Spectrometer Solenoid Fabrication Update Steve Virostek Lawrence Berkeley National Lab NFMCC at LBNL January 25, 2009

Upload: others

Post on 27-Apr-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

Spectrometer Solenoid

Fabrication Update

Steve VirostekLawrence Berkeley National Lab

NFMCC at LBNL

January 25, 2009

Page 2: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 2

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

MICE Cooling Channel Layout

Spectrometer Solenoid #1

(partially disassembled)

Spectrometer Solenoid #2

(final assembly under way)

Page 3: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 3

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

•First magnet completed after several delays caused by both design and assembly errors; testing revealed several issues

•Operational issues: helium lines blocked by N2, poor thermal connection to radiation shield, insufficient cold mass venting

•Mechanical issues: cold mass alignment, support stand height, iron shield support pads, support stand offset

•Issues resolved on 2nd magnet: new cooling circuit, added relief vent line, LN reservoir added, thermal shield connection improved, support stand modified, portable CMM measurements during assembly will ensure proper alignment

•Second magnet in process of final assembly (w/modifications)

•First magnet has been partially disassembled to add fixes

Summary

Page 4: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 4

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Vacuum Vessel and Thermal Shield

Shield assembled w/5 mm Al sheet

Vacuum vessel ready for cold mass installation

Page 5: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 5

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Completion of Cold Mass Installation

Page 6: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 6

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

HTS Leads and Cryocooler Sleeves

Page 7: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 7

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Service Tower Components Wrapped in MLI

Prior to final welding of tower covers

Page 8: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 8

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Completed Service Tower

Page 9: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 9

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Installed Cryocoolers

Removable panel allows access to leads

Page 10: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 10

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Instrumentation Feedthroughs

All instrumentation wired and ready for testing

Page 11: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 11

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Completed First Magnet

Page 12: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 12

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Vacuum Leak Check

No leaks found in cold mass or vacuum vessel

Page 13: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 13

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Ice on magnet leads during cool down

Magnet Cool Down

Vent line Fill line

Page 14: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 14

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

•Magnet cool down accomplished using LN2 & LHe

•Cool down is done slowly to control cold mass thermal stress (i.e. limited ∆∆∆∆T)

•Cold mass of first magnet reached LHe temperature with no observed problems

•Cold mass temperature could not be maintained with the cryocoolers (no re-condensing)

•Apparent blockage in LHe cooling line occurred due to freezing of LN2 or air

•All five coils reached a maximum current of 196 amps during training

Magnet Cool Down and Training

Page 15: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 15

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Cryocooler Testing

All 4 coolers met the mfg. specifications

Page 16: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 16

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Magnet Discharge Circuit

•Rapid discharge circuit designed and fabricated by Wang NMR

•Each magnet will have a dedicated circuit

•The water cooled, aluminum body assemblies are rack mounted

•Water cooling is only needed during and after a discharge

Page 17: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 17

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Cryocooler Compressors

Page 18: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 18

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Magnet Power Supply Racks

Racks include 300 A power supplies, 60 A power supplies and magnet discharge circuits

Page 19: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 19

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Spectrometer Solenoid Iron Shields

•Design integrated with TOF shield; force analysis and fabdrawings done at Oxford

•Two sets of shields needed at FNAL and one initial set at RAL

•Fabrication at JK Mfg near FNAL for two shield sets (420 and 600 mm holes) almost complete

•One set of shields to be sent to RAL with the first magnet

Page 20: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 20

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Issues with Magnet #1

• Frozen N2 in cold mass helium lines prevented proper operation (mainly procedural, partially design issue)

• Inadequate thermal connection between 1st stage of cold heads and cold mass radiation shield

• Venting of cold mass during quench is not sufficient due to crowding of vent line with instrumentation wires

• No provision for direct cooldown of cold mass radiation shield (i.e. long cooldown time)

• Mechanical issues: magnet alignment in vacuum vessel, support stand height, iron shield support pads, support stand offset

Page 21: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

•Opening of Magnet #1 will be required to perform fixes

•Helium line blockage: connect cryocooler condenser directly to cryostat, improve cooldown procedure

•Radiation shield connection: increase thermal conduction with thick 1100 Al flexible straps (instead of thin Cu)

•Cold mass vent: add additional vent line to the cold mass

•Radiation shield cooldown: incorporate LN reservoir to allow direct cooldown of shield

•Mechanical issues: realign Magnet #1 during reassembly (also corrects support stand height), modify end supports to clear iron shield mount pad (Magnets #1 and #2)

Design Solutions

Page 22: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 22

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Cryostat and Cooling System Mods

Improved radiation shield connection

Cold head 1st stage

Radiation shield

Additional vent line

Direct cryostat connection option

Page 23: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 23

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Magnet Cooling Modifications

Original Design

Cold head

Condenser

Cold massshell

Coils

LHe line

Trap

A written cooldown procedure will be generated to ensure that all N2 is purged from the system prior to introducing liquid helium

Modified Design

Directconnection

LHe

Page 24: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 24

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Liquid Nitrogen Reservoir

LN reservoir

Radiation shield

Thermal plate connection

1st stage cooler connection

Vent/fill lines (3) •Reservoir provides direct LN cooldown of shield

• Inproves thermal connection between 1st

stage of cryos and shield

•Frozen mass of nitrogen protects leads in event of power failure

•Reservoir may be empty during operation due to safety concerns

Page 25: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 25

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

•Vacuum vessel end flange center is not well aligned to coil centers (up to 10mm offset w/current install procedure)

Mechanical Issues

Cold mass center is not well aligned with the vacuum vessel shell

Page 26: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 26

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

• Support stand cross beams interfere with the planned location ofthe lower iron shield saddle plate

Mechanical Issues

Area of interference

Page 27: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 27

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Support Stand Modification

Vacuum vessel

Support braces

Iron shield lowermount and pad

Original Design Modified Design

New braces

Vendor will modify both ends of both magnets

Page 28: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 28

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Modified Cooling System

Liquid/vapor helium accumulator base and cold mass connections

Direct cryocoolersleeve connection with bellows

Page 29: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 29

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Leak Test of Cooling System

Liquid/vapor He accumulator and cryocooler sleeves

Page 30: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 30

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Liquid Nitrogen Reservoir

LN2 reservoir to be thermally connected to radiation shield

Plates welded to radiation shield for reservoir connection

Page 31: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 31

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Radiation Shield Thermal Connection

Thicker 1100 series aluminum connection to thermal shield (previously thin copper)

Page 32: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 32

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Fully Assembled Cold Mass and Shield

Page 33: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 33

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

2nd Magnet Cold Mass Ready to Install

•Helium cooling circuit modification is complete

• Improved thermal shield connection added

• Second cryostat vent tube added

•LN reservoir installed on thermal shield

Page 34: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 34

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

2nd Magnet Components before Assembly

Vacuum vessel

Radiation shield

Cold mass

Page 35: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 35

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Vacuum Vessel and Support Stand

End flange

Modified support stand legs

Page 36: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 36

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Second Magnet Assembly

• Second magnet in process of final assembly

•All system improvements have been incorporated

Page 37: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 37

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Second Magnet Assembly

Page 38: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 38

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

Magnet CMM Measurement

•Portable CMM will be used to align cold mass to end flange and set height

•Tooling balls will also be added to vessel and fiducialized to cold mass

Page 39: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 39

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

First Magnet Partially Disassembled

Page 40: Spectrometer Solenoid Fabrication Update · 2009-01-25 · MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 21 Steve Virostek --Lawrence Berkeley National Laboratory

MICE Spectrometer Solenoid Fabrication Update and Schedule Status Page 40

Steve Virostek -- Lawrence Berkeley National Laboratory -- January 25, 2009

•The plan to move forward involves completing and testing Magnet #2 first (with all modifications)

•Several technical issues as well as negotiation with the vendor on change costs has delayed completion

•Completion of the first unit (Magnet #2) and start of cooldown & testing is expected to occur in late February

•Upon shipping of Magnet #2 (now first unit), Magnet #1 will be modified (reconfigure cooling, add vent line, LN reservoir, etc.)

•The reassembly and completion of Magnet #1 should follow by two to three months

Schedule