programmable melting/boiling point apparatus model

4
Programmable Melting/Boiling Point Apparatus Model : µThermoCal25 | Make : ANALAB | HSN Code : 90278090 Supplied with 250ml Beaker, Teon Lid, Tubular Heater, RTD Sensor, Silicon Oil (200g Bottle), Melting Capillary One End Sealed (Pack of 100 Nos.), Boiling Tube, Teon Coated Magnetic Paddle, Sample Tapping Rod, Two Melting Standards 49ºC and 238 ºC, Power Cord, Operational & Instruction Manual with SOP, Calibration Test Certicate (Without Traceability Document) and Warranty Card. www.analab.co.in

Upload: others

Post on 18-Dec-2021

6 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Programmable Melting/Boiling Point Apparatus Model

Programmable Melting/Boiling Point Apparatus

Model : µThermoCal25 | Make : ANALAB | HSN Code : 90278090

Supplied with 250ml Beaker, Teon Lid, Tubular Heater, RTD Sensor, Silicon Oil

(200g Bottle), Melting Capillary One End Sealed (Pack of 100 Nos.), Boiling Tube,

Teon Coated Magnetic Paddle, Sample Tapping Rod, Two Melting Standards 49ºC

and 238 ºC, Power Cord, Operational & Instruction Manual with SOP, Calibration

Test Certicate (Without Traceability Document) and Warranty Card.

www.analab.co.in

Page 2: Programmable Melting/Boiling Point Apparatus Model

Technical Specication:

www.analab.co.in

Temperature Controller : Microcontroller Based

Heating Media : By Silicon Oil

Temperature Range : +5°C above ambient to 300ºC

Readability : 0.1°C

Temperature Accuracy : ± 0.5°C, +5°C above ambient to 300°C

Temperature Heating : Automatic depending upon heating requirement

Overheating Protection : In-built - Set by user in method parameter

Temperature Calibration : On-site with calibrated thermometer OR temperature indicator OR

Pharmacopeia Standard.

Heating Before EMP* : Automatic heating depending upon temperature difference

Heating Rate : 0.2, 0.5, 1.0, 2.0 & 5.0 ºC/min - Selectable after Set Point (EMP)

Display : 20 x 4 Line Alphanumeric Backlit LCD

Magnetic Stirrer : Built-in with soft touch Key speed controller

Display of Melting Range : Simultaneous MP1# & MP2#

Keypad : Membrane waterproof polycarbonate soft touch keypad compliance to

UL/ROHS

Viewing of Melting Capillary : 10x Magnifying Glass

Method : 10 analysis method & 1 for Unknown sample with View, Print & Delete

facility

Calibration : Six Calibration Standards (As per Pharmacopeia AND /OR Customer's

known standards)

Melting Capillary Port : Three

Glass Melting Capillary : OD: 1.4 to 1.6mm | ID : 1.0mm| Length : 100mm |One End Sealed

Sample Filling Height : 3mm

Boiling Tube : One

Password Protection : Date & Time, Calibration Mode, Delete Stored Analysis Data and

Restore Factory Set Calibration Data

Temperature Sensor : Simplex PT-100

Page 3: Programmable Melting/Boiling Point Apparatus Model

www.analab.co.in

Calibration Data Storage : Non Volatile Memory which can store, print & view calibration date &

time, company name, standard name with temperature, actual reading

of standard temperature, observed temperature, calibration done by,

instrument serial number/ID, model number etc.

Analysis Data Storage : More than 500 analysis data can be stored with analyses date & time,

record number, batch/ ID number, sample name, users name, heating

rate, Melting Point (MP1 & MP2), upper & lower calibration standards,

calibration date & time, instrument serial number/ID, model number,

calibration point, calibration date & time, analysis date & time, standard

based on sample temperature, heating rate, melting range (MP1 &

MP2), signature etc. Secondly, analysis data can print directly from

analysis mode without saving analysis data in instrument.

Printer Interface : Centronics Parallel Port / Serial (RS-232 C) Port

Printer : 40 / 80 Column Dot Matrix Printer

(Printer have to procure optionally)

Power Supply : 230 VAC ±10%, 50 Hz

Page 4: Programmable Melting/Boiling Point Apparatus Model

Speci�cation & design are subject to change without prior notice for our continuous development.

www.analab.co.in

Plot No. 361, Sayajimarket,

N. H. No. 8 By-Pass,

At & Post : Sayajipura,

Vadodara - 390019, Gujarat, India.

+91 265 2514404, 9714503393

[email protected]

www.analab.co.in, www.analabscienti�c.com

CIN: U33110GJ2005PTC046555

AN ISO 9001 : 2015 CERTIFIED CO.

Optional

1. IQ / OQ / PQ / DQ Documents

Note:

1) Please provide Customer's Name & Address prior to dispatch.

2) To perform onsite documentation, visit charge will be extra depending upon place of Installation.

3) Traceability documents along with installation report.

4) Standards / Chemicals / Buffers required to perform documents are in customer's scope.

5) We don't send documents by email.

2. Traceability documents

3. Calibration Standard Kit for Melting Point Apparatus - (ANALAB Certied)

Consist of 06 Melting Standards as specied bellow, Melting Capillary One End Sealed

100 mm Long (Pack of 100) – 02 Pack, Boiling Tube – 02 Nos., Sample Tapping Rod.

Melting Standards :

a) Benzophenone - Melting Range : 47 - 49 ºC, Capacity : 5gm

b) Vanillin - Melting Range : 81 - 83 ºC, Capacity : 5gm

c) Benzoic Acid - Melting Range : 121.5 - 123.5 ºC, Capacity : 5gm

d) Sulfanilamide - Melting Range : 164 - 167 ºC, Capacity : 5gm

e) Caffeine - Melting Range : 235 - 238 ºC, Capacity : 5gm

f) Phenolphthalein - Melting Range : 258 - 262 ºC, Capacity : 5gm

Note: Melting Range / Melting Point specied on standard bottle is re-pack based on the values of batch

numberspecied by manufacturer and their COA.

4. Keypad Screen Guard