IMPACT ASSESSMENT OF STABILIZATION DEVICES ON CLABSI Mark Rowe MNSc, RNP, VA-BC®
Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors and Masters in Nursing Science. His initial 8 years in the art of nursing was in the specialty of burns/trauma but for the past 26 years has focused on Vascular Access as a clinical/education specialty. Mark, as Senior Vascular Access Specialist, along with a great team; assist with the education of vascular access by hosting Vascular Access Residency program.
Mark was honored to serve as the 2014-2015 Board President of the Association for Vascular Access (AVA) after serving as Director at Large for 4yrs and past Scientific Meeting Chairperson. Mark has presented local, national, and international presentations related to all aspects of vascular access.
Mark is married and they are the proud adopted parents of the best dogs in the world! (seen is upcoming slide!)
![Page 2: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/2.jpg)
Financial Disclosure • Mark Rowe
• Past-President, Association for Vascular Access (AVA)
• Employer: University of Arkansas for Medical Sciences (UAMS)
• Independent Consult/Speaker:
– Interrad Medical, Inc.
– Ethicon, Inc.
– Becton Dickinson and Company, Inc.
– 3M, Inc.
– I will not discuss off label use and/or investigational use in my presentation
![Page 3: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/3.jpg)
Impact Assessment of Securement Devices on CLABSI
Mark Rowe MNSc, RNP, VA-BC®
Sr. Vascular Access Specialist
University of Arkansas for Medical Sciences
CE Code (Attendee Use Only): _______________________________
![Page 4: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/4.jpg)
Objectives • Learner will:
– Understand the difference between subcutaneous engineered securement device (SESD) and
adhesive engineered securement device (AESD)
– Understand retrospective methodology used to accumulate data
– Understand the risk reduction comparison, as related to securement device selection, as it related to CLABSI
![Page 5: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/5.jpg)
Our Story at University Arkansas Medical Sciences
• 500+ all Private beds
• Only level 1 Trauma center in state
• Only high risk birth center in state
• 7 Institutes on campus
![Page 6: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/6.jpg)
Do We Look Alike? Patient Visit 2017
ED Visits 60,861
Surgery Cases 19,262
Outpatient Visits 485,121
Infusion Visits 44,655 (122.3/day)
Vascular Access 2017
Vascular Access Procedures 2,603
PICCs 1,748
Ultrasound PIVs 668
Chest Procedures 187
![Page 7: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/7.jpg)
Anyone Remember This?
![Page 8: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/8.jpg)
What is your experience with Suture?
![Page 9: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/9.jpg)
Suture • Multiple punctures to tissue creating infection risk
• Variation in technique as broad as inserter base
• Does not prevent movement long term
• Associated with safety issues, skin tears
• Hinders care and maintenance
• May require replacement, and additional punctures
![Page 10: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/10.jpg)
What is your experience with Adhesive Engineered Securement Devices?
![Page 11: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/11.jpg)
Adhesive Engineered Securement Devices
• Evidence suggests frequent migration & dislodgement up to 20%
• May damage skin, Medical Adhesive Related Skin Injury
• Care and maintenance
– Vulnerable to movement & catheter loss
– Must replace with each dressing change
– Inconsistency of care with patient transitions • Are they being replaced?
• Are they available in community care?
– Material costs over time are burdensome • Are the costs covered for the patient after insertion?
![Page 12: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/12.jpg)
What is your experience with Subcutaneous Engineered Securement Devices?
![Page 13: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/13.jpg)
• Learning curve
• Change is not easy (Never Is!)
• Perceived claims vs substantiated truths
– Pain
– Bleeding
– Looks barbaric
– Causes INFECTIONS!?! (BUT DOES IT?)
What is your experience with Subcutaneous Engineered Securement Devices?
![Page 14: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/14.jpg)
Our Team & Impact Assessment
![Page 15: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/15.jpg)
Methods • Retrospective data analysis of 3 years of hospital PICC data
• Analysis of CLABSI was segmented by: – Patient demographics: age, gender, diagnosis
– Placement Arm
– Device Type
– Dwell Time
– Inserter Type
– CLABSI Organisms
– Securement Type
![Page 16: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/16.jpg)
Sources of CLABSI Contaminated
Infusate Fluid or Medication
Extrinsic Sources
Skin Organisms Endogenous Flora Extrinsic Sources –
Pistoning/Migration Invading wound
Contamination of Device Prior to
Insertion Rare post- bundle
Contamination of Catheter Hub
Extrinsic Sources Endogenous Flora Hematogenous Seeding
From distant infection
![Page 17: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/17.jpg)
Departmental PICC Data Bedside Vascular Access
Using SESD
Interventional Group
Using AESD
2015 1827
2016 1795
2017 1688
TOTAL 5310
2015 272
2016 215
2017 203
TOTAL 690
![Page 18: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/18.jpg)
0
2
4
6
8
10
12
14
16
18
2015 2016 2017
Total CLABSI by Gender
Male Female
Mean Age 51 y
Mean Age 51 y
Mean Age 51 y
![Page 19: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/19.jpg)
Oncology 44%
Medicine 24%
Gastronterology 16%
Surgery 11%
Orthopedic 5%
PICC PATIENT DIAGNOSIS DISTRIBUTION
![Page 20: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/20.jpg)
0
2
4
6
8
10
12
14
16
18
2015 2016 2017
CLABSI by Insertions Side Arm
Right Left
![Page 21: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/21.jpg)
0
2
4
6
8
10
12
14
16
18
2015 2016 2017
CLABSI Occurance by Lumens
Single Lumen Double Lumen Triple Lumen (OSH)
![Page 22: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/22.jpg)
0
10
20
30
40
50
60
70
80
2015 2016 2017
Mean Dwell Days for PICC CLABSI by Securement Type
SESD AESD
![Page 23: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/23.jpg)
Candida 59%
Kleb 9%
Entracoc 8%
Ecoli 8%
VRE 8%
Staph Epi 8%
2015 CLABSI ORGANISMS
![Page 24: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/24.jpg)
Candida 22%
Enterab 11%
Entracoc 11%
Ecoli 22%
VRE 34%
2016 CLABSI ORGANISMS
![Page 25: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/25.jpg)
Candida 6%
Kleb 18%
Entracoc 23%
Ecoli 23%
MRSA 6%
Pseudo 6%
Strep 6%
VRE 6%
Acinetobacter 6%
2017 CLABSI ORGANISMS
![Page 26: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/26.jpg)
2015
SESD Infections with SESD AESD Infections with AESD
1.47%
0.44%
![Page 27: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/27.jpg)
2016
SESD Infections with SESD AESD Infections with AESD
1.40%
0.33%
![Page 28: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/28.jpg)
2017
SESD Infections with SESD AESD Infections with AESD
1.97%
0.77%
![Page 29: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/29.jpg)
3 Year Total
SESD Infections with SESD AESD Infections with AESD
1.59%
0.51%
![Page 30: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/30.jpg)
CLABSI per 100 Securement Devices
-0.44
-0.50
-0.49 -0.48
Relative Risk Reduction
![Page 31: MPACT ASSESSMENT OF STABILIZATION DEVICES ON …...Mark Rowe MNSc, RNP, VA-BC® Mark is a graduate of the University of Arkansas for Medical Sciences (UAMS) with both his Bachelors](https://reader036.vdocuments.mx/reader036/viewer/2022070212/6101aeb4c0ac8f5b41086189/html5/thumbnails/31.jpg)
In Conclusion • SESD does not increase chance of CLABSI…all indications
(Relative Risk) is that it decreases!
• SESD has become international recognized as a patient centered securement
• Securement device will continue to evolve… but so must we!
• The risk of having a CLABSI with an SESD is consistently about half as much for a Subcutaneous device as it is with an Adhesive securement device.