  • CALIBRATION PROCEDURE

    PXIe-5632This document contains the verification and adjustment procedures for the PXIe-5632 VectorNetwork Analyzer.

    Refer to ni.com/calibration for more information about calibration solutions.

    ContentsRequired Software.....................................................................................................................2Related Documentation.............................................................................................................2Test Equipment..........................................................................................................................2Test Conditions..........................................................................................................................6Initial Setup............................................................................................................................... 7Test System Characterization....................................................................................................7

    Zeroing and Calibrating the Power Sensor....................................................................... 7Performing User Calibration on the PXIe-5632....................................................................... 7

    Automatically Calibrating Port One Reflection................................................................8Manually Calibrating Port One Reflection....................................................................... 9Manually Calibrating Transmission Frequency Response..............................................10

    As-Found and As-Left Limits................................................................................................. 11Verification.............................................................................................................................. 11

    Verifying Frequency Accuracy........................................................................................11Verifying Output Power Accuracy (+0 dBm Set Level Check)......................................13Verifying Source Maximum Leveled Power...................................................................15Verifying Source Harmonics...........................................................................................17Verifying Source Nonharmonic Spurious....................................................................... 20Verifying Trace Noise..................................................................................................... 22Verifying System Dynamic Range..................................................................................23Verifying Corrected Directivity (300 kHz to

  • Appendix C: Third Harmonic Test Frequency List.................................................................41Appendix D: RF Measurement Chart..................................................................................... 41Worldwide Support and Services............................................................................................ 44

    Required SoftwareCalibrating the PXIe-5632 requires you to install the following software on the calibrationsystem: NI-VNA 2.0.1

    You can download all required software from ni.com/downloads.

    Related DocumentationFor additional information, refer to the following documents as you perform the calibrationprocedure: NI PXIe-5632 Getting Started Guide PXIe-5632 Specifications NI Vector Network Analyzers Help

    Visit ni.com/manuals for the latest versions of these documents.

    Test EquipmentThe following table lists the equipment NI recommends for the performance verification andadjustment procedures. If the recommended equipment is not available, select a substituteusing the minimum requirements listed in the table.

    2 | ni.com | PXIe-5632 Calibration Procedure

    http://www.ni.com/downloadshttp://www.ni.com/manuals

  • Table 1. Recommended Equipment for PXIe-5632 Calibration

    Equipment RecommendedModel

    Where Used Minimum Requirements

    PXI Expresschassis

    PXIe-1073 withintegrated MXI-Expresscontroller

    All procedures Three adjacent, empty PXIExpress slots

    Power meter AnritsuML2438A

    Verifications: Output power

    accuracy

    Adjustments: Port 1 ALC Port 2 ALC LO level

    Display resolution: 0.01 dB

    Settling: 0.1%

    Instrumentation accuracy:

  • Table 1. Recommended Equipment for PXIe-5632 Calibration (Continued)

    Equipment RecommendedModel

    Where Used Minimum Requirements

    Spectrumanalyzer

    AnritsuMS2720T-0720

    Verifications: Frequency accuracy Source harmonics Source nonharmonic

    spurious

    Adjustments: Timebase

    Frequency range: 300 kHz to20 GHz

    DANL: -155 dBm

    TOI: +22 dBm, typical

    Amplitude accuracy:100 kHz to 13 GHz = 1.3 dB

    Timebase aging: 1.0 ppm/10years

    Frequency accuracy:0.3 ppm + aging

    Precisioncoaxialcable (1.0 m,K-type)

    NI 781611-01 Verifications: Frequency accuracy Source harmonics Source nonharmonic

    spurious

    Adjustments: Timebase

    High-quality, (m)-to-(m) RFcable

    K-typecalibrationdevice(open/short/load)

    AnritsuOSLKF50(included inNI 781590-01)

    Verifications: System dynamic

    range

    No substitute

    K-typeairline

    Anritsu SC7594(included inNI 781590-01)

    Verifications: Corrected directivity

    (5 GHz to 8.5 GHz) Corrected source

    match(5 GHz to 8.5 GHz)

    No substitute

    4 | ni.com | PXIe-5632 Calibration Procedure

  • Table 1. Recommended Equipment for PXIe-5632 Calibration (Continued)

    Equipment RecommendedModel

    Where Used Minimum Requirements

    K-type maleshort

    Anritsu 23K50(included inNI 781590-01)

    Verifications: Corrected source

    match (300 MHz to16 dB

    Torquewrench

    NI 747832-01 All procedures Refer to Test Conditions fortorque wrench specifications.

    PXIe-5632 Calibration Procedure | National Instruments | 5

  • Table 1. Recommended Equipment for PXIe-5632 Calibration (Continued)

    Equipment RecommendedModel

    Where Used Minimum Requirements

    K-type thruline

    NI 781607-01 Manually calibratingtransmission frequencyresponse

    Verifications: Source maximum

    leveled power Trace noise System dynamic

    range

    Return loss: >16 dB

    AutomaticVNAcalibrationkit

    NI 782364-01 Automatically calibratingport one reflection

    K-type, male-female, 70 MHzto 9 GHz

    ManualVNAcalibrationkit

    NI 781587-01 Manually calibrating portone reflection

    K-type

    Test ConditionsThe following setup and environmental conditions are required to ensure the PXIe-5632 meetspublished specifications. Keep cabling as short as possible. Long cables and wires act as antennas, picking up extra

    noise that can affect measurements. Verify that all connections, including front panel connections and screws, are secure. Maintain an ambient temperature of 23 C 5 C. Keep relative humidity between 10% and 90%, noncondensing. Allow a warm-up time of at least 45 minutes after the chassis is powered on and NI-VNA

    is loaded and recognizes the PXIe-5632. The warm-up time ensures that the PXIe-5632and test instrumentation are at a stable operating temperature.

    In each verification procedure, insert a delay between configuring all instruments andacquiring the measurement. This delay may need to be adjusted depending on theinstruments used but should always be at least 1,000 ms for the first iteration, 1,000 mswhen the power level changes, and 100 ms for each other iteration.

    6 | ni.com | PXIe-5632 Calibration Procedure

  • Ensure that the chassis fan speed is set to HIGH, that the fan filters, if present, are clean,and that the empty slots contain filler panels. For more information, refer to the MaintainForced-Air Cooling Note to Users document available at ni.com/manuals.

    Use an appropriate torque wrench or torque screwdriver and SMA driver bit to tighten allmodule RF connectors (SMA or 3.5 mm). NI recommends use of 1 N m (9 lb in.).

    Use a 3.5 mm torque wrench (8 lb in., NI 781615-01 or equivalent) to tighten 3.5 mm or2.92 mm (K) connections without an SMA connector.

    Initial SetupRefer to the PXIe-5632 Getting Started Guide for information about how to install thesoftware and the hardware and how to configure the device in Measurement & AutomationExplorer (MAX).

    Test System CharacterizationThe following procedures characterize the test equipment used during verification andadjustment.

    Caution The connectors on the device under test (DUT) and test equipment arefragile. Perform the steps in these procedures with great care to prevent damagingany DUTs or test equipment.

    Zeroing and Calibrating the Power Sensor1. Ensure that the power sensor is not connected to any signals.2. Zero and calibrate the power sensor using the built-in functions in the power meter,

    according to the power meter documentation.

    Performing User Calibration on the PXIe-5632Allow a 45 minute warm-up period before you begin a user calibration.

    The PXIe-5632 includes precise internal circuits and references used during user calibration toadjust for any errors caused by short-term fluctuations in the environment.

    Note You must perform different user calibrations before verifying certainPXIe-5632 specifications.

    The following verification procedures require you to perform either an automatic or manualport one reflection user calibration or a transmission frequency response user calibration priorto the completion of the verification procedure: System Dynamic Range Corrected Directivity (300 MHz to

  • Corrected Directivity (5 GHz to 8.5 GHz) Corrected Source Match (5 GHz to 8.5 GHz)