NEL’s EPAT Facility –
Experiences,
Challenges and Future
Developments
Chris Mills, NEL
25 April 2018
![Page 2: NEL’s EPAT Facility - tuv-sud.co.uk€¦ · –CTSp and CPSp both used –CTLp and CPLp (CTPL) are not used –API values are not robust for NEL test fluid V BPV CTSp CPSp mmp prove](https://reader031.vdocuments.mx/reader031/viewer/2022011802/5b0dba477f8b9a02508e2fe6/html5/thumbnails/2.jpg)
– INTRODUCTION
– EPAT FACILITY
– CHALLENGES
– EXPERIENCES
– FUTURE DEVELOPMENTS
Content
![Page 3: NEL’s EPAT Facility - tuv-sud.co.uk€¦ · –CTSp and CPSp both used –CTLp and CPLp (CTPL) are not used –API values are not robust for NEL test fluid V BPV CTSp CPSp mmp prove](https://reader031.vdocuments.mx/reader031/viewer/2022011802/5b0dba477f8b9a02508e2fe6/html5/thumbnails/3.jpg)
Brief history
• 2011: NEL completed R&D with Coriolis meters for a number of O&G
operators at elevated temperatures.
• 2012: NEL formulated a Joint Industrial Project (JIP) exploring influence of T,
P, and viscosity.
• 2014: JIP completed and concluded that:
– calibration under conditions similar to the field required
– lack of traceable calibration facilities that can operate at elevated
temperature, pressure, and viscosity
• 2014 – 2016 NEL designed & built a new flow loop operating at Elevated P
and T (EPAT) & Fully accredited to UKAS
• 2016 - present: This presentation focus now on Experiences, Challenges
and Future Developments
Introduction
![Page 4: NEL’s EPAT Facility - tuv-sud.co.uk€¦ · –CTSp and CPSp both used –CTLp and CPLp (CTPL) are not used –API values are not robust for NEL test fluid V BPV CTSp CPSp mmp prove](https://reader031.vdocuments.mx/reader031/viewer/2022011802/5b0dba477f8b9a02508e2fe6/html5/thumbnails/4.jpg)
Current regulations
• UK Oil & Gas Authority (OGA) Guidelines
Introduction
![Page 5: NEL’s EPAT Facility - tuv-sud.co.uk€¦ · –CTSp and CPSp both used –CTLp and CPLp (CTPL) are not used –API values are not robust for NEL test fluid V BPV CTSp CPSp mmp prove](https://reader031.vdocuments.mx/reader031/viewer/2022011802/5b0dba477f8b9a02508e2fe6/html5/thumbnails/5.jpg)
The facility:
EPAT Facilities
![Page 6: NEL’s EPAT Facility - tuv-sud.co.uk€¦ · –CTSp and CPSp both used –CTLp and CPLp (CTPL) are not used –API values are not robust for NEL test fluid V BPV CTSp CPSp mmp prove](https://reader031.vdocuments.mx/reader031/viewer/2022011802/5b0dba477f8b9a02508e2fe6/html5/thumbnails/6.jpg)
EPAT Facilities
![Page 7: NEL’s EPAT Facility - tuv-sud.co.uk€¦ · –CTSp and CPSp both used –CTLp and CPLp (CTPL) are not used –API values are not robust for NEL test fluid V BPV CTSp CPSp mmp prove](https://reader031.vdocuments.mx/reader031/viewer/2022011802/5b0dba477f8b9a02508e2fe6/html5/thumbnails/7.jpg)
Specification:
Item Spec
Maximum Pressure: 100 bar g
Minimum Pressure: 4 bar g
Pressure Control: +/- 0.5 Bar
Maximum Temperature: 80 °C
Minimum Temperature: 20 °C
Temperature Control: +/- 0.5 °C
Maximum Flow: 100 l/s
Minimum Flow: 0.5 l/s
Fluid Density: 0.810 kg/l
Maximum Fluid Viscosity: 6 cP
Minimum Fluid Viscosity: 2 cP
Test Section Pipe Size: 1 – 10 inch OD
Test Section Pipe Length: 10 metres
Measurement Uncertainty: ± 0.08 (k=2)
EPAT Facilities
![Page 8: NEL’s EPAT Facility - tuv-sud.co.uk€¦ · –CTSp and CPSp both used –CTLp and CPLp (CTPL) are not used –API values are not robust for NEL test fluid V BPV CTSp CPSp mmp prove](https://reader031.vdocuments.mx/reader031/viewer/2022011802/5b0dba477f8b9a02508e2fe6/html5/thumbnails/8.jpg)
EPAT Facilities
![Page 9: NEL’s EPAT Facility - tuv-sud.co.uk€¦ · –CTSp and CPSp both used –CTLp and CPLp (CTPL) are not used –API values are not robust for NEL test fluid V BPV CTSp CPSp mmp prove](https://reader031.vdocuments.mx/reader031/viewer/2022011802/5b0dba477f8b9a02508e2fe6/html5/thumbnails/9.jpg)
Operation & calculations
• DUT calibrated against Turbine Master Meters
– Master Meters are proved at every flowrate, pressure and temperature
– Typical sequence – (Prove, 3 DUT Repeats, Prove)
• Calculations
– CTSp and CPSp both used
– CTLp and CPLp (CTPL) are not used
CPSpCTSpBPVCTPL
CTPLV
mmp
provemmp
EPAT Facilities
![Page 10: NEL’s EPAT Facility - tuv-sud.co.uk€¦ · –CTSp and CPSp both used –CTLp and CPLp (CTPL) are not used –API values are not robust for NEL test fluid V BPV CTSp CPSp mmp prove](https://reader031.vdocuments.mx/reader031/viewer/2022011802/5b0dba477f8b9a02508e2fe6/html5/thumbnails/10.jpg)
Operation & calculations
• DUT calibrated against Turbine Master Meters
– Master Meters are proved at every flowrate, pressure and temperature
– Typical sequence – (Prove, 3 DUT Repeats, Prove)
• Calculations
– CTSp and CPSp both used
– CTLp and CPLp (CTPL) are not used
– API values are not robust for NEL test fluid
CPSpCTSpBPVVmmp
provemmp
CPSpCTSpBPVCTPL
CTPLV
mmp
provemmp
EPAT Facilities
![Page 11: NEL’s EPAT Facility - tuv-sud.co.uk€¦ · –CTSp and CPSp both used –CTLp and CPLp (CTPL) are not used –API values are not robust for NEL test fluid V BPV CTSp CPSp mmp prove](https://reader031.vdocuments.mx/reader031/viewer/2022011802/5b0dba477f8b9a02508e2fe6/html5/thumbnails/11.jpg)
Prover calculations
– API values are not robust for NEL test fluid
Error may be up to 0.1 % up to 49 bar(a), undefined above, [Glen N., O&G FG,
Dec 16]
Challenges
![Page 12: NEL’s EPAT Facility - tuv-sud.co.uk€¦ · –CTSp and CPSp both used –CTLp and CPLp (CTPL) are not used –API values are not robust for NEL test fluid V BPV CTSp CPSp mmp prove](https://reader031.vdocuments.mx/reader031/viewer/2022011802/5b0dba477f8b9a02508e2fe6/html5/thumbnails/12.jpg)
Density characterisation
– API values are not robust for NEL test fluid
– Requires fluid to be characterised at T&P
– Stable density is critical for facility uncertainty
– Water contamination was an issue
CPSpCTSpBPVVmmp
provemmp
Challenges
![Page 13: NEL’s EPAT Facility - tuv-sud.co.uk€¦ · –CTSp and CPSp both used –CTLp and CPLp (CTPL) are not used –API values are not robust for NEL test fluid V BPV CTSp CPSp mmp prove](https://reader031.vdocuments.mx/reader031/viewer/2022011802/5b0dba477f8b9a02508e2fe6/html5/thumbnails/13.jpg)
Demonstrating equivalence
• Intra-comparison
– Compared EPAT system with NEL gravimetric for low pressure and temperature
– Achieved equivalence but only at low P&T conditions.
Challenges
![Page 14: NEL’s EPAT Facility - tuv-sud.co.uk€¦ · –CTSp and CPSp both used –CTLp and CPLp (CTPL) are not used –API values are not robust for NEL test fluid V BPV CTSp CPSp mmp prove](https://reader031.vdocuments.mx/reader031/viewer/2022011802/5b0dba477f8b9a02508e2fe6/html5/thumbnails/14.jpg)
Demonstrating equivalence
-0.75
-0.50
-0.25
0.00
0.25
0.50
0.75
0.0 2.5 5.0 7.5 10.0 12.5 15.0
%E
rr (R
ef.
Ma
ss
)
Ref. Total MFlow , kg/s
Temperature 20°C, Pressure 5 bar.g
Gravimetric Compact Prover
Challenges
![Page 15: NEL’s EPAT Facility - tuv-sud.co.uk€¦ · –CTSp and CPSp both used –CTLp and CPLp (CTPL) are not used –API values are not robust for NEL test fluid V BPV CTSp CPSp mmp prove](https://reader031.vdocuments.mx/reader031/viewer/2022011802/5b0dba477f8b9a02508e2fe6/html5/thumbnails/15.jpg)
Demonstrating equivalence
• Inter-comparison
– Compared EPAT system with West Lab for high pressure
– Achieved equivalence with manufacturer supplied meter
– Additional customer supplied data allowed comparisons at T&P to be made
Challenges
![Page 16: NEL’s EPAT Facility - tuv-sud.co.uk€¦ · –CTSp and CPSp both used –CTLp and CPLp (CTPL) are not used –API values are not robust for NEL test fluid V BPV CTSp CPSp mmp prove](https://reader031.vdocuments.mx/reader031/viewer/2022011802/5b0dba477f8b9a02508e2fe6/html5/thumbnails/16.jpg)
-0.75
-0.50
-0.25
0.00
0.25
0.50
0.75
0 10 20 30 40 50 60 70 80 90 100
%E
rr (R
ef.
Ma
ss
)
Ref. Total MFlow , Tonne/hr
Temperature 68°C, Pressure 14 bar.g
NEL West Lab
-0.75
-0.50
-0.25
0.00
0.25
0.50
0.75
0 250 500 750 1000
%E
rr (R
ef.
Ma
ss
)
Ref. Total MFlow , kg/s
NEL 10 bar.g NEL 40 bar.g West Lab 10 bar.g West Lab 40 bar.g
Demonstrating equivalence
Challenges
![Page 17: NEL’s EPAT Facility - tuv-sud.co.uk€¦ · –CTSp and CPSp both used –CTLp and CPLp (CTPL) are not used –API values are not robust for NEL test fluid V BPV CTSp CPSp mmp prove](https://reader031.vdocuments.mx/reader031/viewer/2022011802/5b0dba477f8b9a02508e2fe6/html5/thumbnails/17.jpg)
Demonstrating equivalence
-0.75
-0.50
-0.25
0.00
0.25
0.50
0.75
0 10 20 30 40 50 60 70 80 90 100
%E
rr (R
ef.
Ma
ss
)
Ref. Total MFlow , Tonne/hr
Temperature 68°C, Pressure 14 bar.g
NEL West Lab
Challenges
![Page 18: NEL’s EPAT Facility - tuv-sud.co.uk€¦ · –CTSp and CPSp both used –CTLp and CPLp (CTPL) are not used –API values are not robust for NEL test fluid V BPV CTSp CPSp mmp prove](https://reader031.vdocuments.mx/reader031/viewer/2022011802/5b0dba477f8b9a02508e2fe6/html5/thumbnails/18.jpg)
Customer data
• Calibrated large number of Coriolis meters
– All manufacturers
– Variety of sizes and models
Experiences
![Page 19: NEL’s EPAT Facility - tuv-sud.co.uk€¦ · –CTSp and CPSp both used –CTLp and CPLp (CTPL) are not used –API values are not robust for NEL test fluid V BPV CTSp CPSp mmp prove](https://reader031.vdocuments.mx/reader031/viewer/2022011802/5b0dba477f8b9a02508e2fe6/html5/thumbnails/19.jpg)
-2.00
-1.75
-1.50
-1.25
-1.00
-0.75
-0.50
-0.25
0.00
0 50 100 150 200 250 300 350
%E
rr (R
ef.
Ma
ss
)
Ref. Mass Flow, T/hr
Kerosene 20degC 3cSt Gas Oil 20degC 8cSt Gas Oil 20degC 8cSt (R) EPAT 25degC 10bar.g
EPAT25degC 20bar.g EPAT 25degC 20bar.g R EPAT 25degC 40bar.g
Customer data
Experiences
![Page 20: NEL’s EPAT Facility - tuv-sud.co.uk€¦ · –CTSp and CPSp both used –CTLp and CPLp (CTPL) are not used –API values are not robust for NEL test fluid V BPV CTSp CPSp mmp prove](https://reader031.vdocuments.mx/reader031/viewer/2022011802/5b0dba477f8b9a02508e2fe6/html5/thumbnails/20.jpg)
Customer data
-2.00
-1.75
-1.50
-1.25
-1.00
-0.75
-0.50
-0.25
0.00
0 5 10 15 20 25 30 35 40 45
%E
rr (R
ef.
Ma
ss
)
Ref. Pressure, bar.g
Kerosene 20degC 3cSt Gas Oil 20degC 8cSt Gas Oil 20degC 8cSt (R)
EPAT 25degC 10bar.g EPAT 25degC 20bar.g EPAT 25degC 40bar.g
Experiences
![Page 21: NEL’s EPAT Facility - tuv-sud.co.uk€¦ · –CTSp and CPSp both used –CTLp and CPLp (CTPL) are not used –API values are not robust for NEL test fluid V BPV CTSp CPSp mmp prove](https://reader031.vdocuments.mx/reader031/viewer/2022011802/5b0dba477f8b9a02508e2fe6/html5/thumbnails/21.jpg)
Customer data
Linear Pressure Effect
Experiences
![Page 22: NEL’s EPAT Facility - tuv-sud.co.uk€¦ · –CTSp and CPSp both used –CTLp and CPLp (CTPL) are not used –API values are not robust for NEL test fluid V BPV CTSp CPSp mmp prove](https://reader031.vdocuments.mx/reader031/viewer/2022011802/5b0dba477f8b9a02508e2fe6/html5/thumbnails/22.jpg)
Calibration procedure
1. Zero device at operating temperature and pressure
2. Calibrate device at operating temperature and pressure ‘as found’
3. Additional pressure compensation calibration at ± 10 bar.g to derive pressure
compensation
4. Can then perform an ‘as left’ calibration if required
Experiences
![Page 23: NEL’s EPAT Facility - tuv-sud.co.uk€¦ · –CTSp and CPSp both used –CTLp and CPLp (CTPL) are not used –API values are not robust for NEL test fluid V BPV CTSp CPSp mmp prove](https://reader031.vdocuments.mx/reader031/viewer/2022011802/5b0dba477f8b9a02508e2fe6/html5/thumbnails/23.jpg)
EPAT Facility
– MeterVue is fully operational
– Excellent feedback with users from Europe, Middle East and South East
Asia in the past year
– Live uncertainties are calculated for the results
– DAQ system uses EPAT UB for all calculations and reporting
– History for revising facility uncertainty
– Pressure stability ± 0.5 bar.g up to 100 bar
– Increase / decrease temperature by 10°C in 15 minutes
– Increase / decrease pressure by 20 bar in 5 minutes.
Experiences
![Page 24: NEL’s EPAT Facility - tuv-sud.co.uk€¦ · –CTSp and CPSp both used –CTLp and CPLp (CTPL) are not used –API values are not robust for NEL test fluid V BPV CTSp CPSp mmp prove](https://reader031.vdocuments.mx/reader031/viewer/2022011802/5b0dba477f8b9a02508e2fe6/html5/thumbnails/24.jpg)
EPAT Facility
– Meters can be zeroed at T&P but as T drops, P drops significantly.
– RTD on board Coriolis flowmeters measures tube temperature as opposed to
fluid temperature (response lag)
– If zeroing at elevated temperature, it is important to allow time for stabilising
– Zero value can change by end of calibration. Possibly due to expansion /
contraction of facility with T&P over time.
Experiences
![Page 25: NEL’s EPAT Facility - tuv-sud.co.uk€¦ · –CTSp and CPSp both used –CTLp and CPLp (CTPL) are not used –API values are not robust for NEL test fluid V BPV CTSp CPSp mmp prove](https://reader031.vdocuments.mx/reader031/viewer/2022011802/5b0dba477f8b9a02508e2fe6/html5/thumbnails/25.jpg)
• DDAT
– Collating all calibrations (both master meters & customer)
– Enable master meter calibration curves to be used
– Use customer data to revise Coriolis ISO 10790
• Direct Proving
– Not UKAS at present but work is ongoing
– Reduced uncertainty
– Quicker calibration
• Online Density
– Installing a CDM that has been characterised at NEL
Future Developments