POST ABORTAL CARE
(Manual Vacuum Aspiration)
by
M.Kamphinda-BandaUniversity of Malawi
Kamuzu College of NursingMedical Surgical Nursing Department
11 May 2011
![Page 2: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/2.jpg)
Learning Outcomes
• Define Manual Vacuum Aspiration• State the advantages of MVA• State the contraindications for MVA• List the instruments and supplies required for
MVA• Describe the procedure of MVA• Demonstrate post- procedural care following
MVA• List complications of MVA
![Page 3: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/3.jpg)
Definition
A procedure used for uterine evacuation and endometrial sampling consisting of a hand held vacuum syringe and flexible plastic cannullae.
Also referred to as “A Technology that saves women's lives.”
![Page 4: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/4.jpg)
Advantages of MVA
Can be used in outpatient settings, thus increasing women's access to care
A quick and simple method of removing retained products of conception
A safe method because it has rare minimal complications such as severe bleeding, pelvic infection, cervical injury and uterine perforation
![Page 5: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/5.jpg)
Advantages of MVA Cont’
Cheap - does not require electricity, specialized equipment, uses milder levels of medication than dilatation and curettage
Effective for endometrial biopsy because it yields high samples
Increases access to abortion care because it can be performed at lower level health care facilities
![Page 6: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/6.jpg)
Advantages of MVA Cont’
Conserves resources since MVA reduces cost of treating incomplete abortion by decreasing the woman's hospital stay.
![Page 7: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/7.jpg)
Contraindications
• Serious complications resulting from unsafe abortion e.g. shock, uterine perforation or sepsis- should be identified and treated before uterine evacuation is attempted
• Large fibroids
• History of bleeding disorder and severe anemia- MVA should be used with caution in health facilities with full emergency backup
![Page 8: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/8.jpg)
MVA Equipment and Uses
• Single and double valve syringes - used for treatment of incomplete abortion up to 8 weeks from the last menstrual period (LMP) confirmed by bimanual examination
• The double valve syringe - used for MVA in the treatment of incomplete abortion up to 12 weeks LMP
• The single valve syringes are also appropriate for use in obtaining an endometrial biopsy
![Page 9: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/9.jpg)
Preparing MVA Equipment
1. Select cannula according to the assessment of uterine size (weeks LMP)
For treatment of incomplete abortion the cannula should be large enough to allow passage of tissue expected (according to gestation) and to fit snugly through the cervix
Select the largest cannula you can comfortably use
![Page 10: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/10.jpg)
Uterine size VS Cannula size
Uterine size Cannula size
5-6 LMP 5-6 mm
7-8 LMP 7 mm
9-10 LMP 7-10 mm
10-12 LMP 9-12 mm
![Page 11: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/11.jpg)
Cannula size, Adaptor color, Syringe type
Cannula size adaptor color syringe type4,5,6mm No adaptor needed Single6 mm Blue Double7 mm Tan Double8 mm Ivory Double9 mm Dark brown Double10mm Dark green Double12mm No adapter needed Double
![Page 12: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/12.jpg)
Cannula size, Adaptor color, Syringe type Cont’
2. Inspect cannula for cracks or defects. Discard if there are any visible signs of weakness or wear
3. Select syringes and adaptors (if needed) according to the above chart. Useful to prepare two syringes in case one is worn or defective.
Note that the colored dots on the cannulae match the color of the appropriate adapter.
![Page 13: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/13.jpg)
Cannula size, Adaptor color, Syringe type Cont’
4. Inspect syringes. The syringe must be able to hold a vacuum. Discard syringes with any visible cracks or defects, or ones that do not hold a vacuum
5. Attach the adaptor (if required) to the end of the syringe or cannula.
![Page 14: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/14.jpg)
Cannula size, Adaptor color, Syringe type Cont’
6. Check plunger and valve. The plunger should be positioned all the way into the barrel, and the pinch valve should be open with the valve button out
7. Close the pinch valve by pushing the button down and forward until you hear a lock into place.
![Page 15: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/15.jpg)
Cannula size, Adaptor color, Syringe type Cont’
8. prepare the syringe by grasping the barrel and pulling back on the plunger until the arms of the plunger snap outward
9. plunger arms must be fully secured over the edge of the barrel, so the plunger cannot move forward accidentally.
Incorrect positioning of the arms could allow them to slip back inside the barrel, possibly pushing air or the contents into the uterus
![Page 16: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/16.jpg)
Cannula size, Adaptor color, Syringe type Cont’
Never grasp the syringe by the plunger arms
10.Check the syringe for vacuum tightness before use by preparing the syringe as in steps 3-7
Set the arms and leave the syringe that way for several minutes
Then open the pinch valve
You should hear a rush of air into syringe. Indicating there was a vacuum in the syringe
![Page 17: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/17.jpg)
Cannula size, Adaptor color, Syringe type Cont
Where possible, strain and rinse the aspirated tissue to remove blood and clots before placing in a clear receptacle of clean water, saline solution or weak vinegar
Inspect the tissue (villi, fetal membranes, endometrial tissue and fetal parts (after 9 weeks LMP)
![Page 18: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/18.jpg)
Monitor Recovery
In uncomplicated case:Take vital sings while woman is still on the
treatment tableAllow woman to rest comfortably
Check the bleeding at least once before discharge
Check that clamping has diminished. Prolonged clamping is abnormal
![Page 19: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/19.jpg)
Monitor Recovery cont’
If the woman is Rh negative, administer Rh immune globulin
Before the woman leaves the facility she should be provided information about:
Family planning services including counseling and initiation of a contraceptive method, if she wants to use one.
![Page 20: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/20.jpg)
Monitor Recovery Cont’
Signs of normal recovery (some uterine cramping over the next few days, some spotting or bleeding that shouldn’t be more than a normal period)
A normal menstrual period will occur within 4- 8 weeks.
Symptoms that require emergency care (prolonged cramping lasting more than a few days, prolonged bleeding more than 2 weeks)
![Page 21: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/21.jpg)
Monitor Recovery Cont’Bleeding more than a normal period, severe or
increased pain, fever, chills or malaise, fainting)
Take any prescribed medication
Not having sexual intercourse until 3 days after bleeding stops
Many other reproductive health needs may be addressed at this time as well like screening for cervical cancer,
![Page 22: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/22.jpg)
Monitor Recovery Cont’
Identification and treatment of reproductive tract infections
Patient may be discharged as soon as she is stable, can walk without assistance and has received follow up information.
![Page 23: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/23.jpg)
TECHNICAL COMPLICATIONS
Vacuum is lost
In most MVA procedures, the syringe vacuum remains constant until the syringe is approximately 80% full (50cc)
Loss of vacuum during the procedure can be caused by the syringe becoming full or the cannula being accidentally withdrawn from the cervical os
![Page 24: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/24.jpg)
What to do when vacuum is lost
Close the valves
Disconnect the syringe from the cannula. Leave the cannula tip in place and do not push in the plunger
Empty the syringe contents into the receiver by openning the pinch valve and pushing the plunger into barrel
Re establish a vacuum by closing the valves and pulling back and locking the arms of the plunger
![Page 25: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/25.jpg)
Cannula clogged
If no tissue or bubbles are flowing into the syringe , the cannula may be clogged
Remove the syringe and cannula
Examine the tip (openings of the cannula. If there is anything blocking these openings, remove the material using a sterile forcep or gauze sponge
![Page 26: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/26.jpg)
Cannula clogged
Never try to unclog the cannula by pushing the plunger back into the barrel when the cannula is in the woman's body.
![Page 27: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/27.jpg)
PROPERTIESAllow user to leave interaction: After viewing all the stepsShow ‘Next Slide’ Button: Show upon completionCompletion Button Label: Next Slide
![Page 28: POST ABORTAL CARE (Manual Vacuum Aspiration) by M.Kamphinda-Banda University of Malawi Kamuzu College of Nursing Medical Surgical Nursing Department 11](https://reader036.vdocuments.mx/reader036/viewer/2022062519/5697bf921a28abf838c8ed9f/html5/thumbnails/28.jpg)
Complications Related to MVA ProcedureE
Incomplete evacuation
Uterine or cervical perforation
Pelvic infection
Hemorrhage
Air embolism
Anesthetic reaction