SCYON 1

SCYON Orthopaedics AG

Wear Reduction in Artificial Joints (THR) by Geometric Means

SCYON Fossa Cup

Page 2: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 2

Kyon - SCYON

Page 3: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 3

Hip Bearing Solutions

•  Metal on Metal (The Evil) •  Ceramic on Ceramic (The Mean) •  Metal on Poly (The Forgiving)

Page 4: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 4

Metal on Metal Biologic reactions

Page 5: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 5

Metal on Metal Pseudotumors

•  “Pseudotumour” – a cystic or solid mass

•  Diagnose by MRI or ultrasound •  1% of MOM pts develop

pseudotumours by 5 years –  17pts, all female, –  Pandit et al (JBJS, 2008)

•  8% at 5yrs FU indicated by preliminary study from same group (AAOS 2009)

–  Prospective of 97 hips –  6 pseudotumours (5 female, 1 male)

Page 6: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 6

Ceramic - Ceramic

Page 7: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 7

Ceramic - Ceramic

•  Most studies report: –  20% noise, 10% squeaking –  1% persistent/ symptomatic squeaking

•  Varies from 0.7% to 40% •  2 studies report greater incidence

(8%-40%) with Ti-Mo-Zi-Fe alloy stem with V-40 taper neck profile

Page 8: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 8

Metal on UHMWPE (poly)

•  Gamma or ETO sterilized (non cross linked) has been associated to high frequency of mid-term radiographic osteolysis and subsequent clinical failure and revision.

•  1st Generation: two approaches (Melting and annealing) are used to reduce the concentration of residual free radicals after irradiation. Higher incidence of fatigue fractures

Page 9: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 9

What is the solution?

•  Metal/metal: Metallic debris and pseudo tumors •  Ceramic/ceramic: Squeaking •  Metal/highly x-linked poly: Fatigue failures

Page 10: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 10

Solution

•  Use material with longest proven clinical track record – standard gamma sterilized GUR 1020,

•  But address wear by geometry

• Bearing Contact

• Remove material where stress is high

Hertzian contact stress

Page 11: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 11

Solution

Standard geometry SCYON fossa

Page 12: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 12

Finite Element model

28 mm diameter CoCr head

UHMWPE layer

Ti shell support

Either standard liner geometry or SCYON fossa geometry

Page 13: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 13

Fossa Liner von Mises stresses reduced from 9.6 to 4.4 MPa

Standard geometry Point contact condition

Fossa geometry ring contact condition

Page 14: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 14

Stress distribution in Standard Geometry

Page 15: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 15

Stress distribution in Fossa Geometry

Page 16: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 16

Friction Factor

28 mm diameter CoCr head

UHMWPE layer

Either standard liner geometry or SCYON fossa geometry

N.E. Bishop, et. Al., Med. Eng. Physics., 30; 1057-1-64, 2008.

Page 17: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 17

Conventional poly liner geometry

Fluid film is quickly lost at the pole

Page 18: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 18

SCYON Fossa poly liner geometry

* N.E. Bishop, et. Al., Med. Eng. Physics., 30; 1057-1-64, 2008.

Sustained fluid flow out of the fossa maintains lubrication

Under pressure Flow restricted by seal

Page 19: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 19

SCYON Fossa Friction Factor

Zimmer * Friction factor 0.15

Conventional Geometry 0.12

SCYON Fossa 0.067

Page 20: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 20

Fossa Liner Wear testing

Endolab, Rosenheim, DE

Hip simulator according to ISO 14241-1

GUR 1020 UHMWPE liners

Vacuum packed in Nitrogen, gamma sterilized 25-37 kGy

28 mm diameter CoCr heads

liner with standard geometry as control

Compliant double shell and stiff shell backing

Page 21: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 21

Wear testing Wear distribution

Conventional Geometry Fossa Geometry

Page 22: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 22

In-vivo wear by RSA (RadioStereographic Analysis)

•  Precise determination of implant/bone migration and implant/implant displacement

•  RSA most precise method for determining in –vivo subsidence and wear rates

Page 23: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 23

RSA (RadioStereographic Analysis)

•  Spherical Tantalum markers are placed in the implants and bone

• Tantalum is biocompatible, well tolerated by the body, and very radio-opaque

• Markers are placed in the implants, the femur and pelvis

Page 24: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 24

Control points

X-ray tubes Focus 1 Focus 2

Film cassettes

Fiducial marks

Page 25: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 25

92

x

71

111 112

41 41

72

92

Page 26: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 26

Measurements of radiographs

112.344 123.334 112.45 132.45

2D coordinates

Page 27: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 27

RSA (RadioStereographic Analysis)

Page 28: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 28

SCYON head penetration in-vivo RSA

mm

months

+/- 2 std. error

Page 29: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 29

0.00

0.05

0.20

ETO !- Poly XLPE 10 SCYON

0.10 “safe level”

0.15

(Nivbrant , Digas)

mm

Annual in-vivo wear rates, RSA

Page 30: SCYON Orthopaedics AG - KYON Veterinary Surgical … Orthopaedics AG Wear Reduction in Artificial Joints (THR) by Geometric Means SCYON Fossa Cup SCYON 2 Kyon - SCYON SCYON 3 …

SCYON 30

Conclusion

•  Fossa Liner reduces von Mises stress by factor 2

•  Fossa liner improves lubrication which reduces friction factor 2 times

•  Fossa Liner reduces poly wear by factor 3 in hip simulator (with CoCr heads and solid shells)

•  Initial 2 year results are very promising with annual wear rate of 25 microns.