Page 1 of 19 Oxygen Concentrator Procedure Section 1: Investigating Normal Operation Section 2: Determining Cause of Malfunction Appendices A: Using an oxygen sensor B: Using a pressure gauge C: Using a multimeter Warnings Always use proper size and shape of screw driver Note: screws attached to vibrating components might turn in opposite direction. STOP if screwdriver turns and screw does not. Screw will strip and be unusable.

Upload: leliem

Post on 19-Mar-2018

216 views

Category:

Documents


4 download

TRANSCRIPT

Page 1: Oxygen Concentrator Procedure - Wild Apricot. CAUTION: AVOID FANS ON MOTOR AND CENTER OF MACHINE. FANS CAN CAUSE SEVERE INJURY. d. Note minimum and maximum pressures after concentrator

Page 1 of 19  

Oxygen Concentrator Procedure                 

Section 1: Investigating Normal Operation  

Section 2: Determining Cause of Malfunction  

Appendices  

  A: Using an oxygen sensor 

  B: Using a pressure gauge 

  C: Using a multimeter 

 

Warnings  

  Always use proper size and shape of screw driver  

Note: screws attached to vibrating components might turn in opposite                 

direction.  

STOP if screwdriver turns and screw does not.  Screw will strip and be 

unusable.  

 

 

 

 

 

 

 

Page 2: Oxygen Concentrator Procedure - Wild Apricot. CAUTION: AVOID FANS ON MOTOR AND CENTER OF MACHINE. FANS CAN CAUSE SEVERE INJURY. d. Note minimum and maximum pressures after concentrator

Page 2 of 19  

Section 1: Investigating Normal Operation              

In this section you will observe and familiarize yourself with the operation of a working oxygen 

concentrator in order to better locate broken parts of malfunctioning machines.  

Getting started                        

a. Concentrator Specific Safety  a. DO NOT CREATE SPARKS, SMOKE, OR IGNITE FLAMES NEAR OXYGEN 

CONCENTRATORS. High concentrations of oxygen will be leaked into the air.  b. Gather necessary tools  

a. Operation and/or service manuals  b. Screwdriver set  c. Oxygen sensor (Appendix B) d. Flow meter (Appendix A) e. Pressure Gauge (Appendix C) f. Press Gauge Connectors (Female ¼” Thread x 1/8” ID) g. Adjustable wenches   h. Multimeter  

c. Find and Identify Major Components Shown in Pictures Below  (unplug machine and 

remove outside covers)  

 

Page 3: Oxygen Concentrator Procedure - Wild Apricot. CAUTION: AVOID FANS ON MOTOR AND CENTER OF MACHINE. FANS CAN CAUSE SEVERE INJURY. d. Note minimum and maximum pressures after concentrator

Page 3 of 19  

Note: do not touch flow regulator  

 

 

Page 4: Oxygen Concentrator Procedure - Wild Apricot. CAUTION: AVOID FANS ON MOTOR AND CENTER OF MACHINE. FANS CAN CAUSE SEVERE INJURY. d. Note minimum and maximum pressures after concentrator

Page 4 of 19  

 

 

Page 5: Oxygen Concentrator Procedure - Wild Apricot. CAUTION: AVOID FANS ON MOTOR AND CENTER OF MACHINE. FANS CAN CAUSE SEVERE INJURY. d. Note minimum and maximum pressures after concentrator

Page 5 of 19  

 

Note: you will not be able to access this part of the concentrator. Make sure you can 

recognize the parts in the following two pictures.  

Page 6: Oxygen Concentrator Procedure - Wild Apricot. CAUTION: AVOID FANS ON MOTOR AND CENTER OF MACHINE. FANS CAN CAUSE SEVERE INJURY. d. Note minimum and maximum pressures after concentrator

Page 6 of 19  

 

 

 

 

Step 1: Overall Function                    

In this step you will determine if the device release a flow of air when turned on.  

a. Replace all covers  

Page 7: Oxygen Concentrator Procedure - Wild Apricot. CAUTION: AVOID FANS ON MOTOR AND CENTER OF MACHINE. FANS CAN CAUSE SEVERE INJURY. d. Note minimum and maximum pressures after concentrator

Page 7 of 19  

b. Plug in machine 

c. Turn on Oxygen Concentrator  

d. Set both flow rates to 5LMP 

a. Note: measuring ball must sit ON TOP of desired LPM 

e. Make sure:  

a. Alarm sounds for roughly 5 seconds (no longer)  

b. Machine makes reasonable amounts of noise and vibrations  

c. Air can be felt coming out of both outlets  

Step 2: Checking Flow Rate                    

In this step you will determine if the flow rate readings are calibrated correctly.   

a. Set up flow meter  a. Attach barbed output attachment to oxygen output  b. Connect tubing to barbed output attachment  c. Ensure other end of tubing is attached to bottom of flowmeter  d. Read flowmeter in the middle of the ball  e. Note: 0.472LPM=1SCFH 

   b. Plug in machine  c. Turn on concentrator  d. Set both flow rates to 2LPM (4.2 SCFH)  

a. Note: measuring ball must sit in the middle of desired LPM 

b. Ensure set LPM matches meter reading (±1LPM)  

c. Repeat for other oxygen output 

e. Repeat for 5LPM (10.6 SCFH) and 10LPM (21.2 SCFH) 

 

Step 3: Checking the system pressure                  

In this step you will determine if the pressure of the pneumatic system is within the correct 

Page 8: Oxygen Concentrator Procedure - Wild Apricot. CAUTION: AVOID FANS ON MOTOR AND CENTER OF MACHINE. FANS CAN CAUSE SEVERE INJURY. d. Note minimum and maximum pressures after concentrator

Page 8 of 19  

range. The black tubing will test the pressure of the canisters. The blue tubing will test the 

pressure of the air from the compressor.  

a. Turn concentrator off  

b. Depressurize machine  

a. Set both flow rates to maximum  

b. Wait for air to stop flowing from concentrator  

c. Remove right side panel  

d. Locate ¼ in black pressure tubing  

 a. Remove tie and stopper from end of tubing  

b. Connect pressure gauge to black tube  

c. Turn machine on and allow it to run for 5 minutes  

i. CAUTION: AVOID FANS ON MOTOR AND CENTER OF MACHINE. FANS 

CAN CAUSE SEVERE INJURY. 

d. Note minimum and maximum pressures after concentrator has been on for  5 

minutes. Ensure:  

i. Min~0.175MPa 

ii. Max~ 0.220 MPa 

e. turn off machine (do not disconnect pressure gauge yet)  

f. depressurize  

i. set both flow rates to maximum  

ii. wait for air to stop flowing from concentrator  

g. disconnect pressure gauge  

h. replace stopper and tie at end of tubing  

Page 9: Oxygen Concentrator Procedure - Wild Apricot. CAUTION: AVOID FANS ON MOTOR AND CENTER OF MACHINE. FANS CAN CAUSE SEVERE INJURY. d. Note minimum and maximum pressures after concentrator

Page 9 of 19  

e. Locate blue pressure tubing (near black pressure tubing)  

 a. Remove tie from end of tubing 

b. Connect pressure gauge to blue tube  

c. Turn machine on and allow to run for 5 minutes  

i. CAUTION: AVOID FANS ON MOTOR AND CENTER OF MACHINE. FANS 

CAN CAUSE SEVERE INJURY  

d. Ensure maximum pressure reaches 0.047‐0.53MPa 

e. Turn off machine (do not disconnect pressure gauge yet)  

f. Depressurize  

i. set both flow rates to maximum  

ii. wait for air to stop flowing from concentrator 

g. disconnect pressure gauge 

h. Replace tie at end of tubing  

f. Replace outer covers on concentrator   

Step 4: Checking Oxygen Concentration                 

In this step you will determine if the concentrator is producing sufficiently concentrated oxygen 

which will indicate if there are additional problems with the device.  

a. Plug in machine 

b. Turn on Concentrator  

a. Note: machine takes a while to reach full oxygen potential  

c. Set left flow rate to 5LPM, set right flow rate to 0LPM 

Page 10: Oxygen Concentrator Procedure - Wild Apricot. CAUTION: AVOID FANS ON MOTOR AND CENTER OF MACHINE. FANS CAN CAUSE SEVERE INJURY. d. Note minimum and maximum pressures after concentrator

Page 10 of 19  

a. Attach sensor to oxygen output   

b. Ensure concentration is close to 95% (above 85% is acceptable)  

d. Set left flow rate to 0LPM, set right flow rate to 5LMP 

a. Repeat sensor measurement  

Step 5: Checking the Filters and Muffler                 

In this step you will determine if the filters and mufflers are a cause of low flow rates or low 

oxygen concentrations.  

a. Ensure filters are free of debris and don’t appear discolored  

a. Gross particle filter on back cover  

b. Air intake filter  

 c. Air Exhaust filter (you will not be able to check this, but note that there is 

another filter)  

b. Muffler (you will not be able to check this but not that there is a muffler on the exhaust 

to prevent noise)   

Step 6: Checking Power Failure Alarm                   

In this step you will determine if the power failure alarm functions when the device is not 

plugged in.  

Page 11: Oxygen Concentrator Procedure - Wild Apricot. CAUTION: AVOID FANS ON MOTOR AND CENTER OF MACHINE. FANS CAN CAUSE SEVERE INJURY. d. Note minimum and maximum pressures after concentrator

Page 11 of 19  

a. Battery  

a. Plug in machine 

b. Turn machine ON 

c. Ensure: 

i.  Alarm sounds for 5 seconds 

d. Turn machine OFF  

e. Open machine’s shell  

i. Remove Alarm Battery  

ii. Test electrical connection from On/Off Switch to battery  

Step 7: Checking Electrical Connections                  

a. External power is reaching machine 

a. Inspect power cord for wire damage 

b. Ensure circuit breaker is set correctly   

a. Pushed in all the way  

c. Circuit boards and wires  

a. Circuit Boards  

i. Power distribution board 

ii. Power distribution board for fan  

iii. Board containing oxygen sensor  

b. Wires  

i. All visible and accessible wires  

c. Note connectivity of components on all circuit boards and connecting wires for 

later comparison in section 2  

i. Resistors over ~100 ohms should not be continuous 

ii. Resistors under ~100 ohms should be continuous 

iii. Capacitors should not be continuous 

iv. Diodes should not be continuous 

v. Bridge rectifiers should not be continuous 

vi. Wires should be continuous 

d. Pay special attention to oxygen and pressure sensors 

Page 12: Oxygen Concentrator Procedure - Wild Apricot. CAUTION: AVOID FANS ON MOTOR AND CENTER OF MACHINE. FANS CAN CAUSE SEVERE INJURY. d. Note minimum and maximum pressures after concentrator

Page 12 of 19  

i. Oxygen Sensor 

 ii. Pressure sensor   

 d. High Capacitance Capacitor  

a. Directly above motor  

b. IMPORTANT: if you unplug the capacitor you MUST discharge it  

c. Test connections of wires connected to capacitor 

Step 8: Checking Solenoids                      

In this step you will determine if the valves are opening and closing at the appropriate times.  

a. Plug in machine  

b. Open shell  

Page 13: Oxygen Concentrator Procedure - Wild Apricot. CAUTION: AVOID FANS ON MOTOR AND CENTER OF MACHINE. FANS CAN CAUSE SEVERE INJURY. d. Note minimum and maximum pressures after concentrator

Page 13 of 19  

a. CAUTION: AVOID FANS ON MOTOR AND CENTER OF MACHINE. FANS CAN 

CAUSE SEVERE INJURY. 

c. Turn machine on  

d. Observe five LEDs on power distribution board  

 a. Make sure at least 2 LEDs are active at all times  

b. Note:  

i. If any LEDs do not light up 

ii. Which LED(s) is lit if abnormal noise is heard  

e. TURN OFF and UNPLUG the machine  

f. Depressurize  

a. set both flow rates to maximum  

b. wait for air to stop flowing from concentrator 

g. Check resistance of each solenoid  

a. Equalization Valve Solenoid: ~1000‐1100 ohms 

b. Four in the four‐way valve: ~838 ohms  

Step 9: Checking Tubing                    

In this step you will determine if the tubing is secure and not the cause of improper flow rate or 

oxygen concentrations.  

a. Unplug machine  

a. Check for visible damage to tubing and connections  

b. Plug in machine  

c. Open shell  

c. CAUTION: AVOID FANS ON MOTOR AND CENTER OF MACHINE. FANS CAN 

CAUSE SEVERE INJURY. 

d. Turn machine ON 

a. Listen for leaks in tubing or connections  

Page 14: Oxygen Concentrator Procedure - Wild Apricot. CAUTION: AVOID FANS ON MOTOR AND CENTER OF MACHINE. FANS CAN CAUSE SEVERE INJURY. d. Note minimum and maximum pressures after concentrator

Page 14 of 19  

Step 10: Checking the motor                    

In this step you view pictures of the motor and area behind the motor. We will not remove the 

motor because of the fragility of its connections.  

 

Motor  

 

 

Inside of motor  

  

Page 15: Oxygen Concentrator Procedure - Wild Apricot. CAUTION: AVOID FANS ON MOTOR AND CENTER OF MACHINE. FANS CAN CAUSE SEVERE INJURY. d. Note minimum and maximum pressures after concentrator

Page 15 of 19  

Behind motor  

 

 

 

a. Check power connection to motor  

a. Wires going to and from motor  

Step 11: Close and Test                     

In this step you will put the device back together and test to make sure it’s functions are normal.  

a. Close shell  

b. Retest oxygen concentration  

c. Retest flow rate  

d. Retest Alarm  

 

 

 

 

 

 

 

 

 

 

Page 16: Oxygen Concentrator Procedure - Wild Apricot. CAUTION: AVOID FANS ON MOTOR AND CENTER OF MACHINE. FANS CAN CAUSE SEVERE INJURY. d. Note minimum and maximum pressures after concentrator

Page 16 of 19  

Section 2: Determining Cause of Malfunction              

In this section you will troubleshoot and fix malfunctioning devices. Using what was collected from 

Section 1 and the 5 steps of troubleshooting, determine the damaged part(s) of the device. You will 

rotate between three machines with different problems. 

Instructions:  

Read background of the device 

Go through 5 steps of troubleshooting  

Locate damaged part(s)  

Confirm choice with TA 

Replace damaged part(s) 

Reconnect machine  

Check overall function (step 1) 

Check flow rate (step 2) 

Check oxygen concentration (step 3) 

Appendix                          

A. Using an oxygen sensor  

a. Calibrate the sensor  

i. Turn the sensor on 

b. Connected flow diverter (white piece) to top of T adaptor  

c. Connect sensor to flow diverter  

d. Wait for sensor to equilibrate  

e. With nothing connected to the senor press and hold the calibration button (downward 

arrow)  

Page 17: Oxygen Concentrator Procedure - Wild Apricot. CAUTION: AVOID FANS ON MOTOR AND CENTER OF MACHINE. FANS CAN CAUSE SEVERE INJURY. d. Note minimum and maximum pressures after concentrator

Page 17 of 19  

f. Place smaller side of tube against O2 output  

 g. Wait for meter to equilibrate 

B. Using a Pressure Gauge  

a. Buttons  

i. “ON/UNIT/OFF” Button  

1. Press to turn on  

2. Press  to change units   

3. Hold to turn off  

ii. LIGHT/P‐H Button 

1. Press for backlight  

2. Hold to see maximum pressure gauge has been exposed to  

a. Numbers will blink in this setting  

iii. RESET/CLEAR Button  

1. Hold to calibrate the current pressure to 0  

b. Use  

i. Hold down RESET/CLEAR button to calibrate  

ii. Connect appropriate tube adapter to gauge’s port  

iii. Connect tubing end of adapter to tube  

iv. Turn gauge on  

C. Using a Multimeter  

a. To test connectivity   

Page 18: Oxygen Concentrator Procedure - Wild Apricot. CAUTION: AVOID FANS ON MOTOR AND CENTER OF MACHINE. FANS CAN CAUSE SEVERE INJURY. d. Note minimum and maximum pressures after concentrator

Page 18 of 19  

i. Set dial to alarm setting   

ii. Meter will beep if circuit is complete  

iii. Test by touching the red and black probes together  

b. To test voltage  

i. Set meter to AC or DC setting   

ii. Touch probes to opposite sides of voltage being tested  

c. To test Resistance  

Page 19: Oxygen Concentrator Procedure - Wild Apricot. CAUTION: AVOID FANS ON MOTOR AND CENTER OF MACHINE. FANS CAN CAUSE SEVERE INJURY. d. Note minimum and maximum pressures after concentrator

Page 19 of 19  

i. Set meter to ohms   

ii. Touch probes to either side of capacitance being measured  

iii. Note the units if applicable  

D. Using a screwdriver 

a. Make sure the correct size is being used  

b. Apply ample pressure onto the screw 

c. Turn counterclockwise to loosen  

d. Turn clockwise to tighten  

i. Note: screws attached to vibrating components might turn opposite  

e. STOP if screw driver turns and screw DOES NOT 

i. Screw will strip and be unusable