pH Measurement
General Chapters:<791> pH
USP38
ศิริพร เหล่ามานะเจริญ
ส านกัยาและวตัถุเสพติด
กรมวิทยาศาสตร์การแพทย ์
27 พฤศจิกายน 2558
1
![Page 2: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/2.jpg)
2
Outline
● Introduction
● pH measurement system
● Instrument requirements
● Buffer solutions
● Calibration
● Operation
![Page 3: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/3.jpg)
3
Introduction
pH (in an aqueous solution)
→ a measure of hydrogen and hydroxide ions
At 25°C, water contains
- 1x10-7 mol/l of H+ and
- 1x10-7 mol/l of OH-
Kw = [H+][OH-] = 1x10-14
Kw is dissociation constant
![Page 4: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/4.jpg)
4
pH is often expressed as
pH = -log H+
Example:
Hydrogen ion concentration = 1x10-3 mol/l, pH = 3.00
If add base, change the hydrogen ion concentration
to 1x10-11 mol/l, pH = 11.0
![Page 5: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/5.jpg)
5
The Nernst equation:
𝐸 = 𝐸0 + 2.3𝑅𝑇
𝑛𝐹log 𝑎𝐻+
𝐸 = measured potential
𝐸0= constant, standard potential
R = gas constant
T = temperature in Kelvin
n = ionic charge
F = Faraday constant
![Page 6: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/6.jpg)
6
The Nernst equation:
𝐸 = 𝐸0 + 2.3𝑅𝑇
𝑛𝐹log 𝑎𝐻+
Y = a + b X
Slope is the Nernst factor
at 25°C, slope = -59.16 mV/pH unit
![Page 7: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/7.jpg)
7
Theoretically ideal
At 25°C, slope = -59.16 mV/pH and
0 mV at pH =7 (offset or zero point)
![Page 8: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/8.jpg)
8
Temperature influences
Isothermal point of electrode : intersection point
of calibration lines plotted for different temperatures
Ideal electrode:
Isothermal intersection point at zero point (0 mV at
pH 7) for different temperature
Real electrode:
Hardly ever coincides with the zero point of the electrode
![Page 9: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/9.jpg)
9
Different between Real and theoretical isothermal intersection point
↓ Measurement error
T2 > T1
![Page 10: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/10.jpg)
10
pH meter has to be adjusted to
compensate for any deviations due to
non-ideal behaviour of the electrode from
theoretical values by
“Calibration”
2 parameters : offset and slope
![Page 11: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/11.jpg)
11
Offset adjustment
of pH electrode Slope adjustment
of pH electrode
Real behavior Ideal behavior
![Page 12: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/12.jpg)
12
Automatic temperature compensation (ATC)
automatic calculation of pH from the
measured potential and temperature, using
an adjusted slope to the measured
temperature
![Page 13: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/13.jpg)
13
It is recommended to do a calibration
and sample measurement at the same
temperature.
The most accurate pH value
![Page 14: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/14.jpg)
14
pH measurement system
consists of
-measuring electrode
(potential directly proportional to pH)
-reference electrode
(potential independent of pH)
-voltage measurement system with
temperature device
![Page 15: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/15.jpg)
15
Instrument requirements
-capable of performing a 2-point pH calibration
-resolution of pH measurement system at least
0.01 pH
-capable of temperature compensation
(convert mV to pH units at any temperature),
either automatic temperature device or manual
entry
![Page 16: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/16.jpg)
16
-accuracy of temperature measurement
system ±1°C
-resolution of temperature measurement
System at least 0.1°C
-lab-based pH measurement performed at
25 ± 2°C
![Page 17: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/17.jpg)
17
Buffer solutions
-prepared as directed in Table2 by using
purified water
or
-used commercial buffer solutions which is
traceable to NIST or other national authorities,
labeled with a pH value accurate to 0.02 pH
unit
![Page 18: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/18.jpg)
18
![Page 19: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/19.jpg)
19
Buffer solutions
Buffer solutions greater than 11,
-containers are resistant to or reduce CO2
intrusion
-be prepared and used fresh
** CO2 would lower the pH of buffer **
![Page 20: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/20.jpg)
20
Buffer solutions
Buffer solutions lower than 11,
-be prepared at intervals not to exceed 3
months
![Page 21: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/21.jpg)
21
Temperature coefficient
the change in pH of a solution with temperature
(ΔpH/°C)
Ex. Buffer solution pH 8.00
(20°C)
Temperature [°C]
ΔpH
0 +0.015
5 +0.10
10 +0.07
15 +0.004
20 ±0
25 -0.04
30 -0.06
35 -0.08
40 -0.10
50 -0.15
![Page 22: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/22.jpg)
22
Calibration
Standardization
Calibration process Verification
1st buffer 2nd buffer 3rd buffer
cover the expected between calibration
of pH of the sample buffer
![Page 23: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/23.jpg)
23
Calibration process
rinse the electrode with water
rinse the electrode with 1st buffer
Immerse the electrode in 1st buffer
2- point calibrate according to instruction
![Page 24: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/24.jpg)
24
remove the electrode
rinse the electrode with water
rinse the electrode with 2nd buffer
Immerse the electrode in 2nd buffer
2- point calibrate according to instruction
![Page 25: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/25.jpg)
25
verify the slope and offset
Slope from 90-105%
Offset ±30 mV (0.5pH units at 25°C)
If not in acceptance limit,
cleaned, replenished, serviced or replaced
the electrode.
repeat the 2-point calibration
![Page 26: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/26.jpg)
26
Verification
rinse the electrode with water
rinse the electrode with 3rd buffer
Immerse the electrode in 3rd buffer
read the pH according to instruction
![Page 27: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/27.jpg)
27
verify the pH reading
pH reading within ±0.05 pH of the pH
value at the buffer solution temperature
![Page 28: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/28.jpg)
28
Operation
rinse the electrode with water
rinse the electrode with sample
Immerse the electrode in sample
read the pH and temperature
according to instruction
![Page 29: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/29.jpg)
29
-prepare the sample by using purified
water unless otherwise specified in the
monograph
-allow sufficient time for stabilization of
the temperature and pH measurement
![Page 30: pH Measurement General Chapters: pH USP38dmsc2.dmsc.moph.go.th/webroot/drug/km/lab_analysis/pH measurement USP3… · -capable of performing a 2-point pH calibration -resolution](https://reader030.vdocuments.mx/reader030/viewer/2022040317/5e3074a6b495bf7ad854af6c/html5/thumbnails/30.jpg)
30
Reference:
- USP 38
- A guide to pH Measurement- the Theory and
Practice of pH Applications, Metter Toledo