    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 1 of 86 Effective date: December 2008

    Siemens Confidential

    SIEMENS

    Image and Knowledge Management-syngo Workflow SLR

    Siemens Medical Solutions USA, Inc., Health Services Division, Malvern, Pennsylvania, US

    syngo Workflow SLR PC Install Document

    Document ID: 0000002311

    ApplicationVersion:

    syngo Workflow SLR Version 30B

    DocumentVersion:

    0100

    Status: Released

    Owner: Michael Ryan

    Effective Date: December 2008

    Author: Connie Reilly

    Release Date: December 2008

    DocumentAdministrator:

    Michael Ryan

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    2/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 2 of 86 Effective date: December 2008

    Siemens Confidential

    Revision History

    Revision Date Reason for change, and identify alldocuments affected by change

    Minor or MajorChange

    0001 April 2008 Draft None0100 December 2008 Released Minor

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    3/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 3 of 86 Effective date: December 2008

    Siemens Confidential

    EDITION Edition Notice

    Copyright (c) 2008 Siemens Medical Solutions USA, Inc. All rights reserved.This software and its documentation are confidential and proprietary to Siemens and containworks by Siemens' suppliers or open source organizations, in which works these suppliers ororganizations claim copyright. This software may only be used in strict accordance with the termsof the license agreement(s) governing its use.Any reports or other figures that appear in this manual are merely illustrative and do not containnames of or data on real people. Any similarity in names of people, living or dead, or in data isstrictly coincidental and is expressly disclaimed.Siemens does not warrant that the material contained in its documentation is error-free.Documentation supplied to Siemens by third parties and included with this documentation is notwarranted for accuracy or completeness.The information contained in this manual is subject to change. Revisions and updates will beissued from time to time to document changes and/or additions.Siemens and syngo are registered trademarks of Siemens AG.

    Microsoft and Windows are registered trademarks of Microsoft Corporation.

    Dictaphone and PowerScribe are registered trademarks of Nuance Communications, Inc.

    Interlink Electronics is a registered trademark of Interlink Electronics.

    Siemens Medical Solutions USA, Inc., Health Services Division51 Valley Stream ParkwayMalvern, PA 19355Telephone: (610) 219-6300www.usa.siemens.com/medical

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    4/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 4 of 86 Effective date: December 2008

    Siemens Confidential

    What's New in This Edition?

    The following functional and/or technical changes and additions are included in this document.

    Figure 1. What's New in This Edition

    Changes and Additions Topic References

    Throughout the document.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    5/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 5 of 86 Effective date: December 2008

    Siemens Confidential

    Contents

    1. About the PC Install Document..............................................................................................................71.1 Intended Audience...................................................................................................................................71.2 How this Manual Is Organized.................................................................................................................71.3 Related Documentation ...........................................................................................................................81.4 Conventions Used in this Manual............................................................................................................81.5 Distribution Information............................................................................................................................91.6 Questions and Suggestions.....................................................................................................................92. Workstation Requirements...................................................................................................................102.1 New Requirements for V30B.................................................................................................................102.2 General Requirements ..........................................................................................................................102.3 syngo Workflow SLR Workstation Requirements..................................................................................113. Custom/Compact Install - Windows 2000/Windows XP ....................................................................133.1 syngo Workflow SLR Custom/Compact Install......................................................................................133.2 MS Access Runtime - Windows 2000/Windows XP..............................................................................173.3 Sybase 12 - Windows 2000/ Windows XP Installation..........................................................................19 3.4 Configuring Sybase Settings - Windows 2000/ Windows XP................................................................223.5 Formatter Server- Windows 2000/ Windows XP...................................................................................30

    3.5.1 Requirements..................................................................................................................................303.5.2 Installing Print Formatter Server for syngo Workflow SLR.............................................................313.5.3 Word Setting for Print Formatter.....................................................................................................353.5.4 Installing TweakUI...........................................................................................................................363.5.5 Required Microsoft Word XP Change So Documents Do Not Get Corrupted................................36

    3.6 Installing IntegriSign Desktop................................................................................................................383.7 Installing Powerscribe SDK ...................................................................................................................424. Typical Install - Windows 2000/Window XP........................................................................................455. Service Pack Update Detection............................................................................................................51

    6. Release Update Detection ....................................................................................................................527. Install Portal Radiologist Client Software...........................................................................................547.1 General Requirements ..........................................................................................................................547.2 Install Portal Radiologist Client Software ..............................................................................................548. Install IntegriSign (ePad) Signature Software ....................................................................................589. Word Viewer 2003..................................................................................................................................6410. Installing Printers and Print Drivers..................................................................................................6710.1 Printers ................................................................................................................................................6710.2 Bar Codes for Printer...........................................................................................................................6810.3 Install and Select Bar-Code Fonts.......................................................................................................6910.4 Port Setup Instructions for Windows 2000/Windows XP.....................................................................6910.5 Biscom Fax Processor.........................................................................................................................7010.6 Intermec Label Printer .........................................................................................................................7011. Maintenance Mode ..............................................................................................................................7111.1 Modify ..................................................................................................................................................7111.2 Repair ..................................................................................................................................................7111.3 Remove ...............................................................................................................................................72

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    6/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 6 of 86 Effective date: December 2008

    Siemens Confidential

    12. Uninstalling syngo Workflow SLR, MS Access, Sybase 12, Microsoft Word Viewer 2003..........7312.1 syngo Workflow SLR ...........................................................................................................................7312.2 MS Access Runtime (MsAcc10 Platform) ...........................................................................................7312.3 Sybase 12............................................................................................................................................7312.4 Microsoft Word Viewer 2003 ...............................................................................................................7413. Troubleshooting ..................................................................................................................................75

    13.1 syngo Workflow SLR Access or Print Formatter Troubleshooting ......................................................75

    13.1.1 syngo Workflow SLR Access Issues ............................................................................................7513.1.2 Read-Only Files Error ...................................................................................................................7513.1.3 Uninstall Does Not Delete All Files/Directories.............................................................................7613.1.4 Resolution Error ............................................................................................................................7613.1.5 Resolution Setting.........................................................................................................................7613.1.6 Dsedit.exe Error ............................................................................................................................7613.1.7 Print Formatter Problems..............................................................................................................76

    13.2 MS Access Troubleshooting................................................................................................................7713.2.1 Installing MS Access Independent of the syngo Workflow SLR Install ........................................78

    13.3 syngo Workflow SLR Mediator Troubleshooting .................................................................................7813.3.1 Invalid Context ID Error.................................................................................................................7913.3.2 Mediator and Windows XP............................................................................................................7913.4 Unable to Write to Profiles Error..........................................................................................................79

    13.5 Multiple Versions of MS Access ..........................................................................................................8013.6 syngo Imaging Administrator Password ..............................................................................................8013.7 Active X Control Compatibility Problem...............................................................................................8013.8 Release Update Detection Message Box............................................................................................81Glossary .....................................................................................................................................................82

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    7/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 7 of 86 Effective date: December 2008

    Siemens Confidential

    1. About the PC Install Document

    Before you begin, you should be familiar with the installation procedures for your entity. Thisguide often illustrates how to install functionality with examples that are appropriate to most

    entities. In some cases, however, the examples may conflict with the procedures of your entity.This document may present information differently from other Siemens documents you may use.Please take a few minutes to read this introduction before using this document.

    1.1 Intended Audience

    This guide is intended for Installs and Support programmers and experienced systemadministrators involved with the support of syngoWorkflow SLR at their entity.

    1.2 How this Manual Is Organized

    This document contains the following topics.Workstation Requirements in topic 2 Outlines the workstation requirements for

    syngoWorkflow SLR.

    Custom/Compact Install - Windows2000/Windows XP in topic 3

    Provides instructions for a custom softwareinstallation of various syngoWorkflow SLRcomponents for Windows 2000 and WindowsXP environments.

    Typical Install - Windows 2000/Window XP intopic 4

    Provides instructions for a typical softwareinstallation of syngoWorkflow SLR forWindows 2000 and Windows XPenvironments.

    Service Pack Update Detection in topic 5 Describes the update procedure for a servicepack.

    Release Update Detection in topic 6 Describes the update procedure for a release.

    Install Portal Radiologist Client Software in topic7

    Provides instructions for installing the PortalRadiologist client software.

    Install IntegriSign (ePad) Signature Softwarein topic 8

    Provides instructions for installing theIntegriSign Signature software forworkstations that will use an ePad device to

    capture electronic signatures.

    Word Viewer 2003 in topic 9 Provides instructions for installing WordViewer 2003 which enables users to viewMicrosoft Word documents without havingMicrosoft Word installed on the workstation.

    Installing Printers and Print Drivers in topic 10 Contains instructions for installing printers and

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    8/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 8 of 86 Effective date: December 2008

    Siemens Confidential

    print drivers.

    Maintenance Mode in topic 11 Describes how to install syngoWorkflow SLRin Maintenance mode.

    Uninstalling syngo Workflow SLR, MS Access,

    Sybase 12, Microsoft Word Viewer 2003 in topic12

    Provides instructions on how to uninstall

    syngoWorkflow SLR, MS Access, Sybase 12,and Microsoft Word Viewer 2003.

    Troubleshooting in topic 13 Contains troubleshooting instructions.

    Glossary See syngo Workflow SLR Base User's Guidefor a comprehensive list of terms related tosyngoWorkflow SLR.

    1.3 Related Documentation

    Siemens provides extensive product information about the functions, profiles, master tables, andtools in syngoWorkflow SLR. Product information is available in the following formats. Softcopy documentation

    Online Help for PC-based and CRT-based functions

    Classes provided by Siemens Educations Solutions

    Implementation Methodology used when you install syngoWorkflow SLR

    Siemens provides syngoWorkflow SLR-related documentation manuals, user's guides, and setupguides. See syngo Workflow SLR Base User's Guidefor a list of all syngoWorkflow SLRdocuments.Note: Siemens also provides Tip sheets and White Papers on specific information pertaining to

    syngoWorkflow SLR, accessible through the Customers-only Web site.

    1.4 Conventions Used in this Manual

    Menu items, function names, parameter names, and other user-interaction items are shown inbold letters.Safety notices are divided into two classes that relate to potential danger: Warnings andCautions. All safety notices are structured in three parts as follows: source of danger,consequence, and remedy.Notes, cautions, and warnings are specially highlighted:

    Note: Notesprovide user information for optimum operation of the device, the software, andother relevant subjects.

    Caution: Cautionsindicate conditions or consequences that you should pay particularattention to when administering syngoWorkflow SLR but where no direct dangerfor human beings or equipment is involved.

    Warning: Warningsindicate a potential hazard to the health or life of patients or personnel.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    9/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 9 of 86 Effective date: December 2008

    Siemens Confidential

    1.5 Distribution Information

    Please be aware that, due to business and legal considerations, updates to documentation aresent only to the customer information contact. syngoWorkflow SLR manuals, user's guides andsetup guides are available through the Siemens Customers-only Web site. Tip sheets and WhitePapers on specific topics related to syngoWorkflow SLR are also available through the

    Customers-only Web site.

    1.6 Questions and Suggestions

    If you have any questions or suggestions about the documentation, please send your commentsto:Siemens Medical Solutions USA, Inc.Attn: Image and Knowledge Management - syngoWorkflow SLR Department E17Documentation Feedback - syngoWorkflow SLR PC Install Document Version 30B51 Valley Stream ParkwayMalvern, PA [email protected]

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    10/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 10 of 86 Effective date: December 2008

    Siemens Confidential

    2. Workstation Requirements

    2.1 New Requirements for V30B

    Version 3.0 SDK is now supported in syngoWorkflow SLR V30B and higher.

    You cannot use V2.0 SDK with syngoWorkflow SLR V30B or higher. You can only useV2.0 SDK with syngoWorkflow SLR Version 27.2 to Version 30.0.

    2.2 General Requirements

    Video display resolution should be set to a minimum of 1024 x 768 resolution,recommended greater than 256 colors, and either large or small fonts.

    Anti-virus software is strongly recommended.

    Minimum required space to install the syngoWorkflow SLR V30B Production, Trainingand Test environments is approximately 261 MB.

    Minimum required space to install the syngoWorkflow SLR V30B Print Formatter isapproximately 10 MB.

    At least one Windows Workstation or Server with a CD-ROM drive. CD-ROM driveshould be a shared drive for all to access or a Server share with equal access.

    - Siemens strongly suggests creating a share that will be used for all futureService Pack Install/Updates. This share name should not contain spaces. Copythe client install files to this share, and run the client install from this share. Allfuture service pack install/update files should then be copied to this share.

    - A Terminal Emulator is needed on your PC only if you require access to non-GUI(character cell) syngoWorkflow SLR. Any emulator can be used, but SiemensGlobal Services Group will only support the Rumba terminal emulator.

    - When installing third-party software, you should take all default options.

    Due to the new cmbopls1.ocx file, you must uninstall all versions of syngoWorkflowSLR/NOVIUS Radiology prior to V30 loaded on the PC. Use the Add/Remove programand delete the previous syngoWorkflow SLR/NOVIUS Radiology applications prior toV30 before loading V30B software.

    Note: If you use Release Update Detection, you must select the option YES to removeprevious versions of software. After successful installation of V30B, you can thenre-load the previous version, if needed.

    UDT Custom Install requires MS Access 2000 Runtime and Sybase ODBC OpenClient

    You must be logged into a workstation with Administrator privileges to install syngoWorkflow SLR and Sybase OpenClient.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    11/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 11 of 86 Effective date: December 2008

    Siemens Confidential

    To install the Sybase OpenClient ODBC and MS Access Runtime, your workstation musthave 126 MB of free disk space available.

    For disk space requirements, you will need an additional 60 MB, on the "Temp" directorydrive, when installing UDT applications.

    Minimum required space to install the syngoWorkflow SLR V30B Productionenvironment and syngoWorkflow SLR UDT is approximately 244 MB.

    113 MB of additional disk space is needed when installing from the Web.

    Make sure when installing syngoWorkflow SLR software that no other programs arerunning during the installation process.

    All OCX files loaded with Install Shield reside in the Common Files folder found within thesyngoWorkflow V30.0 folder.

    When installing a syngo- Client on a syngoImaging workstation, some environmentvariables need to be manually set to autostart the syngo/Novius-Client, when the syngoImaging application starts up. These environment variables should be set automatically

    during the syngo/Novius-Client Installation to improve the system installation. Theenvironment variables are:

    - MAGICSYNGO_RIS_ROOT = C or the drive letter where the application isinstalled.

    - MAGICSYNGO_RIS_PATH = \Program Files\Siemens\CLN\syngoWorkflow\Production\Readexam.exe

    - MAGIC_SYNGO_BROWSER_TYPE = RIS

    Word Viewer 2003 has replaced Word Viewer 97 as the option you could use to viewUser Defined Reports, in the Management Reporting function if MS Word does not exist

    on the PC.

    2.3 syngo Workflow SLR Workstation Requirements

    Note: Please refer to the current Schedule 1 document for the most up-to-date informationabout syngoWorkflow SLR Workstation required hardware/software. Information listedbelow may be incorrect.

    Required Hardware: 1.4 GHz Pentium 4 (or Pentium compatible)

    512 MB RAM

    4 GB of disk space

    CD-ROM drive

    10/100/1000 MB Network Interface Card

    SVGA Local Bus video adapter or Windows accelerated SVGA graphics adapter

    17-inch SVGA Monitor

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    12/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 12 of 86 Effective date: December 2008

    Siemens Confidential

    Minimum display resolution: 1024 x 768

    Minimum display resolution for workstations that will access Portal Radiologist: 1280 x1024

    Font Setting: Small (96 dpi) or Large (120 dpi)

    Keyboard and Mouse

    A minimum connection speed of 33.6K is required for customers who will use remotetranscription

    Required Software: Windows 2000 Professional or Windows XP Professional

    Microsoft Internet Explorer 5.5, 6.0, 7.0

    A VT emulation package - Required for PCs that will use non-GUI (character cell)functionality. Must include VT220 and key map support.

    Word 2003 or Word 2007 - Required for all workstations where transcription, documentbuilding, electronic signature, or access to the word processing editor is required.

    Microsoft Windows Server Client Access Licensing

    Microsoft SQL Client Access Licensing - Required for users or workstations that willaccess the Voice Servers.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    13/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 13 of 86 Effective date: December 2008

    Siemens Confidential

    3. Custom/Compact Install - Windows2000/Windows XP

    When you select a Custom or Compact Install, you can install Multiple Environments

    simultaneously (PRD, TRN, TST, Print Formatter), MS Access Runtime, Sybase ODBCOpenClient, Powerscribe SDK, on a specific syngoWorkflow SLR workstation.

    3.1 syngo Workflow SLR Custom/Compact Install

    Note: In the future the Compact Install may be used for installing specific files/functionality.Currently, it mirrors the Custom install.

    1. Shut down all Windows programs before running the install. This includes MacAfee, PrintNow, or any program that is running in the background. Check the task bar at the bottomright side of your computer screen for icons that indicate programs are running.

    2. Verify that you are currently logged on to the PC with Local Administrator privileges. Ifunsure, you can check by selecting Start -> Settings -> Control Panel -> Users andPasswords (or User Accounts). Find the current User ID you are using and verify youare listed as a member in the Group Administrators.

    3. If you are installing directly from the delivered CD, load the CD in the CD-ROM drive.Otherwise, map a drive to the network location where the install software is located. Tomap a drive:

    a. From your administrator, get the location of the syngoWorkflow SLR install files.

    b. Open My Network Places. Select Tools->Map Network Drive. Choose a driveletter from the drop-down. Click Browse and locate the folder containing the

    install files. Click Finish. This drive will now display in My Computer on thedesktop under the letter you assigned.

    4. From the Windows Start menu, select Run.

    a. If you are running the install directly from the delivered CD, type the following atthe prompt:[CD:]\Desktop\RAD300\Install\Setup.exe

    b. If you have copied the install software onto a network drive from the CD, run theSetup.exe from that location:[Mapped_Drive:]\Desktop\RAD300\Install\Setup.exe

    Note: If you get an error stating that the directory path is not accessible andthat it has been removed or deleted, verify that the location was enteredcorrectly. If the problem persists, go to Windows Explorer, follow the pathto Setup.exe, and double-click to execute.

    c. The syngoWorkflow Setup - Install Shield Wizard window is displayed.

    5. If previous versions of Novius/syngoRadiology are detected on your computer, thefollowing message box is displayed.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    14/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 14 of 86 Effective date: December 2008

    Siemens Confidential

    6. Select the Yes radio button and click Next. The Release Upgrade detection will removeALL currently installed versions of Novius/syngoWorkflow SLR and in its place, install thenew syngoWorkflow SLR release. The Setup Type Window is displayed.

    7. Select the Custom radio button. The Destination Folder is displayed at the bottom of thescreen. Allow the install to take place in the default directory, (C:\ProgramFiles\Siemens\CLN\syngo Workflow V30.2). If you do not have sufficient space on your[Local Drive:] drive to install the application, use the Browse option to select an alternatedrive and directory.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    15/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 15 of 86 Effective date: December 2008

    Siemens Confidential

    8. Click Next. The Select Components window is displayed.

    9. If you select a Custom Install, you can select to install the following by selecting theappropriate checkboxes.

    Multiple syngoWorkflow SLR environments (PRD, TRN, TST)

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    16/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 16 of 86 Effective date: December 2008

    Siemens Confidential

    Print Formatter

    syngoWorkflow UDT Applications (including MS Access Runtime and SybaseODBC Install Files)

    Caution: Do not install User-Defined Tables applications on the MagicView1000W workstations. The UDT application software conflicts with theMagicView 1000W software, and if loaded, interferes with the use of theMV1000W software. You will be required to re-install the MagicView1000W software.

    10. Select syngoWorkflow Production (and/or Training and/or Test) and syngoWorkflowUDT Applications. Click Next. The Hostname Information screen is displayed.

    11. At the prompt, enter the Hostname of your server.

    Note: In this example, Hostname is RADSERVER. Enter your specific Hostname at theprompt.

    12. Click Next. The Start Copying Files screen is displayed

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    17/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 17 of 86 Effective date: December 2008

    Siemens Confidential

    13. If you want to review or change any settings, click Back. If you are satisfied with thesettings, click Next to begin copying files. The Setup Status window is displayed andsyngoWorkflow SLR begins copying files to the workstation. This step may take severalmoments to complete.

    3.2 MS Access Runtime - Windows 2000/Windows XP

    1. If the MsAcc10 Platform is not already installed on the workstation, the MsAcc10 PlatformSetup window will display. Click OK to continue.

    2. On the MsAcc10 Platform Setup window, the default installation directory should displayas C:\Programs\Siemens\MSAcc10 Platform\.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    18/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 18 of 86 Effective date: December 2008

    Siemens Confidential

    3. Click . The Siemens MSAcc10 Platform - Choose Program Group window is

    displayed. Click Continue to begin the install.

    4. When the Siemens MSAcc10 Platform installation is completed, you will see a message

    box: Siemens MSAcc10 Platform Setup was completed successfully. Click OK. TheSetup Complete screen is displayed.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    19/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 19 of 86 Effective date: December 2008

    Siemens Confidential

    5. The option for placing a shortcut on the Desktop defaults to Yes. If you do not want tohave a shortcut placed on your desktop, uncheck the option.

    6. Click Finish to begin the second part of the UDT installation, Sybase Client Installer. IfSybase is not already installed on the workstation, it will begin to Install. Go to Sybase 12- Windows 2000/ Windows XP Installation.

    Note: As stated on the Setup Complete screen, only a command window (CMD.EXE)will appear. NO user input is required for the Sybase Client Install.

    3.3 Sybase 12 - Windows 2000/ Windows XP Installation

    1. If Sybase is not already installed on the workstation, the Sybase installer willautomatically launch. The C:\WINNT\System32\cmd.exe is displayed and the Sybaseinstallation will begin.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    20/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 20 of 86 Effective date: December 2008

    Siemens Confidential

    2. The installation of Sybase, when near completion, will automatically reboot your PC.When you log back in, you will see an informational screen stating that the Sybase install

    is finish.

    Note: If you are installing from the CD, do not remove the CD.

    3. If this is an upgrade of a workstation that already has Sybase installed and the dsedit andODBC Data Source Administrator values have already been defined, you do not need tocontinue. If not, continue with the next step.

    4. Because the Sybase12 installer installs the links in the wrong folder, the links must bemoved as follows:

    Windows 2000 - In Windows Explorer, move the folder Sybase in\Winnt\Profiles\All Users\Start Menu\Programs\ to Documents and Settings\AllUsers \Start Menu\Programs.

    Windows XP - In Windows Explorer, move the folder Sybase in\Windows\Profiles\All Users\Start Menu\Programs\ to Documents and Settings\AllUsers \Start Menu\Programs.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    21/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 21 of 86 Effective date: December 2008

    Siemens Confidential

    Figure 2. From Location

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    22/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 22 of 86 Effective date: December 2008

    Siemens Confidential

    Figure 3. To Location

    3.4 Configuring Sybase Settings - Windows 2000/Windows XP

    1. From the Windows Start menu, select Programs->Sybase->dsedit.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    23/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 23 of 86 Effective date: December 2008

    Siemens Confidential

    2. On the Select Directory Service window, the DS name Interfaces Driver option should behighlighted. Select OK.

    3. On the dsedit - DSEDIT1 - InterfacesDriver window, select Server Object->Add.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    24/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 24 of 86 Effective date: December 2008

    Siemens Confidential

    4. On the Input Server Name window, type the Hostname of your server (for example,RADSERVER). Click OK.

    5. On the dsedit - DSEDIT1 - InterfacesDriver window, highlight the Server Name justentered and also highlight the Server Address field. Select Server Object->ModifyAttribute.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    25/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 25 of 86 Effective date: December 2008

    Siemens Confidential

    6. On the Network Address Attribute window, click Add.

    7. On the Input Network Address for Protocol window, select TCP from the Protocol drop-down box. Tab to the Network Address box and enter the Hostname and the PortNumber of your server (for example, LABPERF4100,10000). Click OK.

    Note: The Server Name is not case sensitive. However, for consistency use the samecase (upper or lower) here, and below when valuing the Hostname, Data SourceName, and Description on the ODBC Sybase ASE Setup window.

    8. On the Network Address Attribute Window, click OK. The dsedit - DSEDIT1 -InterfacesDriver window is displayed.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    26/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 26 of 86 Effective date: December 2008

    Siemens Confidential

    9. Select Server Object->Ping Server. The Ping window is displayed.

    10. With the server highlighted, click Ping. A dsedit window is displayed with a messageconfirming a successful open and close connection. Click OK. If you have a failure,confirm your server address attributes window and select Ping again. If you continue tohave a problem, contact your IS Administrator. Close the Dsedit windows.

    11. From the Windows Start menu, select Programs->Sybase->ODBC Driver->ODBCAdministrator.

    12. On the ODBC Data Source Administrator window, select the System DSN tab and clickAdd.

    13. On the Create New Data Source window, highlight Sybase ASE ODBC Driver (the optionfor Sybase 12) and click Finish.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    27/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 27 of 86 Effective date: December 2008

    Siemens Confidential

    14. On the ODBC Sybase ASE Setup window, enter the Hostname of your server (forexample, RADSERVER) in the Data Source Name, Description, and Server name fields.The Database Name is an optional field and should be left blank.

    Note: In the example below, RADSERVER is the Hostname, Data Source Name, andDescription. This is site/hospital specific. Consult the Information Servicescontact at your entity for this information.Data Source Name: RADSERVER [your hospital-specific server name should gohere]Description: RADSERVER [your hospital-specific server name should go here]

    Server Name: RADSERVER [your hospital-specific server name should go here]Database Name: leave this field blankThis information is not case sensitive. However, ensure that the Hostname thatyou entered in the Network Address Window above is the same as the Hostnameyou are entering in the ODBC Sybase Driver Setup window.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    28/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 28 of 86 Effective date: December 2008

    Siemens Confidential

    15. Click OK. The ODBC Data Source Administrator window is displayed.

    16. Click OK. The installation of the Sybase OpenClient, Sybase ODBC and MS AccessRuntime is now completed.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    29/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 29 of 86 Effective date: December 2008

    Siemens Confidential

    17. Ensure that your dmq.ini, smsrms,ini and host files contain the correct information(Hostname, DMQ Endpoint, RMS_Sysid, RMS_Version, and Host address).

    DMQ.INI file found in C:\Program Files\Siemens\CLN\syngoWorkflow\.

    In the [Default Server] section, verify that the Hostname and Endpoint settingsare correct.

    The Hostname is the network name of the syngoWorkflow SLR server and mustbe defined either on a Name Server somewhere on the network, or in the localHOSTS file (see below).

    The Endpoint is the BEA Message Queue endpoint defined in the configuration file located on the syngoWorkflow SLR server. Thestandard DMQ Endpoints for a system are: 11100 for PRD, 11105 for TRN, and11110 for TST. Contact your install or support programmer to verify this value, orif you have a non-standard system id, such as in a multi-entity situation.

    SMSRMS.INI file found in C:\Program Files\Siemens\CLN\syngoWorkflow\. If you need to edit this file, make sure the "read only"

    attribute is not on by checking the file's properties.

    In the [Host] section, verify that the RMS_Sysid and RMS_Version settings arecorrect.

    The RMS_Sysid should match the configuration file name(located on the syngoWorkflow SLR server). For the Production environment,this will be PRD, for Training environment it will be TRN, and for the Testenvironment it will be TST. Note: Your values may be different if your

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    30/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 30 of 86 Effective date: December 2008

    Siemens Confidential

    organization has decided to label your environments differently (as is sometimesthe case in a multi-entity environment).

    The RMS_Version should be V30.2

    18. Once the setup is complete, a menu option is created at Start->Programs->Siemenssyngo Workflow->syngo Workflow Production. Select this option to start the

    application. When you log into syngoWorkflow SLR, the Security Warning screen isdisplayed.

    19. Select the check box next to Always trust content from Siemens Medical SolutionsHealth Services Corporation, and click Yes.

    3.5 Formatter Server- Windows 2000/ Windows XP

    3.5.1 Requirements

    Number of Format Servers Required

    A Minimum of (2) Format Servers are required.

    Server Requirement - The first (2) Format Servers can accommodate a total of 400,000annual exams. Each additional Format Server can accommodate 200,000 exams.

    Required Hardware

    500MHz Pentium III (or 100% Pentium compatible)

    1.44 MB 3.5 inch diskette drive

    CD-ROM Drive

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    31/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 31 of 86 Effective date: December 2008

    Siemens Confidential

    10/100 MB network interface card supported by the network

    SVGA Local Bus video adapter or Windows accelerated SVGA graphics adapter

    SVGA Monitor

    Keyboard and Mouse

    PC Memory Requirement: 256 MB, Server Requirement: 1GB

    2.0 GB Disk Space

    Required Software

    Windows 2000 Professional or Windows XP Professional

    Microsoft Office 2003 or Microsoft Office 2007

    Printer Drivers as required by printer manufacturer

    PC Anywhere

    3.5.2 Installing Print Formatter Server for syngoWorkflow SLR

    Note: All Windows programs should be shut down before running this install.

    1. If you are installing directly from the delivered CD, load the CD in the CD-ROM drive.Otherwise, map a drive to the network location where the install software is located.

    2. From the Start menu, select Run.

    If you are running the install directly from the delivered CD, change directory ormap a drive to the CD workstation install software and run setup.exe.[CD:] Desktop\Rad300\install\setup.exe

    If you have copied the install software onto a network drive from the CD, runsetup.exe from that location.

    The Syngo Workflow- Install Shield Wizard screen is displayed.

    3. On the Setup Type screen, select the Custom radio button. The Destination Folder isdisplayed at the bottom of the screen. Allow the install to take place in the default

    directory, (C:\Program Files\Siemens\CLN\syngoWorkflow V30.x). If you do not havesufficient space on your [Local Drive:] drive to install the application, use the Browseoption to select an alternate drive and directory.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    32/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 32 of 86 Effective date: December 2008

    Siemens Confidential

    4. Click Next. The Select Components window is displayed.

    5. Select the Print Formatter checkbox.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    33/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 33 of 86 Effective date: December 2008

    Siemens Confidential

    Note: This will install the Print Formatter in all environments. In the above example,only the Production Print Formatter was selected. You can individually selectwhich environment to install, based on your individual needs.

    6. Click Next.

    7. The Hostname Information screen is displayed. At the prompt, enter the Hostname of

    your server. In this example, Hostname is RADSERVER. Enter your specific Hostnamehere.

    8. Select Next. The Start Copying Files window is displayed.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    34/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 34 of 86 Effective date: December 2008

    Siemens Confidential

    9. If you want to review or change any settings, click Back. If you are satisfied with thesettings, click Next to begin copying files. The Setup Status window is displayed andsyngoWorkflow SLR begins copying files to the workstation. This step may take severalmoments to complete.

    10. When files have been copied, the Setup Complete window is displayed. The defaults forYes, place a shortcut on the Desktop and Yes, I Want To Restart My Computer will beselected. Remove any disks from their drives and click Finish to complete the CustomInstall.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    35/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 35 of 86 Effective date: December 2008

    Siemens Confidential

    11. Once the setup is complete, a menu option is created at Start->Programs->Siemenssyngo Workflow->syngo Workflow Production. Select this option to start theapplication. When you log into syngoWorkflow SLR, the Security Warning screen isdisplayed.

    12. Select the check box next to Always trust content from Siemens Medical SolutionsHealth Services Corporation, and click Yes.

    13. If the print formatter does not start up automatically on boot-up or returns an error

    message, see syngo Workflow SLR Access or Print Formatter Troubleshooting to resolvethe issue.

    Note: Siemens recommends that all customers secure their Format Servers with some type ofVirus Scan software. The customer should load the Virus Scan software in the samecapacity as they would for any other PC that has been deployed, and is operational withintheir facility. However, Siemens does not make any specific recommendation to a specificversion or type of virus scan software to use. This decision is up to the customer, andtheir Information Technology department. Should the customer encounter any problemswith their specific virus scan software they use, it is the customer's responsibility toresolve the issue.

    3.5.3 Word Setting for Print Formatter

    To prevent the comment balloons from printing on Result documents, you must make thefollowing change in Microsoft Word.1. Select Tool->Options->Track Changes.

    2. In the Balloons section, set Use Balloons (Print and Web Layout) to Never.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    36/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 36 of 86 Effective date: December 2008

    Siemens Confidential

    3.5.4 Installing TweakUIFor Windows 2000/ Windows XP, TweakUI can be installed on the Print Formatter to allow thesystem to reboot without prompting for a username and password. This will make it possible forthe formatter to reboot automatically.To install TweakUI:1. Go to the Microsoft web site at http://www.microsoft.com/.

    2. At the Search Microsoft.com for: field, enter Installing TweakUI for Windows (and theOperating System-Example NT, 2000 or XP).

    3. Follow the directions to download the software.

    3.5.5 Required Microsoft Word XP Change SoDocuments Do Not Get CorruptedMicrosoft Word introduced a new RTF command tag series, Revision Save ID (RSID), to allowimproved identification of document changes. However, the initialization code may not properlyhandle RSID tags that appear in the SQL text.To correct this problem, either prevent the insertion of the RSID tags or force Word to removepre-existing RSID tags by including a tagged document into a document that has been opened bywork and configured to NOT use RSID tags.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    37/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 37 of 86 Effective date: December 2008

    Siemens Confidential

    To prevent the initialization problem, alter the default configuration of Microsoft Word as follows:Note: You must have a document open to access Tool->Options (can use a new blank

    document).

    1. In Microsoft Word, go to Tools->Options and click the Security tab.

    2. In the Privacy Options section, uncheck the Store random number to improve mergeaccuracy checkbox.

    3. If you have a document corrupted with RSID tags, you can remove them by doing thefollowing:

    a. Save the corrupted document as an RTF file on the local PC.

    b. Open a new blank document.

    c. Follow step 1 above to go to Tools->Options-> and select the Security tab.Uncheck the Store random number to improve merge accuracy checkbox.Note: If you have already turned off the option, you can skip the next 2 steps.

    d. Close the blank document.

    e. Open a new blank document.

    f. Use Insert->File to insert the corrupted RTF file you saved in step a.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    38/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 38 of 86 Effective date: December 2008

    Siemens Confidential

    g. Save the inserted document as an RTF file under a new name.

    h. Include the cleaned RTF file into syngoWorkflow Word Processing.

    3.6 Installing IntegriSign Desktop

    You can capture electronic signatures (for example, patients, witnesses, guarantors, employees,etc.) in Interactive Documents using IntegriSign Signature software. In order to use this feature,you must purchase the ePad Ink device from Interlink Electronics, which contains the ePad withIntegriSign Signature Software Suite CD. If you plan to also print these documents, use thefollowing procedure to load the required software onto the Print Formatters so that the documentscontaining the signatures can be printed.1. Place the ePad with IntegriSign Signature Software Suite CD in the CD-ROM drive.

    2. From Windows Explorer, select your CD drive.

    3. Double-click the file autorun.exe. The IntegriSign screen is displayed.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    39/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 39 of 86 Effective date: December 2008

    Siemens Confidential

    4. Select the option Install IntegriSign Desktop. The IntegriSign Desktop -Install ShieldWizard screen is displayed.

    5. Click Next to display the IntegriSign - License Agreement screen.

    6. Read agreement and click Yes to accept terms. The IntegriSign - Customer Informationscreen is displayed.

    7. Click Next. The IntegriSign - Select Components screen is displayed.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    40/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 40 of 86 Effective date: December 2008

    Siemens Confidential

    8. Select all options listed and click Next. The IntegriSign - Choose Destination Locationscreen is displayed.

    9. The default location is listed under Destination Folder. Accept the default path and clickNext. The IntegriSign - Select Program Folder screen is displayed.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    41/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 41 of 86 Effective date: December 2008

    Siemens Confidential

    10. Select Interlink Electronics and click Next. The IntegriSign software installation willbegin. The install percent complete is displayed on the Setup Status screen.

    11. When the installation is complete, the IntegriSign - Install Shield Wizard Complete screenis displayed. Select the Yes option and click Finish. Your PC will reboot, and theinstallation will be complete.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    42/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 42 of 86 Effective date: December 2008

    Siemens Confidential

    3.7 Installing Powerscribe SDK

    syngoWorkflow SLR has an embedded third-party speech recognition system to translate speechto text for the creation of radiology result reports. If you plan to use Powerscribe SpeechRecognition SDK, install the software as follows:1. Shut down all Windows programs before running the install. This includes MacAfee, Print

    Now, or any program that is running in the background. Check the task bar at the bottomright side of your computer screen for icons that indicate programs are running.

    2. Verify that you are currently logged on to the PC with Local Administrator privileges. Ifunsure, you can check by selecting Start -> Settings -> Control Panel -> Users andPasswords (or User Accounts). Find the current User ID you are using and verify youare listed as a member in the Group Administrators.

    3. If you are installing directly from the delivered CD, load the CD in the CD-ROM drive.Otherwise, map a drive to the network location where the install software is located. Tomap a drive:

    a. From your administrator, get the location of the syngoWorkflow SLR install files.

    b. Open My Network Places. Select Tools->Map Network Drive. Choose a driveletter from the drop-down. Click Browse and locate the folder containing theinstall files. Click Finish. This drive will now display in My Computer on thedesktop under the letter you assigned.

    4. From the Windows Start menu, select Run.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    43/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 43 of 86 Effective date: December 2008

    Siemens Confidential

    a. If you are running the install directly from the delivered CD, type the following atthe prompt:[CD:]\Desktop\RAD300\Install\Setup.exe

    b. If you have copied the install software onto a network drive from the CD, run theSetup.exe from that location:[Mapped_Drive:]\Desktop\RAD300\Install\Setup.exe

    Note: If you get an error stating that the directory path is not accessible andthat it has been removed or deleted, verify that the location was enteredcorrectly. If the problem persists, go to Windows Explorer, follow the pathto Setup.exe, and double-click to execute.

    c. The syngoWorkflow Setup - Install Shield Wizard window is displayed.

    5. The Setup Type window is displayed. Select the Custom radio button. The DestinationFolder is displayed at the bottom of the screen. Allow the install to take place in thedefault directory, (C:\Program Files\Siemens\CLN\syngo Workflow V30.2). If you do nothave sufficient space on your [Local Drive:] drive to install the application, use theBrowse option to select an alternate drive and directory.

    6. Click Next. The Select Components window is displayed.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    44/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 44 of 86 Effective date: December 2008

    Siemens Confidential

    7. Select the PowerscribeSDK Install Files checkbox, and click Next.

    8. The Destination Folder screen is displayed. The default path to Install DictaphonePowerscribeSDK Components is: C:\WINNT\system32\. Click Next to accept the defaultpath.

    9. On the Install Shield Wizard Completed screen, click Finish complete the install ofPowerscribe SDK.

    10. The Syngo Workflow Setup Complete screen is displayed. Click Finish.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    45/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 45 of 86 Effective date: December 2008

    Siemens Confidential

    4. Typical Install - Windows 2000/Window XP

    A typical install will install the Production Environment ONLY for a syngoWorkflow SLRworkstation. You may want to do this if you need to install the syngoWorkflow SLR software only,but the user does not have a need to access the UDT functions, such as a Radiologists

    Workstation.1. Shut down all Windows programs before running the install. This includes MacAfee, Print

    Now, or any program that is running in the background. Check the task bar at the bottomright side of your computer screen for icons that indicate programs are running.

    2. Verify that you are currently logged on to the PC with Local Administrator privileges. Ifunsure, you can check by selecting Start -> Settings -> Control Panel -> Users andPasswords (or User Accounts). Find the current User ID you are using and verify youare listed as a member in the Group Administrators.

    3. If you are installing directly from the delivered CD, load the CD in the CD-ROM drive.Otherwise, map a drive to the network location where the install software is located. Tomap a drive:

    a. From your administrator, get the location of the syngoWorkflow SLR install files.

    b. Open My Network Places. Select Tools->Map Network Drive. Choose a driveletter from the drop-down. Click Browse and locate the folder containing theinstall files. Click Finish. This drive will now display in My Computer on thedesktop under the letter you assigned.

    4. From the Windows Start menu, select Run.

    a. If you are running the install directly from the delivered CD, type the following atthe prompt:[CD:]\Desktop\RAD300\Install\Setup.exe

    b. If you have copied the install software onto a network drive from the CD, run theSetup.exe from that location:[Mapped_Drive:]\Desktop\RAD300\Install\Setup.exe

    Note: If you get an error stating that the directory path is not accessible andthat it has been removed or deleted, verify that the location was enteredcorrectly. If the problem persists, go to Windows Explorer, follow the pathto Setup.exe, and double-click to execute.

    5. The set-up application begins and the Welcome window is displayed. Click Next.

    6. The syngoWorkflow SLR Setup begins and the syngoWorkflow - Install Shield Wizardwindow is displayed.

    7. If previous versions of Novius/syngoRadiology are detected on your computer, thefollowing message box is displayed.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    46/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 46 of 86 Effective date: December 2008

    Siemens Confidential

    8. Select the Yes radio button and click Next. The Release Upgrade detection will removeALL currently installed versions of Novius/syngoWorkflow SLR and in its place, install thenew syngoWorkflow SLR release. The Setup Type window is displayed.

    9. Select the Typical radio button. The Destination Folder is displayed at the bottom of thescreen. Allow the install to take place in the default directory, (C:\ProgramFiles\Siemens\CLN\syngo Workflow V30.2). If you do not have sufficient space on your

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    47/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 47 of 86 Effective date: December 2008

    Siemens Confidential

    [Local Drive:] drive to install the application, use the Browse option to select an alternatedrive and directory.

    10. Click Next. The Hostname Information screen is displayed.

    11. At the prompt, enter the Hostname of your server.

    Note: In this example, Hostname is RADSERVER. Enter your specific Hostname at the

    prompt.12. Click Next. The Start Copying Files screen is displayed

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    48/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 48 of 86 Effective date: December 2008

    Siemens Confidential

    13. If you want to review or change any settings, click Back. If you are satisfied with thesettings, click Next to begin copying files. The Setup Status window is displayed andsyngoWorkflow SLR begins copying files to the workstation. This step may take severalmoments to complete.

    14. When files have been copied, the Setup Complete window is displayed. The defaults forYes, place a shortcut on the Desktop and Yes, I Want To Restart My Computer will be

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    49/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 49 of 86 Effective date: December 2008

    Siemens Confidential

    selected. Remove any disks from their drives and click Finish to complete the TypicalInstall.

    15. Ensure that your dmq.ini, smsrms,ini and host files contain the correct information(Hostname, DMQ Endpoint, RMS_Sysid, RMS_Version, and Host address).

    DMQ.INI file found in C:\Program Files\Siemens\CLN\syngoWorkflow\.

    In the [Default Server] section, verify that the Hostname and Endpoint settingsare correct.

    The Hostname is the network name of the syngoWorkflow SLR server and mustbe defined either on a Name Server somewhere on the network, or in the localHOSTS file (see below).

    The Endpoint is the BEA Message Queue endpoint defined in the configuration file located on the syngoWorkflow SLR server. Thestandard DMQ Endpoints for a system are: 11100 for PRD, 11105 for TRN, and

    11110 for TST. Contact your install or support programmer to verify this value, orif you have a non-standard system id, such as in a multi-entity situation.

    SMSRMS.INI file found in C:\Program Files\Siemens\CLN\syngoWorkflow\. If you need to edit this file, make sure the "read only"attribute is not on by checking the file's properties.

    In the [Host] section, verify that the RMS_Sysid and RMS_Version settings arecorrect.

    The RMS_Sysid should match the configuration file name(located on the syngoWorkflow SLR server). For the Production environment,this will be PRD, for Training environment it will be TRN, and for the Testenvironment it will be TST. Note: Your values may be different if your

    organization has decided to label your environments differently (as is sometimesthe case in a multi-entity environment).

    The RMS_Version should be V30.2.

    16. Once the setup is complete, a menu option is created at Start->Programs->Siemenssyngo Workflow->syngo Workflow Production. Select this option to start theapplication. When you log into syngoWorkflow SLR, the Security Warning screen isdisplayed.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    50/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 50 of 86 Effective date: December 2008

    Siemens Confidential

    17. Select the check box next to Always trust content from Siemens Medical SolutionsHealth Services Corporation, and click Yes.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    51/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 51 of 86 Effective date: December 2008

    Siemens Confidential

    5. Service Pack Update Detection

    Service pack update detection is only used during the upgrade from one service pack to another.1. When you log into syngoWorkflow SLR after a service pack (SP) upgrade, the Incorrect

    Software message box will display.

    2. If Yes is selected, the application automatically searches for the source at the originalinstallation setup (the path is stored in the SMSRMS.ini file), and updates the workstationwith the new files.

    Note: The SP installation software should be placed in the same location from whichthe original install was performed. If Yes is selected and the installation source isnot located automatically, the application will prompt you to browse for theinstallation. In the CHOOSE PATH window, select the location of the newsoftware. When the folder that contains the new software modules is selected,the box labeled Found setup.exe will be checked. Select OK.

    a. The Install Shield Wizard box is displayed briefly followed by the syngoWorkflowUpdate Program screen. Click Next. The Setup Status screen is displayed andupdate process begins.

    b. For Windows 2000/Windows XP. When the update is complete, the Setup Status

    screen closes, and you will be ready to log into syngoWorkflow SLR.

    3. If No is selected, the update will not occur and you will NOT be able to log into syngoWorkflow SLR until the update occurs.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    52/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 52 of 86 Effective date: December 2008

    Siemens Confidential

    6. Release Update Detection

    Release update detection is only used during the upgrade from one software release to another.For example, a customer at a software level of V27.0 SpX, V27.1 SpX, V27.2 SpX , V28.0SpX,V28.1SpX, V30.0 SpX who is upgrading to a software level V30B SpX is performing a release

    update. SpX equals all service pack levels for the Release Level.Release updates eliminate the need for you to have to manually uninstall an old release from thePC prior to loading the new software. The Install Shield mechanism completes this task for you.In a release upgrade situation, you sign on to syngoWorkflow SLR as usual, and the PC isupgraded with the new executables/files from the V30B SPx release. (SPx equals service packlevel greater than SP0).1. When you log into syngoWorkflow SLR after a Release upgrade, the Incorrect Software

    message box will display.

    2. If Yes is selected, the application will automatically search for the source at the originalinstallation setup (the path is stored in the SMSRMS.ini file), and updates the workstationwith the new files.

    Note: The SP installation software should be placed in the same location from whichthe original install was performed. If Yes is selected and the installation source isnot located automatically then the application will prompt you to browse for theinstallation. In the CHOOSE PATH window, choose the location of the newsoftware. When you have chosen the folder that contains the new software

    modules, the box labeled "Found setup.exe" will be checked. Select OK.

    a. The Install Shield Wizard box is displayed briefly. If previous versions ofRadiology are detected on your computer, you will be prompted with the syngoWorkflow Setup message box asking whether you want to remove the version(s)from your computer.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    53/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 53 of 86 Effective date: December 2008

    Siemens Confidential

    b. Select the Yes option, then click Next. The Release Upgrade detection willremove ALL currently installed versions of Novius/syngoWorkflow SLR and in itsplace, install the new syngoWorkflow SLR Release.

    3. If No is selected, the update will not occur, and you will NOT be able to login tosyngoWorkflow SLR until the update occurs.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    54/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 54 of 86 Effective date: December 2008

    Siemens Confidential

    7. Install Portal Radiologist Client Software

    7.1 General Requirements

    Portal Radiologist requires a monitor resolution setting of 1280 x 1024.

    Anti virus software is strongly recommended.

    Refer to the current Schedule 1 document for the syngoWorkflow SLR workstationrequired hardware/software.

    7.2 Install Portal Radiologist Client Software

    Use the following procedure to load the Portal client software on a client workstation.1. On client workstation, log into Windows as the administrator

    2. Start Internet Explorer and select Tools=>Internet Options.

    3. On Internet Properties screen, click Security tab, then Custom Level button.

    4. On the Security Settings screen, make sure following options are set as indicated:

    Miscellaneous

    - Use Pop-up Blocker - Disable

    ActiveX controls and Plug-ins

    - Download signed ActiveX controls - Prompt or Enabled

    - Initialize and script ActiveX controls not marked as safe - Prompt orEnabled

    - Run ActiveX controls and plug-ins - Enabled

    - Script ActiveX controls marked as safe - Enabled

    5. Click OK on Security Settings and Internet Options screens.

    6. In Internet Explorer, enter the URL address of your specific webserver and website inhttp://webserver/website format. The Siemens syngoWorkflow SLR Portal Radiologistscreen is displayed.

    7. Click Install. The File Download - Security Warning screen is displayed.

    8. Click Run.

    9. If Internet Explorer - Security Warning screen displays, click Run.

    10. A SyngoClient screen similar to the one shown below is displayed if the PC does nothave the proper version of the .NET Framework installed.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    55/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 55 of 86 Effective date: December 2008

    Siemens Confidential

    Figure 4. NET Framework Installation Request

    If you select No, the installation will stop.

    If you select Yes, you will be directed to the Microsoft .NET FrameworkDeveloper Center.

    a. Select the option to download .NET Framework for a 32-bit machine.

    b. Follow the instructions to download and install .NET Framework.

    c. When installation is complete, the SyngoClient - Confirm Installationscreen is displayed.

    11. On the Confirm Installation screen, click Next to begin installation of the SyngoClient.The Installing syngoWorkflow Client screen is displayed with a progress bar.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    56/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 56 of 86 Effective date: December 2008

    Siemens Confidential

    Figure 5. Confirm Installation

    12. When finished, the Installation Complete screen is displayed. Do one of the following:

    Note: If you are using a third-party dictation system, do NOT install Powerscribe SDK.

    a. If the PC does not have Powerscribe SDK installed, the option to install it isdisplayed. Click Powerscribe SDK to begin the installation. The DictaphonePowerscribeSDK Components - InstallShield Wizard screen is displayed. Followthe instructions for installation of Powerscribe SDK.

    b. Click Close if you do not want to install Powerscribe SDK.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    57/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 57 of 86 Effective date: December 2008

    Siemens Confidential

    Figure 6. Installation Complete

    13. On the Siemens syngoWorkflow SLR screen, click Continue. The syngoWorkflow SLRPortal Radiologist screen is displayed.

    14. Enter a valid User Name and Password to log onto the Portal. See syngo Workflow SLRPortal Radiologist User's Guidefor instructions on using Portal Radiologist.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    58/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 58 of 86 Effective date: December 2008

    Siemens Confidential

    8. Install IntegriSign (ePad) Signature Software

    You can capture a Patient Signature electronically using IntegriSign Signature software. To usethis function, customers must purchase the ePad Ink device from Interlink Electronics, whichcomes with the ePad with IntegriSign Signature Software Suite CD. Every PC that will be using

    an ePad must install the ePad Drivers from the CD.Note: All PCs which plan to use the Signature Software must also load the IntegriSign Desktop

    software. Please follow the install instructions found in Installing IntegriSign Desktop.

    1. Place the ePad with IntegriSign Signature Software Suite CD in the CD-ROM drive.

    2. From Windows Explorer, select your CD drive.

    3. Double-click the file autorun.exe. The IntegriSign screen is displayed.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    59/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 59 of 86 Effective date: December 2008

    Siemens Confidential

    4. Select the option Install ePad Drivers (Required First). The Interlink Electronics ePadInstall Shield Wizard screen is displayed.

    5. Click Next to display the Interlink Electronics - Customer Information screen.

    6. Enter the appropriate User Name and Company Name. Select the Anyone who usesthis computer (all users) radio button, then click Next.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    60/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 60 of 86 Effective date: December 2008

    Siemens Confidential

    7. Select all options listed and click Next. The Interlink Electronics - Setup Type screen isdisplayed.

    8. Select the Typical radio button and click Next. The Interlink Electronics ePad - ChooseDestination Location screen is displayed.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    61/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 61 of 86 Effective date: December 2008

    Siemens Confidential

    9. The default location is listed under Destination Folder. Accept the default path and clickNext. The Interlink Electronics ePad - Select Components screen is displayed.

    10. Select ePad-ink USB and click Next. The Interlink Electronics ePad - Select Featuresscreen is displayed.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    62/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 62 of 86 Effective date: December 2008

    Siemens Confidential

    11. Click Next. The Interlink Electronics ePad - Start Copying Files screen is displayed.

    12. Click Next. The install will begin. When the installation is complete the IntegriSign - InstallShield Wizard Complete screen is displayed.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    63/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 63 of 86 Effective date: December 2008

    Siemens Confidential

    13. Click Finish to complete the installation.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    64/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 64 of 86 Effective date: December 2008

    Siemens Confidential

    9. Word Viewer 2003

    1. Place the installation CD in the CD-ROM drive.

    2. From Windows Explorer, select your CD drive.

    3. Select the Word Viewer Folder (DESKTOP\RAD300\WordViewer2003).

    4. Double-click the file wdviewer.exe. The Microsoft Office Word Viewer 2003 Setup screenis displayed.

    5. Click checkbox to select the I accept the terms in the License Agreement option.

    6. Click Next . The Install Word View screen is displayed.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    65/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 65 of 86 Effective date: December 2008

    Siemens Confidential

    Note: If you wish to install the files in another location, select Browse and find thedesired location.

    7. The default install path C:\Program Files\Microsoft Office is displayed. Click Install.The Install of Word Viewer 2003 will begin.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    66/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 66 of 86 Effective date: December 2008

    Siemens Confidential

    8. When installation is complete, the following dialog is displayed. Click OK to complete theinstallation.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    67/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 67 of 86 Effective date: December 2008

    Siemens Confidential

    10. Installing Printers and Print Drivers

    You must set up each printer with spooling disabled, and the output of each printer must bedirected to a file rather than to an actual local or network printer queue. See the detailed

    instructions below for how to do this with Windows 2000/Windows XP.Caution: Be sure to use the use the most current print drivers for your printer. Do notassume that the print drivers delivered with your hospital's printers are the latestprint drivers. For proper operation, the print driver used must be capable ofdirecting the output to a file on the PC's local hard drive.

    Note: Please consult the Customer Only Website/syngoWorkflow / Device Manual forinformation about printers.

    10.1 Printers

    The most current model Printer uses a Windows-based print driver that may need additionaldrivers to support Barcode printing. For this printer, and other Windows-based printers that needthe barcode driver, follow these generic steps. Consult with your Installation Consultant for themost current printer.1. From the Windows Start menu, select Settings->Printers .

    2. Double-click on the Add Printer option and use the Install Wizard to install the printer.

    3. On the Add Printer Wizard Welcome window, click Next.

    4. On the Local or Network Printer window, select the Local Printer radio button, anddeselect the Automatically detect and install my Plug and Play Printer checkbox. ClickNext.

    5. On the Select the Printer Port window, select FILE from the Use the following port list

    box. Click Next.

    6. On the Install Printer Software window, select the manufacturer (for example, HP) andprinter from the corresponding list boxes. Click Next.

    7. On the Name Your Printer window, enter the name for the Printer. Click Next.

    Note: This name will be used when defining the print drivers in the UDT Maintenancefunction - Print Destination Definition. (The names must match exactly.)

    8. On the Printer Sharing window, select the Do not Share this Printer radio button. ClickNext.

    9. On the Print Test Page window, check the No radio button. Click Next.

    10. Click Finish to complete the Add Printer Wizard process.

    11. Right-click on the printer you just installed and select Properties.

    12. Click the Fonts tab. Make sure the BARCODES AND MORE entry is highlighted in thelist of cartridges. If BARCODES AND MORE is not displayed in theCARTRIDGES/SIMMS box, you need to install the BARCODES AND MORE fonts:

    a. Select FONTS.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    68/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 68 of 86 Effective date: December 2008

    Siemens Confidential

    b. Select INSTALL PRINTER FONTS.

    c. Click Add Fonts. The system prompts you for the font directory.

    d. Insert the FONTS diskette into Drive A and enter A:\ or enter the drive where thefonts have been downloaded. The proper diskettes may be delivered with theprinter or downloaded from the manufacturer. Check your printer manual.

    e. Click OK.

    f. Highlight HP:BARCODES AND MORE and select ADD, then click OK when thesystem displays the directory where the fonts will be located.

    g. Select EXIT.

    h. Be sure that HP:BARCODES AND MORE is still highlighted and selectCONNECT.

    i. If you are installing the printer on a Word Processing Formatter PC, click theDetails tab. Click Spool Settings. Make sure the Print Directly to the printeroption is selected. If not, select it. You do not want to use spooling on the Word

    Processing Formatter PCs.

    j. ClickOK.

    10.2 Bar Codes for Printer

    To print bar codes on your printer, you must install a bar code DIMM into a memory slot in theprinter. Please refer to the documentation provided by the manufacturer of your printer. To installthe driver on the PC:1. Insert the Printer CD into CD-ROM drive.

    2. In Windows Explorer, select the CD-ROM drive.

    3. Double-click on Setup.exe from the top-level directory.

    4. Click Continue on the Printer CD-ROM Install Utility screen.

    5. Click Next on the Welcome screen.

    6. Close all open applications and click Next on the Close Other Applications screen.

    7. Click Next on the Download Latest Files from the Internet screen.

    8. Select and click the Custom Installation bullet-button, then click Next on the Method ofInstallation screen.

    9. Select the Printer from the printer list, then click Next on the Select Printer screen.

    10. Select a port from the port list, then click Next on the Select Port screen.

    11. Make sure the following components are selected (checked):

    Printer Driver component Printer

    FontSmart component

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    69/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 69 of 86 Effective date: December 2008

    Siemens Confidential

    Printer ReadMe component

    12. Select or deselect the other components as you desire.

    13. Click Next on the Select Components screen.

    14. Click Exit on the Setup Complete screen.

    10.3 Install and Select Bar-Code Fonts

    1. From the Windows Start menu, select Settings->Control Panel, then select Printers.

    2. Right-click the Printer from the printer list and select Properties.

    3. Click the Device Settings tab.

    4. In the installed Font Cartridge section, click Slot #1 and select HP: Bar Code and More.

    10.4 Port Setup Instructions for Windows 2000/WindowsXP

    If you are installing the driver on a Print Formatter PC, you need to change the port to print to alocal print file.1. From the Windows Start menu, select Settings->Printers .

    2. Right-click on the printer you just installed and select Properties.

    3. Click the Advanced tab. Make sure the Print Directly to the printer option is selected. Ifnot, select it. You do not want to use spooling on the Word Processing Formatter PCs.

    4. Click the Ports tab.

    5. In the Print to the following port list box, select C:\Program Files\Siemens\CLN\syngoWorkflow\Production\PRNTSRV.PRN. If you installed the RMS PC software on a differentdrive or directory, use the drive letter and directory you used for the install. If this doesnot show up in the list, you will have to add it. You will need to add the port once performatter PC.

    6. To add the port:

    a. Click Add Port.

    b. Click Local Port in the list box

    c. Click New Port.

    d. At the Enter a port name prompt, enter C:\Program Files\Siemens\Cln\syngoWorkflow\Production\PRNTSRV.PRN

    e. Click OK.

    f. Click Close.

    7. Click Close to close the Properties window.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    70/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 70 of 86 Effective date: December 2008

    Siemens Confidential

    10.5 Biscom Fax Processor

    The Biscom Fax Processor accepts a PCL file as input and faxes it to the specified phonenumber. You should use a printer driver which produces PCL output. The FAXCOM 3000supports PCL up to level 4, so you should use a driver that produces PCL 3 or 4. You can use aHP LaserJet 4M driver, for example. If the documents being faxed are also designed to be printed

    to a laser printer, such as an HP LaserJet, you may be able to use the same driver for the faxprocessor as you use for the printer. The procedure of installing the driver for a Biscom FaxProcessor is the same as if you were installing it for use by a printer.

    10.6 Intermec Label Printer

    The Intermec Label Printer uses a Windows printer driver. You should use the standard Generic /Text Only driver provided with Windows when using the Intermec Label Printer.1. From the Windows Start menu, select Settings->Printers .

    2. Double-click on the Add Printer option and use the Install Wizard to install theGeneric/Text Only printer.

    3. Select the Local printer option.

    4. Select any port. Leave the printer name as Generic/Text Only.

    5. Select Generic as the manufacturer and Generic/Text Only as the printer.

    6. Select No for using this printer as your Windows default printer.

    7. ClickNext.

    8. Select No for the option to print a test page.

    9. Click Next.

    10. Select Finish on the Completing the Add Printer Wizard screen.

    11. In the Printers window, right-click on the printer you just installed (Generic/Text Only) andselect Properties.

    a. Select the Ports tab

    b. In the Print to the following port(s) section, select the Add Port option.

    c. At the Enter a port name prompt, enter C:\Program Files\Siemens\CLN\syngoRadiology \Production\PRNTSRV.PRN

    d. Click OK.

    12. Click OK to close the Properties window.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    71/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 71 of 86 Effective date: December 2008

    Siemens Confidential

    11. Maintenance Mode

    After syngoWorkflow SLR has been installed, if you rerun the install, Install Shield will proceed inMaintenance Mode.During the install of syngoWorkflow SLR, a record of the install is placed in the following

    directory:C:\Program Files\InstallShield Installation InformationEach syngoWorkflow SLR Install Shield Package will create a subdirectory similar to {B3351260-F9FE-44F7-8898-320356D0C3C6}. The files in this directory contain the information that is usedto run different options in Maintenance Mode. If this directory is removed, the Maintenance Modewill not be triggered. Please note that this is a hidden directory and can only be viewed if the viewaccess to hidden files is granted.

    11.1 Modify

    The Modify option allows you to add or remove selected components based on what is checked(add) or unchecked (remove).1. From the Maintenance Mode window, select Modify. The Select Components window

    opens with the current installed syngoWorkflow SLR already selected (checked).

    2. If you do not wish to uninstall the already-installed components, leave them selected. Ifyou wish to remove one already-installed component, deselect it. To install a newcomponent, select it.

    3. Click Next. The install or remove will begin.

    4. When the modify is complete, you will be prompted to reboot.

    5. Click Finish to complete setup.

    Caution: If an application is not selected and is already installed, if you deselect it and runmodify, it will uninstall the application.Example: If syngoWorkflow SLR is already installed in Production and you wishto install the same in Training, when the selection window opens, Production willalready be selected. If you deselect it and select Training, then syngoWorkflowSLR will be installed in Training and removed from Production.

    11.2 Repair

    The Repair option allows you to repair or reinstall any already-installed component (for example,syngoWorkflow Production, syngoWorkflow Training or syngoWorkflow Test).You will not be prompted for the components to repair/reinstall. The application will be reinstalledback to the image that Install Shield previously installed. If any hot fixes have previously been

    applied, they will need to be reapplied.1. From the Maintenance Mode window, select Repair.

    2. Click Next. The reinstall of the application will begin. When the repair completes, you willbe prompted to reboot.

    3. Click Finish to complete the setup.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    72/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 72 of 86 Effective date: December 2008

    Siemens Confidential

    11.3 Remove

    The Remove option will remove all installed components, except Sybase and MS AccessRuntime. You will not be prompted for which components to uninstall.1. From the Maintenance Mode window, select Remove.

    2. Click Next. The removal of the application will begin. You will be prompted to confirm FileDeletion.

    3. Click OK to continue with the uninstall. When the remove completes, you will beprompted to reboot.

    4. Click Finish to complete the uninstall.

  • 8/2/2019 Syngo Workflow SLR PC Install Doc 30B

    73/86

    Siemens MED HS syngo Workflow SLR PC Install Document

    0000002311 Page 73 of 86 Effective date: December 2008

    Siemens Confidential

    12. Uninstalling syngo Workflow SLR, MS Access,Sybase