socorex self-refilling syringes quick guide guide... · 2018. 8. 20. · socorex isba sa -...
TRANSCRIPT
SOCOREX® SELF-REFILLING SYRINGESQuick Guide
Detach plunger, clean and let dry.
1
Unscrew body and remove.
521 43
5
2
1
43
521 43
521 43
21
521 43
521 43
521 43
521 43
21
Volume setting
Unscrew locking nut. Adjust plunger to desired volume. Tighten locking nut. Adjust body / grip position using locking screw on handle side.
Body setting
Installing accessories: Tubing Vial holder
Attach tubing to aspirationcannula, insert into bottle.
Pierce rubber stopper with vent cannula.
Attach tubing to syringe. Insert connecting tube and needle, assemble parts.
Pierce bottle rubber stopper, lock bottle.
Priming In case of clogged valves
Activate handle until liquid reaches needle.
Activate in vertical position to expell liquid and eliminate any air bubble.
Introduce small pin and push to loosen ball in valve.
If insufficent, disassemble valves and clean.
Daily maintenance: Purge, disassembling and cleaning
Rinse with warm water and purge completely.
Clean residues in glass barrel with a brush.Reassembling
Carefully reassemble glass barrel into body. Attach plunger to handle rod.
Gently insert plunger into glass barrel then screw body.
Activate one or two timesto check proper functioning.I.SYR.E B0818
Socorex Isba SA - Switzerland - www.socorex.comTel. +41 (0)21 651 6000 - [email protected]