For Medical Education Use Only 1 Point Of Care Cardiac Marker Testing Emergency Medicine View William E. Maloney, MD, FACEP Chair, Dept of Emergency Medicine Advocate Condell Medical Center Libertyville, IL December 15, 2011

Upload: others

Post on 03-Oct-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

For Medical Education Use Only 1

Point Of Care

Cardiac Marker Testing Emergency Medicine View

William E. Maloney, MD, FACEP Chair, Dept of Emergency Medicine Advocate Condell Medical Center

Libertyville, IL December 15, 2011

Page 2: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

2 For Medical Education Use Only

Learning Objectives

Overview of ED cardiac marker POC testing

Some of the key studies

Recent ACC/AHA & ACEP guidelines

Strategies for success

Page 3: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

3

Challenges in Chest Pain Triage Over admissions and Missed MIs

US statistics

Rx/PTCA/Other

CCU

CCU/CPEC/Other Treat or discharge

Page 4: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

4 For Medical Education Use Only

Issues in Diagnosis of NSTEMI

1. NSTEMI is a time dependent disease 2. Troponin values are the critical diagnostic test 3. Troponin does not elevate until >6 hours post AMI onset

CRITICAL CHALLENGE: Can earlier markers of necrosis predict

subsequent elevations in troponin?

Page 5: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

5 For Medical Education Use Only

0 10 20 30 40 50 60 70 80

0 2 4 6 8 12 18 24 32 48 72 Hours After Onset of MI

CKMB

Myoglobin

TnI

Appearance of Three Markers of Myonecrosis

Page 6: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

6 For Medical Education Use Only

The Value of Point-of-Care Cardiac Marker Testing

Class I Recommendation:

Institutions that cannot consistently deliver cardiac marker TAT’s of < 1 hour should implement POC testing devices.

Session I, page 1105

Wu et al. Clinical Chemistry 1999;45:1104-1121.

Page 7: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

For Medical Education Use Only 7

Ninety-Minute Exclusion of Acute Myocardial Infarction By Use of Quantitative Point-of-Care

Testing of Myoglobin and Troponin I

J. McCord; RM Nowak; PA McCullough; C Foreback; S Borzak; G Tokarski; MC

Tomlanovich; G Jacobsen; WD Weaver

Page 8: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

8 For Medical Education Use Only

Background & Methods

817 consecutive chest pain patients presenting to urban ED

Blood markers drawn in those with nondiagnostic ECG

65 patients (9%) were determined to have AMI

27 patients (36% of AMI patients had normal TnI at presentation)

Patients presenting with AMI are often missed in the critical

early hours.

McCord et al: AJC, Vol. 94. pp 864-867: October, 2004

Page 9: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

9 For Medical Education Use Only

90 minute AMI Exclusion Study

Hypothesis: Multimarker strategy (MMS) can exclude MI in

under 3 hours

Methods: 817 consecutive patients presenting to ED for

evaluation of possible ACS Point of care measurement of myoglobin, TnI, and

CK-MB done at presentation, 90 min, 3 and 9 hours Standard laboratory CK-MB done at same time Triage in ED done by physicians unaware of the

point of care results

McCord J et al. Circulation 2001;104:1483-88

Page 10: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

10 For Medical Education Use Only

Myoglobin in Early Hours of AMI (Prior to Troponin Release)

0

100

200

300

400

500

600

Baseline 90 Min 120 Min

Patient 1 Patient 2 Patient 3 Patient 4 Patient 5

All Patients Confirmed AMI via Subsequent Troponin positivity

McCord et al: AJC, Vol. 94. pp 864-867: October, 2004: Table 2 Adapted

Page 11: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

11 For Medical Education Use Only

0

50

100

150

200

250

300

350

400

ng/ml

0 HR 90 MINS 3 HR

AMI(- cTnI) Noncardiac

McCord et. al. Circulation 2001 ; 104:1483 - 1488

Value of Myoglobin with Troponin Negative at Presentation

AMI vs. Non-Cardiac Origin

Page 12: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

12 For Medical Education Use Only

Henry Ford Trial: Time to Laboratory Reporting

0

10

20

30

40

50

60

70

80

Time (min)

Central Lab Point of Care Device

71

24

McCord J et al. Circulation 2001;104:1483-88

Page 13: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

13 For Medical Education Use Only

The combination of Myo and cTnI at 0 and 90 mins had the highest early negative predictive value and sensitivity

The 3 hour blood sampling did not improve early sensitivity or negative predictive value

99.5% 99.6% NPV 96.9% 96.9% Sensitivity

0, 90 min, 3 hrs

0, 90 min

McCord et. al. Circulation 2001 ; 104:1483 - 1488

Results

Page 14: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

14 For Medical Education Use Only

2006 ACEP Guidelines re: Cardiac Biomarkers

(Level B Recommendations)

“..single negative CK-MB mass, Troponin I, or Troponin T measured 8 to 12 hours after…onset”

“..negative myoglobin in conjunction with a negative CKMB mass, or negative Troponin when measured at baseline and 90 minutes in patients presenting less than 8 hours after symptom onset”

“..negative 2-hour delta CK-MB mass…with a negative 2-hour delta Troponin in patients presenting less than 8 hours after symptom onset”

Fesmire FM et al. Ann Emerg Med 2006;48:270-301.

Page 15: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

15 For Medical Education Use Only

2006 ACEP Guidelines re: Cardiac Biomarkers

“No single serum marker used alone has sufficient sensitivity or specificity to reliably identify or exclude AMI within 6 hours after symptom onset.”

Fesmire FM et al. Ann Emerg Med 2006;48:270-301

Page 16: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

16 For Medical Education Use Only

2006 ACEP Guidelines Conclusions

Within the first 8 hours of symptoms: Multi-marker

After 8 hours: Single CK-MB or Troponin

Fesmire FM et al. Ann Emerg Med 2006;48:270-301

Page 17: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

17 For Medical Education Use Only

2007 ACC/AHA Guidelines (Regarding UA/Non-STEMI)

Class I - Cardiac biomarkers all patients

Troponin is preferred marker

“..exact timing of serum marker measurement should take into account the uncertainties often present with the exact timing of onset of pain and the sensitivity, precision, and institutional norms of the assay being utilized as well as the release kinetics of the marker being measured.”

Anderson et al. ACC/AHA 2007 Guidelines. Circulation. 2007:116:3148-e304.

Page 18: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

18 For Medical Education Use Only

2007 ACC/AHA Guidelines (Risk Stratification)

Class IIb ...”within 6 hours of onset…

an earlier marker (e.g., myoglobin) in conjunction with a late marker (e.g., troponin) ..

2-h delta CK-MB mass in conjunction with 2-h delta troponin..

myoglobin in conjunction with CK-MB mass or troponin when measured at baseline and 90 min” Anderson et al. ACC/AHA 2007 Guidelines. Circulation.

2007:116:3148-e304.

Page 19: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

19 For Medical Education Use Only

2007 ACC/AHA Guidelines (Risk Stratification)

Class IIb - “Measurement of B-type natriuretic peptide (BNP) or NT-pro-BNP may be considered to supplement assessment of global risk in patients with suspected ACS”

An isolated, early elevation [of myoglobin] in patients with non-dx ECG should not be relied on to make the diagnosis of MI

Anderson et al. ACC/AHA 2007 Guidelines. Circulation. 2007:116:3148-e304.

Page 20: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

A rapid point-of-care cardiac marker testing strategy facilitates the rapid diagnosis and management of chest

pain patients in the emergency department.

Straface AL, Meyers JH, Kirchick HJ, Blick KE. Am J Clin Pathol 2008;129:788-795.

Page 21: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

Data from three HCA hospitals collected February 9 through May 31, 2004

1, 2, and 3 hour marker intervals If any of the three criteria below were met, the patient was

considered positive for an MI: A TnI ≥ 0.4 ng/mL on any draw A ∆ myoglobin of ≥ 50% between sequential draws with any

detectable TnI by the last draw A ∆ myoglobin of ≥ 50% between sequential draws with a

concomitant increase in CK-MB without detectable TnI on any of the draws

Multimarker Algorithm

Straface AL et al (2008) Am. J. Clin. Path. (2008) 129:788-795

Page 22: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

Straface, et al. – Results

Serial draw multi-marker algorithm 99.9% negative predictive value 99.7% accuracy

+ - Total

+ 145 12 157

- 3 5041 5044Total 148 5053 5201

95% CISensitivity 98.0% 95.7% - 100.2%Specificity 99.8% 99.6% - 99.9%Accuracy 99.7% 99.6% - 99.9%

PPV 92.4% 88.2% - 96.5%NPV 99.9% 99.9% - 100.0%

DiagnosisA

lgor

ithm

Res

ults

Page 23: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

+ - Total

+ 144 252 396

- 4 4801 4805

Total 148 5053 5201

95% CISensitivity 97.3% 94.7% - 99.9%Specificity 95.0% 94.4% - 95.6%Accuracy 95.1% 94.5% - 95.7%

PPV 36.4% 31.6% - 41.1%NPV 99.9% 99.8% - 100.0%

0.05 ng/mL (~99th %tile)

Diagnosis

TnI R

esul

ts

+ - Total

+ 101 12 113

- 47 5041 5088

Total 148 5053 5201

95% CISensitivity 68.2% 60.7% - 75.7%Specificity 99.8% 99.6% - 99.9%Accuracy 98.9% 98.6% - 99.2%

PPV 89.4% 83.7% - 95.1%NPV 99.1% 98.8% - 99.3%

0.40 ng/mL (WHO ROC)

Diagnosis

TnI R

esul

ts

Straface, et al. – Results Serial draw Troponin only algorithm

Poor Sensitivity Too Many False Negatives

Poor Specificity Too Many False Positives

Page 24: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

Another Study with Similar Results

Hours Post Admission

0-6 6-12 12-24

% Sensitivity AMI cutoff: 0.30 ng/mL 60.0 78.6 91.7

99th Percentile: 0.012 ng/mL 97.1 96.4 95.8

% Specificity AMI cutoff: 0.30 ng/mL 95.4 94.6 96.5

99th Percentile: 0.012 ng/mL 59.2 56.8 54.9

http://www.abbottdiagnostics.com/pubs/2006/2006_AACC_Wisener_TnI1.pdfs

Page 25: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

Non-ACS/HF TnI Elevations Acute Disease

Cardiac and Vascular Acute aortic dissection Cerebrovascular accident

Ischemic stroke Intracerebral hemorrhage Subarachnoid hemorrhage

Respiratory Acute PE ARDS

Muscular Damage

Cardiac Inflammation Endocarditis Myocarditis Pericarditis

Infectious Sepsis Viral illness

Kelley, Januzzi and Christenson. Clin Chem 2009; 55:2098-2112

Page 26: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

Non-ACS/HF TnI Elevations Other Acute Disease

Kawasaki disease Apical ballooning

syndrome Thrombotic

thrombocytopenic purpura

Rhabdomyolysis Birth complications in

infants (low birth weight, preterm)

Acute complications of inherited disorders (neurofibromatosis, Duchenne muscular dystrophy, Klippel-Feil syndrome)

Environmental Exposure Carbon monoxide Hydrogen sulfide Colchicine

Kelley, Januzzi and Christenson. Clin Chem 2009; 55:2098-2112

Page 27: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

Non-ACS/HF TnI Elevations

Chronic disease ESRD Cardiac infiltrative

disorders Amyloidosis Sarcoidosis Hemochromatosis Scleroderma

Hypertension Diabetes Hyperthyroidism

Iatrogenic disease Invasive procedures

Heart transplant Congenital defect repair Lung resection ERCP RFCA

Noninvasive procedures Cardioversion Lithotripsy

Pharmacological sources Chemotherapy Other medications

Kelley, Januzzi and Christenson. Clin Chem 2009; 55:2098-2112

Myocardial Injury Blunt chest injury Endurance

athletes Envenomation

Snake Jellyfish Spider Centipede Scorpion

Page 28: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

• 99th Percentile Value - Apparently healthy population, results are typically close to zero • AMI Cutoff - Value that determines if a patient is having an AMI based on WHO criteria

AMI Cutoff 99th percentile

Normal Range for the Assay

NACB guidelines: ANY value above the 99th percentile is abnormal and is now defined as an MI if accompanied with

signs and symptoms of ischemia.

Serial draws are recommended to detect temporal rise and fall of troponin-I levels characteristic of MI, and should be used in conjunction with other information such as other cardiac markers, ECG, clinical symptoms, etc.

AMI per WHO criteria

So How Do I Best Use Troponin-I Results?

Page 29: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

Universal Definition of MI (Joint ESC/ACCF/AHA/WHF Task Force)

MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia

Rise and/or fall of cardiac biomarkers (preferably troponin) above the 99th percentile with evidence of ischemia Symptoms of ischemia ECG changes: ST-T, new LBBB or pathological Q waves Imaging evidence: loss of myocardium or wall motion abnormality

Sudden unexpected cardiac death involving cardiac arrest PCI patients: Cardiac marker elevations of 3 x 99th

percentile CABG patients: Cardiac marker elevations of 5 x 99th

percentile Thygesen. et al Eur. Heart J (2007) 28:2525-2538

Page 30: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

Clinical Application of Sensitive Troponin Assays

Morrow, NEJM 2009

Accuracy for AMI improved Increased sensitivity Reduced specificity and PPV Muscle injury versus ischemia (MI) Prognostic implications of low-level increases Additional studies looking at changes in troponin vs. short and long-term outcome

30 For Medical Education Use Only

Page 31: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

Recent Patient EKG #1 at 3:31pm

For Medical Education Use Only

Page 32: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

EKG #2 3:50pm

For Medical Education Use Only

Page 33: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

First Markers Time: arrival

For Medical Education Use Only

CKMB < 1.0 Myo 47.2 Trop < 0.05 BNP < 5

Page 34: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

EKG #3 3:54pm

For Medical Education Use Only

Page 35: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

Second Markers Time: 90 minutes

For Medical Education Use Only

CBMB 1.6 Myo 200 Trop < 0.05 BNP < 5

Page 36: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

EKG #4 5:24pm

For Medical Education Use Only

Page 37: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

EKG #5 6:10pm

For Medical Education Use Only

Page 38: Point Of Care Cardiac Marker Testing · 99.9% negative predictive value ... MI is: Myocardial necrosis in a clinical setting consistent with myocardial ischemia ... Thygesen. et al

Third Markers Time: 240 minutes

For Medical Education Use Only

CKMB 8.1 Myo 222 Trop 0.40 BNP < 5