The plunger cap for the syringe should fit in the gapbetweenthebracketandthepusherblock.

7. Tighten the bracket clamp knob for the plunger retaining brackettosecurethesyringeplungerinplace.

8. Movetheflangeretainingbracketflushagainsttheflangeof the syringe barrel and tighten the bracket clamp knobs tosecurethesyringebarrelflangeinplace.

Page 21: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

21

TheFusionseriesofpumpscangenerateasignificantamountoflinearforce.Whilethisisextremelyusefulfordispensingviscousfluidsathigherflowrates,thehigherlinearforceoftheFusionseriescanalsopotentiallydamagethesyringesinuse.Onesourceofpotentialdamageofasyringemayoccurwhentheplungerofthesyringeiscompletedepressed,andthepusherblockisstillpushingagainstthesyringe.Dependingonthematerialofthesyringe(i.e.,glassorplastic)andthethicknessoftheplunger,thepressurefromthepusherblockcouldbend,distort,orjustbreakthesyringe.

To protect against this potential problem, Fusion syringe pumps use safety collars (Fusion 100 and 200) and safety bars (Fusion400)tophysicallystopthepusherblockbeforeitcancompletelydepressthesyringeplunger.Whenapusherblockhits the safety mechanism and can no longer move forward, the pump will stop running, emit a long beep, and enter pump-stalledmode.

Adjusting the Safety Nut

Adjusting the safety collar requires the use of the hex key, conveniently stored on the back left of the pump, right above the power switch. Loosening the hex screw on the safetycollar allows for the collar to be slid freely along the guide rod and then securely tightened to set the position of the safety collar.Thepositionofthesafetycollarshouldbesetslightlyfurther from the syringe-block holder than the top of syringe plungerwhenfullydepressed.

The Fusion 400 uses a safety bar instead of a collar to limit thedistance thepusherblockmay travel.By loosening thesafety bar clamp knob on top of the pusher block, the safety barcanbeslidleftorright.Thesafetybarshouldbesetsuchthat the length of the bar exposed on the right side of the pusher block is longer than the distance from the top of a fully depressed syringe plunger to the edge of the syringe-holderblock.Oncethesafetybarhasbeensettotheproperlength, tightening the safety bar clamp knob will keep the barfrommoving.

Fusion 100 and 200

Fusion 400

Caution:On the Fusion 200, during withdrawal, a syringe plunger can potentially be pulled out from the syringe, which could cause the contents of the syringe to leak out. Users should be especially aware of the length of the plunger and the point where the withdrawing pusher block will exceed that length. There are no safety collars on the Fusion 200 to prevent too much withdrawal.

Safety bar adjustment

Safety collar adjustment

Page 22: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

22

TheLCDscreendisplaysagraphicaluserinterface(GUI)fromwhichallthepumpsettingsandrunparametersareentered.Wheneveran itemiscurrentlyselected,thecolorschemeoftheelementtypicallyturnstoadarkershadeofthecolor.Forexample, the light blue buttons turn dark blue when selected, and the white background surrounding an unselected editable textfieldturnsbluewhenthefield isselectedandavailable forediting.Also,whenaGUIelement isselected, thetopbluebarwillshowtheactivityperformed(buttonsonly)orshowthelimitstothevaluesthatmaybeenteredfortheselectedfield(editablefieldsonly).

Selecting the itemsthatwillbemodifiedcanbedonetwoways: (1)using theLCDtouchscreenor (2)using thenavigationkeypad.

AlloftheelementsoftheGUIcanbeinteractedwithbytouch.Typicaltouchinteractionsaretappingafingeronabuttonortextfieldorslidingafingertoscroll.Asaresistivetouchscreen,theseinteractionsarenotrestrictedtoonlyunglovedfingertips.Many types of objects can be used to select items on the screen with only very slight pressure, but avoid using sharp objects, as theymaypermanentlydamagethescreen.

Elements of the GUI may also be selected using the navigation keypad on the front of the instrument to move between the elements.

Toactivate(or“press”)abuttonusingthenavigationkeypad,selectthebuttonelementsuchthatitishighlightedandthenpressthe ENTERkeyatthecenterofnavigationkeypad.Editablefieldscanimmediatelybeediteduponselectionwithouttheneedto press the ENTERkey.

Pump OperationOperating InterfaceLimitsforselectedfields

Selected Fields

Unselected Field

Tip:Users may find the navigation keypad easier to use for scrolling through items in a scrollable list, such as the list of syringes in the syringe library.

Page 23: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

23

AllofthescreensintheFusionSeriesGUIexistinahierarchy,withtheModeSelectionscreenatthetop.Enteringanewscreenisperformedsimplybyactivating,ortapping,abuttonoriconthatopensthenewscreen.

Mosteditablefieldswillrequiretheusertoenternumbers,whichisdoneusingthenumerickeypadonthefrontoftheinstrument.

Uponfirstselectinganeditablefield,pressinganynumberwillfirstcleartheoldvalueandreplaceitwiththenewvalue.Thisonlyapplieswhenfirstselectingthefield.Onafieldthathas already been selected and edited, any further edits will just append the number to the end.Insuchacase,toclearthecurrentvalueinthefield,pressthewhiteC key on the numeric keypad.

Valuesenteredinaneditablefieldareaccepteduponpressingthe ENTER key, selecting a new element, or pressing the Start key.Valuesthatexceedthe limitsof theeditablefield(shown in the blue bar at the top of the screen) are automatically corrected to be within the limitsuponacceptingthevalue.

Therearesomeeditablefieldsthatareusedtocreatenames,suchastheSave Run Settings screen.Whiletherearenolettersindicated on the physical numeric keypad, letters can be entered on the GUI numeric keypad or the physical numeric keypad using the lettering scheme

typicallyseenonphonekeypads.Thefirstpressofanumberkeywillenterthenumber in thefield.Eachsubsequentquickpressof thenumberkeywill cyclethrough the letters associated with that number. Once all the letters for thatnumber have been cycled through, the next key press will start back over with the numberagain.Notethatthesekeypressesmustbeinrapidsuccessiontocyclethroughthenumberandletters;anypauseinkeypresseswillcausethenextkeypress toaddanewcharacter to thefield.Theclearkey (C)orbuttonactsasadeleteorbackspacekey,whichwillremovethepreviouscharacter.

Toreturntothepreviousscreen,taporactivatethered“X”buttoninthetoprightofthescreen.Thiswill“close”thecurrentscreen,andreturntothescreenthatwasusedtoentertheclosedscreen.ThiscanalsobeaccomplishedbypressingtheStopkeyonthefrontoftheinstrumentwhenthepumpisnotrunning.

Navigating Between Screens

Entering Numbers

Entering Letters

Physical Numeric Keypad

Digital Numeric Keypad w/ Letters

Page 24: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

24

Mode Selection ScreenThe Mode Selection screen is the top level menufortheFusionseriespumps.Thisscreenallows users to select the Basic or Multi-Step modes, load a previously saved method, and adjustsystemsettingsorpowersettings.

The Mode Selection screen consists of two screens.Itemsoneachscreencanbetappedbytouchtoentertheselectedoption.Tomaneuverbetween the screens, the white triangular arrow on the edge of the screen can be tapped to go betweenthetwoscreens.Userscanalsoswipeafingeronthescreentoswitchbetweenscreens.Additionally, the navigation keypad can also be used to select an item, and the ENTER key is usedtoaccesstheselectedoption.

For a description of the items listed in the Mode Selection screen, please refer to the associated sectioninthismanual.

Mode Selection Screen 1

Mode Selection Screen 2

Page 25: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

25

Basic Mode (or Single-Step Mode) is used for applications that do not require advanced automation, variable volume andratesettings,steplooping,orrampingrates.

The Basic Mode screen can be reached by tapping on the Basic icon on the Mode Selectionscreen.

Basic Mode allows users to select the syringe being used, to set a valid volume and rate, and to choose whether to infuse orwithdraw(Fusion200only).

The pump can only be started in Basic Mode while the Basic Modescreenisdisplayed.

Basic Mode

Syringe Selection

Basic Mode Screen

Foreveryapplication,thefirstrequiredstepforsetupistoselecttheinnerdiameter(ID)ofthesyringe(s)beingused.Thisvalueis important for the syringe pump to accurately calculate the transfer rate and the total transferred volume for the liquid in the syringe.Therearetwoapproachestoenteringthisvalue:(1)enteracustominnerdiameteror(2)usethebuilt-insyringelibrarytoimportaninnerdiameter.

Entering a custom inner diameter (ID) is as simple as typing in the ID in millimeters in the Syringeeditablefield.Themaximumand minimum values for the ID are displayed in the blue bar at the top of the screen while the Syringeeditablefieldisselected.Thisvalueisalsolimitedtoonlythreedecimalplaces,whereanyadditionaldecimalsortrailingzeroesaredropped.IfthevalueoftheVolumefieldissmaller/largerthanthemin/maxvolumeallowedfortheinputinnerdiameter,thevalueintheVolumefieldwillbeadjustedtobewithinrange.Otherwise,nochangestothevolumewilloccur.

Tapping the Find Syringe button to the right of the Syringe editablefieldwillopenanewSyringe Libraryscreen.Syringemanufacturers are found in the list on the left. Selecting amanufacturer will display all of the syringes available from thatmanufacturerinthelistontheright.Thelistofsyringestypicallyindicatesthevolumeofthesyringefirst,followedbytheinnerdiameterofthesyringe.Bothlistscanbescrolledthroughbydraggingafingeronthescrollbar;however,theeasiest approach in this screen may be to use the navigation keypadtoscrollthroughthelists.AlistofthesyringesinthelibrarycanbefoundinAppendixA.

Syringe Library Screen

Page 26: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

26

When a particular syringe is selected from the list on the right, an Accept/Enterbuttonwillappearnexttotheselecteditem.Tapping the Accept/Enter button (or pressing the Enter key on the navigation keypad) will import the inner diameter associated with the selected syringe and exit the Syringe Libraryscreen.

Upon import of a syringe from the syringe library, the Syringeeditablefieldwillbeupdatedtothenewinnerdiameter,andtheVolumeeditablefieldwillbeupdatedtothevolumeoftheimportedsyringe.Pleasenotethatthevolumeunitsandvaluesforthe Volume and RatefieldsmaychangetomatchtheunitsofvolumedisplayedforthesyringeintheSyringe Library.

VolumeThe Volumefieldrepresentsthetotalvolumeofliquidthatthepumpwilltransfer(infusionorwithdrawal)beforestopping.Theamountoftimethatthepumpwillrunwiththesettargetvolumecanbedeterminedbydividingthevolumebytheflowrate.

Thevalueenteredcanbeanyintegerordecimalwithinthelimitsforthesyringe.Themin/maxvolumelimitscanbefoundinthebluebaratthetopofthescreenwhentheVolumefieldisselected.Anyenteredvaluebeyondtheselimitswillbeautomaticallyadjustedtotheminimumormaximumlimit,whicheveriscloser.Thetotalnumberofdecimalplacesislimitedtofive.

ThevolumecanbestatedinunitsofmLorμL.TheunitsusedforthevolumearedependentuponthesetunitsofvolumefortheflowrateintheOptionsscreen.

To continuously run the pump and use the entire volume of the syringe, users can set the volume to the total volume of the syringeortothemaximumlimit.However,usersarestronglyencouragedtosetthesafetycollar/bar(seeAdjustingtheSafetyCollar)sothatthepumpdoesnotpotentiallyruinthesyringe.

Tip:Because changing the values of the syringe ID may possibly change the values and units for volume and rate, selecting a syringe or entering the syringe ID should always be the first step when setting up a run.

Important Note:If more than one syringe will be used at the same time, it is recommended that the all of the syringes be of the same inner diameter. Because Fusion pumps calculate the rate that the pushing block moves based on the input syringe inner diameter and flow rate, syringes with larger or smaller inner diameters will transfer more/less volume than indicated by the GUI.

Important Note:The maximum volume limit for a syringe assumes that the length of the syringe is the length of the pump. Therefore, this maximum volume limit is almost always larger than the actual volume the syringe can hold.

Page 27: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

27

Example:To set a delay of 1 min and 30 seconds, press the “1” key, then the decimal key, and finally the “3” and “0” keys. To set a delay of 45 seconds, press the decimal key, then the “4” and “5” keys.

Infusion/Withdrawal (Fusion 200 only)

Start-Time Delay

Flow Rate

TheFusion200iscapableofbothinfusionandwithdrawal.OnFusion200systems,anInfusion/Withdrawal toggle button shouldappeartotherightoftheVolumefield.Thetextandarrowdirectiononthebuttonindicatesthemodethatiscurrentlysetanddirectionthatthepusherblockwillmoveuponstartingtherun.Tappingthebuttonwillswitchbetweenthetwomodes.

The Delayfieldrepresentstheamountoftime(inminutesandseconds)todelaystartingthepumpafterpressingtheStart button.

When entering this value with the numeric keypad, the decimal key is used to indicate the separation between minutes and seconds.Thefirsttwovaluesenteredwillbeminutesunlessazeroisthefirstvalueoradecimalisentered,theoccurrenceofwhichcausesthenextvaluestobeinseconds.

To remove the delay, simply press the clear key (C)orsetthetimetozero.

The Flow Ratefieldindicateshowfastthepusherblockwillmove.Theamountoftimethatthepumpwilltrytoruncanbedeterminedbydividingthesetvolumebytheflowrate.

Thevalueenteredcanbeanyintegerordecimalwithinthelimitsforthesyringe.Themin/maxflowratelimitscanbefoundinthe blue bar at the top of the screen when the Ratefieldisselected.Anyenteredvaluebeyondtheselimitswillbeautomaticallyadjustedtotheminimumormaximumlimit,whicheveriscloser.Thetotalnumberofdecimalplacesislimitedtofive.

Theratecanbestatedwithfourdifferentunits:mL/min,mL/hr,μL/min,andμL/hr.Adjustingtheunitswillaffectthemin/maxlimits for the Ratefield.TheunitsforratecanbesetinOptions.

Caution:Flow rate is often set for the needs of the experiment; however, users should be aware that setting high flow rates, especially for viscous fluids, may require high linear force beyond the limits of the pump which may cause the pump to stall out. Additionally, high flow rates can exert tremendous pressure within a syringe that may cause glass or plastic syringes to burst. In the case of high-pressure experiments, users should consider using Chemyx Stainless-Steel Syringes and the high-pressure Chemyx Nexus 6000 pump.

Page 28: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

28

Flow-Rate Units

Saving Run Parameters

TheFusion200 is capableofboth infusionandwithdrawal.OnFusion200 systems, an Infusion/Withdrawal togglebuttonshouldappeartotherightoftheVolumefield.Thetextandarrowdirectiononthebuttonindicatesthemodethatiscurrentlysetanddirectionthatthepusherblockwillmoveuponstartingtherun.Tappingthebuttonwillswitchbetweenthetwomodes.

Theflow-rateunitsdeterminewhichunitsofvolumeandtimewillbeusedfortheVolume and Ratefields.Italsodeterminesthe units for the Priming/Bolus Rate.ThissettinginBasic Mode also changes the units for Multi-Step Mode.

All of the current settings in Basic Modecanbesavedforfutureuse.Thename can be any combination of the alphanumeric characters, up to a maxofeightcharacters.Ifanamealreadyexistsinmemory,theuserwillbeasked tooverwrite thepreviousmethod. If theuserentersauniquenameorselects“Yes”totheoverwritequestion,thesettingswillbesavedto the overwrite question, the settings will be saved to the pump, and the screen will return to the Basic Modemainscreen.Iftheuserselects“No”to the overwrite question, the screen will return to the Save Asnamefieldfortheusertoeditthename.TheFusionseriespumpscanstoreuptoamaximumof20differentmethods.

Note:Even if the units have changed, the values within the Volume and Rate fields will not change, which will result in significant changes in the flow rate and volume transferred. Additionally, the min/max limits for these fields will also be changed, so if the current values for volume and rate exceed the new limits, they will be adjusted to be within those set limits.

Save Run Settings screen

Save Run Settings screen

Options

Page 29: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

29

Options

Enable Volume Lock

Priming allows for the user to temporarily run the pump to remove dead/empty volume in the syringe needle or tubing prior torunningtheexperiment.

The rate at which priming occurs is set in the Priming/Bolus Rate field. Thevalueenteredcanbeany integerordecimalwithinthelimitsforthesyringe.Themin/maxflowratelimitscanbefoundintheorangebaratthetopofthescreenwhenthe Priming/Bolus Ratefieldisselected.Anyenteredvaluebeyondtheselimitswillbeautomaticallyadjustedtotheminimumormaximumlimit,whicheveriscloser.Thetotalnumberofdecimalplacesislimitedtofour.

The units for the Priming/Bolus Rate are determined by the Flow-RateUnitssettinginOptions.Adjustingtheunitswillaffectthe min/max limits for the Priming/Bolus Ratefield.

To enter priming mode on Fusion pumps, press and hold the Start key while in the main screen of Basic Mode.

Thisfunctionhasbeendeprecatedandisnolongersupported.

Caution:Flow rate is often set for the needs of the experiment; however, users should be aware that setting high flow rates, especially for viscous fluids, may require high linear force beyond the limits of the pump which may cause the pump to stall out. Additionally, high flow rates can exert tremendous pressure within a syringe that may cause glass or plastic syringes to burst. In the case of high-pressure experiments, users should consider using Chemyx Stainless-Steel Syringes and the high-pressure Chemyx Nexus 6000 pump.

Page 30: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

30

Multi-Step Modeisusedformoreadvancedapplicationsthatmayrequirenumeroussteps,multipledifferentvolumeandratesettings,steplooping,orrampingrates.

The Multi-Step Mode screen can be reached by tapping on the Multi-Step icon on the Mode Selectionscreen.

The pump can only be started in Multi-Step Mode when the Multi-Step Mode screen isshowing.Thisalsoholds true forcomputercontrolofthepumpthroughaserialconnection;thepumpmustbeswitchedtotheMulti-Step Mode screen on the pumpinterfacefirst.

Multi-Step Mode

Setup

Syringe Selection

In Multi-Step Mode, the left side of the screen shows a list of selectableitems.Thefirst(topmost)iteminthislistisSetup.TappingSetup will take users to the Multi-Step Method Setup screen, where the syringe inner diameter (ID), total number of steps, number of loops,andflow-rateunitscanbeset.

Syringe selection for Multi-Step Mode works in much the same manner as seen in Basic Mode.

ChangesinthesyringeIDwillaffectthemin/maxvaluesfortheVolumefieldinallsteps.However,unlikeBasic Mode, importing a syringe from the Syringe Library will not change the volumes in each step to the volume of the syringe unless they exceed the volumelimitsofthenewsyringe.Allvaluesineachstepthatareoutsidethesenewlimitswillbeadjustedtobewithinthelimits.

Multi-Step Mode Setup screen

Tip:Because changing the values of the syringe ID may possibly change the values and units for volume and rate, selecting a syringe or entering the syringe ID should always be the first step when setting up a run.

Page 31: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

31

Total Steps

Loop All

Sub-Loops

The Total Stepsfielddetermineshowmanystepswilloccurinthemethod.EachstephasindependentvaluesforVolume, Rate, Delay, and Infusion/Withdrawal.Uponenteringthenumberofsteps,thelistontheleftofthescreenwillupdatetoshowallofthesteps.Eachoneofthestepswillhavetheirownsetofparametersthatwillneedtobeset(seeStepParameters).

Aminimumof1andmaximumof99stepscanbeenteredasamethod.

Adding steps is accomplished by increasing the Total Steps,whichwilladdthenewstepsattheendofthelist.DecreasingtheTotal Stepswillremovestepsfromtheendofthelist.Ineithercase,thevaluesintheStepParametersforeachexistingstepdoesnotchangeupontheadditionordeletionofsteps.

In the case that the number of steps in the list on the left side of the screen exceeds the available space on the screen, a blue trianglearrowwillappearbelowthelistindicatingthatitcanscroll.Thislistcanscrolltwoways:(1)theusercandragtheirfingerupanddownonthelisttoscrollthelistandthentapontheitemofinterest;or(2)theusercanusetheup/downarrowsinthenavigationkeypadtoscrollthroughtheitems.

The Loop Allfielddetermineshowmanyadditionaltimesthepumpwillperformallofthesteps.Thisisthehighestlevelloopandisthelastlooptoiterateafterallthestepsandsteploopshavecompleted.

Thisfunctionhasbeendeprecatedandisnotsupported.Please leave these fields empty.

The Loop Allfieldtakesanyintegerfrom0to99.Avalueof0indicatesthatthemethodwillrunthroughtheentiretyofstepsonlyonce.Anyvalueabove0representsadditionaltimestorunallofthesteps.

In the pump running screen, the current loop number and total number of loops are indicated in the blue bar at the top of the screen.

Example:In a multi-step method of three steps (1, 2, 3) where the Loop All is set to “2”. The method will run the following steps sequentially: 1, 2, 3, 1, 2, 3, 1, 2, 3. As shown, the method will loop through all three of the steps two additional times.

Page 32: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

32

Flow-Rate UnitsTapping the Optionsbuttonopensascreenthatallowsflow-rateunitstobeset.Theflow-rateunitdetermineswhichunitofvolume and time will be used for the Volume and Ratefieldsforallofthesteps.ThissettinginMulti-StepModealsochangestheunitsforBasicMode.

Every step in the multi-step method will have a selectable entry in the list on the left side of the screen after Setup.Tapping the step will bring users to the Step Parameters screenforthatstep.Inthisscreen,thetransfervolume,flowrate,timestartdelay,andnumberofloopscanbeset.

Important Note:Even if the units have changed, the values within the Volume and Rate fields will not change, which will result in significant changes in the flow rate and volume transferred. Additionally, the min/max limits for these fields will also change, so if the current values for volume and rate exceed the new limits, they will be adjusted to be within the limits.

Multi-Step Mode Options screen

Step Parameters screen

Step Parameters

Page 33: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

33

VolumeThe Volumefieldrepresentsthetotalvolumeofliquidthatthepumpwillattempttotransfer(infusionorwithdrawal)beforecompletingthestep.Theamountoftimethatthepumpwillattempttorunwiththesetvolumecanbedeterminedbydividingthevolumebytheflowrate(aslongastheflowrateisstatic).

Thevalueenteredcanbeanyintegerordecimalwithinthelimitsforthesyringe.Themin/maxvolumelimitscanbefoundinthebluebaratthetopofthescreenwhentheVolumefieldisselected.Anyenteredvaluebeyondtheselimitswillbeautomaticallyadjustedtotheminimumormaximumlimit,whicheveriscloser.Thetotalnumberofdecimalplacesislimitedtofive.

ThevolumecanbestatedinunitsofmLorμL.TheunitsusedforthevolumearedependentuponthesetunitsofvolumefortheflowrateintheOptionsscreenduringthemulti-stepsetup.

To continuously run the pump and use the entire volume of the syringe, users can set the volume to the total volume of the syringeortothemaximumlimit.However,usersarestronglyencouragedtosetthesafetycollar/bar(seeAdjustingtheSafetyCollar)sothatthepumpdoesnotpotentiallyruinthesyringe.

Important Note:Even if the units have changed, the values within the Volume and Rate fields will not change, which will result in significant changes in the flow rate and volume transferred. Additionally, the min/max limits for these fields will also change, so if the current values for volume and rate exceed the new limits, they will be adjusted to be within the limits.

Infusion/Withdrawal (Fusion 200 only)TheFusion200iscapableofbothinfusionandwithdrawaloneachstep.OnFusion200systems,anInfusion/Withdrawal togglebuttonwillappeartotherightoftheVolumefield.Thetextandarrowdirectiononthebuttonindicatesthemodethatiscurrentlysetandthedirectionthatthepusherblockwillmoveuponstartingtherun.Tappingthebuttonwillswitchbetweenthetwomodes.

Page 34: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

34

Flow Rate

Static Flow Rate

Variable Flow Rate

The Flow Ratefieldsindicateshowfastthepusherblockwillmove.Theamountoftimethatthepumpwillruncanbedeterminedbydividingthesetvolumebytheflowrate.

In Multi-Step Mode,theflowrateduringthestepcanbestaticorvariable,whichissetbythetworatefields(Rate 1 and Rate 2)ineachstep.Rate 1representstheflowratewhenthestepstartswhileRate 2representstheflowratewhenthestepends(seeStaticFlowRateandVariableFlowRateformoredetails).

Foreachfield,thevalueenteredcanbeanyintegerordecimalwithinthelimitsforthesyringe.Themin/maxflowratelimitscanbefoundinthebluebaratthetopofthescreenwhentheRatefieldisselected.Anyenteredvaluebeyondtheselimitswillbeautomaticallyadjustedtotheminimumormaximumlimit,whicheveriscloser.Thetotalnumberofdecimalplacesislimitedtofive.

Theratecanbestatedwithfourdifferentunits:mL/min,mL/hr,μL/min,andμL/hr.Adjustingtheunitswillaffectthemin/maxlimitsfortheRatefield.TheunitsforratecanbesetinOptionsduringthemulti-stepsetup.

Astaticflowrateduringastepmeansthattheflowratedoesnotchangeduringthestep.Staticflowratesaresetsimplybyediting the Rate 1field.Anyeditofthe Rate 1fieldwillautomaticallychangeRate 2tobeequivalent.WheneverRate 1 and Rate 2areequivalent,theflowratewillbestatic.

Avariableflowrateduringastepmeansthattheflowratewillchangeduringthestep.Variableflowratesaresetsimplybyediting the Rate 2fieldtoavaluethatisdifferentfromRate 1.Duringastep,theflowratewillstartatRate 1 and will linearly increase or decrease such that Rate 2isreachedattheendofthestep’srun.

Caution:Flow rate is often set for the needs of the experiment; however, users should be aware that setting high flow rates, especially for viscous fluids, may require high linear force beyond the limits of the pump which may cause the pump to stall out. Additionally, high flow rates can exert tremendous pressure within a syringe that may cause glass or plastic syringes to burst. In the case of high-pressure experiments, users should consider using Chemyx Stainless-Steel Syringes and the high-pressure Chemyx Nexus 6000 pump.

Page 35: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

35

Step LoopThe Step Loopfielddetermineshowmanyadditionaltimestosequentiallyperformtheparticularstep.Thisisthelowest-levelloopthatiterateswheneverthatparticularstepisreachedandcompleted.

The Step Loopfieldtakesanyintegerfrom0to99.Avalueof0indicatesthatthemethodwillrunthesteponlyonce.Anyvalueabovezerorepresentsadditionaltimestorunthestep.

In the Pump Runningscreen,thenumberofremainingloopsofthecurrentstepappearsontheright,nexttotherepeaticon.

Start-Time DelayTheDelayfieldrepresents theamountof time (inminutesandseconds) todelaystarting thepumpafter thecurrentstepbegins.

When entering this value with the numeric keypad, the decimal key is used to indicate the separation between minutes and seconds.Thefirsttwovaluesenteredwillbeinminutesunlessazeroisthefirstvalueoradecimalisentered,theoccurrenceofwhichcausesthenextvaluestobeinseconds.

Toremovethedelay,simplypresstheclearkey(C)orsetthetimetozero.

Example:To set a delay of 1 min and 30 seconds, press the “1” key, then the decimal key, and finally the “3” and “0” keys. To set a delay of 45 seconds, press the decimal key, then the “4” and “5” keys.

Example:In a multi-step method of five steps (1, 2, 3, 4, 5) where Step 4 has a Step Loop of “2”. The method will run the following steps sequen-tially: 1, 2, 3, 4, 4, 4, 5. When the step loop is set to “2”, Step 4 is repeated twice more. If the entire method also has a Loop All setting of “1”, the method would run the following steps sequentially: 1, 2, 3, 4, 4, 4, 5, 1, 2, 3, 4, 4, 4, 5.

Page 36: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

36

Thepump’sactionsarecontrolledbytheStop, Pause, and Start keysonthefront.Inordertostartarun,thescreenshouldbeshowing either the Basic Mode (for basic runs) or Multi-Step Mode(formulti-stepruns)mainscreens.PressingtheStartkeyineitherotherthesescreenswillstartarunwiththecurrentlysetparameters.

Amount Functions

Start

Pressing the Start button once will start a new run with the currently displayed parametersorresumeapausedrun.Thisbuttonisonlyfunctionalwhenaruniscurrently paused or when the display is showing the Basic Mode or Multi-Step Mode screensthatshowthecurrentparameters.

Pressing the Start button while a delay countdown is occurring will immediately skip theremainingdelaytimeandstarttherun/step.

Pressing and holding the Start button will begin priming, where the pump will continuouslyrunatthePrime/BolusRateaslongasthebuttonisbeingpressed.ThePrime/BolusRateissetinBasicMode>Options.PrimingisonlyavailableinBasicMode.

ThegreenLEDlightontheStartbuttonwillbeonwhilethepumpisrunning.

Pause

While the pump is running, pressing the Pausebuttonwillpausethecurrentrun.Pausing allows the pump to continue the same run from the point when the pause occurred.

The yellow LED light on the Pause button will be on when the current run has been paused.

Stop

During a run, pressing the Stop button will immediately stop the run, and return the display to the Basic Mode or Multi-Step Modescreen.

Whileoutsideofanactiverun,pressingtheStopbuttonwillexitthecurrentscreen.Thisresultisthesameasactivatingthe“X”buttoninthetoprightcornerofthescreenbytouchornavigationalkeypad.

The red LED light on the Stop button will be on whenever the pump is neither actively runningnorpaused.Thelightwillalsobeonforactiverunsthatarecurrentlyinadelaybeforestartingastep.

TheredLEDlightwillblinkifthepumphasstalled.

Upon starting a pump run, the screen will switch to the System Statusscreen.Therearetwodifferentversionsofthisscreendepending upon whether the pump was started in Basic Mode or Multi-Step Mode.

Pump Control

Page 37: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

37

Run Status

The System Status screen for Basic Modehighlightsthecurrentactivityofthepumpduringarun.

The Elapsed Timefieldindicateshowmuchtimehaspassedsincethepumphasstartedphysicallyrunning,notincludinganystart-timedelay.The Infusion Timefieldshowsthecalculatedtotaltimeforthecurrentruntofinish,barringanypauseorstallintheinstrument.

Thetworectangularblocksonthescreenillustratethetransferofvolumeduringtherun.Theleftrectanglerepresentsthevolumeinthesyringewhiletherightrectanglerepresentsthevolumetransferredinoroutofthesyringe.Whilethepumpisrunning,asmallarrowwillappearbetweenthetworectanglesindicatingthedirectionofflow.Dependinguponthedirectionofthearrow,thenumberswithineachrectanglewithchangeasvolumeistransferred.

Thecoloredtextdirectlyunderthevolumetransfergraphicindicatesthecurrentstatusofthepump:green-running;yellow-pausedordelayed;andred-stoppedorstalled.

The Syringe, Volume,andRatefieldsindicatetheparametersthatweresetforthecurrentlyrunningmethodintheBasic Mode mainscreen.

Tapping the Rate Adjustbuttonallowsuserstochangetheflowrateofthepumpinthemiddleoftherun.

Basic Mode System Status screen

Page 38: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

38

The Load Run screen allows users to recall Basic Mode methods that were saved in the Optionsscreen.

The Load Run screen can be reached by tapping on the Load folder icon on the Mode Selectionscreen.

The left side of the screen shows a list of all the saved methods (amaximumof20).Tappingandselectingamethodinthislistwill display the parameters for the method in the top right of thescreen.

To load a method, select the method and then press the Load button on the bottom right of the screen. The screen willimmediately change to the Basic Mode main screen with all oftheparametersofthesavedmethod.

The System Settings screen can be reached by tapping on the Settings icon on the Mode Selectionscreen.

Load Run

System Settings

Basic Mode System Status screen

System Settings screen

To delete a method, select the method and then press the Deletebuttonontherightsideofthescreen.Apop-upscreenwillasktheusertoconfirmdeletion,anduponpressingYes,themethodwillberemovedfromthesystem’smemoryandfromthelistontheleft.

Touchscreen TogglePressing the Touch On/Off toggle button will enable/disable the touchscreen capabilities of the LCD display, respectively. If thetouchscreenisoff,onlythephysicalarrowkeyscanbeusedtoselectitems on the screen. Conversely, theTouch On/Off toggle button is always selectable by touch regardless of the current state of the toggle, sothatyoumayreturntousingthetouchscreencapabilities.

Page 39: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

39

Thebaudrateistherateatwhichinformationistransferredthroughtheserialconnection.Thisrateisoftenlimitedbythetypeofserialconnectionused.

TheFusionSeriespumpsprovidetwopossiblebaudrates:9600and38400.

SerialconnectionsusingtheUSB-Bportonthebackofthepumpshouldusethefasterbaudrateof38400.

SerialconnectionsusingtheDB9portonthebackofthepumpwillneedtousetheslowerbaudrateof9600.

SeePumpControlbyComputerformoreinformation.

Baud Rate

Page 40: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

40

Power Settings

Power Settings screen

The Power Settings screen can be reached by tapping on the Power icon on the Mode Selectionscreen.

TheFusionSeriespumpscangenerateuptoamaximumof50lbs(35lbsfortheFusion100)oflinearforce.Somesyringesmadeofmorefragilematerialsmaynotbeabletowithstandthefulllinearforceofthepump.OntheFusionSeriespumps,themotor power, and thus the linear force, can be decreased in the Power Settingsscreen.

To adjust this setting by touch, tap the section of the percentage bar that represents the desired motor power percentage (0%onfarleftand100%onfarright).Tomorepreciselyadjustthepercentage,usersshouldslowlydragtheirfingeroverthepercentagebar.

Toadjustthissettingusingthekeypad,selectthemotor-powerpercentagebarwiththenavigationalkeypadorbytouch.Thepercentagebarshouldbeblueincolorwhenselected.Presstheclearkey(“C”onthenumpad)toclearthecurrentvalue,andthenusethenumerickeypadtotypeinthedesiredpercentage;thepercentagebarwillupdateasvaluesareadded.

ThissettingsetstheamountoftimeuntiltheLCDdisplaygoestosleepafterthelastinteractionwiththepumpcontrols.Whenthedisplaygoestosleep,thepumpisstillon.Itwillcontinuetorunthecurrentmethodandisabletoreceivecommandsviacomputerconnection.

Motor Power

LCD Power-Off Time

Important Note:Even if the units have changed, the values within the Volume and Rate fields will not change, which will result in significant changes in the flow rate and volume transferred. Additionally, the min/max limits for these fields will also change, so if the current values for volume and rate exceed the new limits, they will be adjusted to be within the limits.

Page 41: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

41

Pump Control by ComputerTheFusionSeriessyringepumpscanbecontrolledwithanexternalcomputerusinganRS232serialconnection.

Two types of connections are available for external computer control: DB9 and USB, cables for each are available for purchase throughChemyx(call1-281-277-5499orvisitwww.chemyx.com/contact-us).

Mostcomputeroperatingsystems (e.g.,Windows,macOS,Linux,etc.)havebuilt-in terminalprograms thatcanbeused totypeinthecommandsthatrunthepump.OnWindowsXPandearlier,HyperTerminalcanbeusedtocontroltothepump.OnmacOSandLinux,thescreencommandinTerminalcanalsobeusedtocontrolthepump.Therearealsonumerousthird-partyprogramsforserialconnectionsthatexistforeachoperatingsystemthatmaybeeasiertouseandprovidemorefeatures.Athird-partyprogramwillbenecessarytoconnectwithcomputersrunningWindowsVistaandlater.

For tipsonoperating thebuilt-in terminalprogramsand informationonavailable third-partyprograms,please visitwww.chemyx.com/support.

FortheDB9serialconnection,aMale-to-FemaleDB9cablemustbeused.Donotuseany“nullmodem,”“crossover”,or“crossed-over”cables.Thecablecanbeofanylengthupto50ft;however,thespeedoftheconnectionisoftendetrimentallyaffectedatlengthsover15ft.

ForaUSBserialconnection,aUSB-AtoUSB-Bcableshouldbeused.ThemaximumpossiblelengthofasingleUSBcableis15ft;however,USBcablesshouldbelessthan10ftforoptimumperformance.MultipleUSBcablescanbechainedtogethertomakealongerconnectionaslongasthehubconnectingeachindividualcableisconnectedtoapowersupply.

AUSB-to-DB9serialadaptercanbeusedtoconnectaDB9serialconnectiontotheUSBportonacomputer.

Alloftheseconnectionsrequiretheproperdriverstobeinstalledonthecomputer.Typically,mostcurrentoperatingsystemsalreadycontaindriversfortheUSBserialconnection.TheymayalsoalreadyhavedriversfortheDB9serialconnection.USB-to-DB9adapterstypicallyneedtohaveaUSB-to-serialdriverinstalled.Formoreinformationondownloadingandinstallingdriverspleasevisitwww.chemyx.com/support.

DB9 Serial Cable (Male-to-Female) USB-A to USB-B cable

Cable Requirements

Connecting

Page 42: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

42

Regardless of the operating system being used, the terminal program controlling the serial connection should use the following connection settings:

The baud rate setting for the serial connection must match the baud rate set in the pump settings GUI (See Baud Rate in System Settings).

Generally,thepumpshouldbesettousethehigherbaudrate(38400)ifaUSB-onlyconnectionisused.However,ifaDB9-onlyorUSB-to-DB9connectionisused,thebaudratemustbesetto9600.

Typically, if the baud rate is not set correctly, there will be garbled or no response to any commands

COM Port Settings

Baud Rate 9600 or 38400

Data Bits 8

Parity None

Stop Bits 1

Flow Control None

Page 43: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

43

Regardless of the operating system being used, the terminal program controlling the serial connection should use the following connection settings:

Command Description

help Displaysalistofavailablecommands.

start Startsthepumpwiththecurrentlysetparameters.Theparametersthatwillrundependuponwhich mode (Basic Mode or Multi-Step Mode)thepumptouchscreenisshowing.

pause Pausesthecurrentpumprun.Theruncanberesumedfromthispointusingthe start com-mand.

stop Endsthecurrentrun.Theruncannotberesumedandcanonlyberestartedfromthebegin-ning.

set diameterX

Setstheinnerdiameter(ID)ofthesyringeinmillimeters.

ThissettingsimultaneouslyaffectsthesyringeIDusedinbothBasic Mode and Multi-Step Modesettings,whichwillalsoaffectthevolumeandrateparametersinbothmodes.

Xcanbeanintegerordecimal(uptothreedecimalplaces)from0.103mmto40.000mm.

Example set diameter 11.73

Returnsthediameter.

setunitsX

Setstheunitsusedfortherateofinfusion/withdrawal.Thisparameteralsosetstheunitsfortransfer volume, where the transfer volume unit will be the same as the volume unit used in therate.

This setting simultaneously changes the units used in both Basic Mode and Multi-Step Mode settings.

Xcanbeanintegerfrom0to3.Eachintegerrepresentsaspecificrateunit.

0 = mL/min1 = mL/hr2=μL/min3=μL/hr

Example set units 2

Returnstheintegercorrespondingtothesetrateunit.

Com

Page 44: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

44

Command Description

set volume X,X,X,...

Setsthetransfervolume.Thepumpwillrununtilthevolumeinfusedorwithdrawnequalsthisvolume.Thiscommandalsosetswhetherthepumpisinfusingorwithdrawing(Fusion200only)andthenumberofstepsusedinMulti-StepMode.

Xmustbeanintegerordecimal(uptofivedecimalplaces)withinthevolumelimitsforthesyringe(Seereadlimitparametercommand).Anegativesigncanbeaddedinfrontofanyvaluetoindicateaswitchtowithdrawalmode(Fusion200only).

ForBasicMode,onlyonevalueisrequired.EnteringonlyonevaluewillautomaticallyswitchthepumptoBasicMode.

Example set volume 25.0

To set volumes in Multi-Step Mode, the pump must be manually switched to Multi-Step Modeonthepump.Foreachstep,avaluemustbeentered,separatedbycommasandnospaces.Thenumberofstepswillbeautomaticallyadjustedtothenumberofvaluesentered(i.e.,enteringonlythreevalueswillchangethenumberofstepstothree).

Example set volume 3,-2,4

Returnsthetransfervolume,transferrate,andtransfertime(inBasicMode).

Returnsonlythetransfervolumes(inMulti-StepMode).

settimeX,X,X,...

Setsthetargettimeforapumprun.Thiswilladjustthetransferratesothattherunwilllastfortheinputtime.Thetargetvolumewillremainthesame.ThiscommandalsosetsthenumberofstepsusedinMulti-StepMode.

Xmustbeanintegerordecimal(uptofivedecimalplaces).Theresultingtransferratemustbewithintheratelimitsforthesyringe(Seereadlimitparametercommand).

ForBasicMode,onlyonevalue[isrequired.EnteringonlyonevaluewillautomaticallyswitchthepumptoBasicMode.

Example set time 1.0

To set target times in Multi-Step Mode, the pump must be manually switched to Multi-StepModeonthepump.Foreachstep,avaluemustbeentered,separatedbycommasandnospaces.Thenumberofstepswillbeautomaticallyadjustedtothenumberofvaluesentered(i.e.,enteringonlythreevalueswillchangethenumberofstepstothree).

Example set time 1.0,0.5,1.3

Returnstargettimeandrate(inBasicMode).

Returnsnovalues(inMulti-StepMode).

Page 45: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

45

Command Description

setrateX,X,X,...

Setsthetransferrateforinfusion/withdrawal.

Xmustbeanintegerordecimal(uptofivedecimalplaces)withintheratelimitsforthesyringe(Seereadlimitparametercommand).

ForBasicMode,onlyonevalueisrequired.

Example set rate 1.0

To set rates in Multi-Step Mode, the pump must be manually switched to Multi-Step Modeonthepump.Foreachstep,avaluemustbeentered,separatedbycommasandnospaces.

Example set rate 1.0,0.5,1.3

If the number of values is less than the number of steps, the remaining steps will gain thesameratevalueasthelastenteredvalue.Anyextravaluesbeyondthenumberofstepswillbetruncated.

VariableratesmayalsobesetinMulti-StepMode.Rate1andRate2foreachstepcanbesetusingabackslash.Eachstepisstillseparatedbyacomma.

Example set rate 0.5/1.0,0.5/1.0,1.3

Returnstheflowrateandtransfertime(inBasicMode).

Returnstheflowrate1andflowrate2foreachstep(inMulti-StepMode).

Page 46: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

46

Command Description

setdelayX,X,X,...

Setsthestarttimedelay(inminutes)priortoeachstepofapumprun.Uponstartingthe a step, the pump will wait until the delay time elapses before starting to transfer volume.

Xmustbeanintegerordecimal(uptofivedecimalplaces)withintheratelimitsforthesyringe(Seereadlimitparametercommand).

ForBasicMode,onlyonevalueisrequired.EnteringonlyonevaluewillautomaticallyswitchthepumptoBasicMode.

Example set delay 1.0

To set rates in Multi-Step Mode, the pump must be manually switched to Multi-Step Modeonthepump.ForeachstepofMulti-StepMode,avaluemustbeentered,separatedbycommasandnospaces.

Example set delay 1.0,0.5,1.3

If the number of values is less than the number of steps, the remaining steps will gain thesamedelayvalueasthelastenteredvalue.Anyextravaluesbeyondthenumberofstepswillbetruncated.

Returnsthedelaytime(inBasicMode).

Returnsthedelaytimeforeachstep(inMulti-StepMode).

read limit parameter

Returnsthemin/maxvaluesthatcanbesetfortransfervolumeandtransferrate.Thereturnedvaluesaredependentonthesyringeinnerdiameter(ID).Thetransfervolume limits are typically greater than the maximum volume of the syringe being used.

Returns: max rate, min rate, max volume, min volume

dispensed volume Returnsthetransferredvolumeforthecurrentrun/step.Ifnorunhasstarted,returnsthetransferredvolumeforthelastrun/step.

elapsed timeReturnstheelapsedtime(inminutes)forthecurrentrun/step.Ifnorunhasstarted,returnstheelapsedtimeforthelastrun/step.Elapsedtimedoesnotincludeanydelaytime.

view parameter

Returnsthecurrentlysetparameters(BasicModeonly).

Returns: rate units, syringe inner diameter, transfer rate, priming rate, time (in whole number minutes), transfer volume, time delay (in whole number minutes)

Page 47: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

47

Command Description

status

Returnsthecurrentstatusofthepump.Thereturnedintegercorrelatestothestatusofthepump.

0: Pump stopped1: Pump running2: Pump paused3: Pump delayed4: Pump stalled

Returns:integerthatrepresentsthecurrentstatusofthepump.

hexw2XXXXXXstart

Setsmultipleparametersinonecommand(BasicModeonly).

Xvariablesinorder:units: see set units commandtransfer mode: 0-infusion, 1-withdrawalsyringe ID: see set diameter commandtransfervolume:seesetvolumecommand;nonegativestransfer rate: see set rate commandstart time delay: see set delay command

ExampleThe following example shows the command using these parameters: units = mL/min; transfer mode = withdrawal; diameter = 23.05 mm; volume = 1.2 mL; rate = 3.0 mL/min; delay = 30 sec (0.5 min) hexw2 0 1 23.04 1.2 3.0 0.5

A start command can be added after the six parameters to immediately start the run withthenewparameters.

Example hexw2 0 1 23.04 1.2 3.0 0.5 start

Allsixparametersdonotneedtobeentered;however,allparametersuptotheparameter(s)ofinterestmustbealsobeincluded.Theotherparametersnotenteredwillremainastheyarecurrentlyset.

Example hexw2 0 1 23.04 2.4

save setting Deprecatedcommand.

Page 48: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

48

Command Description

setprimerateX

Setsthepriming/bolusrate(BasicModeonly).

Xmustbeanintegerordecimal(uptofivedecimalplaces)withintheratelimitsforthesyringe(Seereadlimitparametercommand).

Onlyonevalueisrequired.

Example set primerate 1.0

Returnsthepriming/bolusrate(inBasicMode).

restart Powersoffandthenautomaticallypowersonthepump.

In some cases, depending on the software used to connect to the pump, each command may need to be followed by a line feed charactersuchas“\n”or“\r”toexecute.Ifacommanddoesnotappeartowork,tryadding“\r”or“\n”totheendoftheline.

When the pump does not recognize a command, it will return text similar to the following:

Exampleset units 1 \r

ExampleCommand not recognized-type in “help”and press enter to see a command list.

In addition to the manual control of the pumps through a terminal, Chemyx provides pump controller programs as an unsupportedopen-sourceapplicationswritteninPython,aLabVIEWpackage,andaMATLABpackage.Moreinformationonthesetoolscanbefoundonthewebsiteatwww.chemyx.com/support.

Chemyx Pump Controller Programs

ChemyxprovidesfreedownloadsforLabVIEW™andMATLAB™asacourtesytoourcustomers.Thisservice is not supported by Chemyx, all customers are responsible for solving any technical issues relatedtoLabViewandMATLABcomputercontrol.

Page 49: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

49

Pumplubricationisthemostimportantpartofregularpumpmaintenance.Properlubricationensureslongservicelife,smoothflowperformance,andprotectionagainstcorrosivelabfumes.Thepumpmaynotbeadequatelycleanedandlubricatedifthefollowingoccur:(1)grindingsoundsarecomingfromtheleadscrew;(2)theblockreleasebuttondecouplesundertheloadbeforethepumpstalls;or(3)thepumpstallswithincreasedfrequency.

PleasecontactChemyxcall+1281-277-5499oremailusatinfo@chemyx.comtoconfirmthesafeuseofadifferentlubricant.

1. Cleancontactsurfaceswithacleancloth.Removeanydebrisontheleadscrewandtheguiderods.

2. Applylubricantalongthelengthoftheleadscrewandtheguiderods.

3. Presstheblock-releasebuttonandslidethepusherblocktotheleftedge.

4. Runthepumpend-to-endwithoutasyringetoworkthelubricantintothecomponents.

Pumplubricationshouldbeperformedeverythreemonths.

Pump Maintenance

• Engine/motor oil

• Silicone oil

• Machine oil

• Toolingoilor“WD-40”

• Lithium-based grease

Lubricating the Pump

Lubrication Instructions

Approved Lubricants

Page 50: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

50

The Chemyx Footswitch provides users the ability to start, pause, or stop the pumpfromashortdistancewiththetapofafoot.

The Chemyx 10-Channel Syringe Rack for the Fusion 200 provides users with anexpansionofthenumberofsyringesthatcanbeusedsimultaneously.The 10-channel syringe rack can accommodate up to 10 syringes with volumesfrom1μLto10mL.Thesyringerackcanalsobeusedwithlargersyringes(10–20mL);however,onlyfoursuchsyringeswillfitintherackatthesametime.

The Chemyx Footswitch for Fusion Series pumps has a USB-B cable that plugsintotheTTLportonthebackofthepump.

To use the Footswitch, the pump must be in either the Basic Mode or Multi-StepModescreens.ThebasiccommandsaregivenbyacertainnumberofclicksoftheFootswitch.

A single click of the Footswitch will start the pump run or pause the pump rundependingonwhetherthepumpiscurrentlyrunning.

AquickdoubleclickoftheFootswitchwillcompletelystopthepumprun.

Footswitch

Footswitch

Using the Footswitch

Pump Accessories

Footswitch

10-Channel Syringe Rack

10 Channel Syringe Rack (Fusion 200 Only)

Page 51: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

51

Attaching the Rack

Fusion 200

Syringe Holder with 10 Channels

Flange Retaining Bracket

Clamp Extensions

Clamp Extensions

Plunger Retaining Bracket

Syringe Clamp

Page 52: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

52

Size Diameter

0.50mL 4.64mm

1.00mL 4.64mm

2.00mL 8.93mm

2.50mL 8.66mm

3.00mL 8.93mm

5.00mL 11.73mm

10.00mL 14.68mm

20.00mL 19.60mm

30.00mL 22.70mm

50.00mL 28.03mm

60.00mL 28.60mm

Size Diameter

0.50μL 0.10mm

1.00μL 0.15mm

2.00μL 0.21mm

5.00μL 0.33mm

10.00μL 0.48mm

25.00μL 0.73mm

50.00μL 1.03mm

100.00μL 1.46mm

250.00μL 2.30mm

500.00μL 3.26mm

1.00mL 4.61mm

2.50mL 7.28mm

5.00mL 10.30mm

10.00mL 14.57mm

25.00mL 23.03mm

50.00mL 32.57mm

Size Diameter

1.00mL 4.64mm

3.00mL 4.64mm

5.00mL 12.07mm

10.00mL 14.50mm

20.00mL 19.13mm

30.00mL 21.69mm

60.00mL 26.72mm

Size Diameter

1.00mL 4.80mm

1.00mL 6.70mm

2.00mL 6.70mm

2.00mL 9.20mm

3.00mL 10.30mm

5.00mL 12.20mm

10.00mL 15.00mm

20.00mL 19.00mm

30.00mL 22.50mm

50.00mL 25.50mm

100.00mL 34.00mm

Chemyx Stainless Steel Hamilton Glass

Hoshi

BD Glass

BD Plastic

AppendicesAppendix A: Syringe Volume/Inner Diameter Reference

Size Diameter Inner | Outer

6.00mL 9.5mminner|13mmouter

20.00mL 19.1mminner|23mmouter

50.00mL 28.6mminner|33mmouter

100.00mL 35 mm inner | 38 mm outer

200.00mL 44 mm inner | 48 mm outer

Page 53: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

53

Size Diameter

1.00mL 4.66mm

2.00mL 6.90mm

2.50mL 9.10mm

5.00mL 12.62mm

10.00mL 14.34mm

20.00mL 19.68mm

30.00mL 22.44mm

50.00mL 28.80mm

100.00mL 36.68mm

Size Diameter

0.25mL 3.45mm

0.50mL 3.45mm

1.00mL 4.50mm

2.00mL 8.92mm

3.00mL 8.99mm

5.00mL 11.70mm

10.00mL 14.70mm

20.00mL 19.58mm

30.00mL 22.70mm

50.00mL 29.00mm

Size Diameter

0.25mL 2.60mm

0.50mL 3.20mm

1.00mL 4.30mm

2.00mL 6.30mm

3.00mL 7.30mm

5.00mL 9.50mm

Size Diameter

2.00mL 9.12mm

5.00mL 12.34mm

10.00mL 14.55mm

20.00mL 19.86mm

30.00mL 23.20mm

50.00mL 27.60mm

Size Diameter

1.00mL 4.71mm

3.00mL 9.50mm

5.00mL 12.91mm

10.00mL 15.40mm

20.00mL 19.30mm

30.00mL 23.40mm

50.00mL 29.00mm

Size Diameter

1.00mL 4.70mm

2.50mL 9.70mm

5.00mL 12.48mm

10.00mL 15.89mm

20.00mL 20.00mm

30.00mL 22.50mm

50.00mL 28.90mm

JMC Air-Tite Plastic Norm-Ject Plastic

Popper & Sons Glass

Ranfac

Natsume

Nipro

Page 54: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

54

Size Diameter

25.00μL 0.73mm

50.00μL 1.03mm

100.00μL 1.46mm

250.00μL 2.30mm

500.00μL 3.26mm

1.00mL 4.61mm

2.50mL 7.28mm

5.00mL 10.30mm

10.00mL 14.57mm

Size Diameter

1.00mL 4.70mm

2.00mL 6.40mm

3.00mL 9.30mm

6.00mL 13.10mm

12.00mL 15.40mm

25.00mL 21.00mm

30.00mL 23.00mm

50.00mL 29.00mm

Size Diameter

1.00mL 4.65mm

3.00mL 8.94mm

6.00mL 12.70mm

12.00mL 15.90mm

20.00mL 20.40mm

35.00mL 23.80mm

50.00mL 26.60mm

Size Diameter

1.00mL 4.73mm

3.00mL 9.00mm

5.00mL 13.04mm

10.00mL 15.79mm

20.00mL 20.18mm

30.00mL 23.36mm

60.00mL 29.45mm

Size Diameter

1.00mL 4.73mm

1.00mL 6.50mm

3.00mL 8.65mm

5.00mL 13.00mm

10.00mL 15.80mm

20.00mL 20.15mm

30.00mL 23.10mm

50.00mL 29.10mm

SGE Glass Terumo Japan

TopSherwood Plastic

Terumo

Page 55: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

55

Size Diameter

10.00μL 0.46mm

25.00μL 0.73mm

50.00μL 1.03mm

100.00μL 1.46mm

250.00μL 2.30mm

500.00μL 3.26mm

1000.00μL 4.61mm

Size Diameter

1.00mL 4.65mm

3.00mL 8.94mm

6.00mL 12.70mm

12.00mL 15.90mm

20.00mL 20.40mm

35.00mL 23.80mm

60.00mL 26.60mm

Unimetrics

Kendall Monoject Plastic

Page 56: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

56

Item Description Part Number

Fusion 100 Touch 07100

Fusion 200 Touch 07200

Fusion 400 Touch 07400

Item Description Part Number

Footswitch (w/ USB-B connection cable) 50002

DB9 Serial Cable (Male-to-Female) 50007

USB Serial Cable (USB-B-to-USB-A) 50017

10-Channel Syringe Rack (Fusion 200 only) 50003

Item Description Part Number

Stainless-Steel Syringe (6 mL) SS6

Stainless-Steel Syringe (20 mL) SS20

Stainless-Steel Syringe (50 mL) SS50

O-Ring Replacement 10-pack (6 mL) CP6

O-Ring Replacement 10-pack (20 mL) CP20

O-Ring Replacement 10-pack (50 mL) CP50

Fusion Series Syringe Pumps

Options and Accessories

Stainless-Steel Syringes

Appendix B: Available Parts

Page 57: Syringe Pump Manufacturer | Chemyx - FUSION SERIES · A syringe pump is a small, positive-displacement pump used to gradually transfer precise volumes of fluid.All Chemyx Fusion-series

57

Chemyx Inc.

10905CashRd.

Stafford,TX77477

Webpage:www.chemyx.com

ContactUs:[email protected]

Phone: +1 281-277-5499

Fax: +1 281-277-0045

OfficeHours:Monday–Friday,8:30am–5pm(CentralTime)

Formorearticlesandtutorialvideos,visit:www.chemyx.com/support

Follow Us

LinkedIn:www.linkedin.com/company/chemyx

Twitter:www.twitter.com/chemyxinc

Appendix C: Contact Information