Primary water calorimetry in clinical electron beams, scanned proton beams, and 192 Ir brachytherapy dosimetry Primary water calorimetry in clinical electron beams, scanned proton beams, and 192 Ir brachytherapy dosimetry A. Sarfehnia 1 , K. Stewart 1 , C. Ross 2 , M. McEwen 2 , B. Clasie 3 , E. Chung 1 , H. M. Lu 3 , J. Flanz 3 , E. Cascio 3 , M. Engelsman 3 , H. Paganetti 3 , J. Seuntjens 1 1 Medical Physics Unit, McGill University, Montreal, Quebec, Canada 2 National Research Council of Canada, Ottawa, Ontario, Canada 3 Massachusetts General Hospital and Harvard Medical School, Boston, MA, USA A. Sarfehnia 1 , K. Stewart 1 , C. Ross 2 , M. McEwen 2 , B. Clasie 3 , E. Chung 1 , H. M. Lu 3 , J. Flanz 3 , E. Cascio 3 , M. Engelsman 3 , H. Paganetti 3 , J. Seuntjens 1 1 Medical Physics Unit, McGill University, Montreal, Quebec, Canada 2 National Research Council of Canada, Ottawa, Ontario, Canada 3 Massachusetts General Hospital and Harvard Medical School, Boston, MA, USA International Symposium on Standards, Applications and Quality Assurance in Medical Radiation Dosimetry International Symposium on Standards, Applications and Quality Assurance in Medical Radiation Dosimetry

Upload: vuongthuy

Post on 13-Jul-2018

219 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Primary water calorimetry in clinical electron beams ... water calorimetry in clinical electron beams, scanned proton beams, and 192Ir brachytherapy dosimetry Ir brachytherapy dosimetry

Primary water calorimetry in clinical electron beams, scanned proton

beams, and 192Ir brachytherapy dosimetry

Primary water calorimetry in clinical electron beams, scanned proton

beams, and 192Ir brachytherapy dosimetry

A. Sarfehnia1, K. Stewart1, C. Ross2, M. McEwen2, B. Clasie3, E. Chung1, H. M. Lu3, J. Flanz3, E. Cascio3, M.

Engelsman3, H. Paganetti3, J. Seuntjens1

1 Medical Physics Unit, McGill University, Montreal, Quebec, Canada2 National Research Council of Canada, Ottawa, Ontario, Canada

3 Massachusetts General Hospital and Harvard Medical School, Boston, MA, USA

A. Sarfehnia1, K. Stewart1, C. Ross2, M. McEwen2, B. Clasie3, E. Chung1, H. M. Lu3, J. Flanz3, E. Cascio3, M.

Engelsman3, H. Paganetti3, J. Seuntjens1

1 Medical Physics Unit, McGill University, Montreal, Quebec, Canada2 National Research Council of Canada, Ottawa, Ontario, Canada

3 Massachusetts General Hospital and Harvard Medical School, Boston, MA, USA

International Symposium on Standards, Applications and Quality Assurance in Medical Radiation Dosimetry

International Symposium on Standards, Applications and Quality Assurance in Medical Radiation Dosimetry

Page 2: Primary water calorimetry in clinical electron beams ... water calorimetry in clinical electron beams, scanned proton beams, and 192Ir brachytherapy dosimetry Ir brachytherapy dosimetry

Water Calorimetry• Calorimetry is the only true primary standard for

absorbed dose to water measurement in radiation dosimetry.

• Advantages: • In principle, energy-independent• Measures Dw to a point

• Currently, implemented only for high energy photon and electron beams.

Dw (vr ) = cw, p ⋅ ΔT (vr ) ⋅ ki∏

Page 3: Primary water calorimetry in clinical electron beams ... water calorimetry in clinical electron beams, scanned proton beams, and 192Ir brachytherapy dosimetry Ir brachytherapy dosimetry

Water Calorimetry (experiment)

Page 4: Primary water calorimetry in clinical electron beams ... water calorimetry in clinical electron beams, scanned proton beams, and 192Ir brachytherapy dosimetry Ir brachytherapy dosimetry

Water Calorimetry (experiment)

Page 5: Primary water calorimetry in clinical electron beams ... water calorimetry in clinical electron beams, scanned proton beams, and 192Ir brachytherapy dosimetry Ir brachytherapy dosimetry

Water Calorimetry (experiment)

Time

Tem

pera

tur e

Ris

e

Predrift Irradiation Postdrift

Page 6: Primary water calorimetry in clinical electron beams ... water calorimetry in clinical electron beams, scanned proton beams, and 192Ir brachytherapy dosimetry Ir brachytherapy dosimetry

Water Calorimetry (experiment)

Time

Tem

pera

tur e

Ris

e

Predrift Irradiation Postdrift

Dw (vr ) = cw, p ⋅ ΔT (vr ) ⋅ ki∏

ΔT

Page 7: Primary water calorimetry in clinical electron beams ... water calorimetry in clinical electron beams, scanned proton beams, and 192Ir brachytherapy dosimetry Ir brachytherapy dosimetry

Heat transport correction factor

• Comsol Multiphysics software was used to solve the heat transport problem using Finite Element Method (FEM)

ρcp∂T∂t

+ ∇ ⋅ −k∇T + ρcpvuT( )= Q

∇ ⋅ vu = 0

ρ∂vu∂t

−η∇2 vu + ρ vu ⋅∇( )vu + ∇p =vF

kht =ΔTw, ideal

calc

ΔTw, actualcalckht =

ΔTw, idealcalc

ΔTw, actualcalc

Page 8: Primary water calorimetry in clinical electron beams ... water calorimetry in clinical electron beams, scanned proton beams, and 192Ir brachytherapy dosimetry Ir brachytherapy dosimetry

AIM

• The aim of our research program is to develop water calorimetry for the following distinct applications

• Clinical high energy electron beams• 192Ir HDR brachytherapy• Scattered/Scanned proton beam

delivery

Page 9: Primary water calorimetry in clinical electron beams ... water calorimetry in clinical electron beams, scanned proton beams, and 192Ir brachytherapy dosimetry Ir brachytherapy dosimetry

High Energy Electrons• Varian Clinac 21EX

• 6, 9, 12, 16, 20 MeV

• In 6 MeV, the thermistors are positioned only 13.4 mm below water surface.

Page 10: Primary water calorimetry in clinical electron beams ... water calorimetry in clinical electron beams, scanned proton beams, and 192Ir brachytherapy dosimetry Ir brachytherapy dosimetry

High Energy Electrons Dw

Q (vr ) = M ⋅ND,w

60 Co ⋅ kecal ⋅ kR50

' ⋅PgrQ

Page 11: Primary water calorimetry in clinical electron beams ... water calorimetry in clinical electron beams, scanned proton beams, and 192Ir brachytherapy dosimetry Ir brachytherapy dosimetry

High Energy ElectronsPresent Work AAPM TG-51 IAEA TRS-398 Mainegra-Hing et

al*

Estimated from the work of Buckley and Rogers**

PTW Roos 0.895 0.901 0.895 0.896 --

Exradin A12 0.918 0.906 0.910 -- 0.904

R50 {cm}

K’ R

50

* E. Mainegra-Hing, I. Kawrakow, D. W. O. Rogers, MedPhys 30, 179 (2003) ** L. Buckley and D. W. O. Rogers, MedPhys 33, 455 (2006)

K. Zink and J. Wulff, PMB 53, 1595 (2008)

Page 12: Primary water calorimetry in clinical electron beams ... water calorimetry in clinical electron beams, scanned proton beams, and 192Ir brachytherapy dosimetry Ir brachytherapy dosimetry

192Ir brachytherapy

• SK=21000-38000 U• dsrc-det=25-70 mm

Page 13: Primary water calorimetry in clinical electron beams ... water calorimetry in clinical electron beams, scanned proton beams, and 192Ir brachytherapy dosimetry Ir brachytherapy dosimetry

192Ir brachytherapy

A. Sarfehnia and J. Seuntjens, MedPhys 37, 1914 (2010)

Page 14: Primary water calorimetry in clinical electron beams ... water calorimetry in clinical electron beams, scanned proton beams, and 192Ir brachytherapy dosimetry Ir brachytherapy dosimetry

192Ir brachytherapy

MethodRelative Diff from Water Calorimetry

Chamber -0.83%

Radiochormic film 0.83%

TG-43 0.55%

Page 15: Primary water calorimetry in clinical electron beams ... water calorimetry in clinical electron beams, scanned proton beams, and 192Ir brachytherapy dosimetry Ir brachytherapy dosimetry

Proton therapy

• Scanned beam consisted of superposition of 15 Bragg peaks.

• Minimized the dose variation in a 1 cm range around thermistors < 0.4% and 0.2% in scattered and scanned beams.

Page 16: Primary water calorimetry in clinical electron beams ... water calorimetry in clinical electron beams, scanned proton beams, and 192Ir brachytherapy dosimetry Ir brachytherapy dosimetry

Proton therapy

A. Sarfehnia et al, MedPhys 37, 3541 (2010)

Page 17: Primary water calorimetry in clinical electron beams ... water calorimetry in clinical electron beams, scanned proton beams, and 192Ir brachytherapy dosimetry Ir brachytherapy dosimetry

Proton therapy

% diff Calorimetry &T1 Chamber (IAEA TRS-

398)

Scattering 0.34%

Scanning 0.42%

Page 18: Primary water calorimetry in clinical electron beams ... water calorimetry in clinical electron beams, scanned proton beams, and 192Ir brachytherapy dosimetry Ir brachytherapy dosimetry

Conclusions

• Water calorimetry is feasible in high energy electron beams, scattered and scanned proton beams, as well as in Ir- 192 HDR brachytherapy.

• This work paves the way toward establishment of Dw primary standard for these beams at Standard lab level.

Page 19: Primary water calorimetry in clinical electron beams ... water calorimetry in clinical electron beams, scanned proton beams, and 192Ir brachytherapy dosimetry Ir brachytherapy dosimetry

CONCLUSIONSEnergy kht

1-sigmauncertainty (%)

6 MV 1.01 0.46

6 MeV 1.022 1.03

9MeV 1.011 0.52

12 MeV 1.010 0.49

16 MeV 1.010 0.51

20 MeV 1.012 0.52

Scattered (250 MEV) 0.996 0.38

Scanned (128-150 MeV) 0.953 0.64

Ir-192 0.963 1.90

Electron

Photon

Proton

Brachytherapy

Page 20: Primary water calorimetry in clinical electron beams ... water calorimetry in clinical electron beams, scanned proton beams, and 192Ir brachytherapy dosimetry Ir brachytherapy dosimetry

Conclusions

• Water calorimetry is feasible in high energy electron beams, scattered and scanned proton beams, as well as in Ir- 192 HDR brachytherapy.

• This work paves the way toward establishment of Dw primary standard for these beams at Standard lab level.

Page 21: Primary water calorimetry in clinical electron beams ... water calorimetry in clinical electron beams, scanned proton beams, and 192Ir brachytherapy dosimetry Ir brachytherapy dosimetry