Optimizing Your Absorption or
Fluorescence Thermal Melt Run
![Page 2: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/2.jpg)
Thermal Melt Application
• Typically used to follow a change in signal as temperature is
changed
• Absorption or Fluorescent Intensity can be used to monitor
system as temperature is ramped
• Single Wavelength monitoring is generally done
• Scan of a Spectral Region may be desired
• How can you increase Sample throughput ?
Confidentiality Label
November 6, 2012 2
![Page 3: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/3.jpg)
Cary 100
UV-Vis NIR Spectroscopy Solutions
The Agilent Cary Family
Cary 300
Cary 4000
8453
Cary 5000/6000i
Cary 60
Cary
Eclipse
Perf
orm
ance
Routine to Research
“A word from our sponsors”
![Page 4: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/4.jpg)
Thermal Melts using UV-Vis Absorption
Confidentiality Label
November 6, 2012 4
![Page 5: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/5.jpg)
Instrumentation Needed for a Thermal Melt
Collection
Instrument typically Cary 100 or 300 (8453 is a
possibility as well)
Multi-cell Changer with Temperature Control
Temperature Probe
Purge Gas if Initial or Final Temperature 15 deg C or less
Confidentiality Label
November 6, 2012 5
![Page 6: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/6.jpg)
Typical Thermal Melt Curve
Confidentiality Label
November 6, 2012 6
![Page 7: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/7.jpg)
Thermal Melt Curve Expanded
Confidentiality Label
November 6, 2012 7
![Page 8: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/8.jpg)
SetUp Dialog Box and Setting Up Simple Collection
Confidentiality Label
November 6, 2012 8
![Page 9: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/9.jpg)
Using Multiple Wavelengths in a Single Melt
Confidentiality Label
November 6, 2012 9
![Page 10: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/10.jpg)
User Collect Function
Confidentiality Label
November 6, 2012 10
![Page 11: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/11.jpg)
Typical Simple Single Ramp
• 0.1 degrees/minute
• Collect data every 0.1 degrees or 0.5 degrees
• 20-90 degrees C
• Time Required to Finish Data Collection on a Sample
70 deg / 0.1 deg per min = 700 minutes or 11
hours and 40 min
Confidentiality Label
November 6, 2012 11
![Page 12: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/12.jpg)
Advance Collect using Multiple Temperature
Collection Stages
Confidentiality Label
November 6, 2012 12
![Page 13: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/13.jpg)
Optimizing Data Collection Using Multi-Stage
Ramp • 20 to 50 degrees Ramp at 1.0 deg/min
• 50 to 70 degrees Ramp at 0.1 deg/min
• 70 to 90 degrees Ramp at 1.0 deg/min
• Collect Data every 0.1 deg or 0.5 deg
• Time Required to Finish Data Collection
20 to 50 deg 30 deg/ 1 deg per min = 30 min
50 to 70 deg 20 deg/ 0.1 deg per min = 200 min
70 to 90 deg 20 deg/ 1 deg per min = 20 min
Total Time 250 min or 4 hours and 10 min
Confidentiality Label
November 6, 2012 13
![Page 14: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/14.jpg)
Double Beam SetUp
Confidentiality Label
November 6, 2012 14
![Page 15: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/15.jpg)
Where and When Should the Temperature be
Measured
Choices for Monitoring Temperature
Block
Probe inserted into Cuvette
Can Sacrifice one cuvette for data collection if you do not want sample in
contact with probe
Temperature measured before each measurement
Confidentiality Label
November 6, 2012 15
![Page 16: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/16.jpg)
QNW T2 Single Cell Peltier Accessory
Confidentiality Label
November 6, 2012 16
![Page 17: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/17.jpg)
QNW T2 Temperature Controller
Confidentiality Label
November 6, 2012 17
![Page 18: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/18.jpg)
Temperature Probe In Semi-Micro Cell
Confidentiality Label
November 6, 2012 18
![Page 19: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/19.jpg)
Temperature Lag from Block to Solution Heating
Confidentiality Label
November 6, 2012 19
![Page 20: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/20.jpg)
Temperature Lag from Block to Solution Cooling
Confidentiality Label
November 6, 2012 20
![Page 21: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/21.jpg)
Effect of Signal Averaging Time on Noise
Confidentiality Label
November 6, 2012 21
![Page 22: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/22.jpg)
Additional Capabilities for Collecting Temperature
Related Data
Possible to Collect a Scan at User Define Temperatures
While at each User Defined Temperature, it is possible to
collect a number of scans at a user define time interval
before going to the next temperature
These capabilities are possible through ADL, which is the
controlling language of the software. User have access to
an editor to able to write/edit macros
Confidentiality Label
November 6, 2012 22
![Page 23: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/23.jpg)
Initial Screen after Program Started
Confidentiality Label
November 6, 2012 23
![Page 24: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/24.jpg)
Temperature SetUp for Scans at Temperature
Confidentiality Label
November 6, 2012 24
![Page 25: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/25.jpg)
Trace Preferences Dialogbox Showing Trace
Names
Confidentiality Label
November 6, 2012 25
![Page 26: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/26.jpg)
Scanning at Temperatures
Confidentiality Label
November 6, 2012 26
![Page 27: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/27.jpg)
Thermal Melts using Fluorescence
Confidentiality Label
November 6, 2012 27
![Page 28: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/28.jpg)
Cary Eclipse Fluorescence Spectrometer
The Power of Xenon…
• Unique Xe flashlamp technology
• Measure small volume samples
• Fiber optics
• Room light immunity: unique,
Varian patented technology
• Eliminates photo-degradation
• Long lamp lifetime
Application focus
• Biochemical applications
• Academia
• Industrial chemistry
![Page 29: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/29.jpg)
Fluorescent Thermal Melt Curve
Confidentiality Label
November 6, 2012 29
20 40 60 80
20
40
60
80
100
Temperature (°C)
Inte
nsity (
a.u
.)
DABCYL Labelled PNA and
Fluoreceine Labelled DNA
Two Ramps Each
![Page 30: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/30.jpg)
Simple Single Ramp Collection
Confidentiality Label
November 6, 2012 30
![Page 31: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/31.jpg)
Sample Cell Selection and Temperature Monitoring
Choice
Confidentiality Label
November 6, 2012 31
![Page 32: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/32.jpg)
Advance Collection Choices
Confidentiality Label
November 6, 2012 32
![Page 33: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/33.jpg)
Multiple Wavelength Pairs
Confidentiality Label
November 6, 2012 33
![Page 34: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/34.jpg)
Advance Collection SetUp for Fluorescence
Thermal going from High to Low Intensity
Confidentiality Label
November 6, 2012 34
![Page 35: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/35.jpg)
Example of Fluorescence Thermal Melt going from
High to Low Intensity
Confidentiality Label
November 6, 2012 35
50 60 70 80 90
0
100
200
300
400
Temperature (°C)
Inte
nsity (
a.u
.)
![Page 36: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/36.jpg)
Importance of Choice of Cuvette
Confidentiality Label
November 6, 2012 36
40 60 80
0
20
40
60
80
100
Temperature (°C)
Inte
nsity (
a.u
.)
40 60 80
0
20
40
60
80
100
Temperature (°C)
Inte
nsity (
a.u
.)
Plastic Cuvette Black
Quartz Cuvette Red
Temperature Monitor Block
![Page 37: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/37.jpg)
Importance of Choice of where Temperature is
Monitored
Confidentiality Label
November 6, 2012 37
20 40 60 800
20
40
60
80
100
120
Temperature (°C)
Inte
nsity (
a.u
.)
Quartz Cuvette Block Monitor RedQuartz Cuvette Probe Monitor Purple
![Page 38: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/38.jpg)
Overlay of Cuvette and Temperature Monitor
Choices
Confidentiality Label
November 6, 2012 38
40 60 800
20
40
60
80
100
120
Temperature (°C)
Inte
nsity (
a.u
.)
Black Plastic CuvetteBlue Quartz Cuvette Block as monitor
Red Quartz Cuvette Probe as monitor
![Page 39: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/39.jpg)
Determination of Tm
Confidentiality Label
November 6, 2012 39
40 60 80
0
2
4
6
Temperature (°C)
Deriv1
Ist Derivative Tm Calculation RedOriginal Data Black
![Page 40: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/40.jpg)
Calculated Tm’s
Cuvette Type Calculated Tm (deg C)
Plastic 72.34
Quartz (Block used as Temperature
Monitor)
67.00
Quartz (Probe used as Temperature
Monitor
64.00
Confidentiality Label
November 6, 2012 40
![Page 41: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/41.jpg)
Effect of Noise on Tm Calculation
Confidentiality Label
November 6, 2012 41
40 60 80
0
2
4
6
Temperature (°C)
Deriv1
Tm Calculation with and without Smoothing
Purple Original Temperature Data
Olive Green 1st derivative no smoothing
Red 1st derivative with smoothing
![Page 42: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/42.jpg)
Effect of Noise on Tm Calculation
Cuvette Type and Calculation
Conditions
Calculated Tm (deg C)
Quartz Cuvette using Probe as monitor
no smoothing
64.00
Quartz Cuvette using Probe as monitor
with smoothing
65.00
Confidentiality Label
November 6, 2012 42
![Page 43: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/43.jpg)
Example of Optimization of Temperature Ramp and
Data Interval
Confidentiality Label
November 6, 2012 43
40 60 80
20
40
60
80
100
Temperature (°C)
Inte
nsity (
a.u
.)
40 60 80
50
100
Temperature (°C)
Inte
nsity (
a.u
.)
Probe Monitor Red
Block Monitor Black
![Page 44: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/44.jpg)
Additional Capabilities in Collecting Temperature
Related Data
Like UV-Vis, it is possible to Collect a Scan at a series of
user define temperatures with a user define equilibration
time
While not yet available, it is possible to control the
autopolarizers in addition to the temperature to
automatically collect data for anisotropy calculations for a
set of user define temperatures
Confidentiality Label
November 6, 2012 44
![Page 45: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/45.jpg)
Questions
Confidentiality Label
November 6, 2012 45
![Page 46: Optimizing Your Absorption or Fluorescence Thermal Melt Run](https://reader034.vdocuments.mx/reader034/viewer/2022051412/627ddcaf76cbf311e5540975/html5/thumbnails/46.jpg)
Confidentiality Label
November 6, 2012 46