TPS: algorithms ICTP SCHOOL ON MEDICAL PHYSICS Radiation Therapy: Dosimetry and Treatment Planning for Basic and Advanced Applications ICTP, Trieste 2019 Paweł Kukołowicz Medical Physics Department, Warsaw, Poland

Upload: others

Post on 10-Oct-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

TPS:algorithms

ICTP SCHOOL ON MEDICAL PHYSICSRadiation Therapy:

Dosimetry and Treatment Planningfor Basic and Advanced Applications

ICTP, Trieste 2019

Paweł KukołowiczMedical Physics Department, Warsaw, Poland

Page 2: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Paweł Kukołowicz

• Head of Medical Physics Department at Cancer-Center Instituteof Oncology in Warsaw, Poland

• Medical physicist from 1980

• Dosimetry, treatment planning, education

[email protected]

• I our department we have accredited SecondaryStandard Laboratory (calibration with Co60)

Page 3: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Delivered dose does matter!

2/#21

10 20 30 40 50NORMAL TISSUE DOSE (Gy)

0,0

0,1

0,2

0,3

0,4

0,5

0,6

0,7

0,8

0,9

1,0

0 20 40 60 80 100 120

TARGET DOSE (Gy)

PR

OB

AB

ILIT

Y O

F E

FF

EC

T

TCP

NTCP

10 20 30 40 50

Accuracy required and achievable in radiotherapy dosimetry:

Have modern technology and techniques changed our views?

Journal of Physics: conference Series 444 (2013)

David Thwaites

3.5% - 1 st. dev.

Page 4: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

How many patients will receivedose smaller of 7%?

Page 5: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

• 7% = 2 standard deviations

• MEAN ± 2 SD covers 95% of samples• 2.5% will receive dose larger of at least 7%

• 2.5% will receive dose smaller of at least 7%

• Consequences• TCP decreases of about 7% - 14%

• (based on normalizeddose gradient concept)

Page 6: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Clinical trials

5/#22

Peters L J et al. JCO 2010;28:2996-3001

Critical Impact of Radiotherapy Protocol Compliance and Qualityin the Treatment of Advanced Head and Neck Cancer: Results From TROG 02.02

Page 7: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Delivered dose does matter!

6/#22

Treatment deliveryTreatment planning

Quality control of

Treatment Planning

Systems

Pre-treatment imaging

Tumour & OAR Outlining

Courtesy Liz Miles RTTQA

Input

dosimetry dataCallibration of

dosimeters

Measurements of dose

distributions

of therapeutic beams

:

Page 8: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Treatment planning system

• Accuracy of dose distribution calculation

Quality control of

Treatment Planning

Systems

All our clinical decisions are made based

on the treatment plan prepared with a TPS!!!

Page 9: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

What are characteristics of a good TPS?

• High accuracy of dose distribution calculations

• Fast calculations

• Able to prepare plans for all contemporary techniques

• User friendly

• Robust

Varian - Eclipse

Elekta - Monaco

RaySearch – RayStation

Pinnacle

Page 10: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Initial remarks

• In general• what you see is what you get

• You must use the TPS you have!

• Everything you may do is• get to know deeply the system

• to read carefully manual,

• to read papers on quality control issues,

• to have contact with other, more experienced users,

• to learn about the system limitations!!!

Page 11: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

How to build a good model?

Page 12: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

What are characteristics of a good TPS?

• High accuracy of dose distribution calculations

• Fast calculations

• User friendly

• Robust

Algorithms implemented in TPS

Page 13: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Step 1 - exposure

• What radiation reachesthe absorber• fluence and energy fluence

• spectrum of energy fluence

• We call it: primary radiation

Page 14: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Step 1 - exposure• Fluence – F [1/m2]

• the number dN of particles (photons) incident ona sphere of a cross-sectional area of da

• Energy fluence – ψ [J/m2]• the energy dE incident on

a sphere of cross-sectional area of da

F E

da

dNF

Page 15: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Energy spectrum

• Depends on• effective accelerating potential

• target material

• flattening filter material and construction• there are flattenning filter free accelerators

• head (colimator system) material and construction

Page 16: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Energy spectrum

6 MV 15 MV

Page 17: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Energy spectrum calculations

• Reconstruction of spectra by iterative least squaresfitting of narrow beam transmission• it requires very precise measurements of attenuation factors

• Monte Carlo• precise knowledge of the treatment head design

• now this information is usually available

• Fiting routine• a given spectrum is used to calculate PDDs (using a database of

Monte Carlo generated Kernels) and compared with the measuredones• procedure is repeated until expected compliance is obtained

Page 18: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Step 2 – Energy deposition

• Primary and secondary dose

• Primary dose• interaction of primary photon

• energy transfered to chargedparticle (mostly to electron)

• electron transfered its energy to medium

• Secondary dose• interactions of secondary photons (scattered) and so on

Page 19: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Primary and secondary dosePrecise modeling of primary dose is the most important!

Sontag, Med. Phys. 1995, 22 (6)

Page 20: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Energy transfered from photonsto electrons Kerma

Φ=ΔN/ ΔA

Δz

ΔA

r

F

trE

r

F

r

number of interactions

per unit mass

energy transferred

to electrons

Page 21: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Energy transferred to electrons

• KERMA• Kinetic Energy Released per unit mass

F

trE

r

Page 22: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Charged particle equilibrium (CPE)

Photon interaction

Electron enters Dm

Electron leavs Dm

Charged particle equilibrium exists

for the volume V

if each charged particle of a given type

and energy leaving V is replaced

by an identical particle

of the same energy entering

E1,in

E2,in

E1,out

Dm

Etr,3

Etr,2

Etr,1

Etr,1

E2,out

Page 23: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Kerma Collision versus Absorbed Dose

• If CPD exists

Absorbed Dose = Kermacol

Kermacol=Kerma·(1-g)

g – fraction of energy emmited in the form of Bremstrahlung

Page 24: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

CPD never exists• Transient CPD exists

rela

tive e

nerg

y p

er

un

it m

ass

D

Kcolzmax

depth in medium

< 1 > 1

= 1

D=·Kcol Absorbed dose is equall to

Kerma at a little smaller depth.

D=(1+fTCPE)·Kcol

Page 25: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

rela

tive

ener

gy p

er u

nit m

ass

D

Kcolzmax

depth in medium

< 1 > 1

= 1

Is the CPE at dmax?

Page 26: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

How does the fluencedependence on the distance?

Page 27: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Fluence in air – inverse square low

( 2

2

fF

FF

air

fF

air+

FF +

isocenterplane

F

F

airF

isocenterplane

F

fF

air

+F

Page 28: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Uncertainty – errors in distance

• Error in distance of 1 cm approximately leads to:• 1% error

• 2% error

• 3% error

Page 29: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Uncertainty – errors in distance

• Distance to isocenter = 100 cm• D = 992/1002 ≈ 0.98

• 2%

• Precise calibration of telemeter does matter!

Page 30: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Fluence in water – dose in water

( dF

air

fF

water efF

F + +

FF

2

2

F

fF

water

+F

fd

( )1(

2

2

gEefF

FD tr

dF

air

fF

water

+F +

r

Primary dose – dose deposited by electrons

Page 31: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Uncertainty – errors in depth

• Error in depth of 1 cm approximately leads to:• 1% error

• 3% error

• 5% error

Page 32: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Difference of 1 cm of soft tissue• ≈ 0.05 1/cm

• D ≈ exp(-0.05) ≈ 0,95

• 5%

Page 33: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Fluence – real situation

• Radiological depth

• In general

water

rad

hhhd

r

rrr 332211 ++

kwaterkrad hd rr

( )1(

2

2

' gEehF

FD tr

h

k

F

air

hF

Q

kkk

+F +

r

h2

h3

h1

Q’

r2

r3

r1

F

Page 34: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

( )1(

2

2

' gEehF

FD tr

h

k

F

air

hF

Q

kkk

+F +

r

radiologicaldepth

physicaldistance

Page 35: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Another aproach to dose distribution calculation• Total energy released per unit mass

(

+F

r

hdF

airh hefF

FTERMA h

2

2

• What will happen with this released energy?• mostly it will be absorbed as primary and secondary dose

• only a little energy will escape (scattered photons, bremstrahlug)

primary energy fluence

Page 36: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Convolution – monoenergetic caseTERMAh = Th

TERMA

Med.Phys. Papanikolau 1993,5,1327-1336.

Convolution kernel representing

the relative energy deposited

per unit volume for photons

of energy hv;

integral over whole medium

( ( '3''),( rdrrArThrD hvhv

( 'rrAhv r

r

'r

Page 37: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Convolution – polyenergetic(real case)• Integral over space and energy spectrum

( ( dhvrdrrAdhv

rdThrD hv

hv '3''

),(

Mohan, Med.Phys, 1985, 12, 592 – 597.

Page 38: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Kernels – Point Spread FunctionAnders A. Ahnsjö, Med.Phys. 16 (4), 1989

• parameteres generated for beams of spectrum typical for 4Mv, 6MV, 10MV, and 15 MV

• Ө angle with respect to the direction of impinging primary photon

• w – stands for water

2/))exp()exp((),( rrbBraArhw +

bBaA ,,,

primary scattered

Page 39: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Kernels

Energy imparted per cm-3

0,4 MeV

1,25 MeV 10 MeV

The dash-dotted line first scatter terma,

calculated using the Klein-Nishina

cross sections and neglecting other process

than the Compton interaction.

Acta Oncologica, 1987, Ahnesjo

Page 40: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Kernels

Energy imparted per cm-3

0,4 MeV

1,25 MeV 10 MeV

The dash-dotted line first scatter terma,

calculated using the Klein-Nishina

cross sections and neglecting other process

than the Compton interaction.

Acta Oncologica, 1987, Ahnesjo

Page 41: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Convolution – polyenergetic(real case)• Integral over space and energy spectrum

( ( dhvrdrrAdhv

rdThrD hv

hv '3''

),(

Mohan, Med.Phys, 1985, 12, 592 – 597.

(

+F

r

hdF

airh hefF

FTERMA h

2

2

2/))exp()exp((),( ttbBtaAthw +

( 'rrAhv

Page 42: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Approximations

• To allow calculations in a resonable time several approximations are used• treatment planning system dependent

• the same model different results

• polyenergetic monoenergetic (e.g. for mean energy)

• single energy spectrum is used

• collapse cone method

• Kernels generated for water only• scaling with density

Page 43: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

6 MV spectrum

0

0.02

0.04

0.06

0.08

0.1

0.12

0.14

0 1 2 3 4 5 6 7

Spectrum 6 MVMean Energy 1.48 MeV

polyenergetic : monoenergetic

Page 44: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Changes of spectrumlateral softenning

single energy spectrum

Page 45: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Changes of spectrumlateral softenning

higher energylower energy

single energy spectrum

Page 46: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Collapsed Cone Convolutionspeed-up calculations

• 30 x 30 x 30 cm3 water Phantom

• 0.3 cm grid size

• 100 x 100 x 100 calulations point = 1 000 000

• Convolution: contribution from each voxel to each voxel

1 000 000 x 1000 000 = 1 000 000 000 000

Page 47: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Collapsed Cone Convolution

• CCC aproaches assumes that all the energy scattered from one voxel into small coneis absorbed along the line forming the axis of the cone

Page 48: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Collapsed Cone Convolution2D illustration

8 cones

Energy is absorber in blue pixels only.

Energy desposition

decreases very quickly

with distance

Page 49: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Collapsed Cone Convolution2D illustration

• According to Mackie (Teletherapy: Present and Future, Advanced medical Publishing, 1996)• 100 collapsed cones is enough

• Mobius3D – 144 collapsed cones

• Pinnacle – 80 collapsed cones

Page 50: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Approximation

• Scaling depth (distance) with density

h2

h3

h1

Q’

r2

r1

r3

( ( '3''),( rdrrArThrD radhvhv

𝑑𝑟𝑎𝑑 =

𝑘

𝜌𝑘/𝜌𝑤 ⋅ ℎ𝑘

Page 51: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Two spaces -worlds

• World 1• Air – fluence is scaled with distance according

to square factor• Precise

• World 2• Medium – Terma and Kernel are scaled with radiological

distance (density is taken into account)• approximation

Page 52: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Convolution – monoenergetic caseTERMAh = Th

TERMA

r

( ( '3''),( rdrrArThrD radhvhv

r

'r

Square factorand scaling with

density

Page 53: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Summary• Be especially careful with

• Measurements of• small beams

• OF

• Extrapolations• range of data entered into the system

• range of CT made for your patient

• Distributions in the vinicityof two much different matrials (Air – soft tissue)

Page 54: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Summary

• Primary and secondary dose

• Kerma and Kolision Kerma versus Dose

• How to described Kerma by photon fluence

Page 55: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Summary

• It is relatively easy to calculate the dose if• Transient CPE exist

• distance

• radiological depth

• If there is no CPE situations becoms much more difficult• transport of electrons must be considered

• interface of two dosimetrically different absorbers• air-soft tissue, lung-soft tissue, bone-soft tissue

Page 56: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Summary• TCP exist

• Primary dose is at least 80% of total dose• accuracy depends on primary dose calculations

• scale fluence with inverse square factor

• depth scaled with density

• first scatter is much larger than second, third etc.

2/))exp()exp((),( ttbBtaAthw +

primary scattered

Page 57: Prezentacja programu PowerPointindico.ictp.it/event/8651/session/0/contribution/17/... · 2019. 4. 16. · Anders A. Ahnsjö, Med.Phys. 16 (4), 1989 •parameteres generated for beams

Thank you very much for your attention!