panoura 18s with ceph operator training quick guide
TRANSCRIPT
![Page 1: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/1.jpg)
Panoura 18SPanoramic X-Ray with Ceph Operation Training
Quick Guide
![Page 2: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/2.jpg)
2ImageWorks Proprietary – Confidential Information
These instructions serve as supplemental guidance on the Panoura 18S. They are not meant to replace
the installation manual, but rather serve to complement it.
If there are any questions, please do not hesitate to contact us
914-592-6100
Summary
![Page 3: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/3.jpg)
3ImageWorks Proprietary – Confidential Information
1. Unit Overview
2. Powering on and Emergency Stop
3. Pan – Prep Unit for Patient
4. Pan – Position Patient
5. Pan – Initiate Scan
6. Ceph - Prep Unit for Patient
7. Ceph - Position Patient
8. Ceph – Initiate Scan
9. Changing Modes
Contents
![Page 4: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/4.jpg)
4ImageWorks Proprietary – Confidential Information
Forehead and temple supports
Bite Stick
Chin rest
Hand Holds
X-ray head
Image sensor
Console
1. Unit Overview
![Page 5: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/5.jpg)
5ImageWorks Proprietary – Confidential Information
Console
1. Unit Overview
Handswitch
![Page 6: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/6.jpg)
6ImageWorks Proprietary – Confidential Information
2. Power Switch
Power Switch
![Page 7: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/7.jpg)
7ImageWorks Proprietary – Confidential Information
Before bringing patient to machine…
Confirm that unit is in Pan mode and sensor is in pan position.In imaging software, make sure you are ready to acquire a panoramic x-ray with the correct device selected.Initiate image acquisition process in imaging software.Dialogue box appears.
Click on the “NEW” button to capture a new X-Ray scan.
3. Pan - Prep Unit for Patient
![Page 8: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/8.jpg)
8ImageWorks Proprietary – Confidential Information
Ensure that the “Radiation Mode Setup” dialog shown below is displayed.
Click on “Panorama” Tab.
Click on Auto Panoramic.
Press Ok.
Ready for exposure screen will appear.
3. Pan - Prep Unit for Patient
![Page 9: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/9.jpg)
9ImageWorks Proprietary – Confidential Information
3. Pan - Prep Unit for Patient
Open the head supports before the patient enters the machine
Move head rest to an approximate position of patient height by pressing Up/Down switch on the console.
Have patient step into machine and rest chin on chinrest.
Recommend patient cross arms and hold stabilizing bars. This drops the shoulders.
Have patient step both feet forward about a foot from the post (holding on to stabilizing bars for support).
Bite bitestick with upper and lower teeth in grooves (place sleeve over bitestick).
![Page 10: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/10.jpg)
10ImageWorks Proprietary – Confidential Information
4. Pan - Patient PositioningHave the patient remove jewelry, eyeglasses, etc and place the apron on the patient.
Press Beam button. Unit will reposition and laser lights will come on.
Close head supports around patient’s head
Utilize the mirror to position the patient. There are 3 beams that must be aligned:
1) Medial/midsagittal ; 2) Frankfurt plane/Oculoauricular; 3) Anteroposterior Plan/Canine tooth.
![Page 11: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/11.jpg)
11ImageWorks Proprietary – Confidential Information
4. Pan - Patient Positioning
1. Medial/midsagittal line should run through the middle point between the left and right incisors and the nose ridge
![Page 12: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/12.jpg)
12ImageWorks Proprietary – Confidential Information
4. Pan - Patient Positioning
2. Frankfurt plane should run from base of eye orbitals to lower portion of ear opening.
If chin too high – flat occlusal plane. Pan will look like a frown
If chin too low – pan will look like smile
![Page 13: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/13.jpg)
13ImageWorks Proprietary – Confidential Information
4. Pan - Patient Positioning
3) Anteroposterior Plan/Canine tooth line should align with mandibular third tooth (patient’s left canine tooth). To assist in showing the tooth, patient can be asked to smile. However, have patient close lips before scan.
![Page 14: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/14.jpg)
14ImageWorks Proprietary – Confidential Information
Press RESET on the console Turn beams off (if still on, BEAM button can be pressed)
Confirm ready for exposure screen still appears.If possible the operator should stand where they can observe both the patient in the machine and the computer screen.
5. Pan - Initiate Scan
Note: Unit will not scan when laser beams are on. Either wait until beam go off or press BEAM button to turn them
off.
![Page 15: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/15.jpg)
15ImageWorks Proprietary – Confidential Information
Confirm Ready LED is solid green (not blinking)
Instruct the patient to swallow, place the tongue on the roof of the mouth, and hold still.
Press and hold switch during entirety of scan.
5. Pan - Initiate Scan
Note: If switch is released before exposure is complete, an ER08
warning is displayed. These error codes can be cleared by pressing RESET on panel.
![Page 16: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/16.jpg)
16ImageWorks Proprietary – Confidential Information
As scan is taken, preview screen will appear.
When preview screen shows completed image, system will process and contruct the image.
During this time, patient can step out of machine.
5. Pan - Initiate Scan
![Page 17: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/17.jpg)
17ImageWorks Proprietary – Confidential Information
Acquired image appears in window.
Click OK to transfer the image to the imaging software where it is permanently saved.
5. Pan - Initiate Scan
![Page 18: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/18.jpg)
18ImageWorks Proprietary – Confidential Information
Before bringing patient to machine…
Confirm that unit is in Ceph mode and sensor is in ceph position.In imaging software, make sure you are ready to acquire a ceph x-ray with the correct device selected.Initiate image acquisition process in imaging software.Dialogue box appears.
Then click on the “NEW” button to capture a new X-Ray scan.
6. Ceph - Prep Unit for Patient
![Page 19: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/19.jpg)
19ImageWorks Proprietary – Confidential Information
Ensure that the “Radiation Mode Setup” dialog shown below is displayed.
Click on (or verify set to) “Cephalo” Tab.
Click on desired sub mode (LA or PA).
Press Ok.
Ready for exposure screen will appear.
6. Ceph - Prep Unit for Patient
![Page 20: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/20.jpg)
20ImageWorks Proprietary – Confidential Information
6. Ceph - Prep Unit for Patient
Move the ear rod to an approximate position of patient height by pressing Up/Down switch on the console.
Rotate head piece depending on PA or LA.
Ear rods should match the heights of the patient’s ear canal (porus acusticus).
Spread the ear rods.
![Page 21: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/21.jpg)
21ImageWorks Proprietary – Confidential Information
7. Ceph - Patient Positioning
Have the patient remove jewelry, glasses, etc. and place the apron on the patient.If taking posteroanterior (PA), remove nose support.Insert ear rod on one side and close the ear rod.Gently insert the ear rod on the other side. Make sure the patient’s Frankfurt plane is positioned horizontally.If taking lateral (LA), then adjust the height of the nose support at the center between the patient’s eyebrows and gently place it to fix.
![Page 22: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/22.jpg)
22ImageWorks Proprietary – Confidential Information
Press RESET on the console
Confirm ready for exposure screen still appears.
If possible the operator should stand where they can observe both the patient in the machine and the computer screen.
7. Ceph - Initiate Scan
Note: Unit will not scan when laser beams are on. Either wait until beam go off or press BEAM button to turn them
off.
![Page 23: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/23.jpg)
23ImageWorks Proprietary – Confidential Information
Confirm Ready LED is solid green (not blinking)
Instruct the patient to swallow, place the tongue on the roof of the mouth, and hold still.
Press and hold switch during entirety of scan.
8. Ceph - Initiate Scan
Note: If switch is released before exposure is complete, an ER08
warning is displayed. These error codes can be cleared by pressing RESET on panel.
![Page 24: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/24.jpg)
24ImageWorks Proprietary – Confidential Information
As scan is taken, preview screen will appear.
When preview screen shows completed image, system will process and contruct the image.
During this time, patient can step out of machine.
8. Ceph - Initiate Scan
![Page 25: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/25.jpg)
25ImageWorks Proprietary – Confidential Information
Acquired image appears in window.
Open the ear rods and guide the patient out of the machine.
Click OK to transfer the image to the imaging software.
8. Ceph - Initiate Scan
Note: Default setting
displays head facing to the right.
![Page 26: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/26.jpg)
26ImageWorks Proprietary – Confidential Information
To change machine from Pan to Ceph or from Ceph to Pan, the sensor must be physically moved, and the unit switched to the new mode.
Warning: Sensor system is very robust and the fastening system has multiple support levels to assure
it is secure. However, the sensor can be damaged if dropped, so please use care in moving the sensor.
Hold sensor by handleTurn know to “Unlock” positionPull sensor off of railsMove sensor to new position, and slide onto railsTurn knob to lock into position
9. Changing Mode
![Page 27: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/27.jpg)
27ImageWorks Proprietary – Confidential Information
Press MODE on console.
Toggle to the desired mode (Pan, Ceph LA, Ceph PA).
Press RESET.
Sensor will turn on and system is ready to use in the new mode.
9. Changing Mode
![Page 28: Panoura 18S with Ceph Operator Training Quick Guide](https://reader036.vdocuments.mx/reader036/viewer/2022081517/58823a831a28ab31228b5db9/html5/thumbnails/28.jpg)
28ImageWorks Proprietary – Confidential Information
These instructions serve as supplemental guidance of the Panoura. They are not meant to replace the
manual, but rather serve to complement it.
If there are any questions, please do not hesitate to contact us
914-592-6100
Summary