  • Physikalisch-Meteorologisches Observatorium Davos, World Radiation Center (PMOD/WRC) And is signatory of the Mutual Recognition Arrangement by the Committee of Weights and Measures (CIPM MRA). The PMOD/WRC is the World Radiation Center of the World Meteorological Organisation,
  • Slide 2
  • PMOD/WRC Absolute radiometry (ARS) Infrared radiometry (IRS) Optical depth WORCC UV radiometry (WCC-UV) World Radiation Center
  • Slide 3
  • EMRP-ENV03 Solar UV http://projects.pmodwrc.ch/env03/ o Enhance the reliability of spectral solar UV radiation measured at the Earth surface o Develop new techniques and devices for traceability better than 2% (now 5%) and for cost-effective array-spectroradiometer in UV monitoring networks o Intercomparison Campaigns and Workshops Primary Irradiance Standard Transfer StandardReference Spectroradiometer End-User Devices Calibrated UV Network o Project Coordination: Julian Grbner, Davos o Duration: 2011 - 2014 o Total Budget: 3.9 M o 8 Partners EU-NMI; 2 Industry; 2 Universities; > 3 Collaborators (Stakeholders)
  • Slide 4
  • Cryogenic radiometer 0.01% Spectral irradiance standard ~1-3% cw-Laser Sources Si-trap detector + aperture Filter Radiometer Blackbody + aperture ~0.3% Detector Source Spectroradiometer Spectrally tuneable source Spectroradiometer ~5% Traceability chain for spectral irradiance Wm -2 nm -1 Portable transfer standard (used for stability check) ~4% Grbner & Sperfeld, Metrologia, 2006
  • Slide 5
  • Validation of the QASUME irradiance reference at PTB Blackbody BB3200pg at PTB Grbner J., and P. Sperfeld, Direct traceability of the portable QASUME irradiance scale to the primary irradiance standard of the PTB, Metrologia, 42, 134139, 2005. QASUME Validation at PTB in 2004 and 2013 1.2%
  • Slide 6
  • Quality site audit with the QASUME reference spectroradiometer UV Quality Assurance Program Status 2002 - 2013 63 site visits 34 sites >150 spectroradiometer intercomparisons http://www.pmodwrc.ch/euvc/euvc.php?topic=qasume_audit Ny lesund, Spitzbergen, 2009 RBCC-E, 2005-2013 QASUME Grbner et al., Appl. Opt. 2005 Grbner et al., Metrologia, 2006
  • Slide 7
  • RBCC-E @ El Arenosillo The sunny place 1999, 2005, 2007, 2009, 2011, 2013 2005
  • Slide 8
  • Daily ratios to QASUME: The ideal case
  • Slide 9
  • Brewer cosine error
  • Slide 10
  • Slide 11
  • Single Brewer Stray light
  • Slide 12
  • Brewer Temperature dependency 6% Temp. coefficient -0.3 %/K
  • Slide 13
  • Summary from 6 th RBCC-E (2011)
  • Slide 14
  • Wavelength scale
  • Slide 15
  • What needs to be addressed? Temperature dependency Absolute calibration Wavelength dispersion Single Brewer Stray Light Angular Response Apply Correction, or better: Entrance optic modification
  • Slide 16
  • Presentations, Guidelines, Publications can be found at the project web-site: http://projects.pmodwrc.ch/env03/ Spectral solar UV Intercomparison at Davos & Final ENV03 Workshop, 7-16 July 2014 International Pyrheliometer Comparison (IPC) 2010
  • Slide 17
  • Quality site audit with the QASUME reference spectroradiometer UV Quality Assurance Program Status 2002 - 2013 63 site visits 34 sites >150 spectroradiometer intercomparisons http://www.pmodwrc.ch/euvc/euvc.php?topic=qasume_audit -1 14 % Relative Diff. UVB (95% coverage)
  • Slide 18
  • QASUME irradiance reference stability: 2004 to 2012
  • Slide 19
  • QASUME irradiance reference Based on 7 transfer standards directly calibrated against the primary irradiance normal of the Physikalisch-Technische Bundesanstalt (PTB), in Germany. Primary standard PTB Transfer standard 1000 W FEL lamp Transfer standard 1000 W FEL lamp Transfer standard 1000 W FEL lamp PTB EUVC/QASUME 250 W portable lamp Portable QASUME irradiance reference
  • Slide 20
  • Spectral UV irradiance site audit procedure PMOD/WRC Spectroradiometer Brewer#163 Portable QASUME reference spectroradiometer Regular intercomparisons Measuring Station Quality Assurance Local Spectroradiometer QASUME Audit report and calibration certificate sent to local operator and published on www.pmodwrc.ch/euvc/euvc.html QASUME irradiance reference traceable to PTB Test report
  • Slide 21
  • Comparison of spectroradiometers matSHIC QASUME TEST Ratio TEST/QASUME Same wavelength scale Uniform slit width (1 nm default)
  • Slide 22
  • Wavelength scale