Calibration of a Carbon ARTEXSplitex Plate Set Carbon (216100C)

Page 2: Splitex Calibration Carbon Light

Seite 2

(1)Magnet Lifter

(2)Assembly Key

(3)Assembly Plate

(4)Magnet Pot

(5)Socket-head cap screw

(6)Maxilla – Base Plate

(7)Mandibular – Calibration

Plate

(8)Glue Plate

Page 3: Splitex Calibration Carbon Light

Seite 3

Standard Setting for Calibration

> Incisal Pin set to „0“. > Articulator settings in average :

> Benett angle 10-15°> Condyle inclination 30-35°

Page 4: Splitex Calibration Carbon Light

Seite 4

>Take out the model plates of Artex articulator.

Page 5: Splitex Calibration Carbon Light

Seite 5

>Take out the magnets using the magnet lifter.

Page 6: Splitex Calibration Carbon Light

Seite 6

>Loosen and unscrew the magnet pot with assembly key.

Page 7: Splitex Calibration Carbon Light

Seite 7

>Screw in the mounting plate instead of the magnet pot down to the lock.

Page 8: Splitex Calibration Carbon Light

Seite 8

>Unscrew the magnet pot from upper plate>Place the upper plate to the upper arm of the Artex

Page 9: Splitex Calibration Carbon Light

Seite 9

>Bolt on the upper plate with the socket-head cap screw to the mounting plate in the upper arm of the articulator.

Page 10: Splitex Calibration Carbon Light

Seite 10

>Screw in the magnet pot to the Splitex upper plate.

Page 11: Splitex Calibration Carbon Light

Seite 11

>Replace the magnet

Page 12: Splitex Calibration Carbon Light

Seite 12

>Take out the model plates of Artex articulator.

Page 13: Splitex Calibration Carbon Light

Seite 13

>Take out the magnet

Page 14: Splitex Calibration Carbon Light

Seite 14

>Loosen and unscrew the magnet pot with assembly key.

Page 15: Splitex Calibration Carbon Light

Seite 15

>Screw in the mounting plate instead of the magnet pot down to the lock.

Page 16: Splitex Calibration Carbon Light

Seite 16

>Isolate the Carbon-surface in the area of the glue plate with a thin layer of vaseline.

Page 17: Splitex Calibration Carbon Light

Seite 17

>Set in the glue plate with chamfering to the rear> (Important for later separation of the glueing!)

chamfering

Page 18: Splitex Calibration Carbon Light

Seite 18

>Bolt on the glue plate with socket-head cap screw.

Page 19: Splitex Calibration Carbon Light

Seite 19

>Put the Artex-Articulator upside down and open the Centric.

Page 20: Splitex Calibration Carbon Light

Seite 20

>Place the Splitex-Key (216010C) on the upper plate TOP to TOP.

Page 21: Splitex Calibration Carbon Light

Seite 21

>Place the lower plate with the magnet on the key.

Page 22: Splitex Calibration Carbon Light

Seite 22

>Clean the glue plate and the lower side (milling) of the lower plate (grease-free)

Page 23: Splitex Calibration Carbon Light

Seite 23

>Spray the activator into the milling of the lower plate.

>Spray the activator on the glue plate.

Page 24: Splitex Calibration Carbon Light

Seite 24

>Put on two tracings of glue - left and right of the drillings. Not into the drillings!

Page 25: Splitex Calibration Carbon Light

Seite 25

>Close the articulator.

>Close the centric.

Page 26: Splitex Calibration Carbon Light

Seite 26

>Press the articulator frontal for approx. 5 sec. smoothly against the incisal pin.

>leave the articulator in position for 1 h00 before you can unmount the key

>For best results, use after 24 to 48 hours folowing the calibration

Page 27: Splitex Calibration Carbon Light

Seite 27

>Mounting plates