siemens acuson x300

32
© Siemens AG The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Print No.: Siemens Medical Solutions Ultrasound ejt ACUSON X300 System Software US 2006 US11-101.816.01.01.02 Replaces: n.a. English Doc. Gen. Date: 09.06 n.a. Part No.: sd Loading Software

Upload: ricardo-salazar

Post on 10-Jul-2016

528 views

Category:

Documents


83 download

DESCRIPTION

US11-101.816.01.01.02

TRANSCRIPT

Page 1: Siemens Acuson x300

© Siemens AGThe reproduction, tranof this document or its contents is notpermitted without express writtenauthority. Offenders will be liable fordamages. All rights, including rightscreated by patent grant or registrationof a utility model or design, arereserved.

Siemens Medical SolutionsUltrasoundejt

ACUSON X300

System

Software

US

smission or use

2006

Print No.: US11-101.816.01.01.02Replaces: n.a.

ED

nglishoc. Gen. Date: 09.06

n.a.

Part No.: sd

Loading Software

Page 2: Siemens Acuson x300

2 Revision / Disclaimer

1Revision / Disclaimer

Copyright

Copyright © 2006 by Siemens Corporation. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, stored in retrieval systems, or translated into any language or computer language, in any form or by any means, electronic, mechanical, magnetic, optical, chemical, manual, or otherwise, without the prior written permission of Siemens Corporation. Siemens Corporation reserves the right to change its products and services at any time. In addition, this manual is subject to change without notice. Siemens Corporation welcomes customer input on corrections and suggestions for improvements to this manual.Although Siemens Corporation has attempted to ensure accuracy through-out this manual, Siemens Corporation assumes no liability for any errors or omissions, nor for any damages resulting from the application or use of this information.

Trademarks

ACUSON, Sequoia, Cypress, AEGIS, Aspen, 128XP, XP, AcuNav, CV70, SONOLINE, SONOVISTA, Adara, Antares, G20, G40, G50, G60S, X300, X500, Omnia, Premier and KinetDx are trademarks of Siemens Corporation registered in the U.S. Patent and Trade-mark Office. CWS3000, DS3000, WS3000, and WebPro are trademarks of Siemens Cor-poration.

Windows XP, 2000 and NT are registered trademarks of Microsoft Corporation. Internet Explorer is a trademark of Microsoft Corporation. Netscape and Netscape Navigator are registered trademarks of Netscape Communications Corporation. Java and the JavaScript name are registered trademarks of Sun Microsystems, Inc. Adobe Acrobat is a registered trademark of Adobe Systems, Inc.

Other products and brand names are trademarks of their respective owners.

Disclaimer

The service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Sie-mens or one of its affiliates to provide such services.

Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro-cedures.

License Agreement

All computer programs copyright 1990-2006 by Siemens Corporation or its suppliers. Pro-grams are licensed under the following agreement:

Siemens or its suppliers retain(s) ownership of and title to any computer program supplied with the equipment and to the trade secrets embodied in such computer programs. Sub-ject to the Buyer’s acceptance and fulfillment of the obligations in this paragraph, Siemens grants the Buyer a personal, non-transferable, perpetual, non-exclusive license to use any computer program supplied with the Equipment that is necessary to operate the Equip-ment solely on the medium in which such program is delivered for the purpose of operat-ing the equipment in accordance with the instructions set forth in the operator’s manuals supplied with the Equipment and for no other purpose whatsoever. Buyer may not reverse-assemble, reverse-compile or otherwise reverse-engineer such computer pro-grams nor may Buyer make a copy of such program or apply any techniques to derive the

ACUSON X300 US11-101.816.01.01.02 Siemens09.06 sd

Page 2 of 32Medical Solutions USA, Inc.

Page 3: Siemens Acuson x300

Revision / Disclaimer 3

trade secrets embodied therein. In the event of a failure by Buyer to comply with the terms of this license, the license granted by this paragraph shall terminate. Further, because unauthorized use of such computer programs will leave Siemens without an adequate remedy at law, Buyer agrees that injunctive or other equitable relief will be appropriate to restrain such use, threatened or actual. Buyer further agrees that (i) any of Siemens"s suppliers of software is a direct and intended beneficiary of this end-user sublicense and may enforce it directly against Buyer with respect to software supplied by such supplier, and (ii) No supplier of Siemens shall be liable to buyer for any general, special, direct, indi-rect, consequential, incidental or other damages arising out of the sublicense of the com-puter programs supplied with the equipment.

Siemens US11-101.816.01.01.02 ACUSON X30009.06 sd

Page 3 of 32Medical Solutions USA, Inc.

Page 4: Siemens Acuson x300

4 Table of Contents

1- 0Table of Contents

1 _______ Loading System Software_________________________________________ 5

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Disk Partition Map . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Setup Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Tools and Materials. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Reference Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Time Required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

2 _______ Verify System Software___________________________________________ 7

Verify the System Software Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

3 _______ Loading System Software_________________________________________ 9

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

Setup Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Tools and Materials. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Reference Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Time Required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

Verify the System Software Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

Backing Up Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

Installing the System Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13Configuring Peripheral Video and Language Format . . . . . . . . . . . . . . . . . . . . . . . . 21

Verifying Software Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Check/Update the Firmware and BIOS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Verify the Software Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Confirming the Product Serial Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24Restore the Presets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

Sleep Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27System Configuration: Hardware Version 2.0 - Software Version 2.0. . . . . . . . . . . . 27System Configuration: Hardware Version 1.0 - Software Version 2.0. . . . . . . . . . . . 28

4 _______ Changes to Previous Version_____________________________________ 31

Changes to Previous Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

ACUSON X300 US11-101.816.01.01.02 Siemens09.06 sd

Page 4 of 32Medical Solutions USA, Inc.

Page 5: Siemens Acuson x300

Loading System Software 5

2- 1Loading System SoftwareOverview 0

The instructions below describe how to load system software on the system.

Observe the following Caution alerts before proceeding.

CAUTION If the system software will be upgraded (or downgraded to an ear-lier version) for testing or to resolve an operational problem, thefeature keys for options that are not supported by the installed ver-sion will be lost.

To avoid this problem, uninstall all feature keys before upgradingor downgrading the system software. See (US07-102.892.01 / Unin-stalling the Product Feature Keys) for details.

Be sure that all feature keys are properly re-installed after upgrading (or downgrading) software.

Disk Partition Map 0

The table shows the disk partition map.

Tab. 1 System Disk Partition Map

Partition Name Partition Siz File System Contents

C: Remains NTFS • Windows XP

• Software (in-cluding DI-MAQ, syngo, SCR 3D)

D: 60 GB NTFS Patient data and image files (including Quick Set and Imag-ing Preset).

Siemens US11-101.816.01.01.02 ACUSON X30009.06 sd

Page 5 of 32Medical Solutions USA, Inc.

Page 6: Siemens Acuson x300

6 Loading System Software

Setup Requirements 0

Tools and Materials 0

Standard service tools

Installation CD / DVD

Reference Documents 0

None

Time Required 0

Approximately 30 minutes is required to load software. More time will be required if the customer’s data is backed up.

ACUSON X300 US11-101.816.01.01.02 Siemens09.06 sd

Page 6 of 32Medical Solutions USA, Inc.

Page 7: Siemens Acuson x300

Verify System Software 7

3- 2Verify System SoftwareVerify the System Software Version 0

Use the following instructions to verify the system software version of ACUSON X300 ultrasound system.

1. Power on the system and wait for the system to boot up.

2. Press Presets. The Preset Main Menu displays.

Fig. 1: Preset Main Menu

Siemens US11-101.816.01.01.02 ACUSON X30009.06 sd

Page 7 of 32Medical Solutions USA, Inc.

Page 8: Siemens Acuson x300

8 Verify System Software

3. Select Service. The Main Service Menu displays.

Fig. 2: Service Main Screen

4. Verify the system SW version information.

5. Record the system SW Version information.

6. Select Cancel to exit.

NOTE Before starting a software installation, make a backup copy of theuser-defined settings and the patient database. After loading thesystem software, customize the setting of video and language for-mat by following the procedure in US11-101.816.01 Installing theSystem Software.

ACUSON X300 US11-101.816.01.01.02 Siemens09.06 sd

Page 8 of 32Medical Solutions USA, Inc.

Page 9: Siemens Acuson x300

Loading System Software 9

4- 3Loading System SoftwareOverview 0

The instructions below describe how to load system software on the ACUSON X300.

Observe the following Caution alerts before proceeding.

CAUTION If the system software will be upgraded (or downgraded to an ear-lier version) for testing or to resolve an operational problem, thefeature keys for options that are not supported by the installed ver-sion will be lost.

To avoid this problem, uninstall all feature keys before upgradingor downgrading the system software. See (US11-102.816.01 / Li-cense Key Installation) for details.

Be sure that all feature keys are properly re-installed after upgrading (or downgrading) software.

Feature keys do not have to be uninstalled. Experienced customer service engineers (CSEs) may choose to leave the keys installed.

The software generally consists of system Presets and system software.

• System Presets (on CD-1)

• System software (on CD-1 and CD-2)

Siemens US11-101.816.01.01.02 ACUSON X30009.06 sd

Page 9 of 32Medical Solutions USA, Inc.

Page 10: Siemens Acuson x300

10 Loading System Software

Setup Requirements 0

Tools and Materials 0

Standard service tools

Reference Documents 0

None

Time Required 0

Approximately 30 minutes is required to load software. More time will be required if the customer’s data is backed up.

ACUSON X300 US11-101.816.01.01.02 Siemens09.06 sd

Page 10 of 32Medical Solutions USA, Inc.

Page 11: Siemens Acuson x300

Loading System Software 11

Verify the System Software Version 0

Use (Verify System Software / p. 7)to verify the software version before proceeding.

Siemens US11-101.816.01.01.02 ACUSON X30009.06 sd

Page 11 of 32Medical Solutions USA, Inc.

Page 12: Siemens Acuson x300

12 Loading System Software

Backing Up Data 0

• Back up the data, see (US11-104.816.01 / Backing Up Data)

• Back up the presets, see (US11-104.816.01 / Backing Up Data)

• Back up the customers data, see ((US11-104.816.01 / Backing Up Data)

ACUSON X300 US11-101.816.01.01.02 Siemens09.06 sd

Page 12 of 32Medical Solutions USA, Inc.

Page 13: Siemens Acuson x300

Loading System Software 13

Installing the System Software 0

1. Power on the system and wait for it to boot up completely.

2. Press Preset. The Preset Main Menu displays.

Fig. 3: Preset Main Menu

Siemens US11-101.816.01.01.02 ACUSON X30009.06 sd

Page 13 of 32Medical Solutions USA, Inc.

Page 14: Siemens Acuson x300

14 Loading System Software

3. Select Service from the menu on the left side of the screen.

Fig. 4: Service Main Menu

ACUSON X300 US11-101.816.01.01.02 Siemens09.06 sd

Page 14 of 32Medical Solutions USA, Inc.

Page 15: Siemens Acuson x300

Loading System Software 15

4. Select Service from the Main Menu.

Fig. 5: Select Service from the Service Menu

5. The Authentication Screen displays.

Siemens US11-101.816.01.01.02 ACUSON X30009.06 sd

Page 15 of 32Medical Solutions USA, Inc.

Page 16: Siemens Acuson x300

16 Loading System Software

6. Enter the password, then select OK. The Service Main Menu displays.

Fig. 6: Authentication Screen

ACUSON X300 US11-101.816.01.01.02 Siemens09.06 sd

Page 16 of 32Medical Solutions USA, Inc.

Page 17: Siemens Acuson x300

Loading System Software 17

7. Select Utilities to display the Utilities menu.

Fig. 7: Select Utilities from the Service Menu

Siemens US11-101.816.01.01.02 ACUSON X30009.06 sd

Page 17 of 32Medical Solutions USA, Inc.

Page 18: Siemens Acuson x300

18 Loading System Software

8. Select Install Software to install software.

Fig. 8: Select Install Software from the Utilities Menu

The following message displays while the software is installing.

Fig. 9: Installation Message

9. Insert the installation CD / DVD into the CD / DVD-R/RW drive.

ACUSON X300 US11-101.816.01.01.02 Siemens09.06 sd

Page 18 of 32Medical Solutions USA, Inc.

Page 19: Siemens Acuson x300

Loading System Software 19

10. Click OK. The system automatically reboots.

Fig. 10: Reboot Screen

11. After the automatic reboot, the system prompts for the type of upgrade to perform.

Fig. 11: Software Install DOS Screen

- Upgrade/Update Installation retains the current presets.

- Clean/Factory Installation initializes the presets and all patient studies are erased.

Siemens US11-101.816.01.01.02 ACUSON X30009.06 sd

Page 19 of 32Medical Solutions USA, Inc.

Page 20: Siemens Acuson x300

20 Loading System Software

NOTICE Selecting Clean/Factory Installation erases all patient studies fromthe hard disk drive.

Be sure to back up all patient studies before selecting Clean/Factory Installation.

12. Choose the desired upgrade. If a choice is not made in 30 seconds, the system will de-fault to Upgrade/Update Installation.

13. After approximately 7-8 minutes, the system will display a pop-up window asking for the second CD. The CD tray will open. Remove the first CD and insert the second.

Fig. 12: Symantec Ghost CD Install

14. After approximately five more minutes, the upgrade is complete, the system will reboot. After the system completes the reboot, verify the system software version as described in (Verify the System Software Version / p. 7).

15. Press Preset, then select Service --> Configuration --> License & Components to confirm newly installed software version.

ACUSON X300 US11-101.816.01.01.02 Siemens09.06 sd

Page 20 of 32Medical Solutions USA, Inc.

Page 21: Siemens Acuson x300

Loading System Software 21

Configuring Peripheral Video and Language Format 0

1. Press Preset and select Basic System 1 from the left side of the Preset Main screen.

Fig. 13: Language Format Selection

2. Select the Language Format from the Format pane according to customer need.

Be sure that all feature keys are properly re-installed after upgrading (or downgrading) software.

3. Select Save.

Siemens US11-101.816.01.01.02 ACUSON X30009.06 sd

Page 21 of 32Medical Solutions USA, Inc.

Page 22: Siemens Acuson x300

22 Loading System Software

4. Select Peripheral from the left side of the Preset Main screen.

Fig. 14: Peripherals Main Menu

5. Select the Video format from the Peripheral pane according to customer need.

6. Select Save.

7. Restart the system.

ACUSON X300 US11-101.816.01.01.02 Siemens09.06 sd

Page 22 of 32Medical Solutions USA, Inc.

Page 23: Siemens Acuson x300

Loading System Software 23

Verifying Software Operation 0

Use the following instructions to verify that the software is operational and to restore data.

Check/Update the Firmware and BIOS 0

During this last phase of the software installation, the system checks the revision level of the firmware and BIOS, updating them if necessary.

Verify the Software Installation 0

Before continuing, verify the system is now running the new software.

1. Power on the system and wait for the system to boot up.

2. Press the Preset key to display the Preset Main Menu.

Fig. 15: Preset Main Menu

3. Select Service to display the Service Main Menu.

Siemens US11-101.816.01.01.02 ACUSON X30009.06 sd

Page 23 of 32Medical Solutions USA, Inc.

Page 24: Siemens Acuson x300

24 Loading System Software

4. The system information section of the service preset menu displays the current soft-ware version. Verify the system software version.

Fig. 16: Verify System Software Version

Confirming the Product Serial Number 0

1. Confirm that no warning message is displayed on the monitor.

2. Press the Preset key on the keyboard.

3. Select Service in the left pane of the Preset menu.

4. Verify that the correct software version is displayed.

ACUSON X300 US11-101.816.01.01.02 Siemens09.06 sd

Page 24 of 32Medical Solutions USA, Inc.

Page 25: Siemens Acuson x300

Loading System Software 25

5. At the product information screen, verify that the correct product Serial No. is dis-played: the number displayed on the product Information screen should be identical to the product Serial Number recorded on the label attached to the rear of the product.

Fig. 17: Location of System Serial Number

6. If necessary, restore the Presets according to (US11-104.816.01 / Backing Up Data) and perform the instructions in reverse fashion.

7. If necessary, restore the service data. Refer to (US11-104.816.01 / Backing Up Service Data) and perform the instructions in reverse fashion.

8. If necessary, restore the Feature keys according to the instructions found in (US11-102.816.01 / License Key Installation).

Restore the Presets 0

NOTE Note: This procedure is only required when if a backup CD wascreated prior to software installation.

Siemens US11-101.816.01.01.02 ACUSON X30009.06 sd

Page 25 of 32Medical Solutions USA, Inc.

Page 26: Siemens Acuson x300

26 Loading System Software

1. Display the Service menu in the Preset.

- Press the Preset key.

- Select Service in the left pane of the Preset menu.

- Select Go.

- Select Backup & Restore icon in the Service menu window.

- Select Restore. The system will request the back-up disc you have made.

2. Insert the Backup CD in the CD-R/W drive and wait approximately ten seconds for the CD to spin up.

3. Select Restore on the pop-up window to start. The system will display “Restore is completed” message.

4. Select Ok and remove the backup CD.

- The system will display “Shutdown before the system starts restoring” message.

5. Select Ok to confirm.

6. Close Backup &Restore and select Exit twice to return to the preset menu.

7. Shutdown the system manually and wait for it to shut down

8. Restart the system.

ACUSON X300 US11-101.816.01.01.02 Siemens09.06 sd

Page 26 of 32Medical Solutions USA, Inc.

Page 27: Siemens Acuson x300

Loading System Software 27

Sleep Mode 0

The Sleep Mode option is a part of software version 2.0. It can be installed on systems configured with hardware versions 1.0 or 2.0. The Sleep Mode setup is different for each system hardware version. Use the set of instruction that apply to the system software and hardware configuration.

System Configuration: Hardware Version 2.0 - Software Version 2.0 0

NOTE After installing software version 2.0, the default status of SleepMode is inactive. To activate Sleep Mode, select the Show field ofthe Preset Sleep Mode Option.

1. Press the Preset key. The system displays the Preset Main Menu.

2. Select Service from the Presets menu. The system displays the Service Authentica-tion screen.

3. Enter the password and select Go. The system displays the Service Main Menu.

4. Select Configuration. The system displays the Configuration screen.

5. Select System from the Configuration menu.

6. Select Preset Sleep Mode Option. See (Fig. 18 / p. 27) for location of the option.

Fig. 18: System Information Screen > Preset Sleep Mode Option

Siemens US11-101.816.01.01.02 ACUSON X30009.06 sd

Page 27 of 32Medical Solutions USA, Inc.

Page 28: Siemens Acuson x300

28 Loading System Software

The system will display the “Enable Sleep Mode” option on Preset Main Menu, Basic System 1 screen.

7. Select the Save button. The system saves the changes.

8. Select the Close button. The system displays to the Preset Main Menu.

9. Verify the Enable Sleep Mode option is selected. See (Fig. 19 / p. 28) for location of the option.

Fig. 19: Preset Menu > Basic System 1 > Enable Sleep Mode

System Configuration: Hardware Version 1.0 - Software Version 2.0 0

NOTE Systems with hardware configuration version 1.0 do not have thesleep mode function. Therefore, there is no need to display the“Enable Sleep Mode” function on Preset Main Menu screen.

If the system hardware includes the DC/DC power supply(MN: 10010385) and control panel firmware, the Sleep Mode func-tion can be deactivated by deselecting “Enable Sleep Mode”.

1. Press the Preset key. The system displays the Preset Main Menu.

ACUSON X300 US11-101.816.01.01.02 Siemens09.06 sd

Page 28 of 32Medical Solutions USA, Inc.

Page 29: Siemens Acuson x300

Loading System Software 29

2. Select Service from the Preset Menu. The system displays the Service Authentication screen.

3. Enter the password and select Go. The system displays the Service Main Menu.

4. Select Configuration. The system displays the Configuration screen.

5. Select System from the Configuration menu.

6. Deselect the Preset Sleep Mode Option.

Fig. 20: System Information Screen > Preset Sleep Mode Option

The system hide the “Enable Sleep Mode” option unavailable on Preset Main Menu, Basic System 1 screen.

7. Select the Save button. The system saves the changes.

8. Select the Close button. The system displays to the Preset Main Menu.

Siemens US11-101.816.01.01.02 ACUSON X30009.06 sd

Page 29 of 32Medical Solutions USA, Inc.

Page 30: Siemens Acuson x300

30 Loading System Software

9. Verify the Enable Sleep Mode option is not displayed. See (Fig. 21 / p. 30) for location of the hidden option.

Fig. 21: Preset Menu > Basic System 1 > Disable Sleep Mode

ACUSON X300 US11-101.816.01.01.02 Siemens09.06 sd

Page 30 of 32Medical Solutions USA, Inc.

Page 31: Siemens Acuson x300

Changes to Previous Version 31

5- 4Changes to Previous VersionChanges to Previous Version 0

Change(s) Description

New document Initial document.

Siemens US11-101.816.01.01.02 ACUSON X30009.06 sd

Page 31 of 32Medical Solutions USA, Inc.

Page 32: Siemens Acuson x300

32 Changes to Previous Version

ACUSON X300 US11-101.816.01.01.02 Siemens09.06 sd