Surgical manual 5.0ICX templant®
optimized surgical protocol
SilverSponSor of the DGi
Bronzesponsor of the eAoYears
20 Years
Guaranteed Spare Part A
vaila
bilit
yOsse
ointegreation
W a r r a n t y
Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e2
Explanation of the color schemes and the color codingof the ICX templant® drills
ICX-templant drills for implant diameter 4,1 belong to the GREEN color scheme.
Single drills have yellow color codes subject to the bone quality (very soft, soft, hard).
1 green marker for very soft bones D4
2 green markers for soft/medium hard bones D3/D2
3 green markers for hard bones, D1
ICX-templant drills for implant diameter 4,8 belong to the bluE color scheme.
Single drills have yellow color codes subject to the bone quality (very soft, soft, hard).
1 blue marker for very soft bones D4
2 blue markers for soft/medium hard bones D3/D2
3 blue markers for hard bones, D1
Ø 4.1mm
Ø 4.8mm
ICX-templant drills for implant diameter 3,75 belong to the RED color scheme.
Single drills have yellow color codes subject to the bone quality (very soft, soft, hard).
1 red marker for very soft bones D4
2 red markers for soft/medium hard bones D3/D2
3 red markers for hard bones, D1
Ø 3.75mm
ICX-plus templant drills for implant diameter 3,45 belong to the yEllow color scheme.
Single drills have yellow color codes subject to the bone quality (very soft, soft, hard).
1 yellow marker for very soft bones D4
2 yellow markers for soft/medium hard bones D3/D2
3 yellow markers for hard bones, D1
Ø 3.45mm
Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e 3
Explanation of the color schemes and the color codingof the ICX templant® drills
Explanation of the color schemes and the color codingof the ICX templant® drills
ICX-templant drills for implant diameter 4,1 belong to the GREEN color scheme.
Single drills have yellow color codes subject to the bone quality (very soft, soft, hard).
1 green marker for very soft bones D4
2 green markers for soft/medium hard bones D3/D2
3 green markers for hard bones, D1
ICX-templant drills for implant diameter 4,8 belong to the bluE color scheme.
Single drills have yellow color codes subject to the bone quality (very soft, soft, hard).
1 blue marker for very soft bones D4
2 blue markers for soft/medium hard bones D3/D2
3 blue markers for hard bones, D1
Ø 4.8mm
Color coding sample:Green = 3,45 implant diameter1 green marker = drill for very soft bone D4
Color coding sample:Blue = 3,45 implant diameter1 blue marker = drill for very soft bone D4
Ø 4.1mm
ICX-templant drills for implant diameter 3,75 belong to the RED color scheme.
Single drills have yellow color codes subject to the bone quality (very soft, soft, hard).
1 red marker for very soft bones D4
2 red markers for soft/medium hard bones D3/D2
3 red markers for hard bones, D1
Color coding sample:Red = 3,45 implant diameter1 red marker = drill for very soft bone D4
Ø 3.75mm
ICX-plus templant drills for implant diameter 3,45 belong to the yEllow color scheme.
Single drills have yellow color codes subject to the bone quality (very soft, soft, hard).
1 yellow marker for very soft bones D4
2 yellow markers for soft/medium hard bones D3/D2
3 yellow markers for hard bones, D1
Color coding sample:Yellow = 3,45 implant diameter1 yellow marker = drill for very soft bone D4
Ø 3.45mm
Ø 2,9Surgical procedure for ICX-mini templant® Ø 2.9mm
Service-Fon: +49 02643 902000-06
Ø 2.9mm from 25 rpm
ICX templant-Implant rpm
from 400 rpmICX templant-Drill
rpm
drills for very soft
to medium hard bones,
D2/D3/D4
drills for
hard bones,
D1
10.0mm
12.5mm
8.0mm6.5mm
Ø 2,9Surgical procedure for ICX-mini templant® Ø 2.9mm D2/D3/D4 drilling sequence for ICX-mini templant® Ø 2.9mm
Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e 7
Ø 2.9mm from 25 rpm
ICX templant-Implant rpm
from 400 rpmICX templant-Drill
rpm
Drilling sequence for very soft to medium hard bones.
Please only use drills with 2 markers.
Important Note: drills for very soft to medium hard bones D2/D3/D4 are labeled with 2 white markers
ICX -mini templant®Ø2.9mm
ICX guide drill ICX D2/D3/D4-drill,
2 markers, white
D1-drilling sequence for ICX-mini templant® Ø 2.9mm
Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e8
Drilling sequencefor hard bones.
Please only use drills with 3 markers.
Important Note: drills for hard bones D1, are labeled with 3 white markers
Explanation of the color schemes and the color coding of the ICX-mini templant®– Ø 2.9mm drills
Ø 2.9mm from 25 rpm
ICX templant-Implant rpm
from 400 rpmICX templant-Drill
rpm
ICX -mini templant®Ø2.9mm
ICX guide drill
ICX D1-drill, 3 markers, white
w w w . m e d e n t i s . d e 9
Explanation of the color schemes and the color coding of the ICX-mini templant®– Ø 2.9mm drills
ICX-mini templant drills for implant diameter 2,9 belong to the white color scheme.
Single ICX mini templant drills have different white color codes subject to the bone quality (very soft, soft, hard).
2 white markers for soft/medium hard bones D3/D2
3 white markers for hard bones, D1
Color coding sample:White = 2,9 implant diameter2 white markers = drill for soft to medium-hard bones D4/D3/D2
Ø 2.9mm
Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e8
Important note for the prosthetic provision of all ICX-plus implants.
both the open and the closed impression posts fit on the ICX and ICX-plus implants. only the universal lab Analogs differ.
Especially for a very low mucosa height, it is mandatory that the technician is aware, which implant is an ICX-plus implant. we recommend invariably informing the technician whether it‘s ICX or ICX-plus implants.(this is not mandatory. The prosthetic devices fit both on the ICX as well as on the ICX-plus implants.)
Especially for ICX-plus, we have developed the straight titanium abutment with 0 mm gingiva height.It only fits on the ICX-plus implants - all other prosthetic components fit both on ICX and on ICX-plus implants.
Important Note + Important Note!
Specificity for all ICX-plus implants.
Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e 9
Important note for the prosthetic provision of all ICX-plus implants.
For ICX-plus implant Ø 3.45 mm, we recommend the ICX-plus universal lab Analog. Art-no.: C-006-010002.
Art-no.: C-006-010001 we do not recommend forICX-plus implant Ø 3.45mm!
Important Note + Important Note!
Specificity for all ICX-plus implants.
Ø 3,45
Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e
Surgical procedure for ICX-plus templant® Ø 3.45mm
12
Drill for
very soft bone,
D4
Drill for soft/
medium-hard bones,
D3/D2
Drill for
hard bones,
D1
ICX stop drill,
mandatory use
10.0mm
12.5mm
15.0mm
8.0mm
6.5mm
Ø 3.45mm from 25 rpm
ICX templant-Implant rpm
from 400 rpmICX templant-Drill
rpm
Ø 3,45Drilling sequence for very soft bone
Please only use drills with 1 color marker.
Important Note: drills for very soft bones D4,are labeled with 1 yellow marker
w w w . m e d e n t i s . d e
Surgical procedure for ICX-plus templant® Ø 3.45mm D4 drilling sequence for ICX-plus templant® Ø 3.45mm
13
transgingival = collar does not sink subgingival = collar sinks
ICX Stop-drill, yellow
ICX Stop-drill, yellow
ICX-plusØ3.45mm
ICX-plusØ3.45mm
ICX D4-drill, 1 marker,yellow
ICX guide drill
ICX D4-drill, 2 markers, white
Note:If the collar is sunk, the cavity needs to be 1.7mm deeper.
Standard Optionally
ICX stop drill,
mandatory use
Ø 3,75mm from 25 rpm
ICX templant-Implant rpm
from 400 rpmICX templant-Drill
rpm
Ø 3.45mm
Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e
D2/D3 drilling sequence for ICX-plus templant® Ø 3.45mm
14
Drilling sequence for soft to medium hard bones
Please only use drills with 2 color markers.
Important Note: drills for soft/medium-hard bones D2/D3,are labeled with 2 yellow markers
transgingival = collar does not sink subgingival = collar sinks
ICX Stop-drill, yellow
ICX Stop-drill, yellow
ICX-plusØ3.45mm
ICX-plusØ3.45mm
ICX D2/D3-drills, 2 markers, white
ICX guide drill
ICX D2/D3-drills, 2 markers, yellow
Note:If the collar is sunk, the cavity needs to be 1.7mm deeper.
Standard Optionally
Ø 3.45mm from 25 rpm
ICX templant-Implant rpm
from 400 rpmICX templant-Drill
rpm
w w w . m e d e n t i s . d e
D2/D3 drilling sequence for ICX-plus templant® Ø 3.45mm D1 drilling sequence for ICX-plus templant® Ø 3.45mm
15
Drilling sequence for hard bones Please only use drills with 3 color markers.
Drilling sequence for soft to medium hard bones
Please only use drills with 2 color markers.
Important Note: drills for soft/medium-hard bones D2/D3,are labeled with 2 yellow markers
Important Note: drills for hard bones D1, are labeled with 3 yellow markers
transgingival = collar does not sink subgingival = collar sinks
ICX Stop-drill, yellow
ICX Stop-drill, yellow
ICX-plusØ3.45mm
ICX-plusØ3.45mm
ICX D1-drill, 3 markers, yellow
ICX guide drill
ICX D1-drill, 3 markers, white
Note:If the collar is sunk, the cavity needs to be 1.7mm deeper.
Standard Optionally
Ø 3,75mm from 25 rpm
ICX templant-Implant rpm
from 400 rpmICX templant-Drill
rpm
Ø 3.45mm
Ø 3,75Drilling sequence for very soft bone
Please only use drills with 1 color marker.
Surgical procedure for ICX–templant® Ø 3.75mm
Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e16
Drill for
very soft bone,
D4
Drill for soft/
medium-hard bones,
D3/D2
Drill for
hard bones,
D1
ICX stop drill,
mandatory use
12.5mm
15.0mm
10.0mm
8.0mm 6.5mm
Ø 3.75mm from 25 rpm
ICX templant-Implant rpm
from 400 rpmICX templant-Drill
rpm
Ø 3,75Drilling sequence for very soft bone
Please only use drills with 1 color marker.
Important Note: drills for very soft bones D4,are labeled with 1 red marker
D4 drilling sequence for ICX–templant® Ø 3.75mm
w w w . m e d e n t i s . d e 17
ICX implantØ3.75
ICX Stop-drill, red
ICX D4-drill, 1 marker,red
ICX guide drill
ICX D4-drill, 2 markers, white
Ø 3.75mm from 25 rpm
ICX templant-Implant rpm
from 400 rpmICX templant-Drill
rpm
Service-Fon: +49 02643 902000-0
D2/D3 drilling sequence for ICX–templant® Ø 3.75mm
Drilling sequence for soft to medium hard bones
Please only use drills with 2 color markers.
Important Note: drills for soft/medium-hard bones D2/D3,are labeled with 2 red markers
Important Note: drills for hard bones D1,are labeled with 3 red markers
18
ICX implantØ3.75
ICX Stop-drill, red
ICX D2/D3-drills, 2 markers, red
ICX guide drill
ICX D2/D3-drills, 2 markers, white
Ø 3.75mm from 25 rpm
ICX templant-Implant rpm
from 400 rpmICX templant-Drill
rpm
19Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e
D2/D3 drilling sequence for ICX–templant® Ø 3.75mm D1 drilling sequence for ICX–templant® Ø 3.75mm
Drilling sequence for soft to medium hard bones
Please only use drills with 2 color markers.
Drilling sequence for hard bones Please only use drills with 3 color markers.
Important Note: drills for hard bones D1,are labeled with 3 red markers
ICX implantØ3.75
ICX Stop-drill, red
ICX D1-drill, 3 markers, red
ICX guide drill
ICX D1-drill, 3 markers, white
Ø 3.75mm from 25 rpm
ICX templant-Implant rpm
from 400 rpmICX templant-Drill
rpm
Ø 4,1
20 Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e
Surgical procedure for ICX–templant® Ø 4.1mm D4 drilling sequence for ICX–templant® Ø 4.1mm
Drilling sequence for very soft bone Please only use drills with 1 color marker.
Drill for
very soft bone,
D4
Drill for soft/
medium-hard bones,
D3/D2
Drill for
hard bones,
D1
ICX stop drill,
mandatory use
12.5mm
15.0mm
10.0mm
8.0mm 6.5mm
Ø 4.1mm from 25 rpm
ICX templant-Implant rpm
from 400 rpmICX templant-Drill
rpm
Ø 4,1
w w w . m e d e n t i s . d e
D4 drilling sequence for ICX–templant® Ø 4.1mm
Drilling sequence for very soft bone Please only use drills with 1 color marker.
Important Note: drills for very soft bones D4, are labeled with 1 green marker
21
ICX implantØ4.1
ICX D4-drill, 1 marker,green
ICX D4-drill, 1 marker,red
ICX guide drill
ICX D4-drill, 2 markers, white
ICX Stop-drill, green
ICX stop drill,
mandatory use
Ø 4.1mm from 25 rpm
ICX templant-Implant rpm
from 400 rpmICX templant-Drill
rpm
Service-Fon: +49 02643 902000-0
D2/D3 drilling sequence for ICX–templant® Ø 4.1mm
Drilling sequence for soft to medium hard bones
Please only use drills with 2 color markers.
Important Note: drills for soft/medium-hard bones D2/D3,are labeled with 2 green markers
Important Note: drills for hard bones D1,are labeled with 3 green markers
22
ICX implantØ4.1
ICX Stop-drill, green
ICX D2/D3-drills, 2 markers, green
ICX D2/D3-drills, 2 markers, red
ICX guide drill
ICX D2/D3-drills, 2 markers, white
Ø 4.1mm from 25 rpm
ICX templant-Implant rpm
from 400 rpmICX templant-Drill
rpm
Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e
D2/D3 drilling sequence for ICX–templant® Ø 4.1mm D1 drilling sequence for ICX–templant® Ø 4.1mm
Drilling sequence for soft to medium hard bones
Please only use drills with 2 color markers.
Drilling sequence for hard bones Please only use drills with 3 color markers.
Important Note: drills for hard bones D1,are labeled with 3 green markers
23
ICX ImplantatØ4.1
ICX Stop-drill, green
ICX D1-drill, 3 markers, green
ICX D1-drill, 3 markers, red
ICX guide drill
ICX D1-drill, 3 markers, white
Ø 4.1mm from 25 rpm
ICX templant-Implant rpm
from 400 rpmICX templant-Drill
rpm
Ø 4,8
Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e
Surgical procedure for ICX–templant® Ø 4.8mm D4 drilling sequence for ICX–templant® Ø 4.8mm
24
Drill for
very soft bone,
D4
Drill for soft/
medium-hard bones,
D3/D2
Drill for
hard bones,
D1
ICX stop drill,
mandatory use
12.5mm
15.0mm
10.0mm
8.0mm 6.5mm
Ø 4.8mm from 25 rpm
ICX templant-Implant rpm
from 400 rpmICX templant-Drill
rpm
Ø 4,8
w w w . m e d e n t i s . d e
D4 drilling sequence for ICX–templant® Ø 4.8mm
Drilling sequence for very soft bone Please only use drills with 1 color marker.
Important Note: drills for very soft bones D4,are labeled with 1 blue marker
25
ICX D4-drill, 1 marker,blue
ICX D4-drill, 1 marker,red
ICX implantØ4.8
ICX guide drill
ICX D4-drill, 2 markers, white
ICX Stop-drill, blue
ICX stop drill,
mandatory use
Ø 4.8mm from 25 rpm
ICX templant-Implant rpm
from 400 rpmICX templant-Drill
rpm
Service-Fon: +49 02643 902000-026
D2/D3 drilling sequence for ICX–templant® Ø 4.8mm
Drilling sequence for soft to medium hard bones
Please only use drills with 2 color markers.
Important Note: drills for soft/medium-hard bones D2/D3,are labeled with 2 blue markers
Important Note: drills for hard bones D1, are labeled with 3 blue markers
ICX D2/D3-drills, 2 markers, blue
ICX implantØ4.8
ICX Stop-drill, blue
ICX guide drill
ICX D2/D3-drills, 2 markers, white
ICX D2/D3-drills, 2 markers, red
Ø 4.8mm from 25 rpm
ICX templant-Implant rpm
from 400 rpmICX templant-Drill
rpm
27Service-Fon: +49 02643 902000-0 w w w . m e d e n t i s . d e
D2/D3 drilling sequence for ICX–templant® Ø 4.8mm D1 drilling sequence for ICX–templant® Ø 4.8mm
Drilling sequence for soft to medium hard bones
Please only use drills with 2 color markers.
Drilling sequence for hard bones Please only use drills with 3 color markers.
Important Note: drills for soft/medium-hard bones D2/D3,are labeled with 2 blue markers
Important Note: drills for hard bones D1, are labeled with 3 blue markers
ICX D1-drill, 3 markers, blue
ICX implantØ4.8
ICX guide drill
ICX D1-drill, 3 markers, white
ICX Stop-drill, blue
ICX D1-drill, 3 markers, red
Ø 4.8mm from 25 rpm
ICX templant-Implant rpm
from 400 rpmICX templant-Drill
rpm
Service-Fon: +49 02643 902000-0 w w w . t e m p l a n t . d e
Handling/Insertion of the ICX templant®
Compatibly to elbow and hand adapter
4 5
1 2 3
10
6
7 8 9
Sterile ICX-packageExtraction of the ICX templant®with the ICX torque wrench driver
ICX torque wrenchDriver short and medium
ICX torque wrenchInserted in the dental motor
ICX torque wrenchDriver inserted in hand-driver
Finally insertedcover cap
Insertion of the ICX templant®
Extraction of the ICX templant®with the ICX torque wrench driver
Extraction of the ICX templant®with the ICX torque wrench driver