suzannah sorin, mspas, pa-c narus february 16, 2018 · suzannah sorin, mspas, pa-c robotics...

47
TROUBLESHOOTING SCENARIOS- As seen from the assistant Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Upload: others

Post on 21-Feb-2020

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

TROUBLESHOOTING SCENARIOS- As seen from the assistant

Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center

NARUS February 16, 2018

Page 2: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

Financial Disclosure

• J&J Ethicon / Google / Verb - robotics consultant

•Medtronic - robotics consultant

•CSATS - independent video reviewer

2

Page 3: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

What to do when the unexpected happens

3

phot

o cr

edit:

http

://w

ww

.bea

rhan

dsar

t.com

/wp-

cont

ent/u

ploa

ds/2

012/

05/th

e-sc

ream

.jpg

Edv

ard

Mun

ch

•We have a tendency to want to shut down, be quiet, not make waves

•That is when it is MOST important to communicate!

•The console can be a physical barrier and an emotional detachment

•Talk to the surgeon, so they feel connected to the room and to the patient

Page 4: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

Your Back-up Plan• To ensure that your are readily equipped with all of the tools needed to keep your Da

Vinci system functioning, you should have a back-up for all of the times listed below:

•ETHERNET CABLE (Xi/Si)

•CAMERA CABLE AND CAMERA HEAD – DA VINCI Si

•EMERGENCY INSTRUMENT WRENCH Si/Xi

•LIGHT GUIDE FOR DA VINCI Si SYSTEM

•BLUE FIBEROPTIC CABLE (Xi/Si DAVINCI SYSTEM)

•MONOPOLAR (GREEN CORD) & BIPOLAR (BLUE CORD) CORDS HAVE 100 USES. HAVE SEVERAL OF EACH CORD STERILE PEEL PACKED TO REPLACE IF CORD STOPS WORKING DURING CASE.

4

Page 5: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

More tips on troubleshooting

•Work in a stepwise fashion

•Don’t try changing everything all at once

•Communicate to the rest of the room your steps working though the problem

•Ongoing updates to the team if the case progresses while you continue to troubleshoot

5

Page 6: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

SCENARIO #1

You are the assistant nurse and are about to start setting up your first case for the day. You press the power ON button for your dV Xi or Si system and continue to assist your scrub nurse opening supplies for the case. A few moments later you notice that the system has not powered on. What should you do?

6

Page 7: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

SCENARIO #1 SOLUTIONS

• Check all components: Surgeon console, patient cart, and vision tower are all plugged into a working outlet. Ensure that the outlets are active and all dV components are on with orange power light and blue light at blue fiberoptic outlet site. Did you know that plugging all components into a single circuit can cause an overload? Try to plug into different circuits. Check the AC power connections

• Check the circuit breaker switches on all components on the dV system. Make sure the breaker switch is in the I position.

• On the patient cart make sure that the EPO- emergency power off red button is not depressed

• If the system is not powering up, you may need to "hard cycle" power system. To hard cycle, turn the breaker switch off (O position) on each dV component and press the EPO switch on the back of the patient cart. Turn off for 10 second and then restart the system

7

Page 8: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

SCENARIO #2

You are about to dock the patient cart when you notice that you have a lot of resistance to drive the cart forward. You check the cart and notice the small door in front of the cart is open, so you try to close it but the door won't close. What should you do?

8

Page 9: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

SCENARIO #2 SOLUTIONS

• Check the red lever and ensure that the lever is not in the manual mode. Turn the lever to the left to the powered steering mode. The handle may not have fully engaged power mode

• Do not force the door closed if its in manual mode

• Ensure there are not any cords or anything

obstructing the door or the feet under the patient cart.

• The patient cart weighs 1850 lbs. Refrain from

placing cart on sloped surfaces.

• When powering up the system, don’t try to drive the cart or move the arms until the system boot-up sequence is complete.

9

Page 10: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

SCENARIO #3

You are docking the patient cart and as you drive forward with the patient cart you notice "Error Code 23" appears on the touch screen of the patient cart, seconds later the code changes to "Error 252" and the screen reads "unrecoverable fault". What should you do to fix this?

10

Page 11: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

SCENARIO #3 SOLUTIONS• Check the blue fiberoptic cable connections at the patient cart and the vision

tower. It was probably not fully seated into the fiberoptic outlet or it got caught on something and pulled out accidentally

• Ensure that the blue fiberoptic cable is not damaged (ie. pins bent) If it is

compromised, retrieve a new cable.

• Power system down and restart.

• Place blue fiberoptic cable back into patient cart, ensuring that it is seated properly with the red indicator facing at the 12 o'clock position. Listen for the audible click once the blue cable is inserted. Give the cable a slight tug to ensure a connection is made. power the system back on.

11

Page 12: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

SCENARIO #4

You are in the middle of a robotic prostatectomy and the surgeon reports the image is foggy. What should you do?

12

Page 13: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

SCENARIO #4 SOLUTIONS• Wipe tip of scope with a dry gauze. Check for any soiling or smudges

• Recommended practice is 55 Celcius (130 F) water to clean/heat scope tip. Its

recommended to not keep the scope sitting in the warmer longer than 10 seconds or submerged in warmer, which can damage the scope. Remember the chip of the camera is in the tip of the scope

• Fogging can occur because of a temperature difference between the cold scope

tip and the warm body cavity. Or a warmed scope tip that has cold CO2 gas insufflating too close to the scope tip.

• Ensure the endoscope is beyond the cannula tip

• Connect insufflation to a non-camera port

• Recommended practice is to plug in scope at least 30 minutes prior to insertion

• Its possible that smoke within the cavity may be mistaken for fogginess. Suction

out any residual smoke caused by cautery and use smoke evacuator (various types).

13

Page 14: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

Smoke Evacuator Options: Active evacuation “Poor Man”

14

Don’t attach to your camera port

Page 15: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

Smoke Evacuator Options: active evacuation Suction machine settings

15

Stryker Neptune

Page 16: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

Active smoke evacuation with charcoal filtration

16

Page 17: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

Smoke Evacuator Options: Passive auto Filtration

17

Pall Laparoshield

Page 18: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios 18

SCOPE WARMER and SURFACTANT

http://www.medtronic.com/content/dam/covidien/library/global/en/product/trocars-and-access/clearify-visualization-system-b.jpg

Page 19: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

JUST FOR FUN…

•What are some funny names you use for the scope warmer? •Hot hot (why do we say it twice?) • Fred • Igloo • Snail •Gary

19

http://vignette2.wikia.nocookie.net/spongebob/images/4/41/Gary_Movie.png/revision/latest?cb=20150603022800http://triplefatgoose.com/blog/wp-content/uploads/2016/04/Triple-FAT-Goose-How-To-Build-Igloo-1170x647.jpg

http://www.deroyal.com/designmaster/img/deroyalcom/contentpages/medicalproducts/surgicalandacutecare/ac-surgical/images/47_fmt2.jpg

Page 20: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

SCENARIO #5

You are about to start a robotic nephrectomy and you notice that as soon as the scope is inserted past the cannula tip, the field of view is dim.

20

Page 21: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

SCENARIO #5 SOLUTIONS

• Check the brightness settings. Use the brightness slider on the touchscreen or console touchpad to adjust to the desired level

• Always check the scope plug to ensure there

isn't any water or moisture that could potentially damage the endoscopic controller

• Remove the scope and examine tip for any

debris or damage

• Place scope into warmer for no longer than 10

seconds and wipe with a dry gauze

• Check cannula to ensure its not dislodged

• Change endoscope if the problem persists

21

Page 22: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

SCENARIO #6

The surgeon is about to start a robotic prostatectomy and the surgeon reports no energy coming from the monopolar scissors. You also notice that the grounding pad is not showing an illuminated lightning bolt sending energy to the instrument.

22

Page 23: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

SCENARIO #6 SOLUTIONS

• Its recommended practice to is to turn off the ERBE in between cases. The ERBE is like a computer software if overloaded the system stops working

• You can power cycle the ERBE without affecting the dV system itself • When the grounding pad is not detected, check the grounding pad. Ensure

that the long edge of the grounding pad points toward the operating field. If poor contact with the grounding pad due to patient with excessive lotion on skin, wash with a CHG scrub sponge and dry before replacing with a new grounding pad

• Check the instrument cord for any damage. Check instrument for any

damage. May need to replace cord or instrument. Each cord has 100 uses. • Ensure the surgeon footplate is deployed

• Ensure that the arrows on the energy cords are facing upward position and

matching the generator

23

Page 24: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios 24

• Disconnected or bad connection

• Arrows lined up

• Correct connection

Page 25: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

SCENARIO #7

The surgeon is ready to use the tilepro feature and they activate tilepro on the surgeon console, but it is not working. What should you do?

25

Page 26: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

SCENARIO #7 SOLUTIONS

• Connect the s-video, SDI, or DVI cord to the column labeled "Tilepro" on the back of the surgeon console or vision tower

• Make sure the video cable in connected in the correct column. ie. not in the auxiliary

column. • Inspect the cord for damage, usually found at the pin sites from excessive pulling or

bending • If using the dV Si system, once the cord is connected to "video source" on the vision

tower. Ensure that the word "Tilepro" is illuminated blue. On the dV Xi system, tap the home button on the vision tower touchscreen and the word "tilepro" should be illuminated white.

• Check the energy source ie. Ultrasound or laparoscopic tower is turned on. Sometimes

connecting the ultrasound pribe before powering on the dV system may not read probe. May need to re-boot the ultrasound and then reconnect the ultrasound probe.

• It may be helpful to apply color-coded tape the ends of each cord to help find what is

connected to each input/output

26

Page 27: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

Tile-Pro is not connected - greyed out

27

Page 28: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

Color Code Blue Fiberoptic Cords

28

Page 29: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

SCENARIO #8

You power on your dV Si system and notice that the vision tower has a flickering image on the screen. You check the surgeon console and notice the same flickering image. What should you do?

29

Page 30: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

SCENARIO #8 SOLUTIONS

• Check the brightness settings on the surgeon console and/or the vision tower. If Si system, press the arrow and to "advanced video adjustments" and select "restore factory settings"

• Confirm that the brightness indicator is not up in

the orange area on the touch screen

• If an Si system, tighten the camera cable on the

vision tower and camera head. Make sure there is not yellow seen, you should only see blue

• If the flickering image persists, you may need to

change the camera head cable.

30

Page 31: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

Si camera cable

31

Yellow visible

Page 32: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

SCENARIO #9

On the dV Si system you have placed the scope into the camera head and inserted the camera into the camera arm. You notice that the image is blurry. Your surgeon complains of blurry vision in the left eye on the surgeon console. What should you do?

32

Page 33: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

SCENARIO #9 SOLUTIONS• Check the endoscope for debris or smudges. Wipe scope and check clarity

• During the setup, look into the surgeon console to make sure both eyes in view

are the same. ie. One eye should not be darker or different from the other • Check the blue fiberoptic cable for a bent pin (make sure power is off if going to

disconnect it) • Rotate the scope 180 degrees. Did the blurred image move to the other eye? If

yes, the scope is your culprit; check both ends of the scope and clean if dirty. If its is not dirty or did not move eyes, change the scope.

• Check the lens of the camera head to see if smudged, dirty, or scratched. Wipe

camera head lens with a Q-tip if needed

• May need to change the camera head

33

Page 34: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

SCENARIO #10

You are in the middle of a case and the bedside assistant or surgeon reports that the instruments are locked in place while grasping tissue or that tissue will not release from the instrument. The instrument arm will not move. What should you do?

34

Page 35: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

SCENARIO #10 SOLUTIONS• Prepare to use the emergency tool wrench by placing the robot into a fault

state- • First press the red button. Never use the wrench on a system that's not in

a fault state. • Wrench should be sterile and readily accessible. • Bedside assistant will use the wrench and insert it in the emergency grip

release socket, turn counter clockwise to open the jaws of the endowrist instrument.

• Once instrument is released, check the instrument. May need to be replaced.

35

Page 36: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

SCENARIO #11

While clutching and advancing the monopolar scissors through the robotic cannula you meet resistance. What should you do?

36

Page 37: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

SCENARIO #11 SOLUTIONS

• Check the port. Make sure that the port is fully through the cannula wall at the remote center line

• If not, use the obturator to replace the port

• Check the monopolar scissor to ensure it is not bent or compromised. If so, replace the instrument

• Check the scissor guard protector to ensure that the energy protector is not pushed past the orange safety zone. Adjust the scissor guard or replace the scissor guard as needed

• Reseat the sterile adaptor. Check if you encounter the same resistance with a different instrument

37

Page 38: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios 38

cover all orange, but not beyond it

Page 39: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

You are in a robotic assisted cystoprostatectomy with bowel resection for ileal conduit. The surgeon is ready to use the robotic stapler. She clamps the stapler across the small bowel. The closed clamp reads at 75% on the davinici monitor. What should you do?

39

SCENARIO #12

Page 40: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

• The stapler should remain clamped in place for 10-15 seconds and allow the nearby edematous tissue to push away from the staple line. It should then be unclamped and re-clamped in the same exact spot. With each successive clamp attempt you should gain progress until it allows a 100% clamp, then it may be fired.

• This technique works for various thick tissue types: lung resection, small bowel, inflamed large bowel, gastrectomy, pancreatectomy, etc.

40

SCENARIO #12 SOLUTIONS

Page 41: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios 41

SCENARIO #13

You are in a robotic assisted prostatectomy with the patient in steep trendelenberg position. When the nurse drives the patient cart forward to the bed, the boom on the cart collides with the OR light boom. What should you do?

Page 42: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios 42

SCENARIO #13 SOLUTIONS

•Unlock and adjust the table position prior to docking. Move it north/south/east/west on the floor as needed. Be sure to lock it again! • Adjust the table height down • Adjust the patient cart boom lower

Page 43: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

SCENARIO #14

The surgeon has removed the robotic cannulae and trocars, and is about to close fascia, while you are starting to count. What should you have on your mayo stand to prepare for closure so that the surgical team can progress with closure while you count?

43

Page 44: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

SCENARIO #14 SOLUTIONS

Closure readiness example:

• May straight scissor • Needle driver x2 • Vicryl suture (UR6) • Monocryl suture x2

• S retractors / Army-Navy / Narrow Richardson retractors

• Lap Pad or 4x4s

44

Page 45: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

Brief Before and Debrief After Each Case

45

Page 46: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

Remember to keep your cool and communicate

46

Page 47: Suzannah Sorin, MSPAS, PA-C NARUS February 16, 2018 · Suzannah Sorin, MSPAS, PA-C Robotics Coordinator Physician Assistant Manager NYU Langone Medical Center NARUS February 16, 2018

Troubleshooting Scenarios

Any Questions?

Thank you!

47