slit lamp biomicroscope
TRANSCRIPT
Slit Lamp Bimicroscope
Done by : Isa Mushtaha
To: Dr. Hamed AL – Khodari.
Slit lamp assessment is considered to be the gold standard device for the assessment of the anterior segment of the eye in clinical practice
This is because they provide…Excellent image qualityStereoscopic imageFlexible illuminationFlexible magnification
Therefore there are many different usesEven more when attachments are added
What are the uses of slit lamp
own On their With accessoriesRoutine
examination of anterior segmentAdnexa through
to anterior vitreous
Problem-based examination of anterior segment
GonioscopyFundoscopyOcular
photographyContact
tonometry (Goldmann)
Direct contact goniolenses
• The lamp facilitates an examination of the anterior segment, or frontal structures of the human eye which includes the : 1. Eyelid &eye lashes 2. Cornea3. Sclera4. Conjunctiva5. Iris6. Aqueous7. Natural crystalline lens and 8. Anterior vitreous.
Basic Design1. Viewing arm
Biomicroscope Adjustable focus eyepieces Magnification dial
2. Illumination arm The “slit lamp” Slit size, shape and filter controls Variable size, shape, colour and brightness
3. Biomicroscope and illumination are mechanically coupled around central pivot point (copivotal)
Both focus at the same point (parfocal) Both arms can swing independently 180º along
horizontal – there is a scale in degrees Both always central regardless of angle (isocentric)
4. Moveable base plate and joystick control
Types of illumination incorporated in slit lamp1. Slit width
Wide- survey globe/corneaNarrow- depth, width & position of small
abnormalitiesbeam as wide as cornea is thickforms a parallelepiped volume: a box of illuminated tissue is seen
Thin (slit)- narrowest beam forms an optical section so thin it's just discernible التمييز سهلvaluating small changes in clarity & pinpointing الدقة depth of pathology
2. Light-source intensityMedium to high: most purposes High: optical section
3. Filtersneutral, cobalt blue (for fluorescein),
red-free and red free filter (green filter).4. Magnification
low power (~10x) is used for surveymedium to high (16-40x) for optic
section & parallelepiped Higher than (40x) for specular reflection
normally, light is focused at same point as microscope (“parfocal”)
1. Start with neutralizing the eye piece at zero if the examiner has no refractive error or wearing his glasses & lower powered objective lens .
2. Use lowest voltage setting on transformer
ensure open aperture 3. Select the longest slit length4. Adjust chin rest :
Patient's eyes approximately with level of the marker on head rest .
5. Lamp height of the slit beam centered vertically on Patient's medial canthus
6. Focus by moving joystick
How the examiner can adjust and use the slit lamp
locking nut: loose for free movementOcular focus to 0
adjust beam height for tall, narrow vertical beam
adjust width for narrow beam w/ good illumination
Methods of examination of slit lamp without attachments
There are six basic methods of illumination1. Diffuse illumination 2. Direct focal illumination 3. Indirect illumination 4. Sclerotic scatter 5. Retro-illumination6. Specular reflection
1. Diffuse Illumination : This method is done by using a wide
slit which slightly out of focus and this method is used in examination of the iris and the adnexa of the eye.
The slit should be wide and the magnification should be low as possible to prevent loss of field and enable
large field of view .
2. Direct focal illumination : •Illumination and observation are focused
in the same plane •slit width narrow to broad
•Illumination angle 45° to 60° •Magnification 10x-40x
This method is used to examine the cornea in details , the anterior chamber , crystalline lens , anterior part of the vitreous and in determine if there is flare , pus or blood cells in the anterior chamber .
3. Indirect illumination : •The beam is focused in an area
adjacent to ocular tissue to be observed
•Decentered beam •Illumination 2 to 4mm slit
•Magnification: Low to medium (depending upon object size) .
This method is used to determine if there is infiltrates رشح , corneal scars , deposits ترسبات and epithelial and stromal defects .
4. Sclerotic scatter : •Light incident on the limbus
with 2-4mm slit at an angle of 45° - 60°
•Decentered slit •The microscope focused
centrally •Total internal reflection of the
incoming light at inner corneal boundaries (endothelium and epithelium).
This method is used in examination of scars, foreign bodies, corneal defects and
irregularities in the cornea .
5. Retro-Illumination from the Fundus : This technique is used to observe media clarities
and opacitiesصفاء •The pupil is dilated
•the slit beam and microscope are made co-axial and light strikes the fundus and creates a glow behind the opacity in the media
•The media opacity creates a shadow in the glow Applications
•abnormities in the anterior vitreous, lens, anterior chamber, cornea
6. Specular Reflection: it is useful slit lamp technique for scrutinsing the corneal endothelium , the tear film surface , crystalline lens and other surfaces .
A bright reflection will be observed in the anterior surface and less bright reflection will be observed in posterior surface of the cornea .The illumination is narrow parallelepiped and the microscope is placed directly in front of the eye with the light source 25degree from
the microscope . The endothelium will appear clear and rough surface ( due to individual endothelial cells ) .
Attachments of the Slit lamp There are many different attachments and accessories,of these :1. video attachment :used to facilitate
teaching of slit lamp .2. Goldman tonometer :used in measuring
the intra ocular pressure of the anterior chamber .
3. Gold 3-mirror lens : used in measuring the angle of the anterior chamber and center of the retina .
4. Volk double aspheric lens (+60,+78,+90) :used in examination the fundus os the patient indirectly .
5. Direct contact goniolenses : for examination of the periphrey of the fundus .
...
...Thank you for listening
?؟Any questions