shower study for si-w prototype ecal with asic readout

17
Shower study for 1-5 GeV energy electron beam with ASIC SKIROC readout Elmaddin Guliyev LLR – Ecole Polytechnique, CNRS/IN2P3 Palaiseau, France 26.09.2012- Analysis meeting

Upload: eguliyev

Post on 27-Jul-2015

107 views

Category:

Education


4 download

TRANSCRIPT

Page 1: shower study for Si-W prototype ECAL with ASIC readout

Shower study for 1-5 GeV energy electron beam with ASIC

SKIROC readout

Elmaddin Guliyev

LLR – Ecole Polytechnique, CNRS/IN2P3

Palaiseau, France

26.09.2012- Analysis meeting

Page 2: shower study for Si-W prototype ECAL with ASIC readout

e- beam

Detector Interface (DIF)

Very front-electronics (FVE8)ASIC+Si wafer

to DAQ (LDA)

Tungsten plate

e- beam energy: 1-5 GeVRate: 1 KHz – 5 Hz

6 layer of detector=6*(Si wafer+4ASIC+DIF)

Page 3: shower study for Si-W prototype ECAL with ASIC readout

1 GeV e- hit with 6 layer of detector: between the layers and in front of 1st layer W plate placed.

All the run was same: 3600 sec.

Page 4: shower study for Si-W prototype ECAL with ASIC readout

Example of MIP for one chip, channel,column

Pedestal

1st MIP

2nd MIP

Page 5: shower study for Si-W prototype ECAL with ASIC readout

MIP position variety:

Page 6: shower study for Si-W prototype ECAL with ASIC readout

MIP position variety:

Page 7: shower study for Si-W prototype ECAL with ASIC readout

Shower profile for 6 layer/detector for 1 GeV electron / X0

Page 8: shower study for Si-W prototype ECAL with ASIC readout

Beam alignment or leakage in configuration????

Page 9: shower study for Si-W prototype ECAL with ASIC readout

Hit per event for different radiation length

Page 10: shower study for Si-W prototype ECAL with ASIC readout

X profile GAUS fit

Use to estimate:

Shower size-sigma from gausBeam alignment-mean from gaus

Page 11: shower study for Si-W prototype ECAL with ASIC readout

Deviation ==> beam alignment

Mean value of Gaus fit of shower on X profile

Page 12: shower study for Si-W prototype ECAL with ASIC readout

Shower size for different radiation length: Size of shower determined from sigma of Gaus fit of shower profile.

Show

er size (sigma gaus)

Page 13: shower study for Si-W prototype ECAL with ASIC readout

e-

Shower study in radial direction for different radiation length.

Page 14: shower study for Si-W prototype ECAL with ASIC readout

Efficiency for the layers (detectors)

1 GeV electron beam

X0 and 6X0 configuration

Page 15: shower study for Si-W prototype ECAL with ASIC readout

Hit for one event in same BCID

Page 16: shower study for Si-W prototype ECAL with ASIC readout

Summary:

Hit distribution for different radiation length analysed.The size of shower estimated for different radiation length.

It seems beam not aligned proper or we had a shower leakage due to particle loose, which gives the discrepancy in end of X0 and beginning of3X0 configuration (same for 3X0 and 6X0).

Need to verify it with same geometry with simulation!

Page 17: shower study for Si-W prototype ECAL with ASIC readout

Hold calibration

Optimized value: 100-110 uDelay