Page | 1 Olympus VS120-L100 Slide Scanner – Standard Operating Procedure Startup 1) Red power bar switch (behind monitor) 2) Computer 3) Login: UserVS120 account (no password) 4) Double click: WAIT FOR INITIALIZATION TO COMPLETE Shutdown 1) Transfer data to external drive and delete from computer 2) Remove ALL slides from cassettes 3) Close VS-ASW FL software 4) Shut down computer 5) Switch off power bar ______________________________________________________________________________________________________________________________________________ MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION CONTENTS BRIGHTFIELD (BF) SLIDE SCANNING SINGLE SLIDE............................................................................................. 2 FLUORESCENCE (FL) slide scanning – Single Slide ............................................................................................ 4 BRIGHTFIELD (BF) slide scanning – Batch Scans .............................................................................................. 7 FLUORESCENCE (FL) slide scanning – Batch Scans ........................................................................................... 9 Troubleshooting .......................................................................................................................................... 11 FILTER SPECS Multi band (Switch channels before moving XY) Single band (Move XY before switching channels) DAPI-FITC-TRITC-Cy5 CFP-YFP-RFP** DAPI-FITC-TRITC-Cy5 Ex (nm) Em (nm) Ex (nm) Em (nm) Ex (nm) Em (nm) DAPI 391/32 435/30 CFP 436/28 474/24 DAPI 375/28 460/50 FITC 479/33 520/30 YFP 506/21 540/24 FITC 480/30 535/40 TRITC 554/24 594/32 RFP 578/24 642/80 TRITC 540/25 605/55 Cy5 638/31 694/60 Cy5 638/31 694/60 Notes: ** Filter change needed, please contact AMP staff (If you require a filter set suited to mCHERRY/Texas Red, use the CFP/YFP/RFP multiband)

Upload: others

Post on 26-Jun-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Olympus VS120 L100 Slide Scanner – Standard Operating ... · 3) Highlight each slide and fill out its properties in the middle column Slide Properties then click Next [ when done

Page | 1

Olympus VS120-L100 Slide Scanner – Standard Operating Procedure

Startup

1) Red power bar switch (behind monitor)

2) Computer

3) Login: UserVS120 account (no password)

4) Double click:

WAIT FOR INITIALIZATION TO COMPLETE

Shutdown

1) Transfer data to external drive and delete

from computer

2) Remove ALL slides from cassettes

3) Close VS-ASW FL software

4) Shut down computer

5) Switch off power bar

______________________________________________________________________________________________________________________________________________

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

CONTENTS

BRIGHTFIELD (BF) SLIDE SCANNING – SINGLE SLIDE............................................................................................. 2

FLUORESCENCE (FL) slide scanning – Single Slide ............................................................................................ 4

BRIGHTFIELD (BF) slide scanning – Batch Scans .............................................................................................. 7

FLUORESCENCE (FL) slide scanning – Batch Scans ........................................................................................... 9

Troubleshooting .......................................................................................................................................... 11

FILTER SPECS

Multi band (Switch channels before moving XY)

Single band (Move XY before switching channels)

DAPI-FITC-TRITC-Cy5 CFP-YFP-RFP** DAPI-FITC-TRITC-Cy5

Ex (nm) Em (nm) Ex (nm) Em (nm) Ex (nm) Em (nm) DAPI 391/32 435/30 CFP 436/28 474/24 DAPI 375/28 460/50

FITC 479/33 520/30 YFP 506/21 540/24 FITC 480/30 535/40

TRITC 554/24 594/32 RFP 578/24 642/80 TRITC 540/25 605/55

Cy5 638/31 694/60 Cy5 638/31 694/60

Notes: ** Filter change needed, please contact AMP staff

(If you require a filter set suited to mCHERRY/Texas Red, use the CFP/YFP/RFP multiband)

Page 2: Olympus VS120 L100 Slide Scanner – Standard Operating ... · 3) Highlight each slide and fill out its properties in the middle column Slide Properties then click Next [ when done

Page | 2

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

BRIGHTFIELD (BF) SLIDE SCANNING – SINGLE SLIDE 1)

a. For first time users, click on the “Scan Projects” tab and choose ‘VS120 – Brightfield’ project. This

will populate the parameters with the most common settings (20x detailed scan, VSI file format).

You can modify settings during the scan setup.

b. For all other users, you may apply your own Scan Project as above or click on:

under Single Slide Scan column.

2) If you have more than one slide, carefully load them into the cassettes as instructed by staff.

3) Choose ‘Manual Load’ or ‘Automatic Load’. If ‘Manual Load’ is chosen, CAREFULLY lift stage access

cover NO MORE THAN 90 DEGREES OR YOU WILL HIT THE CAMERA. Pull back round tab and place

slide sideways/horizontal with label on the LEFT.

4) If ‘Automatic Load’ is chosen, the sensor will automatically detect where your slides are loaded in the

cassettes. Once the sensor has detected the slide positions, choose the slide you wish to scan.

5) Fill out ‘Slide Properties’ left column as needed.

6) Set naming and saving parameters in ‘Automatic Naming and Saving’ middle column.

7) Click ‘Overview’

8) Once overview image is completed, choose objective magnification for detailed scan. ‘Z-Scan Mode’

should be set to ‘Scan one Z-plane’. If you require more than one Z-plane, please see a member of

staff.

9) Click ‘Edit Scan Area’

10) Under ‘Scan Area’ in the left column. You can either click on ‘Create Scan Area’ button to allow

the program to automatically detect and draw Scan Regions around your samples or you can choose

`Remove All Scan Areas` to draw your own. Use mouse wheel to zoom.

11) Under ‘Detailed Scan Settings’, make sure ‘Scan whole area and focus sample only’ (middle button)

option is selected.

12) Next, click on to put down a focus map. Choose the Focus Grid Density in the left

column. You can manually add more Focus points with a mouse click, move existing focus points or

select and delete those not required. NOTE: The more focus points you put down the longer the scan

will take. Flat (fold and bubble free) samples with good contrast do not need a large amount of focus

points while samples with less contrast and/or uneven surfaces require more points.

13) When done click on

14) If ready to proceed, click on ‘Scan Now’. If you require more focusing options, click on ‘Next’

Page 3: Olympus VS120 L100 Slide Scanner – Standard Operating ... · 3) Highlight each slide and fill out its properties in the middle column Slide Properties then click Next [ when done

Page | 3

a. In the ‘Focus Settings’ menu, you may choose the Focus Logic. Options are Automatic

(Default), Semi-Automatic (user confirms automatic focusing at each point) and Manual (user

manually sets each focus point).

b. You can now click on the ‘Scan Now’ button.

15) Once completed you have several options:

a. You may Add another Detailed Scan to the existing slide

b. Scan a new slide with the current parameters

c. Remove/Unload current slide

d. Save Project (HIGHLY RECOMMENDED*)

e. Or finish the experiment.

*To preserve your scan parameters, it is highly recommended that you Save your Project so you do not

have to worry about accidentally using a previous users settings.

16) If not saving a projecty, Click ‘Finish’ and save your scans if you have not done so.

17) To remove the slide, click on and choose ‘Manual Load’.

Page 4: Olympus VS120 L100 Slide Scanner – Standard Operating ... · 3) Highlight each slide and fill out its properties in the middle column Slide Properties then click Next [ when done

Page | 4

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

FLUORESCENCE (FL) slide scanning – Single Slide 1)

a. For first time users, click on the “Scan Projects” tab and choose one of the following options to

populate the parameters with the most common settings (20x detailed scan, VSI file format, typical

exposure etc.). You can modify settings during the scan setup.

i. ‘VS120 – 4 Channel Single Band’. This option uses separate discrete filters for each channel

and moves the stage in XY before switching channels. Less crosstalk but slower. (See page

1 for filter specs)

ii. ‘VS120 – 4 Channel Multi Band’. This option uses a single multiband filter for collecting 4

emission bands simultaneously. Best for speed but more crosstalk. (See page 1 for filter

specs)

iii. ‘VS120 – 3 Channel FP Multiband’. This option uses a single multiband filter for collecting 3

fluorescent protein (CFP/YFP/RFP) emission bands simultaneously. Best for speed but

more crosstalk. FILTER CHANGE NEEDED – SEE MEMBER OF STAFF. (See page 1 for filter

specs)

b. For all other users, you may apply your own Scan Project as above or click on:

under Single Slide Scan column.

2) If you have more than one slide, carefully load them into the cassettes as instructed by staff.

3) Choose ‘Manual Load’ or ‘Automatic Load’. If ‘Manual Load’ is chosen, CAREFULLY lift stage access cover

NO MORE THAN 90 DEGREES OR YOU WILL HIT THE CAMERA. Pull back round tab and place slide

sideways/horizontal with label on the LEFT.

4) If ‘Automatic Load’ is chosen, the sensor will automatically detect where your slides are loaded in the

cassettes. Once the sensor has detected the slide positions, choose the slide you wish to scan.

5) Fill out ‘Slide Properties’ left column as needed.

6) Set naming and saving parameters in ‘Automatic Naming and Saving’ middle column.

7) Click ‘Next’

8) Under ‘Overview Type’, choose the method you want to collect your overview.

a. BF (Brightfield): Quickest method but only if your sample has enough contrast. Click drop down

arrow ↓ to choose Normal, Dim or Faint Sample to correspond to level of contrast.

b. Tissue: To scan a fluorescence overview image on tissue samples. Click drop down arrow ↓ to

choose filter set.

c. Cells: To scan a fluorescence overview image on tissue samples. Click drop down arrow ↓ to

choose filter set.

9) Choose objective lens for overview image (2x is usually sufficient).

10) If you chose option (a) or (b) above, click ‘Start Live’ button under ‘Exposure Time’ column to set exposure.

Use joystick to navigate to sample.

11) Click ‘Overview’ button.

12) Once overview image is completed, choose objective magnification for detailed scan. ‘Z-Scan Mode’ should

be set to ‘Scan one Z-plane’. If you require more than one Z-plane, please see a member of staff. Click

‘Next’ button.

Page 5: Olympus VS120 L100 Slide Scanner – Standard Operating ... · 3) Highlight each slide and fill out its properties in the middle column Slide Properties then click Next [ when done

Page | 5

13) Click ‘Edit Scan Area’

14) Under ‘Scan Area’ in the left column. You can either click on ‘Create Scan Area’ button to allow the

program to automatically detect and draw Scan Regions around your samples or you can choose `Remove

All Scan Areas` to draw your own. Use mouse wheel to zoom.

15) Under ‘Detailed Scan Settings’, make sure ‘Scan whole area and focus sample only’ (middle button) option

is selected.

16) Next, click on to put down a focus map. Choose the Focus Grid Density in the left column.

You can manually add more Focus points with a mouse click, move existing focus points or select and

delete those not required. NOTE: The more focus points you put down the longer the scan will take. Flat

(fold and bubble free) samples with good contrast do not need a large amount of focus points while

samples with less contrast and/or uneven surfaces require more points.

17) Click ‘Next’.

18) If you chose one of the preset Scan Projects, the list of filters under ‘Channel Settings’ should be filled with

the default choices.

19) To add a channel, click and choose to add a single or multiband filter set. To remove a channel, click

.

20) Select Channel 1 (this is the channel that will be used for focusing so it should be your brightest and most

widespread) and click

21) In the Stage Navigator window in the far right, make sure the yellow square is in the middle of your

Detailed Scan area. If not, click inside your region of interest to move the stage.

22) Refer to the histogram in the top right window. Change the Exposure Time so that you achieve at least

1500 counts in the mean intensity reading.

The Orca Flash 4.0 fluorescence camera is a 16-bit camera, allowing for a high dynamic range (65536 bit

depth). It is recommended to use as much of this range as possible (without excessive exposure times

causing photo-bleaching) to capture as much data as possible for potential post-scan analyses.

23) If ready to proceed, click on ‘Scan Now’. If you require more focusing options, click on ‘Next’

24) In the ‘Focus Settings’ menu, you may choose the Focus Logic. Options are Automatic (Default), Semi-

Automatic (user confirms automatic focusing at each point) and Manual (user manually sets each focus

point).

25) You can now click on the ‘Scan Now’.

Page 6: Olympus VS120 L100 Slide Scanner – Standard Operating ... · 3) Highlight each slide and fill out its properties in the middle column Slide Properties then click Next [ when done

Page | 6

26) Once completed you have several options:

a. You may Add another Detailed Scan to the existing slide

b. Scan a new slide with the current parameters

c. Remove/Unload current slide

d. Save Project (HIGHLY RECOMMENDED*)

e. Or finish the experiment.

*To preserve your scan parameters, it is highly recommended that you Save your Project so you do not

have to worry about accidentally using a previous users settings.

27) If not saving a project, Click ‘Finish’ and save your scans if you have not done so.

28) To remove the slide, click on and choose ‘Manual Load’.

Page 7: Olympus VS120 L100 Slide Scanner – Standard Operating ... · 3) Highlight each slide and fill out its properties in the middle column Slide Properties then click Next [ when done

Page | 7

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

BRIGHTFIELD (BF) slide scanning – Batch Scans 1) Carefully load slides into the cassettes starting from cassette #1 (closest to you) before populating

cassette #2 (furthest from you). Take note of your slide descriptions and positions in the cassettes for

your own references. Carefully return cassettes back into the machine in the same order, ensuring that

they are seated correctly.

2) To scan multiple brightfield slides, click on: under Batch Scan column. A sensor will

now detect the positions of your slides within the cassettes. Once complete the next screen shows the

positions of your slides in the cassettes under the ‘Slide Loader’ column. Use the dropdown menu to

choose between cassette #1 and #2.

3) Highlight each slide and fill out its properties in the middle column ‘Slide Properties’ then click ‘Next’

when done. A template can be used to pre-fill these properties. Talk to staff if needed.

4) Under the ‘Slide Loader’ column, you are now able to load one or more previously saved Scan

Project(s) or use the provided ‘VS120 – Brightfield’ project for typical parameters.

5) In the ‘Scan Settings’ column and ‘Magnification and Image Document’ tab, select desired

magnification and Z-scan mode.

6) a) Click the ‘Apply settings to all slides of all cassettes’ button to save these settings to every

slide in every cassette. All slides in the loader should now be highlighted green.

b) If you only want to scan a selection of your slides, hold the Ctrl-key and click the positions in the

cassette to select only the slides you wish to scan. Once the selection is complete, repeat step 5) and

click the ‘Apply settings to selection’ button to save settings to just the selected slides. The

selected slides in the loader should now be highlighted green.

c) If you want to scan different slide selections with different parameters/projects, repeat steps 4 – 6

for each slide selection.

7) Set naming and saving parameters in ‘Automatic Naming and Saving’ column and click ‘Overview’.

8) If for whatever reason you need to scan a single slide NOT loaded in the cassettes, click on ‘Priority

Scan’ to interrupt the batch process.

9) Once overviews are completed, you can review each slide in the ‘Batch Slide Selection’ column. With a

slide selected, click on ‘Edit Scan Area’ button.

10) Under ‘Scan Area’ in the left column. You can either click on ‘Create Scan Area’ button to allow

the program to automatically detect and draw Scan Regions around your samples or you can choose

`Remove All Scan Areas` to draw your own. Use mouse wheel to zoom.

11) Under ‘Detailed Scan Settings’, make sure ‘Scan whole area and focus sample only’ (middle button)

option is selected.

Page 8: Olympus VS120 L100 Slide Scanner – Standard Operating ... · 3) Highlight each slide and fill out its properties in the middle column Slide Properties then click Next [ when done

Page | 8

12) Next, click on to put down a focus map. Choose the Focus Grid Density in the left

column. You can manually add more Focus points with a mouse click, move existing focus points or

select and delete those not required. NOTE: The more focus points you put down the longer the scan

will take. Flat (fold and bubble free) samples with good contrast do not need a large amount of focus

points while samples with less contrast and/or uneven surfaces require more points.

13) When done click on

14) Select next slide in the ‘Batch Slide Selection’ column and repeat steps 10 – 14.

15) If you do not wish to scan a particular slide(s) in the batch, select the slides in the list and click on

‘Exclude slides from the batch scan’ button to exclude from batch scan.

16) If ready to proceed, click on ‘Scan Now’.

Note: If you need to scan a single slide NOT loaded in the cassettes, click on ‘Priority Scan’ to interrupt the

batch process.

Page 9: Olympus VS120 L100 Slide Scanner – Standard Operating ... · 3) Highlight each slide and fill out its properties in the middle column Slide Properties then click Next [ when done

Page | 9

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

FLUORESCENCE (FL) slide scanning – Batch Scans 1) Carefully load slides into the cassettes starting from cassette #1 (closest to you) before populating

cassette #2 (furthest from you). Take note of your slide descriptions and positions in the cassettes for

your own references. Carefully return cassettes back into the machine in the same order, ensuring that

they are seated correctly.

2) To scan multiple fluorescence slides, click on: under Batch Scan column. A sensor

will now detect the positions of your slides within the cassettes. Once complete the next screen shows

the positions of your slides in the cassettes under the ‘Slide Loader’ column. Use the dropdown menu

to choose between cassette #1 and #2.

3) Highlight each slide and fill out its properties in the middle column ‘Slide Properties’ then click ‘Next’

when done. A template can be used to pre-fill these properties. Talk to staff if needed.

4) Under the ‘Slide Loader’ column, you are now able to load one or more previously saved Scan

Project(s) or use the provided default projects (outlined on Page 4 - 1a) for typical parameters.

5) In the ‘Scan Settings’ column and ‘Magnification and Image Document’ tab, select desired

magnification and Z-scan mode. You may also modify the ‘Overview Settings’, ‘Illumination Settings’

and ‘Focus Settings’ tabs. However, you will have a chance to modify these settings later.

6) a) Click the ‘Apply settings to all slides of all cassettes’ button to save these settings to every

slide in every cassette. All slides in the loader should now be highlighted green.

b) If you only want to scan a selection of your slides, hold the Ctrl-key and click the positions in the

cassette to select only the slides you wish to scan. Once the selection is complete, repeat step 5) and

click the ‘Apply settings to selection’ button to save settings to just the selected slides. The

selected slides in the loader should now be highlighted green.

c) If you want to scan different slide selections with different parameters/projects, repeat steps 4 – 6

for each slide selection.

7) Set naming and saving parameters in ‘Automatic Naming and Saving’ column and click ‘Overview’.

8) Once overviews are completed, you can review each slide in the ‘Batch Slide Selection’ column. With a

slide selected, click on ‘Edit Scan Area’ button.

9) Under ‘Scan Area’ in the left column. You can either click on ‘Create Scan Area’ button to allow

the program to automatically detect and draw Scan Regions around your samples or you can choose

`Remove All Scan Areas` to draw your own. Use mouse wheel to zoom.

10) Under ‘Detailed Scan Settings’, make sure ‘Scan whole area and focus sample only’ (middle button)

option is selected.

Page 10: Olympus VS120 L100 Slide Scanner – Standard Operating ... · 3) Highlight each slide and fill out its properties in the middle column Slide Properties then click Next [ when done

Page | 10

11) Next, click on to put down a focus map. Choose the Focus Grid Density in the left

column. You can manually add more Focus points with a mouse click, move existing focus points or

select and delete those not required. NOTE: The more focus points you put down the longer the scan

will take. Flat (fold and bubble free) samples with good contrast do not need a large amount of focus

points while samples with less contrast and/or uneven surfaces require more points.

12) When done click on

13) Select next slide in the ‘Batch Slide Selection’ column and repeat steps 10 – 14.

14) In the ‘Scan Settings’ column, set the parameters required for the ‘Magnification and Image

Document’ tab, ‘Illumination Settings’ tab and ‘Focus Settings’ tab if required.

15) If you do not wish to scan a particular slide(s) in the batch, select the slides in the list and click on

‘Exclude slides from the batch scan’ button to exclude from batch scan.

16) If ready to proceed, click on ‘Scan Now’.

Note: If you need to scan a single slide NOT loaded in the cassettes, click on ‘Priority Scan’ to interrupt the

batch process.

Page 11: Olympus VS120 L100 Slide Scanner – Standard Operating ... · 3) Highlight each slide and fill out its properties in the middle column Slide Properties then click Next [ when done

Page | 11

Troubleshooting Sample not recognized

For weakly stained samples with poor contrast (for example, brightfield slides with weakly staining dyes, very

sparse stained areas OR brightfield overviews of fluorescently stained samples).

During the editing scans dialog you will see the following settings in the left column:

Figure 1.

To increase the threshold of sample detection, modify the Sample detection sensitivity slider. In the example

below (Figure 2) with the slider at ‘0’, you see only a small percentage of the spinal cord tissue sections

detected by the system as indicated by the white, highlighted background. With the slider set to +10 (Figure 3),

almost all the sections are highlighted. Sample detection is also important when creating focus maps on scan

areas. Under ‘Detailed Scan Settings’, if the ‘Scan and focus sample only’ or ‘Scan whole area and focus sample

only’ (first and second button from left) is selected. Focus maps will only be placed on

samples that have been detected (as described above). If your sample is still too faint to be detected, you will

have to choose the 3rd button ‘Scan and focus whole area’ after manually drawing the scan area and/or

manually add focus points.

There are various options in the ‘Sample Detection Settings’ and ‘Advanced Sample Detection Settings’ menu

to assist in fine tuning detection of your samples.

When your samples are successfully highlighted, you can now click the ‘Create Scan Area’ button

which will automatically put a scan area around the detected sections (Figure 4). This is especially useful if you

have many sections on the one slide to avoid manually drawing scan areas.

Page 12: Olympus VS120 L100 Slide Scanner – Standard Operating ... · 3) Highlight each slide and fill out its properties in the middle column Slide Properties then click Next [ when done

Page | 12

Figure 2. Brightfield overview image of fluorescently stained mouse spinal cord cross sections, sample

detection sensitivity set to Zero.

Figure 3. As above with sample detection sensitivity set to +10.

Figure 4. As above after clicking on ‘Create Sample Area’ button.

Page 13: Olympus VS120 L100 Slide Scanner – Standard Operating ... · 3) Highlight each slide and fill out its properties in the middle column Slide Properties then click Next [ when done

Page | 13

If you choose Brightfield overview scan for your fluorescent sample, there are various options in the drop

down arrow menu. Choosing BF (brightfield) you have a choice of Normal, Dim and Faint samples.

Depending on the choice, the program will manipulate the brightness and gamma to improve the contrast of

the samples on your slide, giving you a greater chance of sample detection (mentioned above).

BF – Normal Samples

BF – Dim Samples

BF – Faint Samples

Page 14: Olympus VS120 L100 Slide Scanner – Standard Operating ... · 3) Highlight each slide and fill out its properties in the middle column Slide Properties then click Next [ when done

Page | 14

Focusing Issues

After setting focus maps on your samples and clicking ‘Scan Now’, the system will go to each focus point to set

the focus. If the focus is found, the focus point square will change colour to green. If the focus could not be

found, the focus point square will change to grey. The scan will proceed if there is at least ONE green focus

point square.

Various focusing issues and possible solutions:

1. Samples are weakly stained (either brightfield or fluorescent).

o In the case of weakly stained brightfield samples, go into ‘Live’ mode before proceeding with

the scan to manually set exposure times.

o In the case of weak fluorescent signals, increase the exposure time of the focusing channel

(channel on the top of the list) so the mean intensity on the histogram is at least 1500 counts

(Page 5). Always choose the strongest and most widespread fluorescent channel to be the

focusing channel.

2. Samples are too thick or too thin.

o Thin samples <5 microns will have a smaller volume of fluorescent signal. Ensure that the stage

is close to the focus by going into ‘Live’ mode before proceeding with the scan.

o Thick samples >50 microns will have many points of focus. The algorithm will usually choose

the top of the sample.

3. Stage is too far away from sample or focusing algorithm gets ‘lost’.

o Pertaining to both points 1 and 2, double check in ‘Live’ mode to make sure the stage is at

least close to the focus of your sample.

o Avoid putting focus maps on empty areas where there is no sample. Focus maps on a glass

slide can often bring the stage far from the focus.