recommendations to decrease risk of laser damage to suspension wires

1
Raab - Decrease Laser Risk 1 LIGO-G020301-00-W Recommendations to Decrease Risk of Laser Damage to Suspension Wires Prepare baffles for mounting on sensitive optics during next vent opportunity Incorporate bias slider limiters on IO and RM To be locked in once nominal bias established Use a preset misalignment button if procedural misalignments are required, not a slider Incorporate PSL shutter closure based on shadow-sensor or optical lever triggers Epics not fast enough to stop a direct hit of a 5-10 W beam onto wire, but such an immediate direct hit is unlikely IO steering optics should accommodate a tight threshold

Upload: ahmed-haney

Post on 30-Dec-2015

18 views

Category:

Documents


0 download

DESCRIPTION

Recommendations to Decrease Risk of Laser Damage to Suspension Wires. Prepare baffles for mounting on sensitive optics during next vent opportunity Incorporate bias slider limiters on IO and RM To be locked in once nominal bias established - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Recommendations to  Decrease  Risk of Laser Damage to Suspension Wires

Raab - Decrease Laser Risk 1LIGO-G020301-00-W

Recommendations to Decrease Risk of Laser Damage to Suspension Wires

• Prepare baffles for mounting on sensitive optics during next vent opportunity

• Incorporate bias slider limiters on IO and RM• To be locked in once nominal bias established

• Use a preset misalignment button if procedural misalignments are required, not a slider

• Incorporate PSL shutter closure based on shadow-sensor or optical lever triggers• Epics not fast enough to stop a direct hit of a 5-10 W beam onto wire,

but such an immediate direct hit is unlikely

• IO steering optics should accommodate a tight threshold