Qualitative Determination of Pesticides in a Roma Tomato,
Iceberg Lettuce, and a Macintosh Apple
Karen SandersChemistry 481
12/4/08
1
Varian Saturn 3800-2200 GC/MS
• Similar to the Varian Saturn 3800-2000 GC/MS used in this experiment.
2
Restek Pesticide Standard
Manufactured by Restek Company
Labeled, “Organochlorine Pesticide Mix AB#2”
Held in a solvent mixture of 50:50
Hexane/Toluene.
Varian Saturn 3800-2000 GC-MS Parameters for Restek Standard and Roma Tomato
Separation
Carrier Gas: Helium
Volume Injected: 1 mL
Flow Rate: 1.8 ml/min
Stationary Phase:
5% Phenyl-95% polydimethylsiloxane
Column Temp: 80 °C for 1 minute, increasing to 240 °C at 10 °C/min,
increasing to 265 °C at 10 °C/min, and holding for 10 min.
Oven Temp: 275°C
Injector Port Temp:
270 °C
Solvent Delay: 5 minutes
Column Size: 30 m x 0.25 mm ID, 0.25 um
Mode: Split-less
Restek Standard of Organochlorine
Pesticide Mix AB#2: Chromatogram
18/20 Pesticides Found
Roma Tomato
Kroger Brand
Originated from California, U.S.A.
Unwashed prior to analysis
Reagents
Florisil, also known as activated magnesium silicate had a 30-60
particle mesh size and was manufactured by Fluka Analytical.
CH2Cl2 was of the Aldrich Brand and Reagent Grade.
Na2SO4 and alumina were manufactured by Fisher Scientific and were both reagent grade.
Pesticide Extraction Procedure for the Roma Tomato
“Development of Matrix Solid Phase Dispersion (MSPD) Method for the Determination of Pesticide Residues in Tomatoes by GC/MS”. Published in the Journal of the Brazilian Chemical Society. (2006)
3
2.0 g of Roma tomato were ground with 0.5 g of alumina with a mortar and pestle.
The mixture was separated by column chromatography. Mobile phase: 20 mL of CH2Cl2
Stationary phase: 0.5 g Florisil and 3.0 g of Na2SO4.
Pentadecanoic acid, 13-methyl ester
Dibutyl Pthalate
Tomato Extract: Total Ion Current Chromatogram
Ion Search of Mass Spectrum
Pentadecanoic acid, 13-methyl ester from the carbon components of the plant
Dibutyl Pthalate a plasticizer
No pesticides; known for low pesticide content.
Solvent seems to be interfering with peaks therefore, solvent delay time was increased.
Iceberg Lettuce
Ocean Mist Brand
Originated from the Ocean Mist Farms in California, U.S.A
Unwashed prior to analysis
Pesticide Extraction Procedure for the Iceberg Lettuce
5.9 g of iceberg lettuce was ground in a mortar and pestle with 2 mL of CH2Cl2.
The mixture was gravity filtered and the solvent was then subjected to the same column chromatography procedure as was utilized for the tomato.
Carrier Gas: Helium
Flow Rate: 1.8 ml/min
Injection Volume 2 mL
Column Temp: 80 °C for 1 minute, increasing to 240 °C at 10 °C/min,
increasing to 265 °C at 10 °C/min, and holding for 10 min.
Oven Temp: 275°C
Injector Port Temp:
270 °C
Solvent Delay: 10 minutes
Total Time of Run 50 minutes
Mode: Split
Varian Saturn 2000 GC-MS Parameters for Iceberg Lettuce Separation
Iceberg Lettuce Extract: Total Ion Current Chromatogram
Ion Search of Mass Spectrum
Pentadecanoic Acid Linoleic Acid Ether Ester Pthalate Long Chain Alchols
No Pesticides
Macintosh Apple
Apple Ridge Brand Macintosh Apple
Originated from Jack Brown Produce in Sparta, MI U.S.A.
Unwashed before analysis
Pesticide Extraction Procedure for the Macintosh Apple
One apple was ground in a blender.
4 g was mixed with CH2Cl2.
CH2Cl2 was decanted and concentrated down to 1 mL.
Subjected to the same column chromatography parameters as before; except doubling of the reagents.
Varian Saturn 2000 GC-MS Parameters for Macintosh Apple Separation
Carrier Gas: Helium
Flow Rate: 1.8 ml/min
Injected Volume: 3 mL
Column Temp: 80 °C for 1 minute, increasing to 240 °C at 10 °C/min,
increasing to 265 °C at 10 °C/min, and holding for 10 min.
Oven Temp: 275°C
Injector Port Temp:
270 °C
Solvent Delay: 10 minutes
Total Time of Run 30 minutes
Mode: Split
Macintosh Apple Extract: Total Ion Current Chromatogram
Ion Search of Mass Spectrum
Endosulfin Endosulfin II Endrin Ketone Linoleic Acid Ethyl Ester (essential fatty
acid) 1-Hexacosene (hydrocarbon)
Pesticides!
Sources of Error
Used Florisil which had a particle mesh size of 30-60.
However, several sources state that a 60-100 particle mesh size is suitable for pesticide residue analysis (SIGMA)4
References
1. National Pesticide Information Center. npic.orst.edu (12/01/08)
2. University of Central Missouri. www.ucmo.edu (12/01/08)
3. Filho, M. Adalberto; Navickiene, Sandro; Dorea, S. Haroldo. “Development of the MSPD Method for the Determination of Pesticide Residues in Tomatoes by GC-MS”. J. Braz. Chem. Soc. 2006, 17, 1-8.
4. Sigma-Aldrich. www.sigmaaldrich.com. (12/01/08)