Steam Sterilizer “TIMO” “TIMO- PRINT” - Code n. 88944 - 88957 International pbi S.p.A. « Timo » « Timo Print » Instructions manual Rev. 1 1 INSTRUCTIONS MANUAL International PBI S.p.A. Via Novara, 89 – 20153 Milano – Italy – Tel + 39 02 48779-1 – Fax + 39 02 40090010 E-mail: [email protected] www.internationalpbi.it

Upload: others

Post on 12-Oct-2020

32 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Steam Sterilizer “TIMO” “TIMO- PRINT” - Code n. 88944 - 88957shop.skolebutik.dk/images/pdf/timo.pdf · International pbi S.p.A. « Timo » « Timo Print » Instructions manual

Steam Sterilizer “TIMO” “TIMO-PRINT” - Code n. 88944 - 88957

International pbi S.p.A. « Timo » « Timo Print » Instructions manual Rev. 1 1

IINNSSTTRRUUCCTTIIOONNSS MMAANNUUAALL

International PBI S.p.A. Via Novara, 89 – 20153 Milano – Italy – Tel + 39 02 48779-1 – Fax + 39 02 40090010

E-mail: [email protected] – www.internationalpbi.it

Page 2: Steam Sterilizer “TIMO” “TIMO- PRINT” - Code n. 88944 - 88957shop.skolebutik.dk/images/pdf/timo.pdf · International pbi S.p.A. « Timo » « Timo Print » Instructions manual

INSTRUCTIONS MANUAL READ CAREFULLY THIS HANDBOOK SAFETY INSTRUCTIONS • Use the device just for the indicated purposes. • The Device must be correctly installed according to thes instructions,before starting any operation. • Never disconnect the device by pulling electrical cable from the socket; . • Replace immediately electrical cable when damaged ; never use a damaged or worn electrical cable. • On removing the item connected to the electrical system pay attention to the electrical connection cable • Always disconnect from electrical system on:

A. Reparing or maintenance; these operations must be carried out by qualified staff; B. Cleaning the unit; C. Use original spare parts and accessories for any replacement;

• Do not use this device in the presence of gaseous explosive.

.

INSTALLATION GENERAL CAUTION

CAUTION

When the electrical system socket is not compatible with device socket, replace it paying attention to codified cable colours. Brow: live Blue: neutral Yellow/green: earth

CAUTION

Be aware that the voltage and frequence of the electrical system are compatible with device. The device must be connected to a ground wire. For any question please contact an electrician.

CAUTION

Before starting-up the device be sure that ventilation grids are totaly free to operate and not covered by any barriers.

International pbi S.p.A. « Timo » « Timo Print » Instructions manual Rev. 1 2

Page 3: Steam Sterilizer “TIMO” “TIMO- PRINT” - Code n. 88944 - 88957shop.skolebutik.dk/images/pdf/timo.pdf · International pbi S.p.A. « Timo » « Timo Print » Instructions manual

International pbi S.p.A. « Timo » « Timo Print » Instructions manual Rev. 1 3

SUMMARY

1. Description of unit …………………………………………….……..... pg. 4

2. Electrical supply ……………………………………………………….. pg. 5

3. Before using for the first time ….……………..….…………….…….. pg. 5

4. Preliminary operations ……………………………….………………. pg. 6

5. Sterilization temperature programming ….……..………….……….. pg. 6

6. Sterilization time programming …………………….……….……….. pg. 6

7. Sterilization cycle …………………………………………….……….. pg. 6

8. Panel commands ……………………………………...……………….. pg. 7

9. Cleaning after use ……………………...…………………………….... pg. 7

10. General maintenance ………………….…………………………….. pg. 8

11. Replacement of the heating element ………….…………………….. pg. 8

12. Temperature/Pressure rate ………….…………..………………….. pg. 8

13. Part list ………………………………………….…………………….. pg. 9

14. Visual codes of the programmer and description of the related

parameters ………….…………………………………………………….. pg. 9

15. Connection and disconnection of the safety key of the selected

parameters ………………………………………….…………………….. pg. 10

16. Printer (optional)…………………………………………………….. .pg.10

Page 4: Steam Sterilizer “TIMO” “TIMO- PRINT” - Code n. 88944 - 88957shop.skolebutik.dk/images/pdf/timo.pdf · International pbi S.p.A. « Timo » « Timo Print » Instructions manual

1. DESCRIPTION OF UNIT (FIG. 1) The steam sterilizer “TIMO” consist of: PRESSURE GAUGE (Pos.1) The pressure gauge registers pressure in both pounds (outer scale) and metric measure (inner scale). The pointer moves around the dial indicating precisely the pressure within the unit. PRESSURE GAUGE ADAPTOR The pressure gauge adaptor connects the pressure gauge to the vessel cover. PRESSURE REGULATOR (Pos. 2) The pressure regulator is a solid one-piece unit. Accurate pressure readings can be controlled and maintained by adjusting heat settings. AIR VENT PIPE The pressure regulator fits over the vent pipe and allows excess pressure to be released. AIR VENT COVER LOCK The air vent/cover lock automatically "vents" or exhausts air from the vessel and acts as a visual indication of pressure in the unit. LOCKING BRACKET The locking bracket on the inside of the vessel body engages with the air vent/cover lock to prevent the cover from being opened when there is pressure in the unit. SEALING RING (Pos. 3) The sealing ring fits into the vessel cover and forms a pressure-tight seal between cover and body during sterilisation. OVERPRESSURE PLUG 2 1The overpressure plug is located in the vessel cover. It will automatically pop out and release steam in case the vent pipe becomes clogged and pressure cannot be released normally.

3"MEDIUMBOX" CYLINDRICAL BASKET 4 (Pos. 4) VESSEL AND LID WITH TWO PERIFERAL HANDLES ELECTRONIC DEVICE FOR THERMOREGULATION (Pos.5) MAIN SWITCH (Pos. 6) OPTIONAL PRINTER (Pos. 7)

7 5

6

International pbi S.p.A. « Timo » « Timo Print » Instructions manual Rev. 1 4

Page 5: Steam Sterilizer “TIMO” “TIMO- PRINT” - Code n. 88944 - 88957shop.skolebutik.dk/images/pdf/timo.pdf · International pbi S.p.A. « Timo » « Timo Print » Instructions manual

International pbi S.p.A. « Timo » « Timo Print » Instructions manual Rev. 1 5

2. ELECTRICAL SUPPLY

220 - 240 Vac SINGLE PHASE 1300 W. This unit must be earthed. It is delivered with a mains plug with safety contacts. If replacing the mains plug, the mains lead colored green and yellow must be connected to the earth terminal on the new plug. If using a mains supply without a protective earth, the unit must be connected to a safe earth.

3. BEFORE USING FOR THE FIRST TIME 3.1 Remove the air vent/cover lock from the vessel cover. The cup portion of the air vent/cover lock, on the inside of the cover, unscrews from the pin. 3.2 Remove the sealing ring by simply pulling from the sealing ring groove. 3.3 Wash the cover and body with hot, soapy water. Do not submerge the cover in water or let the gauge come in contact with any liquid. The inside mechanism of the gauge may be damaged if it is submerged or if water is allowed to run over it. Scrub the sealing ring groove to remove the manufacturing oils. Scour the inside of the canner body and cover with a nonalkali, aluminum cleaner or a steel wool pad. Rinse with warm water and dry. 3.4 Reinsert the air vent/cover lock making sure the small gasket is in place. 3.5 Replace the sealing ring in the sealing ring groove. The sealing ring is pre-lubricated, therefore, it is not necessary to apply oil to it. Frequent oiling may cause swelling of the sealing ring. 3.6 If the cover is difficult to open or close use a pastry brush, a piece of cloth or your fingertips to apply a light coating of oil to the underside of the lugs on the vessel body. The cover should open and close easily. Be sure the vent pipe is open before each use by holding the cover up to the light and looking through the vent pipe. If the vent pipe is blocked, excess pressure cannot be released through it. Pressure will continue to build until the over pressure plug pops out of its cover opening. Pressure is indicated on the dial gauge which registers the pressure inside the unit at all times. The pointer will move across the dial as pressure is developed or reduced within the vessel. The one-piece regulator is designed to rock only at 15 pounds pressure. Do not strike the rim of the vessel body with any utensil as this will cause nicks which may damage the rim and allow steam to escape. Pouring water into a dry overheated vessel may crack the metal. The vessel has been designed so the cover will go on in only one position and rotate only in the directions indicated for closing and opening. Press down on the cover handles to compress the sealing ring and make the cover easier to close. The cover has the words CLOSE / OPEN embossed on the top surface near a cover handle. Align the "V" on the cover with the mark on the body handle for proper cover and body alignment. Turn the cover in the direction indicated to close until the cover handles are centered directly above body handles. Do not rotate cover beyond this point. When the cover is rotated to the closed position, the air vent/cover lock passes under the locking bracket. When the handles are aligned, one over the other, the pressure regulator is placed on the vent pipe, and heat is applied, pressure will begin to build within the vessel. At this point, the air vent/cover lock lifts and locks the unit. The cover will remain locked as long as there is pressure in the vessel. When pressure is completely reduced, the air vent/cover lock drops allowing the unit to be opened.

Page 6: Steam Sterilizer “TIMO” “TIMO- PRINT” - Code n. 88944 - 88957shop.skolebutik.dk/images/pdf/timo.pdf · International pbi S.p.A. « Timo » « Timo Print » Instructions manual

International pbi S.p.A. « Timo » « Timo Print » Instructions manual Rev. 1 6

4. PRELIMINARY OPERATIONS 4.1 This steam sterilizer must only be used for a maximum pressure of 1,08 bar (121 °C). 4.2 The steam sterilizer should be positioned in an area where there is adequate means for removal of steam. 4.3 Connect plug to mains supply. 4.4 Introduce distilled water or demineralized water into the sterilization chamber to cover the heating system. The use of distilled water avoids the production of calcareous deposits on the heating element. 20 minutes of sterilizing will use approximately 500 ml of water. 4.5 Two displays are lighted on time/temperature command panel. 4.6 Before programming, it is necessary to erase the protection of the parameters (safety keys), see paragraph 15.

5. STERILIZATION TEMPERATURE SETTING 5.1 Press “ENTER” four times in quick sequence to obtain “SP1” on lower display. 5.2 Select the sterilization temperature using the arrows. The sterilization temperature will be shown on the upper display. 5.3 Press “ENTER” to memorize.

6. STERILIZATION TIME SETTING 6.1 Press “ENTER”. The “t in 1” will be shown on lower display. 6.2 Select the sterilization time using the arrows. The sterilization time will be shown on the upper display. 6.3 Press “ENTER” to memorize. 6.4 The starting values will be shown on the display after 30 seconds (vessel actual temperature and the word “OFF”).

7. STERILIZATION CYCLE 7.1 Press the “START/STOP” button for 3 seconds to start the cycle. The following indications will be shown:

• flashing green led until the selected sterilization temperature is reached on lower display; • vessel actual temperature on upper display; • programmed temperature on lower display; • “HEATING” led on, during heating time; • “AUTOCONTROL” led on for automatic control of selected parameters.

7.2 When the sterilization temperature has been reached, the time left to reach end of the cycle is shown on the lower display. 7.3 The end of the sterilization cycle is indicated by an acoustic alarm for ten seconds. The “CYCLE END” led will light up for ten seconds. 7.4 The upper display will show the actual vessel temperature; the lower display will show the word “OFF”. 7.5 At the end of vessel cooling, open the lid to take out the sterilized material.

Page 7: Steam Sterilizer “TIMO” “TIMO- PRINT” - Code n. 88944 - 88957shop.skolebutik.dk/images/pdf/timo.pdf · International pbi S.p.A. « Timo » « Timo Print » Instructions manual

International pbi S.p.A. « Timo » « Timo Print » Instructions manual Rev. 1 7

8. PANEL COMMANDS

START/STOP To start/stop the program. ▲▼ Arrows to select the values. ENTER Multifunctional button:

- to memorize the selected parameters - to switch from a function to another - to control the selected program

GREEN LED It flashes until the selected temperature has been reached CYCLE AND LED It is on at the end of the sterilization cycle for ten seconds HEATING LED It is on during heating time AUTOCONTROL LED

It is on to indicate that the automatic system is controlling the selected parameters.

LED °C The temperature is indicated in Celsius degrees. NOTES All the programming operations must be performed in quick sequence. Otherwise the programming device is zeroing. The program can be stopped by pressing the “START/STOP” button for more than five seconds. The only parameters to be selected are (SP1) temperature and (t in 1) time. The selected parameters remain memorized also after the instrument is turned off.

9. CLEANING AFTER USE 9.1 The outside surface may be kept bright and shiny by cleaning with a good silver polish or simply by washing with soap and water. Iron and various minerals in water may darken the inside but this discoloration will in no way affect sterilization. These stains are harmless and may be removed by using a solution of water and cream of tartar. For each quart of water, use one table-spoon cream of tartar. Pour enough solution into the vessel to cover the discoloration (do not fill over 2/3 full), then close cover securely. Place regulator on vent pipe and heat until 15 pounds pressure is reached. Switch off and allow the unit to stand two to three hours. Remove regulator, open and empty contents. Scour thoroughly with a soap impregnated steel wool cleaning pad; wash, rinse and dry. 9.2 Each time the vessel is washed, remove the sealing ring and wash in warm, soapy water, rinse, dry and replace in cover. 9.3 The air vent/cover lock may be removed for occasional cleaning or for replacing the small gasket. Wash all parts in warm, soapy water. Use a soft cloth or small nylon brush to clean the cover hole. The small gasket must be in place when reassembling the air vent/cover lock. 9.4 The overpressure plug can be removed for cleaning by pulling it out of its opening from the bottom part of the cover. After cleaning, reinsert the plug by pushing the round side of the plug into the opening from the bottom part of the cover. When the overpressure plug is properly installed the word (TOP) will be visible when viewing the outside of the cover. The dented portion of the overpressure plug is visible when the bottom part of the cover is viewed. If the overpressure plug is ever forced out of its cover opening due to excess pressure while sterilising, it is important to have the vessel properly inspected before further use. Do not attempt to replace the overpressure plug and ask for servicing. 9.5 Be sure the vent pipe is open before each use. Clean the vent pipe with a small brush or pipe cleaner. 9.6 When your unit is not in use, invert the cover on the body and store in a dry place. Storing the unit with the cover locked on may deform the sealing ring.

Page 8: Steam Sterilizer “TIMO” “TIMO- PRINT” - Code n. 88944 - 88957shop.skolebutik.dk/images/pdf/timo.pdf · International pbi S.p.A. « Timo » « Timo Print » Instructions manual

International pbi S.p.A. « Timo » « Timo Print » Instructions manual Rev. 1 8

10. GENERAL MAINTENANCE 10.1 CAUTION: To ensure safe operation and satisfactory performance, replace the overpressure plug every time you replace the sealing ring or sooner if it becomes hard or deformed. In any case replace the overpressure plug every two years. Failure to follow these instruction could result in bodily injury or property damage. 10.2 If the body or cover handles become loose, tighten with a screwdriver. 10.3 If leakage of moisture or steam develops while using your unit, check the following possible causes: The formation of a small amount of moisture under the pressure regulator is normal when heating first begins. This condensation is a result of the temperature of the pressure regulator being lower than the rest of the vessel. If excess condensation continues, the vent pipe may be loose and should be tightened with an adjustable wrench. Leakage between the cover and body is usually caused by shrinkage of the sealing ring after prolonged use. Replace the sealing ring, overpressure plug and pressure gauge adaptor. A slight amount of leakage around the air vent/cover lock is normal when sterilising first begins. If leakage continues, the cover handles may not be fully aligned with the body handles and, therefore, the cover lock cannot engage. Clean the air vent/cover lock occasionally to assure that it operates correctly. A small amount of steam or moisture may be visible around the overpressure plug as sterilisation begins. This will stop when the overpressure plug seals. If leakage continues, clean or replace the overpressure plug. 10.4 If the Pressure vessel becomes difficult to open or close and the sealing ring has expanded, replace the sealing ring. 10.5 IMPORTANT: In time, the sealing ring may shrink and the air vent/cover lock, small gasket, overpressure plug and pressure gauge adaptor may become hard and deformed. When this happens, replace the sealing ring, small gasket, overpressure plug and pressure gauge adaptor. The sealing ring, overpressure plug and pressure gauge adaptor should be replaced at least every two years. Before inserting a new sealing ring, clean the sealing ring groove with a brush. To insure safe operation of your unit, it is recommended that the pressure gauge adaptor and overpressure plug be replaced whenever the sealing ring is replaced.

11. REPLACEMENT OF THE HEATING ELEMENT Disconnect the electric supply, unscrew the two fixing nuts on the outside bottom of the sterilisation chamber and insert a new heating element. To re-assemble, repeat the same steps in reverse order. Ensure that the water level in the sterilisation chamber always covers the heating element.

12. TEMPERATURE/PRESSURE RATE Temperature °C 100 102 105 109 112 115 119 120 128 134 Pressure bar 0 0.1 0.2 0.4 0.5 0.7 0.9 1.0 1.5 2.0

Page 9: Steam Sterilizer “TIMO” “TIMO- PRINT” - Code n. 88944 - 88957shop.skolebutik.dk/images/pdf/timo.pdf · International pbi S.p.A. « Timo » « Timo Print » Instructions manual

International pbi S.p.A. « Timo » « Timo Print » Instructions manual Rev. 1 9

13. SPARE PARTS cod. 32828 MANOMETER cod. 64703 HEATING ELEMENT 200 V 1300 W (110 V ON REQUEST) cod. 53303 ELECTRONIC DEVICE FOR THERMOREGULATION (TIMO) cod. 8174 SEALING AWG FOR COVER cod. 19113 OVERPRESSURE PLUG cod. 67101 MAIN SWITCH cod. 67041 FUSE HOLDER cod. 67038 FUSE 10 A. cod. 64194 BUZZER cod. 53056 NOISE FILTER cod. 53301 RELAY SS cod. 66048 THERMOPROBE cod. 58013 SACRIFICAL BODY cod. 19114 FIXING COVER DEVICE cod. 22721 PRESSURE REGULATOR cod. 77709 ELECTRONIC DEVICE FOR THERMOREGULATION (TIMO PRINT)

14. VISUAL CODES OF THE PROGRAMMER AND DESCRIPTION OF THE RELATED PARAMETERS In normal conditions the programmer should show on the display the function parameters as listed in the table. The parameters can be shown by the multifunctional key “ENTER”. The only values that can be modified by the operator are: “nnn” = parameters protection key “SP1” = sterilization temperature “t in t” = sterilization time

FUNCTIONS PARAMETERS DESCRIPTION Snrt ON Autom. control of parameters nnn ON/OFF Parameters protection key dELY 0,00 Starting cycle waiting time SP1 Steriliz. temp. Sterilization temperature Grd1 Inf Temperature speed increment t in 1 Steriliz.Time Sterilization time SP2 0 Not in use Grd2 inf Not in use t in 2 0,00 Not in use rPt 0 Cycle repetition AL1 0 Not in use HSA1 0,1 Not in use CY1 4 Not in use rL 0 °C lower limit rH 125 °C heating limit OLH 100 Max heating limit %

Page 10: Steam Sterilizer “TIMO” “TIMO- PRINT” - Code n. 88944 - 88957shop.skolebutik.dk/images/pdf/timo.pdf · International pbi S.p.A. « Timo » « Timo Print » Instructions manual

15. CONNECTION AND DISCONNECTION OF THE SAFETY KEY OF THE SELECTED PARAMETERS The selected parameters should be protected to avoid tampering: • press “ENTER” several times until the word “nnn” is shown on the display; • select a number different from 3 using the arrows and press “ENTER”; The protection of the parameters is erased by: • press “ENTER” several times until the word “nnn” is shown on the display; • select number 3 using the arrows and press “ENTER”;

16. PRINTER RECORDER (OPTIONAL) The autoclave is provided with a connection for a temperature printer . Printer records: a) Date b) Batch number c) The temperature recorded during the segment of time programmed d) Sterilization progressive number carried out automatically. 16.1. How to insert the paper

Pressing the two levers (E), open the pannel and insert paper as shown.

International pbi S.p.A. « Timo » « Timo Print » Instructions manual Rev. 1 10

Page 11: Steam Sterilizer “TIMO” “TIMO- PRINT” - Code n. 88944 - 88957shop.skolebutik.dk/images/pdf/timo.pdf · International pbi S.p.A. « Timo » « Timo Print » Instructions manual

International pbi S.p.A. « Timo » « Timo Print » Instructions manual Rev. 1 11

If the paper is correctly installed the led pos F will be on. Flashing led indicates not properly installed paper By pressing button G you can scroll paper. 16.2. Printer settings NOTE: The printer is turned on with the main switch. • HOW TO MODIFY THE DATE: - Press ENTER twice display will show OFF YEAR - You can change with - Confirm with ENTER - Do the same for MONTH and DAY • HOW TO MODIFY BATCH NUMBER: (at the end of the cycle, to assign a new nuber to next cycle) - Press twenty times ENTER The display will show 1 n bat - Bunch: you can change with - Confirm with ENTER • HOW TO MODIFY PRINTER INTERVAL: - After having modified the batch, the display will show: 1.00 P.int - You can change with - Confirm with ENTER - Display will show: OFF TEST • TO EXIT PRESS FIVE TIME START-STOP

Page 12: Steam Sterilizer “TIMO” “TIMO- PRINT” - Code n. 88944 - 88957shop.skolebutik.dk/images/pdf/timo.pdf · International pbi S.p.A. « Timo » « Timo Print » Instructions manual

International pbi S.p.A. « Timo » « Timo Print » Instructions manual Rev. 1 12

To the attention of: Technical department International PBI S.p.A. Via Novara, 89 - 20153 - Milano - Tel + 39 02/48.779.1-Fax +39 02/400.900.10

AUTHORISATION RETURN NUMBER TO BE ASKED TO THE EXPORT DPT

Company:

Address:

City: Zip Code State

Phone Fax Email

Name to contact:

Lab/Division:

return request of technical service

Equipment:

Model Serial Number:

for calibration repair other (specify)

NOTE OF MALFUNCTIONING

ELECTRICAL PART

ELECTRONICAL PART

MECHANICAL PART

HYDRAULICS PART

NOTES:

ATTENTION: THE INSTRUMENT MUST BE SENT TO

INTERNATIONAL PBI C/O STATIM VIA SAN GIUSTO 85 20153 MILANO ITALY

SIGNATURE NAME DATE

Mod. PQ09.9

Page 13: Steam Sterilizer “TIMO” “TIMO- PRINT” - Code n. 88944 - 88957shop.skolebutik.dk/images/pdf/timo.pdf · International pbi S.p.A. « Timo » « Timo Print » Instructions manual

International pbi S.p.A. « Timo » « Timo Print » Instructions manual Rev. 1 13

Page 14: Steam Sterilizer “TIMO” “TIMO- PRINT” - Code n. 88944 - 88957shop.skolebutik.dk/images/pdf/timo.pdf · International pbi S.p.A. « Timo » « Timo Print » Instructions manual

International pbi S.p.A. « Timo » « Timo Print » Instructions manual Rev. 1 14

WARRANTY

WARRANTY CONDITIONS International PBI warrants this product to be free of defects in materials and functionality The repairs will be effected by International PBI or by authorized dealers as follows: - Faulty components free replacement The labour is always at customer's expense - The equipment will be sent at buyer's expenses and risk and the return will be ex-works - This warranty does not apply to damages resulting from tampering, abnormal use and installation and does not

cover transport damages - International PBI shall not be liable for any commercial damages, whether incidental, consequential or

otherwise derived from the use of instruments

ATTENTION TO OUR NEW CUSTOMER As a user of our equipment you are covered for a 12 months guarantee for all spare parts as a result of mechanical failure of the machine during normal use. In order to provide you with first class service, it is essential for us to know all your references and which equipment you have. We kindly ask you therefore to fill in the attached guarantee and return the card to us by first class mail. A NOTRE NOUVEAU CLIENT En tant que utilisateur de notre appareil vous êtes couvert par une garantie de 12 mois pour toutes pièces de recharge eventuelles resultant de defauts mecaniques en utilisation normale. Dans le but de vous fournir un service impeccable il est essenciel que nous sachions qui vous êtes, ou vous êtes situes et quel equipment vous possedez. Voudriez vous remplir cette carte de garantie et nous la retourner. Le service PBI peut vous proposer un service d’entretien vous assurant un fonctionnement parfait rapide en cas de panne. Votre distributeur local peut vous fornir tous renseignements au sujet de ce contrat d’entretien et des prix appliqués. FUR UNSERE NEUEN KUNDEN Für Benutzer unserer Apparate sind alle ersatzteilen die sich aus mechanischen Fehlern in normalen Gerbrauch ergeben., für zen Monate garantiert. Damit wir Ihnen einem erstklassigen Service dienen konnen, ist es unbedingnotwending, dass wir wissen wo Sie sind und welches Gerat Sie haben. Darum bitten wir Sie, die beiliegende Garantiekarte auszufullen, und Sie uns per Post zuzuschicken. PBI bietet Ihnen auch einen vorbeugenden Kundendienst, um dafur zu sorgen, dass Ihre Apparate zu allen Zeiten in Ordun sind, dass Sie regelmassig kontrolliert werden, einen Express –Krisen-Service, wenn ein Apparat ausfallt. Ihr Verkaufs Agent hat alle Details des Service Kontraktes und kann Ihnen Preise offerieren. A TODOS NUESTROS NUEVOS CLIENTES El equipo adquirido por Vd. Está garantizado por doce meses contra cualquier defecto de fabricación que pudiera producir avería durante el uso normal. Con el fin de poderle ofrecer un servicio eficiente, es esencial que sepamos su nombre o razón social, dirección y equipo adquirido. Para ello, rogamos que por favor rellene el certificado de garantía adjunto y nos lo envíe por correro certificado. PBI también puede ofrecer un servicio postventa de mantenimiento preventivo para mantener el perfecto estado de funcionamento de su equipo, asì como ofrecerle una atencion normal y un rapido servicio en el caso de que su equipo falle.

Page 15: Steam Sterilizer “TIMO” “TIMO- PRINT” - Code n. 88944 - 88957shop.skolebutik.dk/images/pdf/timo.pdf · International pbi S.p.A. « Timo » « Timo Print » Instructions manual

DECLARATION OF CONFORMITY

International PBI S.p.A. Via Novara, 89 - 20153 Milano - Italy

HEREBY DECLARES under its sole responsibility that the following product(s)

“TIMO” autoclave code 88944 - 88957

♦ conform(s) with the technical requirements of the following standard(s) CEI 66-5 (EN 61010-1), EN 55011, CEI EN 50081-1, CEI EN 50082-1

♦ EC Directive 2002/95/EC of the European Parliament: “Restriction of the use of

certain hazardous substances” (RoHS) for electric and electronic devices The EC RoHS Directive has been released for the restriction of the use of certain hazardous substances in electric and electronic devices and basically says that with July 1st 2006 new

productions of electric and electronic devices may not contain lead, mercury, cadmium, hexavalent chromium, polypromined biphenyl (PBB), respectively polybromined diphenylether (PBDE).

As a consequence of this EC Directive we have to introduce unleaded technology in our production and we have to make adaptations concerning the use of lead, mercury, cadmium, hexavalent

chromium, PBB, PBDE according to the defined permissible values. The adaptation of our products to the RoHS conformity is confirmed.

♦ EC Directive 2002/96/EC of the European Parliament: “Disposal of old products”

Our products are designed and manufactured with high quality materials and components which can be recycled and reused according to the EC Directive 2002/96/EC. The user should inform himself about the separate collection system for electrical and electronic products. He should act according

to the local rules and do not dispose of the old products with the normal household waste. The correct disposal of old products will help to prevent potential negative consequences for the

environment and human health.

♦ per the provisions of 73/23/EEC, 89/336/EEC and 97/23/EEC Directives

President

Roberto Ligugnana Milan 2007

International pbi S.p.A. « Timo » « Timo Print » Instructions manual Rev. 1 15