nerve stimulator-guided cervical plexus block for carotid endarterectomy
TRANSCRIPT
gauge that had earlier been used on this
patient. We were fortunate that inci-
dent happened in a stable patient
undergoing a routine procedure as the
consequences might, otherwise, have
been much worse.
B. M. Cornforth
Salisbury District Hospital
Salisbury SP2 8CJ, UK
E-mail: [email protected]
National survey of the routinemeasurement of tracheal tubecuff pressure in ICU
Over-inflation of a tracheal tube cuff
can lead to tracheal rupture, ischaemia
and necrosis of the tracheal mucosa,
tracheo-oesophageal fistulas, sore
throat, laryngeal stenosis and tracheo-
carotid artery erosion [1]. Under-infla-
tion can result in micro-aspiration [1]
and increase the risk of nosocomial
infections. Excessive cuff pressures
should be prevented by objective
measurement of the pressure [2]. In
the absence of clear guidelines, many
clinicians consider it reasonable to
maintain cuff pressures in the range of
20–30 cmH2O [1]. We undertook an
audit in our ICU which showed that
66% of cuff pressures measured were
found to lie in the recommended range
of 20–30 cmH2O. Twenty-two per-
cent were inflated to > 30 cmH2O, and
12% were inflated to < 20 cmH2O.
The highest recorded cuff pressure was
70 cmH2O.
Following this audit we conducted a
telephone survey of 79 ICUs across the
UK and asked the following questions:
‘are tracheal tube cuff pressures rou-
tinely measured in your unit’; if yes,
‘how often are measurements done’;
and ‘what is the target cuff pressure in
your unit’? We found that 68% of ICUs
routinely measure tracheal tube cuff
pressures. The frequency of measure-
ment varied from once per day to once
per shift. There was no uniformity in
the target cuff pressure. The target
pressure of some units was the ‘green
zone’, the value of which sometimes
varied with the brand of manometer
used, whereas the range in other units
varied from 14 to 35 cmH2O.
Viyas et al. surveyed 24 ICUs in the
Northern and Yorkshire regions in
2002 and found that 75% did not
routinely measure cuff pressures [3].
The Royal College of Anaesthetists
compendium of audit recipes suggests
that all cuffs should be correctly
inflated, that the correct cuff pressure
is that which is required to prevent a
leak, and that no cuff pressure should
exceed 35 cmH2O [4]. Our audit
revealed that this target is not yet
being met in our ICU and is not met
in at least 32% of ICUs nationally. We
therefore suggest that cuff pressures
should be measured routinely, the
recommendations of the Royal Col-
lege of Anaesthetists about cuff pres-
sures should be followed, and that
there should be a separate a cuff
pressure measuring device for each
critical care bed to avoid cross infec-
tion
N. Burke
R. Baba
A. Moghal
C. Hosahalli Vasappa
Mayday University Hospital
Croydon CR7 7YE, UK
E-mail: [email protected]
References1 Sengupta P, Sessler DI, Maglinger P,
et al. Endotracheal tube cuff pressure in
three hospitals, and the volume
required to produce an appropriate cuff
pressure. BMC Anesthesiology 2004; 4:
8.
2 Stewart SL, Secrest JA, Norwood BR,
Zachary R. A comparison of endotra-
cheal tube cuff pressures using estima-
tion techniques and direct intracuff
measurement. Journal of the American
Association of Nurse Anesthetists 2003; 71:
443–7.
3 Vyas D, Inweregbu K, Pittard A.
Measurement of tracheal tube cuff
pressure in critical care. Anaesthesia
2002; 57: 275–7.
4 Nightingale P, Griffiths H, Clayton J.
Tracheal tube cuff pressures. In: Raising
the Standard: a Compendium of Audit
Recipes for Continuous Quality Improve-
ment in Anaesthesia. London: Royal
College of Anaesthetists, 2006:
204–5.
Nerve stimulator-guidedcervical plexus block for carotidendarterectomy
Regional anaesthesia for carotid artery
surgery allows the patient to remain
awake so that the neurological status
can be assessed during cross-clamping.
However, this technique is unfamiliar
to many anaesthetists and, even in
experienced hands, failure may occur
[1, 2]. We describe a simple modified
technique based on a three-injec-
tion technique as described by Moore
[3].
The patient is placed in the semi-
sitting position with their head turned
slightly away from the side to be
blocked. The transverse processes of
C2, C3 and C4 are located approxi-
mately 1 cm posterior to the posterior
border of the sternomastoid muscle,
and intradermal infiltration of lidocaine
1% 0.25 ml for each level is performed.
The deep cervical plexus block is
performed using a short-bevelled nee-
dle (50 mm-Stimuplex; B Braun, Mel-
sungen, Germany) connected to a
nerve stimulator (Stimuplex DIG, B
Braun), and the needle inserted per-
pendicular to the skin, aiming in a
slightly caudal direction at the C2 level
to elicit neck muscle contractions. The
same technique is repeated at C3 and
C4. The tip of the needle is considered
to be correctly positioned when a
current intensity of 0.5 mA elicits a
neck muscle response. Five ml of local
anaesthetic mixture (bupivacaine 0.5%
and lidocaine 2%) is injected over 2–
3 min after a negative aspiration test.
The technique is completed by per-
forming a superficial cervical plexus
block by infiltration at the midpoint of
the sternomastoid muscle with 7 ml of
the same mixture and infiltration of 3–
5 ml of local anaesthetic mixture along
the inferior border of the mandible to
block the afferent branches from the
cranial nerves. This injection along the
mandible appears to reduce the pain
associated with prolonged use of a
retractor. We have found that patients
are rarely distressed or uncomfortable
during performance of these blocks.
We have obtained excellent results
Anaesthesia, 2007, 62, pages 289–300 Correspondence......................................................................................................................................................................................................................
� 2007 The Association of Anaesthetists of Great Britain and Ireland 299
with this technique in a number of
patients and further evaluation is
ongoing.
The use of a nerve stimulator for
deep cervical plexus blockade has been
previously reported for carotid surgery.
Mehta and Juneja [4] and Merle et al.
[5] used a single-injection technique,
guided by a nerve stimulator. In both
reports the technique was not com-
pletely successful and required supple-
mental intravenous analgesia or local
anaesthetic infiltration, particularly dur-
ing retractor placement and carotid
artery dissection. In our technique, the
identification of nerves of the cervical
plexus was more precise, requiring neck
muscle contraction prior to each of the
three injections.
Phrenic nerve palsy is frequent (up to
90%) after deep cervical plexus block.
The use of a nerve stimulator can elicit a
diaphragmatic muscle response which
helps to avoid the administration of the
local anaesthetic directly into the area of
the phrenic nerve. It has been reported
that the use of a nerve stimulator
decreases the peak serum concentration
(Cmax) and significantly slows the time
to reach peak concentration (Tmax) of
the local anaesthetic [5], both of which
are major determinants of systemic
toxicity.
A. Zeidan
F. Hayek
Sahel General Hospital
Beirut, Lebanon
E-mail: [email protected]
References1 Lee KS, Davis CH, McWhorter JM.
Low morbidity and mortality of carotid
endarterectomy performed with re-
gional anesthesia. Journal of Neurosurgery
1988; 69: 483–7.
2 Murie JA, Morris PJ. Carotid endar-
terectomy in Great Britain and Ireland.
British Journal of Surgery 1986; 73: 867–
70.
3 Moore DC. Regional Block: a Hand-
book for the Use in the Clinical Practice
of Medecine and Surgery, 4th edn.
Springfield: Charles C Thomas, 1978:
21–4.
4 Mehta Y, Juneja R. Regional analgesia
for carotid artery endarterectomy by
Winnie’s single-injection technique
using a nerve detector. Journal of
Cardiothoracic and Vascular Anesthesia
1992; 6: 772–3.
5 Merle JC, Mazoit JX, Desgranges P,
et al. A comparison of two techniques
for cervical plexus blockade: evaluation
of efficacy and systemic toxicity.
Anesthesia and Analgesia 1999; 89:
1366–70.
Rebreathing due to incorrectlyfitted seal on an Aestiva S/5
An Aestiva S ⁄ 5 anaesthesia work station
(Datex Ohmeda, GE Healthcare, Hat-
field, UK) displayed a high carbon
dioxide reading (consistent with
rebreathing) during routine use. Fol-
lowing a detailed search the fault was
found. The seal on the carbon dioxide
bypass unit (whose function is to allow
removal and refilling of the soda lime
canister during use) was incorrectly
fitted inside out (Fig. 5), causing respir-
atory gases to bypass the soda lime
canister. We have been unable to
establish how this occurred. The
rebreathing resolved following correct
positioning of the seal. A correctly
placed seal fits flush with the metal
components of the bypass unit (Fig. 6).
When it is positioned ‘inside out’ the
soda lime canister still appears to fit.
However, significant rebreathing occurs
at low fresh gas flows.
J. Phillips
K. Jones
N. S. Parekh
New Cross Hospital
Wolverhampton WV10 0QP, UK
E-mail: [email protected]
Figure 5 Seal incorrectly fitted (inside out).
Figure 6 Seal correctly fitted.
Correspondence Anaesthesia, 2007, 62, pages 289–300......................................................................................................................................................................................................................
300 � 2007 The Association of Anaesthetists of Great Britain and Ireland