pl-els 2100 ice. pl-els 2100 ice evaporation at sub ambient temperatures the pl-els 2100 ice...

18
PL-ELS 2100 Ice

Upload: kory-horton

Post on 05-Jan-2016

216 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: PL-ELS 2100 Ice. PL-ELS 2100 Ice Evaporation at Sub Ambient Temperatures The PL-ELS 2100 Ice comprises an Integrated Cooled Evaporator in combination

PL-ELS 2100 Ice

Page 2: PL-ELS 2100 Ice. PL-ELS 2100 Ice Evaporation at Sub Ambient Temperatures The PL-ELS 2100 Ice comprises an Integrated Cooled Evaporator in combination

PL-ELS 2100 Ice Evaporation at Sub Ambient Temperatures

The PL-ELS 2100 Ice comprises an Integrated Cooled Evaporator in combination with the ‘evaporation gas’ technology to provide sub-ambient operation

The ‘evaporation gas’ is increased to compensate for the increase in relative humidity at sub-ambient temperatures

This approaches enables the PL-ELS 2100 Ice to operate at sub-ambient temperatures and as low as 10°C.

Page 3: PL-ELS 2100 Ice. PL-ELS 2100 Ice Evaporation at Sub Ambient Temperatures The PL-ELS 2100 Ice comprises an Integrated Cooled Evaporator in combination

Even at ambient temperature the response of semi-volatile, low-molecular weight analytes can be poor.

So the PL-ELS 2100 Ice has been designed to operate at sub-ambient temperature to improve the detection of semi-volatile compounds

Introduction

Page 4: PL-ELS 2100 Ice. PL-ELS 2100 Ice Evaporation at Sub Ambient Temperatures The PL-ELS 2100 Ice comprises an Integrated Cooled Evaporator in combination

PL-ELS 2100 Ice Design Features

Nebulisation

• Same nebuliser as PL-ELS 2100

• Temperature Range: 25-90°C

Evaporation

• Integrated Cooled Evaporator Tube

• Temperature range: 10-80°C

• Active Heating/Cooling

• Same evaporation gas technology as PL-ELS 2100

Page 5: PL-ELS 2100 Ice. PL-ELS 2100 Ice Evaporation at Sub Ambient Temperatures The PL-ELS 2100 Ice comprises an Integrated Cooled Evaporator in combination

Peltier Cooled Evaporator Tube

Page 6: PL-ELS 2100 Ice. PL-ELS 2100 Ice Evaporation at Sub Ambient Temperatures The PL-ELS 2100 Ice comprises an Integrated Cooled Evaporator in combination

PL-ELS 2100 Ice Design Feature for Sub-Ambient Operation

Aqueous eluents can be removed at 15°C

Organic solvents (e.g. THF) can be evaporated as low as 10°C

The benefit of operating below room temperature provides maximum sensitivity for low-molecular weight semi-volatile compounds.

Page 7: PL-ELS 2100 Ice. PL-ELS 2100 Ice Evaporation at Sub Ambient Temperatures The PL-ELS 2100 Ice comprises an Integrated Cooled Evaporator in combination

PL-ELS 2100 Ice Design Feature for Sub-Ambient Operation

The PL-ELS 2100 Ice can heat or cool rapidly, which reduces equilibration time.

• 35°C to 10°C in 10 minutes

• 15°C to 80°C in 11 minutes

Accurate temperature control of evaporator temperature eliminates external temperature effects

Page 8: PL-ELS 2100 Ice. PL-ELS 2100 Ice Evaporation at Sub Ambient Temperatures The PL-ELS 2100 Ice comprises an Integrated Cooled Evaporator in combination

PL-ELS 2100 Ice Design Features

Detection

• Same Optical Bench as PL-ELS 2100

• Digital Output 24bit (10Hz) via RS232

• Faster Time Constant (0.1 Secs) for UPLC

Page 9: PL-ELS 2100 Ice. PL-ELS 2100 Ice Evaporation at Sub Ambient Temperatures The PL-ELS 2100 Ice comprises an Integrated Cooled Evaporator in combination

PL-ELS 2100 Ice

FOR VOLATILE SOLUTES

• Sub-ambient Evaporator temperatures required - 10-30°C

– PL-ELS 2100 Ice

• Gas flow optimised for sub-ambient temperatures and mobile phase

– 100% Water @15°C requires 2.2SLM of gas

• Compounds with high vapour pressures, low MW pharmaceuticals, pesticides, hydrocarbons

Page 10: PL-ELS 2100 Ice. PL-ELS 2100 Ice Evaporation at Sub Ambient Temperatures The PL-ELS 2100 Ice comprises an Integrated Cooled Evaporator in combination

Sub-Ambient Applications

The S/N ratio for thermally sensitive compounds such as Acetanilide is improved at sub-ambient temperatures

In addition, the response of compound’s with high boiling points and high volatility, such as Alkanes, is dramatically improved.

Hence, ELSD temperatures need to be as low as possible, to minimise volatility effects.

Page 11: PL-ELS 2100 Ice. PL-ELS 2100 Ice Evaporation at Sub Ambient Temperatures The PL-ELS 2100 Ice comprises an Integrated Cooled Evaporator in combination

Sub-Ambient Applications: Acetanilide (MW 135)

Page 12: PL-ELS 2100 Ice. PL-ELS 2100 Ice Evaporation at Sub Ambient Temperatures The PL-ELS 2100 Ice comprises an Integrated Cooled Evaporator in combination

Sub-Ambient Applications: Alkanes

Page 13: PL-ELS 2100 Ice. PL-ELS 2100 Ice Evaporation at Sub Ambient Temperatures The PL-ELS 2100 Ice comprises an Integrated Cooled Evaporator in combination

Additional Benefits of the PL-ELS 2100 Ice

Sub-ambient operation down to 10°C for maximum sensitivity to volatile compounds

Active heating/cooling provides rapid equilibration between analyses.

Improved accuracy of quantification.

Page 14: PL-ELS 2100 Ice. PL-ELS 2100 Ice Evaporation at Sub Ambient Temperatures The PL-ELS 2100 Ice comprises an Integrated Cooled Evaporator in combination

Limit of Detection- PEG 106

x40 increase in S/N

Page 15: PL-ELS 2100 Ice. PL-ELS 2100 Ice Evaporation at Sub Ambient Temperatures The PL-ELS 2100 Ice comprises an Integrated Cooled Evaporator in combination

What are you Missing?-Wine Analysis

Page 16: PL-ELS 2100 Ice. PL-ELS 2100 Ice Evaporation at Sub Ambient Temperatures The PL-ELS 2100 Ice comprises an Integrated Cooled Evaporator in combination

PL-ELS 2100 Ice ‘Lock out’ Specs

Sub-ambient operation down to 10°C

High sensitivity to semi-volatile compounds

DMSO transparency at ambient temperature

Extremely low dispersion for high resolution separations

Compatible with eluent flow rates, up to 5.0ml/min

Rapid thermal equilibration

Easy to use

Extremely small footprint, stackable

Page 17: PL-ELS 2100 Ice. PL-ELS 2100 Ice Evaporation at Sub Ambient Temperatures The PL-ELS 2100 Ice comprises an Integrated Cooled Evaporator in combination

ELS2100Ice is the best choice for ELS detection for volatile and semi-volatile compounds.

ELS2100Ice has the widest range of sample detection for any ELSD, therefore it is the most ‘universal’ ELS Detector

ELS2100Ice is the only ELSD with operation at sub-ambient temperatures.

ELS2100Ice sub ambient specification can be used for lock-out specs vs competitive ELSDetectors.

Key Points vs Competition

Page 18: PL-ELS 2100 Ice. PL-ELS 2100 Ice Evaporation at Sub Ambient Temperatures The PL-ELS 2100 Ice comprises an Integrated Cooled Evaporator in combination

PL-ELS 2100/2100 Ice Specifications

200mm

450mm

200mm

415mm

450mm

Light source: LED 480nm

Detector: Photomultiplier tube

Nebuliser temperature: ambient to 90°C in 1°C increments

Evaporator temperature: PL-ELS 2100 Ice - 10 to 80°C (1°C increments)

PL-ELS 2100 - ambient to 120°C (1°C increments)

Gas flow rate: up to 3.25 SLM

Gas pressure: typical 60-100psi, maximum 100psi

Eluent flow rate: 0.2 to 5 ml/min

Output: Analogue: 0-1V FSD

Digital: 24bit, 10Hz using serial connection

Communications: serial I/O (RS232), contact closure, TTL

Power requirements: 90/120V AC or 220/250V AC 50/60 Hz

Dimensions: 200x450x415mm (wxdxh)

Part No. PL0860-0110 (110v), PL0860-0240 (240v)