ViVitro offers a Left Ventricular Assist Device (LVAD) Lab Service and Pulse Duplicator accessory. Ventricular Cannulation Atrial Cannulation LEFT VENTRICULAR ASSIST DEVICE (LVAD) TESTING Product Code: 25205 > Pulse Duplicator Accessory The LVAD accessory uses the ViVitro Pulse Duplicator to test LVAD devices under physiological pulsatile flows. The system can simulate a variety of conditions while the LVAD is connected to the Pulse Duplicator. The effects on the cardiac output of the system can be monitored while varying the output of the LVAD using the Data Acquisition System. Flow sensors can be mounted at any point in the flow loop. An atrial cannulation option is also available. > Features • Ventricle apex connection - ½” barbed fitting • Aortic chamber (aorta) connection - ½” barbed fitting • Tubing and fittings for connecting to your LVAD PULSE DUPLICATOR ACCESSORY LAB SERVICE

Upload: others

Post on 05-Oct-2020

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: PULSE DUPLICATOR ACCESSORY LAB SERVICE LEFT VENTRICULAR ASSIST DEVICE ... · Aortic Cannulation Optional Atrial Cannulation Ventricle > Specifications Flow 0-15 LPM Beat Rate 3-200

ViVitro offers a Left Ventricular Assist Device (LVAD) Lab Service and Pulse Duplicator accessory.

Ventricular Cannulation Atrial Cannulation

LEFT VENTRICULAR ASSIST DEVICE (LVAD) TESTINGProduct Code: 25205

> Pulse Duplicator AccessoryThe LVAD accessory uses the ViVitro Pulse Duplicator to test LVAD

devices under physiological pulsatile flows. The system can simulate a

variety of conditions while the LVAD is connected to the Pulse Duplicator.

The effects on the cardiac output of the system can be monitored while

varying the output of the LVAD using the Data Acquisition System.

Flow sensors can be mounted at any point in the flow loop. An atrial

cannulation option is also available.

> Features• Ventricle apex connection

- ½” barbed fitting

• Aortic chamber (aorta)

connection - ½” barbed fitting

• Tubing and fittings for

connecting to your LVAD

PULSE DUPLICATOR ACCESSORY LAB SERVICE

Page 2: PULSE DUPLICATOR ACCESSORY LAB SERVICE LEFT VENTRICULAR ASSIST DEVICE ... · Aortic Cannulation Optional Atrial Cannulation Ventricle > Specifications Flow 0-15 LPM Beat Rate 3-200

Aortic Cannulation

Optional Atrial CannulationVentricle

> SpecificationsFlow 0-15 LPM

Beat Rate 3-200 BPM

Pressure ≤ 300 mmHg

Waveforms variable systolic duration sine waveforms, linear waveforms, custom

physiological waveforms, custom arrhythmia waveforms

> Lab ServicesWe offer independent LVAD device testing hydrodynamic performance assessment. Our test plans and

reports are designed around approval bodies, thoroughly researched, and justified to ensure they are

regulatory appropriate. Contact us today to discuss your cardiovascular testing needs. We also offer

non-accredited testing for quicker, less costly Preliminary Evaluation and Feasibility Testing.

* By Blausen Medical Communications, Inc. (Donated via OTRS) [CC BY 3.0 (www.creativecommons.org/licenses/by/3.0)], via Wikimedia Commons

*

Ventricle

Ventricular Cannulation

Aorta

To LVAD

From LVAD

ViVi

TRO

LAB

S 20

16 •

viv

itrol

abs.

com

Copyright © 2016 by ViVitro Labs Inc. All rights reserved.

455 Boleskine Road, Victoria B.C. Canada, V8Z 1E7 [email protected] | (250) 388-3531 | www.vivitrolabs.com

// V

IVI-M

KT-0

73