TECHNICAL REPORT ITAK - RT 090/2020
PROFICIENCY TESTING PROGRAM IN
NICKEL ORE ANALYZES
31st ROUND – JUNE 2020
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 2 / 43
1) Introduction
This Technical Report presents the obtained results, its evaluation and interpretation for the
participant laboratory.
The description of the statistical criteria and techniques used in the evaluation and general
information about this proficiency testing is in a separate module and available to each participant
in ITAK PTP Management System.
ITAK does not issue preliminary report. This current report is characterized as Final Technical
Report.
2) Preliminary Information
Tab. 1: Round information and participant laboratory identification
Round Identification: 31st Round – PT in Nickel Ore Analysis
Period of execution: April and May 2020
Laboratory Identification: PT Vale Indonesia Tbk
Unit/Site: Sorowako – Indonesia
Laboratory Code: 05
Test Item Codes for this round: Ni55 and Ni56
Issue Data of this Report: June 15 of 2020
3) Participants of the Proficiency Testing
The participating laboratories of this proficiency testing are listed below:
Participating Laboratory Name Unit/Site
Anglo American Brasil Ltda – Barro Alto Barro Alto/GO - Brazil
Anglo American Brasil Ltda – Niquelândia Niquelândia/GO - Brazil
Nilab Pouembout - New Caledonia
PT Carsurin Kendari Sulawesi Tenggara - Indonesia
PT Vale Indonesia Tbk Sorowako - Indonesia
Vale – Unidade Operacional Onça Puma Ourilândia do Norte/PA - Brazil
4) Appeal
It is afforded to the participant of Proficiency Testing Programs - ITAK the right of appeal as to
the evaluation of its performance given in this Technical Report. For that, the participating
laboratory who have doubts or does not agree with the performance evaluation should register
an appeal on the ITAK PTP Management System. According to a requirement of ISO/IEC 17043 –
Conformity assessment - General requirements for Proficiency Testing, ITAK maintains a
documented procedure governing all the systematic including deadlines for appeal, analysis and
response to participating client.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 3 / 43
5) Performance Commentary
Tab. 2: Performance evaluation of the participating laboratory – consolidated information.
Test
Item
Analyzed
Parameter
Assigned
Value
obtained by
Standard
Deviation
obtained by
Reference
Tables
Z Score
Performance
Related
Picture
Youden
Performance
Related
Picture
Ni55 Al2O3 Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard
Deviation
ISO 5725-2
4 Satisfactory 5
Insufficient
results -
Ni56 Al2O3 5 Satisfactory
Ni55 CaO Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard
Deviation
ISO 5725-2
7 Satisfactory 8
Insufficient
results -
Ni56 CaO 8 Satisfactory
Ni55 Co Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard
Deviation
ISO 5725-2
10 Satisfactory 11
Insufficient
results -
Ni56 Co 11 Satisfactory
Ni55 Cr2O3 Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard
Deviation
ISO 5725-2
13 Satisfactory 14
Insufficient
results -
Ni56 Cr2O3 14 Satisfactory
Ni55 Cu Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard
Deviation
ISO 5725-2
16 Lab has not
participated 17
Insufficient
results -
Ni56 Cu 17
Ni55 Fe Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard
Deviation
ISO 5725-2
19 Satisfactory 20
Insufficient
results -
Ni56 Fe 20 Satisfactory
Ni55 LOI Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard
Deviation
ISO 5725-2
22 Satisfactory 23
Insufficient
results -
Ni56 LOI 23 Satisfactory
Ni55 MgO Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard
Deviation
ISO 5725-2
25 Satisfactory 26
Insufficient
results -
Ni56 MgO 26 Satisfactory
Ni55 MnO Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard
Deviation
ISO 5725-2
28 Satisfactory 29
Insufficient
results -
Ni56 MnO 29 Satisfactory
Ni55 Ni Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard
Deviation
ISO 5725-2
31 Satisfactory 32
Insufficient
results -
Ni56 Ni 32 Satisfactory
Ni55 SiO2 Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard
Deviation
ISO 5725-2
34 Satisfactory 35
Insufficient
results -
Ni56 SiO2 35 Satisfactory
Ni55 TiO2 Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard
Deviation
ISO 5725-2
37 Lab has not
participated -
Insufficient
results -
Ni56 TiO2 38
Ni55 Zn Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard
Deviation
ISO 5725-2
40 Lab has not
participated 38
Insufficient
results -
Ni56 Zn 41
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 4 / 43
6) Mandel’s Graphs
Pic. 1: Graph with Mandel’s h statistics for sample Ni55 for all the involved parameters.
Pic. 2: Graph with Mandel’s h statistics for sample Ni56 for all involved parameters.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 5 / 43
Pic. 3: Graph with Mandel’s k statistics for sample Ni55 for all involved parameters.
Pic. 4: Graph with Mandel’s k statistics for sample Ni56 for all involved parameters.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 6 / 43
7) Results, Methods, Graphs and Interpretation.
Tab. 3 – Assigned Value, Standard Deviations and Standard Uncertainties for Al2O3.
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard Uncertainty
𝒖 (%) Sample
Ni55 11.03 0.51 0.26
Ni56 3.556 0.12 0.059
Tab. 4 – Results reported, average and Z Scores of laboratories participating in the PT Program for Al2O3
analysis in the sample Ni55.
Results for Al2O3 – Sample Ni55
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 Al2O3 10.140 10.250 10.230 10.230 10.110 % 10.192 -1.660 FUS-XRF
Lab 02 Al2O3 10.420 10.420 10.310 9.570 10.130 % 10.170 -1.703 FUS-XRF
Lab 03 Al2O3 11.050 11.140 11.000 11.070 11.000 % 11.052 0.035 FUS-XRF
Lab 04 Al2O3 11.385 11.277 11.235 11.453 11.280 % 11.326 0.576 FUS-XRF
Lab 05 Al2O3 11.130 11.110 11.120 11.050 11.080 % 11.098 0.126 FUS-XRF
Lab 06 Al2O3 11.480 11.550 11.460 11.480 11.540 % 11.502 0.923 FUS-XRF
Tab. 5 – Results reported, average and Z Scores of laboratories participating in the PT Program for Al2O3
analysis in the sample Ni56.
Results for Al2O3 – Sample Ni56
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 Al2O3 3.450 3.440 3.430 3.470 3.490 % 3.456 -0.862 FUS-XRF
Lab 02 Al2O3 3.580 3.590 3.560 3.580 3.610 % 3.584 0.243 FUS-XRF
Lab 03 Al2O3 3.500 3.530 3.500 3.490 3.450 % 3.494 -0.534 FUS-XRF
Lab 04 Al2O3 3.524 3.545 3.572 3.542 3.553 % 3.547 -0.075 FUS-XRF
Lab 05 Al2O3 3.490 3.480 3.480 3.490 3.480 % 3.484 -0.620 FUS-XRF
Lab 06 Al2O3 3.790 3.760 3.760 3.770 3.770 % 3.770 1.848 FUS-XRF
Legend:
FUS-XRF Fusion / X-Ray Fluorescence
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 7 / 43
Pic. 5: Graph with Z Scores for samples Ni55 and Ni56 for Al2O3 Analysis.
Laboratory 05 - PT Vale Indonesia Tbk presented SATISFACTORY performance for Al2O3 analysis in the
samples Ni55 and Ni56, according to the criterion for Z Score evaluating.
Performance Commentary – Confidence Ellipse:
The valid results are not sufficient to build the Confidence Ellipse for the parameter Al2O3.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 8 / 43
Pic. 6: Ranking Graph according to the Mandel’s k statistic to sample Ni55 for Al2O3 Analysis.
Pic. 7: Ranking Graph according to the Mandel’s k statistic to sample Ni56 for Al2O3 Analysis.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 9 / 43
Tab. 6 – Assigned Value, Standard Deviations and Standard Uncertainties for CaO.
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard Uncertainty
𝒖 (%) Sample
Ni55 0.0110 0.0029 0.0015
Ni56 0.1349 0.015 0.0075
Tab. 7 – Results reported, average and Z Scores of laboratories participating in the PT Program for CaO
analysis in the sample Ni55.
Results for CaO – Sample Ni55
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 CaO 0.010 0.010 0.010 0.010 0.010 % 0.010 -0.326 FUS-XRF
Lab 02 CaO 0.010 0.010 0.000 0.000 0.010 % 0.006 -1.687 FUS-XRF
Lab 03 CaO 0.020 0.020 0.020 0.010 0.010 % 0.016 1.716 FUS-XRF
Lab 04 CaO 0.009 0.007 0.008 0.009 0.009 % 0.008 -0.870 FUS-XRF
Lab 05 CaO 0.014 0.014 0.014 0.014 0.014 % 0.014 1.035 FUS-XRF
Lab 06 CaO 0.020 0.010 0.010 0.010 0.010 % 0.012 0.355 FUS-XRF
Tab. 8 – Results reported, average and Z Scores of laboratories participating in the PT Program for CaO
analysis in the sample Ni56.
Results for CaO – Sample Ni56
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 CaO 0.110 0.110 0.120 0.110 0.110 % 0.112 -1.570 FUS-XRF
Lab 02 CaO 0.140 0.130 0.130 0.130 0.130 % 0.132 -0.201 FUS-XRF
Lab 03 CaO 0.140 0.150 0.140 0.140 0.140 % 0.142 0.484 FUS-XRF
Lab 04 CaO 0.137 0.137 0.139 0.136 0.136 % 0.137 0.142 FUS-XRF
Lab 05 CaO 0.133 0.133 0.133 0.126 0.133 % 0.132 -0.228 FUS-XRF
Lab 06 CaO 0.160 0.150 0.150 0.160 0.155 % 0.155 1.374 FUS-XRF
Legend:
FUS-XRF Fusion / X-Ray Fluorescence
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 10 / 43
Pic. 8: Graph with Z Scores for samples Ni55 and Ni56 for CaO Analysis.
Laboratory 05 - PT Vale Indonesia Tbk presented SATISFACTORY performance for CaO analysis in the
samples Ni55 and Ni56, according to the criterion for Z Score evaluating.
Performance Commentary – Confidence Ellipse:
The valid results are not sufficient to build the Confidence Ellipse for the parameter CaO.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 11 / 43
Pic. 9: Ranking Graph according to the Mandel’s k statistic to sample Ni55 for CaO Analysis.
Pic. 10: Ranking Graph according to the Mandel’s k statistic to sample Ni56 for CaO Analysis.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 12 / 43
Tab. 9 – Assigned Value, Standard Deviations and Standard Uncertainties for Co.
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard Uncertainty
𝒖 (%) Sample
Ni55 0.1028 0.018 0.0092
Ni56 0.1095 0.018 0.0093
Tab. 10 – Results reported, average and Z Scores of laboratories participating in the PT Program for Co
analysis in the sample Ni55.
Results for Co – Sample Ni55
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 Co 0.086 0.086 0.085 0.086 0.086 % 0.086 -0.944 FUS-XRF
Lab 02 Co 0.081 0.081 0.081 0.081 0.081 % 0.081 -1.211 FUS-XRF
Lab 03 Co 0.116 0.117 0.117 0.116 0.117 % 0.117 0.770 FUS-XRF
Lab 04 Co 0.112 0.111 0.111 0.112 0.111 % 0.111 0.481 FUS-XRF
Lab 05 Co 0.120 0.119 0.120 0.118 0.118 % 0.119 0.904 FUS-XRF
Lab 06 Co 0.110 0.112 0.108 0.108 0.104 % 0.108 0.314 FUS-XRF
Tab. 11 – Results reported, average and Z Scores of laboratories participating in the PT Program for Co
analysis in the sample Ni56.
Results for Co – Sample Ni56
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 Co 0.085 0.085 0.086 0.085 0.085 % 0.085 -1.335 FUS-XRF
Lab 02 Co 0.088 0.087 0.087 0.089 0.089 % 0.088 -1.181 FUS-XRF
Lab 03 Co 0.123 0.125 0.125 0.122 0.121 % 0.123 0.755 FUS-XRF
Lab 04 Co 0.115 0.115 0.117 0.116 0.117 % 0.116 0.359 FUS-XRF
Lab 05 Co 0.118 0.117 0.117 0.118 0.118 % 0.118 0.447 FUS-XRF
Lab 06 Co 0.125 0.127 0.127 0.128 0.127 % 0.127 0.954 FUS-XRF
Legend:
FUS-XRF Fusion / X-Ray Fluorescence
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 13 / 43
Pic. 11: Graph with Z Scores for samples Ni55 and Ni56 for Co Analysis.
Laboratory 05 - PT Vale Indonesia Tbk presented SATISFACTORY performance for Co analysis in the
samples Ni55 and Ni56, according to the criterion for Z Score evaluating.
Performance Commentary – Confidence Ellipse:
The valid results are not sufficient to build the Confidence Ellipse for the parameter Co.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 14 / 43
Pic. 12: Ranking Graph according to the Mandel’s k statistic to sample Ni55 for Co Analysis.
Pic. 13: Ranking Graph according to the Mandel’s k statistic to sample Ni56 for Co Analysis.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 15 / 43
Tab. 12 – Assigned Value, Standard Deviations and Standard Uncertainties for Cr2O3.
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard Uncertainty
𝒖 (%) Sample
Ni55 5.444 0.18 0.094
Ni56 4.25 0.22 0.11
Tab. 13 – Results reported, average and Z Scores of laboratories participating in the PT Program for Cr2O3
analysis in the sample Ni55.
Results for Cr2O3 – Sample Ni55
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 Cr2O3 5.180 5.190 5.080 5.080 5.170 % 5.140 -1.647 FUS-XRF
Lab 02 Cr2O3 5.520 5.550 5.520 5.600 5.600 % 5.558 0.615 FUS-XRF
Lab 03 Cr2O3 5.610 5.660 5.590 5.610 5.590 % 5.612 0.907 FUS-XRF
Lab 04 Cr2O3 5.497 5.459 5.450 5.486 5.418 % 5.462 0.096 FUS-XRF
Lab 05 Cr2O3 5.330 5.320 5.330 5.310 5.310 % 5.320 -0.673 FUS-XRF
Lab 06 Cr2O3 5.585 5.590 5.557 5.559 5.579 % 5.574 0.702 FUS-XRF
Tab. 14 – Results reported, average and Z Scores of laboratories participating in the PT Program for Cr2O3
analysis in the sample Ni56.
Results for Cr2O3 – Sample Ni56
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 Cr2O3 3.870 3.870 3.930 3.850 3.880 % 3.880 -1.648 FUS-XRF
Lab 02 Cr2O3 4.470 4.450 4.410 4.510 4.550 % 4.478 1.046 FUS-XRF
Lab 03 Cr2O3 4.280 4.280 4.330 4.270 4.260 % 4.284 0.172 FUS-XRF
Lab 04 Cr2O3 4.274 4.275 4.346 4.311 4.311 % 4.303 0.259 FUS-XRF
Lab 05 Cr2O3 4.120 4.120 4.080 4.110 4.110 % 4.108 -0.621 FUS-XRF
Lab 06 Cr2O3 4.417 4.425 4.415 4.430 4.422 % 4.422 0.793 FUS-XRF
Legend:
FUS-XRF Fusion / X-Ray Fluorescence
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 16 / 43
Pic. 14: Graph with Z Scores for samples Ni55 and Ni56 for Cr2O3 Analysis.
Laboratory 05 - PT Vale Indonesia Tbk presented SATISFACTORY performance for Cr2O3 analysis in the
samples Ni55 and Ni56, according to the criterion for Z Score evaluating.
Performance Commentary – Confidence Ellipse:
The valid results are not sufficient to build the Confidence Ellipse for the parameter Cr2O3.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 17 / 43
Pic. 15: Ranking Graph according to the Mandel’s k statistic to sample Ni55 for Cr2O3 Analysis.
Pic. 16: Ranking Graph according to the Mandel’s k statistic to sample Ni56 for Cr2O3 Analysis.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 18 / 43
Tab. 15 – Assigned Value, Standard Deviations and Standard Uncertainties for Cu.
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard Uncertainty
𝒖 (%) Sample
Ni55 0.086 0.024 0.014
Ni56 0.62 0.19 0.11
Tab. 16 – Results reported, average and Z Scores of laboratories participating in the PT Program for Cu
analysis in the sample Ni55.
Results for Cu – Sample Ni55
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 Cu 0.082 0.082 0.080 0.080 0.082 % 0.081 -0.196 FUS-XRF
Lab 02 Cu 0.108 0.111 0.108 0.110 0.110 % 0.109 0.973 FUS-XRF
Lab 03 Cu 0.100 0.102 0.098 0.100 0.101 % 0.100 0.593 FUS-XRF
Lab 04 Cu 0.092 0.092 0.090 0.092 0.092 % 0.091 0.229 FUS-XRF
Lab 05 Cu --- --- --- --- --- % --- --- --- Lab 06 Cu 0.046 0.048 0.046 0.046 0.048 % 0.047 -1.612 FUS-XRF
Tab. 17 – Results reported, average and Z Scores of laboratories participating in the PT Program for Cu
analysis in the sample Ni56.
Results for Cu – Sample Ni56
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 Cu 0.558 0.559 0.557 0.552 0.554 % 0.556 -0.329 FUS-XRF
Lab 02 Cu 0.838 0.835 0.824 0.836 0.829 % 0.833 1.095 FUS-XRF
Lab 03 Cu 0.714 0.721 0.703 0.710 0.708 % 0.711 0.470 FUS-XRF
Lab 04 Cu 0.676 0.677 0.684 0.676 0.683 % 0.679 0.306 FUS-XRF
Lab 05 Cu --- --- --- --- --- % --- --- --- Lab 06 Cu 0.317 0.318 0.322 0.324 0.320 % 0.320 -1.542 FUS-XRF
Legend:
FUS-XRF Fusion / X-Ray Fluorescence
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 19 / 43
Pic. 17: Graph with Z Scores for samples Ni55 and Ni56 for Cu Analysis.
Laboratory 05 - PT Vale Indonesia Tbk has not participated of the PT for the parameter Cu.
Performance Commentary – Confidence Ellipse:
The valid results are not sufficient to build the Confidence Ellipse for the parameter Cu.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 20 / 43
Pic. 18: Ranking Graph according to the Mandel’s k statistic to sample Ni55 for Cu Analysis.
Pic. 19: Ranking Graph according to the Mandel’s k statistic to sample Ni56 for Cu Analysis.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 21 / 43
Tab. 18 – Assigned Value, Standard Deviations and Standard Uncertainties for Fe.
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard Uncertainty
𝒖 (%) Sample
Ni55 43.49 0.28 0.15
Ni56 40.51 0.57 0.29
Tab. 19 – Results reported, average and Z Scores of laboratories participating in the PT Program for Fe
analysis in the sample Ni55.
Results for Fe – Sample Ni55
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 Fe 39.550 39.600 38.810 39.740 39.700 % 39.480 -14.106 FUS-XRF
Lab 02 Fe 40.070 40.210 39.930 40.230 40.300 % 40.148 -11.759 FUS-XRF
Lab 03 Fe 43.610 43.810 43.790 43.770 43.550 % 43.706 0.742 FUS-XRF
Lab 04 Fe 43.353 43.179 42.954 43.289 43.038 % 43.163 -1.167 FUS-XRF
Lab 05 Fe 43.950 43.600 43.840 43.320 43.350 % 43.612 0.412 FUS-XRF
Lab 06 Fe 43.484 43.555 43.463 43.458 43.531 % 43.498 0.012 FUS-XRF
Tab. 20 – Results reported, average and Z Scores of laboratories participating in the PT Program for Fe
analysis in the sample Ni56.
Results for Fe – Sample Ni56
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 Fe 36.480 36.500 36.820 36.130 36.240 % 36.434 -7.144 FUS-XRF
Lab 02 Fe 39.770 39.690 39.430 39.890 39.750 % 39.706 -1.409 FUS-XRF
Lab 03 Fe 41.240 41.290 41.400 41.050 41.100 % 41.216 1.238 FUS-XRF
Lab 04 Fe 40.442 40.393 40.888 40.527 40.878 % 40.626 0.203 FUS-XRF
Lab 05 Fe 40.550 40.330 40.140 40.200 40.440 % 40.332 -0.311 FUS-XRF
Lab 06 Fe 40.659 40.668 40.679 40.669 40.669 % 40.669 0.279 FUS-XRF
Legend:
FUS-XRF Fusion / X-Ray Fluorescence
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 22 / 43
Pic. 20: Graph with Z Scores for samples Ni55 and Ni56 for Fe Analysis.
Laboratory 05 - PT Vale Indonesia Tbk presented SATISFACTORY performance for Fe analysis in the
samples Ni55 and Ni56, according to the criterion for Z Score evaluating.
Performance Commentary – Confidence Ellipse:
The valid results are not sufficient to build the Confidence Ellipse for the parameter Fe.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 23 / 43
Pic. 21: Ranking Graph according to the Mandel’s k statistic to sample Ni55 for Fe Analysis.
Pic. 22: Ranking Graph according to the Mandel’s k statistic to sample Ni56 for Fe Analysis.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 24 / 43
Tab. 21 – Assigned Value, Standard Deviations and Standard Uncertainties for LOI (Loss On Ignition).
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard Uncertainty
𝒖 (%) Sample
Ni55 7.43 0.38 0.19
Ni56 8.43 0.66 0.34
Tab. 22 – Results reported, average and Z Scores of laboratories participating in the PT Program for LOI
analysis in the sample Ni55.
Results for LOI – Sample Ni55
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 LOI 7.460 7.840 7.520 7.600 7.310 % 7.546 0.311 GRAV
Lab 02 LOI 7.470 7.430 7.260 7.520 7.370 % 7.410 -0.050 GRAV
Lab 03 LOI 8.240 7.960 8.100 7.850 7.860 % 8.002 1.520 GRAV
Lab 04 LOI 7.087 7.058 6.984 7.021 7.017 % 7.033 -1.049 TGA
Lab 05 LOI 7.030 7.020 7.080 7.060 7.040 % 7.046 -1.016 GRAV
Lab 06 LOI 7.570 7.501 7.480 7.577 7.551 % 7.536 0.284 GRAV
Tab. 23 – Results reported, average and Z Scores of laboratories participating in the PT Program for LOI
analysis in the sample Ni56.
Results for LOI – Sample Ni56
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 LOI 8.870 8.270 8.720 8.710 9.010 % 8.716 0.440 GRAV
Lab 02 LOI 8.280 8.640 8.560 8.580 8.320 % 8.476 0.076 GRAV
Lab 03 LOI 9.470 9.330 9.410 9.520 9.640 % 9.474 1.589 GRAV
Lab 04 LOI 7.869 7.852 7.944 7.908 7.968 % 7.908 -0.784 TGA
Lab 05 LOI 7.820 7.810 7.880 7.870 7.830 % 7.842 -0.885 GRAV
Lab 06 LOI 8.435 8.336 8.502 8.439 8.428 % 8.428 0.004 GRAV
Legend:
GRAV Gravimetric Method
TGA ThermoGravimetric Analyzer
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 25 / 43
Pic. 23: Graph with Z Scores for samples Ni55 and Ni56 for LOI Analysis.
Laboratory 05 - PT Vale Indonesia Tbk presented SATISFACTORY performance for LOI analysis in the
samples Ni55 and Ni56, according to the criterion for Z Score evaluating.
Performance Commentary – Confidence Ellipse:
The valid results are not sufficient to build the Confidence Ellipse for the parameter LOI.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 26 / 43
Pic. 24: Ranking Graph according to the Mandel’s k statistic to sample Ni55 for LOI Analysis.
Pic. 25: Ranking Graph according to the Mandel’s k statistic to sample Ni56 for LOI Analysis.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 27 / 43
Tab. 24 – Assigned Value, Standard Deviations and Standard Uncertainties for MgO.
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard Uncertainty
𝒖 (%) Sample
Ni55 1.058 0.11 0.058
Ni56 3.06 0.20 0.10
Tab. 25 – Results reported, average and Z Scores of laboratories participating in the PT Program for MgO
analysis in the sample Ni55.
Results for MgO – Sample Ni55
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 MgO 1.010 1.030 1.030 1.050 1.030 % 1.030 -0.242 FUS-XRF
Lab 02 MgO 0.910 0.900 0.900 0.830 0.900 % 0.888 -1.488 FUS-XRF
Lab 03 MgO 1.130 1.210 1.150 1.160 1.140 % 1.158 0.880 FUS-XRF
Lab 04 MgO 1.159 1.137 1.161 1.157 1.147 % 1.152 0.830 FUS-XRF
Lab 05 MgO 1.000 1.000 1.100 1.100 1.100 % 1.060 0.021 FUS-XRF
Lab 06 MgO 1.560 1.560 1.528 1.542 1.542 % 1.546 4.288 FUS-XRF
Tab. 26 – Results reported, average and Z Scores of laboratories participating in the PT Program for MgO
analysis in the sample Ni56.
Results for MgO – Sample Ni56
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 MgO 2.740 2.950 2.910 2.740 2.820 % 2.832 -1.098 FUS-XRF
Lab 02 MgO 3.210 3.220 3.190 3.290 3.240 % 3.230 0.856 FUS-XRF
Lab 03 MgO 2.860 2.900 2.900 2.900 2.850 % 2.882 -0.853 FUS-XRF
Lab 04 MgO 2.864 2.880 2.897 2.864 2.877 % 2.876 -0.880 FUS-XRF
Lab 05 MgO 2.900 3.000 3.000 3.000 3.000 % 2.980 -0.372 FUS-XRF
Lab 06 MgO 3.310 3.302 3.294 3.334 3.310 % 3.310 1.249 FUS-XRF
Legend:
FUS-XRF Fusion / X-Ray Fluorescence
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 28 / 43
Pic. 26: Graph with Z Scores for samples Ni55 and Ni56 for MgO Analysis.
Laboratory 05 - PT Vale Indonesia Tbk presented SATISFACTORY performance for MgO analysis in the
samples Ni55 and Ni56, according to the criterion for Z Score evaluating.
Performance Commentary – Confidence Ellipse:
The valid results are not sufficient to build the Confidence Ellipse for the parameter MgO.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 29 / 43
Pic. 27: Ranking Graph according to the Mandel’s k statistic to sample Ni55 for MgO Analysis.
Pic. 28: Ranking Graph according to the Mandel’s k statistic to sample Ni56 for MgO Analysis.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 30 / 43
Tab. 27 – Assigned Value, Standard Deviations and Standard Uncertainties for MnO.
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard Uncertainty
𝒖 (%) Sample
Ni55 0.891 0.039 0.020
Ni56 0.962 0.042 0.021
Tab. 28 – Results reported, average and Z Scores of laboratories participating in the PT Program for MnO
analysis in the sample Ni55.
Results for MnO – Sample Ni55
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 MnO 0.850 0.850 0.830 0.870 0.830 % 0.846 -1.136 FUS-XRF
Lab 02 MnO 0.880 0.880 0.880 0.880 0.880 % 0.880 -0.274 FUS-XRF
Lab 03 MnO 0.920 0.920 0.920 0.920 0.920 % 0.920 0.740 FUS-XRF
Lab 04 MnO 0.883 0.879 0.874 0.881 0.876 % 0.879 -0.309 FUS-XRF
Lab 05 MnO 0.880 0.870 0.870 0.850 0.870 % 0.868 -0.578 FUS-XRF
Lab 06 MnO 0.946 0.967 0.940 0.946 0.962 % 0.952 1.556 FUS-XRF
Tab. 29 – Results reported, average and Z Scores of laboratories participating in the PT Program for MnO
analysis in the sample Ni56.
Results for MnO – Sample Ni56
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 MnO 0.900 0.900 0.910 0.890 0.900 % 0.900 -1.480 FUS-XRF
Lab 02 MnO 0.980 0.980 0.970 0.990 0.980 % 0.980 0.424 FUS-XRF
Lab 03 MnO 1.000 1.000 1.010 1.000 0.990 % 1.000 0.899 FUS-XRF
Lab 04 MnO 0.959 0.958 0.968 0.961 0.969 % 0.963 0.019 FUS-XRF
Lab 05 MnO 0.930 0.930 0.920 0.930 0.920 % 0.926 -0.861 FUS-XRF
Lab 06 MnO 1.001 1.007 1.007 1.002 1.004 % 1.004 0.999 FUS-XRF
Legend:
FUS-XRF Fusion / X-Ray Fluorescence
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 31 / 43
Pic. 29: Graph with Z Scores for samples Ni55 and Ni56 for MnO Analysis.
Laboratory 05 - PT Vale Indonesia Tbk presented SATISFACTORY performance for MnO analysis in the
samples Ni55 and Ni56, according to the criterion for Z Score evaluating.
Performance Commentary – Confidence Ellipse:
The valid results are not sufficient to build the Confidence Ellipse for the parameter MnO.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 32 / 43
Pic. 30: Ranking Graph according to the Mandel’s k statistic to sample Ni55 for MnO Analysis.
Pic. 31: Ranking Graph according to the Mandel’s k statistic to sample Ni56 for MnO Analysis.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 33 / 43
Tab. 30 – Assigned Value, Standard Deviations and Standard Uncertainties for Ni.
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard Uncertainty
𝒖 (%) Sample
Ni55 0.5498 0.0164 0.0083
Ni56 1.1538 0.016 0.0080
Tab. 31 – Results reported, average and Z Scores of laboratories participating in the PT Program for Ni analysis in the sample Ni55.
Results for Ni – Sample Ni55
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 Ni 0.540 0.550 0.530 0.530 0.530 % 0.536 -0.844 FUS-XRF
Lab 02 Ni 0.530 0.530 0.530 0.540 0.540 % 0.534 -0.966 FUS-XRF
Lab 03 Ni 0.550 0.550 0.550 0.550 0.540 % 0.548 -0.110 FUS-XRF
Lab 04 Ni 0.552 0.548 0.544 0.550 0.547 % 0.548 -0.098 FUS-XRF
Lab 05 Ni 0.560 0.550 0.560 0.560 0.550 % 0.556 0.379 FUS-XRF
Lab 06 Ni 0.574 0.576 0.571 0.576 0.586 % 0.577 1.639 FUS-XRF
Tab. 32 – Results reported, average and Z Scores of laboratories participating in the PT Program for Ni analysis in the sample Ni56.
Results for Ni – Sample Ni56
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 Ni 1.010 1.020 1.030 1.010 1.010 % 1.016 -8.788 FUS-XRF
Lab 02 Ni 1.150 1.150 1.140 1.150 1.150 % 1.148 -0.370 FUS-XRF
Lab 03 Ni 1.150 1.150 1.160 1.140 1.130 % 1.146 -0.497 FUS-XRF
Lab 04 Ni 1.153 1.155 1.167 1.155 1.168 % 1.160 0.370 FUS-XRF
Lab 05 Ni 1.150 1.140 1.130 1.130 1.150 % 1.140 -0.880 FUS-XRF
Lab 06 Ni 1.172 1.179 1.174 1.177 1.175 % 1.175 1.378 FUS-XRF
Legend:
FUS-XRF Fusion / X-Ray Fluorescence
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 34 / 43
Pic. 32: Graph with Z Scores for samples Ni55 and Ni56 for Ni Analysis.
Laboratory 05 - PT Vale Indonesia Tbk presented SATISFACTORY performance for Ni analysis in the
samples Ni55 and Ni56, according to the criterion for Z Score evaluating.
Performance Commentary – Confidence Ellipse:
The valid results are not sufficient to build the Confidence Ellipse for the parameter Ni.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 35 / 43
Pic. 33: Ranking Graph according to the Mandel’s k statistic to sample Ni55 for Ni Analysis.
Pic. 34: Ranking Graph according to the Mandel’s k statistic to sample Ni56 for Ni Analysis.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 36 / 43
Tab. 33 – Assigned Value, Standard Deviations and Standard Uncertainties for SiO2.
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard Uncertainty
𝒖 (%) Sample
Ni55 9.99 0.48 0.24
Ni56 18.22 0.41 0.21
Tab. 34 – Results reported, average and Z Scores of laboratories participating in the PT Program for SiO2
analysis in the sample Ni55.
Results for SiO2 – Sample Ni55
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 SiO2 9.400 9.350 9.420 9.460 9.390 % 9.404 -1.230 FUS-XRF
Lab 02 SiO2 9.510 9.840 9.340 8.930 9.210 % 9.366 -1.309 FUS-XRF
Lab 03 SiO2 10.180 10.370 10.210 10.260 10.220 % 10.248 0.531 FUS-XRF
Lab 04 SiO2 10.029 10.076 10.050 10.108 10.099 % 10.072 0.164 FUS-XRF
Lab 05 SiO2 10.700 10.600 10.600 10.500 10.600 % 10.600 1.265 FUS-XRF
Lab 06 SiO2 9.680 9.548 9.657 9.693 9.639 % 9.643 -0.730 FUS-XRF
Tab. 35 – Results reported, average and Z Scores of laboratories participating in the PT Program for SiO2
analysis in the sample Ni56.
Results for SiO2 – Sample Ni56
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 SiO2 17.740 17.790 17.620 17.830 17.830 % 17.762 -1.101 FUS-XRF
Lab 02 SiO2 17.800 17.840 17.810 17.710 17.630 % 17.758 -1.110 FUS-XRF
Lab 03 SiO2 18.440 18.580 18.570 18.410 18.210 % 18.442 0.540 FUS-XRF
Lab 04 SiO2 18.248 18.254 18.449 18.202 18.341 % 18.299 0.195 FUS-XRF
Lab 05 SiO2 18.300 18.200 18.200 18.200 18.300 % 18.240 0.053 FUS-XRF
Lab 06 SiO2 18.833 18.782 18.769 18.847 18.808 % 18.808 1.423 FUS-XRF
Legend:
FUS-XRF Fusion / X-Ray Fluorescence
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 37 / 43
Pic. 35: Graph with Z Scores for samples Ni55 and Ni56 for SiO2 Analysis.
Laboratory 05 - PT Vale Indonesia Tbk presented SATISFACTORY performance for SiO2 analysis in the
samples Ni55 and Ni56, according to the criterion for Z Score evaluating.
Performance Commentary – Confidence Ellipse:
The valid results are not sufficient to build the Confidence Ellipse for the parameter SiO2.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 38 / 43
Pic. 36: Ranking Graph according to the Mandel’s k statistic to sample Ni55 for SiO2 Analysis.
Pic. 37: Ranking Graph according to the Mandel’s k statistic to sample Ni56 for SiO2 Analysis.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 39 / 43
Tab. 36 – Assigned Value, Standard Deviations and Standard Uncertainties for TiO2.
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard Uncertainty
𝒖 (%) Sample
Ni55 --- --- ---
Ni56 --- --- ---
Tab. 37 – Results reported, average and Z Scores of laboratories participating in the PT Program for TiO2
analysis in the sample Ni55.
Results for TiO2 – Sample Ni55
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 TiO2 --- --- --- --- --- % --- --- ---
Lab 02 TiO2 --- --- --- --- --- % --- --- ---
Lab 03 TiO2 0.230 0.230 0.230 0.230 0.230 % 0.230 --- FUS-XRF
Lab 04 TiO2 --- --- --- --- --- % --- --- --- Lab 05 TiO2 --- --- --- --- --- % --- --- --- Lab 06 TiO2 0.224 0.222 0.223 0.225 0.213 % 0.221 --- FUS-XRF
Tab. 38 – Results reported, average and Z Scores of laboratories participating in the PT Program for TiO2
analysis in the sample Ni56.
Results for TiO2 – Sample Ni56
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 TiO2 --- --- --- --- --- % --- --- ---
Lab 02 TiO2 --- --- --- --- --- % --- --- ---
Lab 03 TiO2 0.110 0.120 0.110 0.110 0.110 % 0.112 --- FUS-XRF
Lab 04 TiO2 --- --- --- --- --- % --- --- --- Lab 05 TiO2 --- --- --- --- --- % --- --- --- Lab 06 TiO2 0.114 0.113 0.110 0.113 0.112 % 0.112 --- FUS-XRF
Legend:
FUS-XRF Fusion / X-Ray Fluorescence
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 40 / 43
Assigned Value, EP Standard Deviation and Standard Uncertainty Calculations:
The valid results are not sufficient to calculate these parameters.
Mandel’s Graphics:
The valid results are not sufficient to build the Mandel’s Graphics for the parameter TiO2.
Z Scores:
The valid results are not sufficient to calculate the Z Scores for the parameter TiO2.
Confidence Ellipse:
The valid results are not sufficient to build the Confidence Ellipse for the parameter TiO2.
Ranking Graphics:
The valid results are not sufficient to build the Ranking Graphics for the parameter TiO2.
.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 41 / 43
Tab. 39 – Assigned Value, Standard Deviations and Standard Uncertainties for Zn.
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard Uncertainty
𝒖 (%) Sample
Ni55 0.0289 0.0018 0.0013
Ni56 0.046 0.014 0.010
Tab. 40 – Results reported, average and Z Scores of laboratories participating in the PT Program for Zn
analysis in the sample Ni55.
Results for Zn – Sample Ni55
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 Zn --- --- --- --- --- % --- --- ---
Lab 02 Zn --- --- --- --- --- % --- --- ---
Lab 03 Zn 0.030 0.030 0.030 0.030 0.030 % 0.030 0.603 FUS-XRF
Lab 04 Zn 0.027 0.027 0.026 0.028 0.027 % 0.027 -1.115 FUS-XRF
Lab 05 Zn --- --- --- --- --- % --- --- --- Lab 06 Zn 0.030 0.031 0.029 0.029 0.030 % 0.030 0.490 FUS-XRF
Tab. 41 – Results reported, average and Z Scores of laboratories participating in the PT Program for Zn
analysis in the sample Ni56.
Results for Zn – Sample Ni56
Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 Zn --- --- --- --- --- % --- --- ---
Lab 02 Zn --- --- --- --- --- % --- --- ---
Lab 03 Zn 0.030 0.030 0.030 0.030 0.030 % 0.030 -1.154 FUS-XRF
Lab 04 Zn 0.054 0.053 0.055 0.054 0.055 % 0.054 0.546 FUS-XRF
Lab 05 Zn --- --- --- --- --- % --- --- --- Lab 06 Zn 0.054 0.055 0.054 0.055 0.055 % 0.055 0.591 FUS-XRF
Legend:
FUS-XRF Fusion / X-Ray Fluorescence
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 42 / 43
Pic. 38: Graph with Z Scores for samples Ni55 and Ni56 for Zn Analysis.
Laboratory 05 - PT Vale Indonesia Tbk has not participated of the PT for the parameter Zn.
Performance Commentary – Confidence Ellipse:
The valid results are not sufficient to build the Confidence Ellipse for the parameter Zn.
PT Report for Nickel Ore Analyzes – 31st Round
Cgcre’s Accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation nº. PEP 0021
Contacts: e-mail: [email protected]; Telephone number: +55 31 3851-3166
Pág. 43 / 43
Pic. 39: Ranking Graph according to the Mandel’s k statistic to sample Ni55 for Zn Analysis.
Pic. 40: Ranking Graph according to the Mandel’s k statistic to sample Ni56 for Zn Analysis.