September 1, 2005 IEEE-EMBS, Shanghai, China Authors: Aveh Bastani, MD 1 ; Hani Kayyali, MS, MBA 2 ; Robert N. Schmidt, MS, MBA, JD, PE 2 ; Rizwan Qadir, MD 1 ; Prasanth Manthena, MD 3 Institutions: 1 Troy Beaumont Hospital, Troy, MI; 2 Cleveland Medical Devices, Inc., Cleveland, OH; 3 Northwestern University, Chicago, IL; Wireless Brain Monitoring in the Emergency Department

Upload: bryan-clark

Post on 16-Dec-2015

216 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: September 1, 2005IEEE-EMBS, Shanghai, China Authors: Aveh Bastani, MD 1 ; Hani Kayyali, MS, MBA 2 ; Robert N. Schmidt, MS, MBA, JD, PE 2 ; Rizwan Qadir,

September 1, 2005 IEEE-EMBS, Shanghai, China

Authors: Aveh Bastani, MD1; Hani Kayyali, MS, MBA2; Robert N. Schmidt, MS, MBA, JD, PE2; Rizwan Qadir, MD1;

Prasanth Manthena, MD3

Institutions: 1 Troy Beaumont Hospital, Troy, MI; 2 Cleveland Medical Devices, Inc., Cleveland, OH;

3 Northwestern University, Chicago, IL;

Wireless Brain Monitoring in the Emergency Department

Page 2: September 1, 2005IEEE-EMBS, Shanghai, China Authors: Aveh Bastani, MD 1 ; Hani Kayyali, MS, MBA 2 ; Robert N. Schmidt, MS, MBA, JD, PE 2 ; Rizwan Qadir,

September 1, 2005 IEEE-EMBS, Shanghai, China

Objective

To evaluate the feasibility, quality and utility of four-channels of electroencephalogram (EEG) telemetered from patients presenting with Altered Metal Status in the Emergency Department.

Page 3: September 1, 2005IEEE-EMBS, Shanghai, China Authors: Aveh Bastani, MD 1 ; Hani Kayyali, MS, MBA 2 ; Robert N. Schmidt, MS, MBA, JD, PE 2 ; Rizwan Qadir,

September 1, 2005 IEEE-EMBS, Shanghai, China

• Many modalities exist to evaluate neurologic status of patients such as: • Bispectral Index (BIS), • Cerebral Oximetry, or • Positron Emission Tomography (PET).

• Despite some optimistic trials in the Emergency Department (ED) to evaluate the neurologic status of patients, none has supplanted the need for the Electroencephalogram (EEG).

• EEG is the gold-standard for objectively evaluating the functional neurologic status of patients.

Background

Page 4: September 1, 2005IEEE-EMBS, Shanghai, China Authors: Aveh Bastani, MD 1 ; Hani Kayyali, MS, MBA 2 ; Robert N. Schmidt, MS, MBA, JD, PE 2 ; Rizwan Qadir,

September 1, 2005 IEEE-EMBS, Shanghai, China

• Only a handful of hospitals in the US routinely perform EEGs in the ED. Subsequently, patients in whom an EEG is required are admitted to the hospital with their potential disorder undiagnosed and untreated for many hours or days.

• This Altered Mental Status group currently makes up 10% or 14 million of the 140 million yearly ED visits in the U.S.

Background - Continued

Page 5: September 1, 2005IEEE-EMBS, Shanghai, China Authors: Aveh Bastani, MD 1 ; Hani Kayyali, MS, MBA 2 ; Robert N. Schmidt, MS, MBA, JD, PE 2 ; Rizwan Qadir,

September 1, 2005 IEEE-EMBS, Shanghai, China

•Most emergency departments (ED) do not perform electroencephalogram (EEG) studies. This is due to several inhibiting factors including:

1.The bulk of the equipment makes it inconvenient to be permanently located in an ED setting,

2.The cost of equipment at $20,000-$40,000 per unit is very expensive for most ED budgets,

3.The time and expertise required to set up and monitor an EEG is typically lacking in the ED, and

4.ED personnel are not trained to read EEGs and a neurologist may not be immediately available to read the EEG.

Background - Continued

Page 6: September 1, 2005IEEE-EMBS, Shanghai, China Authors: Aveh Bastani, MD 1 ; Hani Kayyali, MS, MBA 2 ; Robert N. Schmidt, MS, MBA, JD, PE 2 ; Rizwan Qadir,

September 1, 2005 IEEE-EMBS, Shanghai, China

New EEG Equipment is Needed for the Emergency Department

The Crystal Monitor was developed under US National Institutes of Health Grants to have an ED EEG device Wireless, small, unobtrusive Low Cost, ~ 1/2-1/3 of most EEG

machines Fast, easy to use, limited number of

channels Easy to learn how to use

Page 7: September 1, 2005IEEE-EMBS, Shanghai, China Authors: Aveh Bastani, MD 1 ; Hani Kayyali, MS, MBA 2 ; Robert N. Schmidt, MS, MBA, JD, PE 2 ; Rizwan Qadir,

September 1, 2005 IEEE-EMBS, Shanghai, China

The Study

Guided by NIH recommendations and support, Cleveland Medical Devices Inc. (CleveMed) has created a portable telemetry multi-channel EEG monitor. A four-channel montage (Fp1-C3, Fp2-C4, C3-O1, C4-O2, Gnd FpZ) was used to maximize EEG coverage while minimizing electrode set-up time.

Telemetry allows the patient to be un-tethered and moved about freely while still being monitored, an important requirement for any patient being assessed in the ED. An internet connection allows a neurologist to interpret the EEG from anywhere.

Page 8: September 1, 2005IEEE-EMBS, Shanghai, China Authors: Aveh Bastani, MD 1 ; Hani Kayyali, MS, MBA 2 ; Robert N. Schmidt, MS, MBA, JD, PE 2 ; Rizwan Qadir,

September 1, 2005 IEEE-EMBS, Shanghai, China

Patient with Crystal Monitor 16 (older version)

Page 9: September 1, 2005IEEE-EMBS, Shanghai, China Authors: Aveh Bastani, MD 1 ; Hani Kayyali, MS, MBA 2 ; Robert N. Schmidt, MS, MBA, JD, PE 2 ; Rizwan Qadir,

September 1, 2005 IEEE-EMBS, Shanghai, China

Inclusion Criteria:

The following patients will be eligible for study inclusion:1. Patients with known seizure disorder of any type, but with prolonged (> 1 hr.)

post-ictal mental status change.2. Patients with status epilepticus who have received a muscle relaxant for

intubation to determine the presence of subclinical seizures.3. Patients with brief alteration of mental status of unknown origin. (This group

includes new onset seizure disorder, syncope, “spells,” “blackouts,” etc.)4. Patients with behavioral changes that may indicate nonconvulsive seizures

(impaired consciousness, violent outbursts, unusual behaviors, etc.)5. Acute head injury patients with mental status changes that may indicate

nonconvulsive seizures. 6. Patients with a history of previous head injury presenting with new onset

mental status changes. (Head injured patients are at risk for post-traumatic seizures).

7. Patients with neurological exams that may be consistent with focal or partial nonconvulsive seizure. (Eg: aphasia, Todd’s paralysis, etc.)

Page 10: September 1, 2005IEEE-EMBS, Shanghai, China Authors: Aveh Bastani, MD 1 ; Hani Kayyali, MS, MBA 2 ; Robert N. Schmidt, MS, MBA, JD, PE 2 ; Rizwan Qadir,

September 1, 2005 IEEE-EMBS, Shanghai, China

Exclusion Criteria:

The following patients will be excluded from the above groups.

1. Patients who are convulsing.2. Medically or surgically unstable patients.3. Family member, other authorized representative unable to give

informed consent.4. Patients with a head injury incompatible with the use of EEG (eg:

gunshots, severe scalp abrasions, etc.)

Page 11: September 1, 2005IEEE-EMBS, Shanghai, China Authors: Aveh Bastani, MD 1 ; Hani Kayyali, MS, MBA 2 ; Robert N. Schmidt, MS, MBA, JD, PE 2 ; Rizwan Qadir,

September 1, 2005 IEEE-EMBS, Shanghai, China

Transmission and Reception:After EEG was completed, the data was password encrypted and

transmitted to one of two study neurologists.

• The neurologist would then provide a real-time read for the EEG via telephone conversation or email.

• The neurologist also subjectively evaluated the quality of the EEG utilizing the following four point scale:

4 = Excellent quality/Acceptable

3 = Good quality/Acceptable

2 = Fair quality/Acceptable

1 = Poor quality/Unacceptable

• Patients were followed to either attain their discharge diagnosis from the ED or the hospital in the case of admission.

Page 12: September 1, 2005IEEE-EMBS, Shanghai, China Authors: Aveh Bastani, MD 1 ; Hani Kayyali, MS, MBA 2 ; Robert N. Schmidt, MS, MBA, JD, PE 2 ; Rizwan Qadir,

September 1, 2005 IEEE-EMBS, Shanghai, China

48.6 % of the Subjects had EEG’s that were Abnormal

EEG interpretations (5 unusable EEG’s were not included):

• 37/72 (51.4%) EEG’s were interpreted as normal

• 2/37 were diagnosed as pseudoseizure by the ED physician

• 28/72 (38.9%) EEG’s were interpreted as slowing

• 11/28 were patients who clinically appeared post-ictal

• 7/72 (9.7%) EEG’s identified a sub clinical epileptogenic foci

Correlation with Standard Inpatient EEG:

• 24/77 (31.1%) patients with EEG had an inpatient EEG

• 18/24 (75%) were equivalent to the study EEG

• The six dissimilar results are described below

Page 13: September 1, 2005IEEE-EMBS, Shanghai, China Authors: Aveh Bastani, MD 1 ; Hani Kayyali, MS, MBA 2 ; Robert N. Schmidt, MS, MBA, JD, PE 2 ; Rizwan Qadir,

September 1, 2005 IEEE-EMBS, Shanghai, China

75% of the EEG’s Corresponded with the 32 Channel Clinical EEG

Patient # Study EEG result Inpatient Result

14 Left sided seizure activity Post-Ictal State

21 Sharp waves bi-frontopolar channels, possible artifact

Normal

30 Mostly wake EEG Slightly beyond normal limits, suggestive of mild diffuse cerebral irregularity

50 Mostly wake with movement artifact

Diffuse slowing consistent with mild to moderate encephalopathy

57 Mild slowing bilaterally – consistent with mild sedation

Normal EEG

61 Poor Quality EEG Normal EEG

Six ED EEG’s were different than the EEG Lab results that were performed later

Page 14: September 1, 2005IEEE-EMBS, Shanghai, China Authors: Aveh Bastani, MD 1 ; Hani Kayyali, MS, MBA 2 ; Robert N. Schmidt, MS, MBA, JD, PE 2 ; Rizwan Qadir,

September 1, 2005 IEEE-EMBS, Shanghai, China

Discussion and Future Considerations

Understanding that EEG is a time-sensitive modality, it is important that we perform EEG’s when they can be most useful, i.e. in the acute setting.

No enrolled patient failed to complete an EEG. Only 5 of 77 patients (6.5%) had unusable EEG’s primarily due to combination of muscular artifact and gaps in the data for interference during wireless transmission.

An improved radio that can re-transmit lost packets has been developed (Crystal Monitor Model 20) and will be used for the second half of this study.

Based on this data we believe that ED EEG provides valuable information to the ED physician, which can expedite safe medical care. We do not assert that a four-channel EEG is superior or equivalent to the standard EEG.

We do believe its use as a screening tool in the ED provides the ED physician with the additional information necessary to make a more appropriate disposition from the ED.

Page 15: September 1, 2005IEEE-EMBS, Shanghai, China Authors: Aveh Bastani, MD 1 ; Hani Kayyali, MS, MBA 2 ; Robert N. Schmidt, MS, MBA, JD, PE 2 ; Rizwan Qadir,

September 1, 2005 IEEE-EMBS, Shanghai, China

• Dimensions: 135 mm x 63 mm x 25 mm (5.3” x 2.5” x 1”) (not including antenna)• Weight: 210 grams (6.4 oz.) with batteries• Antenna: 76 mm (3.0“) flexible• Number of Input Channels:

• 8 configurable channels (external sensors) plus: • 1 internal position sensor,• 1 pulse oximeter, • 1 airflow sensor, • 1 DC channel Input Range ± 750µV to ± 2V (configurable)

• Resolution 8, 12, 16 bits, configurable• Sampling Rate 960 Samples per second per channel • Filter Input bandwidth 0.5 Hz - 250 Hz (-3dB attenuation); CMRR 100 dB• Noise < 2 µV peak-to-peak (0.5 Hz – 100 Hz)• Input Impedance > 20 MΩ @ 10 Hz• Input Interface Standard no-touch 1.5 mm connectors• Power Supply 2 AA alkaline batteries, Battery Life 12 hours continuous use

Crystal Monitor 20 Specifications

Page 16: September 1, 2005IEEE-EMBS, Shanghai, China Authors: Aveh Bastani, MD 1 ; Hani Kayyali, MS, MBA 2 ; Robert N. Schmidt, MS, MBA, JD, PE 2 ; Rizwan Qadir,

September 1, 2005 IEEE-EMBS, Shanghai, China

Conclusions

Four-channel telemetry EEG used in the ED is feasible, provides good quality screening EEG’s and was able to diagnose underlying seizure in a significant number of patients.

Page 17: September 1, 2005IEEE-EMBS, Shanghai, China Authors: Aveh Bastani, MD 1 ; Hani Kayyali, MS, MBA 2 ; Robert N. Schmidt, MS, MBA, JD, PE 2 ; Rizwan Qadir,

September 1, 2005 IEEE-EMBS, Shanghai, China

This work was supported by NIH Phase II SBIR Grant No. 5 R44 NS042977-03

National Institute of Neurological Disorders and Stroke

US National Institutes of Health

ACKNOWLEDGMENT

Page 18: September 1, 2005IEEE-EMBS, Shanghai, China Authors: Aveh Bastani, MD 1 ; Hani Kayyali, MS, MBA 2 ; Robert N. Schmidt, MS, MBA, JD, PE 2 ; Rizwan Qadir,

September 1, 2005 IEEE-EMBS, Shanghai, China

Robert N. SchmidtCleveland Medical Devices Inc.4415 Euclid Ave., Suite 400Cleveland, Ohio 44103 [email protected]: 1-877-CleveMed (253-8363)

(US Toll Free)Phone Direct: +01-216-619-5925Fax: 216-791-6744

For Questions Contact

Page 19: September 1, 2005IEEE-EMBS, Shanghai, China Authors: Aveh Bastani, MD 1 ; Hani Kayyali, MS, MBA 2 ; Robert N. Schmidt, MS, MBA, JD, PE 2 ; Rizwan Qadir,

September 1, 2005 IEEE-EMBS, Shanghai, China

Smaller Hardware

BioRadio Jr.

• Non-Programmable (factory settings)• 2 channels• 960 sps• Up to 12 bits

• Input Selections• EEG +/- 1 mV, 0.1-70 Hz.• EKG +/- 5 mV, 0.1-150 Hz.• EMG +/- 50 mV, 0.1-500 Hz.

• Range, 50 ft.• Low Noise , < 1 uV RMS• Low weight, 191 grams (0.42 oz.)• Battery options, 14 hrs to days