Page 1: Status of APD Screening in Bochum · After the initial screening: Irradiation of APDs with 37Gy Annealing for 48 hours at 80 C. Short circuit of the APD pins during irradiation and

Status of APD Screening in BochumPanda Collaboration Meeting 18/3

Jan Reher

Ruhr-Universitat Bochum

Page 2: Status of APD Screening in Bochum · After the initial screening: Irradiation of APDs with 37Gy Annealing for 48 hours at 80 C. Short circuit of the APD pins during irradiation and

Readout of EMC crystals

Readout of crystals for the forward endcap EMC:I 768 crystals in the endcap (very close to the beam pipe)

are read out by VPTTsI Most crystals read out by two APDs each

Full EMC is read out by ≈ 30 000 APDs

8 APDs are supplied by one power supply and need to bematchedProperties vary between APDsProperties are dependant on temperatureValues provided by manufacturer were measured at 20 ◦C

⇒ Screening and matching of all APDs for T = −25 ◦C isrequired!

,1 Jan Reher APD Screening Status

Page 3: Status of APD Screening in Bochum · After the initial screening: Irradiation of APDs with 37Gy Annealing for 48 hours at 80 C. Short circuit of the APD pins during irradiation and

APD screening setup in Bochum

Each screening board provides 16 slots to simultaneouslyscreen 15 APDs and one reference APDAPD signals read out by original Panda preamplifiers andSADCBias voltage and current measurements provided by isegHV modulesIllumination of APDs by DC light and by light pulserresembling the signal shapes of crystalsFull setup with two boards located inside climate chamber

⇒ Capacity to simultaneously screen 30 APDs⇒ Extension of screening capacity by 30 APDs in second

climate chamber in progress

,2 Jan Reher APD Screening Status

Page 4: Status of APD Screening in Bochum · After the initial screening: Irradiation of APDs with 37Gy Annealing for 48 hours at 80 C. Short circuit of the APD pins during irradiation and

APD screening setup in Bochum

,3 Jan Reher APD Screening Status

Page 5: Status of APD Screening in Bochum · After the initial screening: Irradiation of APDs with 37Gy Annealing for 48 hours at 80 C. Short circuit of the APD pins during irradiation and

Measurement process

Every APD is screened both at 20 ◦C and -25 ◦CPrevious studies1 showed: Measurement of dark currentsand pulse shapes can be performed in parallel.

→ Four measurements in total: DC- and dark currents at in awarm and cold environmentDuration for a a full set of measurements ≈ 10 hAt full capacity, screening 60 APDs per day is possible→300 APDs per weekWith duplicated setup: 600 APDs per week.As of today, ≈ 1000 new APDs and ≈ 350 irradiated APDswere screened and evaluated at Bochum

1Leon Knarr, Bachelor’s thesis

,4 Jan Reher APD Screening Status

Page 6: Status of APD Screening in Bochum · After the initial screening: Irradiation of APDs with 37Gy Annealing for 48 hours at 80 C. Short circuit of the APD pins during irradiation and

Exemplary gain curve

0.1

1

10

100

1000

10000

0 50 100 150 200 250 300 350 400

M

U [V]

,5 Jan Reher APD Screening Status

Page 7: Status of APD Screening in Bochum · After the initial screening: Irradiation of APDs with 37Gy Annealing for 48 hours at 80 C. Short circuit of the APD pins during irradiation and

Bias voltages for M = 100 and M = 200 for forward endcap

320 340 360 380 400 420 440U(M=100)[V]

0

50

100

150

200

250

300

AP

Ds

T=20

T=-25

320 340 360 380 400 420 440U(M=200)[V]

0

50

100

150

200

250

300

AP

Ds

T=20

T=-25

,6 Jan Reher APD Screening Status

Page 8: Status of APD Screening in Bochum · After the initial screening: Irradiation of APDs with 37Gy Annealing for 48 hours at 80 C. Short circuit of the APD pins during irradiation and

Comparison to values provided by Hamamatsu atM = 100

Mean 0.6002

Std Dev 1.71

10− 8− 6− 4− 2− 0 2 4 6 8 10 [V]M=100, Bochum - UM=100,HamamatsuU

0

20

40

60

80

100

120

140

160

180

Eve

nts

/ (0.

4 V

)

Mean 0.6002

Std Dev 1.71

Slight deviation from the manufacturers values is observedDeviations are well within parameters explained bytemperature uncertainties (∆T ≤ 2 K)

,7 Jan Reher APD Screening Status

Page 9: Status of APD Screening in Bochum · After the initial screening: Irradiation of APDs with 37Gy Annealing for 48 hours at 80 C. Short circuit of the APD pins during irradiation and

Temperature dependency of bias voltages

0.4 0.5 0.6 0.7 0.8 0.9 1 1.1 1.2 1.3 1.4/dT [V/K]

biasdU

0

50

100

150

200

250

300

350

even

ts /

(0.0

2 V

/K)

M=100

M=150

M=200

Distribution of difference between bias voltages atT = 20◦C and T = −25◦C is narrow→ Linear correlationbetween bias voltage and temperature

,8 Jan Reher APD Screening Status

Page 10: Status of APD Screening in Bochum · After the initial screening: Irradiation of APDs with 37Gy Annealing for 48 hours at 80 C. Short circuit of the APD pins during irradiation and

SADC noise measurements

Noise measured at the SADCs depends on the darkcurrent in the APDMaximum acceptable dark current can be estimated fromthe noise at the SADCNoise limit of 3 · σ(M = 200) ⇒ Idark,max = 10 nA

,9 Jan Reher APD Screening Status

Page 11: Status of APD Screening in Bochum · After the initial screening: Irradiation of APDs with 37Gy Annealing for 48 hours at 80 C. Short circuit of the APD pins during irradiation and

Slope at M = 200

0.02 0.04 0.06 0.08 0.1 0.12 0.141/M dM/dU [1/V]

0

20

40

60

80

100

120

140

160

180

200

220

even

ts /

(0.0

03 1

/V)

,10 Jan Reher APD Screening Status

Page 12: Status of APD Screening in Bochum · After the initial screening: Irradiation of APDs with 37Gy Annealing for 48 hours at 80 C. Short circuit of the APD pins during irradiation and

Selection criteria for forward endcap

Maximum dark current at M = 800:

Idark (M = 800) < 10 nA

Matching criteria:I ∆Ubias at M = 200I dM

dU at M = 200

,11 Jan Reher APD Screening Status

Page 13: Status of APD Screening in Bochum · After the initial screening: Irradiation of APDs with 37Gy Annealing for 48 hours at 80 C. Short circuit of the APD pins during irradiation and

Irradiation effects

After the initial screening: Irradiation of APDs with 37 GyAnnealing for 48 hours at 80 ◦C.Short circuit of the APD pins during irradiation andannealingSecond screening iteration after irradiation and annealingContinuous checks on the effects of irradiation, e.g. on biasvoltages or temperature gradient

,12 Jan Reher APD Screening Status

Page 14: Status of APD Screening in Bochum · After the initial screening: Irradiation of APDs with 37Gy Annealing for 48 hours at 80 C. Short circuit of the APD pins during irradiation and

Summary and outlook

APD screening in Bochum ongoing, current capacity of300 APDs per week, soon to be doubled by second setupResults are in agreement with data from the manufacturerSelection criterium for APDs for the forward endcap:I IDark (M = 800) < 10 nA

Matching criteria:I ∆Ubias at M = 200I dM

dU at M = 200

Work on measurement process focussed on increasingoutput

,13 Jan Reher APD Screening Status