module 10: maintenance

40
Module 10: Maintenance Global Laboratory Initiative – Xpert MTB/RIF Training Package Slides adapted from Cepheid

Upload: kailani-james

Post on 03-Jan-2016

82 views

Category:

Documents


1 download

DESCRIPTION

Module 10: Maintenance. Slides adapted from Cepheid. Global Laboratory Initiative – Xpert MTB/RIF Training Package. Contents of this module. Discussion of the GeneXpert warranty Overview of maintenance Discussion of maintenance tasks: daily, weekly, monthly and annual - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Module  10:  Maintenance

Module 10: Maintenance

Global Laboratory Initiative – Xpert MTB/RIF Training Package

Slides adapted from Cepheid

Page 2: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-2-

Contents of this moduleContents of this module

Discussion of the GeneXpert warranty

Overview of maintenance

Discussion of maintenance tasks: daily, weekly, monthly and annual

Updating an assay definition file (ADF)

Page 3: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-3-

Learning objectivesLearning objectives

At the end of this module, you will be able to:

Describe current Cepheid warranty and service agreements

Explain requirements for GeneXpert maintenance including periodic tasks, calibration of modules and updating the software with a new Assay Definition File

Page 4: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-4-

Each instrument procured after 24 April 2012 has an initial 2-year warranty:◦ The warranty starts on the date of initial installation – that is,

the date on the installation qualification report sent to Cepheid

◦ If the installation qualification report is not sent, the warranty starts from the date of shipment.

If the machine is not calibrated after 1 year, the second year of the warranty is not valid.

GeneXpert warranty (1 of 3)GeneXpert warranty (1 of 3)

Page 5: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-5-

The warranty covers repairs to the GeneXpert instrument and the replacement of any parts.

The initial warranty is provided free of charge except for annual calibration costs:◦ A calibration kit must be procured separately.

As of 2014, an extended warranty may be purchased for US$ 2,900 per year (or for a discounted price as part of a 3-year package) to cover the machine after the initial 2-year warranty had ended:◦ A calibration kit is included in the price of the extended

warranty.

GeneXpert warranty (2 of 3)GeneXpert warranty (2 of 3)

Page 6: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-6-

GeneXpert warranty (3 of 3)GeneXpert warranty (3 of 3)

Qualification

(send immediately to Cepheid)

(send the report immediately to Cepheid)

Page 7: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-7-

If Cepheid has appointed a local authorized service provider (ASP), the provider will offer certain services free of charge; these include:◦ Exchanging or swapping modules after calibration if required; however, the price of

the new modules is not included unless they are under warranty◦ Repairing parts of the GeneXpert instrument; however, the price of parts is not

included unless they are under warranty◦ Placing orders for GeneXpert cartridges◦ On-site installation and 1-day training at the time of installation.

However, travel costs for the ASP (such as flights and accommodation) are NOT included.

Additional training can be provided for a fee.

Authorized service providersAuthorized service providers

Page 8: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-8-

Maintenance overviewMaintenance overview

Frequency Task

Daily • Remove and properly dispose of cartridges• Clean and disinfect work area• Ensure 10 cm clearance around instrument• Put on dust cover when instrument not in use

Weekly • Disinfect the interior of the cartridge bay• Restart the GeneXpert instrument and computer

Monthly • Disinfect plungers• Disinfect the instrument’s surfaces• Clean the instrument’s filter (this applies only to

newer model GeneXpert instruments, which have a white cover)

• Archive and back-up test results

Annually or after 2,000 tests per module

• Calibrate the module

Page 9: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-9-

Materials required for maintenanceMaterials required for maintenance

Sodium hypochlorite solution (bleach with 0.72% active chlorine; use within 1 day after preparation)

70% ethanol or isopropyl alcohol solution (or equivalent) Wipes, tissues, or cotton or PET (polyethylene terephthalate) swabs Disposable gloves Clean water and soap (for washing the filters) Replacement filters for the fan (available from Cepheid)

Page 10: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-10-

1. Dampen the wipe, tissue or swab with the sodium hypochlorite solution.

2. Wipe the surface or element to be cleaned with the wipe, tissue or swab.

3. Discard the used wipe, tissue or swab.

4. Wait 10 minutes.

5. Dampen a new wipe, tissue or swab with the ethanol or isopropyl alcohol solution.

6. Wipe the surface or element to be cleaned with the wipe, tissue or swab.

7. Discard the used wipe, tissue or swab.

8. Repeat steps 5 through 7.

Disinfection proceduresfor the work area, cartridge bay, plunger and instrument surfaces

Disinfection proceduresfor the work area, cartridge bay, plunger and instrument surfaces

Page 11: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-11-

Daily tasksDaily tasks

After testing, remove the cartridges from the instrument . Dispose of cartridges in the appropriate biohazardous waste container. Remove clutter from the work area. Disinfect the work area. Ensure that there is 10 cm clearance around all sides of the instrument. Put on the dust cover when the instrument is not in use, in order to

reduce the amount of dust that may enter the system

Page 12: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-12-

Weekly task: Disinfect the interior of the cartridge bay

Weekly task: Disinfect the interior of the cartridge bay

1. Open the module door of the GeneXpert instrument.2. Clean the surfaces inside the module’s cartridge bay following the

instructions on slide 10.

CAUTION: Do not touch the slit in the back of the cartridge bay into which the cartridge-reaction tube is inserted.

3. Close the module’s door.4. Repeat steps 1 through 3 for each module .

Page 13: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-13-

Weekly task: Restart the instrument and computer

Weekly task: Restart the instrument and computer

Note: This task is necessary only if the instrument and computer are not routinely turned off at the end of each day.

1. On the computer, close the GeneXpert DX software.2. Turn off the GeneXpert instrument.3. Turn off the computer.4. Turn on the GeneXpert instrument.5. Turn on the computer.6. Launch the GeneXpert DX software.

Page 14: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-14-

Monthly task: Disinfect plungers (1 of 2)

1. On the Maintenance menu, select “Plunger Maintenance”.

2. In the “Plunger Maintenance” window, select a module to clean or select “Clean All”.

3. Follow the instructions in the dialog box.

4. Click “OK”.

Page 15: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-15-

Monthly task: Disinfect plungers (2 of 2)

5. The plunger rod in the chosen module will be automatically lowered.

Plunger

Cartridge bay interior

Module door open

6. After cleaning the plungers, click on “Move Up All” and the plungers will return to their original position.

7. Click “Close” .

Page 16: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-16-

Monthly task: Clean the instrument’s filter (1 of 2)

1. Unscrew the 4 screws on the rear grey panel of the GeneXpert instrument.

Page 17: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-17-

Monthly Task: Clean the instrument’s filter (2 of 2)

2. Remove the grey cover from the back of the instrument.

3. Remove the filter (sponge).

4. Wash the filter with soap and water .

5. Allow the filter to dry completely, then replace it in the instrument.

NOTE: Replacement filters are available from Cepheid if necessary. It is important to use only filters provided by Cepheid or the warranty may be compromised.

Page 18: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-18-

Monthly task: Archive and save results

Archiving tests creates copies of test data in .gxx files (see also Module 7).

Archiving allows you to: Back-up your data to ensure it will not be lost if the computer breaks

down Create a copy of the data to send to Cepheid for assistance in

troubleshooting problems. Files should be archived and saved to a CD or other appropriate external

media (preferably an external drive) at least once a month to ensure that no test data are lost.

Page 19: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-19-

Difference between Archive/Backup:how to archive, retrieve, backup and restore data from a backup

Page 20: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-20-

1. Click “Data Management”

2. Click on “Archive test”

3. Choose the tests that need to be archived(or “Select All”)

CAUTION !!! Leave “Delete Archived Tests” UNCHECKED

How to archive results

Page 21: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-21-

5. In the nest dialog box click“Proceed”

6. The files will be saved in thefolder “Export”.In the file name you will seethe date of archiving

7. Click on “SAVE”

8. Click “OK”

It is highly advised to save (write) your archived data on a RW-CD or external drive

How to archive results (contd.)

Page 22: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-22-

1. Click “Data Management”andClick “Retrieve Test”

2. Select the file you want toretrieve

3. Click on ”Open”

How to retrieve results

Page 23: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-23-

4. Select the test you may want toretrieve (or Select All)

5. Click “OK”

6. Click on “Proceed”

7. Click “OK”

How to retrieve results (contd.)

Page 24: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-24-

At the end of the working day when closingthe GeneXpert software, you will have thefollowing screen:

1. Click “YES”

2. Click on “Database Backup”

3. Click on “Proceed”

How to back up data

Page 25: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-25-

The software will create a zip file with allthe results.

The file is saved on the desktop, in theGeneXpert folder -> Backup section

4. Click on “SAVE”

5. Click “OK”

The zip file cannot be unzipped

How to back up data

Page 26: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-26-

In case of computer failure you might have to restore your data afterreinstallation of the system.

1. Click “YES”

2. Click on “Database Restore”

3. Click on “Proceed”

By closing the GeneXpert software, you will have the following screen:

How to restore data from a back up

Page 27: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-27-

A warning message tells you thatcurrent database will be lost

Click « Proceed » to continue

Select the database to be restored andClick « Open » to continue

How to restore data from a back up (contd.)

Page 28: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-28-

A dialog box asks you if you want to create a backup of the current database.Click « Proceed » to create the backup Click « Cancel » to continue Backup Restore

How to restore data from a back up (contd. 2)

Page 29: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-29-

Annual task: Module calibration (1 of 3)

Calibration should take place each year or after every 2,000 runs on each instrument module (whichever comes first)◦ The date is calculated from the initial installation date (or based on the

previous calibration date) , which is the date on the installation qualification report sent to Cepheid.

NOTE:• Order 1 calibration kit for each 4-module GeneXpert instrument.• In 2014, the cost was US$ 450 plus shipping . If an extended warranty is

purchased, a calibration kit is included. A calibration kit is not included in the initial warranty.

• Shipping costs are free if calibration kits are ordered with cartridges.• The shelf-life of a calibration kit is 6-12 months (confirm the expiry date when

you place order).

Page 30: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-30-

Annual task: Module calibration (2 of 3)

Page 31: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-31-

Annual calibration: Module calibration (3 of 3)

Usually, this remote calibration process will be sufficient to readjust the modules for another year of operation.

However, it is possible that the calibration of one or more of the modules may fail.

If calibration fails, then the modules must be exchanged or swapped for new or refurbished modules from Cepheid, and the old modules must be sent to Cepheid.

Page 32: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-32-

Updating an assay definition file (ADF) (1 of 4)

1. Insert the blue CD, which will be supplied with the cartridges, into the CD drive.

2. In the GeneXpert DX software, click “Define Assays”.

NOTE: The ADF is supplied on the CD that comes with the cartridge kit.

3. Click on “Import”.

Occasionally Cepheid updates its software, and this requires an updated ADF to be imported.Occasionally Cepheid updates its software, and this requires an updated ADF to be imported.

Page 33: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-33-

4. The “Import Assay” window will open.

5. Find the CD on the appropriate drive.

Updating an assay definition file (2 of 4)

Page 34: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-34-

6. Open the “GeneXpert Systems” folder.

7. Click on the file “Xpert MTB-RIF.gxa”.

8. Click on “Import”.

NOTE: The name of this file will vary depending on the version of the software. In any case, choose the “.gxa” file, of which there will be only one on the CD.

Updating an assay definition file (3 of 4)

Page 35: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-35-

The assay is now available for use. It will be listed as shown.

Updating an assay definition file (4 of 4)

Page 36: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-36-

Cepheid assistance and support

Before asking for support, have the following items ready:– GeneXpert serial number, error code (or codes), description of the incident and

archived run (if available).

Cepheid HBDC Europe Training and Assistance (Monday to Friday, 8 am - 18 pm Central European time)

– Phone: +33.5.63.82.53.60 (53.78) (53.86)

Cepheid Technical Support (for issues with the instrument and error codes)– Europe (Monday to Friday, 8 am - 18 pm Central European time)

• Phone: +33.5.63.82.53.19• E-mail : [email protected]

– United States (Monday to Friday, 6 am - 21 pm Pacific time)• Phone: +1.888.838.3222; select option 2• E-mail: [email protected]

Page 37: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-37-

Log for documenting maintenance tasks

Microsoft Excel 97-2003 Worksheet

Example of a maintenance log:

To be customized by each country

Page 38: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-38-

Summary

The warranty starts from the initial date of installation or from the date of shipment if the installation qualification report is not sent to Cepheid within 1 year of the shipment date.

If the machine is not calibrated after 1 year, the second year of the warranty is not valid. An extended warranty can be purchased annually (or as a 3-year package) after the end of

the initial 2-year warranty; a calibration kit is included with the extended warranty. Local authorized service providers (or ASPs) can provide specific services free of charge,

including additional training; however, travel costs are NOT included. A number of maintenance procedures must be performed regularly (daily, weekly, monthly

and annually, or after 2,000 tests per module) in order to ensure that the GeneXpert system functions properly.

Cepheid occasionally updates its software; when this happens, the GeneXpert’s assay definition file (ADF) must be updated.

Page 39: Module  10:  Maintenance

Global Laboratory InitiativeXpert MTB/RIF Training Package-39-

Assessment

Why is it important to promptly send the installation qualification report as well as the calibration report to Cepheid?

When should these two reports be sent?

What is included in the warranty and what is excluded?

List the daily, weekly, monthly and annual maintenance tasks, and specify those that are performed using the software.

Describe the disinfection process and list the parts of the instrument that need to be disinfected as well as the disinfecting solutions that should be used for maintenance.

Page 40: Module  10:  Maintenance

Acknowledgements

The Xpert MTB/RIF Training Package has been developed by a consortium of GLI partners, including FIND, KNCV, US CDC, USAID, TB CARE I and WHO, with funding from USAID.

The modules are based on materials originally developed by FIND, KNCV and Cepheid.