respiratory monitoring during mechanical ventilation
TRANSCRIPT
SYMPOSIUM: RESPIRATORY MEDICINE
Respiratory monitoringduring mechanicalventilation
Robinder G Khemani
Robert D Bart III
Christopher JL Newth
Abstract
Techniques to monitor the respiratory system during mechanical
ventilation have evolved significantly over the years. When integrated
with the physical examination, these tools aid the management of
respiratory disease, ultimately leading to safer and more effective care
for all mechanically ventilated children. This review will focus on readily
available methods of respiratory monitoring for children undergoing
mechanical ventilation. In particular, there will be a brief discussion on
gas exchange, capnography, respiratory plethysmography and oesopha-
geal manometry, as well as a more substantial discussion on
pressure–volume and flow–volume loops. Finally, we discuss using all
these tools to help determine optimal ventilator support for a variety of
pulmonary disease states.
Keywords capnography; manometry; mechanical; paediatrics;
plethysmography; respiratory-function tests; ventilation
Introduction
Respiratory failure is one of the most common reasons for
admission to paediatric intensive care units (PICUs), particularly
in the first 2 years of life. While patients are recovering from the
aetiology leading to respiratory collapse, intensivists are charged
with supporting the respiratory system with mechanical ventila-
tion. Paramount to ensuring adequate support is close super-
vision of therapeutic interventions – through both invasive and
non-invasive forms of respiratory monitoring. This discussion
concentrates on common means of respiratory monitoring used
Robinder G Khemani MD is at the Children’s Hospital Los Angeles,
Department of Anesthesiology and Critical Care, 4650 Sunset Blvd., Mail
stop 12, Los Angeles, CA 90027, USA.
Robert D Bart III MD is at the Children’s Hospital Los Angeles, Department
of Anesthesiology and Critical Care, 4650 Sunset Blvd., Mail stop 12, Los
Angeles, CA 90027, USA. Assistant Professor of Clinical Pediatrics,
University of Southern California, Keck School of Medicine.
Christopher J L Newth MB FRCPC FRACP is at the Children’s Hospital Los
Angeles, Department of Anesthesiology and Critical Care, 4650 Sunset
Blvd., Mail stop 12, Los Angeles, CA 90027, USA. Professor of Pediatrics,
University of Southern California, Keck School of Medicine.
PAEDIATRICS AND CHILD HEALTH 17:5 193
in intensive care units. In particular, it will focus on capnography,
work of breathing (WOB) measurements, and flow–volume and
pressure–volume loops. Additionally, substantial sections are
devoted to therapeutic interventions based on integrating
monitoring and the underlying disease state, particularly with
regards to optimising ventilator support.
Gas exchange
Regardless of the process leading children to require mechanical
ventilation, the paediatric intensivist must maintain gas exchange
to keep the body in a state of relative homeostasis. Inadequate
gas exchange manifests either as an inability to deliver adequate
oxygen or to eliminate carbon dioxide effectively.
Oxygenation
Manipulations of mechanical ventilator support for hypoxaemic
respiratory failure are targeted at maintaining oxygen delivery.
Fundamentally, two types of hypoxaemic respiratory failure exist.
Type I hypoxaemic respiratory failure results from ventilation–-
perfusion (V/Q) mismatch. In particular, it refers to areas of
intrapulmonary shunt generated from perfused alveoli that are
inadequately ventilated (due to atelectasis, consolidation, oede-
ma, etc.). Characteristic of this type of respiratory failure are large
differences between pAO2 (Alveolar oxygen tension) and pAO2
(arterial oxygen tension). This results in an increase in the
Alveolar–arterial (A–a) oxygen gradient. Under normal circum-
stances, this oxygen difference is less than 10 mm Hg. However,
with severe restrictive pulmonary disease, the A–a oxygen
gradient is often several hundred mm Hg.1 One goal of positive
pressure ventilation is to diminish intrapulmonary shunt by
recruiting atelectatic or consolidated lung units, thereby improv-
ing V/Q matching and reducing the A–a oxygen gradient.
Type II hypoxaemic respiratory failure is less common than
type I failure, and classically results from hypoventilation
without increase of the A–a oxygen gradient. For the most part,
pure Type II respiratory failure presents less of a challenge for
management as modest improvements in ventilation or small
increases in supplemental oxygen can often overcome the
hypoxaemia.
Ventilation
Positive pressure ventilation is commonly initiated for respiratory
failure that manifests by elevated PaCO2 (arterial carbon dioxide
tension), and is often associated with a disease causing an
increased A–a oxygen gradient. In the vast majority of cases this
results from insufficient alveolar minute ventilation, although in
rare cases, PaCO2 may rise from metabolic derangements and
increased production of CO2. The determinants of alveolar
minute ventilation are two-fold. Recall:
Minute ventilation ðVEÞ ¼ ½respiratory rate ðf Þ�½tidal volume ðVTÞ�;
where VT ¼ volume of dead space (VD)+alveolar volume (VA);
combining VE ¼ f(VD+VA), solving alveolar minute ventilation
(MVA) ¼ f[VT–VD].
VD is the volume within the respiratory system that does not
participate in gas exchange. In healthy individuals, this is made
up almost exclusively of anatomic dead space (oropharynx,
trachea, large conducting airways), with little contribution from
lung units. However, as pulmonary disease worsens, and V/Q
r 2007 Elsevier Ltd. All rights reserved.
2 (
mm
Hg
)
III
a’
a
70
60
50
40
30
SYMPOSIUM: RESPIRATORY MEDICINE
mismatch ensues, dead space increases (this includes anatomic
dead space and lung units that are ventilated, but not perfused).
In normal individuals the dead space:tidal volume ratio (VD/VT)
is estimated to be 0.2–0.3. However, patients with significant lung
disease requiring positive pressure ventilation often have VD/VT
ratios exceeding 0.6.2 Therefore, elevated levels of PaCO2 can
come from three major mechanisms: increased CO2 production,
inadequate minute ventilation from low tidal volumes or
respiratory rates, or increases in dead space.
CO
Time (seconds)
I
II
IV20
10
0
0 2 4 6 108
Figure 1 Time-based normal capnograph (solid line) and capnograph of
obstructed airways (dotted line). (I) Contribution from anatomic dead
space- where CO2 content is near zero. (II) Mixture of dead space and
alveolar ventilation-abrupt rise in CO2. (III) CO2 plateau phase-
characterized by pure alveolar ventilation. (IV) Inspiration – a rapid fall
in CO2 (a & a’) End Tidal CO2. Note the upsloping of phase III from airway
disease.
Monitoring
Oxygenation
The gold standard for monitoring oxygenation is the arterial
blood gas (ABG). The ABG measures PaO2 directly from blood,
and uses values of temperature, pH, PaCO2 and PaO2 to calculate
the percent oxyhaemoglobin saturation. Only arterial samples
will suffice for estimations of PaO2. Capillary blood gases do not
reliably predict arterial PaO2.3 However, the advent of pulse
oximetry in the 1980s has provided an alternative for inexpen-
sive, accurate and continuous monitoring of arterial oxygen
saturation. Classic pulse oximeters are very accurate when
oxyhaemoglobin concentrations are greater than 60%. However,
they can underestimate oxygen saturation when perfusion to the
extremity where the probe is located is compromised, as is often
encountered with shock states, the use of vasoactive medica-
tions, peripheral oedema or peripheral vascular disease.4 Pulse
oximeters may also be inaccurate if other forms of haemoglobin
(i.e methaemoglobin or carboxyhaemoglobin) that absorb light at
similar wavelengths as oxyhaemoglobin or deoxyhaemoglobin
are present.5
Ventilation
Similar to PaO2 with oxygenation, the PaCO2, determined from an
ABG, best measures ventilation. Unlike PaO2, however, a free
flowing capillary sample adequately estimates PaCO2. Central
venous samples may accurately reflect pH, but peripheral venous
samples are inaccurate measures of systemic pH and PaCO2.3
However, short of repeated blood sampling, capnography and
transcutaneous CO2 detectors provide non-invasive alternatives
to monitor ventilation.
Capnography: capnography waveforms (either time or volume
based) provide a wealth of information. In addition to detecting
end-tidal CO2 (ETCO2), they provide information about respira-
tory rate and rhythm, dead space calculations, cardiac output,
confirmation of endotracheal tube (ETT) placement, mechanical
failure of the ventilator from kinking, displacement or obstruc-
tion of the ETT, patient–ventilator asynchrony and the presence
of obstructive airway disease.6
ETCO2 detectors are positioned at the adaptor end of the ETT,
and detect exhaled CO2 using an exhalation chamber, an infrared
light source and a detector. The module can be placed directly on
the end of the ETT or on the side via an aspirating catheter. The
absorption of CO2 is unique, and can be illustrated graphically as
a function of time or volume.
Time-based capnographs consist of four phases. Phases I–III
comprise exhalation, and Phase IV represents inhalation
PAEDIATRICS AND CHILD HEALTH 17:5 194
(Figure 1). The level of CO2 detected during the plateau phase
gradually approaches the PaCO2 during normal physiologic
conditions, but never reaches it. The inflection point just prior
to inspiration is where ETCO2 is measured. Under normal
circumstances, without a significant ETT leak and during tidal
respiration, the ETCO2 estimates PaCO2 within 2–5 mmHg.7
An increased gradient between ETCO2 and PaCO2 signals
increased dead space, and may represent worsening underlying
disease, pulmonary vascular abnormalities, worsening cardiac
output or iatrogenic pulmonary overdistension. The capnogram
can be used to estimate the relative dead space:tidal volume ratio
(VD/VT):
VD=VT ¼ ½PaCO2 � ECO2�=PaCO2,
where ECO2 ¼ mean expired CO2, graphically represented as the
area under the curve of the time-based capnogram.8 This number
can be calculated using automated software with a pneumota-
chograph device, or crudely estimated from the difference
between ETCO2 and PaCO2. In reality, VD/VT ratios are more
easily calculated using volumetric capnograms (vide infra).
In addition to the numerical ETCO2 value, the graphic pattern
of a time-based capnogram can provide a clue to important
disease states and abnormalities. Obstructive airway disease
(Figure 1) shows a classic upslope of Phase III as restriction to
expiratory flow delays emptying of alveolar gas from the different
regions of the lung. In fact, the degree of upslope correlates well
with severity of airway obstruction.9 Rising baseline levels of CO2
on the capnogram may represent rebreathing of CO2 from
insufficient mechanical ventilatory support, inadequate exhala-
tion time or dead-space ventilation.10 Acute respiratory distress
syndrome (ARDS) and other diseases affecting the lung par-
enchyma may cause markedly widened A–a carbon dioxide
gradients (20–30 mm Hg) at the peak of the lung disease, and
r 2007 Elsevier Ltd. All rights reserved.
SYMPOSIUM: RESPIRATORY MEDICINE
subsequent narrowing occurs with disease resolution. Impaired
pulmonary blood flow from low cardiac output, pulmonary
embolus or severe elevations in pulmonary vascular resistance
may have low-amplitude capnogram tracings. The extreme
example of this is during cardiac arrest when lack of pulmonary
blood flow results in a very low ETCO2 value and a high PaCO2.
This combination immediately raises concern that the ETT is not
in the airway, but in reality the ETCO2 is low because there is no
pulmonary blood flow.
The volumetric capnogram traces CO2 concentration against
exhaled volume, and is particularly useful for performing dead
space calculations, determining optimal positive end-expiratory
pressure (PEEP), and titrating drug therapy for states of increased
airway resistance. It is composed of three phases (Figure 2a). In
addition, knowing the measured PaCO2 from an ABG allows easy
computation of VD/VT for each breath (Figure 2b). Just as with
the time-based capnogram, the slope of phases III will increase
with increased airway resistance from obstructive disease.11
FeC
O2
FeC
O2
Z
Y
X
Exhaled tidal volume
Art
erial P
CO
2
End-t
idal C
O2
Exhaled tidal volume
Phase IIPhase I Phase III
b
a
Figure 2 Volumetric capnogram (single breath) (a) Phase I represents
anatomic dead space. Phase II transition from early alveolar emptying,
and phase III the alveolar plateau with slowly rising CO2 from different
lung units; (b) explains the calculation of the VD/VT ratio when PaCO2 is
known. Here VD/VT ¼ (Y+Z)/(X+Y+Z), where X is the area of alveolar
ventilation, Z anatomic dead space, and Y imposed dead space from
poor perfusion or worsening lung disease.
PAEDIATRICS AND CHILD HEALTH 17:5 195
While capnography is useful in the PICU, it has limitations.
First, it is designed for tidal breathing at relatively normal
respiratory rates; therefore, it is best when mechanically
ventilating at relatively slow rates. As such, spontaneously
breathing neonates and infants who normally have a rapid
shallow breathing pattern often lack the plateau phase on the
time-based capnogram. However, the numerical ETCO2 value is
accurate and predicts PaCO2. The tracing is also affected by large
air leaks around the ETT. In addition, care must be taken with
smaller diameter tubes not to kink the tube with the weight of the
end-tidal module. Even if there is a system error in the end-tidal
acquisition unit (such as a large leak around the ETT), the ETCO2
will always read lower than the PaCO2. Therefore, an elevation in
ETCO2 invariably means that the PaCO2 is at least as high, and
should be treated very seriously.
Transcutaneous CO2 monitoring: in situations where capnogra-
phy cannot be employed (e.g. high-frequency ventilation, non-
intubated patients), transcutaneous CO2 monitors (TCOM)
provide an alternative. The TCOM heats the skin, vasodilating
the capillary beds and increasing diffusion of CO2 across the skin.
The probe then detects the CO2 electrochemically. TCOMs are
generally accurate within 6–10 mmHg, and are particularly useful
to follow trends, especially in thinner patients and neonates.
Obesity, oedema, and hypoperfusion may make the TCOM less
accurate.12
Work of breathing measurements
In addition to normalising gas exchange, another goal of
mechanical ventilation is to support the patient while he/she is
recovering from the inciting event that led to respiratory failure.
To this end, minimising work of breathing (WOB) during
recovery by optimal selection of ventilator support allows the
patient the highest likelihood of weaning to extubation. This is
particularly useful for patients with obstructive airway disease.
Oesophageal manometry
Oesophageal manometry has long been used in animals and
adults to study the mechanics of breathing. An oesophageal
manometer can be used to subdivide measurements of respira-
tory system compliance into contributions from the chest wall
and those from the lung. Recall, respiratory system work is the
pressure applied to yield a change in volume of the system. In
reality, this work is comprised of elastic, resistive, inspiratory,
expiratory, lung and chest wall components.13 The most
significant and dynamic component of WOB is best estimated
by the change in pleural pressure needed to generate a change in
volume. An oesophageal manometer, placed correctly in the
lower third of the oesophagus, closely approximates pleural
pressure.14
Oesophageal manometry can monitor patient initiated work on
various levels of support. A good example of this is having a
spontaneously breathing patient on continuous positive airway
pressure (CPAP) and pressure support (PS) where optimal
ventilator support is titrated based on the patient’s WOB and
comfort. Modern ventilators increasingly have built-in software
to analyse WOB when paired with an oesophageal catheter.
r 2007 Elsevier Ltd. All rights reserved.
SYMPOSIUM: RESPIRATORY MEDICINE
Respiratory plethysmography
During normal tidal breathing, without significant pathology,
diaphragm contraction leads to inspiration followed quickly by
chest wall muscle activity. However, as respiratory pathology
develops, regardless of etiology, there is often an increasing
amount of thoraco-abdominal asynchrony (TAA), resulting from
worsening respiratory (initially intercostal) muscle fatigue. Such
TAA can be measured using respiratory plethysmography.
Elastic bands are placed around the rib cage (RC) and
abdomen (ABD). The output of these movements is then
downloaded to a computer program that displays characteristic
oscillatory patterns which approximate sine waves. The level of
synchrony between these two bands, the phase angle (Y), is then
calculated:
sin Y ¼ m=s,
where m is the length of the midpoint of the RC excursion and s is
the length depicting the ABD excursion.15
During normal breathing, the phase angle is less than 251
(mean 81), and creates a very tight, counterclockwise loop
(Figure 3).16 However, as TAA worsens, the loop opens and
widens, with larger phase angles. Additional information is
learned from the direction of rotation of the loop. If the RC leads
instead of the ABD, then the loop will take on a clockwise
rotation, commonly seen with bilateral diaphragm paralysis. If
one hemi-diaphragm or hemi-thorax is ineffective, the loop will
take on a ‘figure of 8’ appearance.17 As such, continuous phase
angle monitoring allows clinician’s to see the effect of medical
interventions (e.g. increased CPAP), monitor disease progression
(e.g. acute upper airway obstruction) and help wean patients
from mechanical ventilation.
Rib cage
Abdomen
Rib cage
Abdomen
Phase shift
Rib cage
Abdomen
Figure 3 Respiratory plethysmography with the Respitrace. On the left-hand
right-hand side graphically represents the phase angles and direction of the lo
breathing moves from synchronous to asynchronous, the size of the phase an
with diaphragm paralysis (paradoxical), the RC and ABD are exactly 180 deg
PAEDIATRICS AND CHILD HEALTH 17:5 196
Monitoring respiratory mechanics
Ventilator management of respiratory failure needs to cater to the
underlying disease pathology. Paramount to optimal manage-
ment is not only initially selecting the correct mode and
ventilator settings for the underlying disease, but also monitoring
physiologic changes that occur from the disease state, or in
response to therapeutic interventions.
Flow–volume loops
Flow–volume loops are particularly useful in diagnosing the type
of respiratory disease present (restrictive versus obstructive). In
the case of lower airway obstruction they have a characteristic
shape, which may change in response to bronchodilators.
Moreover, in large airways (i.e. above the carina) they can help
identify the type of obstruction (fixed or variable) if obtained
while spontaneously breathing or if the ETT lies above the level
of obstruction.
For mechanically ventilated patients, loops vary depending on
the mode of ventilation selected. In volume ventilation, tidal
volume delivered is set with a constant inspiratory flow, and peak
and plateau airway pressures vary. In pressure ventilation,
inspiratory pressure delivered is set and volume varies; the flow
pattern is decelerating. In general, this decelerating flow pattern
yields lower peak inspiratory pressures than volume-limited
ventilation (Figure 4a).
It is necessary to understand the scales and axes when
interpreting flow–volume loops on ventilators and those
of usual pulmonary function testing. Classically, flow–volume
loops produced by spontaneously breathing patients are
inspiration negative and exhalation positive, moving
either clockwise or counterclockwise. However, with the display
RC
ABD synchronous
m/s = 0
φ = 0°
RC
ABD asynchronous
m/s = 0.71
φ = 45°m
s
RC
ABD paradoxical
m/s = 0
φ = 180°
side, RC and ABD movements are plotted as a function of time, and the
op. Note in the usual situation where the diaphragm leads inspiration, as
gle loop widens, although preserving counterclockwise rotation. However,
rees out of phase, with clockwise movement.
r 2007 Elsevier Ltd. All rights reserved.
Flo
w (
L/m
in)
Tid
al vo
lum
e (
ml)
Pressure (cm H2O)
Volume control Pressure control
Exhalation
Inspiration
Inspiration
40
30
20
10
0500
500
400
300
200
100
0
–10
–20
–30
–40
400 300 200 100
0 40302010
Tidal volume (ml)
Flow-volume loop of normal lung
Pressure-volume loop of normal lung
Exhalation
b
a
Figure 4 Normal flow–volume (a) and pressure–volume (b) curves during
volume control and pressure control ventilation. (a) Flow–volume. By
convention, inspiration is on the bottom of the y-axis. As inspiration
ensues, (clockwise), volume gradually increases in a relatively constant
(volume control) or decelerating (pressure control) manner until end
inhalation. At inspiratory flow cessation, volume is exhaled passively
with a typical flat or slightly convex shape to the descending portion of
the expiratory flow limb. (b). Pressure-volume. Pressure rises to deliver a
preset volume (volume-limited) or pressure (pressure-limited), repre-
sented with counterclockwise rotation of the curve. Once inspiration is
terminated, pressure gradually returns to baseline as volume is exhaled.
SYMPOSIUM: RESPIRATORY MEDICINE
on most mechanical ventilators the flow pattern is in
a clockwise direction and inspiration is on the positive
aspect of the y-axis, and exhalation on the negative. For the
purposes of this review all flow-volume loops are expressed in
the conventional way, with inspiration negative and exhalation
positive.
PAEDIATRICS AND CHILD HEALTH 17:5 197
Pressure–volume loops
Pressure–volume loops are useful to help to determine optimal
lung recruitment, compliance and overdistension (Figure 4b).
Here, pressure is on the x-axis, and volume on the y-axis, a
reversal of classic nomenclature.
Special considerations
Precision of measurements
Before discussing characteristic patterns of various loops for
different disease states, it is necessary to discuss the accuracy and
reproducibility of these measurements. First, the presence of a
significant leak around the ETT will underestimate returned tidal
volumes and increase calculated resistance and compliance
values.18 This is easily overcome with a cuffed ETT. While there
is some reluctance by some of the critical care and anaesthesia
community to use cuffed ETTs for paediatric patients, good data
show that current low-pressure, high-volume cuffed tubes are no
different from uncuffed tubes in the incidence of acute or long-
term post-extubation complications.19 As with uncuffed ETTs,
cuffed tubes should be placed with an initial leak at less than
25 cm H2O.
Second, in order to obtain accurate flow–volume and
pressure–volume loops, as well as measurements of tidal volume,
it is important for measurements to occur as close to the tip of the
ETT as possible. While modern ventilator technology has
improved significantly, there are often large discrepancies (up
to 200%) between measurements taken at the tip of the ETTwith
a pneumotachograph and those recorded at the ventilator,
regardless of the software package in use. This is particularly
relevant with smaller patients, where the relatively compliant
ventilator tubing dead space volume may be greater than a small
patient’s tidal breath.
Restrictive lung disease
Restrictive lung disease from pneumonia, ARDS, pleural effu-
sions and similar entities constitutes the majority of ICU
admissions for respiratory failure. The hallmarks of restrictive
lung disease include decreased compliance, with end-expiratory
lung volumes that fall below normal functional residual capacity
(FRC). Moreover, alveolar closing pressures lie above end-
expiratory lung volume resulting in alveolar collapse. Therefore,
effective management of restrictive lung disease requires normal-
isation of end-expiratory lung volumes to a level closer to FRC.
The pressure–volume curve helps to this end.
The inspiratory limb of the pressure–volume curve of a patient
with restrictive lung disease (classically ARDS) has three main
segments, separated by upper and lower inflection points
(Figure 5). The goal is to maintain ventilation within the two
inflection points on the pressure–volume curve. This minimises
ventilator-induced injury from alveolar recruitment and de-
recruitment (below the lower inflection point), and overdisten-
sion (above the upper inflection point). End-expiratory lung
volumes are maintained above the lower inflection point with
positive end-expiratory pressure (PEEP). While recommenda-
tions differ, one strategy is to select a PEEP 2 cm H2O above the
lower inflection point, and attempt to stay in the zone of best
compliance by selecting a relatively conservative tidal volume
(6 ml/kg).20 If a low compliance zone is seen again on the
r 2007 Elsevier Ltd. All rights reserved.
Tid
al vo
lum
e (
ml)
Zone of overdistension
Maximal compliance
Zone of under recruitment
Upper reflection point
Lower reflection point
500
400
300
200
100
0
0 10 20 30 60
ΔV
C = ΔV/ΔP
ΔP
5040
Pressure (cm H2O)
Figure 5 Pressure–volume loop in ARDS. Note the two inflection points
on the inspiratory limb of the pressure–volume loop. The first segment
shows relatively low compliance (manifest by the slope of the line on the
pressure–volume loop) with a flat appearance. This represents areas of
under recruitment. As recruitment improves and end-expiratory lung
volumes exceed the critical opening pressure of most alveoli,
compliance improves (lower inflection point). Compliance stays relatively
stable within this segment of ventilation, but may decrease again
towards the upper limit of a pressure control breath (upper inflection
point). This flattening of the curve represents relative overdistension. The
goal for mechanical ventilation is to keep tidal breaths between these
two inflection points, in the zone of maximal compliance.
Tid
al vo
lum
e (
ml)
Pressure (cm H2O)
500
400
300
200
100
0
0 10 20 30 40
Tid
al vo
lum
e (
ml)
500
400
300
200
100
0
0 10 20 30 40
Normal lung Low compliance lung
Pressure (cm H2O)
Normal lung Overdistended lung
b
a
Figure 6 Pressure–volume loops in states of low compliance. (a)
Pressure–volume loop of lungs with poor compliance. Note the lower
slope of the pressure–volume curve, and higher pressure needed to
generate less volume. (b) Pressure–volume curve of overdistended lungs
from excessive PEEP. Note the shift in end-expiratory lung volumes to the
right, and lower slope of the curve.
SYMPOSIUM: RESPIRATORY MEDICINE
pressure–volume curve (beaking), then the pressure or volume
should be decreased to prevent overdistension. Less commonly,
overdistension from excessive PEEP may occur. This will not
manifest with the classic inflection points, but rather simply as a
pressure–volume curve that is shifted to the right with a smaller
slope (Figure 6). Other measures of overdistension derived from
mathematical examination of the pressure–volume curve21,22
have not been validated in the paediatric population.
In contrast to the pressure–volume curve, the flow–volume
loop for a patient with significant restrictive lung disease will
often have the same characteristic shape as a patient without
lung disease, but with smaller amplitude for a given flow.
Subsequent to changes in compliance through optimising PEEP
and improvements in the underlying disease, in concert with
conservative lung protective pressure-limited ventilation, the
amplitude of the flow�volume loops should return towards
normal (Figure 7).
Clinical entities causing obstructed airways
Medium and small airway disease: in contrast to restrictive lung
disease where disease pathology typically affects the alveoli,
obstructive disease affects airways. Significant airway obstruc-
tion affecting medium (asthma) or small (bronchiolitis) airways
allows the development of air trapping. Here end-expiratory lung
PAEDIATRICS AND CHILD HEALTH 17:5 198
volumes reside above functional residual capacity, and airway
resistance is increased. The flow–volume loops are characteristic
of flow limitation on the expiratory limb. The flow–volume loop
therefore demonstrates lower peak expiratory flows, smaller tidal
volumes, as well as a hallmark ‘scooped-out’ or concave app-
earance to the descending limb of expiratory flow. Moreover, if
the obstruction is variable or reversible, the effect of a therapeutic
intervention (i.e. bronchodilator for an asthmatic) can be
followed. Here the flow–volume loop demonstrates improved
r 2007 Elsevier Ltd. All rights reserved.
Flo
w (
L/m
in)
40
30
20
10
0500
–10
–20
–30
–40
400 300 200 100
Tidal volume (ml)
Figure 7 Flow–volume loop for restrictive lung disease. The shape of the
loop for a patient with significant restrictive disease like ARDS (solid line)
looks identical to the normal loop (dotted line), but with smaller volumes
and generally lower flows at each volume. However, there is no limitation
to flow.
Flo
w (
L/m
in)
40
30
50
20
10
0500
–10
–20
–30
–40
400 300 200 100
Flo
w (
L/m
in)
40
30
20
10
0500
–10
–20
–30
–40
400 300 200 100
Flo
w (
L/m
in)
40
30
20
10
0500
–10
400 300 200 100
Tidal volume (ml)
Tidal volume (ml)
Tidal volume (ml)
a
b
c
After response to
bronchodilator Normal
Diseased
SYMPOSIUM: RESPIRATORY MEDICINE
peak expiratory flow and a return of a flat or slightly convex
shape to the decelerating limb of expiratory flow (Figure 8a).
Large airway disease: a spontaneously breathing, non-intubated
patient with extrathoracic obstruction (e.g. croup, laryngomala-
cia, tracheomalacia, vocal cord dysfunction, etc.) classically has
limitation of flow during inspiration. This flow limitation
manifests with a flattening of the inspiratory limb of the
flow–volume loop (Figure 8b). This can usually be overcome
by bypassing the obstruction with an ETT.
If the obstruction is not bypassed by the ETT and is in the
trachea distal to the tube (i.e. intrathoracic), the flow–volume
loop will show either flattening on exhalation (variable lesion
such as tracheomalacia) or flattening on both inspiration and
exhalation (fixed lesion) (Figure 8c). This latter situation is
commonly seen with foreign body aspiration, tracheal stenosis or
accidental kinking of an ETT.
Clinical application in medium and small airway disease:
when mechanically ventilating patients with significant obstruc-
tive airway disease, care must be taken to prevent dynamic
hyperinflation and the creation of auto-PEEP. Whenever possible,
–20
–30
–40
Figure 8 Various levels of obstruction and flow limitation. Normal loops
are dotted, diseased loops solid. (a) Variable intrathoracic obstruction,
from asthma. Note the diminished peak expiratory flow, smaller tidal
volumes, as well as the scooped out appearance of the decelerating limb
of expiratory flow. Note the increase in peak flow, and normalization of
the ‘scooped out’ shape after bronchodilator therapy. (b) Extrathoracic
obstruction, with predominance of flow limitation in inspiration. (c) Fixed
airway obstruction in a large airway, with limitation to flow throughout
inspiration and exhalation.
PAEDIATRICS AND CHILD HEALTH 17:5 199 r 2007 Elsevier Ltd. All rights reserved.
Time
V
Auto-PEEP
olume trapped
InspirationExhalation
Vo
lum
eP
ressu
reF
low
Figure 9 Obstructed medium and small airways with mechanical ventilation and progressive dynamic hyperinflation from insufficient expiratory time.
Note the failure of flow to return to zero (arrow) during exhalation on the flow versus time graph (top). The pressure-time graph (middle) displays
progressive air trapping and generation of auto-PEEP. The volume-time graph (bottom) displays a progressive rise in lung volumes as air is trapped with
each breath.
SYMPOSIUM: RESPIRATORY MEDICINE
spontaneous breathing with CPAP and PS is preferable. This
allows patients the ability to set their own inspiratory:expiratory
ratio and CPAP is carefully applied to match the intrinsic level of
auto-PEEP that the patient has generated.23 This strategy can
minimise WOB, and PEEP can gradually be removed as the
airway obstruction improves.24 However, if such a strategy
cannot be employed and the patient is either paralysed or heavily
sedated, then care must be taken to allow sufficient expiratory
time to prevent further air trapping. This can be monitored by
examining the flow versus time curves, paying particular
attention to expiratory flow returning to zero before the initiation
of a subsequent breath. Evidence of dynamic hyperinflation may
be observed on the pressure or volume versus time curves, where
increased air trapping will manifest as gradual increases in end-
expiratory lung volumes and pressures (Figure 9).25 If total
ventilatory support is employed (no spontaneous breathing),
then extrinsic PEEP will worsen air trapping. However, if the
patient is completely spontaneously breathing, then the
application of extrinsic PEEP to match the patient’s
level of intrinsic or auto-PEEP (some advocate 80% of auto-
PEEP) will help minimise respiratory muscle fatigue. Of course,
the level of support should ultimately be determined by patient
comfort and WOB.
Conclusion
Firm understanding of respiratory monitoring tools is invaluable
for all physicians taking care of critically ill children. It allows for
titrating therapeutic interventions to the patient’s disease state,
and if used correctly can facilitate optimal respiratory support
and aid in eventual weaning to endotracheal extubation. More-
over, with close monitoring, aberrations or changes in physiolo-
PAEDIATRICS AND CHILD HEALTH 17:5 200
gic states can be detected before disease progression, allowing for
early interventions and prevention of worsening disease. ~
REFERENCES1 Dantzker DR, Brook CJ, Dehart P, et al. Ventilation�perfusion
distribution in the adult respiratory distress syndrome. Am Rev Respir
Dis 1979; 120: 1039–52.
2 Hubble CL, Gentile MA, Tripp DS, et al. Deadspace to tidal volume
ratio predicts successful extubation in infants and children. Crit Care
Med 2000; 28: 2034–40.
3 Courtney SE, Weber KR, Breakie LA, et al. Capillary blood gases in the
neonate. A reassessment and review of the literature. Am J Dis Child
1990; 144: 1287–88.
4 Schnapp LM, Cohen NH. Pulse oximetry: uses and abuses. Chest
1990; 98: 1244.
5 Barker SJ, Tremper KK, Hyatt J. Effects of methemoglobinemia on
pulse oximetry and mixed venous oximetry. Anesthesiology 1989;
70: 113.
6 Meliones J, Wilson BG, Cheifetz IM, et al. Respiratory monitoring. In:
Rogers, et al. (eds): Textbook of pediatric intensive care. Baltimore:
Williams and Wilkins, 1996. p. 338.
7 Sivan Y, Eldadah MK, Cheah TE, Newth CJ. Estimation of arterial
carbon dioxide by end-tidal and transcutaneous PCO2 measurements
in ventilated children. Pediatr Pulmonol 1992; 12: 153–7.
8 Martin L. Mechanical ventilation, respiratory monitoring, and the
basics of pulmonary physiology. In: Tobias JD. (eds): Paediatric
critical care: The essentials., Armonk, NY: Fugura Publishing Co, Inc,
1999. p. 57–105.
9 Krauss B, Deykin A, Lam A, et al. Capnogram shape in obstructive
lung disease. Anesth Analg 2005; 100: 884–8.
10 Thompson JE, Jaffe MB. Capnographic waveforms in the mechanically
ventilated patient. Respir Care 2005; 50: 100–9.
r 2007 Elsevier Ltd. All rights reserved.
SYMPOSIUM: RESPIRATORY MEDICINE
11 Yaron M, Padyk P, Hutsinpiller M, Cairns CB. Utility of the expiratory
capnogram in the assessment of bronchospasm. Ann Emerg Med
1996; 28: 403–7.
12 Epstein MF, Cohen AR, Feldman HA, Raemer DB. Estimation of PaCO2
by two non-invasive methods in the critically ill newborn infant.
J Pediatr 1985; 106: 282.
13 Benditt J. Esophageal and gastric pressure measurements. Respir
Care 2005; 50: 68–77.
14 Milic-Emili J. Measurements of pressures in respiratory physiology:
techniques in the life sciences. Shannon: Elsevier Scientific, 1984.
p. 1–22.
15 Agostoni E. Deformation of chest wall during breathing effort. J Appl
Physiol 1966; 21: 1827–32.
16 Hammer J, Deakers TW, Newth CJ. Lissajous figure analysis in infants
with thoracoabdominal asynchrony due to neuromuscular diseases.
Am J Respir Crit Care Med 1994; 149: A36.
17 Willis BC, Graham AS, Wetzel RL, Newth CJ. Respiratory inductance
plethysmography used to diagnose bilateral diaphragmatic paralysis.
A case report. Pediatr Crit Care Med 2004; 5: 399–402.
18 Main E, Castle R, Stocks J, James I, Hatch D. The influence of
endotracheal tube leak on the assessment of respiratory function in
ventilated children. Intensive Care Med 2001; 27: 1788–97.
19 Newth CJ, Rachman B, Patel N, Hammer J. The use of cuffed versus
uncuffed endotracheal tubes in pediatric intensive care J Pediatr
2004; 144: 333–7.
20 ARDS Network. Ventilation with lower tidal volumes as compared
with traditional tidal volumes for acute lung injury and the acute
respiratory distress syndrome. N Engl J Med 2004; 42: 1301–8.
21 Neve V, de la Roque ED, Leclerc F, et al. Ventilator-induced
overdistension in children: dynamic versus low-flow inflation volume-
pressure curves. J Respir Crit Care Med 2000; 162: 139–47.
22 Fisher JB, Mammel MC, Coleman JM, et al. Identifying lung
overdistention during mechanical ventilation by using volume-
pressure loops. Pediatr Pulmonol 1988; 5: 10–4.
PAEDIATRICS AND CHILD HEALTH 17:5 201
23 Smith TC. Impact of PEEP on lung mechanics and work of breathing in
severe airflow obstruction. J Appl Physiol 1988; 65: 1488–99.
24 Graham AS, Chandrashekharaiah G, Citak A, Wetzel RC, Newth CJL.
Positive end-expiratory pressure and pressure support in peripheral
airways obstruction. Intensive Care Med 2007; 33: 120–7.
25 Blanch L, Bernabe F, Lucangelo U. Measurement of air trapping,
intrinsic positive end-expiratory pressure and dynamic hyperinflation
in mechanically ventilated patients. Respir Care 2005; 50: 110–24.
Practice points
� Capnography generally provides accurate estimates of arterial
CO2, in addition to helping detect obstructive airway disease,
the adequacy of pulmonary blood flow and increases in dead
space
� ETCO2 will always be lower than PaCO2, so elevations in ETCO2
should be taken seriously
� Oesophageal manometry and respiratory plethysmography are
useful surrogate objective measures of work of breathing
� Flow–volume loops are particularly useful for distinguishing
obstructive airway disease and restrictive lung disease on
mechanical ventilation
� With restrictive lung disease, pressure–volume loops help the
clinician optimise ventilator support, particularly with regard
to positive end-expiratory pressure, pressure support and
peak inspiratory pressure
� Pulse oximeter numbers lag the clinical situation by 15–20 s.
Values are not accurate unless the pulse rate determined by
the oximeter is the same as that determined from the bedside
monitor ECG recording
� Measurements of pulmonary function should occur as close to
the tip of the ETT as possible
r 2007 Elsevier Ltd. All rights reserved.