Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source for Use in a Conformal Superficial Brachytherapy
Device
Brent Rogers, M.S.
![Page 2: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/2.jpg)
Overview
2
Conformal Superficial Brachytherapy Device
Radioactive Source - 90Y
Monte Carlo Simulation - MCNP
Gafchromic Film Measurements
Conclusions
![Page 3: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/3.jpg)
CONFORMAL SUPERFICIAL BRACHYTHERAPY DEVICE
![Page 4: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/4.jpg)
Conformal Superficial Brachytherapy Device
4
![Page 5: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/5.jpg)
RADIOACTIVE SOURCES90Y
![Page 6: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/6.jpg)
Activated by Neutron Bombardmentβˉ Decay 99.983%Q = 2.280 MeVĒβ = 0.934 MeVRCSDA = 1.129 cmt1/2 = 64.1 hours
90Y Source
6
1 mm
2 mm
![Page 7: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/7.jpg)
MONTE CARLO SIMULATIONMCNP
![Page 8: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/8.jpg)
1
10
100
0 0.2 0.4 0.6 0.8 1
r2·D
ose
Rat
e (n
Gy·
cm2 ·
Bq-
1·h-
1)
r (cm)
MCNP ICRU 56
Point Source in an Infinite Water MediumICRU 56
Dmono Monte Carlo (EGS4, ACCEPT)Dpoly = ∫ Dmono dEβStatistical uncertainty ~2%Systematic uncertainty 3-5%
MCNPβ spectrum ICRU 56100 shells*f8 tally (MeV)
8
Ēdeposited ~90%
MCNP vs. ICRU 56 within Ēdeposited ~90%7.35% max difference
3.16 % average difference
0.1 mm
1.0 cm
![Page 9: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/9.jpg)
GAFCHROMIC FILM MEASUREMENTS
![Page 10: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/10.jpg)
1E+04
2E+04
2E+04
3E+04
3E+04
4E+04
4E+04
5E+04
0 250 500 750 1000 1250
Red
Ch
ann
el P
ixel
Val
ue
(16
bit)
Dose (cGy)
Film MeasurementFit95% Confidence Interval
EBT3 Film Calibration
6 MeV electronsRational fit function
X(D) = a + b/(D-c) where X(D) = red channel scanner response
Non-Linear Least Squares FitR2 = 0.9997 Radj
2 = 0.9996
10
![Page 11: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/11.jpg)
0.0E+005.0E-081.0E-071.5E-072.0E-072.5E-073.0E-073.5E-074.0E-074.5E-07
0 0.1 0.2 0.3 0.4 0.5 0.6
Dos
e R
ate
(cG
y s-
1 Bq-
1)
Depth (cm)
MCNP Measurement
90Y Source Depth Dose MeasurementsActivity: 2.5*107 Bq
0.68 mCi
Dmax = 4.6*10-7 cGy/s/Bq
11
![Page 12: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/12.jpg)
Bare Source vs. Tipped Source at the Surface
Bare SourceSlightly narrower profileFaster dose rate
TippedSlightly wider profileSlower dose rate
![Page 13: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/13.jpg)
00.10.20.30.40.50.60.70.80.9
1
-0.5 -0.3 -0.1 0.1 0.3 0.5
Rel
aive
Dos
e
Position (cm)
Measured: Bare Measured: Tipped
MC: Bare MC: Tipped
Bare Source vs. Tipped Source at the SurfaceBareD=9.8 Gy/min/mCi
T[min]=0.11min∙mCi
Gy ∗Dx [Gy]A [mCi]
TippedD=3.89Gy/min/mCi
T[min]=0.28min∙mCi
Gy ∗Dx [Gy]A [mCi]
D(tipped) =40% D(bare)
BareFWHM ≈ 0.18 cm
TippedFWHM ≈ 0.23 cm
![Page 14: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/14.jpg)
Multiple Sources Applied to the Surface with the CSBT Device
14
6 Sources
D = 5.27 Gy/min/mCi
T = 0.19 Dx/Aper Source
6 Sources
D = 4.66 Gy/min/mCi
T = 0.21 Dx/Aper Source
7 Sources
D = 5.05 Gy/min/mCi
T = 0.20 Dx/Aper Source
8 Sources
D = 5.10 Gy/min/mCi
T = 0.20 Dx/Aper Source
Max dose rate is weakly dependent on the number of sources.
![Page 15: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/15.jpg)
CONCLUSIONS
![Page 16: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/16.jpg)
Conclusion
16
Tips decrease the dose rate by 40%
Bare Source Dose Rate ≈ 10 Gy/min/mCi
6, 7, & 8 Source Dose Rate ≈ 5 Gy/min/mCi
Dose rate is independent of the number of sources for many sources.
![Page 17: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/17.jpg)
References
17
Ferreira, C., at al. (2017). A novel conformal superficial high-dose-rate brachytherapy device for the treatment of nonmelanoma skin cancer and keloids. Brachytherapy, 16(1), 215-222. doi:10.1016/
Alexis Lazarine Reed, A. L. (2007). Medical Physics Calculations with MCNP: A Primer. LA-UR-07-4133
Cross, W. G., at al. (1992). Beta-Ray Dose Distributions From Point Sources In An Infinite Water Medium. Health Physics, 63(2), 160-171. doi:10.1097/00004032-199208000-00003
![Page 18: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/18.jpg)
THANK YOU
![Page 19: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/19.jpg)
MCNP Simulation Benchmarking
19
0.1 mm 1 cm
1
10
100
0 0.2 0.4 0.6 0.8 1
Dos
e Ra
te
(nG
y·cm
2·Bq-
1·h-
1)
Depth (cm)
MCNP ICRU 56
![Page 20: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/20.jpg)
0.1 mm
1.0 cm
0.1 mm 1 cm
![Page 21: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/21.jpg)
21
Rational fit function X(D) = a + b/(D-c) where X(D) = red channel scanner response
Non-Linear Least Squares FitR2 = 0.9997 Radj
2 = 0.9996
1E+04
2E+04
2E+04
3E+04
3E+04
4E+04
4E+04
5E+04
0 500 1000
Red
Cha
nnel
Pix
el V
alue
(16
bit)
Dose (cGy)
Film Measurement Fit
![Page 22: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/22.jpg)
AA
22
0.0E+005.0E-081.0E-071.5E-072.0E-072.5E-073.0E-073.5E-074.0E-074.5E-07
0 0.1 0.2 0.3 0.4 0.5 0.6Dos
e R
ate
(cG
y s-
1 B
q-1
)
Depth (cm)
MCNP Measurement
![Page 23: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/23.jpg)
AA
23
![Page 24: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/24.jpg)
Bare Source vs. Tipped Source at the Surface
0
0.5
1
-0.5 -0.3 -0.1 0.1 0.3 0.5Rel
aive
Dos
e
Length (cm)
Relative Surface Dose Profile: Bare Source vs. Tipped Source
Bare Source Tip and RodMCNP: Bare Source MCNP: Tip and Rod
Film Piece: A2; Max Dose: 481 cGy
Film Piece: A28; Max Dose: 258 cGy
BareSlightly narrower profileFaster dose administration
D = 4.30E-07 cGy/s/Bq5 Gy time: 0.52 mCi*minT(min) = 0.10 (min*mCi)*D(Gy)/A(mCi)
TippedSlightly wider profileSlower dose administration
D = 1.82E-07 cGy/s/Bq5 Gy time: 1.24 mCi*min
Bare Source Placed on Surface
Source, Tip, & Rod Placed on Surface
![Page 25: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/25.jpg)
Max dose rate is weakly dependent of the number of sourcesRelatively large distance between sourcesRelatively close proximity to surfaceOne source does not contribute greatly to the area beneath neighboring sources
Multiple Sources Applied to the Surface with the CSBT Device
25
6 SourcesDose Rate: 8.78 cGy/s/mCiTime5Gy: 0.95 mCi*min
6 SourcesDose Rate: 7.76 cGy/s/mCiTime5Gy: 1.07 mCi*min
7 SourcesDose Rate: 8.42 cGy/s/mCiTime5Gy: 0.99 mCi*min
8 SourcesDose Rate: 8.50 cGy/s/mCiTime5Gy: 0.98 mCi*min
![Page 26: Monte Carlo Simulation and Validation of a Novel Yttrium-90 Source …chapter.aapm.org/nccaapm/z_meetings/2019-04-26/04_Agenda... · 2019-04-26 · Bare Source vs. Tipped Source](https://reader034.vdocuments.mx/reader034/viewer/2022050404/5f81548d4de0fa4dc206b520/html5/thumbnails/26.jpg)
Bare Source vs. Tipped Source at the SurfaceBareD=9.8 Gy/min/mCi
T[min]=0.11min∙mCi
Gy ∗Dx [Gy]A [mCi]
TippedD=3.89Gy/min/mCi
T[min]=0.28min∙mCi
Gy ∗Dx [Gy]A [mCi]
00.10.20.30.40.50.60.70.80.9
1
-0.5 -0.3 -0.1 0.1 0.3 0.5
Rel
aive
Dos
e
Position (cm)
Measured: Bare Measured: Tipped
MC: Bare MC: TippedD(tipped) =40% D(bare)
FWHM ≈ 0.18 cm FWHM ≈ 0.23 cm