panels for quality assurance and quality - who · cv coefficient of variation ... qa quality...

57
Technical Guidance Series (TGS) for WHO Prequalification – Diagnostic Assessment Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–6 Draft for comment 22 May 2017

Upload: truongquynh

Post on 07-Jul-2018

215 views

Category:

Documents


0 download

TRANSCRIPT

Technical Guidance Series (TGS)

for WHO Prequalification – Diagnostic Assessment

Panels for quality

assurance and quality

control of in vitro

diagnostic medical

devices

TGS–6

Draft for comment 22 May 2017

© World Health Organization 2017 All rights reserved. Publications of the World Health Organization can be obtained from WHO Press, World Health Organization, 20 Avenue Appia, 1211 Geneva 27, Switzerland (tel.: +41 22 791 3264; fax: +41 22 791 4857; email: [email protected]). Requests for permission to reproduce or translate WHO publications – whether for sale or for non-commercial distribution – should be addressed to WHO Press, at the above address (fax: +41 22 791 4806; email: [email protected]). The designations employed and the presentation of the material in this publication do not imply the expression of any opinion whatsoever on the part of the World Health Organization concerning the legal status of any country, territory, city or area or of its authorities, or concerning the delimitation of its frontiers or boundaries. Dotted lines on maps represent approximate border lines for which there may not yet be full agreement. The mention of specific companies or of certain manufacturers’ products does not imply that they are endorsed or recommended by the World Health Organization in preference to others of a similar nature that are not mentioned. Errors and omissions excepted, the names of proprietary products are distinguished by initial capital letters. All reasonable precautions have been taken by the World Health Organization to verify the information contained in this publication. However, the published material is being distributed without warranty of any kind, either expressed or implied. The responsibility for the interpretation and use of the material lies with the reader. In no event shall the World Health Organization be liable for damages arising from its use. Contact: Irena Prat, EMP Prequalification Team Diagnostics WHO – – 20 Avenue Appia – – 1211 Geneva 27 Switzerland

WHO Prequalification – Diagnostic Assessment: Technical Guidance Series

The WHO Prequalification Programme is coordinated through the Department of

Essential Medicines and Health Products. WHO prequalification of in vitro diagnostic

medical devices (IVDs) is intended to promote and facilitate access to safe, appropriate

and affordable IVDs of good quality in an equitable manner. The focus is on IVDs for

priority diseases and their suitability for use in resource-limited settings. The WHO

Prequalification Programme undertakes a comprehensive assessment of individual IVDs

through a standardized procedure that is aligned with international best regulatory

practice. It also undertakes post-qualification activities for IVDs to ensure ongoing

compliance with prequalification requirements.

Products that are prequalified by WHO are eligible for procurement by United Nations

agencies. The products are then commonly purchased for use in low- and middle-income

countries.

IVDs prequalified by WHO are expected to be accurate, reliable and be able to perform as

intended for the lifetime of the IVD under conditions likely to be experienced by a typical

user in resource-limited settings. The countries where WHO-prequalified IVDs are

procured often have minimal regulatory requirements, and the use of IVDs in these

countries presents specific challenges. For instance, IVDs are often used by health care

workers without extensive training in laboratory techniques, in harsh environmental

conditions, without extensive pre- and post-test quality assurance (QA) capacity, and for

patients with a disease profile different to those encountered in high-income countries.

Therefore, the requirements of the WHO Prequalification Programme may be different to

the requirements of high-income countries, or of the regulatory authority in the country of

manufacture.

The Technical Guidance Series was developed following a consultation – held on 10–

13 March 2015 in Geneva, Switzerland – which was attended by experts from national

regulatory authorities, national reference laboratories and WHO prequalification dossier

reviewers and inspectors. The guidance series is a result of the efforts of this and other

international working groups.

This guidance is intended for manufacturers interested in WHO prequalification of their

IVD. It applies in principle to all IVDs that are eligible for WHO prequalification for use in

WHO Member States. It should be read in conjunction with relevant international and

national standards and guidance.

The TGS guidance documents are freely available on the WHO website.

WHO

Prequalification

– Diagnostic

Assessment

Procurement of

prequalified

IVDs

Prequalification

requirements

About the

Technical

Guidance

Series

Audience and

scope

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

4 Draft for comment 22 May 2017

Contents

Acknowledgements .................................................................................................................... 6

1 Abbreviations and definitions ................................................................................... 7

1.1 Abbreviations .......................................................................................................................7

1.2 Definitions ............................................................................................................................7

2 Introduction ........................................................................................................... 12

2.1 Key concept ...................................................................................................................... 12

2.2 Rationale for the use of QA and QC panels ...................................................................... 12

2.3 Purpose of this document ................................................................................................ 13

2.4 Limitations of this guidance .............................................................................................. 13

3 WHO prequalification requirements ....................................................................... 15

3.1 Manufacturer responsibility ............................................................................................. 15

3.2 Standards .......................................................................................................................... 15

3.3 Suitability for use in Member States ................................................................................ 16

4 Basic principles for developing panels ..................................................................... 17

4.1 Core principle .................................................................................................................... 17

4.2 Quantitation is essential ................................................................................................... 17

4.3 Verification and validation ................................................................................................ 18

4.4 Use of surrogate specimens in panels .............................................................................. 18

5 Use of panels .......................................................................................................... 19

5.1 Release-to-sale panels ...................................................................................................... 19

5.2 In-process panels .............................................................................................................. 19

5.3 Stability panels .................................................................................................................. 19

5.4 Reproducibility during evaluations ................................................................................... 20

5.5 Analytical sensitivity and range: quantitative assays ....................................................... 21

5.6 Control materials provided to users ................................................................................. 21

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

5 Draft for comment 22 May 2016

6 Selection of specimens ........................................................................................... 22

6.1 General comments ........................................................................................................... 22

6.2 In-house panels ................................................................................................................. 23

6.3 Certified reference materials ............................................................................................ 29

6.4 International conventional calibration materials ............................................................. 29

6.5 Measurement variation .................................................................................................... 30

6.6 External regulatory panels ................................................................................................ 30

7 Relationship between panel members and claims ................................................... 33

7.1 Process control charts ...................................................................................................... 34

8 Maintenance of panels ........................................................................................... 35

8.1 Storage .............................................................................................................................. 35

8.2 Replacement ..................................................................................................................... 35

9 References.............................................................................................................. 38

10 Annex 1 – Examples ................................................................................................ 41

10.1 Example 1: Correlation of a QA / QC panel member with a critical specimen ................. 41

10.2 Example 2: Correlation under stress conditions ............................................................... 44

10.3 Example 3: Validation of a release-to-sale panel for anti-HCV ........................................ 47

10.4 Example 4: A release panel for a flow cytometer for the enumeration of CD4 T-cells .... 49

10.5 Example 5: Nucleic acid testing ........................................................................................ 51

10.6 Example 6: Use of imposed specimens as QA / QC panel members ................................ 53

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

6 Draft for comment 22 May 2017

Acknowledgements

The document Panels for quality assurance and quality control of in vitro diagnostic

medical devices was developed as part of the Bill & Melinda Gates Foundation Umbrella

Grant and the UNITAID grant for “Increased access to appropriate, quality-assured

diagnostics, medical devices and medicines for prevention, initiation and treatment of

HIV/AIDS, TB and malaria”. The draft was prepared in collaboration with Dr J Duncan,

London, United Kingdom; Ms K Richards, WHO Geneva, Switzerland and Ms R Meurant,

WHO Geneva, Switzerland and with input and expertise Dr S Hojvat, Maryland, United

States of America; Dr E Cowan, Maryland, United States of America and Dr D Milic, WHO

Geneva, Switzerland. This document was produced under the coordination and

supervision of Kim Richards and Josée Hansen WHO/HIS/EMP, Geneva, Switzerland.

The draft guidance has been posted on the WHO website for public consultation on 22

May 2017.

1

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

7 Draft for comment 22 May 2016

1 Abbreviations and definitions 2

1.1 Abbreviations 3

CE Conformité Européenne (European Conformity)

CV Coefficient of variation

CLSI Clinical and Laboratory Standards Institute

CRM Certified Reference Material

EIA Enzyme-linked immunoassay

GHTF Global Harmonization Task Force

HBsAg Hepatitis B surface antigen

HBV Hepatitis B virus

HCV Hepatitis C virus

HIV Human immunodeficiency virus

IFU Instructions for Use

IS International Standard

ISO International Organization for Standardization

IU International Unit

IVD In vitro diagnostic or in vitro diagnostic device

NAT Nucleic Acid Test, Nucleic Acid Testing

OD Optical density

PEI Paul Ehrlich Institute

QA Quality assurance

QC Quality control

QMS Quality management system

RDT Rapid diagnostic test

R&D Research and development

SI International System of Units/Système International d’Unités

1.2 Definitions 4

The definitions given below apply to the terms used in this document. They may have 5

different meanings in other contexts. 6

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

8 Draft for comment 22 May 2017

Acceptance criteria: A defined set of conditions that must be met to establish the performance of 7 a system. 8

Source: (1) 9 10 Numerical limits, ranges, or other suitable measures for acceptance of the results 11

of analytical procedures. 12 Source: (2) 13 14 Batch/lot: A defined amount of material that is uniform in its properties and has been 15

produced in one process or series of processes. 16

Note: The material can be either starting material, intermediate material or 17 finished product. 18

Source: (1), definition 3.5 19

Component: Part of a finished, packaged and labelled in vitro diagnostic device (IVD). 20

Note: Typical kit components include antibody solutions, buffer solutions, 21 calibrators or control materials. 22

Source: (1), definition 3.12 23

Constituent: Raw materials used to make a component. 24

Source: (1), definition 3.57 25

Control material: A substance, material or article intended by its manufacturer to be used to 26 verify the performance characteristics of a medical IVD. 27

Source: (3), definition 3.4 and (1), definition 3.13 28

Certified reference material (CRM): Reference material, accompanied by a certificate, one or 29 more of whose property values are certified by a procedure that establishes 30 metrological traceability to an accurate realization of the unit in which the 31 property values are expressed, and for which each certified value is accompanied 32 by an uncertainty at a stated level of confidence. 33

Source: (4), definition 3.8 34

Design input: The physical and performance requirements of an IVD that are used as a basis for 35 IVD design. 36

Source: (5), definition f 37

Diagnostic sensitivity: The proportion of patients with a well-defined clinical disorder (or 38 condition of interest) whose test values are positive or exceed a defined decision 39 limit (i.e. a positive result and identification of the patients who have a disease). 40

Note 1: The clinical disorder must be defined by criteria independent of the test 41 under consideration. 42

Note 2: The term "diagnostic sensitivity" (Europe) is equivalent to "clinical 43 sensitivity" (United States) 44

Source: [http://htd.clsi.org] 45

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

9 Draft for comment 22 May 2016

Evidence: Information that can be proved true, based on facts obtained through 46 observation, measurement, test or other means. 47

Source: Modified from (6), definition 3.8.3 48

Instructions for use (IFU): Information supplied by the manufacturer to enable the safe and proper 49 use of an IVD. 50

Note: Includes the directions supplied by the manufacturer for the use, 51 maintenance, troubleshooting and disposal of an IVD, as well as warnings and 52 precautions. 53

Source: (1), definition 3.30 54 WHO comment: In the United States, the acronym IFU occasionally stands for 55 “indications for use”, and the acronym IU stands for “intended use” or 56 “indications for use”. The International Organization for Standardization 57 (ISO) definition and requirements (1) for IFU cover the intended use and the 58 precise method of use. 59

International conventional calibrator: A calibrator whose value of a quantity is not metrologically 60 traceable to the International System of Units (SI), but is assigned by 61 international agreement. 62

Note: The quantity is defined with respect to the intended clinical application. 63

Source: (4) definition 3.11 64

In vitro diagnostic medical device: A medical device, whether used alone or in combination, 65 intended by the manufacturer for the in vitro examination of specimens derived 66 from the human body, solely or principally to provide information for diagnostic, 67 monitoring or compatibility purposes. 68

Note 1: IVDs include reagents, calibrators, control materials, specimen 69 receptacles, software and related instruments or apparatus or other articles; 70 they are used, for example, for diagnosis or to aid diagnosis, screening, 71 monitoring, predisposition, prognosis, prediction and determination of 72 physiological status. 73

Note 2: In some jurisdictions, certain IVDs may be covered by other regulations. 74

Source: (7) 75

International unit: An arbitrary unit assigned to a WHO International Standard by the WHO Expert 76 Committee of Biological Standardisation. 77

Source: (4), Section 4.2.6 78

IVD reagent: Chemical, biological or immunological components, solutions or preparations 79 intended by the manufacturer to be used as an IVD. 80

Source: (1), definition 3.28 81 WHO comment: This document uses the terms “IVD” and “IVD” reagent 82 interchangeably. 83

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

10 Draft for comment 22 May 2017

Life cycle: All phases in the life of a medical device, from the initial conception to final 84 decommissioning and disposal. 85

Source: (8), definition 2.7 86

Metrological traceability: Property of the result of a measurement or the value of a standard 87 whereby it can be related to stated references, usually national or international 88 standards, through an unbroken chain of comparisons, all of which have stated 89 uncertainties. 90

Note 1: Each comparison is affected by a (reference) measurement procedure 91 defined in a calibration transfer protocol. 92

Source: (4) 93

Performance claim: Specification of a performance characteristic of an IVD as documented in the 94 information supplied by the manufacturer. 95

Note 1: The specification can be based on prospective performance studies, 96 available performance data or studies published in the scientific literature. 97

WHO comment: “Information supplied by the manufacturer” includes but is not 98 limited to: statements in the IFU, in the dossier supplied to WHO and /or other 99 regulatory authorities, in advertising, on the internet. 100

Referred to simply as “claim” or “claimed” in this guide. 101

Source: (1), definition 3.51 102

103

Quality assurance (QA): Part of the quality management focused on providing confidence that 104 quality requirements will be fulfilled. 105

106 Source: (6), definition 3.3.6 107 108 Quality control (QC): Part of quality management focused on fulfilling quality requirements. 109 110 Source: (6), definition 3.3.7 111 112 Risk management: The systematic application of management policies, procedures and practices 113

to the tasks of analysing, evaluating, controlling and monitoring risk. 114

Source: (8) 115

Statistical process control: Activities focused on the use of statistical techniques to reduce 116 variation, increase knowledge about the process and steer the process in the 117 desired way. 118

Source: (9), definition 2.1.8 119

Trueness of measurement: Closeness of agreement between the average values obtained from a 120 large series of results of measurements and a true value. 121

Source: (4), definition 3.33 122

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

11 Draft for comment 22 May 2016

Validation: Confirmation by examination and provision of objective evidence that the 123 particular requirements for a specific intended use can be consistently fulfilled. 124

Source: (5), definition z; (6), definition 3.8.13. 125

Verification: Confirmation by examination and provision of objective evidence that specified 126 requirements have been fulfilled. 127

Source: (5), definition aa; (6), definition 3.8.12. 128

129

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

12 Draft for comment 22 May 2017

2 Introduction 130

2.1 Key concept 131

A “panel” is a collection of well-characterized materials and specimens that is used to 132

monitor aspects of the function of an IVD or its components. Probably the most 133

important use of a panel is to verify, before a lot of an IVD can be released-to-sale, that 134

the lot will consistently meet all its quality critical metrics until the end of its assigned 135

shelf life (not just at the time of its release-to-sale, also referred to as batch release). 136

Panels are also used for in-process control, during reproducibility and stability studies and 137

for some aspects of design validation. The same materials might be used for each of these 138

purposes, but would be assigned different acceptance criteria for the different functions. 139

The manufacturer should be able to justify their rationale for assigning specific 140

acceptance limits when panel samples are being tested at any point throughout the 141

product lifecycle. The rationale can be documented as part of the risk management 142

process; alternatively, it can be included in the design control documentation when 143

statistical techniques are used as part of the process for establishing performance 144

characteristics. 145

2.2 Rationale for the use of QA and QC panels 146

There is a regulatory requirement for CE-marked products for the detection of infectious 147

diseases listed in Annex II of the In Vitro Diagnostic Medical Devices Directive 98/79/EC 148

(IVDD) that, “The manufacturer’s release testing criteria shall ensure that every batch 149

consistently identifies the relevant antigens, epitopes, and antibodies” (10: Section 3.4.1). 150

It is expected that this will be shown for all of the relevant specimen types claimed for the 151

IVD (e.g. serum, whole blood and urine), even for rest-of-world products that fall into the 152

high-risk categories C and D of the Global Harmonization Task Force (GHTF) classification 153

(10). However, subject to documented risk evaluation by the manufacturer, this 154

requirement could be relaxed to testing of only the most searching specimen type(s). It is 155

also a regulatory requirement that the manufacturer provides objective, scientifically 156

sound evidence to support all claims made regarding the performance of an IVD (e.g. 157

stability, reproducibility and sensitivity). It is not reasonable to verify all performance 158

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

13 Draft for comment 22 May 2016

metrics for every lot manufactured, but use of a well-designed lot release panel will – 159

subject to appropriate risk assessment and validation work – provide a high degree of 160

assurance that each lot of the IVD will consistently perform as claimed for its assigned life. 161

In addition to testing of the IVD at release, conventional practice is to evaluate 162

intermediates in the of manufacture (referred to as “in-process testing”). Use of carefully 163

chosen panels will provide evidence that the intermediates meet their specifications and 164

that the manufacturing process is in statistical control (9). In-process panels are also used 165

to provide evidence that a lot is homogeneous in performance metrics from the 166

beginning to the end of each production process. 167

Panels are also needed for stability and reproducibility studies used during IVD 168

development, and for aspects of verification at the end of the product’s assigned shelf 169

life. Results over time – as dictated by the developer’s quality management system (QMS) 170

– must be shown to be within predetermined and validated specifications. 171

2.3 Purpose of this document 172

The purpose of this document is to provide IVD manufacturers with guidance on possible 173

approaches to preparing validated panels for QA and QC; for example, choosing the 174

materials, assigning meaningful criteria to them, storing them and replacing them when 175

necessary. It describes the expectations for WHO prequalification in terms of the QA and 176

QC information to be provided in dossiers submitted according to PQDx_018 (11: Section 177

6.2.1). It also provides guidance on information that might be requested during QMS 178

inspections according to PQDx_014 (12: Section 7.2.2), following requirements in 179

ISO 13485:2016 (13: Clauses 7.3.4 and 8.2.6). 180

2.4 Limitations of this guidance 181

This document should not be taken as a prescriptive checklist of what must be 182

performed, but as a guide on how to improve processes and generate the evidence 183

needed to ensure a comprehensive, systematic procedure with an appropriate risk 184

management plan. As explained in Section 3.3, the expectations of the WHO 185

prequalification might be more stringent than the requirements of the users and 186

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

14 Draft for comment 22 May 2017

regulatory authority in the country of manufacture. Wherever possible the guidance 187

attempts to explain the reasons for these additional expectations. Other approaches to 188

accommodating these further expectations can be provided in dossiers submitted for 189

WHO prequalification, if supported by rigorous risk assessment or other evidence. 190

The examples included in this document apply to the principles outlined here only. 191

Manufacturers must perform their own product-specific risk assessment for each of their 192

IVDs. The risk evaluation must be related to the specific product, in its specific format 193

(e.g. antigen sandwich, next-generation sequence methodology, etc.), and to the specific 194

target analyte and the specific, claimed intended use and users. 195

Depending on the particular categorization of the product, additional requirements may 196

apply in particular jurisdictions. Such regulatory and legal requirements are specific for 197

each regulatory authority; they are beyond the scope of this document but should be 198

documented in the design input requirements and their effects should be evaluated by 199

risk analyses. 200

201

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

15 Draft for comment 22 May 2016

3 WHO prequalification requirements 202

WHO requires the following details for prequalification: 203

…any in-process and final product testing ... 204

… an overview of verification, validation and quality control activities for all 205

stages of design and manufacture (including purchased components, 206

in-process products, and finished products). 207

Provide the batch release criteria … 208

Source: (11: Section 6.2.1) 209

The extent of the information provided in the dossier will vary. For in-process, the control 210

points and test methods would probably be noted on process flow diagrams, with a link 211

to the risk assessments that describe the necessity for that control. For stability or for 212

reproducibility work, and especially for release-to-sale (i.e. batch release), a full 213

description of the panel would be expected, including the reason for the inclusion of each 214

panel member, its characterization, the criteria assigned and the validation of the test 215

method. If required during any on-site QMS inspection (12), the full information about 216

each QA and QC control point and test method should be available in the design history 217

file. 218

The information provided must demonstrate the link to the predetermined user 219

requirements and to product development. 220

3.1 Manufacturer responsibility 221

It is a manufacturer’s responsibility to ensure that all claims made regarding the 222

performance of the IVD are supported by evidence that is objective and scientifically 223

sound. 224

3.2 Standards 225

WHO recommends that manufacturers be familiar with the standards and guidance 226

documents listed in the Standards applicable to the WHO prequalification of in-vitro 227

diagnostics (14), and take them into account when planning, assessing risk and 228

developing QA and QC procedures. 229

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

16 Draft for comment 22 May 2017

3.3 Suitability for use in Member States 230

Information on the use and value of panels for QA and QC in the dossiers of a product 231

submitted to WHO must reflect the expected environmental conditions and the normal 232

usage conditions and methods encountered by the users in WHO Member States. The 233

environmental conditions might differ from, and be more extreme than, those in the 234

country of manufacture; for example, more extreme temperature and humidity, and 235

different contaminating microorganisms. Each of these factors must be considered not 236

only in the design input documentation (in the risk management section) but also in 237

validating the IVD and in developing QA procedures to verify adequate performance 238

through the life cycle of the IVD. 239

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

17 Draft for comment 22 May 2016

4 Basic principles for developing panels 240

4.1 Core principle 241

The core principle underlying any testing is that the results of the test are relevant to the 242

investigation; that is, the test methods must be validated (15). For development of QA 243

and QC panels, the materials and panel members chosen must be applicable to the task in 244

hand, and their utility must be validated and documented. A panel member should be 245

chosen for the purpose of showing stability (or sensitivity, reproducibility, etc.) of some 246

aspect of an IVD, so that if the IVD becomes unstable (or insensitive, irreproducible, etc.) 247

in that respect, the test result from that panel member will reflect the potential for 248

generation of an incorrect result, with resulting incorrect patient management. Often, the 249

data provided in WHO prequalification submissions do not adequately support the 250

conclusions drawn because the panel members have not been properly characterized and 251

validated for their stated use. Examples of appropriate (and inappropriate) choice of 252

specimens are given in Annex 1 of this document, and in the guides and sample dossiers 253

available online (16). 254

4.2 Quantitation is essential 255

Panels are used in quantitative, semiquantitative and qualitative processes. It is important 256

to be able to show whether a parameter is different from what is expected and, if so, by 257

how much. That change can then be related to the predetermined limits within which the 258

IVD will function, as shown by the panel validation work (see Section ‎0 of this document). 259

Finding a panel member to be merely positive or negative is often uninformative. Many 260

IVDs are not intended to produce quantitative results; however, for certain situations 261

(e.g. release-to-sale, stability work, robustness studies and process control) it is essential 262

to be able to assign some form of quantitative values to the results from panel members. 263

The methods for doing this should be determined through research and development 264

(R&D). 265

For most rapid diagnostic tests (RDTs), the intensity of the colour of the signal from the 266

IVD can be compared with a calibration scale. Degradation of signal intensity may be a 267

sign that the IVD is degenerating. It is sometimes argued that the relative intensity of the 268

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

18 Draft for comment 22 May 2017

signal from a qualitative IVD antibody test is not related to the antibody content of the 269

specimen. Although that is true when comparing different specimens it is not true for the 270

relative signal generated from an IVD that has degenerated or been manufactured 271

incorrectly (nor for dilutions of a single specimen); in such cases, the signal changes and 272

that change can be used to monitor potential changes in the device. 273

Some IVDs based on a nucleic acid test (NAT) cannot be forced to give a quantitative 274

signal, and some qualitative IVDs need to be assigned a sensitivity (with confidence limits) 275

relative to an international conventional calibrator. In such cases, the parameter that can 276

most easily be used to monitor stability, and verification at release, is the detection limit 277

of each claimed analyte, which is almost always required to be stated either in the IFU or 278

in regulatory submissions. The detection limit (or any other threshold value) is found by 279

probit or logistic analysis of replicates of dilutions of the target analytes in each claimed 280

specimen type, but it is beyond the scope of this document to discuss the use of these 281

techniques in panels. For an introductory text see (17: Section 6.2.6) and (18: Appendix C). 282

The data should be analysed statistically rather than by eye, and a statistical plan should 283

be developed before starting the experiments. 284

4.3 Verification and validation 285

The following discussion often uses the terms “verification” (or “verify”) and “validation” 286

(or “validate”). These terms are defined in Section ‎1.1; however, to reiterate, verification 287

means providing proof of meeting predetermined specifications, whereas validation 288

means providing proof of consistently meeting predetermined needs, one of which is 289

always that the validated factor is robust and reliable. 290

4.4 Use of surrogate specimens in panels 291

Frequently, IVD lots must be verified as being able to detect particular specimens that are 292

often available in only limited quantities, because of either rarity or restriction in supply 293

from an authority. In such cases, surrogate panel members must be found that, when 294

tested in the assay of interest, mimic the rare specimens but are more readily available 295

(see Section ‎6 and Example 1 in Annex 1). 296

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

19 Draft for comment 22 May 2016

5 Use of panels 297

5.1 Release-to-sale panels 298

As noted in Section ‎2.2, for many years it has been accepted as best practice that the 299

release criteria for IVD will ensure that the device consistently performs as claimed over 300

the assigned shelf life of the lot. It is not necessary to verify every aspect of performance 301

at release-to-sale with every specimen type claimed (provided that R&D analytical matrix 302

studies have shown performance equivalence). Nevertheless, the documentation of 303

claims that are not verified at release must include a stringent risk assessment and 304

evidence of validation during the development of the device. 305

5.2 In-process panels 306

QA panels used to verify individual process stages (including testing of incoming 307

materials) and the performance of manufactured intermediates will be varied, and will 308

require critical evaluation and validation. Materials used in these in-process panels might 309

be similar to those used in other panels, particularly the release-to-sale panel, but the 310

criteria assigned will usually be different. For example, the in-process criteria assigned to 311

a panel member used for release of a coated microplate to the next stage of manufacture 312

might be wider than the criteria assigned to the same panel member when used at 313

release-to-sale, because some variability in the coating is accommodated by adjustment 314

of a conjugate. 315

5.3 Stability panels 316

Stability of each critical aspect of the IVD must be evaluated, subject to risk assessment 317

where the necessity to evaluate is determined. Some of the performance claims that are 318

not verified lot by lot at release-to-sale will probably be validated as part of the stability 319

programme. Hence, the QA panels used for stability studies, although similar to those 320

used for releasing lots to sale, are likely to be more comprehensive. 321

It is good practice to establish an ongoing stability monitoring programme where 322

necessary (19: Section 4.1), to verify that lots released-to-sale maintain their claimed 323

performance at the end of their assigned shelf lives. This verification is achieved using 324

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

20 Draft for comment 22 May 2017

IVDs retained by the manufacturer and evaluated at some predetermined time at or after 325

their expiration date, employing a panel designed specifically for the purpose (20). The 326

panel will probably be the same as that used for release-to-sale but it is likely to include 327

additional material to monitor some of the quality metrics validated for stability and 328

therefore not verified at release (e.g. in-use stability, open-pack stability and specificity). 329

The criteria might be different from those of the same panel members in the release-to-330

sale panel. For example, if there is a known and allowable degeneration of signal over the 331

life of the device the requirement at release will be higher than that at end of life so that 332

the device will continue to meet the claim over its assigned life. 333

The manufacturer can extend a product’s shelf-life claim by using IVD lots kept under 334

ongoing stability conditions and evaluated at appropriate intervals. The panel used for 335

this, as with the panel used for validation, will probably be more complex in its 336

composition than the release-to-sale panel. 337

5.4 Reproducibility during evaluations 338

The QA panel used for repeatability and reproducibility studies need not be complex – 339

unless a risk assessment shows otherwise. It is generally sufficient to use dilutions of 340

selected specimens or calibrants with concentrations near decision points (e.g. cut-off 341

values or clinically important concentrations) in each of the claimed specimen types. A 342

significant aspect of precision studies that is frequently absent from submissions to WHO 343

prequalification is the relationship between the repeatability and reproducibility as 344

measured by the developers of the assay and by intended users in their own 345

environment. If there is an important difference between the findings of the developers 346

and of the intended users, the assay needs further development in some respect. Best 347

practice is to give the users involved in external performance evaluation / clinical 348

validation a reproducibility panel that is identical to the one used by the R&D department 349

during development work, and then to ensure that the panel is tested, along with the 350

specimens, each time the assay is performed. The reproducibility data collected during 351

external performance evaluation should cover the variability between the testing sites, 352

the testers and the lots of IVD used. Comparison of these data with similar in-house data, 353

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

21 Draft for comment 22 May 2016

and the development of an experimental design that will enable the data to be collected 354

in the first place, requires statistical advice that is beyond the scope of this guide. 355

5.5 Analytical sensitivity and range: quantitative assays 356

Panel members used in development work for verifying the limits of detection, 357

quantitation and sensitivity at clinically important thresholds and range of quantitative 358

assays are likely to be a subset of the materials eventually used in the release-to-sale 359

panel. For most quantitative assays, certified reference materials or international 360

conventional calibrators will be available. In the worst case, only a manufacturer’s 361

calibrator will be available – ISO 17511 (4: Section 4.2.2g ) provides information about 362

assigning values to such material. These panels are of fundamental importance in 363

verifying the performance of an IVD. 364

5.6 Control materials provided to users 365

Control materials provided to users are intended to assure users that IVD performance is 366

consistent with its intended use and the manufacturer’s claims – a function related to 367

that of QA / QC panels. Such materials are sometimes called “run controls”, and they 368

differ from materials provided as part of external QA programmes. These controls are not 369

part of the manufacturer’s own QA system; nevertheless, the controls chosen and the 370

criteria they must meet are similar to those of the manufacturer’s internal QA panels. 371

Controls may be supplied along with the IVD as individual components or as a line on a 372

membrane (e.g. in the case of an RDT), or they may be available to purchase separately; 373

however, in all cases, ISO 15198 (3) is applicable. The claims given for run controls in IFU 374

need careful validation, and it must be shown that these controls do in fact provide 375

evidence that the IVD has or will function as claimed. Submissions to WHO 376

prequalification rarely document that the control materials meet the relevant claim in the 377

IFU; for example, for serological assays, most run controls merely show addition of a 378

reagent, or flow, but not that the IVD would meet its claimed quality critical performance. 379

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

22 Draft for comment 22 May 2017

6 Selection of specimens 380

6.1 General comments 381

Some IVD require use of fresh specimens; for example, assays for CD4, some NIVDD 382

AT IVDs and some assays requiring capillary whole blood. If stable surrogate specimens 383

cannot be generated and validated then all the required panels must be generated and 384

evaluated before each use. This will require use of a predicate test method, validated for 385

both the purpose and the specific specimen type to be used. The variance of the test 386

method must be proven not to conceal variance in the IVD being verified. This 387

requirement is usually studied as gauge repeatability and reproducibility (R&R), but the 388

process and methodology applies to any measuring system, not just to gauges. Statistical 389

analysis of gauge R&R is well documented (21). 390

Inactivation 6.1.1391

Many specimens used in QA panels are reactive for infectious organisms and must 392

therefore be handled with appropriate caution. All control materials should undergo 393

universal screening; for example, for human immunodeficiency virus (HIV), hepatitis B 394

virus (HBV), hepatitis C virus (HCV), human T-lymphotrophic virus (HTLV- I/II) and syphilis. 395

They should also be shown to be negative for transmissible (i.e. infectious) agents, unless 396

a particular analyte is essential for demonstrating a performance characteristic of an IVD 397

or suitable negative specimens are not available. 398

It is often convenient to inactivate the organisms specific for the assay so that the 399

specimens can be handled with lower risk; for control materials provided to users (see 400

Section ‎5.6), inactivation is generally essential. Thermal inactivation is commonly used for 401

HIV-positive specimens but it is often not realized that the routine conditions for such 402

inactivation (≥56 °C for ~30–60 minutes) do not adequately inactivate virus dried down in 403

blood fractions, as might be found round stoppers or seals (22). Even the conditions 404

commonly used for HBV inactivation (≥65 °C for ~16–20 hours) might not completely 405

inactivate dried specimens containing HIV. Where thermal inactivation is used, it must be 406

performed and documented correctly, using properly calibrated thermal measurements 407

and taking care to ensure that no dried specimen resides on the tops of containers. 408

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

23 Draft for comment 22 May 2016

Chemical inactivation with tri(n-butyl)phosphate–detergent mixtures has been shown to 409

efficiently inactivate enveloped viruses (23, 24), and for some purposes chemical 410

inactivation might be more appropriate than thermal inactivation. 411

Whatever method of inactivation is used, it is important to verify that the treatment does 412

not affect the analyte of interest. For antibodies there is rarely a direct problem; 413

however, heat-treated serum or plasma are well known to cause false reactivity in some 414

assays. For antigen and nucleic acid detection, the risk of affecting the analyte by 415

inactivation is relatively high. Care must be taken to show that the measurement is not 416

affected by the treatment. It must at least be shown that signal and end-point titre from 417

the specimen are not affected, and that both newly manufactured and aged IVD product 418

detect the treated specimen in exactly the same way as they detect the untreated 419

specimen. If the treated specimen is to be used as a control material for more than one 420

IVD, the proof of validity must be documented for each IVD. 421

Dilution 6.1.2422

Most specimens used in QA panels will probably be dilutions of stronger specimens. In 423

such cases, it is important to document that the diluent is appropriate. The diluent must 424

not interfere in the assay in any way and must give a negative signal. In addition, if the 425

diluent is a specimen negative for the analyte of interest, it must be shown not to contain 426

any related antibodies, antigens, polynucleotides or inhibitors, even if it appears not to 427

interact with the IVD under test. When the QA / QC process requires a titration curve 428

(e.g. to monitor sensitivity relative to an international conventional calibrator), the result 429

from the diluent must always be included in the documented measurements. A titration 430

curve in the absence of documentation of the signal from the diluent is invalid. 431

6.2 In-house panels 432

Characterization 6.2.1433

Panel members must be well characterized. Annex 1 of this document provides some 434

examples of characterization for release and stability panels. Normally, specimens are 435

chosen that are non-reactive for infectious agents other than the agent of specific 436

interest, but sometimes this is not possible. If the reagent is reactive for an agent other 437

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

24 Draft for comment 22 May 2017

than that of interest, then it must be proven and documented that the other agents will 438

not affect the primary assay. Where specimens are to be used in routine panels, the 439

process documentation must contain warnings alerting operating staff to the presence of 440

other infectious agents. 441

Verification or validation 6.2.2442

It is incumbent on the manufacturer to develop appropriate QA and QC panels. Risk 443

assessment of each key performance and functionality aspect, and experience with 444

development of the IVD, should indicate which features must be verified at release-to-445

sale, during stability work, in the process of manufacture and for end-of-life security, and 446

which aspects can be validated as routinely fit for purpose and only rarely need to be 447

verified. Efficient development work will minimize the need for verification by providing 448

comprehensive documentary evidence of validation for as many aspects as possible. 449

However, according to regulation or current practice, some functionalities must always be 450

verified at release-to-sale, at least to some extent. The correct, consistent functionality of 451

each critical constituent (enzyme, primer, antibody, antigen, other active biological 452

substances such as protein A and streptavidin, and agents to suppress or monitor false 453

reactivity or immune complex disruption etc.) is normally expected to be verified at 454

release-to-sale. However, the ability to function with each specimen type claimed should 455

be determined unless other strong evidence is available. 456

Sensitivity 6.2.3457

As noted previously, although verification panels for QA at release-to-sale might contain a 458

restricted set of material, the panels used to validate key metrics must be comprehensive 459

and provide the evidence required to allow documentation of the validity. The example 460

that follows illustrates what might be expected for validation of epitope functionality 461

safer. In antibody detection assays, recombinant fusion proteins are frequently used. For 462

example, for syphilis antibody detection, the fusion protein might comprise epitopes from 463

the three treponemal proteins (TpN15, TpN17 and TpN47); for anti-HIV-1 it might 464

comprise epitopes of gp41 and gp120; and for anti-HCV it might comprise epitopes of NS3 465

and core. It must be validated that each of those epitopes will function consistently to the 466

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

25 Draft for comment 22 May 2016

end of the assigned life of the IVD. The development panel will therefore prove that for all 467

lots of the recombinant, in all lots of the complete device, each epitope functions as 468

expected and remains consistently functional throughout stability studies. Testing one lot 469

of recombinant, even in several lots of IVD, is not sufficient to validate the system, 470

because it is known that different purification runs can subtly modify the epitopic 471

efficiency, particularly that of fusion proteins and the various epitopes within them. The 472

development panel used in these studies will therefore specifically and independently 473

monitor the epitopes involved, and show that the activities remain consistent and stable. 474

The R&D department faces the difficult challenge of selecting specimens to provide such 475

proof. Usually, specimens can be found that react strongly by Western blot (or another 476

method) with just one of the epitopes required, and that by dilution can be made 477

essentially specific for that epitope, at least so far as a development panel is concerned 478

( see also Section ‎6.2.5). Once the panel members have been identified (usually 479

throughout the R&D phase of an IVD life cycle) they can be used each time the 480

recombinant is prepared and used. The results, collected and analysed over time, can 481

either support validation of only one of those specimens in the release panel, or show the 482

need for all of them as part of the in-process panels. 483

Prozones 6.2.4484

Prozones, also known as “hook effects” occur when dilution series of specimens with a 485

high concentration of analyte give a maximum signal stronger than the signal from the 486

undiluted specimen. The following relates to assays for which the immunological 487

response profile for a target analyte is not clearly established or in which the analyte is 488

heterogeneous; for example, assays for antibodies or for antigens occurring in multiple 489

forms, such as hepatitis B surface antigen (HBsAg) and histamine rich protein 2 (HRP2) for 490

malaria. Assays for well-characterized, homogeneous analytes (e.g. small metabolites and 491

some hormones) can usually be verified not to prozone with a single artificial specimen 492

containing well above the highest concentration of the analyte found in a clinical setting. 493

The possibility of a prozone depends largely on the format of the assay. If the format 494

chosen is known to prozone, it is imperative to show that the balance of reagents will 495

make prozones rare and prozoning to negative even more rare. Prozones are commonly 496

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

26 Draft for comment 22 May 2017

associated with particle agglutination assays, but can also occur with immunometric 497

assays and with lateral flow devices. Two-step assays – that is, add specimen to capture 498

system, separate (wash), add label, wash – are not susceptible to prozones. However, the 499

normal form of lateral flow device – that is, add specimen to conjugate pad, cause 500

specimen and conjugate together to flow over capture line – will always have the 501

potential to prozone with sufficiently strong specimens. Prozones are very sensitive to the 502

exact concentrations of reagent in the IVD; hence, studies of lot-to-lot variation are 503

necessary. 504

Selection of specimens for the panel to validate consistent lot-to-lot absence of prozone 505

is not difficult, but there are two issues that need to be taken into account. 506

The first issue, as noted, is that of the number of specimens to be chosen. Because 507

prozones are relatively rare, a reasonable number of specimens must be evaluated. The 508

number would need to be obtained from knowledge of the concentration range of 509

analyte likely to be found and a stringent risk analysis – at least 10 specimens would be a 510

suitable minimum. 511

The second issue is the selection of the specimens. It is necessary to choose specimens 512

that have a strong signal and a very high titre in an assay format that cannot prozone. 513

Once the specimens and their numbers have been selected it is a simple matter to test all 514

of them on several lots of the device, to show that there is no increase in signal after 515

dilution of the specimens in a diluent validated not to interfere with the test. Any increase 516

whatever in signal strength for any of the specimens implies a potential prozone and 517

needs further analysis. 518

Rare specimens 6.2.5519

Sometimes it is necessary to validate detection of critical specimens (e.g. seroconversion, 520

rare subtypes or antigen or nucleic acid in immune complexes); however, the genuine 521

specimens might be too valuable to use in lot release-to-sale panels or in stability panels. 522

In such cases, the reactivity or the classes of antibody and epitopic specificity are unlikely 523

to be found in commonly available specimens. Instead, it is necessary to validate 524

surrogate specimens. For immune complex and seroconversion work, near-525

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

27 Draft for comment 22 May 2016

seroconversion specimens can sometimes be shown to mimic the critical individual 526

seroconversion specimens claimed. Such specimens might be available in existing 527

seroconversion series where the later members are not of particular interest in 528

themselves. Immune complexes at later stages of an infection can be important – for 529

example, in HBsAg testing (25) – in that case, specimens can be prepared by dilution of 530

analyte into individual specimens to cover an appropriate range of antibodies (at least ad 531

and ay related, in the HBsAg testing example). For rare subtypes, it might be necessary to 532

prepare surrogates synthetically either biochemically or by evaluating numbers of 533

available specimens until one that mimics the target is found. Further information is given 534

in Example 1 of Annex 1. 535

For instance, a specimen with a rare marker might be tested once with a device and the 536

result recorded. Such a single evaluation does not provide evidence of consistency and it 537

gives no indication of the extent of any lot-to-lot variability that might be present. 538

Therefore, a properly constituted panel is required that can be used on a sufficient 539

number of occasions to prove – using lots that are as varied as possible, and lots that are 540

at the end of their assigned lives – that any variability is within a predetermined 541

acceptable range. 542

Specificity 6.2.6543

Specificity of IVD is largely controlled by the additives in wash solutions and diluents (e.g. 544

detergents, chaotropic agents, cell constituents and masking proteins); monitoring the 545

functioning of such additives is an important role of QA / QC panels. Specificity must be 546

monitored during stability work both for components and for complete devices. 547

It is usual to collect false reactive specimens and interfering specimen types (26) during 548

the R&D phase of device design, and then to monitor and control lot-to-lot variation by 549

using these specimens in release-to-sale panels. When an unusual type of false reactive or 550

interfering substance is identified post-market, either submitted as a complaint or 551

identified in published literature, the specimen (if available in sufficient volume) should 552

be added to release-to-sale panels, and monitored during the statistical process control 553

to ensure that lots do not show excessive reactivity with that specimen. 554

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

28 Draft for comment 22 May 2017

Specimen types 6.2.7555

The claims to usability of different specimen types are usually validated (and 556

documented) in the R&D phase of the IVD life cycle, and verification at lot release is 557

restricted to one or two specific specimen types. This is a particular issue for whole blood 558

specimens with some flow IVD where there is variability between lots and between 559

devices in clearance of the red cell debris from the results window. In general, the whole 560

blood to be tested must be fresh unless otherwise validated, and a sufficient number of 561

individual specimens must be used to verify that the device will operate as expected. 562

Associated materials 6.2.8563

Some materials used in an IVD are not directly regarded as functionality related; for 564

example, antimicrobials, stabilizers for enzymes, agents to quench false reactants and 565

desiccants. Each of these materials must be validated in some way, and a panel of 566

materials, or a chemical method, prepared with which to document acceptable 567

performance. The antimicrobials used in an IVD submitted to WHO prequalification need 568

special consideration, both in terms of the material used (which must be capable of 569

protecting the device against a range of aggressive microorganisms) and stability under 570

harsh conditions. The panels of microorganisms used to validate antimicrobial efficacy 571

can be derived from various pharmacopoeia – for example (27) – and should be used to 572

show efficacy at the end of the assigned shelf life of the IVD. 573

Acceptance criteria 6.2.9574

Acceptance criteria are specific indicators or measures employed in assessing the ability 575

of a component, structure or system to perform its intended function. The manufacturer 576

should be able to justify the rationale for assigning specific acceptance limits when panel 577

samples are being tested at any point throughout the product lifecycle. The rationale can 578

be documented either as part of the risk management process, or as part of the design 579

control documentation when statistical techniques are used as part of the process for 580

establishing performance characteristics. 581

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

29 Draft for comment 22 May 2016

Conclusion 6.2.10582

Selection of specimens for preparation of appropriate panels to verify or support 583

validation of claims for all manufactured lots requires an in-depth knowledge of the 584

particular analyte, the mechanism of the particular assay and the clinical intent of the 585

IVD. Each claim (which will originate from a requirement in the input documentation) and 586

each statement in any documentation of an IVD needs careful consideration as to 587

whether the claim or statement is to be validated or verified for lot release. This 588

consideration needs start early in R&D so that appropriate specimens and procedures can 589

be identified and proper validation performed. 590

6.3 Certified reference materials 591

Certified reference materials (CRM) are available for many IVDs intended to measure 592

analytes that are homogeneous at a molecular level. CRM can frequently be traced to 593

calibrants with an accurate concentration and uncertainty in SI units. Whenever they are 594

available, materials traceable to SI units must be used to validate and verify quantitative 595

IVD. For this type of analyte the production of panels is usually a matter of diluting the 596

pure material into each specimen type required and proving commutability (4). CRM not 597

traceable to SI units used in biological assays are usually international conventional 598

calibrators, as discussed in the sections below. 599

6.4 International conventional calibration materials 600

Many standards from WHO and the Paul Ehrlich Institute (PEI) fall within the category of 601

international calibration materials. Examples include the 3rd International Standard for 602

HBsAg (28), and the 1st International Reference Panel for HBV genotypes for HBsAg-603

based assays (29). Such standards are used for calibration of quantitative assays and for 604

evaluation of analytical sensitivity of qualitative assays. These materials are in too limited 605

supply to use routinely in panels, so it is usual for to prepare manufacturer`s working 606

standards from more readily available specimens, calibrated against the reference 607

materials by methods described in ISO 17511 (4). 608

The 2nd HBsAg WHO international standard is serotype adw2, genotype A, while the 3rd 609

international standard is a mixture of serotypes ayw1 and adw2. Relative reactivity of 610

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

30 Draft for comment 22 May 2017

these standards in HBsAg assays from different manufacturers is different, and so when 611

verifying claims related to this type of standard, it is important to define the version used. 612

The variation can be a result of molecular heterogeneity, which is likely to be detected 613

differently by different antibodies. The same applies for IVDs intended to detect 614

antibodies – if reference materials are changed it is likely that the apparent sensitivity of 615

devices from different manufacturers will also change. 616

6.5 Measurement variation 617

It is expected that any quantitative result will be accompanied by an uncertainty 618

statement (4); therefore, any sensitivity claimed in IFU or assigned to panel members 619

should always be associated with the uncertainty. The uncertainty should take into 620

consideration variation both within and between lots. 621

6.6 External regulatory panels 622

Further details on external regulatory panels are given in Example 6 of Annex 1. Some 623

national agencies and some regulatory authorities provide panels of specimens that the 624

IVD must detect correctly before it can be released-to-sale. These panels cannot control 625

all aspects and all claims about the IVD from all the manufacturers that they regulate 626

because of the variety of reagents in those IVDs and the different specimen types 627

involved. A properly designed regulatory panel might be able to verify some key metrics, 628

but not all that a manufacturer has claimed. For example, some HIV-1 antibody detection 629

systems include only gp41, whereas others include a variety of epitopes from gp41, 630

gp120, gp160 and p24. It is unlikely that a panel provided by a regulator could monitor 631

each of these epitopes in a critical way so as to ensure that the claims related to each 632

could be verified. 633

In the example of the previous paragraph, depending on how the regulator designed and 634

measured the balance of antibodies in a changed panel member, as assay that did not 635

detect anti-p24 could, after a change to a panel member that resulted in reduced anit-636

gp41 but augmented anti-p24, suddenly appear insensitive although it had not changes at 637

all and would continue to perform as claimed. An IVD that detected anti-p24 might 638

appear more sensitive or remain that same on the changed panel member. 639

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

31 Draft for comment 22 May 2016

A different problem might be encountered if correct detection of a regulatory panel were 640

used as the manufacturer’s sole criterion for release-to-sale. Some large national 641

regulatory panels – intended to set a baseline of performance, although each 642

manufacturer’s IVD is unique – set an accuracy requirement of, for example, correct 643

detection of nine out of 10 specimens. Such a criterion is perfectly acceptable if used 644

appropriately; however, if it were used as a sole criterion, a manufacturer could release a 645

lot with an acknowledged 10% false detection rate. 646

Manufacturers must not rely exclusively on external panels to prove compliance but must 647

control products according to their claims while also meeting and using regulatory 648

requirements intelligently. Despite these cautionary comments, from a manufacturer’s 649

point of view, regulatory panels might be useful in maintaining lot-to-lot consistency. 650

For certain high-risk IVD in GHTF class D (10), the European Commission Common 651

Technical Specifications (30: Section 3.4.2) states that, “The manufacturer’s batch release 652

testing for screening assays shall include at least 100 specimens negative for the relevant 653

analyte”, but gives no further comment on choice of specimens or the required result. 654

Presumably, the intent is to detect specificity problems but the (two-sided) 95% 655

confidence interval around zero false reactives in 100 specimens is 0–3.6% false reactive. 656

For the manufacturer of an IVD claiming a typical 0.05% or fewer false reactives, these 657

100 specimens superficially add little value, particularly if the 100 specimens have been 658

pre-selected as negative on that IVD and then stored. Certainly, the results with these 100 659

specimens could not be used as the sole criterion for verification of specificity at release. 660

Again, for a manufacturer, the importance of these types of panels is as part of a 661

performance consistency monitoring programme. 662

For assays with a quantitative output, the 100 specimens might give information if the 663

background were monitored and corrected for minor changes in blank readings. A change 664

in apparent signal for the negative specimens or an increased skewness towards the cut-665

off value might signal that specificity problems might be found with larger numbers of 666

specimens. However, experience also shows that false reactions are likely to be from 667

sporadic high signals, unrelated to the background as a whole but related to impurities in 668

the system, changes in conformations of proteins (leading to new possibility of incorrect 669

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

32 Draft for comment 22 May 2017

epitopic reactions) or changes in the population or individuals tested, such as a 670

vaccination programme (e.g. vaccination against influenza in the autumn). 671

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

33 Draft for comment 22 May 2016

7 Relationship between panel members and claims 672

The particular panel members must fail their criteria if the IVD will not attain the relevant 673

claims at the end of its labelled life. For example, if there is a known change in activity or 674

a drift in signal, the release panel criteria must be set so that each lot will still meet claims 675

at the end of life, despite the drift. Similarly, the end-of-life panel must show that the 676

device has, or has not, met its claims. The signals from the panel members must therefore 677

be correlated with the signals from the critical specimens, providing information on how 678

signals from both the panel member and the critical specimens will change over the life of 679

the IVD. For panel members controlling sensitivity and specificity, the routine R&D work 680

on stability and robustness should have shown the correlation and the change of signal 681

that can be tolerated for as wide a range of conditions as possible; for example, different 682

constituent purification or syntheses, different balances of constituents and various forms 683

of stress (e.g. temperature and humidity) of routine devices. This gives assurance of 684

correlation under conditions that might occur in manufacturing; it also emphasizes the 685

necessity of preparing and validating potential panel members throughout the R&D phase 686

of the IVD life cycle, when these parameters are all accessible. For some components that 687

appear completely stable under routine conditions, extremely stressful conditions must 688

be used to force change of signal in the critical specimens and the panel members to 689

obtain points for the correlation studies (see Examples 1 and 2 in Annex 1). Correlation is 690

the key: the signals from the panel member and the related critical specimen or antigen 691

must follow each other. It is not necessary that the signals be the same, just that the 692

correlation be understood and applied. For example, in many microplate enzyme 693

immunoassays (EIAs) the positive control material is manufactured with an optical density 694

(OD) of around 1.5–2.0, and the criterion in the IFU for a successful assay is >0.8 OD. 695

Thus, the device could have lost more than 50% of its activity and still be declared as 696

efficacious. These same EIA seroconversion series are often claimed have a specimen to 697

cut-off ratio of 1:1. Even if the control material is a dilution of a specimen from a late 698

stage of the infection and bears no relationship to seroconversion, it is unlikely that those 699

seroconversion claims would be upheld if the device had lost more than 50% of its activity 700

on any kind of specimen. It is irrelevant that the device is normally stable and does not 701

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

34 Draft for comment 22 May 2017

lose activity – if the control material is claimed to function at 0.8 OD, then so must the 702

IVD, otherwise the method of validation of the control material would have failed. The 703

same principle applies to all QA / QC panel members. 704

7.1 Process control charts 705

As noted in Section ‎4.2, data from panel members should be made quantitative. Best 706

practice is to plot successive values lot by lot for both release-to-sale and in-process data, 707

generating “process control” charts. There is a large body of literature on how to design 708

and interpret charts to monitor incipient or actual changes in the process, or in the state 709

of the product at release-to-sale, as shown in (31). Such charts for release-to-sale data are 710

expected to be available for inspection (12), given that this is a method of verifying lot-to-711

lot consistency in manufacture, assuming the QA panel used is valid. 712

713

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

35 Draft for comment 22 May 2016

8 Maintenance of panels 714

8.1 Storage 715

Fully characterized and validated specimens are valuable; therefore, appropriate storage 716

is wise. To minimize the possibility of degradation, the stock from which the specimens 717

are to be prepared by dilution is usually split between two locations, and stored frozen at 718

a temperature below its eutectic point (usually between –20 °C and –40 °C), so that liquid 719

water does not occur. Larger stocks of prepared panel members are also usually stored at 720

these temperatures, as are the working size aliquots. Storage of some of the working 721

aliquots at ≤ –70 °C acts as insurance in case of doubt over the stability of those stored at 722

–40 °C. It is unlikely that material stored at both –40 °C and –70 °C would degrade at the 723

same rate; hence, if differences are found, instability could be suspected. Although it is 724

probably impractical to undertake full stability studies on frozen stored specimens, the 725

number of freeze–thaw cycles permissible and the time allowable for storage unfrozen 726

must be documented unless specimens are thawed, used within a short space of time and 727

then discarded. 728

The volume in the stored working aliquots needs to be determined from the rate of use, 729

the stability as a liquid and the validated number of freeze–thaw cycles permissible. 730

Storing in too small a volume in too large a vial should be avoided to minimize any effects 731

from potential freeze drying. 732

The temperature in freezers used for storing panel members must be monitored, ideally 733

electronically, and the temperature record must be retained. Normally, there is an alarm 734

system to alert staff remotely if a freezer shows signs of failure. 735

8.2 Replacement 736

Panel members are key in monitoring and verifying the performance of the IVD and its 737

components; thus, any change must be controlled rigorously. The pre-launch risk 738

evaluation of an IVD should consider the situation that the parent stock of a panel 739

member might not satisfy requirements for the expected commercial life of the product. 740

Factors to be taken into consideration are the likelihood of being able to obtain an 741

equivalent specimen and the extent of characterization already available for the existing 742

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

36 Draft for comment 22 May 2017

specimen. When fresh specimens must be used in QA panels, the risk assessment must 743

also evaluate the validation of the acceptance criteria. As pointed out previously, a 744

thorough, documented understanding of the role of each panel member and of the 745

characteristics required to achieve the role is vital and is an essential part of the 746

development effort of the IVD. This documentation and standard operating procedure 747

(SOP) to control and direct replacement of panel members should be in place from the 748

time the finalized manufacturing documentation of the IVD is prepared (i.e. before 749

verification and validation of the device). 750

Replacement from an existing stock or with well-defined analytes 8.2.1751

Replacement of working vials of panel members is simple if the parent material (e.g. 752

serum pack or manufacturer’s working calibrator (4)) is available. Replacement is simply a 753

matter of dilution into a validated diluent (see Section ‎6.1.2) and assignment of new 754

criteria to the diluted material. To be sure of conformity, it is best to verify that the signal 755

from the newly made panel member is the same as that from a stock previous version, by 756

assaying the new and previous members side by side in sufficient replication and on 757

sufficient lots of the IVD (or component if it is an in-process control), taking into account 758

the issues discussed in the next paragraph. If there are slight differences, a new criterion 759

can usually be applied by proportion of the difference. The new panel members must be 760

prepared under the usual change control system, and it must be possible to trace the 761

change from the identification number of the new working stocks and the records of use 762

of the panels. 763

Changes by comparison are known to lead to drift. It is therefore important to reserve 764

some of the working panel members validated during device development and stored at 765

≤ –70°C for use in the comparisons when panel members are changed, and not merely to 766

compare with the previous version of that panel member. 767

Replacement from a new stock 8.2.2768

Paragraph ‎8.2.1 could apply to replacement of panels of analytes with a well-defined 769

molecular structure (e.g. polynucleotides and some viral proteins), although the parent 770

stock might be exhausted. Even for relatively well-defined entities such as HBV it is clear 771

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

37 Draft for comment 22 May 2016

that the serotype (for HBsAg testing) or genotype (for NAT) must be taken into 772

consideration for replacement panel members, the precise level of identity required must 773

be defined (e.g. ad or ay, ayw or ayr, and ayw1 or ayw3) and the replacement must be 774

shown to correlate with the original in several distinct lots of the IVD. 775

Replacement of panel members from a new parent material is more difficult for 776

heterogeneous analytes. In antibody testing, the spectrum of antibodies (e.g. exact 777

epitopes detected, exact conformation of those epitopes, affinities and composition in Ig 778

classes) varies between specimens, making replacement of like for like tedious. Ideally, in 779

this situation a complete re-validation of the new specimen should be performed, 780

showing correlation with claimed critical specimens for different lots and for accelerated 781

stability. Subject to stringent and documented risk assessment related to product claims 782

and consistency, this might not be necessary; however, it does require detailed 783

consideration of exactly what is required, particularly for specimens intended to control 784

critical epitopes such as in seroconversion to HIV, HCV NS3 or syphilis TpN15. 785

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

38 Draft for comment 22 May 2017

9 References 786

1 ISO 18113–1:2009 In vitro diagnostic medical IVDs – information supplied by the 787 manufacturer (labelling) – Part 1: Terms, definitions and general requirements. Geneva: 788 International Organization for Standardization (ISO); 2009. 789

2 ICH. ICH Harmonised tripartite guideline for elemental impurities Q3D: Current Step 4 790 version. Rockville, MD: International Conference on Harmonisation of Technical 791 Requirements for Registration of Pharmaceuticals for Human Use (ICH); 2014 792 (https://www.ich.org/fileadmin/Public_Web_Site/ICH_Products/Guidelines/Quality/Q3D/793 Q3D_Step_4.pdf, accessed April 2017). 794

3 ISO 15198:2004 Clinical laboratory medicine – in vitro diagnostic medical IVDs – validation 795 of user quality control procedures by the manufacturer. Geneva: International 796 Organization for Standardization (ISO); 2004. 797

4 ISO 17511:2003. In vitro diagnostic medical IVDs – measurement of quantities in biological 798 samples – metrological traceability of values assigned to calibrators and control materials. 799 Geneva: International Organization for Standardization (ISO); 2003. 800

5 HHS U. Code of Federal Regulations Title 21. Sec. 820.3 Definitions. United States 801 Department of Health and Human Services (US HHS); 2016 802 (https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfCFR/CFRSearch.cfm?fr=820.3&Sea803 rchTerm=definitions, accessed April 2017). 804

6 ISO 9000:2015. Quality management systems – fundamentals and vocabulary. Geneva: 805 International Organization for Standardization (ISO); 2015. 806

7 GHTF Steering Committee. GHTF/SG1/N071:2012: Definition of the terms ‘medical device’ 807 and ‘in vitro diagnostic medical device’. Global Harmonization Task Force (GHTF); 2012. 808

8 ISO 14971:2007: Medical IVDs – application of risk management to medical IVDs. Geneva: 809 International Organization for Standardization (ISO); 2007. 810

9 ISO 3534–2:2006 Statistics — vocabulary and symbols — Part 2: Applied statistics. Geneva: 811 International Organization for Standardization (ISO); 2006. 812

10 GHTF Steering Committee. GHTF/SG1/N045: Principles of in vitro diagnostic (IVD) medical 813 devices classification. Global Harmonization Task Force (GHTF); 2008. 814

11 WHO Prequalification of In Vitro Diagnostics Programme. PQDx_018: Instructions for 815 compilation of a product dossier. Geneva: World Health Organization (WHO); 2014 816 (http://www.who.int/diagnostics_laboratory/evaluations/141015_pqdx_018_dossier_instr817 uctions_v4.pdf, accessed April 2017). 818

12 WHO Prequalification of In Vitro Diagnostics Programme. PQDx_014: Information for 819 manufacturers on the manufacturing site(s) inspection (assessment of the quality 820 management system. Geneva: World Health Organization (WHO); 2014 821 (http://www.who.int/diagnostics_laboratory/evaluations/140717_pqdx_014_info_manufa822 cturers_pq_inspections_final.pdf, accessed April 2017). 823

13 ISO 13485:2016: Medical IVDs – quality management systems – requirements for 824 regulatory purposes. Geneva: International Organization for Standardization (ISO); 2016. 825

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

39 Draft for comment 22 May 2016

14 WHO. Technical Guidance Series (TGS): TGS1 – Standards applicable to the WHO 826 prequalification of in vitro diagnostics Geneva: World Health Organization (WHO); 2016 827 (http://www.who.int/diagnostics_laboratory/guidance/technical_guidance_series/en/, 828 accessed April 2017). 829

15 WHO. Technical Guidance Series (TGS): TGS4 – Test method validation for WHO 830 prequalification of in vitro diagnostic devices (draft). Geneva: World Health Organization 831 (WHO); 2016 832 (http://www.who.int/diagnostics_laboratory/guidance/technical_guidance_series/en/, 833 accessed April 2017). 834

16 WHO. In vitro diagnostics and laboratory technology – Guidance and training. Geneva: 835 World Health Organization (WHO); 2017 836 (http://www.who.int/diagnostics_laboratory/guidance/en/, accessed April 2017). 837

17 Magnusson B, Örnemark U, (ed). Eurachem Guide: The fitness for purpose of analytical 838 methods: A laboratory guide to method validation and related topics (2nd edition). 839 Eurachem. 2014 (https://www.eurachem.org/index.php/publications/guides/mv, accessed 840 April 2017). 841

18 CLSI. Evaluation of detection capability for clinical laboratory measurement procedures. 842 CLSI document EP17-A2 (2nd edition). Wayne, PA, Clinical and Laboratory Standards 843 Institute (CLSI). 2012. 844

19 ISO 23640:2011 In vitro diagnostic medical devices – evaluation of stability of in vitro 845 diagnostic reagents. Geneva: International Organization for Standardization (ISO); 2011. 846

20 WHO. Technical Guidance Series (TGS): TGS2 – Establishing stability of an in vitro diagnostic 847 for WHO prequalification (draft). Geneva: World Health Organization (WHO); 2015 848 (http://www.who.int/diagnostics_laboratory/guidance/technical_guidance_series/en/, 849 accessed April 2017). 850

21 Burdick RK, Borror CM, Montgomery DC. Design and analysis of gauge R&R studies: making 851 decisions with confidence intervals in random and mixed ANOVA models. SIAM. 2005. 852

22 Tersmette M, de Goede RE, Over J, de Jonge E, Radema H, Lucas CJ et al. Thermal 853 inactivation of human immunodeficiency virus in lyophilised blood products evaluated by 854 ID50 titrations. Vox Sang. 1986;51(3):239–243 855 (https://www.ncbi.nlm.nih.gov/pubmed/3643679, accessed April 2017). 856

23 Horowitz B, Prince AM, Horowitz MS, Watklevicz C. Viral safety of solvent/detergent 857 treated blood products. Dev Biol Stand. 1993:237–237. 858

24 Piet MP, Chin S, Prince AM, Brotman B, Cundell AM, Horowitz B. The use of tri(n-859 butyl)phosphate detergent mixtures to inactivate hepatitis viruses and human 860 immunodeficiency virus in plasma and plasma's subsequent fractionation. Transfusion. 861 1990;30(7):591–598 (https://www.ncbi.nlm.nih.gov/pubmed/2402772, accessed April 862 2017). 863

25 Brunetto MR. Chance and necessity of simultaneous HBsAg and anti-HBs detection in the 864 serum of chronic HBsAg carriers. J Hepatol. 2014;60(3):473–475. 865

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

40 Draft for comment 22 May 2017

26 CLSI. Interference testing in clinical chemistry. CLSI document EP07-A2. Wayne, PA, Clinical 866 and Laboratory Standards Institute (CLSI). 2005. 867

27 USP-NF. US Pharmacopeial Convention. Rockville, MD: United States Pharmacopeia and 868 National Formulary (USP-NF); 2008 (http://www.usp.org/usp-nf, accessed April 2017). 869

28 NIBSC. 3rd International Standard for HBsAg (HBV genotype B4, HBsAg subtypes 870 ayw1/adw2) – NIBSC code: 12/226. Potters Bar, Hertfordshire: National Institute for 871 Biological Standards and Control (NIBSC). 872

29 Chudy M, Scheiblauer H, Hanschmann K-M, Kress J, Nick S, Wend U et al. Performance of 873 hepatitis B surface antigen tests with the first WHO international hepatitis B virus genotype 874 reference panel. J Clin Virol. 2013;58(1):47–53. 875

30 GHTF Steering Committee. GHTF/SG1/N068:2012: Essential principles of safety and 876 performance of medical devices. Global Harmonization Task Force (GHTF); 2012. 877

31 CLSI. Statistical quality control for quantitative measurement procedures principles and 878 definitions; approved guideline. CLSI document C24-A3 (3rd edition). Wayne, PA, Clinical 879 and Laboratory Standards Institute (CLSI). 2006. 880

881

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

41 Draft for comment 22 May 2016

10 Annex 1 – Examples 882

The anecdotal examples that follow are not comprehensive but are intended to be 883

illustrative. Manufacturers and their R&D groups know their own assays better than 884

anyone else, so they are in the best position to develop and validate their panels. The 885

suggestions in the examples are based on selection of the panels for confirmation of 886

claims but the same principles apply to all method validations; that is, assurance that the 887

test method is providing a meaningful result (e.g. if it is a stability test, that the output 888

really does prove stability relative to claims and not just stability for a particular, possibly 889

irrelevant, spectrum of reactivity). 890

It is regarded as good practice, subject to the evaluation of residual risk before launch of 891

the product, to verify the validated claims on occasion over the life cycle of the IVD. This 892

will be achieved using the QA panels and test methods initially devised. For some 893

attributes this is a requirement (e.g. whole device stability, ISO 14971 (1) and ISO 23640 894

(2)) but for others it is more flexible (e.g. specificity, by active review of data from users). 895

10.1 Example 1: Correlation of a QA / QC panel member with a critical specimen 896

Background 10.1.1897

Critical specimens are likely to be rare and expensive, and our unlikely to be available 898

over the commercial life of an IVD. It might be possible to use critical specimens at all 899

stages of device characterization and at release-to-sale, but in view of the expense and 900

restricted availability of such specimens, validation and documentation of the methods 901

involved in their replacement becomes crucial. Therefore, the critical specimens 902

themselves are not particularly suited for inclusion in QA / QC panels and substitutes 903

must be found. Dilutions of random specimens from late stages of the infection are 904

unlikely to substitute for seroconversion, because the class, the affinity and even the 905

epitopes involved are likely to be different. Similarly, dilutions of specimens from one 906

stage of an infection or specimen type may well not monitor the behaviour of the IVD for 907

a different stage or specimen type. For NAT, the presence of different amounts of analyte 908

in organisms in immune complexes or in perinatal whole blood could present problems 909

hence, throughout the development work, R&D should be searching for readily available 910

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

42 Draft for comment 22 May 2017

specimens that mimic the rare, critical ones, and proving that the specimens chosen do 911

indeed monitor the expected characteristic. Obtaining the correct panel is not an 912

insignificant task, the main thing is that throughout the development of the device, the 913

potential panel member and the critical specimen type should correlate, and should have 914

been tested together whenever possible. Frequently, several specimens must be used 915

and tested every time the critical specimen is tested – this allows the correlation to be 916

obtained and non-correlating specimens to be rejected. Once the correlation is 917

understood, it is possible to establish the criteria for the signal generated by the IVD for 918

that panel member and hence to have evidence of meeting the corresponding claim. 919

The following example is drawn from the development of an EIA for antibodies to 920

Treponema pallidum when it was known (from the input requirements) that a particular 921

specimen in a particular regulator’s collection had to be detected in order to allow that 922

IVD to be sold in various jurisdictions. The cut-off of the EIA had been fixed at 0.15 OD 923

above the mean value of the control negatives and the regulator’s specimen gave a signal 924

of ~ 0.3 OD above the mean value negatives on routine lots. The specimen was only 925

available to IVD manufacturers in limited quantities but it was in the regulator’s own 926

panel (along with other specimens that presented no challenges). After risk evaluation, it 927

was decided that it was necessary to supplement the QA release-to-sale panel with a 928

mimic of this specimen to ensure consistent compliance with the regulator’s requirement 929

because no other specimen in the panel monitored behaviour with this type of specimen. 930

Development of a valid panel member 10.1.2931

R&D characterization of the specimen showed it to have predominantly IgM antibodies to 932

TpN17 (found by in-house Western blots against a panel of recombinant proteins made 933

for the assay’s development). A search through the in-house collection of specimens 934

found only a limited number that had this characteristic. To be sure that the chosen 935

specimen would mimic the regulator’s specimen, it was tested alongside the regulator’s 936

specimen when the IVD was in development, with several different experimental batches 937

of the recombinant fusion protein and with the lots also severely stressed by accelerated 938

stability-like studies. That led to a Passing & Bablok regression (3), as shown in Fig. A1. 939

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

43 Draft for comment 22 May 2016

Fig. A1 The lower values were from stressed IVD, the higher from R&D IVD with varying

concentrations of recombinant from different purifications

Establishing the criteria for the panel member 10.1.3940

a) The regression appeared reasonable and the putative panel member gave about twice 941

the signal from the regulator’s specimen. 942

b) The regulator’s specimen had a signal of about 0.3 OD on routine lots so it was decided 943

to dilute the in-house specimen twofold in negative serum (in order to have a greater 944

volume available). 945

c) Having shown that the dilution did indeed have half the signal of the undiluted 946

material on several lots of the device, a criterion of 0.45 OD was assigned to the panel 947

member in the release specification. That value was calculated from the observation 948

that the panel member as diluted gave a signal of about 0.35 OD with routine lots 949

(which gave about 0.3 OD with the regulator’s specimen) with a standard deviation of 950

about ± 0.02 OD, and to allow for a known loss of activity of about 10% over the 951

assigned life of the IVD. 952

d) For end-of-life testing criteria, this panel member was assigned a value of 0.4 OD. 953

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

44 Draft for comment 22 May 2017

Comment 10.1.4954

This was an exceptional case – it will not always be possible to perform such complete 955

regression studies – but it serves to illustrate the principle. For qualitative IVD or IVD with 956

discrete steps of signal (e.g. when read from a scale of colour with integral values), 957

regression is much more difficult to show. However, important factors are the proof that 958

the panel member varies with the critical specimens it is to monitor, and that the 959

criterion assigned will ensure the claims are met at end of life. 960

10.2 Example 2: Correlation under stress conditions 961

Background 10.2.1962

In Example 1, correlation between a rare, critical specimen and potential panel member 963

was proven using data obtained from both accelerated stability-like stress and changes in 964

recombinant preparation and concentration. Sometimes, routine accelerated stability 965

conditions are insufficient to effect change in the IVD, and it appears to be completely 966

stable. It is still necessary to show that the panel member and specimen will correlate as 967

the IVD is stressed, thus helping to ensure that unforeseen issues in manufacture are 968

likely to be detected (see Section ‎0). 969

This example is of a stable RDT for anti-HIV based on lateral flow and with a recombinant 970

polymeric gp41 as the constituent for HIV-1 detection. The RDT provided a qualitative 971

result but an in-house step-wise graduated colour chart was developed for QA / QC 972

purposes to quantify line strength. The chart was validated both by serial dilution of 973

specimens and by changes to the concentration balances of recombinants used. 974

Independent readers were in agreement with the assigned classification of results of test 975

devices about 90% of the time, and were never more than one grade-step in 976

disagreement, over the whole range of gradation. On this basis it was concluded the 977

colour chart was valid for routine use, with both ratios and differences of grade being 978

meaningful. 979

Detection of critical specimens in a number of commercial seroconversion series was 980

claimed. Between different recombinant purification lots, the appropriate panel members 981

were shown to correlate well with both IgG-first and IgM-first seroconversions (as 982

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

45 Draft for comment 22 May 2016

reflected by class-specific Western blots, relative activity on second generation (IgG only) 983

and third generation (IgG and IgM) commercial assays, and by protein A-based and 984

protein L-based research assay). There appeared to be no loss of activity either with panel 985

members or the critical seroconversion specimens over the claimed life of the assay (24 986

months claimed, 27 months stability data, 4–40 °C, humid or dry storage). Risk 987

assessment found that because activity of the gp41 component of the assay depended 988

critically on the conformation of the recombinant (4, 5), further correlation to include 989

stress conditions would give more assurance that the panel members could detect subtle 990

changes in the recombinant that might not be detected in R&D conditions, but might 991

affect the manufacturing, perhaps in the long term. 992

Extended correlation work 10.2.2993

Knowledge of the IVD suggested that stressful conditions could include storing the IVD 994

out of its protective pouch for various lengths of time under humid conditions at elevated 995

temperatures or by freezing and thawing it several times, again out of its protective 996

pouch. Trial experiments with the putative panel members showed that both these sets 997

of conditions caused loss of sensitivity: the more the stress, the more the loss. Neither of 998

these conditions were likely to occur either with users or in manufacture; nevertheless, 999

they were judged to give additional, different and useful information on the state of the 1000

IVD beyond simply using different lots and concentrations of the recombinant. 1001

It was not practical to test the critical specimens more than once under each condition 1002

because of rarity value, but the panel members could be tested several times at each 1003

condition. The work was done with three lots of the IVD made to final documentation on 1004

routine equipment. Lot A was at the end of its assigned life, whereas Lot B and Lot C were 1005

recently manufactured but from different lots of critical constituents. 1006

Some of the correlation data for the QA panel member (QA 123) and critical 1007

seroconversion specimens 1 and 2 is shown in Fig. A2. The work was done under 1008

isochronous conditions (6: Section 4.3.2), with RDT being taken from routine storage and 1009

placed under stress at different times, starting at 10 hours before testing, then 7 hours 1010

and so on for the data shown, in order that all specimens could be tested in random order 1011

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

46 Draft for comment 22 May 2017

at about the same time. Each number represents the result from an individual test of an 1012

RDT kept under the stated conditions before use. The result was read by two readers 1013

independently and the score relative to the colour scale was recorded for each. 1014

”Correlation” data

Fig. A2. “Correlation” data

A value such as 3.5 indicates that the two independent readers gave different values, but

these were never more than 1 grade different (e.g. 3 and 4 gives 3.5, whereas 1 and 0

gives 0.5).

Comments 10.2.31015

Although the analysis of these data are not statistically rigorous, the general trends can 1016

easily be seen; for example, the panel member was inactivated at about the same rate as 1017

this pair of critical specimens (which happened to be IgG-first seroconversions). Data from 1018

IgM-first seroconversions and the relevant QA / QC panel member and from the 1019

corresponding freeze–thaw experiments were equivalent to these results. 1020

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

47 Draft for comment 22 May 2016

This study was approved by the risk assessment group as sufficient to show satisfactory 1021

relationships between the QA panel members and the critical specimens, when taken 1022

together with the similar data obtained with different lots of the recombinant, as stated 1023

above. Approval to manufacture and release the IVD was given in relation to these panel 1024

members, which were considered to have a low risk of failing to detect non-compliant 1025

product. 1026

The panel member QA 123 was assigned a release-to-sales criterion of ≥3 but ≤5 on the 1027

basis of this and other evidence, and of ≥3 for end-of-life verification testing (for which 1028

the IVD is kept at a constant 40 °C for more than its assigned life). 1029

10.3 Example 3: Validation of a release-to-sale panel for anti-HCV 1030

This example applies both to RDTs with line intensity related to a graduated scale and to 1031

EIAs. 1032

Background 10.3.11033

As noted previously (Section 5.1), a release-to-sale panel must be validated to provide 1034

evidence that the tested lot will meet all the claims for the device for at least the entirety 1035

of its labelled shelf life. The claims will be derived from user input requirements and will 1036

generally include specificity, sensitivity to seroconversion, sensitivity to genotypes, 1037

capability to work on whole blood (for RDTs) and on serum and various plasma types, a 1038

functional internal control and so on. A review of the input documentation and the risk 1039

analyses is important, and is likely to produce a number of other attributes that will be 1040

claimed for the IVD but are not directly performance related (e.g. microbiological stability 1041

and operating temperature ranges). R&D should have developed the device with these 1042

and the other input requirements in mind. During the development, R&D will probably 1043

have noticed that at least NS3 and core antigens from HCV are required, and that some 1044

lots of antigen (especially NS3 1) detect some seroconversion series later than other lots. 1045

They will probably have noticed also that some lots of antigen are less specific than 1046

1 NS3 (proteases and RNA helicase) is a known to be a difficult antigen to use in product development; it seems to include both conformational and linear epitopes, and the exact conditions of recombinant culture and purification are critical to functionality – see, for example, European patent EP 1 471 074 (7) and Mondelli et al. (1994) (8).

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

48 Draft for comment 22 May 2017

others, especially NS5, if that is included in the IVD. Depending on the format of the assay 1047

there might be separate components to detect IgG and IgM. Whole blood is known to 1048

cause flow problems in many IVD, and there should be an input requirement related to 1049

invalid rates; the risk analysis should have reviewed this and R&D should have developed 1050

the device to overcome the problem. A competent design input risk assessment should 1051

have identified a range of other factors related to safe usage and other factors that will 1052

bring “added value” to the users of the device and so bring commercial advantage. 1053

Development of a valid panel 10.3.21054

For release-to-sale panels, not all claims associated with the IVD need to be addressed, 1055

depending on the risk assessment (e.g. microbiological stability, which can be validated 1056

during R&D from antibiotic efficacy and stability studies). For HCV, genotype detection 1057

can probably be validated in R&D and not verified at release. Risk evaluation must occur 1058

continuously in R&D to minimize the amount of work at release by proven and 1059

documented validation of as many factors as possible. 1060

Characteristics of IVD that are known from R&D studies to vary between lots must be 1061

verified at release (e.g. sensitivity, specificity or invalid result rate), and lot variance must 1062

be shown to be within acceptable limits. Also, it is generally accepted that concentrations 1063

and functions of critical constituents (e.g. antigens, antibodies, and biologicals such as 1064

protein A and streptavidin) must be verified at release. The number of specimens to be 1065

tested can be minimized by making appropriate choices, so that each specimen can 1066

monitor the condition of more than one thing (e.g. anti-NS3 first seroconversion and the 1067

NS3 antigen in the system). 1068

Once the claims to be verified have been decided, the specimens used to monitor them 1069

can be selected and evaluated as in Example 1 of Annex 1. A comparison of the published 1070

commercial line-assay results for critical specimens with those from [dilutions of] 1071

potential panel members might help in deciding on the initial choices for an anti-HCV IVD. 1072

For specificity monitoring, it is usual to include known falsely reactive specimens found 1073

during R&D; some falsely reactive specimens are available from the commercial suppliers 1074

for inclusion in QA / QC panels. Methods for monitoring invalid result rates must be 1075

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

49 Draft for comment 22 May 2016

devised and validated on a device-by-device basis because fresh specimens must normally 1076

be used for these studies, particularly if there is a whole blood claim. 1077

Suggested anti-HCV QA / QC panel members with these considerations in mind are given 1078

in the TGS2 guidance document for establishing stability of in vitro diagnostics assays and 1079

components (9). 1080

Establishing the criteria for panel members 10.3.31081

The particular panel members must fail their criteria if the device will not attain any of 1082

the claims at the end of its labelled life. The signals from the panel members must 1083

therefore be correlated with the signals from the critical specimens, as noted in Example 1084

1. For panel members controlling sensitivity and specificity, the routine development, 1085

stability and robustness work should have shown the correlation and the loss (or gain) of 1086

signal that can be tolerated; it should also have shown that the test method is 1087

satisfactorily less variable than the variability between lots of the device itself (see gauge 1088

R&R in Section ‎6.1). 1089

10.4 Example 4: A release panel for a flow cytometer for the enumeration of 1090

CD4 T-cells 1091

Background 10.4.11092

The user input requirements for a quantitative assay generally include: 1093

a specified limit of detection (LoD), with or without a specified limit of 1094

quantitation (LoQ); 1095

a measure of accuracy at clinically relevant thresholds; 1096

(possibly) a specified number of calibrator points, and hence provision of a 1097

calibrator solution) and 1098

(probably) a specified linear range or, more scientifically, a range with a specified 1099

accuracy and precision. 1100

For a CD4+ T-cell enumerating assay, there will almost certainly be a specificity 1101

requirement for CD4+ T-cells but not for other cells that might carry the CD4 determinant 1102

(e.g. monocytes). The design input risk analysis should have identified other well-known 1103

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

50 Draft for comment 22 May 2017

CD4-related problems, including potential analyte-specific interference (e.g. tuberculosis 1104

with its effects on CD4-bearing monocytes and, for some RDTs, malaria with its effects on 1105

red blood cell haemolysis). A competent design input risk assessment should have 1106

identified a range of other factors related to safe usage, and others that will bring “added 1107

value” to the users of the device and so bring commercial advantage. Many of these 1108

factors might lead to claims that require verification at release in addition to evaluation 1109

during device characterization; see, for example, the WHO sample product dosser for CD4 1110

IVD (10). 1111

Development of a valid panel 10.4.21112

For release-to-sale panels, not all claims associated with the IVD need to be addressed, 1113

depending on the risk assessment. For example, monoclonal antibodies from a reputable 1114

source are unlikely to change in avidity. However, device specificity could conceivably 1115

change from lot to lot, depending on the precise purity of the monoclonal antibodies; 1116

also, there could easily be variation between lots in the exact proportions of the 1117

monoclonal antibodies used in manufacture, which could lead to differences in ratios of 1118

cell counts and effects from analyte-specific interference. As noted previously, risk 1119

evaluation must occur continuously in R&D to minimize the amount of work at release, by 1120

proven and documented validation of as much as possible; in particular, any 1121

characteristics of the IVD that are known by R&D studies to be variable between lots must 1122

be verified at release. 1123

Selection of QA / QC panels for an analysis such as CD4+ T-cell enumeration presents 1124

problems because the analyte cannot easily be stored – even short periods of time at 2–1125

8 °C can affect the apparent cell count. It might be possible to design the device so that 1126

blood could be stored, or some form of stabilized or artificial “cells” might be developed. 1127

In each of those cases, the relationship between the behaviour of real blood and the 1128

substitute must be proven rigorously throughout device development. As in the HCV 1129

example, it must be proven and documented that there is satisfactory correlation 1130

between the real specimens and the substitutes. 1131

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

51 Draft for comment 22 May 2016

Where fresh blood is used at release-to-sale, the relationship between the device being 1132

tested and the device used as the calibration method must be validated for all lots of the 1133

tested device. A previous batch of the device is unlikely to adequately meet the 1134

requirements as a validated test method because it could lead to a drift in device 1135

performance. 1136

Establishing the criteria for panel members 10.4.31137

The general comments from the previous examples are applicable; however, for the 1138

quantitative aspects of testing the criteria must be established directly from the 1139

numerical claims (e.g. LoQ: 20 CD4+ T-cells/µL, accuracy: ±25 at 350 CD4+ T-cells/µL). Of 1140

course, the criteria must be set so that the claims will be met at end of life, so the criteria 1141

might well be different from the claim. The criteria for any substitutes for fresh blood 1142

must be shown to correlate with the same attributes in the blood in the same way as the 1143

substitutes for critical specimens in Section ‎10.3.3. 1144

Criteria must be established and documented during R&D for panel members monitoring 1145

claims that are not related directly to quantitation of the analyte (e.g. specificity on 1146

specimens from patients with tuberculosis and invalidity rate on specimens from patients 1147

with malaria). Similarly, criteria must be established and documented for panel members 1148

monitoring appropriate reactivity of the critical, potentially variable, components of the 1149

device. 1150

10.5 Example 5: Nucleic acid testing 1151

Background 10.5.11152

There is a wide variety of NAT methodology, qualitative and quantitative, based on DNA 1153

and RNA, and with intended use ranging from blood donor screening to infant diagnosis. 1154

The main issues relating to the assay concern specimen collection and preparation, 1155

inhibitory substances varying by specimen, stability of the enzyme systems involved and 1156

contamination during manufacture or in use. Occasionally, sub-genotypes present 1157

problems with some systems. Claimed specimen types usually include whole blood, often 1158

in the form of dried blood spots, in addition to buccal fluid, various types of respiratory 1159

samples, cerebrospinal fluid, stool, serum and plasma. Control materials, including 1160

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

52 Draft for comment 22 May 2017

calibrators for quantitative assays, are normally supplied with the IVD. As for all devices, 1161

each of the issues found during design input and process risk management processes 1162

must be dealt with and the methodology developed to minimize any effects. The nature 1163

and type of specimens to control all the hazards must be defined during the R&D phase of 1164

the work. 1165

Development of a valid panel 10.5.21166

As usual, each of the claims that cannot be validated during R&D for all lots over the 1167

commercial life of the device must be verified lot by lot at release-to-sale. In particular, 1168

the panels must demonstrate maintenance of sensitivity or LoD, or accuracy and precision 1169

for all claimed specimen types and all claimed genotypes if the latter cannot be validated 1170

in R&D. There are international conventional calibrators against which in-house QA / QC 1171

panel members can be standardized for most organisms of concern to WHO1 (see also 1172

Section ‎6.4), and the derived secondary standards should be enough to monitor most 1173

claims for both qualitative and quantitative assays. ISO 17511 (14) describes the types of 1174

calibrators and methods for tracing secondary, in-house, standards to internationally 1175

accepted calibrators that are expensive and in restricted supply. For genotype claims, it 1176

might be necessary to refer to collections such as the 1st WHO International Reference 1177

Panel for HBV Genotypes for NAT-Based Assays (15), or the 1st International Reference 1178

Panel HIV-1 RNA Genotypes (16). However, it is likely that the manufacturer will be more 1179

aware of the difficult subtypes from in-house and user testing than might be revealed by 1180

these panels. In that case, critical specimens will probably need to be included in the 1181

panels, at least for in-house validation of attributes such as stability and inhibition. 1182

Whenever in-house panels are made, the correlation under varying conditions with the 1183

internationally accepted materials must be demonstrated, as for any substitute specimen; 1184

proof of ISO 17511 (14) metrological traceability alone is not sufficient. 1185

1 For example, the 3rd HIV 1 International Standard (11), the 4th International Standard for Hepatitis C Virus for Nucleic Acid Amplification Techniques (12) and the Third International Standard for HBsAg (13).

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

53 Draft for comment 22 May 2016

Proof of functionality of extraction, especially from dried blood spots, and removal of 1186

inhibitory substances will almost certainly need to be verified lot by lot. This might 1187

require specimens in the panel or simply chemical analysis of the reagents. 1188

Establishing the criteria for panel members 10.5.31189

Refer to previous examples. 1190

10.6 Example 6: Use of imposed specimens as QA / QC panel members 1191

Background 10.6.11192

Some regulatory authorities require that specimens supplied by them (with criteria also 1193

supplied by them) be detected appropriately before each lot of an IVD can be released-to-1194

sale. Regulators cannot know the claims of every IVD within their purview, so such 1195

specimens cannot act as a valid release-to-sale panel. They are a requirement and, as 1196

such, will be in the design input documentation, but they should not be used as a sole 1197

basis for the manufacturer to verify release-to-sale. 1198

Reasons for not using imposed panels for QA / QC as the main panel members 10.6.21199

See also Section ‎6.6. 1200

a) A regulator’s panel might not control all components. The following are actual 1201

examples. A regulator provided a panel for HCV release-to-sale that contained 1202

virtually no NS3 antibodies, so the lot could still be released even though the NS3 1203

component of a device could be lacking. Another regulator changed an HIV release 1204

panel member from one that contained virtually no anti-p24 to one that contained 1205

high levels of anti-p24, both dilutions of positive specimens. The signals on some 1206

EIAs appeared to be equivalent, but not those on others, and the specimen did not 1207

correlate with seroconversion sensitivity – in fact it contributed little and certainly 1208

did not monitor critical aspects of devices. 1209

b) A panel member might be changed by the regulator, as in point (a), so that 1210

although still being detected appropriately as “positive”, it could allow a change in 1211

assay sensitivity to occur without being detected; alternatively, the device 1212

manufacturer could change his manufacturing to meet the new imposed panel 1213

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

54 Draft for comment 22 May 2017

member acceptance criteria with no control over the influence of the 1214

manufacturing change on specificity or sensitivity to meet his established claims. 1215

c) The requirements on the panel might be such that the device is not controlled 1216

within its claims. For example, if 19 of 20 panel members were required to be 1217

detected as negative, then a lot with about 5% false reactives could be released 1218

despite a claim of a much better specificity. 1219

d) If a lot meets the regulator’s requirements but the manufacturer decides the lot 1220

would not meet its claims and so does not release it, the manufacturer would be 1221

expected to document why the lot was not released as part of its QMS, to 1222

establish an auditable trail. 1223

e) Even international calibrators are not necessarily valid for release of a device 1224

against a sensitivity claim. When the first international HBsAg standard was 1225

replaced by the second, this changed the apparent sensitivity of a number of 1226

assays (17). Claims must always be cited against particular versions of 1227

international conventional calibrators. 1228

f) A manufacturer is responsible for confirming and maintaining his claims over the 1229

commercial life cycle of the IVD, despite external regulatory changes. 1230

Appropriate uses of regulators’ panels by manufacturers 10.6.31231

It might be possible to validate the regulator’s panel as in Section ‎10.3.3, and use it 1232

appropriately. However, the correlation work would be the same as the manufacturer 1233

choosing the specimens and there would be no security against the regulator changing 1234

the panel without notice. 1235

Using the regulator’s panel alongside the manufacturer’s own panel to monitor changes 1236

by routine trend analysis might occasionally be informative – usually about changes by 1237

the regulator. However, panel testing is done independently of the manufacturer by a 1238

regulatory authority. For example, for NAT devices, the panels are designed to test for a 1239

lot’s ability to detect a minimally acceptable number of copies/mL, defined by a 1240

regulatory authority. Any adverse comments on the composition of a regulator’s panel 1241

should always be discussed with that authority. 1242

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

55 Draft for comment 22 May 2016

1243

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

56 Draft for comment 22 May 2017

1 ISO 14971:2007: Medical IVDs – application of risk management to medical IVDs. Geneva: 1244 International Organization for Standardization (ISO); 2007. 1245

2 ISO 23640:2011 In vitro diagnostic medical devices – evaluation of stability of in vitro 1246 diagnostic reagents. Geneva: International Organization for Standardization (ISO); 2011. 1247

3 Passing H, Bablok W. A new biometrical procedure for testing the equality of 1248 measurements from two different analytical methods. Application of linear regression 1249 procedures for method comparison studies in clinical chemistry, Part I. J Clin Chem Clin 1250 Biochem. 1983;21(11):709–720. 1251

4 Pinter A, Honnen WJ, Tilley SA, Bona C, Zaghouani H, Gorny MK et al. Oligomeric structure 1252 of gp41, the transmembrane protein of human immunodeficiency virus type 1. J Virol. 1253 1989;63(6):2674–2679 (https://www.ncbi.nlm.nih.gov/pubmed/2786089, accessed April 1254 2017). 1255

5 Yuan W, Li X, Kasterka M, Gorny MK, Zolla-Pazner S, Sodroski J. Oligomer-specific 1256 conformations of the human immunodeficiency virus (HIV-1) gp41 envelope glycoprotein 1257 ectodomain recognized by human monoclonal antibodies. AIDS Res Hum Retroviruses. 1258 2009;25(3):319–328 (https://www.ncbi.nlm.nih.gov/pubmed/19292593, accessed April 1259 2017). 1260

6 CLSI. Evaluation of stability of in vitro diagnostic reagents; approved guideline. CLSI 1261 document EP25-A. Wayne, PA, Clinical and Laboratory Standards Institute (CLSI). 2009. 1262

7 Innogenetics N.V. Patent EP1471074: Methods for improving the conformation of proteins 1263 by means of reducing agents. 2004 1264 (https://register.epo.org/application?number=EP04103238&tab=main, accessed April 1265 2017). 1266

8 Mondelli MU, Cerino A, Boender P, Oudshoorn P, Middeldorp J, Fipaldini C et al. 1267 Significance of the immune response to a major, conformational B-cell epitope on the 1268 hepatitis C virus NS3 region defined by a human monoclonal antibody. J Virol. 1269 1994;68(8):4829–4836 (https://www.ncbi.nlm.nih.gov/pubmed/7518528, accessed April 1270 2017). 1271

9 WHO. Technical Guidance Series (TGS): TGS2 – Establishing stability of an in vitro diagnostic 1272 for WHO prequalification (draft). Geneva: World Health Organization (WHO); 2015 1273 (http://www.who.int/diagnostics_laboratory/guidance/technical_guidance_series/en/, 1274 accessed April 2017). 1275

10 WHO Prequalification of In Vitro Diagnostics Programme. Sample product dossier for WHO 1276 prequalification. Geneva: World Health Organization (WHO); 2014 1277 (http://www.who.int/diagnostics_laboratory/guidance/160613_simu_poc_cd4_dossier_w1278 eb.pdf?ua=1, accessed April 2017). 1279

11 NIBSC. 3rd International Reference Panel for HIV-1 RNA Genotypes – NIBSC code: 10/152. 1280 Potters Bar, Hertfordshire: National Institute for Biological Standards and Control 1281 (NIBSC);(http://www.nibsc.org/documents/ifu/10-152.pdf, accessed April 2017). 1282

12 NIBSC. 4th International Standard for Hepatitis C Virus for Nucleic Acid Amplification 1283 Techniques – NIBSC code: 06/102. Potters Bar, Hertfordshire: National Institute for 1284

Panels for quality assurance and quality control of in vitro diagnostic medical devices TGS–5

57 Draft for comment 22 May 2016

Biological Standards and Control (NIBSC);(http://www.nibsc.org/documents/ifu/06-1285 102.pdf, accessed April 2017). 1286

13 NIBSC. 3rd International Standard for HBsAg (HBV genotype B4, HBsAg subtypes 1287 ayw1/adw2) – NIBSC code: 12/226. Potters Bar, Hertfordshire: National Institute for 1288 Biological Standards and Control (NIBSC). 1289

14 ISO 17511:2003. In vitro diagnostic medical IVDs – measurement of quantities in biological 1290 samples – metrological traceability of values assigned to calibrators and control materials. 1291 Geneva: International Organization for Standardization (ISO); 2003. 1292

15 WHO. WHO 1st International Reference Panel: Hepatitis B virus genotype panel for NAT-1293 based assays – PEI code 5086/08. 1294

16 NIBSC. 1st International Reference Panel for HIV-1 RNA Genotypes – NIBSC code: 01/466 1295 Potters Bar, Hertfordshire: National Institute for Biological Standards and Control 1296 (NIBSC);(http://www.who.int/bloodproducts/publications/IFU_03-1961.pdf, accessed April 1297 2017). 1298

17 Ferguson M, Health A, Lelie N, Nubling M, Nick S, Gerlich W et al. WHO Working Group on 1299 Hepatitis and HIV Diagnostic Kits: report of a collaborative study to 1) assess the suitability 1300 of candidate replacement international standard for HBsAG and a reference panel for 1301 HBsAG and 2) to calibrate the candidate standard in IU. 1302 2003;(http://apps.who.int/iris/handle/10665/68581, accessed April 2017). 1303

1304