evaluation of spectrum measurement devices for operational use

7
Nuclear Instruments and Methods in Physics Research A 476 (2002) 416–422 Evaluation of spectrum measurement devices for operational use Robert T. Devine*, Leonard L. Romero, Devin W. Gray, David T. Seagraves, Richard H. Olsher, Jeff P. Johnson Los Alamos National Laboratory, MS G761 ESH-4, P.O. Box 1663, Los Alamos, NM 87545, USA Abstract Several neutron spectrometers manufactured by Bubble Technology Industries (BTI) were tested and evaluated in a variety of neutron fields. Findings and conclusions are presented for the following BTI instruments: a modification of the Rotational Spectrometer (ROSPEC) that includes a thermal and epithermal capability, the Simple Scintillation Spectrometer that is used in conjunction with the ROSPEC to extend its high-energy range, and the MICROSPEC N-Probe which is capable of providing a crude spectrum over the energy range from thermal to 18 MeV. The main objective of these measurements was to determine the accuracy of both the energy spectrum and dose equivalent information generated by these devices. In addition, the dose response of the Wide-Energy Neutron Detection Instrument (WENDI-II) was measured in all neutron fields relative to a bare 252 Cf calibration. The performance of the WENDI-II rem meter was compared to the dose information generated by the neutron spectrometers. The instruments were irradiated to bare 252 Cf and 241 AmBe sources, and in a series of moderated 252 Cf fields using a standard D 2 O sphere and a set of polyethylene spheres. The measured spectra were benchmarked with a set of detailed Monte Carlo calculations with the same energy bin structure as that of the instruments under test. These calculations allowed an absolute comparison to be made with the measurements on a bin by bin basis. The simulations included the effects of room return and source anisotropy. r 2002 Elsevier Science B.V. All rights reserved. PACS: 87.53.Qc Keywords: Neutron dosimetry measurements 1. Introduction Four devices were investigated: Rotational Spectrometer [1] (ROSPEC). This consists of four spherical gas counters arrayed on a rotating table with a range 50–4.5 MeV. The particular version used is enhanced with two additional detectors to measure the range from thermal to 50 keV (bare 6 LiI scintillator for thermal neutrons, and another 6 LiI scintillator shielded by 10 B for detection of epithermal neutrons). Simple Scintillation Spectrometer [2] (SSS). This is an array of small plastic scintillators coupled to a photomultiplier tube with a range of 4–15 MeV. The ROSPEC and SSS are used in combination. N-Probe [3]. This device uses a boron shielded 6 LiI for the range thermal to 0.8 MeV and a *Corresponding author. Tel.: +1-505-667-4254; fax: +1- 505-665-6071. E-mail address: devine [email protected] (R.T. Devine). 0168-9002/02/$ - see front matter r 2002 Elsevier Science B.V. All rights reserved. PII:S0168-9002(01)01481-4

Upload: robert-t-devine

Post on 02-Jul-2016

214 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Evaluation of spectrum measurement devices for operational use

Nuclear Instruments and Methods in Physics Research A 476 (2002) 416–422

Evaluation of spectrum measurement devices foroperational use

Robert T. Devine*, Leonard L. Romero, Devin W. Gray, David T. Seagraves,Richard H. Olsher, Jeff P. Johnson

Los Alamos National Laboratory, MS G761 ESH-4, P.O. Box 1663, Los Alamos, NM 87545, USA

Abstract

Several neutron spectrometers manufactured by Bubble Technology Industries (BTI) were tested and evaluated in a

variety of neutron fields. Findings and conclusions are presented for the following BTI instruments: a modification ofthe Rotational Spectrometer (ROSPEC) that includes a thermal and epithermal capability, the Simple ScintillationSpectrometer that is used in conjunction with the ROSPEC to extend its high-energy range, and the MICROSPECN-Probe which is capable of providing a crude spectrum over the energy range from thermal to 18MeV. The main

objective of these measurements was to determine the accuracy of both the energy spectrum and dose equivalentinformation generated by these devices. In addition, the dose response of the Wide-Energy Neutron DetectionInstrument (WENDI-II) was measured in all neutron fields relative to a bare 252Cf calibration. The performance of the

WENDI-II rem meter was compared to the dose information generated by the neutron spectrometers. The instrumentswere irradiated to bare 252Cf and 241AmBe sources, and in a series of moderated 252Cf fields using a standard D2Osphere and a set of polyethylene spheres. The measured spectra were benchmarked with a set of detailed Monte Carlo

calculations with the same energy bin structure as that of the instruments under test. These calculations allowed anabsolute comparison to be made with the measurements on a bin by bin basis. The simulations included the effects ofroom return and source anisotropy. r 2002 Elsevier Science B.V. All rights reserved.

PACS: 87.53.Qc

Keywords: Neutron dosimetry measurements

1. Introduction

Four devices were investigated:Rotational Spectrometer [1] (ROSPEC). This

consists of four spherical gas counters arrayed on arotating table with a range 50–4.5MeV. Theparticular version used is enhanced with two

additional detectors to measure the range fromthermal to 50 keV (bare 6LiI scintillator for thermalneutrons, and another 6LiI scintillator shielded by10B for detection of epithermal neutrons).

Simple Scintillation Spectrometer [2] (SSS).This is an array of small plastic scintillatorscoupled to a photomultiplier tube with a rangeof 4–15MeV. The ROSPEC and SSS are used incombination.

N-Probe [3]. This device uses a boron shielded6LiI for the range thermal to 0.8MeV and a

*Corresponding author. Tel.: +1-505-667-4254; fax: +1-

505-665-6071.

E-mail address: devine [email protected] (R.T. Devine).

0168-9002/02/$ - see front matter r 2002 Elsevier Science B.V. All rights reserved.

PII: S 0 1 6 8 - 9 0 0 2 ( 0 1 ) 0 1 4 8 1 - 4

Page 2: Evaluation of spectrum measurement devices for operational use

NE213 scintillator for the range 0.8MeV to about18MeV.

WENDI II [4]. This a neutron rem meterincorporating a shell of tungsten powder insidea polyethylene moderator, surrounding a 3Hedetector.

2. Experimental and analytical methods

LANL Low-Scatter Neutron Facility was usedfor the exposures. This facility consists of a sheetmetal structure (8m� 5.3m� 6m) housing a plat-form made of an aluminum lattice. The primaryscattering boundary is located 3m below thealuminum platform. Detectors were positioned ata height of 60 cm above the platform. Monte Carlosimulations were done with MCNP [5]. The fluencewas calculated using the same energy bin structureas that of the spectrometers under test for all of thereference fields. The appearance of the ROSPEC/

SSS and the N-Probe spectra will differ since thebins used in the calculation correspond to theoutput bins of the instruments which are not thesame. The calculations were based on the actualexposure conditions in that the total flux for eachindividual exposure was input to the MCNPcalculation. The effects of room return and sourceanisotropy were included in the simulations. Allcalculations were carried out to 5% uncertainty.

3. Results and discussion

3.1. ROSPEC/SSS

For bare 252Cf, D2O moderated 252Cf and241AmBe the spectra shown in Fig. 1 with goodagreement with calculation in the 50 keV–13.8MeV region. They are referred to in Figs. 1and 3 and the Tables as BCF, MCF, and AmBe,respectively. The fluence results in three energy

Fig. 1. ROSPEC/SSS measured and calculated spectra: BCF, MCF and AmBe (ROSPEC/SSS, solid line, MCNP dotted line, see text

for abbreviations).

R.T. Devine et al. / Nuclear Instruments and Methods in Physics Research A 476 (2002) 416–422 417

Page 3: Evaluation of spectrum measurement devices for operational use

bins (Slow (ThermalF53.5 keV, thermal andepithermal 6LiIdetectors), Intermediate(53.5 keV–4.53MeV, ROSPEC) and Fast (4.53–13,2MeV, SSS), is given in Table 1. For bare252Cf, the total fluence is over reported by 25%,due to the over response in the Slow bin. However,

the dose equivalent contribution from the Slow binis 1% and H � ð10Þ agrees with the calculated datawithin 6%. There is a factor of two overestimate ofthe total fluence for the D20 moderated spectrumagain mainly due to over response in the Slow binwhich in this case contributes 17% of the dose.

Table 1

ROSPEC/SSS fluence results (see text for definitions and abbreviations)

Reference field MCNP calculated fluence (n/cm2) ROSPEC/SSS measured fluence (n/cm2)

Slow Intermediate Fast Total Slow Intermediate Fast Total

BCF 8.21e5 5.73e7 5.41e6 6.35e7 1.93e7 5.40e7 5.75e6 7.91e7

MCF 3.62e7 1.74e7 1.50e6 5.51e7 8.82e7 1.74e7 1.68e6 1.07e8

Ambe 2.11e5 6.59e6 3.16e6 9.96e6 3.28e6 6.03e6 3.94e6 1.33e7

p2 4.18e6 4.97e7 4.65e6 5.86e7 2.71e7 4.74e7 5.03e6 7.95e7

p3 1.04e7 4.39e7 4.01e6 5.83e7 4.04e7 3.94e7 4.37e6 8.41e7

p5 2.00e7 3.09e7 2.99e6 5.39e7 5.58e7 2.87e7 3.06e6 8.75e7

p8 1.86e7 1.65e7 1.91e6 3.70e7 3.07e7 1.43e7 2.34e6 4.74e7

p10 1.35e7 1.05e7 1.42e6 2.55e7 2.74e7 9.08e6 1.33e6 3.78e7

p12 8.96e6 6.73e6 1.05e6 1.67e7 1.33e7 6.20e6 1.72e6 2.12e7

Fig. 2. ROSPEC/SSS measured and calculated spectra for polyethyene spheres (ROSPEC/SSS, solid line, MCNP dotted line, see text

for abbreviations).

R.T. Devine et al. / Nuclear Instruments and Methods in Physics Research A 476 (2002) 416–422418

Page 4: Evaluation of spectrum measurement devices for operational use

The AmBe total fluence is overestimated by 34%by a large Slow bin contribution but in this casethe Slow bin contribution to the dose is again 1%.If the ROSPEC alone is used for detection ofAmBe the dose is under reported by 27% whichwas to be expected.

Spectra of 252Cf moderated by polyethylenespheres of 2, 3, 5, 8, 10, and 12 in. diameter areshown in Fig. 2. These are referred to in Figs. 2and 4 and the Tables as p2; p3; p5; p8; p10; andp12; respectively. The successively larger modera-tion yields spectra that differ from bare califor-nium by larger contributions to the thermal andepithermal regions. There is good agreement withthe calculated spectrum over the energy rangefrom 0.1 to 13MeV with unfolding artifactsbetween 50 and 100 keV and at 4.5MeV causedby the joining of the spectra calculated from thevarious components of the ROSPEC/SSS combi-nation. An anomaly in the spectra for the 10- and12-in. polyethylene is zero fluence 1 eV to 10 keV.The manufacturer attributes this to an algorithmthat reports zero when the fluence falls below agiven value. Estimated dose equivalents are givenin Table 2. The results for these fields are within8% [H � ð10Þ] of the calculated dose equivalentand the contribution from the Slow bin is less than6%.

Several measurements (with identical counttimes) of a D2O moderated 252Cf spectrum wereperformed over a period of two months. Thesemeasurements indicate minimal drift and repro-ducible performance over all energy bins. The SP6

detector covers the upper energy range of thestandard ROSPEC. Over the past two years, thisdetector has had a history of becoming noisy forno apparent reason. The problem is intermittentand may clear up several days after the system isturned off. The problem has not been resolvedeven after a trip to the factory for repair. Artifactscreated by the algorithm in the juncture of thelower end of the spectra obtained from the SSSand the high end of ROSPEC are more pro-nounced and the total reported fluence is reducedsignificantly. It is clearly essential to confirm thatall of the detector channels are noise free prior toand immediately after each data collection.

3.2. MICROSPEC/N-Probe

Bare 252Cf, D2O moderated 252Cf, and 241AmBespectra are shown in Fig. 3 and spectra forpolyethylene spheres shown in Fig. 4. Fluenceresults in three energy bins (Slow (Ther-malF50 keV), Intermediate (50keV–5MeV andFast (5MeV–16MeV), are given in Table 3. Forbare 252Cf, the total fluence is over reported by 6%,due to the over response in the Slow bin. However,the dose equivalent contribution from the Slow binis 1% and H � ð10Þ agrees with the calculated datawithin 6%. There is factor of two overestimate ofthe total fluence for the D20 moderated spectrumagain mainly due to over response in the Slow binwhich in this case contributes 18% of the dose. TheAmBe total fluence is overestimated by 11% alsoby overestimate in the Slow bin but in this case the

Table 2

Results. Normalized to the MCNP calculation for each spectrum or bare Cf (see text for definitions and abbreviations)

Reference field ROSPEC H � ð10Þ: ROSPEC/SSS H � ð10Þ N-probe H � ð10Þ WENDI H � ð10ÞMCNP=1 MCNP=1 MCNP=1 Cf=1.00

BCF 0.96 1.06 1.06 FMCF 1.18 1.28 1.34 1.29

Ambe 0.73 1.24 1.15 0.98

p2 0.98 1.08 1.05 1.00

p3 0.94 1.04 1.01 1.01

p5 0.98 1.08 0.99 1.05

p8 0.91 1.05 0.95 1.01

p10 0.84 1.02 1.07 1.00

p12 0.98 1.02 1.01 0.95

R.T. Devine et al. / Nuclear Instruments and Methods in Physics Research A 476 (2002) 416–422 419

Page 5: Evaluation of spectrum measurement devices for operational use

Slow bin contribution to the dose is 2%. With theexception of D2O moderated californium the overresponse in the Slow bin results in an overestimateof the total fluence that ranges from �26% to+13%. Dose reported is shown in Table 2 with avariation in H � ð10Þ of �5% to +34%

3.3. WENDI-II neutron rem meter

Normalized dose response summarized in Table2 with H � ð10Þ variation of –5% to +29%. Thetraditional energy response (response per unit doseequivalent) of all moderated rem meters is far fromideal in the energy range from 10 eV to 50 keV,with an over response as large as a factor of nine atan energy of 5 keV. This means that in unusualoperational cases where a significant fraction ofthe dose is contributed by sub-50 keV neutrons,rem meters may over respond by a factor of two ormore.

4. Conclusions

ROSPEC (without the SSS) accurately measuresneutron spectra from 0.1 to 4.5MeV. Its resolutionis exceptional and exceeds that of the traditionalBonner sphere spectrometer. Thermal and epither-mal energy bins are significantly over estimated.Dose contribution in this region for most work-place spectra [6] is no more than 10%; the practicaleffect of this over response is minor in terms of thereported dose.

SSS provides good resolution and accuracy overits range of operation (4–15MeV). SSS greatlyimproves the dosimetric accuracy of the standardROSPEC for measurements in hard spectra.

N-Probe gives a much coarser spectrum. Thetransition region between the fast and thermalscintillators is in many cases inaccurate with asignificant bias in the 10 keV to 0.8MeV energyrange. However, its portability and ease of usemake it attractive as a ‘‘smart’’ neutron rem meter

Fig. 3. N-Probe measured and calculated spectra: BCF, MCF and AmBe (N-probe, solid line, MCNP dotted line, see text for

abbreviations).

R.T. Devine et al. / Nuclear Instruments and Methods in Physics Research A 476 (2002) 416–422420

Page 6: Evaluation of spectrum measurement devices for operational use

for field dose determinations. Its dosimetricperformance was similar to that of the LANL/ROSPEC and the WENDI-II rem meter.

In many workplace environments, where thedose is dominated by the spectral componentsabove 0.1MeV, the WENDI-II Rem Meter canprovide accurate dosimetric information, compar-

able to that available from much more expensivespectrometers.

The results of these evaluations should allow aninformed choice to be made of which instrumentsto use in particular situations based on the spectraldetail required by the operational conditions andthe difficulties of handling the equipment.

Fig. 4. N-Probe measured and calcuated spectra for polyethyene spheres (N-Probe, solid line, MCNP dotted line, see text for

abbreviations).

Table 3

N-probe fluence results (see text for definitions and abbreviations)

Reference field MCNP Calculated fluence (n/cm2) N-Probe measured fluence (n/cm2)

Slow Intermediate Fast Total Slow Intermediate Fast Total

BCF 2.11e6 8.58e6 6.33e5 1.13e7 3.04e6 8.50e6 5.13e5 1.20e7

MCF 2.01e7 4.32e6 4.85e5 2.49e7 4.47e7 4.52e6 4.40e5 4.96e7

AmBe 1.55e6 5.78e6 2.63e6 9.96e6 2.45e6 5.84e6 2.77e6 1.11e7

p2 2.96e6 7.14e6 5.47e6 1.06e7 0.00e0 7.40e6 4.46e5 7.84e6

p3 4.20e6 6.03e6 4.75e5 1.07e7 5.40e6 5.97e6 3.80e5 1.18e7

p5 5.50e6 4.11e6 3.55e5 9.97e6 6.92e6 4.10e6 2.81e5 1.13e7

p8 4.45e6 2.22e6 2.31e5 6.90e6 4.04e6 2.13e6 1.74e5 6.34e6

p10 1.58e7 7.98e6 1.06e6 2.48e7 1.02e7 7.51e6 1.04e6 1.88e7

p12 1.04e7 5.19e6 7.95e5 1.64e7 6.85e6 4.77e6 8.10e7 1.24e7

R.T. Devine et al. / Nuclear Instruments and Methods in Physics Research A 476 (2002) 416–422 421

Page 7: Evaluation of spectrum measurement devices for operational use

References

[1] H. Ing, T. Clifford, T. McLean, W. Webb, T. Cousins,

J. Dhermain, Radiat. Prot. Dosim. 70 (1997) 273.

[2] Bubble Technology Industries, Inc, Simple Scintillation

Spectrometer BTI-SSSt for Neutron Spectroscopy to

15MeV, Revision 4 (July 13, 1999).

[3] H. Ing, Personal Communication from Harry Ing to Robert

Devine (October 18, 1999).

[4] R.H. Olsher, H.H. Hsu, A. Beverding, J.H. Kleck, W.H.

Casson, D.G. Vasilik, R.T. Devine, Health Physics 79

(2000) 170.

[5] Transport Methods Group, Los Alamos National Labora-

tory, MCNPFA General Monte Carlo N-Particle

Transport Code Version 4B (March 20, 1997).

[6] W.H. Harvey, F. Hajnal, Radiat. Prot. Dosim. 50 (1993) 13.

R.T. Devine et al. / Nuclear Instruments and Methods in Physics Research A 476 (2002) 416–422422