radiation dosimetry in digital breast tomosynthesis/media/downloads/uk/services/dose...

55
Imagination at work. Imagination at work. Radiation Dosimetry in Digital Breast Tomosynthesis March, 2015 William J. O’Connel, Dr. Ph, Senior Medical Physicist

Upload: others

Post on 22-Mar-2020

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Imagination at work. Imagination at work.

Radiation Dosimetry in Digital Breast Tomosynthesis

March, 2015 William J. O’Connel, Dr. Ph, Senior Medical Physicist

Page 2: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Syllabus

2

1. Introduction

2. Dosimetry in Mammography

3. Dosimetry in Digital Breast Tomosynthesis

4. Methodology

5. Conclusion

Page 3: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Introduction

3

Page 4: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Introduction

4

Radiation dosimetry in digital breast tomosynthesis:

Report of AAPM Tomosynthesis Subcommittee Task Group 223

(2014)

• Ioannis Sechopoulos • John M. Sabol

• Johan Berglund • Wesley E. Bloch • Libby Brateman • Emmanuel Christodoulou • Michael Flynn • William Geiser • Michael Goodsitt

• A. Kyle Jones • Joseph Y. Lo

• Andrew D.A. Maidment • Kazuyoshi Nishino • Anita Nosratieh • Baorui Ren • W. Paul Segars • Miriam Von Tiedemann

Page 5: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Introduction

5

• Over a lifetime, 1 in 8 women will develop breast cancer

• Annual mammography for women age 40 and older*

• Complimentary imaging procedures (MRI, US, NUC) used at increasing frequency

Page 6: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Introduction

6

Page 7: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Introduction

7

History of X-Ray Breast Imaging

1950’s

no-screen direct exposure film

high dose clinical benefits?

1960’s

Xeroradiography

better resolution contrast/dose WIP

1985

NEXT Survey poor IQ in

36% of mammograms

1987

ACR Mammography Guidelines

1992

MQSA

2000

FFDM

DBT

Page 8: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Introduction

8

• Digital Breast Tomosynthesis (DBT) is a new screening tool

• Need to understand, estimate and communicate patient dose

• Radiation dose to glandular tissue is primary consideration

in x-ray based breast imaging

• Cancer risk to adipose tissue is considered minimal

Page 9: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Introduction

9

Image Quality Requirements:

• Masses with irregular or spiculated borders

• Clusters of micro-calcifications

• Architectural distortions of breast structures

Page 10: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

10

relative exposure

Op

tic

al d

en

sity

air

skin margin

fatty tissue

dense tissue

Introduction

Page 11: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Introduction

11

high % adipose tissue high % glandular tissue

Page 12: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Introduction

12

• Small attenuation differences between normal and cancerous tissue

• Low energy x-rays can help to highlight the difference

Page 13: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Introduction

13

μ

kV

fat

glandular

tumor

Page 14: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Introduction

14

kV

glandular

Ca5(PO4)3 OH (0.1 mm)

co

ntr

ast

Page 15: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Dosimetry in Mammography

15

Page 16: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

16

focal spot

diaphragm

compression paddle

grid

screen - film

Dose in Mammography

supporting table

Page 17: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Entrance Skin Exposure

17

Entrance Skin Exposure (XESE )

free-in-air ionization chamber measurement at the point where the x-ray beam first interacts with the skin

Typical XESE values for a 4.5 cm breast are in the range of 500-1,000 mR

Page 18: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Dose in Mammography

18

Entrance dose

1 cm ≈ 50%

2 cm ≈ 25%

3 cm ≈ 10%

X-Ray

Page 19: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Mean Glandular Dose

19

Mean Glandular Dose (Dg ) preferred dose index

glandular tissue is the most probable site of carcinogenesis

Difficult to calculate Dg since doses vary by depth below skin

Dg does not represent individual patient dose

Page 20: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Mean Glandular Dose

20

Dg is dependent upon: • composition of the breast • breast thickness • Half value layer • kVp

Page 21: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Normalized Mean Glandular Dose

21

Normalized Mean Glandular Dose ( DgN ) relates XESE to Dg

DgN provides a conversion factor in units of mGy/R

DgN depends on kVp, HVL, target material, filter material, breast thickness, and tissue composition

Dg = DgN * XESE

Page 22: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Mean Glandular Dose

22

kVp

HVL (mm) 25 26 27 28 29 30 31 32

0.25 1.22

0.26 1.26 1.28

0.27 1.30 1.32 1.34

0.28 1.34 1.36 1.38 1.39

0.29 1.39 1.41 1.42 1.43 1.44

0.30 1.43 1.45 1.46 1.47 1.48 1.49

0.31 1.47 1.49 1.50 1.51 1.52 1.53 1.54

0.32 1.51 1.53 1.54 1.55 1.56 1.58 1.59 1.60

0.33 1.55 1.57 1.58 1.59 1.60 1.62 1.63 1.64

0.34 1.60 1.61 1.62 1.63 1.64 1.66 1.67 1.68

0.35 1.64 1.66 1.67 1.68 1.69 1.70 1.71 1.72

0.36 1.68 1.70 1.71 1.72 1.73 1.74 1.75 1.76

0.37 1.74 1.75 1.76 1.77 1.78 1.78 1.79

0.38 1.79 1.80 1.81 1.82 1.82 1.83

0.39 1.84 1.85 1.86 1.86 1.87

0.40 1.89 1.90 1.91 1.92

mGy/R (for Mo/Mo and 4 cm compressed 50/50 breast)

Page 23: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

kVp

HVL (mm) 25 26 27 28 29 30 31 32

0.25 1.22

0.26 1.26 1.28

0.27 1.30 1.32 1.34

0.28 1.34 1.36 1.38 1.39

0.29 1.39 1.41 1.42 1.43 1.44

0.30 1.43 1.45 1.46 1.47 1.48 1.49

0.31 1.47 1.49 1.50 1.51 1.52 1.53 1.54

0.32 1.51 1.53 1.54 1.55 1.56 1.58 1.59 1.60

0.33 1.55 1.57 1.58 1.59 1.60 1.62 1.63 1.64

0.34 1.60 1.61 1.62 1.63 1.64 1.66 1.67 1.68

0.35 1.64 1.66 1.67 1.68 1.69 1.70 1.71 1.72

0.36 1.68 1.70 1.71 1.72 1.73 1.74 1.75 1.76

0.37 1.74 1.75 1.76 1.77 1.78 1.78 1.79

0.38 1.79 1.80 1.81 1.82 1.82 1.83

0.39 1.84 1.85 1.86 1.86 1.87

0.40 1.89 1.90 1.91 1.92

Mean Glandular Dose

23

mGy/R (for Mo/Mo and 4 cm compressed 50/50 breast)

For 28 kVp, 0.35 mm HVL XESE = 1.231 R Dg = 1.68 mGy/R * 1.231 R

Dg = 2.07 mGy

Page 24: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Factors Affecting Breast Dose

24

Comp Breast ESE (mR) MGD (mGy)

2 cm 250 0.7

4 cm 1,000 1.8

6 cm 1,500 2.4

• Increased breast thickness requires increased dose

• Compression lowers breast dose by reducing breast thickness

• Higher kVp lowers XESE and

Dg (decreases contrast)

breast thickness

MG

D

25 kV

30 kV

35 kV

Page 25: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Factors Affecting Breast Dose

25

Increased breast glandularity requires

increased mAs

Page 26: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Conclusion

26

8 cm

2 cm

keV

Dg

• Rapid decrease in dose with increase in beam energy

• Increasing kV from 17.5 to 30 kV lowers Dg by a factor of 30 for 8 cm breast

• Dg increases by a factor

of 17 from 2 cm breast to 8 cm breast 10 20 30 40

Page 27: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

MQSA

27

• MQSA limits the average glandular dose to 3 mGy for a compressed breast thickness of 4.2 cm and a breast

composition of 50% adipose tissue and 50% glandular tissue (using the MQSA phantom)

• If the average glandular dose for the MQSA phantom exceeds 3 mGy, mammography can not be performed

• The average glandular dose for the MQSA phantom is typically 1.5 to 2.2 mGy per view

Page 28: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

28

Full Field Digital Mammography (FFDM)

The shift from screen film mammography to digital mammography MQSA resulted in improved image quality while providing opportunities for dose reduction Acquisition and display can be optimized independently

Page 29: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

29

Full Field Digital Mammography (FFDM)

Page 30: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

30

Full Field Digital Mammography (FFDM)

Page 31: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Full Field Digital Mammography (FFDM)

31

film∙screen

relative exposure

op

tic

al d

en

sity

Wide dynamic range (1000:1) screen film mammography (40:1)

dynamic image manipulation post processing

CAD 3D imaging

Page 32: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Full Field Digital Mammography (FFDM)

32

Page 33: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Full Field Digital Mammography (FFDM)

33

Page 34: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Full Field Digital Mammography (FFDM)

34

Page 35: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Dosimetry in Digital Breast Tomosynthesis

35

Page 36: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

36

Dosimetry in Digital Breast Tomosynthesis

Page 37: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

37

Dosimetry in Digital Breast Tomosynthesis

employ multiple lines of response acquired over various angles

similar acquisition and processing paradigms employed in computed tomography – including iterative reconstruction dose indices may be different from mammography

Page 38: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

38

Dosimetry in Digital Breast Tomosynthesis

Page 39: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

39

Dosimetry in Digital Breast Tomosynthesis

Page 40: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

40

Dosimetry in Digital Breast Tomosynthesis

Page 41: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

41

Dosimetry in Digital Breast Tomosynthesis

Page 42: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

42

Dosimetry in Digital Breast Tomosynthesis

Page 43: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

43

Dosimetry in Digital Breast Tomosynthesis

A B

C D

7

7

6 8 5 5

2 5

4 3

A + B = 7 A + C = 6 A + D = 5 B + C = 9 B + D = 8 C + D = 7

Iterative

reconstruction

Page 44: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Dosimetry in Digital Breast Tomosynthesis

44

• The changing geometry in tomosynthesis may result in variations in dose

• Source to skin distance and XESE may change as the tube

(and possibly the detector) rotate

• Relative Glandular Dose (RGD) describes the change in mean glandular dose between the 0° projection (mammography) and a projection acquisition for a non-zero angle

Page 45: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Dosimetry in Digital Breast Tomosynthesis

45

Relative Glandular Dose is a function of: • Projection angle

• Breast size

• Breast thickness

Page 46: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

46

Dosimetry in Digital Breast Tomosynthesis

DgN (0°) is the same as standard mammography.

DgN (α) for different angulations is a fraction of DgN (0°) Fraction is described by RGD

Different RGD for each angulation

0° α

Page 47: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

47

Dosimetry in Digital Breast Tomosynthesis

By convention, MGD (0°) has a value of 1.000. Assume DgN (0°) = 1.68 Suppose α = 12.5° and DgN for that angulation is only 1.61

RGD (12.5°) = 1.61/1.68 RGD (12.5°) = 0.96

0° α

Page 48: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

48

Dosimetry in Digital Breast Tomosynthesis

Projection angle

0 5 10 15 20 25 30

RG

D (α

)

1.00

0.98

0.96

0.94

0.92

0.90

0.88

0.86

0.84

0.82

Page 49: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

49

Dosimetry in Digital Breast Tomosynthesis

• Similar to the example in mammography (DgN), the goal is to develop a Roentgen-to-mGy conversion factor for digital breast tomosynthesis

• DgNTOMO is the Normalized Glandular Dose for a complete tomosynthesis acquisition

• Estimation of the Normalized Glandular Dose requires an RGD correction for every angulation in the tomosynthesis acquisition

• The use of an average RGD value is possible in certain cases.

Page 50: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Methodology

50

Page 51: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

51

thickness (cm)

Angle (degree) RGD

0 3.125 6.25 9.375 12.5

2 1.000 0.999 0.996 0.990 0.982 0.994

3 1.000 1.000 0.994 0.988 0.979 0.992

4 1.000 0.999 0.993 0.986 0.974 0.990

5 1.000 0.999 0.992 0.983 0.970 0.989

6 1.000 0.999 0.993 0.983 0.969 0.989

7 1.000 0.997 0.991 0.980 0.966 0.987

8 1.000 0.998 0.991 0.979 0.963 0.986

9 1.000 0.998 0.991 0.978 0.961 0.986

Methodology

RGD and RGD values for GE Healthcare SenoClaire DBT System

Page 52: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

Conclusion

52

Page 53: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

53

• improved low contrast visibility

• potential for lower dose

• reduced interference from overlapping tissue

• re-evaluate compression requirements • fewer patient recalls

Conclusion

Page 54: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

54

• Report provides methodology to estimate breast glandular dose in breast tomosynthesis in a manner consistent with established mammography methodology

• For standard breast tomosynthesis acquisition, the normalized glandular dose appears to be in line with values from mammography

• Acquisition protocols with variance in tube current exposure time or with asymetric tube angulations may not fit the generic model

Conclusion

Page 55: Radiation Dosimetry in Digital Breast Tomosynthesis/media/downloads/uk/services/dose management... · Dosimetry in Digital Breast Tomosynthesis • Similar to the example in mammography

55

• Estimated doses are not patient specific

• Patient specific models not possible without knowledge of true distribution of glandular tissue throughout the breast

• Need better accounting for heel effect

Conclusion