Optical Coherence Tomography

When Black and White TV Just Isn’t Good Enough

Jeffrey A Southard, MD

UC Davis Medical Center May 5, 2012

Page 2: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

• Consultant for Boston Scientific, Cordis, St Jude

Page 3: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

Imaging

Presenter
Presentation Notes
He will try to convince you that A is as good as B
Page 4: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

What is OCT?

4

Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images of tissue microstructure.

Key Features: Micrometer-level resolution Subsurface imaging High tissue contrast Safe, non-ionizing radiation

Presenter
Presentation Notes
There is no question this provides better images. These are the facts and they are not disputable.
Page 5: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

C7-XR Dragonfly™ Imaging Catheter

Insertable length: 135 cm with

hydrophilic coating

Purge port with 3ml syringe

2.7 Fr tip diameter

(distal 28 cm) with hydrophilic coating Minirail tip compatible with .014”

guidewire

Optical fiber quick-connect

Presenter
Presentation Notes
Pulls back within a hollow tube to prevent injury.
Page 6: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

Intravascular OCT

Flexible fiberoptic catheter used for light delivery

Catheter rotates to create image frames

Catheter pulls back to map vessel segment

Lesion analysis, stent planning, post-stent assessment, follow-up

6

Presenter
Presentation Notes
Mention that blood clearing is required
Page 7: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

Image Generation Measure echo time

delay of reflected light waves

One pixel 5 x 19 um One axial line 1024

pixels One frame 500 axial

lines Optical resolution 15

axial, 20 to 40 um transverse

7

Page 8: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

Pullback Generation

One pullback 270 frames

8

Presenter
Presentation Notes
2.5 seconds
Page 9: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

Intima Media

Adventitia

Normal Coronary Artery

Black and White TV

Page 10: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

Normal, Mild Intimal Thickening

Page 11: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

Calcified Plaque Signal poor (darker) Low backscatter Low attenuation Sharp edges

Circumferential Calcium C

Calcium Nodule

Page 12: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

Calcified Plaque with Rupture

Page 13: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

Lipid Top is bright Body is dark

High attenuation Diffuse edge

L

L

L L

Page 14: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

Fibrous Plaque

14

Homogeneous High backscatter, bright Finely textured Low attenuation

F

F

Page 15: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

Metallic stent struts (BMS & DES)

Bright inner surface reflection

Shadows behind

Stent Images

Page 16: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

F/U DES at 1 Mo F/U BMS

DES compared to BMS in OCT

Page 17: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

Red Thrombus

Page 18: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

Stent

Dissection

Edge Dissection

Page 19: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

Small Intimal Tear

Page 20: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

Zoom Feature (Magnify)

Presenter
Presentation Notes
Another Zoom feature, or magnification, is easily viewed by a simple drag and drop on the region of interest.
Page 21: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images
Presenter
Presentation Notes
What did Jason always say. Too slow, crappy picture. What am I looking at? What are physicians looking for?
Page 22: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

Tissue Penetration

22

Page 23: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

- 61 intermediate stenosis by angiography were assessed by FFR - FFR less than 0.80 was considered significant - OCT vs IVUS on these vessels for anatomic assessment - In vessels less than 3 mm in diameter OCT was superior to IVUS in

identifying functionally significant lesions - MLA for IVUS was 2.42 mm2 - MLD for IVUS was 1.55 mm2

Page 24: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

40 Mhz IVUS

Dissection – OCT vs. IVUS

Page 25: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

FD-OCT vs IVUS

25

Edge dissection during stent implantation

Neointimal growth on previously implanted stent at follow-up

Page 26: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

Stent malapposition: 10atm

Stent Underexpansion & Malapposition: OCT vs. IVUS

Presenter
Presentation Notes
OCT vs IVUS: 30% stent malapposition vs 5% with IVUS. What are we missing?
Page 27: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

Technical Comparison

27

Resolution Axial 15 µm 100 - 200 µm

Transverse 20 – 40 µm 200 - 300 µm Imaging Speed Rotational 100 frames/s 30 frames/s

Pullback 20 mm/s 0.5 - 1 mm/s

Scan Diameter 10 mm 15 mm (var)

Tissue Penetration 1 - 2 mm 6 -10 mm

OC

T

IVU

S

Presenter
Presentation Notes
OCT provides about a 10X improvement in axial and lateral resolution and at least a 20X improvement in pullback speed. The C7 system scans a 50mm arterial segment in 2.5 seconds. It takes approximately 1.5 to 2 seconds of contrast flush (@4ml/sec) to clear the artery of blood prior to the pullback being initiated. So the entire run can be completed in less than 5 seconds!
Page 28: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

What I Can’t Do With IVUS

1. Clearly see what I am doing 2. Image thrombus 3. Determine thin cap fibroatheroma (TCFA)

and hence possible vulnerable plaque 4. Visualize bio-absorbable stents 5. Visualize spontaneous dissection well 6. Determine measurements in small vessels < 3 mm with accuracy 7. Image the vessel in a timely fashion

Page 29: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

29

Page 30: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

Quotes

• “The high resolution OCT images provide a glimpse at structural detail not seen before by other imaging modalities. We believe that OCT has the potential to identify thin-cap fibro-atheromas, clearly highlight intraluminal thrombus, and assess the complete healing of novel drug-eluting stent designs. Volcano is committed to exploring the various clinical benefits of our OCT platform…”

Michael Lussier- President, Volcano Europe 2008

Page 31: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

• "Volcano is committed to being the leader in intravascular imaging…Our investment in this OCT product line underscores that commitment, and opens a series of new clinical applications and opportunities…”

• “a breakthrough that is the result of a 10 year development effort and an investment of over $150 million dollars…”

Scott Huennekens, President and Chief Executive Officer of Volcano 2008, 2009

Page 32: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

Final Thoughts

• IVUS is used more than OCT right now because OCT is new…

• Physicians don’t know how to interpret the OCT images at the present time

• Financial concerns • What do I do with what I find? - I can’t lie to my staff anymore…

Page 33: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

Beer Goggles… OCT vs. IVUS I’m Not Very Sure I Understand What I’m Seeing

Page 34: Optical Coherence Tomography - UC Davis Health · Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images

You Win! OCT is BETTER!