peak performer 1 pdd - 泰利司系統股份有限公司peak performer 1 pdd the peak performer 1...
TRANSCRIPT
www.peaklaboratories.com
2330 Old Middlefield Way, Suite 10, Mountain View, CA 94043 USA Ph: 650-691-1267 Fax: 650-691-1047
Peak Performer 1 PDD The Peak Performer 1 PDD uses a highly sensitive Pulse Discharge Detector (PDD) which has been designed specifically for process applications in UHP gases. The PDD is used to measure trace impurities down to part per billion levels, while offering a common operating platform for simple, accurate measurements. The PP1 technology has proven to offer a more cost-effective and user friendly alternative to other process GC’s. In addition, Peak Labs practical experience and ability to customize to suit your applications makes Peak Labs your partner, not just your supplier. Please contact us at Peak Laboratories or your local sales representative for custom applications and further information on our standard applications. The Peak Performer 1 PDD gas chromatograph (GC) has been optimized for nitrogen as an impurity in UHP gases to simplify the chromatography and consistency of the process application. Typical applications are in UHP process gases such as, Ar, O2, He, and H2. Performance * Some standard detection limits (in parts per billion)
Matrix Gas He Ar O2 H2 N2 N2O Impurity
N2: Nitrogen 1** 1** 1** 1** n/a 1- 5 Ar: Argon 1** n/a 1** 1** 1** n/a
- * All performance specifications can be optimized at best efforts for specific applications. - ** Contact local sales agent for quote if < 1 ppb LDL is required
• Unless specified, carrier gas is purified research grade helium.
www.peaklaboratories.com
2330 Old Middlefield Way, Suite 10, Mountain View, CA 94043 USA Ph: 650-691-1267 Fax: 650-691-1047
Accuracy • + / - detection limit or 10% of reading,
whichever is higher. Range • 10000 : 1
Examples, 1 ppb – 10 ppm w/1 cc sample loop
Operation Run time: • 400 seconds < (depending on application) Operating Temperature: • 55 - 85 °F (13 - 30 °C ) Gas Supply Requirements: 1. Carrier: • Purified research grade helium • Supply pressure 70 – 110 psig with 5% stability
to purifier • Total flow ~ 200 sccm (plus ~ 50 cc per
actuation) 2. Actuator
• Purified research grade helium or equivalent regulated @ 70 psig
Data Collection / Communication (Standard) • Backlit 3” x 5.25” touch screen • Visual chromatogram and numerical results • On-board integration with re-run capability • Chromatograms collected and archived via
standard 9 – pin com-port in .csv format (data is excel compatible)
Dimensions / Electrical • 27” L x 17” W x 7” H • 25 lbs. • 115 VAC, 50 – 60 Hz / 220 VAC, 50-60 Hz • 1.5 amp maximum Accessories and other features • trend outputs (VDC or mA available) • analysis averaging • auto start/stop capability
Example Ar / N2 analysis in Helium:
PDHID Low Blend Step Challenge
y = 3965.5x + 1813.9(Ar) R2 = 0.9977
y = 3648.4x + 1613.1(N2) R2 = 0.9929
0
10000
20000
30000
40000
50000
60000
70000
80000
0 5 10 15 20Concentration (PPB)
Mea
sure
d A
rea
Cou
nts
Ar
N2