s4000 optical chilled mirror hygrometers - · pdf filechilled mirror hygrometer...

4
Chilled Mirror Optical Hygrometers the ultimate standard in moisture measurement technology

Upload: vocong

Post on 06-Feb-2018

237 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: S4000 Optical Chilled Mirror Hygrometers -  · PDF fileCHILLED MIRROR HYGROMETER SPECIFICATIONS Optidew Optidew Optidew Optidew Optisure Optisure Optisure RS S4000TRS Wall

Chilled Mirror Optical Hygrometers

the ultimate standard inmoisture measurement

technology

Page 2: S4000 Optical Chilled Mirror Hygrometers -  · PDF fileCHILLED MIRROR HYGROMETER SPECIFICATIONS Optidew Optidew Optidew Optidew Optisure Optisure Optisure RS S4000TRS Wall

The Ultimate Standard in Moist :

Product ScopeKahn offers two optical (chilled mirror) hygrometer products, Optidew and Series 4000, to meet therequirements of a broad range of dewpoint measurement applications. Each product is available in a varietyof models to suit the user’s specific needs. Kahn’s 20+ years of experience in chilled mirror technology hasproduced extremely sensitive (part per billion), accurate and drift-free instrumentation for measurement of gasdewpoint. All Kahn hygrometers offer measurement traceability to national and international standards.

OPTIDEWCompact, Economical & Rugged

OPTISURE/SERIES 4000 Ultimate in Dewpoint Monitoring

Principle of OperationKahn optical hygrometers use a fundamentaland therefore highly accurate, reliable and drift-free method of continuously measuring thedewpoint of a gas sample. A mirror is chilled bya Peltier thermoelectric heat pump; a light isfocused on the mirror; a photodetector meas-ures the amount of light being reflected or scat-tered off the mirror; a platinum resistance ther-mometer (PRT) measures the temperature ofthe mirror which is dewpoint or frostpoint, bydefinition.

Mirror and Cooling SystemA chemically resistant, polished metal mirror isthermally bonded to a multi-stage Peltier ther-moelectric heat pump. Direct current deliveredto the heat pump allows the mirror to cool untilcondensation begins to form.

Electro-Optical Detection SystemA single or dual optical detection systemconsisting of a light source and photodetec-tor(s) measures reflected and/or scattered lightoff the mirror. Closed-loop temperature controlensures rapid equilibrium between condensa-tion and evaporation at the mirror surface.

Temperature Measurement SystemThe temperature of the mirror surface is meas-ured directly with a highly accurate PRTembedded within the mirror, and is displayed inselectable units on the instrument’s front panel.

Contamination CompensationContamination can compromise the measure-ment accuracy of an optical hygrometer. AllKahn optical hygrometer products feature eitherDynamic Contamination Correction (DCC) orAutomatic Balance Compensation (ABC), whichensures continuous optimal operation of thesensor in contaminated environments. Thecontamination compensation system periodi-cally heats the mirror to vaporize the accumu-lated dew or frost. The instrument then read-justs itself to allow for altered reflectivity due tocontaminants and returns to normal measuring.

NOTE: The information included herein was correct at the time ofpublication and supercedes all previous data. It is our policy to con-tinually improve our products to insure even better performance.Consequently current Kahn products may incorporate modificationsnot shown on these pages.

SampleIn

Light SourceReflected LightDetector

Peltier Heat Pump

SampleOut

Mirrorand PRT

ScatteredLightDetector

ViewingPort

The Optidew Dewpoint and RH Hygrometer is a com-pact, sturdy and economical instrument that providescontinuous dewpoint measurement, display and output.

*The S4000TRS also features our unique “speed pipe” technology that improves the response speed at trace moisture levels. The “speed pipe” concen-trates the formation of ice crystals on the mirror surface and can reduce response time at trace moisture levels by a factor of four times.

ModelsOptidew-Wall

� Wall mount transmitter with integral sensors

Optidew-Remote� Transmitter with remote sensors and 6 foot cables

Optidew-Probe� Transmitter with probe type sensor

Optidew-Bench� Bench mount hygrometer with monitor, carry

handle, remote sensors and 6 foot sensor cables

ModelsOptisure Integrale

� Dewpoints to -60°C with internal sensor� Intuitive touchscreen display

Optisure Remote� Remote sensor with up to 10 meter cable� Dewpoints to +90°C

Optisure RS� Intuitive touchscreen display� Dewpoints to -90°C

S4000TRS*� Temperature-controlled sensor body for dewpoint

measurements to -100°C (13 ppb)� Available with VCR couplings for optimum trace

moisture measurement

The Optisure/Series 4000 Hygrometer family featuresthe most accurate and versatile optical hygrometersavailable in the marketplace today.

Product Highlights� Dewpoint measurements from -100°C to +90°C

� Accuracy of ±0.1°C

� Dewpoint, temperature, %RH measurements

� 2-Stage or 3-stage Peltier cooling

� Automatic contamination compensation

� Current, voltage, USB Modbus and RS232 outputs

� Application software

� Guaranteed frost information below 0°C

� Microscope to monitor condensation on mirror

� Optional pressure compensation

� Built-in data hold function

Optidew� -50 to +130°C dewpoint measurement range� Accuracy of ±0.2°C� 1-Stage or 2-Stage Peltier cooling� Automatic contamination compensation� Built-in data hold function� Analog and digital outputs� Transmitter or display models� High temperature sensor available

Page 3: S4000 Optical Chilled Mirror Hygrometers -  · PDF fileCHILLED MIRROR HYGROMETER SPECIFICATIONS Optidew Optidew Optidew Optidew Optisure Optisure Optisure RS S4000TRS Wall

Some SatisfiedCustomers

Air ProductsCargillDuke EnergyExxonMobilFord General ElectricHarvard UniversityHoneywellIBMIntelLockheed MartinMerckNASANational Weather ServicePratt & Whitney AircraftQualcommTennessee Valley AuthorityTexas InstrumentsTranswestern PipelineU.S. NavyUnited Technologies

The calibrations of Kahn hygrometers are traceable to the NationalInstitute of Standards and Technology. Sensors are calibrated througha master optical hygrometer which has been calibrated at the NISTand is periodically re-calibrated. A certificate of traceability is provid-ed with all of these instruments.

Applications Count onKahn ExperienceKahn, a leader in pneumatic, hydraulic andelectronic technology for over 60 years,provides innovative solutions to practicalmeasurement problems. Since Kahn’s firstmoisture measurement designs were intro-duced 50 years ago, we have manufac-tured high quality, durable hygrometers formany specialized applications, often underdemanding conditions. Our long-standingsuccess in customer satisfaction and ourexpanding product line ensure that Kahncan provide you with hygrometers to suit allyour needs.

Kahn provides technical support and main-tenance for all of its equipment, from theearliest models to the latest innovations.Our hygrometers are also backed by thefinest warranty in the industry: One full yearon calibration and workmanship for boththe instrument and sensor.

Utility Switchgear Performance MonitoringSulfur hexaflouride (SF6) is used by electric utilities as aninsulating gas in high voltage switchgear. Because humiditycompromises the ability of SF6 to insulate, Kahn opticalhygrometers are used to monitor its dewpoint and preservethe integrity of vital utility company equipment.

Motor Vehicle Emissions TestingMeasurement of motor vehicle emissions such as nitrousoxide, carbon monoxide, lead and residual hydrocarbons isaffected by humidity. By precisely measuring dewpoint, Kahnoptical hygrometers help ensure an accurate determinationof emissions concentrations.

Calibration Reference TestingInstrument manufacturers, calibration laboratories and otherend users utilize Kahn optical hygrometers as referencestandards when verifying the calibration or performance ofrelative humidity probes or dewpoint sensors.

Electronic Component ManufacturingMoisture is an unwanted contaminant in the manufacture ofsemi-conductors, integrated circuits and other electroniccomponents. Kahn optical hygrometers are used to monitordewpoint levels in gases used in production processes and tocontrol humidity in fabrication and assembly areas.

Heat TreatingAnnealing, carburizing and other heat treat processes requireprecise monitoring and control of moisture content. Kahnoptical hygrometers allow manufacturers and independentheat treaters to control the quality of their processes, mini-mize scrap and ensure compliance with quality standardsincluding ISO 9000.

Gas Turbine Inlet Air MonitoringThe moisture content of air drawn into industrial gas turbinesmust be monitored to optimize combustion and minimizeharmful emissions. Durability and accuracy of moisturemeasurement instrumentation are essential. These benefitsare provided by Kahn optical hygrometers to turbine manu-facturers and users worldwide.

ture Measurement Technology

Page 4: S4000 Optical Chilled Mirror Hygrometers -  · PDF fileCHILLED MIRROR HYGROMETER SPECIFICATIONS Optidew Optidew Optidew Optidew Optisure Optisure Optisure RS S4000TRS Wall

CHILLED MIRROR HYGROMETER SPECIFICATIONS

Optidew Optidew Optidew Optidew Optisure Optisure Optisure RS S4000TRSWall Remote Probe Bench Integrale Remote

System Accuracy ±0.2°C* ±0.2°C* ±0.2°C* ±0.2°C* ±0.1°C ±0.1°C ±0.1°C ±0.1°*Optional accuracy: ±0.15°C

Cooling Rate 1°C/Sec 1°C/Sec 1°C/Sec 1°C/Sec 2°C/Sec 1°C/Sec 2°C/Sec 0.5°C/Sec

Digital Display Optional Optional Optional Integrated Touchscreen Color LCD Touchscreen Dual MetersIntegrated Integrated Integrated Screen

Units°C, °F, %RH °C, °F, %RH °C, °F, %RH °C, °F, %RH °C, °F, PPM(V), °C, °F, PPM(V), °C, °F, PPM(V), °C, °F, PPM(V), g/m3, g/kg g/m3, g/kg g/m3, g/kg g/m3, g/kg PPM(W) SF6, %RH, PPM(W), %RH, PPM(W), %RH, PPM(W), %RH,

g/m3, g/Kg or kPa g/m3, g/Kg or kPa g/m3, g/Kg or kPa g/m3, g/Kg or kPa

4-20 mA, 4-20 mA, 4-20 mA, 4-20 mA, 4-20mA,0-20mA 4-20 mA, 4-20mA,0-20mA 4-20 mA,0-20 mA, 0-20 mA, 0-20 mA, 0-20 mA, 0-1V, USB, 0-20 mA, 0-1V mV/°C

Outputs RS232 RS232 RS232 RS232 RS232/RS485 0-1V USB, SD card RS232RS485 opt. RS485 opt. RS485 opt. RS485 opt. Ethernet opt. USB, SD card Ethernet opt.Alarm relay Alarm relay Alarm relay Alarm relay Alarm relays Alarm relays Alarm relays

Sensor Location Integral Remote Integral Remote Integral Remote Integral Integral

Configuration Wall Bench Flange Bench Horizontal bench Horizontal bench Horizontal bench Bench oror panel or panel or 19” rack or 19” rack or 19” rack 19” rack

Pressure Compensation No No No No Yes Yes Yes Yes

Power Requirements 90-264 VAC 90-264 VAC 90-264 VAC 90-264 VAC 85-264 VAC 85-264 VAC 85-264 VAC 115 VAC47-440 Hz 47-440 Hz 47-440 Hz 47-440 Hz 47-63 Hz 47-63 Hz 47-63 Hz 50-60 Hz

DimensionsMonitor Dimensions H x W x D 14.5”x9”x4” 9”x9”x4” 18.4”x10”x22.7” 4”x11.5”x10.3” 7.3”x19”x14.5” 7.1”x15.8”x12.6” 7.0”x17.3”x21.7” 34”x22”x24”

Sensor Dimensions L x Dia. 5”x1.8” 5”x1.8” 5”x1.8” 5”x1.8” includes sensor 5”x1.8” includes sensor includes sensor

Weight 7 lbs. 8 lbs. 13.5 lbs. 6.5 lbs. 24.3 lbs. Sensor 0.7 lbs. 37.5 lbs 110 lbs. Monitor 17.4 lbs.

-40°C to +90°C -40°C to +90°C -40°C to+90°C -40°C to +90°C -60°C to +40°C -40°C to +90°C -80°C to +20°C -100°C to +20°C (Single Stage) (Single Stage) (Single Stage) (Single Stage) OR

-50°C to +90°C -50°C to +90°C -50°C to +90°C -50°C to +90°C -90°C to +20°C (Dual Stage) (Dual Stage) (Dual Stage) (Dual Stage)

-40°C to+130°C -40°C to+130°C -40°C to +130°C -40°C to +130°C(High Temperature) (High Temperature) (High Temperature) (High Temperature)

Operating Pressure Vacuum to Vacuum to Vacuum to Vacuum to Vacuum to Vacuum to Vacuum to Vacuum to3600 PSIG 3600 PSIG 3600 PSIG 3600 PSIG 290 PSIG 3600 PSIG 145 PSIG 150 PSIG

-40°C to +90°C* -40°C to +90°C*-40°C to +90°C* -40°C to +90°C* -20°C to +40°C -10°C to +90°C +5°C to +30°C +10°C to +30°C

-20°C to +50°C -20°C to +50°C -20°C to +50°C -20°C to +50°C -20°C to +40°C -20°C to +40°C +5°C to +40°C -20°C to +50°C*High Temperature Sensor: -40°C to +130°C

Sample Flow Rate 0.1 to 0.1 to 0.1 to 0.1 to 0.1 to 0.1 to 0.5 to 0.1 to 2 L/min. 2 L/min. 2 L/min. 2 L/min. 1 L/min. 2 L/min. 1 L/min. 0.7 L/min.

Catalog no. OpticalPrinted in USA 0116\

885 Wells Road, Wethersfield, CT 06109

Phone: (860) 529-8643 Fax: (860) 529-1895 www.kahn.com [email protected] INSTRUMENTS, INC.

DEWPOINT SENSOR

Measurement RangeDewpoint

Operating TemperatureStandard Sensor

Electronics

GENERAL