-13- Genera l Photon ics Corporat ion 909.590.5473 www.genera lphoton ics .com
Specifications:
Emulation Instruments for System and Network Characterization
Polarization Synthesizer/Analyzer – PolaFlex™
ThePSY-201 isadeterministicpolarizationcontroller thatcangenerateandmaintainanystateofpolarization (SOP), regardlessof the inputSOP. It combinesGeneralPhotonics’ patented polarization controller, in-line polarimeter, and control algorithm into an instrument that functions as both a polarization state generator and a polarization analyzer.ThegeneratedSOPandthecorrespondingPoincaréSphererepresentationcanbedisplayedonacomputerscreenviaUSBinterface.TheoutputSOPcanbespecifiedbyinputtingStokesparametersusingthefrontpanelkeypadorbymanuallytuningtheSOPtoreachaspecificpointonthePoincarésphereortoreachanoptimumvalueofapolarization-dependentmetric.OnceadesiredoutputSOP is found, the instrumentcanautomaticallymaintain thisSOPagainst inputSOPfluctuations.Anotherattractivefeatureisthattheusercangenerateanyof6distinctSOPs(0°, 90°, ± 45°,RHCandLHC) forMuellermatrixcalculations,orselectanyof the6statesat the touchofabutton.Furthermore, the instrumentcangenerateseveralpreprogrammedSOPtraces
thatemulatecertaincommonpolarizationvariations.The instrumentcanalso functionasapolarizationscrambler,generatingSOPscanswithuser-definedpatternandspeed.Finally,withtheinternalpolarizationcontrollerdisabled,PolaFlex™canfunctionasanin-linepolarimeter,displayingtheinstantaneousSOPandDOPoftheinputlightbeam.ThePSY-201replacesthepopularPSY-101,withimprovedaccuracy,speed,andfunctionalities.Newfeaturesincludelong-termSOPmonitoring,SOPmarkersforanglemeasurement,anda"SOPreplay"functioninspheredisplaymode,aswellasextendedtriggeringcapabilityinoscilloscopemode.Itputsallofthetools necessary for polarization management at your fingertips.
Note:Lossspecificationsarereferencedwithoutconnectors.Unlessotherwisenoted,specificationslistedintableapplyforstandard1480-1620nmor1280-1340nmoperationat23±5°C,atpowerlevels>-25dBm.1.Forinputpower>-10dBm.Atlowerpowerlevels,bandwidthmaychangeduetoautomaticgaincontrol.
Features:.4MHzSOPsamplingrate.1MHzanalogbandwidth.45dBinputpowerdynamic range.Real-timePoincaréSphere display.High-speedSOPgeneration and tracking.HighspeedanalogoutputofSOP&DOP
Applications:.Receiverpolarizationsensitivity analysis.SystemSOP/DOPmonitoring.PERmeasurement. Polarization generation and stabilization.Sensorsystemcharacterization. 100G system polarization characterization
OperatingWavelengthRange 1480to1620nmor1280to1340nm
SamplingRate(max.) 4.0MSOPsamples/s
AnalogBandwidth1 1MHz
SOPSettlingTime 1msatstableinputSOP
SOPStability(InputPower>-25dBm,DOP>95%)
0.1°withstableinputSOP0.5°withinputSOPvariation<2π/s2°withinputSOPvariation<10π/s
SOPMeasurement/GenerationUncertainty ±0.25°afterusercalibration,withinput>-25dBm
DOPUncertainty±2%usingbuilt-incalibration,withinput>-25dBm±0.5%afterusercalibration,withinput>-25dBm
InputStokesParameterResolution 0.001
OpticalPowerUncertainty ±0.25dB
InsertionLoss 1.6dBmax.atcenterwavelength
ReturnLoss55dB(APCconnector),45dB(PCconnector)
PDL <0.25dB
PMD < 0.1 ps
OperatingPowerRange -35dBmto+10dBm
OpticalPowerDamageThreshold 300mW
OperatingTemperature 0°Cto40°C
StorageTemperature -20°Cto60°C
FrontPanelDisplay GraphicOLED
CommunicationInterfacesHighSpeedUSB2.0(30MB/sdatarate)forPolaViewsoftwareRS-232,Ethernet,GPIB
Analog Output0 to 5 V max range, user configurableMonitorvoltageforDOP,S1,S2,S3,powerordREF
PowerSupply 100–240VAC,50–60Hz
Software PolaViewTM(included)
Dimensions 2U,19”halfrackwidth14"(L)x8.5"(W)x3.5"(H)
INST
RUME
NTS
Genera l Photon ics Corporat ion 909.590.5473 www.genera lphoton ics .com
MOD
ULES
OCT P
RODU
CTS
SPEC
IAL P
OLAR
IZAT
ION
COM
PONE
NTS
ACCE
SSOR
IES
APPL
ICAT
ION
GUID
EFA
QSIN
STRU
MEN
TS
-14-
TypicalPerformanceData:Polarization stabilization
Figure1. Inputpolarizationpattern: trianglewavescrambleat1Hz,takenover20sec
Specialpolarizationstate/tracegeneration
Figure3.Poincaréspherepolestategeneration
Figure2.OutputpolarizationstabilizedbyPSY-201 against the same polarization-scrambledinput,takenover20sec
Figure4.TraceScans
Scrambling
Figure5.Trianglescramblingtrace,1Hzafter 1 minute
Figure6.Discretescrambling,100Hzafter1 minute
Emulation Instruments for System and Network Characterization
Polarization Synthesizer/Analyzer – PolaFlex™
Accessories:NoTailTM Polarizer p. 90NoTailTMIsolator p.91RackMountKitp.83
OrderingInformation:
PSY—201——
Wavelength:15=1550nm13=1310nm
ConnectorType:FC/PC,FC/APCSC/PC,SC/APC
ApplicationExample:CoherentReceiverPolarizationSensitivityTest
Samplesetupforacoherentreceiverperformancetestusingapolarizationstabilizer(POS-202)andapolarizationsynthesizer(PSY-201).
1.Usepolarization stabilizer (POS-202) to lock thepolarizationofonereceiverinput(localoscillatorinput).2.Usepolarizationsynthesizer(PSY-201)tocontrolthepolarizationoftheotherreceiverinputtofindtheSOPthat maximizes the receiver power reading.3.Lock thePSY-201outputat thatSOP toeliminatepolarization fluctuations in theSMfiber.Test receiverperformance.4.UsePSY-201 to findor switch to theorthogonalSOP(minimizereceiverpowerreading).5. Lock PSY-201 output at that SOP to eliminatepolarizationfluctuations.Testreceiverperformance.