n line analysis - kinetics · 2 max at choice, threshold, water /analyser failure… operation...
TRANSCRIPT
The cost / benefit ratio is a key
asset for your operations but the
quality / efficiency pair is of the
essence, then CRISTALLITE is your
choice within the CRISTAL range
CRISTALLITE, a new analyser following in
the footsteps of CRISTAL:
• for your basic measurements needs
(single stream, single parameter),
• for reduced capital costs & operating
expenses,
• for many parameters / pollutents in water.
CRISTALLITE Water Quality
www.seres-france.com
[email protected] Cristallite_Brochure_UK_12032014 - Non contractual document
ADVANTAGES & APPLICATIONS Depending on parameters, various analysis
methods have been developed on the
CRISTALLITE, the only limit being the number of
pumps available (max. 6):
COLORIMETRY: • Ammonium, Free or Total Chlorine, Chromium VI
• Total Iron, Hydrazine, Phenol
• Phosphates (Orthophosphates), Silica, Hardness
TITRIMETRY:
• TH, TAC, TA
POTENTIOMETRY (pending):
• Ammonium, Cyanides, Chlorides, Fluorides
ABSORPTION:
• Colour
RANGES & OTHER PARAMETERS : PLEASE CONSULT
Compact, reliable, user-friendly
Automatic on line monitoring
Limited maintenance
Low operating costs
Many parts identical to those used in the
CRISTAL range (pumps, measuring
vessels, …)
Benefits of CRISTALLITE:
� Compact
� Easy to use
� Economical
� Efficient
METHODS & PARAMETERS
ON LINE ANALYSIS
EVOLVING WITH THE TIMES
Raw and drinking water : alert stations
Waste water : treatment stations
Urban and industrial waste
Cooling water, process water
SiO2 colorimetry version - Non contractual
TECHNICAL SPECIFICATIONS
CRISTALLITE Water Quality
SERES environnement Tel : +33 (0)4 4297 3737 360 rue Louis de Broglie Fax : +33 (0)4 4297 3030
La Duranne - BP 20087 Email : [email protected]
13793 AIX EN PROVENCE Cedex 3 - France Internet : www.seres-france.com
CONSTRUCTION & ENVIRONMENT
Dimensions (L x H x P)
Wall cabinet : 460 x 606 x 226 mm Overall dimensions (with sample inlet valve) : 460 x 689 x 226 mm
Weight Analyser (reagents bottles excl.) : 7 kg
Materials
Cabinet : ABS, white Front panel : polycarbonate, transparent Side panels : polycarbonate, violet
Protection IP55
Environment
Installation in safe and sheltered area, away from corrosive atmosphere
Ambient T° 5 to 40°C (depending method)
Relative humidity 10 to 80%
ELECTRICAL UTILITIES
Power supply 110 / 230 VAC 50 Hz / 60 Hz
Consumption Typical 150 VA - Maximum 300 VA
ANALYSIS
Parameters Refer to list overleaf / Consult
Ranges Depending on parameter / Consult
Method
Colorimetry, Titrimetry, Potentiometry, Absorption - Depending on parameter / Consult
Number of stream 1 stream of analysis
Accuracy & Repeatability
1 to 3% of end of range
Cycle duration 15 min on average
CONNECTIONS, ALARMS & COMMUNICATION
User interface
Colour LCD display 4.3‘’, touch screen Windows interface
Data retrieval
Data transfer by USB port
Output signal 1 output 4-20 mA, galvanic insulation
Relays Alarms
2 relays, dry contacts 2 max at choice, threshold, water /analyser failure…
OPERATION
Reagents Depending on parameter and method
Zero Automatic at each measurement cycle (optical)
Calibration
Manual with syringe (Seres supply) :
- at each replacement of reagents
- otherwise, depending on the method
SAMPLE
Preparation Filtration as needed
Pressure Inlet : 0.5 bar to 2 bar + 0.2 b / Free outlet
Vessel volume 8 to 10 ml (Colorimetry, Titrimetry) 25 ml (potentiometry)
Inlet T° 5 to 45°C
Inlet / Outlet Inlet : 1/4’’ BSP female / Outlet : tubing 25 x 28 mm
Dilution Depending on parameter / Consult
Example Measurement principle—Chrome VI
COLORIMETRY VERSION : Cabinet & Principle Drawings (example)
CRISTALLITE benefits from the technological advances of
CRISTAL : measuring vessel and PCB, micropumps, ana-
lysis methods, Window interface, etc.
Analyses requiring a maximum of 6 micropumps have
been adapted into this new simple & compact monitor.
OPAL Water Quality
www.seres-france.com
[email protected] Opal_Brochure_UK_21062013 - Non binding document
Water & Hydrocarbons, a sphere of
expertise at Seres.
The new OPAL infra-red is the best
solution for an early detection of oil
traces in water. OPAL, A NEW GENERATION DETECTOR, • on line, real time,
• IR scattering measurement,
• reagent free,
TO MONITOR SUSPENDED HYDROCARBON
IN WATER.
ADVANTAGES The fields of applications & references of the
OPAL will match those of its forerunner, the
Bwam S751.
All types of water
• condensate water, thermal exchangers,
• industrial water,
• urban & industrial waste water
Wide range of fields
• Onshore : refineries, oil drilling plants,
energy, petrochemical and other
industries, ...
• Offshore : oil platforms, ships, ...
Compact system, fast & efficient
IR light diffusion detection,
automatic, on line
Intuitive, touchscreen user interface
Extended choice of inputs & outputs
Easy, cost-efficient operation
No reagent, no cleaning product
Automatic cleaning of measuring vessel
Engineered solutions
Main benefits of OPAL :
� Simple : installation &
operation made easy
� Economical : low capital
& maintenance costs
� Flexible : choice of
options & customized
versions
APPLICATIONS
HYDROCARBONS
UNDER CONTROL
Option ATEX
SERES environnement Tel : +33 (0)4 4297 3737 360 rue Louis de Broglie Fax : +33 (0)4 4297 3030
La Duranne - BP 20087 Email : [email protected]
13793 AIX EN PROVENCE Cedex 3 - France Internet : www.seres-france.com
OPAL Water Quality
TECHNICAL DATA
CONSTRUCTION & ENVIRONMENT
Dimensions (W x H x D)
Basic model on wall skid : 800 x 1055 x 250 mm Detector : 435 x 290 x 195 mm
Weight Basic model on skid : 30 kg (45 kg with cooler)
Material
Wall skid : SS 304 / Detector : high impact PS Measuring vessel : Delrin & PVC Hydraulic circuit : flexible thermoplastic piping
Protection & Environment
IP65 Installation in safe & sheltered area, away from dust and corrosive atmospheres
Ext. temperature 5 to 55°C
Relative humidity 10 to 90%
ELECTRICAL UTILITIES
Power supply 110 / 240 VAC 50 Hz / 60 Hz
Consumption Typical 150 VA - Maximum 300 VA (OPAL basis)
ANALYSIS
Method & Parameter
IR scattering beam measurement Suspended hydrocarbons
Range 0 -10 up to 0 - 500 ppm, others on request
Streams 1 stream of analysis (multistream on option)
Accuracy & Repeatability
1 to 2 % of end of range (depending on range)
Response time Instantaneous, T90%C < 3 sec.
OPAL Water Quality
CONNECTIVITY & ALARMS
User interface Colour LCD graphic display 4.3‘’, touch screen
Data transfer 1 sealed USB connection for transfer on key
Output signal
7 dry contact output relays, 4 outputs TOR 24 V, 1 digital outputs RS232 1 digital outputs RS485 (JBUS protocol) 1 output 4-20 mA
Input signal 3 inputs TOR
Alarms Hi/Lo thresholds, flow failure, analyzer failure, ...
OPERATION
Calibration Quick calibration, optical checking device
Zero On clean, fresh water
Vessel cleaning Automatic with recurrently actuated wiper jack
Clean water 200 l/h for 10 min, once a month, 0.5 to 3 bar
SAMPLE
Preparation Sample conditionning pump & filtration if needed
Interferent Turbidity (auto-compensated on option)
Suspended Max. 100 ppm (above with hydrocyclone option)
Pressure / Flow Inlet : 0.5 to 3 bar maxi / Outlet = inlet / 200 l/h
Inlet T° 5 to 50°C (above with cooler option - consult)
Inlet / Outlet 1/2’’ BSP female
OPTIONS : Cooler, protective or pressurized enclosure, ...
SYSTEM OVERVIEW - Scope of supply
OPAL BONUS FEATURES :
� GRAPHIC INTERFACE : DISPLAY OF
CONCENTRATION & FLOW, EASY SET-
TING, FUNCTIONAL TESTS, ETC…
� COMPACTNESS : LOWER WEIGHT &
DIMENSIONS.
� COMMUNICATION : WIDER SELECTION
OF INPUTS / OUTPUTS.
� OPTIONS : COOLER, HYDROCYCLON,
MULTISTREAM, ENGINEERING, ...
User interface
TOPAZ Water Quality
www.seres-france.com
[email protected] TOPAZ_Brochure_UK_12032014- Document non contractuel
TOPAZ, the obvious choice for
automatic, on line monitoring of a
large variety of chemical
compounds in all types of water.
TOPAZ (CRISTAL Range) SERES environnement solution :
• innovative concept and functionalities,
• designed to achieve the best quality /
efficiency & cost / benefit ratios.
Several measurement methods are available on
the TOPAZ :
COLORIMETRY
• Ammonium, Free and/or Total Chlorine,
Hydrazine, Morpholine, Phenol, Sulphates
• Colour, Silica, Phosphates (Orthophosphates),
Hardness
• Aluminium, Chromium VI, Copper, Iron, Nickel,
Lead, Zinc
TITRIMETRY
• TH, Alkalinity
POTENTIOMETRY
• Ammonium, Chlorides, Cyanides, Fluorides, ...
Specific, customized methods can be adapted
on TOPAZ for the surveillance of process water &
brines :
• Peracetic acid, VFA, Ca Mg, NH4, etc...
Benefits for the user :
� Reduced running costs
� Strict surveillance of water quality
METHODS & PARAMETERS
Automatic, on line measurement
1 to 6 streams of analysis
Intuitive & efficient user interace
Data storage & communication
Lower reagents’ consumption
Routine maintenance made easy
OTHER PARAMETERS : PLEASE CONSULT
TOPAZ key assets :
� Accuracy
� Reliability
� Flexibility
ON LINE ANALYSIS
Drinking & surface water : alert stations
Waste water : sewage works
Urban & industrial wastes
Boiler, cooling water
Process water & brines
ADVANTAGES & APPLICATIONS
TOPAZ Water Quality
SERES environnement Tel : +33 (0)4 4297 3737 360 rue Louis de Broglie Fax : +33 (0)4 4297 3030
La Duranne - BP 20087 Email : [email protected]
13793 AIX EN PROVENCE Cedex 3 - France Internet : www.seres-france.com
The essential design requisites of TOPAZ were :
� automatic, on line analysis
� easily configurable, modular, intuitive & user friendly, multifunctional
� multi-parameter, multi-stream (options)
TECHNICAL SPECIFICATIONS
CONCEPTION & OPERATION OVERVIEW
CONSTRUCTION & ENVIRONMENT
Dimensions
Weight & Material
Wall cabinet stainless steel 316L: 610 x 825 x 400 mm
(W x H x D) 20 kg approx - Stainless steel 316L
Environment
& Protection
Installation in safe and sheltered area, away from
corrosive atmosphere. IP55.
Ambient T° 5 to 40°C (depend method)
ELECTRICAL UTILITIES
Power supply 110 - 240 VAC 50 / 60 Hz
Consumption Typical 150 VA - Maximum 300 VA
ANALYSIS
Parameters Refer to list on reverse page / Consult
Range Depend on parameter / Consult
Method Continuous, on line measurement
Colorimetry, titrimetry, potentiometry or absorption Selection based on parameter and/or range
Number of streams 1 to 6 on option (above, please consult)
Multiparameter Single or multiparameter analyser (consult)
Durée du cycle 15 min on average
Accuracy
Repeatability
± 1 to 2% end of range (colo, titri, pot.)
± 1 to 2% end of range (colo, titri), ± 3 to 5% (pot.)
Relative humidity 10 to 80%
Reagents Depend on parameter and method
CONNECTIVITY, ALARMS & COMMUNICATION
User interface Colour LCD display, 5.7‘’, 160 x 230 mm, touch-screen
Windows interface
Data storage
and retrieval
Data storage in analyser memory
Transfer via USB port
Input / Output &
Communication
4 - 20 mA, dry contacts—Jbus/Modbus RS232
On option : support converter RS485
Alarms Thresholds per stream (HI-LO), sample & analyser failure
Remote control JBus/ModBus protocole or dry contact: end of cycle
stop,
SAMPLING
Preparation Filtration if needed / Dilution, depending on application
Sample inlet Flow : min 30 l/h - optimum 46 l/h (4 l/h with water saver)
Pressure : 0.1 to 3 bar maximum Temperature : 5 to 45°C
Hydraulic
connections
Sample : Inlet 1/4’’BSP F / Outlet soft tubing D INT 9
Waste : soft tubing D INT 12
Volume of vessel 25 ml for potentiometry, othervise 8 to 10 ml
OPERATION
Zero Automatic at end of each measurement cycle
Semi-automatic calibration
Required upon renewal of reagents
Otherwise : depends on method
Cleaning Mechanical wiper on option, if needed
The various modules and their features :
� User interface : smart & intuitive interface enabling all the analyser controls and status reports
� Measurement : emission & reception directly on the PCB gathering all programs driving the whole measurement process whatever the parameter. Increased accuracy resulting from the association of the measuring and its fiber optics system
� JBus/ModBus module : retrieval data / steering
� Supervision : management of data and JBus/ModBus « slave » protocole, , execution of cycles & measurement PCB control, data storage
Potentiometry
Colorim
etry / Titrim